Seitron NOVO Combustion Analyzer Quick Guide

This guide provides essential information for operating the Seitron NOVO Combustion Analyzer.

Product Overview

The Seitron NOVO is a combustion analyzer designed for flue gas analysis, calculating stack heat loss and efficiency, ambient CO measurement, and more. It is compliant with EN 50379-1 and EN 50379-2 standards.

Key Features:

  • Flue gas analysis
  • Calculating stack heat loss and efficiency
  • Ambient CO measurement
  • Measuring differential pressure
  • Draft measurement
  • Pressure measurement in the gas supply pipe
  • Tightness test
  • Black smoke test and average value calculation
  • QR code display for downloading acquired measurements
  • Ticket printing with measurement data (built-in or optional Bluetooth® printer)

Instrument Controls and Functions

The instrument features several controls for analysis and operation:

Control #Description
1Configuration of the main analysis parameters.
2Acquires current measure or starts the combustion analysis according to set parameters.
3Displays TIME ELAPSED (time since last measure, running nº-: number of acquired measures) or TIME LEFT (time for next sample acquisition). Also shows LAST ANALYSIS date/time.
4Puts the current combustion analysis on hold.
5Interrupts the analysis and returns to the starting phase.
6Stores the acquired measures.
7Shows the QR code related to the performed analysis.
8Starts the current analysis printing phase, based on home page settings.
9Turns the smoke pump on/off. If the smoke pump is off, current measures refresh stops.

Visual Description: The instrument display shows analysis screens with parameters like Combustible (e.g., Propane), O2 (e.g., 20.9%), CO2 (e.g., 50%), T GAS, T AIR, Q, NO, H2, H2S, and elapsed time indicators.

Warning

WARNING: The magnets on the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators, and other devices sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 25cm away from these devices.

Operation and Connections

The Seitron NOVO analyzer requires specific connections for operation:

  • Smoke Probe (AJ SJ01): Connects to the instrument for gas sampling.
  • Gas Inlet/Pressure Hose (AJ TB01): Used for pressure measurements.
  • Accessory Connections (AJ KA02): For optional accessories or data transfer.
  • Other Probes (AA SA08): Used for specific gas measurements.

Visual Description: A diagram illustrates the setup: the analyzer connected via hoses and cables to a smoke probe inserted into a flue, and potentially to a laptop (AJ KA02) for data management. Extension cords can be used, up to three with a total length of 3 meters each, between the inputs and the smoke probe.

Suggested connections are used for measures such as O2, various gases (CO, NO, SO₂, CxHy, NO₂, H₂, CO₂, NH₃), efficiency calculation, combustion analysis, and simultaneous measurement of gas pressure at the burner.

Powering On and Off

Instrument ConditionActionFunction
OFF (powered off)Press and hold the key indicated by the arrow for a long time (> 3sec.).The instrument powers on and starts the automatic auto-zero phase.
ON (powered on)Press and hold the key indicated by the arrow for a long time (> 3sec.).The instrument turns off after completing the cleaning cycle for the set time.

Note: At the first start, the instrument will guide the user step-by-step through the initial configuration.

Main Menus

The instrument's main menu provides access to various functions:

  • 1. Analysis: Performs the complete combustion analysis. Accesses the analysis configuration menu to set parameters.
  • 2. Operator: Enters the Operator menu to select the operator's name for the analysis report.
  • 3. Measurements: Accesses accessory measurements the instrument can perform.
  • 4. Settings: Accesses the configuration menu to set instrument reference parameters.
  • 5. Print: Accesses the Print menu.
  • 6. Archive: Accesses the Archive menu to display stored measurements.

Visual Description: The main menu screen displays numbered icons or options, typically arranged in a grid or list, allowing users to navigate through different functions.

Models: EN 50379-1, EN 50379-2, EN 50379-1 Novo Combustion Analyzer, Novo Combustion Analyzer, Combustion Analyzer, Analyzer

PDF preview unavailable. Download the PDF instead.

037468 j100000000se eng Adobe PDF Library 15.0 Adobe InDesign 16.3 (Windows)

Related Documents

PreviewSeitron NOVO Gas Analyzer: Use and Maintenance Manual
Comprehensive guide for the Seitron NOVO Gas Analyzer, covering operation, maintenance, technical specifications, and troubleshooting for professional flue gas analysis.
PreviewSeitron NOVO Combustion Analyzer: Use and Maintenance Guide
Comprehensive guide detailing the operation, features, safety precautions, and maintenance procedures for the Seitron NOVO Combustion Analyzer, designed for professional flue gas analysis.
PreviewSeitron NOVO: Manual de Uso y Mantenimiento del Analizador de Combustión
Manual detallado de Seitron para el analizador de combustión NOVO. Cubre operación, mantenimiento, especificaciones técnicas y solución de problemas para profesionales.
PreviewManuel d'Utilisation et d'Entretien Analyseur de Combustion Seitron NOVO
Guide complet pour l'analyseur de combustion Seitron NOVO. Couvre le fonctionnement, la sécurité, l'entretien, les mesures de gaz, la connectivité et les spécifications techniques pour les professionnels du CVC et de l'industrie.
PreviewSeitron NOVO Analyseur de Combustion : Manuel d'Utilisation et d'Entretien
Manuel complet pour l'Analyseur de Combustion Seitron NOVO, couvrant l'utilisation, l'entretien, la configuration, les mesures, et le dépannages pour une performance optimale.
PreviewSeitron Be Safe MG Portable Multi-Gas Detector: Use and Maintenance Manual
Comprehensive guide for the Seitron Be Safe MG portable multi-gas detector, covering its use, maintenance, technical features, operation, safety precautions, and troubleshooting.
PreviewSeitron be safe SG: Use and Maintenance Manual for Single Gas Personal Detector
Comprehensive guide for the Seitron be safe SG single gas personal detector, covering operation, safety, calibration, and maintenance for reliable gas detection in hazardous environments.
PreviewSeitron Catalogo 2024: Analizzatori di Combustione, Sicurezza Gas e Termoregolazione
Esplora il catalogo Seitron 2024 per analizzatori di combustione professionali, rilevatori di gas per uso domestico e industriale, strumenti per la termoregolazione, climatizzazione e smart home. Scopri soluzioni innovative per la sicurezza e l'efficienza.