Proper cleaning and maintenance of the Miris Ultrasonic Processor are crucial to ensure its efficient performance. Here is a detailed guide on how to clean and maintain the instrument based on the user manual:
1. **Probe Cleaning:**
- Wipe off the probe after each milk sample using a tissue.
- Clean the probe with a cloth dampened with Miris Cleaner™ after each day of use.
- Polish the tip of the probe with an emery cloth after each use.
- Ensure even polishing of the probe tip, avoiding beveled edges, and completely remove any milk residues.
2. **Instrument Surface Cleaning:**
- Wipe off spillages immediately.
- Clean the instrument surface and inside using a cloth dampened with Miris Cleaner™.
- If necessary, a mild disinfectant can be used to clean the probe and instrument surfaces.
3. **Cleaning Routines:**
- Miris Cleaner™ and emery cloth are essential for proper maintenance of the instrument.
- Miris Cleaner™ is a cleaning agent that should be diluted with distilled or deionized water as per instructions.
- Emery cloth is used for polishing the probe tip.
4. **Required Consumables:**
- **Miris Cleaner™:** A cleaning agent for the probe and surfaces.
- Dilute 50 ml concentrate with 950 ml distilled or deionized water.
- Store the concentrate in a dark, sealed container at room temperature.
- **Emery Cloth:** Fine grit emery cloth for polishing the probe tip.
5. **Cleaning of the Instrument:**
- To maintain homogenization efficiency, clean and polish the probe after each use.
- Remove any erosion on the probe tip by careful polishing with fine grit emery cloth.
- Polish off only the eroded part to prevent quick wearing out of the probe.
By following these cleaning and maintenance procedures as outlined in the manual, you can ensure the optimal performance and longevity of your Miris Ultrasonic Processor.