WatchGas POLI Multi-Gas Detector - Quick Start Guide

This guide provides essential information for the quick setup and operation of the WatchGas POLI multi-gas detector.

Important Warnings

Read Before Use: The complete user guide for the POLI portable multi-gas monitor must be carefully read by all individuals who will use, calibrate, or maintain this product. The product will function as designed only if used, maintained, and serviced according to the manufacturer's instructions.

Caution: Change the battery only in a safe area (non-ATEX). Use only original spare parts, for example, replace the WatchGas battery with part number: M004-3002-000 (12.01.09.0001 or 12.01.09.0003) when depleted.

WARNING: Never use the monitor with the cover removed. Remove the monitor cover only in a safe area (non-ATEX). Use only the WatchGas lithium battery, part number M004-3002-000 (12.01.09.0001 or 12.01.09.0003). This instrument has not been tested in a gas/air explosive atmosphere with an oxygen concentration above 21%.

The replacement of components compromises intrinsic safety suitability and voids the warranty.

It is recommended to perform a BUMP TEST with a known gas concentration to confirm the instrument is functioning correctly before use.

Before use, ensure the colorless ESD layer on the display is not damaged or peeling. (The blue protective film can be removed.)

User Interface

The device features two function buttons, four sensor inputs, a large LCD display, eight alarm LEDs, a buzzer, and two vibration alarms.

Operation

Powering On

Press and hold the bottom key for 3 seconds until the buzzer sounds and the red LED flashes. After the warm-up and self-test sequence, the unit enters normal reading mode, displaying real-time gas concentrations.

Powering Off

In normal reading mode, press and hold the Down key for 5 seconds until "Power off" is displayed.

Charging

Always fully charge the battery before use. Preferably, use the WatchGas USB charger for charging the POLI. The charger can be plugged into the bottom of the POLI. The battery icon on the display shows the charging status. During active charging, the icon changes from empty to full, and a yellow alarm light remains on, turning green when charging is complete.

Charging/Communication Cable

The USB cable allows charging via a personal computer. The USB cable must be used for correct data transfer when communicating with the WatchGas Suite. It can also be used for partial or slow charging. A locally purchased cable may work for charging, but for correct data transfer with the WatchGas Suite, the WatchGas cable must be used.

Normal Mode

After warm-up, the unit enters normal reading mode and continuously displays instantaneous gas concentrations. Other values such as STEL, TWA, Peak, Min, Data, LEL gas, etc., can be viewed by pressing the Right key to cycle through these screens. The Left key "Test" checks the alarm function. Two text boxes and symbols at the bottom of the LCD are assigned to the Left key [+/OK] and Right key [↓]. The text or symbol in the boxes varies depending on the menu.

Indications of Left and Right key functions:

Configuration Mode

Press and hold both the Left and Right keys simultaneously for 3 seconds to enter Configuration Mode. The mode is password-protected with a default password "0000". Enter all four digits and press Left (OK) to access the configuration menu.

Configuration Menu Options:

Exiting Configuration Mode:

Repeatedly press the Right key until the "Exit" icon is highlighted, then press the Left key to exit.

Numeric Value Scrolling:

When modifying numeric values like span values or alarm limits, the function of the Left key [+/OK] can be switched between increasing (Inc/+) and decreasing (Dec/-) the number by pressing both keys simultaneously.

Modifying Alarm Limits

CAUTION: Disabling or modifying any alarm device prevents the notification of dangerous gas concentrations and can cause serious injury or death!

All alarm limits can be customized, including High, Low, STEL, and TWA alarms.

To enter alarm limit settings: Configuration Mode → Alarm Menu → Alarm Type.

To modify a value: Press the Right key to highlight the desired digit and press the Left key to change it. Hold the Left key to scroll through numbers continuously.

To save and exit: Press the Right key until "Done" appears in the bottom left corner, then press the Left key to exit. If changes were made, a message will appear to confirm saving the new values. Press "Yes" to save, "No" to cancel.

Bump Test

A bump test is a quick 30-second check to verify the functionality of sensors and alarms, without performing a precise calibration.

To enter the bump test: Configuration Mode → Calibration Menu → Bump Test.

Bump Test Results:

Clean Air Calibration

Zero calibration sets the baseline for the sensor and is performed in normal clean air or another clean air source.

To enter clean air calibration: Configuration Mode → Calibration Menu → Clean Air Calibration.

Span Calibration

Span calibration uses known gas concentrations to define the response of the sensor(s) to the gas or gases. Multiple sensors can be calibrated simultaneously using a gas mixture. Use a fixed flow regulator of approximately 0.5 LPM. For units with a pump, a variable flow rate regulator can also be used.

To perform span calibration: Configuration Mode → Calibration Menu → Span Value.

To perform span calibration: Configuration Mode → Calibration Menu → Span Calib.

Maintenance and Support

For units with a pump, regularly replace the inlet filter, especially if used in dusty environments. For battery or sensor replacement and other maintenance or support, contact WatchGas or refer to the user manual.

Menu Symbol Indications

PDF preview unavailable. Download the PDF instead.

Quick-Start-Guide-POLI-V1.4-IT Canva Canva

Related Documents

PreviewWatchGas PDM+ User Manual: Portable Gas Detector Guide
Comprehensive user manual for the WatchGas PDM+ single sustainable gas detector. Covers product overview, operation, calibration, maintenance, specifications, and safety certifications.
PreviewWatchGas PDM Multi Dock Manual: Bump & Docking Station
Comprehensive manual for the WatchGas PDM Multi Dock, a bump and docking station for gas detection monitors. Covers setup, configuration, logging, specifications, and warranty information.
PreviewWatchGas SST4 Series Multi Gas Detectors User Manual
Comprehensive user manual for the WatchGas SST4 series multi-gas detectors, including SST4 Micro, SST4 Mini, SST4 Pump, and SST5. Covers product features, safety information, setup, configuration via the WatchGas Device Link App, operation, maintenance, and troubleshooting.
PreviewWatchGas SST1 Single Gas Detector User Manual and Specifications
Comprehensive user manual for the WatchGas SST1 single gas detector, covering features, operation, safety, specifications, warranty, and compliance. Detects H2S, CO, SO2, O2, and more.
PreviewWatchGas UNI Sustainable Single Gas Detector Quick Start Guide
Quick start guide for the WatchGas UNI Sustainable Single Gas Detector, covering unit operation, user interface, display details, calibration, maintenance, and disposal.
PreviewWatchgas TOX Point 1000 User Manual
Comprehensive user manual for the Watchgas TOX Point 1000 Toxic Gas Detector, detailing specifications, installation, calibration, and safety precautions for industrial gas monitoring.
PreviewWatchGas QGM Multi Gas Detector User Manual
User manual for the WatchGas QGM portable multigas detector. This document provides detailed information on its features, operation, calibration, specifications, and warranty.
PreviewWatchGas SST4 Pump - Quick Start Guide
This guide provides essential information for the WatchGas SST4 Pump gas detector, covering safety warnings, operation, specifications, calibration, charging, and warranty details.