HS100-B Blood Glucose Monitoring System User Manual

˙User Manual ˙Quick Reference Guide ˙Test Strip Package Insert ˙Control Solution Package Insert ˙Log Book ˙Carrying case 03. About Tyson Bio HS100-B Blood Glucose Meter 1. Test Port 2. Strip Ejector 3. Display To display measuring results and information in numbers and icons 4.

It is accurate, easy to use, and quick in response time. This User Manual contains all information needed to operate and maintain Tyson Bio HS100-B Blood Glucose Monitoring System. Please read carefully before use. Tyson Bio HS100-B Blood Glucose meter can be used with: ˙Tyson Bio HS100-B Blood Glucose Test Strip Intended Use Statement

TysonBio HS100-B- UserMan

TYSON BIORESEARCH, INC. HS100B Blood Glucose Monitoring System 2AJZ6HS100B 2AJZ6HS100B hs100b

PDF preview unavailable. Download the PDF instead.

GetApplicationAttachment.html?id=3898142
HS100-B Blood Glucose Monitoring System
User Manual

Limitations
The Tyson Bio HS100-B Blood Glucose Test Strips give accurate results when the following limitations are observed. Use only capillary whole blood from nger, palm, and forearm. Do not use for testing neonatal blood. Altitude higher than 10745 feet may cause inaccurate results. Severe dehydration and excessive water loss may lead to inaccurate blood glucose test results.
Consult your physician immediately if you believe you are su ering from severe dehydration. For single-patient use only and should not be shared. Should not be used to test critically ill patients. Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low results may occur for individuals experiencing a hyperglycemic-hyperosmolar state, with or without ketosis. For In vitro diagnostic only. For over the counter use. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Alternative site measurements should never be used to calibrate continuous glucose monitors (CGMs). Alternative site measurements should never be used for insulin dosing calculations.
Warning hThis device is not intended for use in healthcare or assisted-use settings such as hospitals, physician o ces, or long-term
care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. hUse of this device on multiple patients may lead to transmission of Human Immunode ciency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.

Table of Content

Limitations

Introduction

01

Important Safety Instructions

02

About Tyson Bio HS100-B Blood Glucose Monitoring System

03

About Tyson Bio HS100-B Blood Glucose Meter

04

About Tyson Bio HS100-B Blood Glucose Meter Display

05

About Tyson Bio HS100-B Blood Glucose Test Strip

07

About Tyson Bio HS100-B Blood Glucose Test Strip Label

08

Setting Up Tyson Bio HS100-B Blood Glucose Meter

09

Installing/Replacing Batteries

09

Date/Time Setup

10

Reminder Alarms Setup

11

HI/LO Alarms Setup

12

Important Information

13

Using Tyson Bio HS100-B Blood Glucose Test Strip

14

Obtaining a Blood Sample

15

Applying a Blood Sample

17

HI and LO Readings

19

Table of Content

Available Alternate Sites Testing

20

Control Solution

21

Performing a Control Solution Test

21

Control Solution Results

23

Memory Features

24

View Day Average Results

26

Bluetooth Data Transfer

28

Cleaning and Disinfection Procedure

29

Care and Storage

32

Display Messages

33

Error Messages

36

Trouble Shooting

39

Speci cations

40

Meter Storage

42

Icon Description

42

Service and Warranty

43

Federal Communications Commission Interference Statement

44

RF Exposure Warning

45

Introduction
Thank you for choosing Tyson Bio HS100-B Blood Glucose Monitoring System to monitor your blood glucose (sugar) level. It is accurate, easy to use, and quick in response time. This User Manual contains all information needed to operate and maintain Tyson Bio HS100-B Blood Glucose Monitoring System. Please read carefully before use.
Tyson Bio HS100-B Blood Glucose meter can be used with: Tyson Bio HS100-B Blood Glucose Test Strip
Intended Use Statement The Tyson Bio HS100-B Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the ngertips, forearm, or palm. Alternate site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the e ectiveness of diabetes control. It should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use.
The Tyson Bio HS100-B Blood Glucose Monitoring System is comprised of the Tyson Bio HS100-B Blood Glucose Meter and Tyson Bio HS100-B Blood Glucose Test Strip.
01

Important Safety Instructions
All parts of the system should be considered potentially infectious and are capable of transmitting blood-borne pathogens between patients and healthcare professionals.
The meter and lancing device are for single person use. Do NOT share them with anyone including other family members!
All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after you have performed cleaning and disinfection.
Cleaning Solution and Disinfecting Solution: Clorox Bleach Germicidal Wipes (Clorox Professional Products Company. EPA Reg. No. 67619-12).
Reference: "FDA Public Health Noti cation: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting
Bloodborne Pathogens: Initial Communication" (2010)http://www.fda.gov/MedicalDevices/Safety/AlertsandNotic es/ucm234889.htm "CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Blood-borne Pathogens" (2010) http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html "CDC: Infection Prevention during Blood Glucose Monitoring and Insulin Administration" (2010) https://ww w.cdc.gov/injectionsafety/blood-glucose-monitoring.html
02

About Tyson Bio HS100-B Blood Glucose Monitoring System
The complete kit contains: HS100-B Blood Glucose Meter HS100-B Blood Glucose Test Strip HS100-B Control Solution Level 2 Adjustable Lancing Device Sterile Lancets User Manual Quick Reference Guide Test Strip Package Insert Control Solution Package Insert Log Book Carrying case
Note: 25-strips & 50-strips are sold separately. The actual composition is marked on the box. Some of items may not be included.
03

About Tyson Bio HS100-B Blood Glucose Meter

HS100-B Blood Glucose Meter
2AJZ6HS100B Tyson Bioresearch, Inc. Zhunan, Taiwan

1. Test Port

Insert test strip here

2. Strip Ejector Push to eject the test strip

3. Display

To display measuring results and information in

numbers and icons

4.

Button For the operation of the meter. See more in the

further chapter

5. Meter Label Manufacturer information

6. Serial Numbers The meter manufacturer serials

7. Battery Cover To slide open and replace battery

AL001-0239000N (01)

04

About Tyson Bio HS100-B Blood Glucose Meter Display

1. Memory Icon

Appears during memory mode

2. Year/Date

To display year/date

1

3. Control Solution Icon

2

Appears during a Control Solution test and indicates its result

3

9

as a Control Solution test result

4

10 4. Alarm Icon

5

11

Appears when an alarm is set

6

12 5. Test Result

7

13

To display test results

14 6. Buzzer Icon

8

15

Appears while sound setting

16 7. Meal Indicator Icon

Pre-meal and post-meal

8. 14 day Average

14 day average result

05

9. Time

To display time.

1

10. Blood Drop Icon

2

Flashes when it's ready to perform a test and collect a blood sample.

3

9 11. Warning Icon

4

10

Appears when a result is out of range.

5

11 12. Test Strip Icon

6

12

Appears when the meter is ready for a test.

7

13 13. Temperature Icon

14

Appears when the meter exceeds the range of operating temperature.

8

15 14. Measurement Units

16

To identify test result units (mg/dL)

15. Battery Icon

Appears when battery power is low.

16. Bluetooth

Bluetooth connection.

Note:

The mg/dL is the standard unit in the United States. Meters used in the United States must be set to read in mg/dL.

06

About Tyson Bio HS100-B Blood Glucose Test Strip

Tyson Bio HS100-B Blood Glucose Test Strip is only for Tyson Bio HS100-B Blood Glucose Meter.

Use each strip ONCE ONLY!

1

2
A test strip consists of the following parts:

1. Target Area

Where a blood drop or control solution to be applied

2. Con rmation Window

To con rm su cient blood sample and control solution

3. Contact Bar

To be inserted into the test port and to activate the meter (top side up)

3 top side

07

About Tyson Bio HS100-B Blood Glucose Test Strip Label

HS100-B Blood Glucose Test Strips

Store at 39 ~ 86°F(4 ~ 30°C), 10~90% RH. Do not refrigerate. Do not store test strips outside this vial. Re-cap immediately. WARNING: Always close cap tightly. For testing of glucose in capillary whole blood.
Tyson Bioresearch, Inc. 5F., No. 16, 18, 20, 22, Kedong 3rd Rd., Zhunan Township, Miaoli County 35053, Taiwan Tel: +886-37-585998 Customer Service U.S.A. +1-800-948-3894
Test Strips

Use within 6 months after first opening. Do not use past the expiration date. Use Tyson Bio HS100-B meter only !
2018-12-25 Control Solution Range

Test Strip Vial Label

08

Date opened:_____/_____/_____ AL001-0237000N (01) Rev. date: 04/2018

HS100-B Blood

Store at 39 ~ 86°F(4 ~ 30°C), 10~90% RH. Do not refrigerate. Do not store test strips outside this vial. Re-cap immediately. WARNING: Always close cap tightly. For testing of glucose in capillary whole blood.
Tyson Bioresearch, Inc. 5F., No. 16, 18, 20, 22, Kedong 3rd Rd., Zhunan Township, Miaoli County 35053, Taiwan Tel: +886-37-585998 Customer Service U.S.A. +1-800-948-3894
Test Strips

Use within 6 mont Do not use past th Use Tyson Bio HS
2018-12-2 Control

Test Strip Vial

Setting Up Tyson Bio HS100-B Blood Glucose Meter
Make sure the batteries are properly installed The default is set to January 1st, 0:00 Please follow instructions to properly setup the meter

Installing/Replacing Batteries

e Test Strips

opening. 1. Slide and remove the battery cover (Fig.1).

date. r only !

2. Install batteries with the + and ­ ends matching indication marks on the battery compartment (Fig.2).

3. Slide battery cover back into place (Fig.3).

Range

HS100-B Blood Glucose Meter
2AJZ6HS100B Tyson Bioresearch, Inc. Zhunan, Taiwan

HS100-B Blood Glucose Meter
2AJZ6HS100B Tyson Bioresearch, Inc. Zhunan, Taiwan

HS100-B Blood Glucose Meter
2AJZ6HS100B Tyson Bioresearch, Inc. Zhunan, Taiwan

Date opened:_____/_____/_____ AL001-0237000N (01) Rev. date: 04/2018
AL001-0239000N (01) AL001-0239000N (01)
AL001-0239000N (01)

(Fig.1)

(Fig.2)

(Fig.3)

09

Date/Time Setup
1. When the meter is o , press and hold for 4 seconds to enter this setting mode. 2. Press or to select "24hr" or "12hr", and press to con rm (Fig.4). 3. The numbers of "year" will appear and ash. Press or to adjust and Press to con rm (Fig.5). 4. The number of "month" will appear and ash. Press or to adjust and press to con rm (Fig.6). 5. The numbers of "day" will appear and ash. Press or to adjust and press to con rm (Fig.7). 6. The numbers of "hour" will appear and ash. Press or to adjust and press to con rm (Fig.8). 7. The numbers of "minute" will appear and ash. Press or to adjust, and press to con rm (Fig.9). 8. The set date and time will now be displayed
on screen (Fig.10, Fig.11). Press to turn o the meter. 9. The meter date/time setting is complete.

(Fig.4)

(Fig.5)

(Fig.6)

(Fig.7)

(Fig.8)

(Fig.9)

(Fig.10)

(Fig.11)

10

Reminder Alarms Setup
Tyson Bio HS100-B Blood Glucose Meter provides 4 reminder alarms. All reminder alarms are OFF as default. Please follow these steps to setup reminder alarms.
1. When the meter is OFF, press and hold for 4 seconds or longer to enter setting. 2. The rst reminder alarm will appear on the display (Fig.12). 3. Press or to select "ON" or "OFF". 4. If "OFF" is chosen and pressed, the meter will proceed to the next reminder alarm setting. If "ON" is chosen (Fig.13)
and pressed, the number of "hour" will start ashing (Fig.14). Press or to adjust and press to con rm. 5. The numbers of "minutes" will ash (Fig.15). Press or to adjust and press to con rm. The meter will go to the
next reminder alarm setting. 6. Repeat steps 3 to 5 to set the rest of reminder alarms.

(Fig.12)

(Fig.13)

(Fig.14)

(Fig.15)

11

HI/LO Alarms Setup
1. Press to enter the HI alarm setting after completing the reminder alarm setup. 2. Press or to set "ON" and press to con rm. If "OFF" is chosen, the meter will skip this setting (Fig.16,17). 3. To change the alarm setting, press or until you reach your desired HI setting and then press to con rm.
(Fig.18) The HI alarm value range is 100 ~ 400mg/dL and the default value is 180mg/dL. 4. Continue to the LO alarm setting (Fig.19), repeat Step 2 and 3 (Fig.20,21) to con rm or to change LO alarm setting.
The LO alarm value range is 40 ~ 90mg/dL and the default value is 70mg/dL. 5. Press to nish setting and the meter will be turned o .

(Fig.16)

(Fig.17)

(Fig.18)

(Fig.19)

(Fig.20)

(Fig.21)

NOTE: The hypoglycemic (LO) and hyperglycemic (HI) alarm glucose values should be selected in consultation with your health care provider.
12

Important Information
Before Testing Always keep test strips in the original vial. The vial contains drying agents to protect the test strips from moisture. Tightly close the vial cap immediately after removing a test strip. Use each strip immediately after removing it from the vial. Use each test strip ONCE ONLY. Do not use the test strips or Control Solution beyond the expiration date printed on the package. This may cause
inaccurate results. Do not use the test strips or Control Solution if the package is damaged. Store your test strips and the Control Solutions between 39°F ~ 86°F (4°C ~ 30°C)
Temperature beyond this range can damage test strips and lead to inaccurate results. Any change of medications based on the Tyson Bio HS100-B Blood Glucose Monitoring System results without
advices of a doctor is NOT RECOMMENDED. The lancing device is intended only for single user and should NOT be shared. Users should wash hands thoroughly with soap and water after handing the meter, lancing device, or test strips. Only Tyson Bio HS100-B Blood Glucose Test Strip and Control Solution can be used with the Tyson Bio HS100-B Blood
Glucose Meter. For accurate results, testing must be done within the operating temperature range (50°F ~ 104°F/10°C ~ 40°C).
If the meter is moved to an area appropriate conditions from where is out of operating temperature range, place the meter for 30 minutes before conducting the testing.
13

Using Tyson Bio HS100-B Blood Glucose Test Strip

You need the following items to test Tyson Bio HS100-B Blood Glucose Meter Tyson Bio HS100-B Blood Glucose Test Strip Adjustable lancing device and sterile lancet

Preparing the Test Strip

1. Wash hands using soap and warm water. Rinse and dry thoroughly (Fig.22).

2. Please remove protective wrap completely before opening the vial.

3. Take a test strip from the vial and re-cap immediately (Fig.23).

4. Insert the test strip top side up into the test port (Fig.24). The meter will be automatically turned on.

5. System checking will appear (Fig.25).

6. A beep will sound and following by a ashing " " icon. You can press or to select the meal indicator for

marking the test as (Fig.26): Pre-Meal Test Post-Meal Test No marker as default

7. A blood sample can now

be obtained (refer to P.15).

(Fig.22)

(Fig.23)

(Fig.24)

(Fig.25)

(Fig.26)

14

Obtaining a Blood Sample
1. Unscrew the lancing device cap (Fig.27). 2. Place a lancet into the lancing device (Fig.28). 3. Twist and pull to remove the protective lancet cap to expose the sterile needle. Save the cap for later use (Fig.29). 4. Screw the lancing device cap back on. Select the skin penetration depth preferred (Fig.30). 5. Pull the barrel back until you hear a 'click' (Fig.31). 6. Place the lancing device against the nger, palm, or forearm. Press the trigger to obtain a blood sample (Fig.32).

(Fig.27)

(Fig.28)

(Fig.29)

(Fig.30)

(Fig.31)

(Fig.32)

15

To unload the lancing device 1. After sampling, carefully remove the cap containing the used lancet. 2. Push the exposed tip of the lancet into its Protective Cap (Fig.33).
Slide the Lancet Ejector forward and dispose the used lancet in an appropriate container.
For further instructions please see the insert provided with the lancing device.
(Fig.33)
CAUTION: To reduce chances of infection: Lancets are for single usage only. Wash hands thoroughly with soap and water before and after handling the meter, lancing device, and test
strips. Please refer to section "Cleaning and Disinfection Procedure" (refer to P.28) for detail instructions of meter
maintenance.
16

Applying a Blood Sample
SAMPLE MAY BE OBTAINED FROM FINGER, PLAM, OR FOREARM 1. Introduce the tip of the test strip to the drop of blood. Blood sample will be automatically drawn into the test strip
(Fig.34). Discard the rst drop of blood sample to prevent contamination. 2. Hold the tip of the test strip touching the blood drop until the meter beeps. 3. The meter will start to countdown as soon as the blood has completely lled the con rmation window of the test strip
If applying an insu cient blood sample, the error code (E-7) will appear on the display (Fig35). 4. The test result will appear on the display in 5 seconds (Fig.36). All results will be automatically stored into the
meter memory (Fig.37). 5. The meter will automatically turn o after 2 minutes. 6. After completing the test, push the strip ejector (Fig.38) or use a tissue paper to remove the test strip.
Su cient Blood

(Fig.34)

Insu cient blood (Fig.35)

(Fig.36)

(Fig.37)

(Fig.38)

17

Warning: DO NOT point the test strip at anyone when ejecting. Used lancets and test strips are considered bio-hazardous. Dispose of used lancets and test strips in a clinical waste
container Normal Glucose Values Before a meal (fasting) : <100 mg/dL 2 h after a meal: < 140mg/dL. These ranges were cited from the American Diabetes Association, Classi cation and Diagnosis of Diabetes, 2017. Diabetes Care 40(Suppl. 1):S11­S24
18

HI and LO Readings
Tyson Bio HS100-B Blood Glucose Meter is designed to display test results between 20 ~ 600 mg/dL. If "HI" (Fig.39) appears on the display, it indicates the reading is higher than 600 mg/dL. If "LO" (Fig.40) appears on the display, it indicates the reading is lower than 20 mg/dL.

(Fig.39)

(Fig.40)

We suggested the testing procedure should be reviewed and the test should be repeated using a new test strip to recon rm the test result.

19

Available Alternate Sites Testing

Palm and Forearm Tyson Bio HS100-B Blood Glucose Monitoring System provides you alternate sites testing (AST). This system allows you to test on the palm and the forearm with the equivalent results to ngertip testing. Physiological di erences in the circulation between the nger and other test sites may result in di erences in blood glucose measurements. Changes in blood glucose may be observed in nger blood samples sooner than blood samples from the alternate sites. If you are testing for hypoglycemia (low blood glucose), we recommend that you test on your ngertips.
AST ONLY in the following intervals: In a pre-meal or fasting state (more than 2 hours since the last meal) Two hours or more after taking insulin Two hours or more after exercise

Palm Forearm

DO NOT use AST if: You think your blood glucose is low You are unaware of hypoglycemia Your AST results do not match the way you feel You are testing for hyperglycemia Your routine glucose results are often uctuating 20

There are limitations for doing AST. Please consult your healthcare professional before you do AST. AST measurements should not be used to calibrate continuous glucose monitors (CGMs). AST measurements should not be used for insulin dose calculations.

Control Solution

The Control Solution is to ensure that Tyson Bio HS100-B Blood Glucose Monitoring System is working properly and you are performing a test correctly.

When to Perform a Control Solution Test

1. You use the meter for the rst time. 2. You use a new vial of test strips. 3. You suspect that the meter or test
strips are not working properly.

4. Your test results are not consistent with how you feel. 5. You think your test result is not accurate. 6. You damage your meter. 7. An advice from your healthcare professional.

Note: Please read the Control Solution package insert for more detail information.

Performing a Control Solution Test
1. Be sure of the Control Solution is at a room temperature between 59°F ~ 95°F (15°C ~ 35°C) before testing. Control solution that is not at room temperature when tested may cause out of range results.
2. Insert a Tyson Bio HS100-B Blood Glucose Test Strip, top side up, contact bar end rst, into the test port. The meter will be automatically turned ON (Fig.41).
3. A beep will sound and the system checking will appear followed by a ashing " " icon (Fig.42).

(Fig.41) (Fig.42)

21

4. Press and hold for 2 seconds to change to the Control Solution test mode. " " will appear on the display (Fig.43). Press again for 2 seconds and the meter will switch back to the normal testing mode.
5. Discard the rst drop of the Control Solution. Squeeze a small drop on a clean nonabsorbent surface e.g. wax paper. Do not apply the Control Solution to the test strip directly from the bottle because contamination may occur (Fig.44)
6. Introduce the tip of the test strip to the Control Solution drop. The Control Solution will be automatically drawn into the strip. Hold until the meter beeps. The meter will now start to count down and the Control Solution test result will appear on the LCD screen (Fig.45).
Caution: Follow the above instructions when performing a Control Solution test in order to prevent contamination.

(Fig.43)

(Fig.44)

(Fig.45)

22

Control Solution Results
The Tyson Bio HS100-B Blood Glucose Monitoring System is working properly only if the Control Solution test result is within the speci ed range printed on the test strip vial. Please repeat the test when the test result is out of range. Out of range result may be caused by: Incorrect steps while performing the test The Control Solution temperature is lower than 59°F (15°C) or higher than 95°F (35°C) Expired or contaminated Control Solution Expired or contaminated test strips Meter malfunction Note: The result will not be included in any day average calculation when the Tyson Bio HS100-B Blood Glucose
meter is set in the Control Solution test mode. DO NOT use Tyson Bio HS100-B Blood Glucose meter if the problem persists. Please contact customer service
immediately.
23

Memory Features
How to view results stored in the memory 1. When the meter is o , press button to turn on the meter. The Data and Time will be shown on (At the left-upper
corner. Firstly the year is shown for 1 second, and then Month-Day will be shown on in the same area). 2. Press to view previous results. "MEM" icon will be displayed on the top center area. Initially, the "year" will be
display on the top left corner (Fig.46). After 1 second "month-day" will then be displayed on the top left corner and "time" will be displayed on the top right corner (Fig.47). The most recent memory results will be displayed rst. 3. You can press or to scroll forwards and backwards through the results. 4. Press button to exit the stored test results or the meter will be shut o automatically after 2 minutes without any action.

(Fig.46)

(Fig.47)

24

How to view the control solution test result in the memory 1. When the meter is o , press button to turn on the meter. 2. Press button again, and the lastest control solution test result will be displayed (Fig.48). 3. Press button to turn o the meter or the meter will be shut o automatically after 2 minutes without any action.
(Fig.48)
Note: 1. A result with or symbol indicates that the reading was taken out of the meter's specified operating
temperature range and may not be accurate. This value is excluded from the 7/14/30 day average.
25

View Day Average Results
1. Press to turn the meter on, then Press button to view day average results.
2. 7/14/30 day average results marked with a no meal / pre-meal / post-meal and will be displayed on screen in sequence when is pressed (Fig.49 ~ 57).
3. When the meter displays 7 days with no meal status assigned, press button and the meter will then change to view stored test results in memory.
4. Press button to turn o the meter or the meter will be shut o automatically after 2 minutes without any action.

(Fig.49) (Fig.52)

(Fig.50) (Fig.53)

(Fig.51) (Fig.54)

(Fig.55)

(Fig.56)

(Fig.57)

26

1

3

4

2

5

1. day average 2. meal ag (no meal / pre-meal / post-meal ) 3. The number used for day average 4. The average calculated based on day average (upper left), number of results (upper right), meal ag (lower left). 5. The average calculated from all 14 day test results
Note: 1. The consecutive 7/14/30 day average is calculated from the test results obtained during the last
consecutive 7/14/30 day periods.

27

Bluetooth Data Transfer
Tranfer Memory Data 1. Press to turn on the meter (Fig.58). 2. When the meter is switched ON, hold for 4 seconds to enter Bluetooth mode. will appear and ash
on the meter when meter is not connected with a mobile device (Fig.59). 3. Turn Bluetooth on your mobile device, and BLE tool APP (with Blood Glucose Test) on. 4. Select the device shown (ex.Tysonbio-BGM) in APP list and click "Connect" . stops blinking when the
meter & mobile device is connected. 5. Press the button on APP to transfer data. The meter will display while data are being transferred (Fig.60).

(Fig.58)

(Fig.59)

(Fig.60)

28

Cleaning and Disinfection Procedure
Cleaning Cleaning removes visible soil, organic material, and most importantly, blood products. Always clean before disinfecting. The meter should be cleaned whenever it is visibly dirty. Disinfection Disinfection generally refers to the use of chemical and physical agents, such as chemicals, to destroy the potential inactivity of a material. The meter should be disinfected periodically, such as twice per week. Material Clorox Bleach Germicidal Wipes (Clorox Professional Products Company. EPA Reg. No. 67619-12). Signs of deterioration hPerformance: The control solution measurements are out of range which is listed on the test strip label. hPhysical: Polymer crazing (thin silver streaks appear), cracking, swelling, dissolving, softing, or becoming brittle,
and the symbols are not clear in LCD display. If these signs of deterioration are noted, users should stop using meter and contact customer service. Customer Service
U.S.A. +1-800-948-3894 24 hours a day, 7 days a week Taiwan Taiwan +886-37-585998 GMT+8, 8:30 ~ 17:30, Monday ~ Friday Note: Users should wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips should be included.
29

Cleaning and Disinfection Procedure

(Step 1) (Step 5) (Step 9)

(Step 2)

(Step 3)

(Step 6)

(Step 7)

(Step 10)

(Step 11)

(Step 4) (Step 8) (Step 12)

Cleaning: The meter should be cleaned prior to disinfection. Switch the meter OFF. Cleaning solution and disinfecting solution:
Clorox Bleach Germicidal Wipes Use a new cleaning wipe. The meter is cleaned follow by the illustrations
(Step1~Step15). Wipe the entire surface of the meter to clean
blood and other body uids. Please avoid using alcohol or organic solvents in
cleaning. Do not immerse the meter in water. Do not allow
water or cleaning uids into the meter, test port, and data port.

(Step 13) 30

(Step 14)

(Step 15)

Cleaning and Disinfection Procedure

(Step 1)

(Step 2)

(Step 3)

(Step 4)

(Step 5)

(Step 6)

(Step 7)

(Step 8)

(Step 9)

(Step 10)

(Step 11)

(Step 12)

Disinfection: Cleaning solution and disinfecting solution:
Clorox Bleach Germicidal Wipes  Use a new disinfection wipe.  The meter is cleaned follow by the illustrations
(Step1~Step15). Wipe the entire surface of the meter to remove
bloodborne pathogens. Dispose of the used wipe in a trash bin. Allow exteriors to remain wet for the 1 minute
contact time. Wipe surface using a towel and allow to air dry.

(Step 13)

(Step 14)

(Step 15)

31

Care and Storage
Please handle the meter with care. Dropping the meter may result in damage. Do not expose the meter, test strips, and the Control Solution to extreme conditions such as high humidity, heat,
freezing cold, or dust. DO NOT STORE IN DIRECT SUNLIGHT OR AREAS WITH HIGH HUMIDITY AND/OR DUST. It is advised that you store all parts of Tyson Bio HS100-B Blood Glucose Monitoring System in the carrying case
provided.
32

Display Messages

Display

Description
The system is ready to accept a blood sample.

Action
Apply a blood sample.

The result is HIGHER than the "HI" alert setting.
The result is LOWER than the "LO" alert setting.

The default setting is 180 mg/dL and it can be changed according to HI/LO Alarms Setup
The default setting is 70 mg/dL and it can be changed according to HI/LO Alarms Setup

33

Display

Description
Test result is HIGHER than 600 mg/dL.

Test result is LOWER than 20 mg/dL.

Temperature is TOO HIGH during the test procedure.

Action
Repeat the test by using a new test strip. If the same message displays again, the result is con rmed to be higher than the meter test range. Please consult your healthcare professional immediately for further advice.
Repeat the test by using a new test strip. If the same message displays again, the result is con rmed to be lower than the meter test range. Please consult your healthcare professional immediately for further advice.
Large variation may occur between results due to high or low temperature. Move to an environment (50°F ~ 104°F or 10°C ~ 40°C) and wait for 30 minutes before re-testing.

Temperature is TOO LOW during the test procedure.

Large variation may occur between results due to high or low temperature. Move to an environment (50°F ~ 104°F or 10°C ~ 40°C) and wait for 30 minutes before re-testing.

34

Display

Description
The preset reminder alarm is ringing

Action
Press any button to turn o the alarm or wait for the alarm to be turned o automatically in 30 seconds.

Battery power is LOW. Meter will only provide proximately 10 more measurements.

Replace with 2 AAA batteries.

An error message indicates that the hematocrit of the hematocrit of the blood sample is too high for the system to perform a test.

Follow the instructions and try again with a new test strip. Please contact customer service if the problem persists.

1. An error message indicates that the hematocrit of the blood sample is too low for the system to perform a test.
2. Use control solution to check the meter and test strips.

1. Follow the instructions and try again with a new test strip. Please contact customer service if the problem persists.
2. Follow the instructions to enter control solution mode, and try again with a new test strip.
35

Error Messages

Display

Description

Temperature is TOO HIGH to perform a test.

Temperature is TOO LOW to perform a test.

Battery power is depleted for any further usage.
This message is displayed when your meter has NOT been coded.
36

Action
Move to an environment between 50°F ~ 104°F or 10°C ~ 40°C and wait for 30 minutes before re-testing.
Move to an environment between 50°F ~ 104°F or 10°C ~ 40°C and wait for 30 minutes before re-testing.
Replace with 2 AAA batteries immediately
Please contact customer services for assistance.

Display

Description

Action

An error message indicates a problem with Review the instructions and try again with a new test strip.

the test strip.

Please contact customer service if the problem persists.

An error message indicates a problem with Review the instructions and try again with a new test strip.

the test strip or meter.

Please contact customer service if the problem persists.

An error message indicates a problem with the meter.

Remove batteries and wait for 1 minute. Reinstall batteries to see if meter is working properly. Please contact customer service if the problem persists.

An error message indicates a problem with the test strip.

Review the instructions and try again with a new test strip. Please contact customer service if the problem persists.

37

Display

Description

Action

An error message indicates a problem with the meter.

Remove batteries and wait for 1 minute. Reinstall batteries to see if meter is working properly. Please contact customer service if the problem persists.

An error message indicates a problem with Review the instructions and try again with a new test strip.

the test strip.

Please contact customer service if the problem persists.

An error message indicates insu cient volume of blood sample for the test.

Follow the instructions and try again with a new test strip, then applying a su cient blood sample to the test strip. Please contact customer service if the problem persists.

An error message indicates a problem with the meter.

Remove batteries and wait for 1 minute. Reinstall batteries to see if meter is working properly. Please contact customer service if the problem persists.

38

Trouble Shooting
Other problems which may occur A test strip has not been inserted into the meter properly.
Defective test strip.

Action Review the instructions and re-insert a test strip correctly.
Replace with a new test strip.

A test strip remains in the test port for more than 2 minutes before testing.

Review the instructions and re-insert a test strip correctly.

LCD display on the meter is blank when trying to perform a test.

Please contact customer services for assistance.

The meter does not function after new batteries are Remove batteries and wait for 1 minute. Reinstall

installed.

batteries to see if meter is working properly.

39

Specifications
Operation Temperature Operation Humidity Hematocrit Test Sample Sample Volume Measuring Unit Measuring Range Test Time Memory Capacity Average
40

50°F ~ 104°F (10°C ~ 40°C) 10 ~ 90% RH 10 ~ 65% Capillary Whole Blood ( ngertip, palm, and forearm) 0.7 µL mg/dL 20 ~ 600 mg/dL 5 seconds countdown 500 most recent results 7/14/30 days average

Power supply Battery Life Dimension Weight

2 AAA batteries Approximately 1000 tests 108L x 55W x 17H mm 60 g without batteries

Note: Please refer to the Test Strip Insert for the performance of system accuracy and precision.
The Meter meets the electromagnetic compatibility requirements as per IEC 60601-1-2:2014 and EN 60601-1-2:2015.
Caution: Do not use the meter close to the electronic equipment which is a source of strong electromagnetic radiation, which may interfere with proper operation of meter.

41

Meter Storage
Transport and store at temperature -4°F ~ 122 °F (-20°C ~ 50°C), less than 90% RH. Precision instrument, do not drop or crush when being transported.
Icon Description

Temperature limitation

Manufacturer

Batch code

In vitro diagnostic medical device

Use by

Consult instructions for use

Do not reuse

Su cient for

Serial Number

Catalogue number

Caution
42

Control

Service and Warranty
IMPORTANT The Tyson Bio HS100-B Blood Glucose Meter, Tyson Bio HS100-B Blood Glucose Test Strip and Tyson Bio HS100-B
Control Solution are in conformity with the CFR, Title 21 and IVDD 98/79/EC. Tyson Bio HS100-B Blood Glucose Monitoring System manufacturer warranty is valid only when used properly
within the guidelines of this User Manual provided. The Lancing device and Lancets are in conformity with the CFR, Title 21 and MDD 93/42/EEC.
Manufacturer Warranty Tyson Bioresearch Inc. o ers 3 years guarantee on this product. Our company shall repair or replace any Tyson Bio
HS100-B Blood Glucose Monitoring System found defective with a new one. This warranty does not apply to the performance of a Tyson Bio HS100-B Blood Glucose Monitoring System that
has been accidentally damaged, altered, misused, tampered with or abused in anyway. In no event shall our company be liable to the purchaser or any other person for any incidental, consequential, or punitive damages arising from or in anyway connected with the purchase or operation of Tyson Bio HS100-B Blood Glucose Monitoring System. For manufacturer warranty services, purchaser must contact Tyson Bioresearch, Inc. for help.
43

Federal Communications Commission Interference Statement
15.105 This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit di erent from that to which the receiver is connected. -Consult the dealer or an experienced radio/ TV technician for help.
15.21 Any changes or modi cations not expressly approved by the grantee of this device could void the user's authority to operate the equipment.
44

RF Exposure Warning
The equipment complies with FCC RF exposure limits set forth for an uncontrolled environment. The equipment must not be co-located or operating in conjunction with any other antenna or transmitter. 15.19 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
45

Manufacturer of Blood Glucose Meter, Test Strip and Control Solution
Tyson Bioresearch, Inc. 5F, No.16, 18, 20, 22, Kedong 3rd Rd., Zhunan Township, Miaoli County 35053, Taiwan +886-37-585998 www.tysonbio.com
Customer Service U.S.A. +1-800-948-3894 24 hours a day, 7 days a week Taiwan +886-37-585998 GMT+8, 8:30 ~ 17:30, Monday ~ Friday
When you call our customer service, please have all components of the product available. This will allow us to answer your questions with speed and e ciency.
Manufacturer of Lancet and Lancing Device Shandong Lianfa Medical Plastic Products Co., Ltd. No.1 Shuangshan Sanjian Road 250200 Zhangqiu City, Jinan, Shandong PEOPLE'S REPUBLIC OF CHINA
SteriLance Medical (Suzhou) Inc. No.68 Litanghe Road,Xiangcheng Suzhou, Jiangsu 215133 P.R.China
46

Tyson Bioresearch, Inc. 5F., No. 16, 18, 20, 22, Kedong 3rd Rd., Zhunan Township, Miaoli County 35053, Taiwan +886-37-585998 www.tysonbio.com
Customer Service U.S.A. +1-800-948-3894 24 hours a day, 7 days a week Taiwan +886-37-585998 GMT+8, 8:30 ~ 17:30, Monday ~ Friday

AF004-0068000N (01) Rev. date: 04/2018

