truflo TIP-Series Insertion Paddle Wheel Flow Meter Sensor Instruction Manual

TIP-Series Insertion Paddle Wheel Flow Meter Sensor

Product Information

Specifications:

  • Operating Range: 0.1 to 10 m/s
  • Pipe Size Range: DN15 to DN600
  • Linearity: Provided
  • Repeatability: Provided
  • Wetted Materials: PVC (Dark), PP (Pigmented),
    PVDF (Natural), 316SS, FKM, EPDM, FFKM, Zirconium Ceramic
    (ZrO2)
  • Electrical: Frequency – 49 Hz per m/s nominal,
    15 Hz per ft/s nominal, Supply Voltage – Provided, Supply Current –
    Provided

Product Usage Instructions:

Safety Information:

Before using the unit, ensure you de-pressurize and vent the
system. Confirm chemical compatibility and do not exceed maximum
temperature or pressure specifications. Always wear safety goggles
during installation. Do not alter the product construction.

Installation:

Hand tighten the unit to avoid damaging product threads. Do not
use tools during installation to prevent any damages.

Personal Protective Equipment (PPE):

Take caution with pressurized systems and ensure to vent the
system before installation or removal to prevent equipment damage
or injury.

Product Description:

  • High Impact NEMA 4X Enclosure
  • Vivid LED Display for Flow and Total
  • Flow Rate and Total Display
  • Pulse and RS485 Outputs (Optional)
  • M12 Quick Connection with True Union Design
  • Zirconium Ceramic Rotor and Bushings for increased wear
    resistance

FAQ:

Q: What should I do if the unit is under pressure?

A: Vent the system prior to installation or
removal to avoid equipment damage or injury.

Q: Can I use tools during installation?

A: It is advised not to use tools as it may
damage the product beyond repair and void the warranty.

Q: How do I avoid damaging the product threads?

A: Hand tighten the unit to prevent
over-tightening that may lead to thread damage.

“`

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor
Quick Start Manual

Read the user’s manual carefully before starting to use the unit. Producer reserves the right to implement changes without prior notice.

24-0500 © Icon Process Controls Ltd.

1

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Safety Information
De-pressurize and vent system prior to installation or removal Confirm chemical compatibility before use DO NOT exceed maximum temperature or pressure specifications ALWAYS wear safety goggles or face-shield during installation and/or service DO NOT alter product construction

Warning | Caution | Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death.

Hand Tighten Only
Over tightening may permanently damage product threads and lead to failure of the retaining nut.

Note | Technical Notes
Highlights additional information or detailed procedure.

Do Not Use Tools
Use of tool(s) may damage produced beyond repair and potentially void product warranty.

WARNING

Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during installation and service of Truflo® products.
Pressurized System Warning
Sensor may be under pressure. Take caution to vent system prior to installation or removal. Failure to do so may result in equipment damage and/or serious injury.

24-0500 © Icon Process Controls Ltd.

2

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Product Description
The TI Series insertion plastic paddle wheel flow meter has been engineered to provide long-term accurate flow measurement in tough industrial applications. The paddle wheel assembly consists of a engineered Tefzel® paddle and micro-polished zirconium ceramic rotor pin and bushings. High performance Tefzel® and Zirconium materials have been selected due to their excellent chemical and wear resistant properties.

*
Rotates 330° *Optional
High Impact NEMA 4X Enclosure

TIP Thermal Plastic
Vivid LED Display
(Flow & Total)

Features

? ½” ­ 24″ Line Sizes

? Flow Rate | Total

? Pulse | RS485 Outputs (Optional)

TI3P 316 SS

New ShearPro® Design
? Contoured Flow Profile ? Reduced Turbulence = Increased Longevity ? 78% Less Drag than Old Flat Paddle Design*
*Ref: NASA “Shape Effects on Drag”

Tefzel® Paddle Wheel ? Superior Chemical And Wear Resistance vs PVDF

M12 Quick Connection
True Union Design
vs. Flat Paddle

Zirconium Ceramic Rotor | Bushings
? Up to 15x the Wear Resistance ? Integral Rotor Bushings Reduce Wear
and Fatigue Stress

360º Shielded Rotor Design
? Eliminates Finger Spread ? No Lost Paddles

TIP Thermal Plastic

TI3P 316 SS

vs. Competitor

24-0500 © Icon Process Controls Ltd.

3

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Technical Specifications

General

Operating Range Pipe Size Range Linearity Repeatability

0.3 to 33 ft/s ½ to 24″ ±0.5% of F.S @ 25°C | 77°F ±0.5% of F.S @ 25°C | 77°F

0.1 to 10 m/s DN15 to DN600

Wetted Materials

Sensor Body O-Rings

PVC (Dark) | PP (Pigmented) | PVDF (Natural) | 316SS FKM | EPDM* | FFKM*

Rotor Pin | Bushings

Zirconium Ceramic | ZrO2

Paddle | Rotor

ETFE Tefzel®

Electrical

Frequency

49 Hz per m/s nominal

15 Hz per ft/s nominal

Supply Voltage

10-30 VDC ±10% regulated

Supply Current

<1.5 mA @ 3.3 to 6 VDC

<20 mA @ 6 to 24 VDC

Max. Temperature/Pressure Rating ­ Standard and Integral Sensor | Non-Shock

PVC

180 Psi @ 68°F | 40 Psi @ 140°F

12.5 Bar @ 20°C | 2.7 Bar @ 60°F

PP

180 Psi @ 68°F | 40 Psi @ 190°F

12.5 Bar @ 20°C | 2.7 Bar @ 88°F

PVDF

200 Psi @ 68°F | 40 Psi @ 240°F

14 Bar @ 20°C | 2.7 Bar @ 115°F

316SS

Consult Factory

Operating Temperature

PVC

32°F to 140°F

0°C to 60°C

PP PVDF

-4°F to 190°F -40°F to 240°F

-20°C to 88°C -40°C to 115°C

316SS

-40°F to 300°F

-40°C to 148°C

Output

Pulse | RS485*

Display

LED | Flow Rate + Flow Totalizer

Standards and Approvals

CE | FCC | RoHS Compliant

See Temperature and Pressure Graphs for more information

* Optional

Model Selection

PVC | PP | PVDF

Size ½” – 4″ 6″ – 24″ 1″ – 4″ 6″ – 24″ 1″ – 4″ 6″ – 24″

Part Number TIP-P-S TIP-P-L TIP-PP-S TIP-PP-L TIP-PF-S TIP-PF-L

Material PVC PVC PP PP PVDF PVDF

Add Suffix `E’ – EPDM Seals

`R’ – RS485 Communication Output

316 SS

Size ½” – 4″ 6″ – 24″

Part Number TI3P-SS-S TI3P-SS-L

Material 316 SS 316 SS

Add Suffix `E’ – EPDM Seals

`R’ – RS485 Communication Output

24-0500 © Icon Process Controls Ltd.

4

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor
Display Characteristics
LED Display

Flow Total

Selected Unit
See Pg.10 for details
M12 Connection
Dimensions (mm)

Flow Rate
Unit | Output Indicators

91.7

91.7

106.4 210.0
179.0

24-0500 © Icon Process Controls Ltd.

5

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor
Wiring Diagram

1 7

8

2 3

6

4

5

Terminal 1 2 3 4 5 6

M12 Female Cable
Description 10~30 VDC Pulse Output
– VDC Pulse Output
RS485A RS485B

Brown | 10~30VDC Black | Pulse Output (OP2) White | Pulse Output (OP1) Grey | RS485B Blue | -VDC Yellow | RS485A
Color Brown White
Blue Black Yellow Grey

Wiring – SSR* (Totalizer)
Set “Con n” in Pulse Output Control (Refer Pulse Control Programmming, Page 11)

Wire Color Brown White Blue

Description + 10~30VDC Pulse Output
-VDC * SSR – Solid State Relay

Wiring – One Pulse/Gal | Con E

Set “Con E” in Pulse Output Control (Refer Pulse Control Programmming, Page 11)

Wire Color Brown White Blue

Description + 10~30VDC Pulse Output
-VDC

Wiring – SSR* (Flow Rate)
Set any “Con” in Pulse Output Control (Refer Pulse Control Programmming, Page 11)

Wire Color Brown Black Blue

Description + 10~30VDC Pulse Output
-VDC * SSR – Solid State Relay

Wiring – To Flow Display | Con F
Set “Con F” in Pulse Output Control (Refer Pulse Control Programmming, Page 11)

Wire Color Brown White Blue

Description + 10~30VDC Paddle Pulse
-VDC

24-0500 © Icon Process Controls Ltd.

6

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Installation

Retention Cap

Very Important

Lubricate O-rings with a viscous lubricant, compatible with the materials of construction.

Using an alternating | twisting motion, carefully lower the sensor into the fitting. | Do Not Force | Fig-3

Ensure tab | notch are parallel to flow direction | Fig-4
Hand tighten the sensor cap. DO NOT use any tools on the sensor cap or the cap threads or fitting threads may be damaged. | Fig-5

Lubricate with silicone inside of insertion fitting

Fig – 1

Fig – 2

Retention Cap
Flow Process Pipe

Fig – 3

Locating Pin
Ensure O-rings are well lubricated Notch
1¼” G
Sensor Blade Ensure tab is parallel to flow direction

Fig – 4 Top View

Correct Sensor Position

0011

Tab

Notch

VERY IMPORTANT Lubricate O-rings with a viscous 02 lubricant, compatible with the system 03

Fig – 5

Notch

Hand tighten using the retention cap

DO NOT use display to tighten

Locate the flow meter positioning tab and clamp saddle notch.
24-0500 © Icon Process Controls Ltd.

Engage one thread of the sensor cap, then turn the sensor until the alignment tab is seated in the fitting notch. Ensure tab is parallel to flow direction.

· Hand tighten the screw cap · DO NOT use any tools — threads may
be damaged · Ensure meter is firmly in place
7

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Correct Sensor Position Setup
TI Series flow meters measure liquid media only. There should be no air bubbles and the pipe must always remain full. To ensure accurate flow measurement, the placement of the flow meters needs to adhere to specific parameters. This requires a straight run pipe with a minimum number of pipe diameters distance upstream and downstream of the flow sensor.

Flange
Inlet

Outlet

2x 90º Elbow

Inlet

Outlet

Reducer

Inlet

Outlet

10xID

5xID

25xID

5xID

15xID

5xID

90º Downward Flow

90º Elbow Downward Flow Upward

Inlet

Outlet

Inlet

Outlet

Ball Valve

Inlet

Outlet

40xID

5xID

Installation Positions
Figure – 1

20xID

5xID

Figure – 2

50xID

5xID

Figure – 3

Good if NO SEDIMENT present

Good if NO AIR BUBBLES present

*Maximum % of solids: 10% with particle size not exceeding 0.5mm cross section or length

Preferred installation if SEDIMENT* or AIR BUBBLES
may be present

24-0500 © Icon Process Controls Ltd.

8

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Fittings and K-Factor
TEE FITTINGS

CLAMP-ON SADDLES

CPVC SOCKET WELD-ON ADAPTERS

Tee Fitting

IN

DN

½” (V1) 15

½” (V2) 15

¾”

20

1″

25

1½”

40

2″

50

2½”

65

3″

80

4″

100

K-Factor

LPM
156.1 267.6 160.0 108.0 37.0 21.6 14.4
9.3 5.2

GPM
593.0 1013.0 604.0 408.0 140.0
81.7 54.4 35.0 19.8

Sensor Length
S S S S S L L L L

TEE FITTINGS (V2)

Size
½” ¾” 1″ 1½” 2″

K-Factor
282.0 196.0 136.0 43.2 23.2

Pressure vs. Temperature

Bar Psi 15.2 220

= PVC

= PP

13.8 200

12.4 180

11.0 160

9.7 140

8.3 120

6.9 100

5.5 80

4.1 60

2.8 40

1.4 20

00

°F 60

104

140

175

°C 20

40

60

80

= PVDF

212

248

100

120

Note: During system design the specifications of all components must be considered. | Non-Shock

Clamp Saddles

K-Factor

IN

DN

LPM GPM

2″

50

21.6

81.7

3″

80

9.3

35.0

4″

100

5.2

19.8

6″

150

2.4

9.2

8″

200

1.4

5.2

Sensor Length
S S S L L

Rotates 330°*

PVC PP PVDF

316SS

* Optional

Weld On Adapter

IN

DN

2″

50

2½”

65

3″

80

4″

100

6″

150

8″

200

10″

250

12″

300

14″

400

16″

500

18″

600

20″

800

24″

1000

K-Factor

LPM
14.4 9.3 9.3 5.2 2.4 1.4 0.91 0.65 0.5 0.4 0.3 0.23 0.16

GPM
54.4 35.5 35.0 19.8 9.2 5.2 3.4 2.5 1.8 1.4 1.1 0.9 0.6

Sensor Length
S S S S L L L L L L L L L

Min/Max Flow Rates

Pipe Size (O.D.)
½” | DN15 ¾” | DN20 1″ | DN25 1 ½” | DN40 2″ | DN50 2 ½” | DN60 3″ | DN80 4″ | DN100 6″ | DN150 8″ | DN200

LPM | GPM LPM | GPM

0.3m/s min. 10m/s max

3.5 | 1.0

120.0 | 32.0

5.0 | 1.5

170.0 | 45.0

9.0 | 2.5

300.0 | 79.0

25.0 | 6.5

850.0 | 225.0

40.0 | 10.5 1350.0 | 357.0

60.0 | 16.0 1850.0 | 357.0

90.0 | 24.0 2800.0 | 739.0

125.0 | 33.0 4350.0 | 1149.0

230.0 | 60.0 7590.0 | 1997.0 315.0 | 82.0 10395.0 | 2735.0

316SS PC

PVC

PP PVDF

24-0500 © Icon Process Controls Ltd.

9

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Programming

STEPS

1

Home Screen

+

3 Sec.

2

Lock

3

Flow Unit

4

K Factor

Select/Save/Continue
DISPLAY

Move Selection Left
OPERATION

Home Screen

Change Digit Value

Lock Settings Factory Default: Lk = 10 Otherwise meter will enter Lockout Mode*
Flow Unit Ut.1 = Gallons (Factory Default) Ut.0 = Liters | Ut.2 = Kiloliters
K Factor Value Enter K Factor value depending on pipe size. Refer to Page 9 for K-Factor Values

Setting Output Limits (SSR*)

Select/Save/Continue

Move Selection Left

STEPS

DISPLAY

1

Home Screen

Home Screen

OPERATION

Change Digit Value
Current Value (CV) Set Value (SV)

2 Flow Rate Pulse Output (OP1) 3 Totalizer Pulse Output (OP2)

Flow Rate Pulse Output (OP1) Limit Enter Flow Rate Pulse Output Value CV SV : Flow Rate Output (OP1) ON CV < SV : Flow Rate Output (OP1) OFF
Refer Page 6 for SSR* Wiring
Totalizer Pulse Output (OP2) Limit Enter Totalizer Pulse Output Value CV SV : Totalizer Output (OP2) ON CV < SV : Totalizer Output (OP2) OFF Note: Refer Pulse Control Programming (Pg 11)
Refer Page 6 for SSR* Wiring
*SSR – Solid State Relay

24-0500 © Icon Process Controls Ltd.

10

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Pulse Control Programming

Select/Save/Continue

Move Selection Left

Change Digit Value

STEPS

DISPLAY

1

Home Screen

3 Sec.

Home Screen

OPERATION

2

Pulse Output Control

3 OP2 Auto Reset Time Delay

4

Alarm Mode Setting

Pulse Output Control Con = n : OP2 Manual Reset (When Totalizer = Set Value (SV)) Con = c | r : OP2 Auto Reset after (t 1) Secs Con = E : One Pulse/Gal (Default) Con = F : Paddle Pulse — Frequency Max 5 KHz (For TVF)
OP2 Auto Reset Time Delay Factory Default: t 1 = 0.50 | Range: 0 ~ 999.99 Secs (Displayed only when Con r | Con c is selected) Note: OP2 = Totalizer Output
Alarm Mode Setting Factory Default: ALT = 0 | Range: 0 ~ 3 Refer to Alarm Mode Selection

5

Hysterisis

Hysterisis Factory Default: HYS = 1.0 | Range: 0 ~ 999.9 (Hysterisis is a buffer around the Programmed Set Point)

6 OP1 Power On Time Delay

OP1 Power On Time Delay Factory Default: t2 = 20 | Range: 0 ~ 9999 Secs Note: OP1 = Flow Rate Output

Alarm Mode Selection

ALt No.

Description

ALt = 0 CV SV — Relay ON | CV < [SV – Hys] — Relay OFF

ALt = 1 CV SV — Relay ON | CV > [SV + Hys] — Relay OFF

ALt = 2 [SV + Hys] CV [SV – Hys] — Relay ON : CV > [SV + Hys] or CV < [SV – HyS] — Relay OFF

ALt = 3 [SV + Hys] CV [SV – Hys] — Relay OFF: CV > [SV + Hys] or CV < [SV – HyS] — Relay ON

Hys = Hysteresis — Acts like a buffer ± around (OP1) pulse output

CV: Current Value (Flow Rate) | SV = Set Value

24-0500 © Icon Process Controls Ltd.

11

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Totalizer Reset

STEPS

1

Home Screen

+

3 Sec.

2

Totalizer Reset

DISPLAY

Home Screen

OPERATION

Totalizer Value will Reset to Zero

Rotor Pin | Paddle Replacement

1
Line up pin with hole

2
GENTLY tap

Small Pin

3
Tap until pin is 50% out

Pin Hole

4
Pull out

5

6

Pull out Paddle

Insert new paddle in flow meter

7
Push in pin approx. 50%

8
GENTLY tap

9
Congratulations! Replacement procedure is complete!

Ensure holes are aligned

24-0500 © Icon Process Controls Ltd.

12

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor

Installation Fittings

SA
Clamp-On Saddle Fittings
· PVC Material · Viton® O-Rings · Available in Metric DIN · Will Accept Signet® Type Flow Meter

Size 2″ 3″ 4″ 6″ 8″

PVC
Part Number SA020 SA030 SA040 SA060 SA080

PT | PPT | PFT
Installation Fittings
· PVC | PP | PVDF · Socket End
Connections · Will Accept Signet® Type
Flow Meter · True-Union Design

PVDF

PVC

Size ½” ¾” 1″ 1½” 2″

Part Number PFT005 PFT007 PFT010 PFT015 PFT020

Part Number PT005 PT007 PT010 PT015 PT020

Add Suffix `E’ – EPDM Seals `T’ – NPT End Connectors `B’ – Butt Fused End Connections for PP or PVDF

PP
Part Number PPT005 PPT007 PPT010 PPT015 PPT020

SAR
Clamp-On Saddle Fittings (SDR Pipe)
· PVC Material · Viton® O-Rings · Available in Metric DIN · Will Accept Signet® Type Flow Meter

Size 2″ 3″ 4″ 6″ 8″ 10″ 12″ 14″ 16″

PVC
Part Number SAR020 SAR030 SAR040 SAR060 SAR080 SAR100 SAR120 SAR140 SAR160

CT
CPVC Tee Installation Fitting
· 1″-4″ Pipe Sizes · Easy to Install · Will Accept Signet®
Flow Meter

CPVC

Size 1″
1 ½” 2″ 3″ 4″

Part Number CT010 CT015 CT020 CT030 CT040

Add Suffix `E’ – EPDM Seals `T’ – NPT End Connectors `B’ – Butt Fused End Connections for PP or PVDF

PG
Glue-On Adapter
· 2″-24″ Pipe Sizes · Easy to Install · Will Accept Signet® Flow Meter

Glue-On Adapter ­ CPVC

Size

Part Number

2″- 4″

PG4

6″- 24″

PG24

24-0500 © Icon Process Controls Ltd.

13

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor
SWOL
Weld-On Adapter
· 2″-12″ Pipe Sizes · 316SS Weld-o-let with PVDF insert · Easy to Install · Will Accept Signet® Flow Meter

Weld-On Adapter – 316 SS

Size 3″ 4″ 6″ 8″ 10″ 12″

Part Number SWOL3 SWOL4 SWOL6 SWOL8 SWOL10 SWOL12

SST
316SS TI3 Series NPT Tee Fittings
· Will Accept Signet® Type Flow Meter

Threaded Tee Fitting – 316 SS

Size

Part Number

½” ¾” 1″ 1 ½” 2″ 3″ 4″

SST005 SST007 SST010 SST015 SST020 SST030 SST040

SSS
316SS TI3 Series Sanitary Tee Fittings
· Will Accept Signet® Type Flow Meter

Sanitary Tee Fitting – 316 SS

Size

Part Number

½” ¾” 1″ 1 ½” 2″ 3″ 4″

SSS005 SSS007 SSS010 SSS015 SSS020 SSS030 SSS040

SSF
316SS TI3 Series Flanged Tee Fittings
· Will Accept Signet® Type Flow Meter

Flanged Tee Fitting – 316 SS

Size

Part Number

½”

SSF005

¾”

SSF007

1″ 1 ½”
2″ 3″ 4″

SSF010 SSF015 SSF020 SSF030 SSF040

24-0500 © Icon Process Controls Ltd.

14

Truflo® — TIP | TI3P Series
Insertion Paddle Wheel Flow Meter Sensor
Warranty, Returns and Limitations
Warranty
Icon Process Controls Ltd warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely and exclusively limited to the repair or replacement, at Icon Process Controls Ltd option, of the products or components, which Icon Process Controls Ltd examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Icon Process Controls Ltd must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the one year from the date of replacement.
Returns
Products cannot be returned to Icon Process Controls Ltd without prior authorization. To return a product that is thought to be defective, go to www.iconprocon.com, and submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Icon Process Controls Ltd must be shipped prepaid and insured. Icon Process Controls Ltd will not be responsible for any products lost or damaged in shipment.
Limitations
This warranty does not apply to products which: 1. are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures
outlined above; 2. have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use; 3. have been modified or altered; 4. anyone other than service personnel authorized by Icon Process Controls Ltd have attempted to repair; 5. have been involved in accidents or natural disasters; or 6. are damaged during return shipment to Icon Process Controls Ltd
Icon Process Controls Ltd reserves the right to unilaterally waive this warranty and dispose of any product returned to Icon Process Controls Ltd where: 1. there is evidence of a potentially hazardous material present with the product; 2. or the product has remained unclaimed at Icon Process Controls Ltd for more than 30 days after Icon Process Controls Ltd
has dutifully requested disposition.
This warranty contains the sole express warranty made by Icon Process Controls Ltd in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL Icon Process Controls Ltd BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF Icon Process Controls Ltd. This warranty will be interpreted pursuant to the laws of the province of Ontario, Canada.
If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty.
For additional product documentation and technical support visit:
www.iconprocon.com | e-mail: sales@iconprocon.com or support@iconprocon.com | Ph: 905.469.9283

by

Phone: 905.469.9283 · Sales: sales@iconprocon.com · Support: support@iconprocon.com

24-0500 © Icon Process Controls Ltd.

15

Documents / Resources

truflo TIP-Series Insertion Paddle Wheel Flow Meter Sensor [pdf] Instruction Manual
TIP-Series, TIP-Series Insertion Paddle Wheel Flow Meter Sensor, Insertion Paddle Wheel Flow Meter Sensor, Paddle Wheel Flow Meter Sensor, Wheel Flow Meter Sensor, Flow Meter Sensor, Meter Sensor

References

Leave a comment

Your email address will not be published. Required fields are marked *