Answer
Dec 13, 2025 - 02:30 AM

If the cryo probe is not getting cold enough, there could be several reasons for this issue. Here are some troubleshooting steps based on the AtriCure 2797 cryoICE Box User Guide:
1. **Possible Cause:** Pressure gauge less than 700psi(4.8MPa) and cylinder warm.
**Action to Take:** Replace the cylinder with a full one.
2. **Possible Cause:** Pressure gauge less than 700psi (4.8MPa) and cylinder cold.
**Action to Take:**
- Verify that the heater band is working (warm to the touch).
- Check if the heater band icon is off; if so, verify the heater band connection.
3. **Possible Cause:** Pressure gauge is above 700psi (4.8MPa) in freeze mode.
**Action to Take:**
- Unplug the cryoICE BOX Orange tube. If the temperature drops to -65°C(-85°F), the exhaust filter may be clogged, and you should return the cryoICE BOX for service.
- If the temperature doesn't drop to -65°C(-85°F) after unplugging the Orange tube at the cryoICE BOX, the CryoProbe may be clogged, and you may need to replace the probe.
By following these troubleshooting steps, you can address the issue of the cryo probe not getting cold enough during the cryoablation procedure.