Calibrating the using external weights is essential to maintain accuracy. Here is a step-by-step guide based on the provided manual:
1. **Access Calibration Mode:** In normal weighing mode, press and hold the [FUNC/→] button for 3 seconds until the display shows the first parameter 'S1 Un.' Use the [Tare/↑] key to scroll through the menu until you reach 'S8 CAL.'
2. **Initiate Calibration:** Press the [FUNC/→] button to access the calibration parameter. The display will show 'UnLoAd.'
3. **Set Calibration Weight:** Press the key, and the display will show the last calibration weight with the first digit flashing. Use the [FUNC/→] and [Tare/↑] keys to modify the value to match the calibration weight you have. Press the key to accept the calibration mass.
4. **Apply Calibration Weight:** Place the calibration weight on the pan. When the stable indicator is on, press the key. The display will show "PASS," indicating successful calibration. The scale will return to normal weighing mode after 3 seconds.
5. **Verify Calibration:** If the calibration weight is outside the previous calibration value by more than 5%, the display will show "FAIL L" or "FAIL H," indicating that the new calibration weight was too low or too high compared to the previous calibration value.
By following these steps, you can calibrate your Adam ABW 32S Weighing Scale accurately using external weights as per the manufacturer's instructions.