PST GSB Electronics plus Sensor User Guide

PST GSB Electronics plus Sensor

PST GSB Electronics plus Sensor

Mounting, housing and cable pairing

The quick start guide is intended for factory calibrated systems, i.e. systems delivered with pre-installed sensor

The GSB is supplied as a PCB. Housing and mounting should be provided, to prevent ESD or short circuit issues.
For systems with cables, it is important to understand that the calibration data of each sensor is stored on the GSB, so it is not possible to connect a different sensor without recalibration. The orientation of the cable connection is only verified via an electronic polarity check. The GSB will stay in contact check mode until a correctly connected sensor is detected. See Figure 1 for correct cable connector orientation. In case of an onboard mounted sensor (TO8 or TO39 housing), no cable sensor must be connected.

Type Label Information

EGa0202std-D0010A300
C809912-003
S: 9110225H03 05/25

  1. Line: System Type e.g. EGa0202std-D0010A300
  2. Line: Batch + Ser. No. e.g. C809912 – 003
  3. Line left: Sensor engraving e.g. 9110225H 03
    ID of the factory-calibrated sensor.
    System will only correctly operate when this sensor is connected!
  4.  Line right: calibration date e.g. 05/25 => week: 05 year: 25

Figure 1 GSB with cabled sensor, orientation!
(System delivered without mounting accessories)
Type Label Information

GSB dimensions in mm
Type Label Information

Electrical interface and O2 signal output

The recommendation is to use a dedicated power supply with 12Vdc/500mA, which is also applied during calibration and offers the optimal performance in regards of accuracy and thermal load of the board. The power supply must ensure protection against hazardous voltages, e.g. SELV, NEC Class 2

Analogue current output 4-20mA (nominal)

Figure 2 connection circuit

Analogue current output 4-20mA (nominal)

Analogue voltage output 0-5V (nominal)

Analogue voltage output 0-5V (nominal)

The calculation of the O2 concertation is based on the analogue output and the O2 full-scale value
Analogue voltage output 0-5V (nominal)

Over range capability of analogue outputs and the O2 measurement: 

The analogue current output displays over-ranges up to 125% full scale, i.e. 24mA
The analogue voltage output displays over-ranges up to 120% full scale, i.e. 6V
Exception is the 96% O2 sensor range, there the outputs are limited to 100% O2, respectively 20mA or 5V.

Behavior after power-up and indication via LED, 4-20mA / 0-5V Output

Figure 3 Power-up states and indication via LED and analogue output
Figure 3 Power-up states and indication via LED and analogue output

Figure 4 GSB main interfaces and DIP-switches
Behaviour after power-up and indication via LED, 4-20mA / 0-5V Output

GSB Parameter Presets / DIP Switch

For factory calibrated systems there is no need to interfere with the DIP-switch setting. The O2 range is defined by the sensor type, the DIP-switches do not change that, they will just select the matching FW-parameters.

Warning: operating the calibration DIP-switch will void/overwrite the factory calibration.

O2 range (min / max) GSB preset DIP switch bit GSB full scale (output) O2-calibration concentration Typical Accuracy Sensor Voltage
1 2 3
10-1000ppm 1 OFF ON ON 1000 vol.-ppm 1000 vol.-ppm +/- 20ppm 0.70 Vdc
0.01-1vol.% 2 ON OFF ON 1 vol.-% 1 vol.% +/- 0.01vol.% 0.75 Vdc
0.01-2vol.% 3 OFF OFF ON 2 vol.-% 2 vol.% +/- 0.2vol.% 0.75 Vdc
0.05-5vol.% 4 ON ON OFF 5 vol.-% 5 vol.% +/- 0.5vol.% 0.80 Vdc
0.1-25vol.% 5 OFF ON OFF 25 vol.-% 20.9 vol.% +/- 0.5vol.% 0.85 Vdc
50-96vol.% 6 ON OFF OFF 100 vol.-% 20.9 vol.% +/- 1vol.% 1.60 Vdc
1-96vol.% 7 OFF OFF OFF 100 vol.-% 20.9 vol.% +/- 1vol.% 1.00 Vdc

Table 1 GSB parameter presets as define by DIP switch setting

Thermal consideration

The GSB, in general, is specified to be operated between 0 and 50°C (ambient). For sensors directly mounted on the GSB, the thermal load on the board is higher, ensure reasonable heat exchange to avoid heat buildup at the board.

Symbol Warning: during operation sensor casings might reach temperatures, which must be rated as “hot surface”. Consider the placing of a warning (sticker) if the sensor is accessible by the end user.

CUSTOMER SUPPORT

ProcessSensing.com
Selection guide for calibrated systems
250217 ega0202std quickstart guide
LogoLogo

Documents / Resources

PDF thumbnailGSB Electronics plus Sensor
User Guide · EGa0202std-D0010A300, C809912-003, GSB Electronics plus Sensor, Electronics plus Sensor, Sensor

References

Ask a Question

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Ask a Question

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.