Catalyst One Analyzer Components

The Catalyst One analyzer features several key components:

  • Side door
  • Status LED
  • Start/power button
  • Lock light
  • Sample drawer
  • Waste drawer

External components include:

  • Pipette tips
  • Other reagent consumables
  • Whole blood separator
  • Sample cup
  • PHBR reagent consumable/automated dilution cups

Note: The sample drawer photo is for reference purposes only. Not all consumables should be loaded for a single run (for example, never load both a whole blood separator and a sample cup).

Sample Volume Required

The minimum sample volume required varies depending on the number of slides being run.

Sample TypeMinimum Volume
Whole BloodFill to lowest line on separator (0.7 cc [700 μL]).
Serum/Plasma/Urine300 μL is the recommended sample volume when running a Chem 17 CLIP and electrolytes.

Running a Sample

  1. Enter the sample information on the IDEXX VetLab* Station.
  2. Load the slides and sample in the sample drawer.
  3. Close the sample drawer and press the Start button on the analyzer.

Dilutions

The Catalyst One analyzer supports automated dilutions (the analyzer mixes the sample and diluent for you) and manual dilutions (you prepare the dilution outside of the analyzer). Do not dilute samples that are undergoing ammonia, phenobarbital, fructosamine, total T4, SDMA, pancreatic lipase, progesterone, bile acids, or electrolyte testing.

Running Quality Control

  1. Tap the Catalyst One icon on the IDEXX VetLab Station Home screen.
  2. Tap Maintenance and then tap Quality Control.
  3. Tap the quality control lot number you are using and then tap Run QC.
  4. Follow the on-screen instructions for preparing and running quality control.

Cleaning the Internal Components (Monthly)

  1. Tap the Catalyst One icon on the IDEXX VetLab Station Home screen.
  2. Tap Maintenance, tap Clean, and follow these on-screen instructions:
    1. Open the side door on your analyzer.
    2. Raise the carousel cover until the green lever magnetizes itself to the inside of the analyzer.
    3. Remove the carousel.
    4. Using an IDEXX-supported alcohol prep pad, wipe the incubator ring and optics window.
    5. Clean the reference tile on the carousel.
    6. Using an IDEXX-supported dry optical tissue, dry the optics window and reference tile.
    7. Replace the carousel, lower the carousel cover until it firmly clicks in place, and close the side door.
    8. Tap Done.

Catalyst CLIPs and Slides

The following table lists available chemistries and their compatibility with different Catalyst CLIPs and Individual Slides.

ChemistryAbbreviationChem 17 CLIPChem 15 CLIPChem 10 CLIPEquine 15 CLIPNSAID 6 CLIPUPC Panel+Lyte 4 CLIPSDMA and TT4 KitQC CLIPIndividual Slides
AlbuminALB
Alkaline phosphataseALKP
Alanine aminotransferaseALT
AmylaseAMYL
Aspartate aminotransferaseAST
Bile acidsBA
Blood urea nitrogenBUN
CalciumCa
CholesterolCHOL
Creatine kinaseCK
CreatinineCREA
ChlorideCl
C-reactive protein+CRP
Fructosamine+FRU
Gamma-glutamyltransferaseGGT
GlucoseGLU
PotassiumK
LactateLAC
Lactate dehydrogenaseLDH
LipaseLIPA
MagnesiumMg
SodiumNa
AmmoniaNH3
Phenobarbital+PHBR
Inorganic phosphatePHOS
Pancreatic lipase+PL
ProgesteronePROG
Symmetric dimethylarginine+SDMA
Total bilirubinTBIL
Total proteinTP
Total T4+TT4
TriglyceridesTRIG
Urine creatinineUCRE
Urine proteinUPRO
Uric acidURIC

+Validated reference intervals for equine and "other" species are unavailable.

#Validated reference intervals for feline, equine, and "other" species are unavailable.

Models: Catalyst One, Catalyst One Chemistry Analyzer, Catalyst One, Chemistry Analyzer, Analyzer

PDF preview unavailable. Download the PDF instead.

catalyst-one-quick-reference-guide-en Adobe PDF Library 17.0 PDFsam Enhanced 7

Related Documents

PreviewCatalyst SDMA Test Quick Reference Guide
A quick reference guide for the Catalyst SDMA Test, covering storage, handling, and frequently asked questions about its use with Catalyst One and Catalyst Dx analyzers.
PreviewCatalyst Pancreatic Lipase (PL) Test: A Quick Reference Guide
This guide provides essential information for running the Catalyst Pancreatic Lipase (PL) Test, including storage and handling requirements, sample types, species validation, and frequently asked questions for the Catalyst One and Catalyst Dx Chemistry Analyzers.
PreviewCatalyst SmartQC Control Quick Reference Guide
A quick reference guide for Catalyst SmartQC Control, detailing storage and handling requirements, and frequently asked questions about its use with Catalyst One and Catalyst Dx analyzers.
PreviewCatalyst Dx Chemistry Analyzer Quick Reference Guide
A concise guide for operating and maintaining the IDEXX Catalyst Dx Chemistry Analyzer, covering sample preparation, analysis, dilutions, and cleaning procedures.
PreviewSNAP Pro Analyzer Troubleshooting Guide
A comprehensive guide to troubleshooting common issues encountered with the SNAP Pro Analyzer, covering connection problems, test ejection errors, sample flow detection, and interpreting status icons.
PreviewIDEXX ProCyte Dx Hematology Analyzer Quick Reference Guide
A concise guide for operating the IDEXX ProCyte Dx Hematology Analyzer, covering analyzer components, sample preparation, running patient samples and quality control, exiting standby mode, monthly rinse procedures, powering on/off, and changing reagent kits and stain packs.
PreviewIDEXX VetLab UA Analyzer Operator's Guide - Veterinary Urinalysis Testing
Comprehensive operator's guide for the IDEXX VetLab UA Analyzer, detailing installation, operation, maintenance, and troubleshooting for veterinary urinalysis testing.
PreviewIDEXX SNAP Test Quick Reference Guide: Veterinary Diagnostic Tests
A comprehensive quick reference guide for IDEXX SNAP tests, covering various diagnostic assays for companion animals, including Heartworm, FIV/FeLV, cPL, fPL, Parvo, Giardia, Feline proBNP, Lepto, and Foal IgG tests. Details sample preparation, test procedures, and result interpretation.