IQC Checking

IQC timeout at 8 AM daily

Roche cobas h 232 System fig 1
Power on the h 232Press Control TestInsert the IQC test strips (Low and High)
Roche cobas h 232 System fig 2
The measurement begins automaticallyThe result is displayed as Pass or FailReturn the test strip quickly back into the container

■ Daily Cleaning (70% alcohol)

Roche cobas h 232 System fig 3
A) Clean the meter surface with moistened tissue B) Remove and wipe the green coverClean the inside area with cotton swab* Be gentle & caution when cleaning the detection window on the side

Routine Testing

Roche cobas h 232 System fig 4
Use heparinized venous whole blood onlyGently swirl the tube for 10 timesPower on the h 232
Roche cobas h 232 System fig 5
Press the “Patient Test” ButtonEnter or scan the patient IDInsert the test strip
Roche cobas h 232 System fig 6
The test strip will warm up automaticallyDraw exactly 150 μL sample using a Roche CARDIAC Pipette
in between the 2 blue arrows
Apply the sample within 5
minutes. Then press the check
button.*

Roche cobas h 232 System fig 7

* Hold the pipette vertically over the sample application area. Slowly apply the sample to avoid spillage and losing of samples.
The Roche CARDIAC Pipette tip should not touch the sample application area of the test strip.

Roche cobas h 232 System QR codehttps://vimeo.com/757443233
Technical Support:
+852 5729 6447 (Max Tsang)
+ 852 5729 3919 (Jeffrey Chan)
Roche Diagnostics (Hong Kong) Limited
Video GuideRoche logo

Documents / Resources

Roche cobas h 232 System [pdf] User Guide
cobas h 232 System, cobas h 232, 232 System, cobas

References

Leave a comment

Your email address will not be published. Required fields are marked *