Introduction
Thank you for selecting the easy@Home arm type Blood Pressure Monitor (EBP-095). This monitor measures blood pressure, pulse rate, and stores results. It is designed for two years of reliable service.
Readings from the EBP-095 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual provides important safety and care information, along with step-by-step instructions for product use. Please read it thoroughly before use.
Features:
- 60mm x 80mm Digital LCD display
- Maximum 60 records
- Measuring during inflation technology
Safety Information
The following symbols may appear in the user manual, labeling, or other components, indicating compliance with standards:
- ℹ Symbol for "THE OPERATION GUIDE MUST BE READ"
- ⚒ Symbol for "MANUFACTURER"
- SN Symbol for "SERIAL NUMBER"
- ☑ For indoor use only
- F1 T1A/250V Φ3.6*10CCC Fuse rating
- ★ Symbol for "TYPE B APPLIED PARTS" (indicating protection against electric shock)
- ♻ Symbol for "ENVIRONMENT PROTECTION" (electrical waste products should be disposed of according to local guidelines)
- ⚡ Symbol for "DIRECT CURRENT"
- ⅀ Symbol for "Class II Equipment"
- ⏱ Symbol for "MANUFACTURE DATE"
⚠ CAUTION
- This device is intended for adult use only.
- It is for non-invasive measuring and monitoring of arterial blood pressure, not for use on extremities other than the arm or for functions other than blood pressure measurement.
- Do not confuse self-monitoring with self-diagnosis. This unit monitors blood pressure; do not begin or end medical treatment without physician's advice. Consult your physician regarding medication and measurement timing.
- If cuff pressure exceeds 40 kPa (300 mmHg), the unit will automatically deflate. If it doesn't, detach the cuff and press START/STOP.
- Read this manual carefully to avoid measurement errors.
- The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
- The operator should not touch battery/AC adapter output and the patient simultaneously.
- Do not wind the air tube around the neck.
- Use only accessories and detachable parts specified/authorized by the manufacturer.
- Luer lock connectors are not used; do not change any provided connectors.
- Manufacturer will provide circuit diagrams, component parts list etc. upon request.
- WARNING: No modifications of this equipment are allowed.
- This unit is not suitable for continuous monitoring during medical emergencies or operations. Continuous monitoring may cause the patient's arm and fingers to become anesthetic, swollen, or purple due to lack of blood.
- Use the device within the environment specified in the user manual to avoid impacting performance and lifetime.
- Cuff materials comply with ISO 10993-5:2009 and ISO 10993-10:2010, and will not cause allergic reaction or contact injury.
- The device does not require calibration within two years of reliable service.
- Dispose of accessories, detachable parts, and ME EQUIPMENT according to local guidelines.
- For patients with common arrhythmias (atrial or ventricular premature beats, atrial fibrillation), results may deviate.
- This device may provide contradictory results for pregnant females; effects on the fetus are unknown.
- Clinical evaluation confirms measurements are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method, within American National Standard limits.
- For problems with setup, maintenance, or use, contact easy@Home service personnel. Do not open or repair the device yourself. Report any unexpected operation or events to easy@Home.
LCD Display Signal
The LCD displays various symbols and values:
- SYS: Systolic blood pressure (high pressure result)
- DIA: Diastolic blood pressure (low pressure result)
- Pul/min: Pulse (pulse/minute)
- ↓: Deflating (cuff air is exhausting)
- ⏱ ⏰: Current Time (year:month:day:hour:minute)
- MEM: Memory (displayed values are from memory)
- kPa mmHg: Measurement unit (1mmHg = 0.133kPa; 1kPa = 7.5mmHg)
- ⚠ Lo+: Low battery (batteries are low and need replacement)
- ♥: Irregular heartbeat (detection of irregular heartbeat)
- ●: Grade (grade of blood pressure)
- ♡: Heartbeat (heartbeat detection during measurement)
- ⚠: Shocking reminder (shocking will result in inaccurate readings)
- ♟1: User 1 (start measurement for user 1 and save result automatically)
- ♟2: User 2 (start measurement for user 2 and save result automatically)
- ↻: Data Enquiry Mode (recall records)
Monitor Components
The blood pressure measuring system includes:
- Blood Pressure Monitor (EBP-095)
- Cuff (22cm-32cm, Type B applied part)
- 4 x AA Batteries
- User manual
Diagram shows the monitor with LCD display, START/STOP button, SET/DOWN button, MEM/UP button, and battery compartment. The cuff is connected via an air hose and air connector plug.
Before You Start
The Choice of Power Supply
- Battery powered mode: Uses 6VDC 4*AA batteries.
- AC Power adaptor (optional, not included): Input: 100-240VAC 50/60Hz 0.3A Max; Output: 6V --- 1000mA (Conforms to UL certificate). The power adaptor hole is shown on the device.
Installing and Replacing the Batteries
- Slide off the battery cover.
- Install batteries matching correct polarity (+/-).
- Replace the cover.
Replace batteries when: the ⚠ Lo+ symbol shows, the display dims, or the display does not light up.
Setting Date, Time and Measurement Unit
It is important to set the clock for time stamps on stored records (year: 2011-2050, 12 H format).
- When off, hold "SET" for 3 seconds to enter year setting mode.
- Press "MEM" to change the [YEAR]. Each press increases the numeral by one in a cycling manner.
- Once the correct year is set, press "SET" to confirm and proceed to [MONTH] setting.
- Repeat steps 2 and 3 to set the [MONTH] and [DAY].
- Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
- Repeat steps 2 and 3 to set the [MEASUREMENT UNIT].
- After [MEASUREMENT UNIT] is set, the LCD will display "dOnE" and then turn off.
Measurement
Tie the Cuff
- Tie the cuff on your upper arm. Position the tube off-center toward the inner side of the arm, in line with the little finger.
- The cuff should be snug but not too tight. One finger should fit between the cuff and your arm.
- Sit comfortably with your arm resting on a flat surface.
- Patients with hypertension should sit with correct posture:
- Bare your arm or wear tight clothing only when measuring.
- Sit comfortably with legs uncrossed, feet flat on the floor, back and arm supported.
- The center of the cuff should be at the same level as the right atrium of the heart.
- Rest for 5 minutes before measuring.
- Wait at least 3 minutes between measurements for blood circulation to recover.
- For meaningful comparison, measure under similar conditions (e.g., same time, same arm) or as directed by a physician.
Diagrams illustrate proper cuff placement on the upper arm, showing the tube aligned with the little finger and a 2-3cm gap from the elbow. Another diagram shows a person sitting correctly with arm resting on a table for measurement.
Start the Measurement
Before starting, press the SET button to choose User 1 or User 2 as the User ID. Press START/STOP to confirm.
- After selecting the user, press "START/STOP" to begin measurement. The device will adjust to zero, then inflate and measure. The results will be displayed and saved. The backlight color indicates the blood pressure grade.
- Press "SET", "MEM", or "START/STOP" to power off, otherwise it will turn off within 1 minute.
Data Management
Recall the Records
- When the monitor is off, press "MEM" to display the average value of the latest three records.
- Press "MEM" or "SET" to rotate through the records. Up to 60 records are stored under each user ID. Date and time will be shown alternately with pulse.
Delete the Records
To delete all results for the selected user:
- Hold "MEM" for 3 seconds when the monitor is in memory mode; "DEL ALL" will show.
- Press "MEM" to confirm deleting. The monitor will turn off.
- If you do not want to delete records, press "START/STOP" or "SET" to escape.
- If there are no records, the display will show dashes when recalling.
Information for User
Tips for Measurement
Measurements may be inaccurate if taken in the following circumstances:
- Within 1 hour after dinner or drinking.
- Immediately after tea, coffee, or smoking.
- Within 20 minutes after taking a bath.
- While talking or moving your fingers.
- In a very cold environment.
- When needing to discharge urine.
Diagrams illustrate situations to avoid for accurate measurement, such as eating, drinking, smoking, bathing, talking, moving fingers, being cold, or needing to urinate.
Maintenance
To ensure best performance, follow these instructions:
- Store in a dry place, away from sunshine.
- Avoid touching water; clean with a dry cloth.
- Avoid intense shaking and collisions.
- Avoid dusty and unstable temperature environments.
- Use wet cloths to remove dirt.
- Avoid washing the cuff.
Diagrams illustrate maintenance tips, such as keeping the device dry, avoiding water, avoiding impacts, avoiding dust/temperature extremes, cleaning with a damp cloth, and not washing the cuff.
About Blood Pressure
What are systolic pressure and diastolic pressure?
Systolic pressure is the maximum blood pressure value when ventricles contract and pump blood out of the heart. Diastolic pressure is the minimum blood pressure value when ventricles relax.
Diagrams show blood discharging from the artery (systolic) and blood entering the vein (diastolic).
What is the standard blood pressure classification?
The chart below reflects blood pressure categories defined by the American Heart Association (AHA).
Backlight on BPM | Blood Pressure Category | Systolic mmHg (upper#) | Diastolic mmHg (lower#) |
---|---|---|---|
Blue | Normal | less than 120 | and less than 80 |
Blue | Prehypertension | 120-139 | or 80-89 |
Yellow | High Blood Pressure (Hypertension) Stage 1 | 140-159 | or 90-99 |
Red | High Blood Pressure (Hypertension) Stage 2 | 160 or higher | or 100 or higher |
Red | Hypertensive Crisis (Emergency care needed) | Higher than 180 | or Higher than 110 |
AHA Home Guideline for Upper Limit of Normal BP: SYS 135 mmHg, DIA 85 mmHg.
Irregular Heartbeat Detector
An irregular heartbeat is detected when the rhythm varies during measurement. If the calculated standard deviation is ≥ 15, the irregular heartbeat symbol (♥) appears with the results.
Why does my blood pressure fluctuate throughout the day?
- Individual blood pressure varies multiple times daily and is affected by cuff tying and measurement position. Measure under consistent conditions.
- Medication can cause pressure variations.
- Wait at least 3 minutes between measurements.
Diagram shows a person taking blood pressure, indicating daily fluctuations.
Why do I get a different blood pressure at home compared to the hospital?
Blood pressure differs throughout the day due to weather, emotion, exercise, etc. The "white coat" effect often causes blood pressure to increase in clinical settings.
What to pay attention to when measuring at home: Ensure the cuff is properly tied (not too tight or too loose) on the upper arm. Avoid anxiety. Take 2-3 deep breaths before measuring. Relax for 4-5 minutes until calm.
Is the result the same if measuring on the right arm?
It is acceptable to measure on both arms, but results may differ between individuals. It is suggested to measure on the same arm every time for consistent comparison.
Diagram shows a person measuring blood pressure on their arm.
Troubleshooting
This section lists common problems and their remedies. If the product does not operate as expected, check here before seeking service.
Problem | Symptom | Check This | Remedy |
---|---|---|---|
No power | Display is dim or will not light up. | Batteries are exhausted. | Replace with new batteries. |
Batteries are inserted incorrectly. | Insert the batteries correctly. | ||
AC adaptor is inserted incorrectly. | Insert the AC adaptor tightly. | ||
Low batteries | ⚠ Lo+ shows on the display. | Batteries are low. | Replace with new batteries. |
Error message | E 1 shows | The cuff is not secure. | Refasten the cuff and then measure again. |
E 2 shows | The cuff is very tight. | Refasten the cuff and then measure again. | |
E 3 shows | The pressure of the cuff is excess. | Relax for a moment and then measure again. | |
E 10 or E11 shows | The monitor detected motion while measuring. | Movement can affect measurement. Relax for a moment and then measure again. | |
E20 shows | The measurement process does not detect the pulse signal. | Loosen the clothing on the arm and then measure again. | |
E21 shows | The treatment of the measurement failed. | Relax for a moment and then measure again. | |
EExx shows on the display. | A calibration error occurred. | Retake the measurement. If problem persists, contact retailer or customer service. Refer to warranty for contact and return instructions. |
Specifications
Feature | Details |
---|---|
Power supply | Battery powered: 6VDC 4*AA batteries. AC adaptor powered: INPUT: 100-240VAC 50/60Hz 0.3A Max; OUTPUT: 6V --- 1000mA (Not Included). |
Display mode | Digital LCD V.A.60mm*80mm (2.36"*3.15") |
Measurement mode | Oscillographic testing mode |
Measurement range | Pressure: 0mmHg~300mmHg (0kPa-40kPa); Pulse value: (40-199)times/minute |
Accuracy | Pressure: 5℃-40℃ (41℉-104℉) within ±3mmHg (0.4kPa); Pulse value: ±5% |
Normal working condition | Temperature: 5℃ to 40℃ (41℉ to 104℉); Relative humidity: ≤85%RH; Atmospheric pressure: 86kPa to 106kPa |
Storage & transportation condition | Temperature: -20℃-60℃ (-4℉ to 140℉); Relative Humidity: 10%RH-93%RH; Atmospheric Pressure: 50kPa-106 kPa |
Measurement perimeter of the upper arm | About 22cm~32cm (8"-12") |
Net Weight | Approx. 388g (13.69oz) (Excluding the batteries) |
External dimensions | Approx. 102mm*143mm*73mm (4.02"*5.63"*2.87") |
Attachment | 4*AA batteries, one storage bag, user manual |
Mode of operation | Continuous operation |
Degree of protection | Type B applied part |
Protection against ingress of water | IPX0 |
Software Version | V01 |
Authorized Components
The following components are authorized for use with this device:
- easy@Home authorized adapter (Not Included)
- Storage bag
Diagrams show the authorized adapter and the storage bag.
Complied Standards List
The device complies with various international standards:
Standard Category | Standard Reference |
---|---|
Risk management | ISO 14971:2007 |
Labeling | EN 980:2008 |
User manual | EN 1041:2008 |
General Requirements for Safety | IEC 60601-1:2005 |
Electromagnetic compatibility | IEC 60601-1-2:2007 |
Non-invasive Sphygmomanometers General Requirements | ASSI/AAMI SP10:2002/A1:2003/A2:2006/ (R)2008 |
Software Lifetime | EN 62304:2006/AC:2008 |
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
EMC Guidance
This section provides guidance and manufacturer's declaration regarding electromagnetic emissions and immunity for the ME EQUIPMENT and ME SYSTEMS, ensuring the device operates correctly in specified electromagnetic environments.
Electromagnetic Emissions
The device uses RF energy only for its internal function, resulting in very low RF emissions that are unlikely to cause interference with nearby electronic equipment. It complies with CISPR 11 Group 1 and Class B for RF emissions, and is not applicable for Harmonic emissions (IEC 61000-3-2) or Voltage fluctuations/flicker emissions (IEC 61000-3-3).
Electromagnetic Immunity
The device is designed to operate in typical commercial or hospital environments. It maintains performance under various electromagnetic disturbances:
- Electrostatic discharge (ESD) (IEC 61000-4-2): Complies with ±6 kV contact and ±8 kV air discharge. Floors should be wood, concrete, or ceramic tile, or if synthetic, relative humidity should be at least 30%.
- Electrical fast transient/burst (IEC 61000-4-4): Complies with ±2 kV for power supply lines and ±1 kV for input/output lines. Mains power quality should be typical of commercial or hospital environments.
- Surge (IEC 61000-4-5): Complies with ±1 kV line(s) to line(s) and ±2 kV line(s) to earth. Mains power quality should be typical of commercial or hospital environments.
- Voltage dips, short interruptions and voltage variations on power supply input lines (IEC 61000-4-11): Complies with specified dips and interruptions. If continuous operation during power mains interruptions is required, the device should be powered from an uninterruptible power supply or a battery.
- Power frequency (50/60Hz) magnetic field (IEC 61000-4-8): Complies with 3 A/m. Magnetic fields should be characteristic of a typical commercial or hospital environment.
Recommended Separation Distances
To prevent electromagnetic interference, maintain minimum separation distances between portable/mobile RF communications equipment and the device. The required distance depends on the transmitter's maximum output power and frequency.
For example, for a transmitter with 1W output power:
- 150 kHz to 80 MHz: 1.167 meters
- 80 MHz to 800 MHz: 1.167 meters
- 800 MHz to 2.5 GHz: 2.333 meters
Field strengths from fixed RF transmitters (e.g., base stations, broadcast radio/TV) should be less than the compliance level. If measured field strength exceeds the level, observe the device for normal operation; if abnormal performance occurs, re-orient or relocate the device.
Contact Information
For more information, visit Healthcare-Manager.com, or call easy@Home Customer Support toll-free at 1-855-822-6999 (M-F 9 a.m.-5 p.m. CST). Email: service@healthcare-manager.com. easy@Home will serve you anytime for usual problems and customer downloads.