Abstract
Analysis of small particles such as exosomes can be extremely challenging and frustrating. Electronic noise, voltage adjustment, sheath purification, and population separation are all aspects of exosome detection by flow cytometry that pose significant challenges. However, utilizing the exceptional range of the ZE5 Cell Analyzer alleviates these challenges. This document compares flow cytometry-based exosome detection methods using the ZE5 and demonstrates that exosomes can be directly detected without being attached to beads. It also highlights further advantages of the ZE5 that lead to experimental reliance.
Introduction
Exosomes are a type of secreted membrane vesicle of approximately 50–100 nm in diameter. They are mainly involved in intracellular communication but also have key roles in several processes such as metastasis and antigen presentation. Since exosomes are secreted in all major bodily fluids, including blood and urine, and carry important cellular information, they have recently emerged as a novel source of potential biomarkers. However, because of their small size in comparison to cells, they can be difficult to characterize and study.
Traditional exosome analysis by flow cytometry requires manual hardware adjustments, advanced instrument calibration, hours of sheath purification, and data manipulation. The ZE5 Cell Analyzer eliminates these hurdles of current exosome research by creating a streamlined process for easier analysis (as depicted in Figure 1). By using the built-in capabilities of the ZE5, any lab can study exosomes and other small particles. Regardless of the staining method, whether for surface or internal markers, the ZE5 displays optimal sensitivity for small particle identification.
This document shows that surface exosome marker proteins such as CD63 and CD81, as well as intravesicular markers ALIX and TSG101, can easily be detected using the ZE5 Cell Analyzer. The ZE5 has many unique features that, in addition to studying large cancer cells, allow it to be used in small particle analysis.
Figure 1: Exosome detection workflow with the ZE5 Cell Analyzer
This figure illustrates a workflow for exosome detection. The steps are: 1. Culture your cells, 2. Stain with markers, 3. Analyze with the ZE5, and 4. Visualize with data analysis software.
Methods and Results
Sample Preparation
MCF-7 cells were cultured in minimum essential media (MEM) supplemented with 10% FBS and 0.01 mg/ml insulin until they were 70-80% confluent. The cells were washed twice with PBS and incubated with exosome-free media for 12-72 hr. The media was collected and filtered through a 0.22 µm membrane to remove cells and debris. The filtered media was mixed thoroughly with 0.5 volumes of Total Exosome Isolation Reagent (Life Technologies). The mixture was incubated at 4°C overnight followed by centrifugation at 10,000 x g for 1 hr at 4°C. The supernatant was aspirated, and the pellet was washed with PBS filtered through a 0.1 µm membrane. The protein concentration and size of the exosomes were determined by the NanoDrop Spectrophotometer (Thermo Scientific Inc.) and Zetasizer Nano ZSP Dedicated Zeta Analyzer (Malvern Pananalytical) as 185-300 nm.
Instrument Setup
ProFlow Sort Grade Water (Bio-Rad Laboratories) was used to analyze samples on the ZE5 Cell Analyzer (Bio-Rad Laboratories). QC was performed using ZE Series QC Beads (Bio-Rad Laboratories). A mixture of yellow fluorescent size beads ranging from 0.22–1.35 µm (Spherotech) was used to set up instrument voltages and thresholds (as shown in Figure 2). For small particle setup, single or dual thresholds can be used to set up the instrument. For direct exosome detection of surface markers, forward side scatter (FSC) from the 405 nm laser and fluorescent signal were used as thresholds. FSC from the 405 nm laser was used to set up the threshold for intravesicular marker and bead-based exosome detection. Figure 2 illustrates the ZE5 Cell Analyzer setup for exosome analysis.Figure 2: Instrument setup
Non-Bead-Based Surface Staining of Exosomes
All buffers used for exosome detection were filtered through a 0.1 µm membrane. A titrated amount of fluorochrome-labeled antibody was used for surface and intravesicular staining. For exosome surface staining, 2 µg of exosomes were added to 50 µl of PEB buffer (PBS, 5 mM EDTA, and 0.5% BSA) and incubated with FITC anti-human CD63 (Bio-Rad Laboratories) and APC anti-human CD81 (BioLegend) for 30 min at room temperature. After incubation, the samples were washed with PEB and re-suspended in 500 µl of PEB. Exosome samples were filtered through a 10 µm filter prior to loading on the instrument. As shown in Figure 3, exosome surface markers were easily detected compared to an unstained control.
Figure 3: Purified exosomes
Figure 3 presents flow cytometry scatter plots demonstrating the detection of purified exosomes and their surface markers.
- A: Shows the gating strategy using FSC from the 405 nm laser with the SPD and SSC from the 488 nm laser on an area log scale to identify exosomes.
- B: Displays an unstained control used to set the quadrant gating for CD63-FITC vs. CD81-APC.
- C: Shows a single-color control for CD63-FITC staining.
- D: Shows a single-color control for CD81-APC staining.
- E: Presents a dual-stained sample showing both CD63 and CD81 expression on purified exosomes.
Intravesicular Staining
The Intracellular Fixation and Permeabilization Buffer Set (eBioscience) was used to fix and permeabilize exosomes for intravesicular staining. Exosomes (2 µg) were fixed in 100 µl of IC Fixation Buffer and incubated in the dark at room temperature for 20 min. After incubation, the samples were washed twice with 1 ml of permeabilization buffer and then resuspended in 100 µl of the same buffer. The samples were then incubated with PE anti-human TSG101 (Abcam) and anti-human ALIX conjugated to 494/516 excitation/emission labeling dye, using the ReadiLink Antibody Labeling Kit (Bio-Rad Laboratories), for 30 min at room temperature. After incubation, the samples were washed twice with 1 ml of permeabilization buffer and resuspended in 500 µl of PEB buffer (as shown in Figure 4). Figure 4 displays histograms for the detection of intravesicular exosome marker proteins ALIX and TSG101.Figure 4: Exosome intravesicular detection
Traditional Bead-Based Exosome Detection
Anti-human CD63 antibody (10 µg, Bio-Rad Laboratories) was incubated with 100 µl of SureBeads Protein G Magnetic Beads (Bio-Rad Laboratories) with rotation for 10 min at room temperature. Coated SureBeads were then incubated with 10 µg of exosomes for 1 hr at room temperature. Exosome-coated SureBeads were washed three times with PBS containing 0.1% Tween 20, and resuspended in 250 µl of PEB-0.1% Tween 20 (PBS, 5 mM EDTA, 0.5% BSA, and 0.1% Tween 20). Exosome-coated SureBeads (50 µl) were incubated with FITC anti-human CD63 (Bio-Rad Laboratories) and APC anti-human CD81 (BioLegend) for 30 min at room temperature and washed twice with PEB-0.1% Tween 20. Figure 5 shows successful characterization of exosome surface markers using bead-based flow cytometry on the ZE5 Cell Analyzer. Figure 5 illustrates the characterization of exosome surface markers using bead-based flow cytometry on the ZE5 Cell Analyzer.Figure 5: Bead-based exosome capture
Conclusions
- The direct detection of exosomes is possible on the ZE5 Cell Analyzer.
- Traditional bead detection methods for flow cytometry can also be performed on the ZE5.
- Distinct intravesicular exosome marker proteins such as ALIX and TSG101 can be quantified on the ZE5.
- The exosome workflow is made simpler by eliminating preparation steps on the ZE5.
Advantages of Using the ZE5 Cell Analyzer
Exosome research can be challenging, from sheath purification to gating in data analysis. To address these challenges in an easy-to-use system, the ZE5 was designed with the input of flow cytometry researchers. As a result, it contains built-in capabilities that allow any lab to study exosomes and other small particles. The ZE5 is equipped with extremely low-noise electronics and offers a small particle PMT detector using a 100 mW 405 nm laser. Together, this aspect, coupled with dual parameter thresholding, solves the problem of distinguishing exosomes or other small particles from debris and noise. Furthermore, the ZE5 uses DI water as its sheath and has an inline sheath filter to eliminate the need to pre-filter. This, together with filtering of samples, leaves few other cellular particulates to populate the plots. This boosts confidence in gating on the exosome population. Although Bio-Rad offers bead conjugation methods, the ZE5 Cell Analyzer is able to detect exosomes without beads. Using size beads, the ZE5 can quickly be set up to easily collect purified exosomes.
Visit bio-rad.com/exosome-appnote for more information.
Bio-Rad is a trademark of Bio-Rad Laboratories, Inc. in certain jurisdictions. All trademarks used herein are the property of their respective owners.