ULTR Small ­ Ultrasonic Bath with Timer
Please read the User Manual carefully before use, and follow all operating and safety instructions!
user manual
english

User Manual
ULTR Small ­ Ultrasonic Bath with Timer
Preface
Users should read this Manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
Service
When help is needed, you can always contact the Service Department of manufacturer for technical support: www.labbox.com / e-mail: info@labbox.com
Please provide the customer care representative with the following information: · Serial number · Description of problem · Your contact information
Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 12 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation. For claim under the warranty please contact your supplier.
2

1. Safety Instructions
Warning! · Read the operating instructions carefully
before use. · Ensure that only trained staff works with the
instrument. · Do not work for long time when you use
flammable liquid like alcohol! · You will get a better cleaning effect if a
cleaning solution is added to the ultrasonic cleaner. · Do not operate if the tank is empty, or the ultrasonic generator will be damaged. · Do not turn on the machine if the tank is empty or the transducer will be damaged. · Do not relocate the machine when there is fluid in the tank (to avoid overflow). · Water-soluble liquids are suggested for the ultrasonic cleaner. Strong acids or flammable cleaners are strictly forbidden. If a solvent type cleaner is added in the tank, please stop working for a while after the 30 minute cycle. · Connect the device to an earthed power supply to ensure the safety of the machine and the experiment; connect the power as the machine requires it. · Refitting is strictly prohibited! · Ensure the earth wire is well connected before starting.
Table 1
· The control panel will be eroded by organic solution, strong acid and strong alkali. · Activate key of system should be on position of " ON/OFF " button. · Please close the lid to reduce noise and pay attention to water and steam in case of burning skin while open the lid.
Do not use the machine in any of the environments below: · A place where temperature changes fiercely · A place where vibration or impact is strong · A place where water oil or chemicals splash · A place where humidity is too high and it is easy to produce dew · A place where there is corrosive gas or dust · A place where there is explosive and flammable gas
3

2. Inspection
2.1 Receiving Inspection Unpack the equipment carefully and check for any damages which may have arisen during transport. Please contact your supplier for technical support.

2.2 Listing of Items

Note: If there is any apparent damage to the system, please do not plug it into the power line.

The package includes the following items:

2.3 Specifications:
Item Model Tank Capacity Transducer Input Voltage Ultrasonic Power Frequency Tank Size Unit Size

Parameter ULTR Small 0,8 L 1 unit AC 220-240V, 50-60 Hz 50 W 40 KHZ 150 x 85 x 65mm 178 x 110 x 130mm
4

3. Operation

Operation steps:

Check the machine Plug the machine to be sure the machine power is on, in this case the power indicator and LED display says "300"

Uncover the lid and fill the tank with tap water. No water operation will damage the machine, please put the cleaning object in the water. Fully immerse the part to be cleaned in the water, but do not put water above the "Max" Line.
Time Setting Press TIMER multiple times in order to choose the amount of time you wish the items to be cleaned. There are 18 options: 90s 120s 150s 180s 210s 240s 540s 720s 900s 18m 21m 24m 27m and 30m.

Reset Press RESET button when setting time or cleaning. Time will reset to "300" and machine will suspend work.

Finish Work Please do unplug the machine first after working, and then clean the tank, Be sure no liquid is in the tank if need long-term storage.
5

4. Maintenance

· Regular maintenance, electrical control system with high pressure air blow off the dust. · Clean the tank at least once a week

Trouble clearing

No. Problems

Possible Cause

Solutions

1 No ultrasonic A. Power supply not connected A. Check and plug power switch

activity

B. Fuse broken

B. Check the fitted power supply and change

C. Cable is short circuited

the same specification fuse

D. Transducer is short circuited C. Connect fitted cable or replace a new

E. PCB Board broken

D. Consult with your supplier

F. Other reasons

E. Check the broken part and change it

F. Consult with your supplier

2

Not well

A. Not strong ultrasonic

A. Connect ultrasonic button and adjust

cleaning

cleaning

B. Adjust liquid into the best surface

B. Too high or too low liquid C. Adjust the temperature into the most fitted

surface

D. Stop and switch off power supply, replace

C. Too high too low

suitable liquid after previous liquid

temperature

cooling down

D. Not suitable cleaning liquid E. Consult with your supplier

E. Other Reasons

3 Electric Leakage A. Customer side not grounded A. Ensure it is grounded

B. Machine not grounded

B. Check if machine earth wire loosens

4 Other problems

Consult with your supplier

6

. 7

www.labbox.com

