BARO Module
Specifications
- Measurement Specifications:
- Barometric Pressure Range: +65 kPa to +105 kPa
- Accuracy: TBD
- Temperature Range: TBD
- Accuracy: TBD
- Communication Specifications:
- Output:
- Analog Output: 0 to 2,000 mV (default), 0 to 1,000 mV
(configurable with tensioVIEW) - Digital Output: SDI-12, tensioLINK, Modbus RTU
- Analog Output: 0 to 2,000 mV (default), 0 to 1,000 mV
- Data Logger Compatibility: Compatible with various data
acquisition systems
- Output:
- Physical Specifications:
- Dimensions: 80 mm (L) x 29 mm (W) x 30 mm (H)
- Cable Length: 1.5 m (standard)
- Connector Types: 4-pin and 8-pin M12 plug connector or stripped
and tinned wires
- Compliance: EM ISO/IEC 17050:2010 (CE
Mark)
Product Usage Instructions
Connection Types
Refer to the equivalent circuit diagram in Figure 2 for
connecting the BARO Module to a data logger using the recommended
SDI-12 specification. Follow the wiring diagrams provided in
Figures 3, 5, 6, and 8 for different connector types.
Precautions
Handle the sensor with care to avoid damage. Proper installation
and protection are essential to prevent warranty voiding. Follow
recommended installation instructions and implement safeguards
against interference.
Sensor Communications
The sensor supports SDI-12, tensioLINK, and Modbus over RS-485
two-wire communication protocols. Ensure correct protocol selection
based on application requirements. Contact METER Customer Support
for assistance.
FAQ
Q: What should I do if I need a nonstandard cable length?
A: Contact Customer Support for assistance with nonstandard
cable lengths.
Q: How do I know which communication protocol to use for my
application?
A: Evaluate the advantages and challenges of each protocol based
on your application needs. If unsure, contact METER Customer
Support for guidance.
18577-00 2025
BARO INTEGRATOR GUIDE
SENSOR DESCRIPTION
The BARO Module is a precise barometer to compensate for matric potential measurements of TEROS 31 and TEROS 32 tensiometers. The BARO Module can be used as a standalone sensor to compensate one or more tensiometers at a measuring site, or as a digital/analog converter to compensate a connected TEROS 31 or TEROS 32 value and convert the SDI-12 signal into an analog voltage output (only 8-pin version). The BARO Module and TEROS 32 combination can be used as a T8 tensiometer replacement. For a more detailed description of how this sensor makes measurements, refer to the BARO Module User Manual.
Figure 1 BARO Module
APPLICATIONS
· Barometric pressure measurement · Barometric compensation of matric potential measurements · Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers · Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32
ADVANTAGES
· Digital sensor communicates multiple measurements over a serial interface · Low-input voltage requirements · Low-power design supports battery-operated data loggers · SDI-12, Modbus RTU or tensioLINK serial communications protocol supported · Analog output supported (only 8-pin version)
METER Group, Inc. 2365 NE Hopkins Court, Pullman, WA 99163 T +1.509.332.2756 F +1.509.332.5158 E info@metergroup.com W metergroup.com
BARO MODULE INTEGRATOR GUIDE
SPECIFICATIONS
MEASUREMENT SPECIFICATIONS
Barometric Pressure
Range
+ 65 kPa to +105 kPa
Resolution ± 0.0012 kPa
Accuracy
± 0.05kPa
Temperature
Range
-30 to + 60 °C
Resolution ± 0.01 °C
Accuracy
± 0.5 °C
COMMUNICATION SPECIFICATIONS
Output
Analog Output (8-pin connector only) 0 to 2,000 mV (default) 0 to 1,000 mV (configurable with tensioVIEW)
Digital Output SDI-12 communications protocol tensioLINK communication protocol Modbus RTU communication protocol
Data Logger Compatibility
Analog Output Any data acquisition system capable of switched 3.6- to 28-VDC excitation and single-ended or differential voltage measurement at a greater than or equal to 12-bit resolution.
Digital Output Any data acquisition system capable of 3.6- to 28-VDC excitation and RS-485 Modbus or SDI-12 communication.
PHYSICAL SPECIFICATIONS
Dimensions Length Width Height
80 mm (3.15 in) 29 mm (1.14 in) 30 mm (1.18 in)
Cable Length 1.5 m (standard)
NOTE: Contact Customer Support if a nonstandard cable length is needed.
Connector Types 4-pin and 8-pin M12 plug connector or stripped and tinned wires
COMPLIANCE
EM ISO/IEC 17050:2010 (CE Mark)
2
BARO MODULE INTEGRATOR GUIDE
EQUIVALENT CIRCUIT AND CONNECTION TYPES
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.
GND DATA R2 100K C1 220PF
L1
R1
10UH
510
GND
Figure 2 Equivalent circuit diagram
Figure 3 M12 4-pin output connector
Power
(brown)
Data output +
(white)
Ground
(blue)
Data output
(black)
Excitation
Digital in
Ground
Data Logger
Figure 4 Wiring diagram SDI-12 4-pin connector
3
BARO MODULE INTEGRATOR GUIDE
Power
(brown)
Data
output
(white)
Data
output
(black)
Ground
(blue)
Excitation
Digital in
Digital in
RS-485 A (+) RS-485 B (-)
Ground
Data Logger
Figure 5 Wiring diagram RS-485 4-pin connector
Figure 6 M12 8-pin output connector
Power
(white)
Data output +
(pink)
Ground
(brown)
Data output –
(blue)
Excitation
Digital in SDI-12
Ground
Data Logger
Figure 7 Wiring daigram SDI-12 8-pin connector
4
Power
(white)
BARO MODULE INTEGRATOR GUIDE
Data
output
(pink)
Data
output
(blue)
Ground
(brown)
Excitation
Digital in
Digital in
RS-485 A (+) RS-485 B (-)
Ground
Data Logger
Figure 8 Wiring diagram RS-485 8-pin connector
PRECAUTIONS
METER sensors are built to the highest standards, but misuse, improper protection, or improper installation may damage the sensor and possibly void the warranty. Before integrating sensors into a sensor network, follow the recommended installation instructions and implement safeguards to protect the sensor from damaging interference.
SENSOR COMMUNICATIONS
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensioLINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantages and challenges. Please contact METER Customer Support if the protocol choice for the desired application is not obvious.
SDI-12 INTRODUCTION
SDI-12 is a standards-based protocol for interfacing sensors to data loggers and data acquisition equipment. Multiple sensors with unique addresses can share a common 3-wire bus (power, ground, and data). Two-way communication between the sensor and logger is possible by sharing the data line for transmit and receive as defined by the standard. Sensor measurements are triggered by protocol command. The SDI-12 protocol requires a unique alphanumeric sensor address for each sensor on the bus so that a data logger can send commands to and receive readings from specific sensors.
Download the SDI-12 Specification v1.3 to learn more about the SDI-12 protocol.
RS-485 INTRODUCTION
RS-485 is a robust physical bus connection to connect multiple devices to one bus. It is capable of using very long cable distances under harsh environments. Instead of SDI-12, RS-485 uses two dedicated wires for the data signal. This allows the use of longer cables and is more insensitive to interference from outside sources, since the signal is related to the different wires, and supply currents do not influence the data signal. See Wikipedia for more details on RS-485.
TENSIOLINK RS-485 INTRODUCTION
tensioLINK is a fast, reliable, proprietary serial communications protocol that communicates over the RS-485 interface. This protocol is used to read out data and configure features of the device. METER provides a tensioLINK PC USB converter and software to communicate directly with the sensor, read out data, and update the firmware. Please contact Customer Support for more information about tensioLINK.
MODBUS RTU RS-485 INTRODUCTION
Modbus RTU is a common serial communications protocol used by Programmable Logic Controllers (PLCs) or data loggers to communicate with all kinds of digital devices. The communication works over the physical RS-485 connection. The combination of RS-485 for the physical connection and Modbus as serial communications protocol allows fast and reliable data transfer for a high number of sensors connected to one serial bus wire. Use the following links for more Modbus information: Wikipedia and modbus.org.
5
BARO MODULE INTEGRATOR GUIDE
INTERFACING THE SENSOR TO A COMPUTER
The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
The BARO Module can also be configured and measured via tensioLINK using METER software tensioVIEW, available to download at meter.ly/software. To connect a BARO Module to a computer a tensioLINK USB converter and a suitable adapter cable is necessary.
METER SDI-12 IMPLEMENTATION
If a BARO Module is connected between a TEROS 31 or 32 tensiometer, both the barometric air pressure and the absolute pressure of the TEROS tensiometer can be read out via Modbus. The compensated matrix potential can be read out via Modbus as well.
METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 aR9 and aRC0 aRC9 ). M , R , and C command implementations are found on pages 89.
Out of the factory, all METER sensors start with SDI-12 address 0 .
SENSOR BUS CONSIDERATIONS
SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com for articles and virtual seminars containing more information.
SDI-12 CONFIGURATION
Table 1 lists the SDI-12 communication configuration.
Table 1 SDI-12 communication configuration
Baud Rate Start Bits Data Bits Parity Bits Stop Bits Logic
1,200 1 7 (LSB first) 1 (even) 1 Inverted (active low)
SDI-12 TIMING
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.
START D0 D1 D2 D3
D4 D5
D6 EP STOP
Figure 9 Example SDI-12 transmission of the character 1 ( 0x31 )
6
BARO MODULE INTEGRATOR GUIDE
DATA LOGGER
Break (at least 12 ms)
Command
SENSOR
Response
Marking (at least 8.33 ms)
Marking (at least 8.33 ms)
Sensor must respond within 15 ms
Maximum time*
*Maximum time is dependent upon the amount of data returned for the command sent.
Figure 10 Example data logger and sensor communication
COMMON SDI-12 COMMANDS
This section includes tables of common SDI-12 commands that are often used in an SDI-12 system and the corresponding responses from METER sensors.
IDENTIFICATION COMMAND ( aI! )
The Identification command can be used to obtain a variety of detailed information about the connected sensor. An example of the command and response is shown in Example 1, where the command is in bold and the response follows the command.
Example 1 1I!113METER BARO
Parameter 1I!
Fixed Character
Length 3
Description Data logger command. Request to the sensor for information from sensor address 1 .
Sensor address.
1
1
Prepended on all responses, this indicates which sensor on the bus is returning
the following information.
13
2
Indicates that the target sensor supports SDI-12 Specification v1.3.
METER
8
Vendor identification string. ( METER and three spaces for all METER sensors)
BARO
Sensor model string.
6
This string is specific to the sensor type.
For the BARO, the string is BARO .
Sensor version.
100
3
This number divided by 100 is the METER sensor version
(e.g., 100 is version 1.00).
BARO-00001
13, variable
Sensor serial number. This is a variable length field. It may be omitted for older sensors.
CHANGE ADDRESS COMMAND ( aAB! )
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A Z , and 0 9 ). An example output from a METER sensor is shown in Example 2, where the command is in bold and the response follows the command.
7
BARO MODULE INTEGRATOR GUIDE
Example 2 1A0!0
Parameter 1A0!
0
Fixed Character
Length 4
1
Description Data logger command. Request to the sensor to change its address from 1 to a new address of 0 .
New sensor address. For all subsequent commands, this new address will be used by the target sensor.
ADDRESS QUERY COMMAND ( ?! )
While disconnected from a bus, the Address Query command can be used to determine which sensors are currently being communicated with. Sending this command over a bus will cause a bus contention where all the sensors will respond simultaneously and corrupt the data line. This command is helpful when trying to isolate a failed sensor. Example 3 shows an example of the command and response, where the command is in bold and the response follows the command. The question mark ( ? ) is a wildcard character that can be used in place of the address with any command except the Change Address command.
Example 3 ?!0
Parameter ?!
0
Fixed Character
Length 2
1
Description Data logger command. Request for a response from any sensor listening on the data line.
Sensor address. Returns the sensor address to the currently connected sensor.
COMMAND IMPLEMENTATION
The following tables list the relevant Measurement ( M ), Continuous ( R ), and Concurrent ( C ) commands and subsequent Data ( D ) commands, when necessary.
MEASUREMENT COMMANDS IMPLEMENTATION Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus.
Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
Table 2 aM! command sequence
Command Response
This command reports average, accumulated, or maximum values.
aM!
atttn
aD0!
a±<Press>±<Temp>+<Status>
Comments
When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output.
If the BARO module is used in standalone <Press> returns the current barometric pressure.
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3 document for more information.
CONCURRENT MEASUREMENT COMMANDS IMPLEMENTATION
Concurrent Measurement ( C ) commands are typically used with sensors connected to a bus. C commands for this sensor deviate from the standard C command implementation. First, send the C command, wait the specified amount of time detailed in the C command response, and then use D commands to read its response prior to communicating with another sensor.
8
BARO MODULE INTEGRATOR GUIDE
Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
Table 3 aC! measurement command sequence
Command Response
This command reports instantaneous values.
aC!
atttnn
aD0!
a±<Press>±<Temp>+<Status>
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is processed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt seconds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.
CONTINUOUS MEASUREMENT COMMANDS IMPLEMENTATION Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command.
aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.
Table 4 aR0! measurement command sequence
Command Response
This command reports average, accumulated, or maximum values.
aR0!
a±<Press>±<Temp>+<Status>
NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.
PARAMETERS
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.
Parameter ± a n nn ttt <TAB> <CR> <LF>
<sensorType>
<Checksum> <CRC>
Unit — — — — s — — —
—
— —
Table 5 Parameter Descriptions
Description Positive or negative sign denoting sign of the next value SDI-12 address Number of measurements (fixed width of 1) Number of measurements with leading zero if necessary (fixed width of 2) Maximum time measurement will take (fixed width of 3) Tab character Carriage return character Line feed character ASCII character denoting the sensor type For BARO Module, the character is ; METER serial checksum METER 6-bit CRC
METER MODBUS RTU SERIAL IMPLEMENTATION
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.
9
BARO MODULE INTEGRATOR GUIDE
Table 6 Modbus communication characters
Baud Rate (bps) Start Bits Data Bits Parity Bits Stop Bits Logic
9,600 bps 1 8 (LSB first) 0 (none) 1 Standard (active high)
Figure 11 RTU Message Frame Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character. The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community. For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.
SUPPORTED MODBUS FUNCTIONS
Table 7 Function Definitions
10
BARO MODULE INTEGRATOR GUIDE
Function Code
Action
Description
01
Read coil/port status
Reads the on/off status of discrete output(s) in the ModBusSlave
02
Read input status
Reads the on/off status of discrete input(s) in the ModBusSlave
03
Read holding registers
Reads the binary contents of holding register(s) in the ModBusSlave
04
Read input registers
Reads the binary contents of input register(s) in the ModBusSlave
05
Force single coil/port
Forces a single coil/port in the ModBusSlave to either on or off
06
Write single register
Writes a value into a holding register in the ModBusSlave
15
Force multiple coils/ports Forces multiple coils/ports in the ModBusSlave to either on or off
16
Write multiple registers Writes values into a series of holding registers in the ModBusSlave
DATA REPRESENTATION AND REGISTER TABLES
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.
Please note that some Modbus dataloggers use addressing with a +1 offset. This sometimes causes confusion and is based on a Modbus specification void. If there are problems in implementing your Modbus program on the datalogger always try testing different register offsets and data types. Using a known value, like temperature, where it’s known what value to expect is a good practice to start testing.
Table 8 Modbus Primary Tables
Register Number
Table Type
1xxx
Discrete Output Coils
2xxx
Discrete Input Contacts
3xxx
Analog Input Registers
4xxx
Analog Output Holding Registers
Access Description Read/Write on/off status or setup flags for the sensor
Read
sensor status flags
Read
numerical input variables from the sensor (actual sensor measurements)
Read/Write
numerical output variables for the sensor (parameters, setpoint values, calibrations etc.)
For example, register 3001 is the first analog input register (first data address for the input registers). The numeric value stored here would be a 16-bit unsigned integer-type variable that represents the first sensor measurement parameter (pressure value). The same measurement parameter (pressure value) could be read at register 3201, but this time as a 32-bit floating-point value with a Big-Endian format. If the Modbus Master (Datalogger or a PLC) supports only 32-bit float-values with a Little-Endian format, then one could read the same measurement parameter (same pressure value) at register 3301. The Virtual Sub-Blocks are meant to simplify the user’s effort in programming the Modbus query of the sensors.
Table 9 Modbus Virtual Sub-Blocks
Register Number Access Size
X001-X099
Read/Write 16 bit
X101-X199
Read/Write 16 bit
X201-X299
Read/Write 32 bit
X301-X399
Read/Write 32 bit
Sub-Table Data Type signed integer unsigned integer float Big-Endian format float Little-Endian format
11
BARO MODULE INTEGRATOR GUIDE
REGISTER MAPPING
Table 10 Holding Registers
41000 (41001*) Modbus Slave Address Detailed Description Read or update the sensor’s modbus address
Data Type
Unsigned integer
Allowed Range
1 – 247
Unit
–
Comments
Updated slave address will be stored in the sensor’s nonvolatile memory
32000 (32001*) Detailed Description Data Type Allowed Range Unit Comments
Table 11 BARO Module Input Registers Soil Water Potential Compensated tension value from tensiometer 32 bit floating Big-Endian -200 to +200 kPa Tensiometer needs to be connected as slave
32001 (32002*) Detailed Description Data Type Allowed Range Unit Comments
Soil Temperature High accuracy on board temperature measurement 32 bit floating Big-Endian -30 to +60 degC Tensiometer needs to be connected as slave
32002 (32003*) Detailed Description Data Type Allowed Range Unit Comments
Sensor Supply Voltage On board supply voltage measurement 32 bit floating Big-Endian -10 to +60 Volts –
32003 (32004*) Detailed Description Data Type Allowed Range Unit Comments
BARO Status Binary status 32 bit floating Big-Endian 0/1 –
32004 (32005*) Detailed Description Data Type Allowed Range Unit Comments
BARO Reference Pressure On board high accuracy barometric pressure measurement 32 bit floating Big-Endian +70 to +120 kPa –
12
BARO MODULE INTEGRATOR GUIDE
Table 11 Baro Module Input Registers (continued)
32005 (32006*)
Tensiometer Pressure
Detailed Description Data Type Allowed Range Unit Comments
Absolute pressure value from tensiometer 32 bit floating Big-Endian -200 to +200 kPa Tensiometer needs to be connected as slave
32006 (32007*) Detailed Description Data Type Allowed Range Unit Comments
BARO Temperature On board temperature measurement 32 bit floating Big-Endian -30 to +60 degC –
*Some devices report Modbus register addresses with an offset of +1. This is true for Campbell Scientific Loggers and Dataker loggers. In order to read the desired register use the number in the parenthesis.
EXAMPLE USING A CR6 DATALOGGER AND MODBUS RTU
The Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbuscompatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.
PROGRAMMING A CR6 DATALOGGER
The Programs running on the CR6 (and CR1000) Loggers are written in CRBasic, a language developed by Campbell Scientific. It is a high-level language designed to provide an easy yet extremely flexible and powerful method of instructing the data logger how and when to take measurements, process data, and communicate. Programs can be created using either the ShortCut Software or edited using the CRBasic Editor, both of which are available for downloading as stand-alone applications on the official Campbell Scientific website (www.campbellsci.com).
ShortCut Software (https://www.campbellsci.com/shortcut)
CRBasic Editor (https://www.campbellsci.com/crbasiceditor)
A typical CRBasic program for a Modbus application consists of the following: · Variables and constants declarations (public or private) · Units declarations · Configuration parameters · Data tables declarations · Logger Initializations · Scan (Main Loop) with all the sensors to be quired · Function call to the Data Tables
CR6 LOGGER RS-485 CONNECTION INTERFACE
The universal (U) terminal of the CR6 offers 12 Channels that connect to nearly any sensor type. It gives the CR6 the ability to match more applications and eliminates the use of many external peripherals.
The Modbus CR6 connection shown in Figure 12 uses the RS-485 (A/B) interface mounted on terminals (C1-C2) and (C3-C4). These interfaces can operate in Half-Duplex and Full-Duplex. The serial interface of the BARO Module used for this example is connected to (C1-C2) terminals.
13
BARO MODULE INTEGRATOR GUIDE
BARO Module to CR6 Datalogger Wiring Diagram
CR6
BARO Module
G
RG
C4
C3
C2
RS-485-A white (for 8-pin M12 pink)
C1
RS-485-B black (for 8-pin M12 blue)
G
12V
Power brown (for 8-pin M12 white)
G
Ground blue (for 8-pin M12 brown)
SW2
SW1
G
Figure 12 RS-485 interface
After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports–the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).
14
BARO MODULE INTEGRATOR GUIDE
EXAMPLE PROGRAMS
‘CR6 Datalogger ‘This is an example program for reading out the BARO Module using a CR6 ‘datalogger and the MODBUS RTU protocol over a RS-485 Bus. The measurement values polled ‘from the sensor will be: Water Potential, Temperature and the Sensor’s Supply voltage.
‘This program runs a scan every 1 Min and stores the data in a 1 Min table.
‘Declare Constants
Const BARO_MB_ADDR=1 Const MB_TIMEOUT= 10 Const MB_RETRIES= 1
‘BARO Module Modbus slave address ‘100ms timeout (value * 0.01sec)
‘Declare Public Variables Public PTemp, batt volt Public mb statu ‘variable used for monitoring the modbus poll status
‘Declare Private Variables Dim Measurements (3) ‘array for holding the measurements values read from the sensor
‘Aliases used for the BARO Module
Alias Measurements (1)= Water Potential Alias Measurements (2)= Temperature Alias Measurements (3)= Sensor Supply
‘Declare Units
Units Water Potential=kPa Units Temperature=degC Units Sensor supply=V
‘Define Data Tables.
DataTable (BARO_Table,1,-1) ‘Set table size to # of records, or -1 to auto allocate. DataInterval (0,1,Min,10) ‘Store new measurement every 1 Minute Minimum (1,batt_volt,FP2,False,False) Sample (1,PTemp,FP2) Sample (3,Measurements(),IEEE4)
EndTable
‘Main Program
BeginProg
SW12(2,1) ‘Switch ON the SW12-2 terminal (if used for powering the BARO Module)
SerialOpen(ComC1,9600,3,2,50,4)
‘open communication port, setup for RS-485 ‘(BaudRate, Format, TXDelay, BufferSize, CommsMode)
Scan (1,Min,0,0) PanelTemp (PTemp,15000) Battery (batt_volt)
‘Scan Loop
‘Read multiple Input registers from the BARO Module using a 32 bit f loat, Big-Endian
‘format ModbusMaster(mb_status,ComC1,9600,BARO_MB_ADDR,4,Measurements(),32001,3,MB_RETRIES,MB_ TIMEOUT,2)
‘Call Output Tables CallTable BARO_Table
NextScan
EndProg
15
BARO MODULE INTEGRATOR GUIDE
CUSTOMER SUPPORT
NORTH AMERICA Customer service representatives are available for questions, problems, or feedback Monday through Friday, 7:00 am to 5:00 pm Pacific time.
Email:
support.environment@metergroup.com sales.environment@metergroup.com
Phone: +1.509.332.5600
Fax:
+1.509.332.5158
Website: metergroup.com
EUROPE
Customer service representatives are available for questions, problems, or feedback Monday through Friday, 8:00 to 17:00 Central European time.
Email:
support.europe@metergroup.com sales.europe@metergroup.com
Phone: +49 89 12 66 52 0
Fax:
+49 89 12 66 52 20
Website: metergroup.com
If contacting METER by email, please include the following information:
Name Address Phone number
Email address Instrument serial number Description of problem
NOTE: For products purchased through a distributor, please contact the distributor directly for assistance.
REVISION HISTORY
The following table lists document revisions.
Revision 00
Date 6.2025
Compatible Firmware Description
1.10
Initial release
16
Documents / Resources
![]() | METER BARO Module [pdf] User Guide TEROS 31, TEROS 32, BARO Module, BARO Module, Module |