Thank you for purchasing the OMRON EVOLV Automatic Upper Arm Blood Pressure Monitor.
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. It is mainly designed for general household use.
Important Safety Information:
- Consult your physician before using this monitor if you have common arrhythmias, arterial sclerosis, poor perfusion, diabetes, pregnancy, pre-eclampsia, or renal disease.
- DO NOT adjust medication based on measurement results.
- DO NOT use on an injured arm or an arm under medical treatment.
- DO NOT wear the arm cuff on your arm while on an intravenous drip or blood transfusion.
- DO NOT apply or inflate the arm cuff around any body parts other than your upper arm.
- DO NOT use this monitor in locations where RF is restricted, such as on an aircraft or in hospitals.
- Contains small parts that may cause a choking hazard if swallowed by children.
- Keep batteries out of reach of children.
- Stop using this monitor and consult your physician if you experience skin irritation or discomfort.
- Note that PATIENT motion, trembling, shivering may affect the measurement reading.
- Do not use this monitor in a moving vehicle.
- Do not drop, or subject this monitor to strong shocks or vibrations.
- Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement.
- Rest for at least 5 minutes before taking the measurement.
- Remove tight-fitting or thick clothing from your arm while taking a measurement.
- Remain still and do not talk while taking a measurement.
- Do not replace the battery while your reading is being transferred to your smart device.
- Do not insert batteries with their polarities incorrectly aligned.
- Use only 4 “AAA” alkaline batteries. Do not use other types of batteries. Do not use new and used batteries together.
- Remove batteries if the monitor will not be used for a long period.
- If battery fluid gets in your eyes, rinse with plenty of clean water and consult a physician immediately.
- If battery fluid gets on your skin, wash with plenty of clean, lukewarm water. Consult a physician if irritation persists.
- Do not use batteries after their expiration date.
Understanding Your Blood Pressure
Blood pressure is the force of blood flowing against the walls of the arteries. Arterial blood pressure changes during the heart’s cycle. The highest pressure is Systolic Blood Pressure, and the lowest is Diastolic Blood Pressure. Both are needed for a physician to evaluate your blood pressure status.
Arrhythmia: A condition where the heartbeat rhythm is abnormal. Symptoms can include skipped heartbeats, premature contraction, or an abnormally fast or slow pulse.
Many factors influence blood pressure, such as physical activity, anxiety, and time of day. A single measurement may not be sufficient for an accurate diagnosis. Measure at the same time each day for consistency. Blood pressure is typically lower in the morning and increases throughout the day. It is lower in summer and higher in winter.
Hypertension (high blood pressure) is a key risk factor for Stroke. Effective treatment can prevent strokes.
Home Blood Pressure Monitoring is endorsed to help manage hypertension effectively.
What's Included
Monitor, battery set, instruction manual, setup instructions, soft case.
Key Symbols
- Connection symbol: ?
- Connection OFF symbol: ??
- Connection ON symbol: ✅?
- Battery symbol (low): ?⬇️
- Battery symbol (depleted): ?❌
- Deflation symbol: ?
- Movement error symbol: ?♂️❌
- Heartbeat symbol: ❤️ (Flashes during measurement.)
- SYNC symbol: ?
- Cuff Wrap Guide Symbol: ?
- Irregular heartbeat symbol: ?
Getting Started
1. Installing Batteries
- Remove the battery cover.
- Insert 4 “AAA” batteries as indicated.
- Replace the battery cover.
Note: When the depleted battery symbol "?❌" appears, replace all 4 batteries with new ones.
2. Pairing Your Monitor with a Smart Device
- Turn on the Bluetooth feature of your smart device.
- Download and install the free “OMRON connect” app from your device's app store. If you already have the app, go to Menu > Devices > Add Device.
- Open the app and follow the set-up and pairing instructions.
- Confirm the monitor is connected successfully. “OK” will appear on the display if connected. Note: If you use any other app, the date and time may not be automatically set.
Compatibility: Review the list of compatible smart devices at: omronconnect.com/devices
For details: Visit omronconnect.com
Taking a Measurement
1. Applying the Arm Cuff
- Remove tight-fitting clothing from your left upper arm. Do not place the arm cuff over thick clothes.
- Apply the arm cuff to your left upper arm. The bottom edge should be 1 to 2 cm above the elbow. Position the display for clear viewing.
- Securely close with the fabric fastener.
Notes:
- Blood pressure can differ between arms. OMRON recommends using the same arm for measurement. Consult your physician if values differ substantially.
- Position the monitor with the [START/STOP] button towards your body for easier use.
2. Positioning for Measurement
Be relaxed and comfortably seated at a comfortable room temperature.
- Sit in a chair with legs uncrossed and feet flat on the floor.
- Sit with your back and arm supported.
- The arm cuff should be placed on your arm at the same level as your heart.
3. Starting the Measurement
- Press the [START/STOP] button. The arm cuff will start to inflate.
- Remain still and do not talk during measurement.
Cuff Wrap Guide:
- “ ” appears: The arm cuff was wrapped snugly enough.
- “ ” appears: The arm cuff was loose or not wrapped correctly.
Stopping the Measurement: Press the [START/STOP] button once to deflate the arm cuff.
Measurement Storage: This device stores 100 measurements. View all measurements using the “OMRON connect” app.
4. Transferring Your Readings
- Ensure your monitor is within 5 m (16 ft.) of your smart device and the “OMRON connect” app is active.
- As soon as the measurement is complete, your reading will be automatically transferred to the app.
- If readings are not automatically transferred, open the app and follow instructions to transfer data. “OK” will flash on the display if successful.
Notes:
- When the SYNC symbol "?" appears, transfer data immediately before readings are deleted.
- Once data is transferred, it cannot be resent.
5. Turning Off the Monitor
Press the [START/STOP] button to turn off the monitor. It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes between measurements for arteries to return to their prior condition.
Using Your Monitor Without Pairing
- Apply the arm cuff.
- Press the [START/STOP] button.
- Write down your SYS/DIA and Pulse readings, along with the date and time, on paper to keep a record.
Note: Without a smart device, you cannot view prior readings stored in memory.
Disabling Wireless Transmission
Turn off the wireless feature in the following areas where use of wireless equipment is prohibited:
- On aircraft
- In hospitals
- While abroad
To turn off wireless transmission:
- Press and hold the connection button for more than 10 seconds until "??" appears.
- Press the [START/STOP] button to turn off the monitor.
Resetting Your Monitor
To reset to factory settings (deletes all stored information):
- Ensure the monitor is turned off.
- While holding the connection button down, press the [START/STOP] button for more than 5 seconds.
- When the “Clr” symbol appears, release both buttons.
- Press the [START/STOP] button to turn off the monitor.
Note: Resetting the monitor does not delete information in the app. You may need to delete the monitor registration in the app first before re-pairing.
Troubleshooting
Error Symbols and Messages
- Irregular heartbeat detected (?): Remove cuff, wait 2-3 minutes, and re-measure. If it persists, consult your physician.
- Movement error (?♂️❌): Remove cuff, wait 2-3 minutes, and re-measure, remaining still.
- Cuff applied too loosely (?): Apply cuff tighter.
- Low batteries (?⬇️): Replace all 4 batteries with new ones.
- Depleted batteries or incorrect polarity (?❌): Replace all 4 batteries with new ones, ensuring correct polarity.
- Connection failure / Data not transmitted: Ensure monitor is within 5m of smart device, Bluetooth is on, and the OMRON connect app is active. Re-pair if necessary.
- Device error (“Err” with number): Contact your local OMRON representative.
- No power / No display: Replace batteries and check installation.
- Measurement values too high or too low: Review proper cuff application and measurement procedures. Blood pressure varies due to stress, time of day, etc.
Specifications
- Model: EVOLV (HEM-7600T-E)
- Display: OLED
- Measurement Method: Oscillometric method
- Pressure Measurement Range: 0 to 299 mmHg
- Pulse Measurement Range: 40 to 180 beats/min.
- Accuracy: Pressure: ±3 mmHg, Pulse: ±5% of reading
- Power Source: 4 “AAA” batteries (1.5V)
- Battery Life: Approximately 300 measurements
- Operating Temperature: 10°C to 40°C / 15 to 90% RH (non-condensing)
- Storage Temperature: -20°C to 60°C / 10 to 90% RH (non-condensing)
- Weight: Approx. 240 g (without batteries)
- Dimensions: Approx. 85 mm × 120 mm × 20 mm (monitor only)
- Arm Circumference: 22 to 42 cm
- Wireless Technology: Bluetooth low energy, 2.4 GHz band
- IP Classification: IP 22
- Applied Part: Type BF (Cuff)
- Manufacturer: OMRON HEALTHCARE Co., Ltd., Japan
Maintenance and Storage
- Store the device and components in a clean, safe location.
- Do not use abrasive or volatile cleaners.
- Do not wash the device or immerse in water.
- Use a soft dry cloth, or a soft cloth moistened with neutral soap to clean the monitor and arm cuff, then wipe with a dry cloth.
- Do not disassemble or attempt to repair the device.
- Store away from extreme temperatures, humidity, direct sunlight, dust, corrosive vapors, vibrations, or shocks.
Warranty
This product is guaranteed by OMRON for 3 years from the date of purchase. The warranty covers defects in construction, workmanship, and materials. It does not cover transport costs, unauthorized repairs, periodic check-ups, wear of optional parts, accidental damage, or misuse. Calibration service is not included.
Contact the dealer or an authorized OMRON distributor for warranty service. Keep your original invoice/cash ticket for warranty claims.
Disposal of This Product
Do not dispose of this product with household waste. Separate it from other wastes and recycle it responsibly to promote sustainable reuse of material resources. Contact your local government office or retailer for details on environmentally safe recycling.
Trademarks
OMRON, the OMRON logo, and OMRON connect are trademarks or registered trademarks of OMRON Corporation.
Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc.
Apple and the Apple logo are trademarks of Apple Inc.
Android, Google Play logo are trademarks of Google Inc.
EMC Information
This product conforms to EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard. Further documentation is available at www.omron-healthcare.com.
This product operates in the 2.4 GHz band. Avoid using it near other wireless devices operating on the same frequency to prevent interference.