Programmable AC/DC Power Source
ASR-2000 Series Quick Start Guide
EFNR

SAFETY INSTRUCTIONS
This section contains the basic safety symbols that may appear on the accompanying User Manual CD or on the instrument. For detailed safety instructions and precautions, please see the Safety Instructions chapter in the user manual CD.
Safety Symbols

These safety symbols may appear in the user manual or on the instrument.

Warning: Identifies conditions or practices that could result Warning
in injury or loss of life.

Caution

Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
DANGER High Voltage

Attention Refer to the Manual

Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.

2

Power Cord for the United Kingdom When using the instrument in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons.

WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT:

The wires in this lead are coloured in accordance with the following code:

Green/ Yellow: Earth

Blue:

Neutral

Brown:

Live (Phase)

As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows: The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol or coloured Green/Green & Yellow. The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black. The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red. If in doubt, consult the instructions provided with the equipment or contact the supplier. This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used. Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.

3

Main Features

Performance · Maximum AC output voltage is 350 Vrms

· Maximum DC output voltage is 500 Vdc

· Maximum output frequency is 999.9 Hz in AC mode · Supported AC+DC waveform application

· DC full capacity output ability

· Output voltage total harmonic distortion is less than 0.5% at all frequency.

· Crest factor reached 4 times high

Features

· Include sine, square, triangle, arbitrary and DC output waveforms

· Variable voltage, frequency and current limiter

· Harmonic voltage and current analysis ability

· Excellent and feature-rich measurement capacity · Sequence and simulate function

· External input amplification

· AC line synchronized output · Preset memory function

· USB memory support

· Remote sense

· OCP, OPP and OTP protection function

4

Interface

· Built-in LAN, USB host and USB device interface · External control I/O · External signal input · Factory option RS232 and GPIB interface

5

Appearance Front Panel Overview
21

20 19 18 17 16 15

14

13

12

6 8 10

22

1

2 3 4 5 7 9 11

Description 1. Power switch button 3. Hardcopy key 5. Menu key/On phase key 7. Shift key 9. Test key/Output waveform key 11. Enter key 13. Scroll wheel 15. Display mode select key 17. Irms/IPK-Limit button 19. V/V-Limit button 21. Air inlet

2. Lock/Unlock button 4. USB interface connector (A Type) 6. Preset key/Local mode key 8. Cancel key/ALM CLR key 10. Output key 12. Arrow keys 14. Function keys (blue zone) 16. Range key/Output mode key 18. F/F-Limit button 20. LCD screen 22. Output socket

6

Rear Panel Overview

6 5

4

3

2

7

81

Description

1. Line input

2. Remote sensing input terminal

3. Exhaust fan

4. External I/O connector

5. USB interface connector (B Type) 6. Ethernet (LAN) connector

7. External signal input/ External 8. Output terminal

synchronized signal input

7

Power up
1. Connect the power cord to the socket on the rear panel. 2. Turn on the power switch on the front panel.

Note

The power supply takes around 15 seconds to fully turn on and shutdown.
Do not turn the power on and off quickly.

How to Use the Instrument

Background

The ASR-2000 AC power supplies generally use the scroll wheel, Arrow keys and Enter keys to edit numerical values or to select menu options.

Menu navigation is performed using the menu keys and function keys on the front panel.

Selecting Menu 1. Turn the scroll wheel to select parameters in

Items

menus and lists. The selected parameter will

be highlighted in orange. The scroll wheel is

also used to increment/decrement setting

values.

2. Press the Enter key to edit the parameter or to enter the selected menu.

Example

The following is an example of the menu list that appears when the Menu key is pressed.

Selected parameter

Using the Arrow Use the Arrow keys to select a digit power and then use

Keys and Scroll the scroll wheel to edit the value by that power.

Wheel

1. Use the Arrow keys to move the cursor to

the digit of the desired value.

2. Turn the scroll wheel to edit the value by the resolution of the selected digit.

Note

Cursor
3. Press the Enter key to edit the parameter or to enter the selected menu. Repeat the steps above for all the relevant digits.
4. Press the Enter key to confirm the edit.
By default the cursor starts at the lowest digit of value.

9

SPECIFICATIONS

The specifications apply when the ASR-2000 series is powered on for more
than 30 minutes.

Electrical specifications

Model

ASR-2050 ASR-2050R

ASR-2100 ASR-2100R

Input ratings (AC rms)

Nominal input voltage

100 Vac to 240 Vac

Input voltage range

90 Vac to 264 Vac

Phase

Single phase, Two-wire

Nominal input Frequency

50 Hz to 60 Hz

Input frequency range

47 Hz to 63 Hz

Max. power consumption

800 VA or less

1500 VA or less

Power factor*1

100Vac 0.95 (typ.)

200Vac 0.90 (typ.)

Max. input current 100Vac 8 A

15 A

200Vac 4 A

7.5 A

*1 For an output voltage of 100 V/200 V (100V / 200V range), maximum

current, and a load power factor of 1.

Model

ASR-2050 ASR-2050R

ASR-2100 ASR-2100R

AC mode output ratings (AC rms) Setting Range*1 0.0 V to 175.0 V / 0.0 V to 350.0 V

Voltage

Setting Resolution

0.1 V

10

Accuracy*2

±(0.5 % of set + 0.6 V / 1.2 V)

Output phase

Single phase, Two-wire

Maximum current*3

100 V 200 V

5 A 2.5 A

10 A 5 A

Maximum 100 V peak current*4 200 V

20 A 10 A

40 A 20 A

Load power factor

0 to 1 (leading phase or lagging phase)

Power capacity

500 VA

1000 VA

Frequency

Setting range

AC Mode: 40.00 Hz to 999.9 Hz, AC+DC Mode: 1.00 Hz to 999.9 Hz

Setting resolution

0.01 Hz (1.00 to 99.99 Hz), 0.1 Hz (100.0 to 999.9 Hz)

Accuracy Stability*5

For 45 Hz to 65 Hz: 0.01% of set For 1 Hz to 999.9 Hz: 0.02% of set ± 0.005%

Output on/off phase

0.0° to 359.9° variable (setting resolution

0.1°)

DC offset*6

Within ± 20 mV (TYP)

*1 100 V / 200 V range

*2 For an output voltage of 17.5 V to 175 V / 35 V to 350 V, sine wave, an

output frequency of 45 Hz to 65 Hz, no load, DC voltage setting 0V

(AC+DC mode) and 23°C ± 5°C

*3 For an output voltage of 1 V to 100 V / 2 V to 200 V. Limited by the

power capacity when the output voltage is 100 V to 175 V / 200 V to 350

V. If there is the DC superimposition, the current of AC+DC mode

satisfies the maximum current. In the case of lower than 40 Hz, and the

ambient temperature is 40°C or higher, the maximum current will be

decrease.

11

*4 With respect to the capacitor-input rectifying load. Limited by the maximum current.
*5 For 45 Hz to 65 Hz, the rated output voltage, no load and the resistance load for the maximum current, and the operating temperature.
*6 In the case of the AC mode and output voltage setting to 0 V.

Model

ASR-2050 ASR-2050R

ASR-2100 ASR-2100R

Output rating for DC mode

Voltage

Setting Range*1

-250.0 V to +250.0 V / -500.0 V to +500.0 V

Setting Resolution 0.1 V

Accuracy*2

±(|0.5 % of set| + 0.6 V / 1.2 V)

Maximum current*3

100 V 200 V

5 A 2.5 A

10 A 5 A

Maximum peak 100 V

current*4

200 V

20 A 10 A

40 A 20 A

Power capacity

500 W

1000 W

*1 100 V / 200 V range

*2 For an output voltage of -250 V to -25 V, +25 V to +250 V / -500 V to -50

V, +50 V to +500 V, no load, AC voltage setting 0V (AC+DC mode) and

23°C ± 5°C

*3 For an output voltage of 1.4 V to 100 V / 2.8 V to 200 V. Limited by the

power capacity when the output voltage is 100 V to 250 V / 200 V to 500

V.

*4 Within 5 ms, Limited by the maximum current.

ASR-2050

ASR-2100

Model

ASR-2050R

ASR-2100R

Output voltage stability

Line regulation*1

±0.2% or less

12

Load regulation*2

±0.15% @45 - 65Hz

±0.5% @DC, all other frequencies

(0 to 100%, via output terminal)

Ripple noise*3

0.7 Vrms / 1.4 Vrms (TYP)

*1 Power source input voltage is 100 V, 120 V, or 230 V, no load, rated output.
*2 For an output voltage of 75 V to 175 V / 150 V to 350 V, a load power factor of 1, stepwise change from an output current of 0 A to maximum current (or its reverse), using the output terminal on the rear panel.
*3 For 5 Hz to 1 MHz components in DC mode using the output terminal on the rear panel.

ASR-2050

ASR-2100

Model

ASR-2050R

ASR-2100R

Output voltage waveform distortion ratio, Output voltage response time, Efficiency

Total harmonic distortion (THD) *1 <0.2 % @50/60 Hz

<0.3 % @<500 Hz

<0.5 % @500.1 Hz to 999.9 Hz

Output voltage response time*2 Efficiency*3

100 s (TYP) 70 % or more

*1 At an output voltage of 50 V to 175 V / 100 V to 350 V, a load power factor of 1, and in AC and AC+DC mode.
*2 For an output voltage of 100 V / 200 V, a load power factor of 1, with respect to stepwise change from an output current of 0 A to the maximum current (or its reverse). 10% ~ 90% of output voltage
*3 For AC mode, at an output voltage of 100 V / 200 V, maximum current, and load power factor of 1 and sine wave only.

13

ASR-2050

ASR-2100

Model

ASR-2050R

ASR-2100R

Measured value display

Note: All accuracy of the measurement function is indicated for 23 °C±5 °C.

Voltage

RMS, AVG value*1

Resolution Accuracy*2

0.1 V
For 45 Hz to 65 Hz and DC: ±(0.5 % of reading + 0.3 V / 0.6 V) For 40 Hz to 999.9 Hz: ±(0.7 % of reading + 0.9 V / 1.8 V)

PEAK value Resolution 0.1 V

Accuracy For 45 Hz to 65 Hz and DC: ±(|2 % of reading| + 1 V / 2 V)

Current

RMS, AVG value

Resolution 0.01 A

Accuracy*3 For 45 Hz to 65 Hz For 45 Hz to 65 Hz

and DC:

and DC:

±(0.5 % of reading ±(0.5 % of reading + 0.02 A / 0.02 A) + 0.04 A / 0.02 A)

For 40 Hz to 999.9 For 40 Hz to 999.9

Hz:

Hz:

±(0.7 % of reading ±(0.7 % of reading + 0.04 A / 0.04 A) + 0.08 A / 0.04 A)

PEAK value Resolution Accuracy*4

0.01 A

For 45 Hz to 65 Hz For 45 Hz to 65 Hz

and DC:

and DC:

±(|2 % of reading| ±(|2 % of reading|

+ 0.2 A / 0.1 A) + 0.2 A / 0.1 A)

Power

Active (W)

Resolution 0.1 / 1 W

Accuracy*5 ±(2 % of reading + ±(2 % of reading +

0.5 W)

1 W)

14

Apparent (VA)

Resolution 0.1 / 1 VA

Accuracy*5*6 ±(2 % of reading + ±(2 % of reading +

0.5 VA)

1 VA)

Reactive (VAR)

Resolution 0.1 / 1 VAR

Accuracy*5*7 ±(2 % of reading + ±(2 % of reading +

0.5 VAR)

1 VAR)

Load power factor Range

0.000 to 1.000

Resolution 0.001

Load crest factor

Range

0.00 to 50.00

Resolution 0.01

Harmonic voltage

Range

Effective value (rms)

Percent (%)

Full Scale

(AC-INT and 50/60 Hz Resolution

only)

Accuracy*8

Up to 100th order of the fundamental wave
175 V / 350 V, 100%
0.1 V, 0.1%
Up to 20th ±(0.2 % of reading + 0.5 V / 1 V)

20th to 100th ± (0.3 % of reading + 0.5 V / 1 V)

Harmonic current

Range

Effective value (rms)

Up to 100th order of the fundamental wave

Percent (%)

Full Scale

(AC-INT and 50/60 Hz Resolution

only)

Accuracy*3

5 A / 2.5 A, 100% 0.01 A, 0.1% Up to 20th

10 A / 5 A, 100% Up to 20th

±(1 % of reading + ±(1 % of reading + 0.1 A / 0.05 A) 0.2 A / 0.1 A)

20th to 100th

20th to 100th

±(1.5 % of reading ±(1.5 % of reading + 0.1 A / 0.05 A) + 0.2 A / 0.1 A)

15

*1 The voltage display is set to RMS in AC/AC+DC mode and AVG in DC mode.
*2 AC mode: For an output voltage of 17.5 V to 175 V / 35 V to 350 V and 23 °C ± 5 °C. DC mode: For an output voltage of 25 V to 250 V / 50 V to 500 V and 23 °C ± 5.
*3 An output current in the range of 5 % to 100 % of the maximum current, and 23 °C ± 5 °C.
*4 An output current in the range of 5 % to 100 % of the maximum peak current in AC mode, an output current in the range of 5 % to 100 % of the maximum instantaneous current in DC mode, and 23 °C ± 5 °C. The accuracy of the peak value is for a waveform of DC or sine wave
*5 For an output voltage of 50 V or greater, an output current in the range of 10 % to 100 % of the maximum current, DC or an output frequency of 45 Hz to 65 Hz, and 23 °C ± 5 °C.
*6 The apparent and reactive powers are not displayed in the DC mode. *7 The reactive power is for the load with the power factor 0.5 or lower. *8 An output voltage in the range of 17.5 V to 175 V / 35 V to 350 V and 23
°C ± 5 °C.

Note

 Product specifications are subject to change without notice.
 The spec aforementioned applies to when slew rate mode is the Time mode.

General Specifications Interface Standard USB
LAN
EXT Control

Type A: Host, Type B: Device, Speed: 1.1/2.0, USB-CDC
MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask
External Signal Input External Control I/O

16

Optional 1 GPIB

SCPI-1993, IEEE 488.2 compliant interface

RS-232C Complies with the EIA-RS-232 specifications

Insulation Between input and resistance chassis, output and
chassis, input and output

500 Vdc, 30 M or more

Withstand Between input and voltage chassis, output and
chassis, input and output

1500 Vac, 1 minute

EMC

EN 61326-1 (Class A) EN 61326-2-1/-2-2 (Class A) EN 61000-3-2 (Class A, Group 1) EN 61000-3-3 (Class A, Group 1) EN 61000-4-2/ -4-3/ -4-4/ -4-5/ -4-6/ -4-8/-4-11 (Class A, Group 1) EN 55011 (Class A, Group1)

Safety

EN 61010-1

Environment Operating environment

Indoor use, Overvoltage Category II

Operating temperature range

0 °C to 40 °C

Storage temperature range

-10 °C to 70 °C

Operating

20 % RH to 80 % RH (no condensation)

humidity range

17

Storage humidity 90 % RH or less (no condensation) range

Altitude

Up to 2000 m

Dimensions (mm)

ASR-2000

ASR-2000R

285(W)×124(H)×480( 213(W)×124(H)×4

D) (not including

80(D) (not

protrusions)

including

protrusions)

Weight

ASR-2000

ASR-2000R

Approx. 11.5 kg

Approx. 10.5 kg

Others

Protections

OCP, OTP, OPP, FAN Fail

Display

TFT-LCD, 4.3 inch

Memory Function

Store and recall settings, Basic settings: 10

Arbitrary Wave

16 (nonvolatile)

4096 words

A value with the accuracy is the guaranteed value of the specification. However, an accuracy noted as reference value shows the supplemental data for reference when the product is used, and is not under the guarantee. A value without the accuracy is the nominal value or representative value (shown as type).

External Signal Input (AC+DC-EXT, AC-EXT Mode)

Specification

Gain setting range 100 V range: 0.0 to 250.0 times

200 V range: 0.0 to 500.0 times

Input terminal

BNC connector

Input impedance

1 M

Input voltage range ±2.5 V (A/D resolution 12 bit)

Factory Default 100 200

18

Nondestructive

±10 V

maximum input voltage

Gain resolution

0.1 times

Accuracy

±5 % (DC, or 45Hz ~ 65 Hz, gain is at initial value, with rate voltage output, no load)

EXT: Output voltage (V) = External signal input (V) x Gain (V/V)

Voltage Setting Signal Input (AC-VCA Mode)

Specification

Factory Default

Gain setting range 100 V range: 0.0 to 250.0 times 100

200 V range: 0.0 to 500.0 times 200

Input terminal

BNC connector

Input impedance

1 M

Input voltage range DC 0 ~ 2.5 V

Nondestructive

±10 V

maximum input voltage

Accuracy

±5 %

External Signal Input (AC+DC-ADD, AC-ADD Mode)

Specification

Factory Default

Gain setting range 100 V range: 0.0 to 250.0 times 100

200 V range: 0.0 to 500.0 times 200

Input terminal

BNC connector

Input impedance

1 M

Input voltage range ±2.5 V (A/D resolution 12 bit)

Nondestructive

±10 V

maximum input voltage

Input frequency range DC to 999.9 Hz (sine wave) DC to 100 Hz (other than sine wave)

19

Gain resolution

0.1 times

Accuracy

±5 % (DC, or 45Hz ~ 65 Hz, gain is at initial value, with rate voltage output, no load)

ADD: Output voltage (V) = External signal input (V) x Gain (V/V) + Internal signal source setting (V)

External Synchronous Signal or Line (AC+DC-SYNC, AC-SYNC)

Specification

Factory Default

Synchronization signal External synchronization signal

source

(EXT) or Power input (LINE)

LINE

Synchronization frequency range

40.00 Hz to 999.9 Hz

Input terminal

BNC connector

Input impedance

1 M

Threshold of input voltage

TTL level

Minimum pulse width 500 us

Nondestructive

±10 V

maximum input voltage

Resolution

0.01 / 0.1 Hz

Accuracy

±0.2 Hz

20

EC Declaration of Conformity

We declare that the below mentioned product

ASR-2050, ASR-2100, ASR-2050R, ASR-2100R

satisfies all the technical relations application to the product within the

scope of council:

Directive: 2014/30/EU; 2014/35/EU; 2011/65/EU; 2012/19/EU. The above

product is in conformity with the following standards or other normative

documents:  EMC 

EN 61326-1 : EN 61326-2-1: EN 61326-2-2:

Electrical equipment for measurement, control and laboratory use ­­ EMC requirements (2013)

Conducted & Radiated Emission Electrical Fast Transients

EN 55011: 2016+A1: 2017 Class A EN 61000-4-4: 2012

Current Harmonics

Surge Immunity

EN 61000-3-2: 2014

EN 61000-4-5: 2014+A1: 2017

Voltage Fluctuations

Conducted Susceptibility

EN 61000-3-3: 2013

EN 61000-4-6: 2014

Electrostatic Discharge

Power Frequency Magnetic Field

EN 61000-4-2: 2009

EN 61000-4-8: 2010

Radiated Immunity

Voltage Dip/ Interruption

EN 61000-4-3: 2006+A2: 2010

EN 61000-4-11: 2004+A1: 2017

 Safety

Low Voltage Equipment Directive 2014/35/EU

Safety Requirements

EN 61010-1: 2010

21

