OEM Tech NBU-1508 Thyristor Based Discharge Circuit User Manual

rev 1.07 / 2025 01 22

NBU-1508 thyristor-based discharge circuit 

User manual

NBU-1508 thyristor-based discharge circuit

Warning! This equipment may be dangerous.

Please read the entire user manual carefully before using the product.

Overview / Appearance

The NBU-1508 discharge circuit is designed for simplification of solid-state  laser systems development. The module utilizes complete discharge of a  relatively small capacitor bank (a typical value is 100uF), which, together  with an external inductor, forms short (tens to hundreds of microseconds) bell-shaped pulses on a flashlamp. The NBU-1508 includes a thyristor, its  driver, protective circuits, simmer and trigger circuits and controls. Typical  applications are Q-switched Nd:YAG lasers or similar.

It’s important to emphasize that the NBU-1508 is not a stand-alone solution. It  requires a capacitor charging power supply, an external inductor and capacitor  bank, as well as some minor parts and controls to operate properly.

Overview-Appearance Figure 1

By default the NBU-1508 includes capabilities both for serial and external  triggering (See Triggering section for details). Simplified as well as more  complicated modifications are available on request.

An important feature of the NBU-1508 is clear and visible separation  between low voltage and high voltage parts. The 24V power input and all  interface circuits (SIM and PWF) are separated from the high voltage circuits  by 4mm creepage distances and solid insulation to 2500VAC test voltage (1  MOOP/MOPP).

Overview-Appearance Figure 2

Cooling

Module is cooled with a built-in fan.

No external cooling is required.

Contents of delivery

By default, the delivery package contains the NBU-1508 module only.

The following standardized set of cables can be purchased optionally:

  • Flashlamp connection cable – 2pcs (150cm each)
  • Capacitor connection cable – 2pcs (30cm each)
  • Inductor connection cable – 2pcs (30cm each)
  • External trigger transformer connection cable with mounted 1:70  trigger transformer – 1pc (150cm)
  • 24VDC cable – 1pc (50cm)
  • SIM cable – 1pc (50cm)
  • PWF cable – 1pc (50cm)

A DR cable and a discharge resistor itself can not usually be supplied.  Customized delivery content is available on request.

Connections, signals, signal descriptions

Connections Figure 1

INDUCTOR CONNECTION: Two M5 terminals

L1, L2 – to the inductor (polarity is not important)

FLASHLAMP CONNECTION: Two M5 terminals

LAMP+ – to the flashlamp anode

LAMP– – to the flashlamp cathode

CAPACITOR BANK CONNECTION: Two M5 terminals

C+ – to the capacitor bank positive

C– – to the capacitor bank negative

CAPACITOR CHARGER CONNECTION CONSIDERATIONS

Note: we recommend connecting the capacitor charging power supply  not to the C+/C– terminals, but directly to the capacitor bank.

Connections Figure 2

EXT (TO THE TRIGGER TRANSFORMER): Molex 39-30-1020

By default, the NBU-1508 supports both serial and external methods of  triggering. To use the NBU-1508 for external triggering, a trigger  transformer should be connected to the EXT connector. If the EXT connector is left unconnected, the NBU-1508 automatically recognizes this  and applies pulses of serial triggering to the LAMP– flashlamp connection  (see also Triggering section for details).

Connections Figure 3

DR (TO DISCHARGE RESISTORS): Molex 39-30-1020

Connections Table 2

We suggest using the NBU-1508 with discharge resistors of your choice.  The resistance and the power rating of discharge resistors must be selected  accordingly to your application. Real operating voltages, your capacitor  bank capacitance and the required quick discharge time are the most  important parameters affecting the choice. You are welcome to contact us  for assistance.

24VDC (TO 24V DC POWER SUPPLY): Molex 39-30-1040

Connections Table 3

PWF (PULSE CONTROL): Molex 39-30-1040

Connections Table 4

PWF CIRCUITS:

Connections Figure 4

SIM (SIMMER SUPPLY CONTROL): Molex 39-30-1040

Connections Table 5

SIM CIRCUITS:

Connections Figure 5

LED:

There is the warning LED indicating that NBU-1508 circuits are under high  voltage.

HV LED (red) – is ON when NBU-1508 circuits are under high voltage

GND: M4 THREAD

The module should be grounded using this thread.

Grounding should be carried out before the module is powered up.

Safety

Warning! This equipment produces high voltages that can be very  dangerous. Be careful around the device.

  • During operation all the protective covers of the equipment must be  securely fixed in place and all electrical connections must be properly  attached
  • The NBU-1508 discharge circuit is designed to be installed inside a  properly grounded metal enclosure. It is the user’s responsibility to  ensure that personnel are prevented from accidentally contacting the  NBU-1508, especially DR, L1/L2, C+/C–, LAMP+/LAMP– connectors and cables. Casual contact could be fatal! 
  • After shutdown, do not handle the capacitance load until it has been  discharged. Use an appropriate meter to check for complete discharge.
  • Disconnect the module from the DC power source before making or  changing electrical or mechanical connections.
  • Don’t remove protective covers! There are no user serviceable parts  inside this equipment.

Operations

A typical start-up sequence of the NBU-1508 is given below:

  1. Connect your capacitor charging power supply, +24V DC power supply, the  inductor, the capacitor and the flashlamp to the NBU-1508 module
  2. Set +5V voltage on PIN1 of PWF connector to switch off the capacitor bank  discharging
Warning! By default, the capacitor discharge is switched on.

Be careful and always switch off capacitor discharging before  turning on your capacitor charging module. Otherwise, it may lead  to failure of discharge resistors used in the system.

  1. Disable simmer supply (PIN3 of SIM connector)
  2. Apply +24V DC power to the module
  3. Enable your capacitor charging power supply (charging module).
    As soon as charging module starts operation, capacitor is charged
  1. Enable simmer supply (set +5V on PIN3 of SIM connector)
  2. Wait a few seconds for Simmer Sensor. If it fails, shut down the system
  3. Disable the capacitor charging power supply prior to pulse
Note: this is important to disable the capacitor charging power  supply. Otherwise, the output current of working capacitor charging  power supply won’t allow the thyristor to close after the flash.
  1. Apply +5V TTL pulse to PIN3 of PWF connector. The rising edge of TTL  pulse will initiate the discharge through the flashlamp. The discharge will  stop automatically once the energy stored in the capacitor bank is emitted.  Pulse length of TTL signal should be a little longer than the actual discharge  time, which depends on the L, C and flashlamp parameters
  2. Enable the capacitor charging power supply after the pulse
  3. Repeat #8-#10

To power down NBU-1508 and discharge capacitor bank:

  1. Turn off or Disable your capacitor charging power supply (important!)
  2. Disable simmer supply
  3. Set 0V on PIN1 of PWF connector to switch on capacitor bank discharging.  Wait for the complete discharge of capacitors.
Warning! If the capacitor bank capacitance is enormously high, this might lead to the overheating of the discharge resistors used in  the system.
  1. Remove +24V DC power from the module.

Specifications

ELECTRICAL

+24VDC INPUT:
Voltage+24V DC
Current consumption5A max, up to 10A peak
PULSE PARAMETERS:
Max. voltage1500V (other on request)
Max. discharge current

(depends on flashlamp

impedance K0)

1000A for pulse width < 1 ms

500A for pulse width > 1 ms

(other on request)

Max. average power2000W (other on request)
Pulse widthDefined by L, C and flashlamp parameters
Max. rep. rate50Hz (up to 100Hz on request)
RECOMMENDED WIRES:
For capacitor bank

connections

FLEXI-2V or similar (>1000V rated voltage,  >4mm2cross-section), short length (30cm recommended)
For flashlamp connectionsFor external triggering – FLEXI-2V or  similar (>1000V rated voltage, >4mm2cross section)

For serial triggering additional insulation  (e.g. with silicone tubing) or spacing (e.g.  with spiral bundle hose) of LAMP– wire is  required

SIMMER PARAMETERS:
Simmer current400mA (200-500mA on request)
Max output voltage300V
Max output power100W
Open circuit voltage1500V
FLASHLAMP TRIGGERING PARAMETERS:
Trigger pulse width~1us
Restrike rateA few Hertz (automatically adjusted)
EXTERNAL TRIGGERING:
Pulse energy / trigger

voltage

~20mJ / 350V positive to the EXT connector  (other on request)
Trigger transformerExternal transformer (purchased separately) Recommended p/n ZS1324-24V LUL1(H) by
Excelitas Technologies (Digikey p/n ZS1324- 24VLUL1(H)-ND)
SERIAL TRIGGERING:
Pulse energy / trigger

voltage

~160mJ / 10kV negative to LAMP– (other on  request)
Trigger transformerIntegrated transformer
Cooling Forced air cooling with built-in fan
Safety1 MOOP/MOPP between high voltage and low voltage parts (4mm creepages /

2500VAC HiPot)

Protections From overheating of internal components
ENVIRONMENT:
Operation temperature0 … +40 °C
Storage temperature-20 … +60 °C
Humidity<90%, non-condensing

MECHANICAL

Size (LxWxH) Approx. 217x177x64mm (see also the  dimensional drawing below)
Weight Approx. 2.0 kg (w/o cables)

Dimensional drawing

Dimensional drawing

How to order? / Options

By default, there is the only standard part number, namely NBU-1508.

The following options are available on request:

  • Integrated current sensor
  • High power / current / voltage thyristor with extended parameters
  • Low power / current / voltage thyristor with reduced cost
  • Digital RS-232 / RS-485 interface
  • Modified cables
  • Modified simmer and / or trigger parameters

Triggering (serial triggering, external triggering, triggering basics)

By default the NBU-1508 supports both serial and external methods of  triggering.

To use the NBU-1508 for external triggering, a trigger transformer should be  connected to the EXT connector.

Triggering Figure 1

Default parameters for external triggering are (other on request):

  • 350V voltage (applied to EXT connector)
  • ~20mJ energy
  • Recommended transformer (sold separately) – 1:70 e.g. ZS1324-24V  LUL1(H) by Excelitas Technologies (Digikey p/n ZS1324-24VLUL1(H)- ND)

If EXT connector is left unconnected, the NBU-1508 automatically  recognizes this and applies serial triggering to the flashlamp.

Default parameters for serial triggering are (other on request):

  • 10kV voltage (applied to LAMP–)
  • ~160mJ energy