PST-UM-2010-EN-02 Minox i Oxygen Transmitter Issue No. Description Date Author Initials 01 Original document issued 03/2022 CdF 02…
User Manual PST-UM-2010-EN-02
Safety information
Please read this manual, ensuring that you fully understand the content before attempting to install, maintain or use the Minox i. Important safety information is highlighted throughout this document as follows:
The warning symbol indicates instructions that must be followed to avoid minor, serious or even fatal injury to personnel.
The caution symbol indicates instructions that must be followed to avoid damage to equipment (hardware and/or software) or the occurrence of a system failure.
NOTE highlights an essential operating procedure, condition, or statement.
AC | Alternating current |
DC | Direct current |
oC | Degrees celcius |
oF | Degrees fahrenheit |
EC | Electrochemical |
g | Grams |
GND | Ground |
kg | Kilogram |
LED | Light emitting diode |
LPM | Liters per minute |
mA | Milliampere |
oz | Ounces |
PC | Personal computer |
PCB | Printed circuit board |
PLC | Programmable logic controller |
PPM | Parts per million |
SCFH | Standard cubic feet per hour |
S-S | Solid-state |
SS | Stainless steel |
This page is intentionally blank.
Introduction to Minox i
This manual is applicable to Minox i oxygen transmitter with electrochemical (EC) or solid-state (S-S) sensors; and KF-40 flange, flow-through or 2” tri-clamp connections.
ATEX, IECEx and UKCA certified, the Minox i is an intrinsically safe, 2-wire, loop-powered transmitter, with a 4…20mA output. It is designed with proven sensor technology to accurately measure O₂ in a variety of gases, in the most demanding applications and hazardous environments. Minox i is available with the three process connections shown below.
NOTE: You must provide suitable mating flange and clamp for KF-40 and 2” tri-clamp connections.
Minox i can be used with a Programmable Logic Controller (PLC) or suitable equipment capable of providing a 4…20 mA loop-power interface. For operation in hazardous Ex-rated areas, the interface must be provided via a suitable isolation barrier device.
The Minox i oxygen transmitter must only be operated in the way set out in this user manual.
Always verify electrical interface circuit compatibility before connection.
1.2 Structure
The Minox i oxygen transmitter consists of an oxygen sensor and inbuilt microprocessor. They are encased in stainless steel, with a process connection on the removable housing and an electrical signal connector on the main housing.
The sensor is replaceable and can be accessed via the removable housing, as shown in Figure 2:
1.3.1 Electrochemical % (EC%)
The Minox i EC% oxygen sensor is a galvanic fuel cell capable of superior performance, accuracy, and stability. Its operational life is typically up to 18 months in ambient air. The sensor is designed to be unaffected by the presence of other background gases.
The Minox i is required to be mounted in a vertical orientation with the electrical signal connector at the top. For gases with a composition of hydrocarbons and/or CO2, the solid-state sensor should be used when % level O2 measurements are required.
1.3.2 Electrochemical ppm (EC ppm)
The Minox i can be specified with a choice of EC ppm oxygen sensors. Both are galvanic fuel cells with different chemistry. With both EC ppm sensors, the Minox i must be mounted vertically with the electrical signal connector at the top.
The OC-60 sensor is the ppm version of the EC% and is suitable for inert, H2 and He gases. The OC-69 sensor is compatible with gases that have a hydrocarbon composition or CO2 element.
1.3.3 Solid-state (S-S)
The Minox i S-S sensor is not gas-dependent or specific, and offers a robust measurement capability on process gas streams or in containment environments. It can operate with or without sample flow, making it ideal for in-situ monitoring, and it is not orientation-specific. With the tri-clamp and KF-40 process connections, the Minox i can be mounted on a T-piece of pipeline or on the vertical panel of a containment system without concern. The transmitter provides a linear output with a sensitivity that is constant throughout its operating range.
Typically, the intrinsically safe Minox i oxygen transmitter can be used in the following applications:
Before installing your Minox i oxygen transmitter, ensure that its specifications are suitable for the process or pipeline where it will be installed.
See “2.2.1 KF-40 flange and 2” tri-clamp process connections with dimensions” on page 5 or “2.2.2 Flow-through process connection with dimensions” on page 6 for more guidance.
This product is not approved for use in applications where the sample gas is greater than 21 % volume O.
2.1 Unpack the Minox i
NOTE: The sensor is packed in a sealed package that has been purged to remove O2 from the packed environment. It should not be removed from the package until the transmitter has been installed at the measurement point and you are ready to perform measurements.
Your Minox i oxygen transmitter pack is comprised of the following equipment:
Leaving a sensor in ambient air conditions will reduce sensor life.
2.2 Mount the Minox i
The Minox i transmitter requires a two-wire industry-standard 4…20 mA loop-powered connection, which you must be provide. It must be installed in the measurement point before the sensor is removed from its sealed package.
For flow-through connection, continue to “2.2.2 Flow-through process connection with dimensions” on page 6.
2.2.1 KF-40 flange and 2” tri-clamp process connections with dimensions
A suitable mating flange and seal/centering ringwill be required to install the KF-40 flange or triclamp Minox i. These are not included in the pack.
2.2.2 Flow-through process connection with dimensions
The flow-through Minox i is designed for measurement on an extracted sample gas. The sample gas should be at atmospheric pressure +/- 100 mbar (xxx psig), and a flow-rate of 250…500 ml (0.5…1.0 SCFH) is required.
If the sample gas is at a pressure regulator, a flow-meter must be placed before the transmitter. The transmitter must be able to vent the gas to atmosphere or an atmospheric vent.
If the sample gas is at sub-atmospheric pressure, the sample must be drawn through the flowthrough block. Ntron can supply systems to aid this installation upon request.
2.2.3 Electrical connection
A suitable galvanic isolation barrier is required if the Minox i is located in an Ex area.
Consult the EC-type examination certificate Baseefa19atex0200 for Parameter Matching.
The Minox i oxygen transmitter operates using a 2-wire industry-standard 4…20 mA loop-powered supply which you must provide.
Once the Minox i is connected to your process, it is ready to be connected to your power supply.
To do this:
2.2.4 Install your sensor
The sensor supplied should only be removed from the sealed package when the transmitter is installed and ready to make a measurement.
Installing the sensor is straightforward, but should be done quickly to limit exposure to ambient O2 levels, this is particularly important for the EC ppm sensors. Prolonged exposure to air will shorten the operational life of the sensor.
As a guideline, you should install the sensor into the Minox i, and the mount the transmitter in the measurement environment in 15-30 seconds. To install the sensor, please see “4.2 Replace the sensor” on page 12.
NOTE: Calibration should be carried out by an appropriately trained or suitably qualified person.
The oxygen sensor output will decrease over time, and periodic calibration will be required. This is a non-intrusive procedure, achieved by using the MagTip one-touch calibration tool included with your Minox i.
Calibration cycles for the Minox i depend on the application, the applied gases, and the sensor type. They can be carried out in-situ or after safely removing the oxygen transmitter from your process in a bench test.
NOTE: For optimal calibration, the atmosphere surrounding the sensor must be ambient air (20.9 % volume O2) with no contaminants present. This can be achieved using an oxygen analyzer or by flushing the system with clean, oil-free compressed air.
The Minox i ppm must be calibrated in a controlled environment. Calibrating the Minox i % in ambient air will significantly reduce sensor life.
When the MagTip tool is applied, the indicator LED will flash six times in rapid succession. This confirms calibration is in progress.
During normal operation, the indicator LED will flash once every two seconds. This slower flash confirms Minox i is powered up and operational.
3.1 In-situ
3.1.1 Minox i ppm/%
Please see above before proceeding.
1. Apply the calibration gas (20.9 % volume O2) for a minimum of 15 minutes.
The pressure flow should replicate that applied to the Minox i during normal operation.
This will ensure calibration is accurate, and avoid over-pressurizing the oxygen sensor.
The following method of air calibration should only be performed on a Minox i %.
4.1 General
To maintain performance, the sensor in Minox i will require replacement. Please check the Technical Specifications for frequency. During sensor replacement, it is recommended that light cleaning of the external housing is carried out, as well as the checks outlined in this section. Never use chemical cleaning agents or high pressure water or steam to clean the equipment. Do not submerge in water.
4.2 Replace the sensor
The Minox i oxygen sensor is replaceable. When it reaches the end of serviceable life, calibration can no longer be performed and the sensor must be replaced.
A regular program of calibration will mitigate against sudden sensor failure. It is advisable to establish a program of preventative maintenance to ensure process downtime is kept to a minimum or avoided.
The sensor supplied should only be removed from the sealed package when the transmitter is installed and ready to make a measurement.
Installing the sensor is straightforward, but should be done quickly to limit exposure to ambient O2 levels, this is particularly important for the EC ppm sensors. Prolonged exposure to air will shorten the operational life of the sensor.
As a guideline, you should install the sensor into the Minox i, and the mount the transmitter in the measurement environment in 15-30 seconds.
To replace your sensor:
3. Place your new EC or S-S oxygen sensor into the removable housing, ensuring the sensor’s brass rings are facing out of the removable housing as shown in Figure 10.
4. Screw the removable housing back onto the main housing, by turning it counter-clockwise.
NOTE: The brass slip rings on the sensor must face into the main housing and out of the removable housing as shown above.
4.3 Check your Minox i
The following checks will ensure your Minox i oxygen transmitter is operating as its optimum.
This is to ensure there is no build-up of particulate matter, and that moisture is not being retained at the orifice.
Typically, the Minox i will require removal from the process to carry out checks, cleaning and sen-sor replacement.
4.4 Operational spares
The only replaceable part for Minox i is the sensor. Sensor model details can be found in ” Appendix A – Technical Specifications” on page 14.
The replacement sensor must be the same as the installed sensor, for example, it is not possible to replace an OC-90 EC sensor with an OC-60 EC sensor.
NOTE: The sensors have a shelf life. Please see ” Appendix A – Technical Specifications” on page 14 for details.
Appendix A – Technical Specifications
Sensor
Solid-state | Electrochemical (%) | Electrochemical (ppm) | Electrochemical (ppm) | |
Model | OC-92 | OC-90 | OC-60 | OC-69 |
Operating temperature | -20 °C…+50 °C (-4 °F…+122 °F) | 5 °C…+45 °C (+31 °F…+113 °F) | +5 °C…+45 °C (+31 °F…+113 °F) | -10 °C…+45 °C (+14 °F…+113 °F) |
Calibration interval | Up to 1 year | Up to 6 months | Up to 3 months | |
Operational life (dependent on O2 and moisture concentration) | Up to 2 years | Up to 18 months | Up to 12 months | |
Shelf life (in original packaging) | Up to 1 years | Up to 6 months | ||
Calibration interval | Up to 1 year | Up to 6 months | Up to 3 months | |
Gas-wetted materials | Acetal, Stycast 2651 mmstainless steel, PTFE, FR4 and gold | HDPVDF, ABS, Teflon, FR4 and gold | HDPVDF, stainless steel, Teflon, FR4 and gold |
Performance
Measuring range | 0…25% O2 | 0…10, 0…100, 0…1000 ppmV O2 |
Output resolution | 0.01 % O2 | 1 ppmV O2 |
LDL (sensitivity) | 0.05 % O2 (500 ppm) | 1 ppmV O2 |
Accuracy (intrinsic error) | +/-2 % of reading @ calibrated temperature and pressure | |
Linearity | +/- 2 % of reading | |
Response time (T90) | < 15 seconds @ 25 °C (77 °F) within selected range | |
Flow rate | 250…500 ml/min (0.5…1.0 SCFH) | |
Humidity | 0…98 %rh non-condensing | |
Operating pressure | Nominally atmospheric (+/-100 mBar) |
Electrical input/output
Output signal | 4…20 mA |
Power supply | 24 V DC +/- 10 % |
Maximum power consumption | 50 mA @ 24 V DC |
Electrical interface | M12 connector (1.5 metre cable supplied) |
Mechanical
Housing material | 303 and 316 stainless steel |
Gas-wetted materials excluding sensor | 316 stainless steel |
O-ring material | Nitrile |
Ingress protection | IP66 (NEMA4) |
Weight | 800 g (28 oz) |
Area | Certification details | Standards |
Europe & Japan (Pending) | IECEx / ATEX / UKCA IECEx BAS 19.0013 / Baseefa 19ATEX0020 / BAS22UKEX0052 Ex ia IIC T4 Ga (-20 °C≤Ta≤+55 °C) Ex ia IIC T200 135°C Da (-20 °C≤Ta≤+55 °C) Ui=28 V Ii=93 mA Pi=0.65 W Ci=12nF Li=705uH | EN IEC 60079-0:2018 EN 60079-11:2012 IEC 60079-0 Ed. 7 2017 IEC 60079-11 Ed. 6 2011 |
North America/Canada (Pending) | Class II Division 1 Groups E-G T135°C Class I Zone 0 AEx ia IIC T4 Ga / Ex ia IIC T4 Ga Class 1 Zone 20 AEx ia IIC T135°C Da / Ex ia IIC T135°C Da Ta = -20°C≤Ta≤+55°C UI/Vmax=28 V II/Imax=93 mA Pi=0.65 W Ci=12nF Li=705uH | CSA C22.2 No. 60079-0:19 CSA C22.2 No. 6007911:14 UL 60079-0:2019 UL 60079-11:2014 CSA C22.2 No. 61010-1-12 (Reference only) |
Appendix C- Quality, recycling and warranty information
The PST Oxygen group of companies – AII, Ntron and SST – comply with applicable national and international standards and directives.
Full information can be found on our website at https://www.processsensing.com/en-us/compliance
The compliance site contains information on the following directives:
This page is intentionally blank.
For PST contact information, visit
ProcessSensing.com
This document is the property of Process Sensing Technologies and may not be copied or otherwise reproduced, communicated to third parties, nor stored in any Data Processing System without the express written authorization of Process Sensing Technologies.
©2022 Process Sensing Technologies
PROCESS SENSING PST-UM-2010-EN-02 Minox i Oxygen Transmitter [pdf] User Manual PST-UM-2010-EN-02 Minox i Oxygen Transmitter, PST-UM-2010-EN-02, Minox i Oxygen Transmitter |
MCK-400 Carbon Heater Product Information: Heater MCK-400 The Heater MCK-400 is a portable electric heater…
HRP-300N3 Series 300W Ultra-High Peak Power Supply Product Information The HRP-300N3 series is a 300W…
H881 Trail Camera Product Information and Usage Instructions Product Information Introduction The trail camera is…
15019 PM06 V2-14S Power Module Power Module Product Information The PM06 V2-14S Power Module is…
JV10413 Coffee Table Product Information The product is made by COSTWAY and has a model…