Answer

Apr 07, 2025 - 09:07 AM
To set the output voltage of the VOLTCRAFT VSP 2403 VSP Laboratory Power Unit, follow these steps:
1. Make sure there are no connected consumers to outputs A and B.
2. Switch on the power unit using the power switch (I=ON / 0=OFF).
3. Press the selection switch "individual" (8) to enable separate adjustment of outputs A and B.
4. Set the current adjustment controls "A" (21 and 18) to the central position.
5. Use the rotary controls "V coarse" and "V fine" to adjust the output voltage for outputs A and B.
- The "V coarse" control is for fast voltage changes.
- The "V fine" control is for precise voltage selection.
6. Verify that the output voltage has been set correctly.
Note: The device operates in constant voltage mode by default, meaning it emits a constant, preset output voltage. This operation is indicated by the "CV" status display (4 or 9) being lit.