How to change pre-filters and HEPA 14 filters

Cleaning a pre-filter

The pre-filter can be rinsed with clean water but must be completely dry before being used again. Any remaining moisture could potentially cause damage to the HEPA 14 filter.

Cleaning and changing times

Please refer to the next 7 steps on how to clean the Rediair pre-filters, air outlet and particle sensor.

NOTE: Before first use, the Rediair filters must be removed from their protective packaging, making sure the retaining straps point outwards before installation.

WARNING: Before dismantling Rediair to perform filter maintenance/replacement, perform hand hygiene. Follow local guidance on the use of appropriate PPE.

WARNING: Always unplug the Rediair before dismantling the device to perform filter maintenance/replacement.

Figure 1: Filter indicator on the LCD touch control panel flashes red repeatedly, indicating it is time to clean the pre-filters.

[Filter Indicator Flashing Red]

Refer to the main Rediair product manual for more information.

Scan to find more resources on our Rediair webpage.

Steps for Cleaning and Maintenance

  1. Remove the external covers (both sides) and the pre-filters (both sides).

    Diagram showing removal of external covers and pre-filters.

    [External Covers and Pre-filters Removed]

  2. The pre-filters can be rinsed with clean water. WARNING: The pre-filters MUST be completely dry before use. Any remaining moisture could damage the HEPA 14 filter.

    Rinsing pre-filters under a tap.

    [Rinsing Pre-filters]

  3. To clean Rediair's air outlet, lift the grill from the recesses indicated by the arrow symbols and clean the area with a dust cloth.

    Cleaning the air outlet grill.

    [Cleaning Air Outlet Grill]

  4. Re-install the external covers and the pre-filters (both sides). Note: Ensure external covers are fitted correctly or the Rediair will not power-on.

    Re-installing external covers and pre-filters.

    [Re-installing Covers and Pre-filters]

  5. To clean the particle sensor (back of Rediair), lift the flap and carefully clean with a cotton swab. This should be done every 2 months or if the LED air indicator consistently shows poor air quality.

    Cleaning the particle sensor.

    [Cleaning Particle Sensor]

  6. To re-set the pre-filter timer, hold the filter button for 5 seconds until it beeps.

    Resetting the pre-filter timer.

    [Resetting Filter Timer]

Filters: Changing and Disposal

Used Rediair filters can be disposed of following local waste guidance policies. If the device has been used in polluted air situations, the filter must be packed into an airtight bag and disposed of per local waste guidance policies. In the case of contamination with viruses, please consult local guidance for the appropriate PPE. Respiratory protection is recommended at a minimum of FFP2.

Please refer to the next 11 steps to dispose and install new Rediair main filters and pre-filters.

Filter indicator on the LCD touch control panel lights up continuously red OR flashes 5 times and then turns continuously red for 5-seconds, indicating it is time to change all filters.

[Filter Indicator Alert]

  1. Switch off device and disconnect the power supply.

    Switching off and disconnecting the power supply.

    [Power Disconnected]

  2. Remove external covers (both sides).

    Removing external covers.

    [Removing External Covers]

  3. Remove pre-filters and main filters, and dispose of them (both sides).

    Removing and disposing of old filters.

    [Removing Old Filters]

  4. Remove new main filters and pre-filters from box and remove the plastic wrap.

    Preparing new filters.

    [Preparing New Filters]

  5. Install the new main filters (both sides), making sure that the tabs are facing outwards.

    Installing new main filters.

    [Installing New Main Filters]

  6. Install the new pre-filters (both sides).

    Installing new pre-filters.

    [Installing New Pre-filters]

  7. Re-install the external covers (both sides). Note: Ensure external covers are fitted correctly or the Rediair will not power-on.

    Re-installing external covers.

    [Re-installing External Covers]

  8. Connect the power supply.

    Connecting the power supply.

    [Connecting Power Supply]

  9. Switch on the Rediair via the LCD touch control panel.

    Switching on the Rediair.

    [Switching On Rediair]

  10. To reset the timer after changing the filters, press and hold the filter indicator for 5 seconds until it beeps, at which point the filter light will stop flashing.

    Resetting the filter timer after replacement.

    [Resetting Filter Timer]

Models: MFB1143, MFB1143 Instant Air Filtration and Purification, Instant Air Filtration and Purification, Air Filtration and Purification, Filtration and Purification

File Info : application/pdf, 4 Pages, 580.43KB

PDF preview unavailable. Download the PDF instead.

Rediair-IFU-for-filter-change-3.0 JBN211164 01-08-2022 Web AW D1

References

Adobe PDF Library 17.0 Adobe InDesign 18.5 (Macintosh)

Related Documents

PreviewRediair Air Decontamination Device: Operation, Installation & Maintenance Manual
Comprehensive guide to operating, installing, and maintaining the Rediair air decontamination device by GAMA Healthcare. Covers setup, features, cleaning, filter replacement, and technical specifications.