UNI-T UT8802E Benchtop Digital Multimeter

UNI-T UT8802E Benchtop Digital Multimeter

Symbol Warning: When the measured voltage is greater than 600V, The instrument cannot be used for measurements in CAT II, CAT Ill and CAT IV environment.

Overview

UT8802E is a manual range, benchtop digital multimeter that features 19999 display counts, large LCD screen with backlight, full scale overload protection and a unique design. This instrument can be used to measure AC and DC voltage, AC and DC current, resistance, frequency, capacitance, transistor, hFE , diode (LED), SCR, continuity, etc.
This manual contains relevant safety and warning information. Please read the contents carefully and strictly follow all warnings and cautions.

Inspection of open box

Open the packing box and take out the instrument. Please check if the following items are deficient or
damaged. Please contact your supplier immediately if any item is deficient or damaged.
User Manual (CD-ROMt—————–1 pc
Test leads ——————–1 set
Alligator clip——————-1 set
Power cable (AC 220 v—————1 pc
Software application CD————— 1 pc
USB interface wire —————————————————————- 1 pc

Safety rules

This instrument strictly follows the EN 61010-1 : 2010, EN 61326: 2013, RoHS, pollution grade II safety standard, CAT II 600V.

Symbol Unplug the power cord when the instrument is not in use.

Symbol A Note: In the case that the instrument is not used in accordance with the operation instructions, the protection provided by the instrument may be weakened or lost.

CLEANING

Be sure meter is turned off and wipe with a clean ,dry lint-free cloth.
Do not use abrasive cleaners or solvents

Power cord specification :

NameDescriptionRatingApproval NO.
CORDH05VVF 3X0.75mm2300/500V116006
PLUGXR-T00216A250-40036455
CONNECTORXR-W00210A250-40040244
  1. Before using the instrument, please check if there is any item which is damaged or behaving abnormally.
    If any abnormal item is found (such as: test lead bared, housing case damaged, LCD broken, etc.), please stop using the instrument. It is strictly prohibited to use an instrument without shell cover.
    Otherwise, there is a danger of electric shock.
  2. If the test lead has been damaged, it must be replaced with the one of the same type or the same electrical specification.
  3. When measuring, do not touch exposed wires, connectors, unused inputs, or the circuits being measured.
  4. When measuring the voltage higher than 60 V de or 36 Vrms, remember not to exceed the finger baffle plate position on the test lead in order to prevent electric shock.
    Before each use, verify operation by testing a known working circuit that is within the rating of this unit.
  5. If the range of the voltage to be measured is unknown, the maximum range should be selected and then gradually decreased.
  6. Never input voltage and current which exceeds the rated range indicated on the instrument housing.
  7. Before switching the function knob to select the test range, make sure to disconnect the test probes with the circuit to be tested. It is strictly prohibited to turn the function knob during the measurement in order to avoid damages to the instrument.
  8. Do not use or store the instrument in high temperature, high humidity, flammable, explosive and strong magnetic field environments.
  9. Do not change the internal circuit of the instrument in order to avoid the damage to the instrument and user.
  10. After the measurements are finished, the power supply should be turned off. If the instrument is not used for a long time, please unplug the power cord.

Comprehensive index

  1. The maximum voltage between input and COM terminal is DC 1 000V or AC 750V
  2. μA, mA input terminal protection: (CE) 400mA, 1 000V fuse, Cl>6.3x32mm
  3. 10A input terminal protection: (CE) F1 (12A, H, 1000V) fast melt fuse Cl>6.3x32mm
  4. 19999 display counts, update rate 2 -3 times per second.
  5. Manual range
  6. Polarity display: Auto
  7. Over range symbol : OL
  8. Operating temperature: 0~ 40°C (32°F~104 °F)
  9. Storage temperature: -10~ 50°C (14 °F~122°F)
  10. Relative humidity: 0°C~30°C =::;75%RH , 30°C~40°C =::;50%RH
  11. Electromagnetic compatibility:
    In the field with less than 1 V/m radio frequency, the total accuracy= designated accuracy+ range of 5%, In the field with more than 1 V/m radio frequency, the accuracy is not specified.
  12. Power Supply: AC 100V/120V/127V/220V/230VAC/240V, 450-440Hz, 28VA max Protection fuse being used: For AC 1 00V/120V/127V,AC 250V T 250mA
    For AC 220V/230V/240V, AC 250V T 125mA μA mA FUSE: 400mA/1 000V
  13. Outer dimension: (320 x 265 x 110) mm
  14. Weight: about 3100g (accessories excluded)
  15. Safety standards: IEC 61010: CAT 11600V
  16. CAT II: It is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
  17. Temperature coefficient: 0.1 X (specified accuracy) /°C ( < 18°C or> 28°C)

Symbols on LCD screen

Symbols on LCD screen

  1. HOLD: Hold mode
    2. MAX: Max value test
    3. MIN: Min value test
    4. USB: USB communication
    5. Function Icon Relative value test
    6.Function IconHigh voltage test
    7. Function Icon Negative voltage
    8. Function Icon:AC voltage test
    9. DC: DC voltage test
    10.Function Icon Diode and SCR Polar
    11. : SCR/duty cycle test
    12.Function Icon Continuity test
    Function Icon Diode test
    Function Icon Transistor hFE test
    13. Digital reading
    14. Measurement unit

Function buttons & terminals introduction

1.Terminals:

RangeInput terminalFunction
Function IconVCOMDC voltage test
Function IconVCOMAC voltage test
Function IconVCOMResistance test
Function IconVCOMContinuity test
Hz%VCOMFrequency, Duty cycle test
FVCOMCapacitance test
Function IconµAmA COM
A COM
 

DC current test

Function IconµAmA COM
A COM
 

AC current test

Function IconVCOM
socket adapter (UT-S03 A)
Diode (LED) test
hFEsocket adapter (UT-S03 A)Transistor amplification test
SCRsocket adapter (UT-S03 A)SCR test

Function buttons:

Function buttons:
Function buttons:

  1. Power On/Off button
  2. LCD display screen
  3. 20 A current input socket
  4. μA and mA current input socket
  5. COM terminal
  6. Input terminal for V, 0 , Diode, Cap, Frequency
  7. Function buttons
    HOLD: Data holding button
    SELECT : Function selection button
    MAX/MIN: MAX/MIN value t button
    REL: Relative value measurement button
    Function Icon: Backlight button
    USB: USB communication button
  8. Function switch
  9. Grounding terminal
  10. Fuse socket
  11. USB interface
  12. Switch to select AC voltage
  13. Socket
Symbol on meter
Function IconPower on
Function IconPower off
Function IconDirect current
Function IconAlternating current
Function IconGround Terminal
SymbolCaution, possibility of electric shock
SymbolWarning or caution, To ensure safe operation and service of this meter, follow all warnings and instructions detailed in this manual.

Symbol

USB port
SymbolDo not place equipment and its accessories in the trash. Items must be properly disposed of in accordance with local regulations.
SymbolComply with European Union Directive
SymbolConforms to UL STD. 61010-1, 61010-030, Certified to CSA STD. C22.2 No. 61010-1, 61010-030.

CATII

It is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.

Measurement operation instructions

Measurement of DC voltage ( see Figure 1)
Measurement operation instructions

  1. Insert the black test lead into the COM socket, the red test lead into the V socket. UNl-“‘I:
  2. Switch the function knob to ”V’ position. And then connect test leads with the power supply (for measuring open circuit voltage) or the load (for measuring load voltage drop), the polarity will be displayed on the screen.

Symbol Note:

  • Do not input any voltage higher than 1 OOOV. Otherwise there is a risk that the instrument might get damaged. When measuring high voltage, pay attention to avoid electric shock.
  • To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Measurement of AC voltage ( see Figure 2)

Measurement of AC voltage

  1. Insert the black test lead into the COM socket, the red test lead into the V socket.
  2. Switch the function knob to “V” position. Connect test leads with the power supply (for measuring open circuit voltage) or the load (for measuring load voltage drop).

Symbol Note:

• Do not input any voltage that is higher than 750V. Otherwise there is a risk that the instrument might get damaged. When measuring high voltage, pay attention to avoid electric shocked.
• To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Measurement of AC /DC current (see Figure 3)
Measurement of AC /DC current

  1. Insert the black test lead into the COM socket, the red test lead into the “μA” “mA” or “A” socket.
  2. Switch the function knob to “A-” or “A~” position, then connect the instrument in series with the circuit to be tested.

Symbol Note:

  • Before measuring current, the power supply of the circuit should be switched off, and discharge all capacitors.
  • If the range of the current to be measured is unknown, the maximum range should be chosen, and gradually decreased.
  • If the current to be tested is greater than 1 0A, the measurement time should be less than 30 seconds and the wait time of performing the next test should be more than 15 minutes.
  • Disconnect probes and remove from input end after all measurements are completed.

Measurement of Resistance (see Figure 4)
Measurement of Resistance

  1. Insert the black test lead into the COM socket, the red test lead into the O socket.
  2. Switch the function knob to O position, then connect test leads with the resistance to be measured .

Symbol Note:

  • Before measuring current, the power supply of the circuit should be switched off, and get the residual charge stored in the high voltage capacitor fully discharged.
  • In low resistance measurement, the test leads will bring about 0.10 to 0.20 resistance measurement error. In order to obtain accurate readings, the relative measurement function can be used. Short circuit the test leads and presst. button to enter the REV test mode.
  • If the measurement in shorting the test leads is more than 0.50, please check the test leads if they are behaving abnormally
  • When measuring the resistance above 1 M ohm, it may take a few seconds to make the readings steady.
    This is the normal phenomenon for measuring high resistance. In order to obtain the steady data quickly, it is recommended to use the short test wire to measure high resistance.
  • Do not input the voltage higher than AC 30Vrms or DC 60V. Otherwise there is a risk that the instrument might get damaged.
  • To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Measurement of capacitance (see Figure 5)
Measurement of capacitance

  1. Insert the black test lead into the COM socket, the red test lead into the Icon socket.
  2. witch the function knob to “F” position, and then connect test leads with the capacitor to be measured.

Symbol Note:

  • If the measurement value is out of range (too small or too large), the “OL” symbol will be displayed on the screen.
  • If the capacitance to be tested is too small , the REL measuring mode should be used in order to avoid the influence coming from distributed capacitance so as to to get the correct reading.
  • If the capacitance to be tested is greater than 600 μF, in order to get the correct reading, it will take a long time to finish the measurement.
  • Before the measurement, make sure that the residual charges stored in the high voltage capacitor are fully discharged in order to avoid the risk that the instrument might get damaged.
  • Do not input the voltage higher than AC 30 Vrms or DC 60V. Otherwise there is a risk that the instrument might get damaged.
  • To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Measurement of Frequency (see Figure 6)
Measurement of Continuity

  1. Insert the black test lead into the COM socket, the red test lead into the “Hz” socket.
  2. Switch the function knob to “Hz” position, and then connect test leads with signal source to be tested.

Symbol Note:

  • Do not input the voltage higher than AC 36 Vrms. Otherwise there is a risk that the instrument might get damaged.
  •  To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Measurement of Continuity (see Figure 7)
Measurement of Continuity

  1. Insert the black test lead into the COM socket, the red test lead into the “O” socket.
  2. Switch the function knob to “Function Icon” position, and then connect test leads with the circuit to be tested.
  3. if the resistance to be tested is less than 500, the buzzer goes off.
  4. if the resistance to be tested is more than 1000, the buzzer doesn’t go off.

Measurement of Diode (see Figure 8, Figure 9)

Method one:

  1. Insert the black test lead into the COM socket, the red test lead into the ” Function Icon ” socket.
  2. Switch the function knob to”Function Icon” position, and then connect test leads with the Diode to be tested.
    When theFunction Icon symbol is displayed on screen, where the red test lead is connecting is the positive, where the black test lead is connecting is the negative. When theIcon symbol is displayed on screen, where the red test lead is connecting is the negative, where the black test lead is connecting is the positive.

Method two:

  1. Insert adaptor UT-S03A into where it is required to be loaded on the instrument (Figure 9).
  2. Insert the Diode to be test into the adaptor UT-S03A
    When the -M- symbol is displayed on screen, the right of the socket is the positive. The left of the socket is the negative.
    When the +I- symbol is displayed on screen, the right of the socket is the negative. The left of the socket is the positive.
    Note:
    If the Diode to be tested is NG, symbol “OL” or “0.000” will be displayed on screen. Before measuring, the power supply for the circuit must be turned off and the residual charge stored in the capacitors should be fully discharged.
    The OCV to be used to test diode is about ±9V.
    Do not input the voltage higher than AC 36 Vrms, DC 48V. Otherwise there is a risk that the instrument might get damaged.
    To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.
    Measurement of Diode

Measurement of Transistor (see Figure 10)

  1. nsert adaptor UT-S03A into where it is required to be loaded on the instrument.
  2. Switch the function knob to “SCR” position.
  3. Insert transistor to be tested into adaptor UT-S03A according with the polarity indicated on the adaptor.
    Measurement of Transistor

Symbol Note:

  • Before measuring, the power supply for the circuit must be turned off and the residual charge stored in the capacitors should be fully discharged.
  • Do not input the voltage higher than AC 36 Vrms, DC 48V. Otherwise there is a risk that the instrument might get damaged.
  • To disconnect probe and measured circuit then remove probe from input end after finishing all measurement operations.

Technical specification

Error limit : ±(% reading + digit), one year guarantee period
Ambient temperature: 18-28 °C
Ambient humidity: no more than 75%RH

DC voltage
FunctionRangeResolutionAccuracy
±(% reading + digit)
DCV200mV10µV±(0.1%+5)
2V100µV±(0.1%+3)
20V1mV
200V10mV
1000V0.1V±(0.2%+5)
  • Input impedance: approximately 10M ohm.
  • Max input voltage: 1000v
AC voltage
FunctionRangeResolutionAccuracy
±(% reading + digit)
DCV2V100µV 

±(0.5%+20)

20V1mV
200V10mV
750V0.1V±(0.8%+40)
  • Input impedance: approximately 10 M ohm
  • Max input voltage : 750Vrms
  • Frequency response : 40Hz~1 KHz
  • Display: sine wave RMS (average response)
  • There would be some residual readings displayed on LCD screen while without input, but this does not affect the measurement accuracy.
DC current
FunctionRangeResolutionAccuracy
±(% reading + digit)
DCA200µA10nA±(0.5%+20)
2mA100nA
20mA1µA
200mA10µA
20A1mA±(1.5%+40)

If the current to be tested is greater than 1 0A.

  • the measuring time must be less than 30 seconds
  • the interval time of must be over 15 minutes.
AC current
 

Function

Range 

Resolution

Accuracy
±(% reading + digit)
ACA
Frequency response:
40~ 400Hz
2mA0.1µA±(0.8%+40)
20mA1µA
200mA10µA
20A1mA±(2.0%+40)
  • Frequency response 45Hz~400Hz
  • If the current to be tested is greater than 1 0A , the measuring time must be less than 30 seconds
  • the interval time of must be over 15 minutes .
Resistance
FunctionRangeResolutionAccuracy
±(% reading + digit)
Ω20000.010±(0.5%+10)
2kO0.10 

±(0.5%+10)

20kO10
200kO100
2MO1000
200MO1kOFor reference
  • If the resistance to be tested is greater than 20M , the measured result is only for the reference .
Capacitance
FunctionRangeResolutionAccuracy
±(% reading + digit)
F20nF1pF±(2.5%+10)
200nF10pF 

 

±(1.5%+10)

2µF100pF
20µF1nF
200µF10nF
2mF100nF
20mF1µF±(10%+10)
100mF10µFFor reference

If the capacitance to be tested is greater than 20F. the measured result is only for the reference .

Frequency / Duty cycle
FunctionRangeResolutionAccuracy
±(% reading + digit)
 

 

Hz

200Hz0.01Hz 

 

±(1%+5)

2kHz0.1Hz
20kHz1Hz
200kHz10Hz
2MHz100Hz
10MHz1kHz
%10Hz~10kHz

5%~ 99%

0_1 %±(1.5%+2)
  • < 100 kHz: 100 m Vrms < Amplitude~ 20Vrms
  • 100 kHz~1 MHz : 200 m Vrms< Amplitude< 20Vrms
  • 1 MHz5~ MHz : 500 m Vrms< Amplitude< 20Vrms
  • 5 MHz~10 MHz : 900 m Vrms< Amplitude< 20Vrms
Diode I Triode I SCR / Continuity
FunctionRangeResolutionAccuracy
±(% reading + digit)
Diode9.0V1mV10%
SCR9.0V1mV10%
Triode hFE20001Not specified
Continuity10000.10Not specified
  • If the measured resistance is greater than 1000, the circuit is regarded as open status.
  • The buzzer will not go off. if the measured resistance is less than 500, the circuit is regarded as in good conduction status, the buzzer will go off.
  • SCR is the abbreviation of “Silicon Controlled Rectifier”

Power supply setting and fuse replacement (see Figure 12)

Power supply settings
  1. Turn the red switch to the corresponding position
  2. Setting steps:
    a. Unplug the power cord
    b. Turn the red switch to corresponding position
    c. Selectable positions are shown below
    PositionVoltageDemonstrationDescription
    1100V Input corresponding voltage
    2120V/127V 
    3220V/230V 
    4240V 

    Figure 12
    Power supply settings:

Fuse replacement
  1. Unplug the test leads from the instrument.
  2. Turn off the power supply for the instrument
  3. Open the fuse housing with a screwdriver.
  4. Replaced the fuse with new one.

UNI-TREND TECHNOLOGY (CHINA) CC.1 LTD.
No6, Gong Ye Bei 1st Road,
Songshan Lake National High-Tech Industrial
Development Zone, Dongguan City,
Guangdong Province, China
Tel: (86-769) 8572 3888
http://www.uni-trend.com