User Guide for Honeywell models including: ToxoRAE Pro LEL

ToxiRAE Pro LEL QuickStart Guide RevD

ToxiRAE Pro LEL, personal wireless monitor, wireless volatile organic compounds, VOC detection, remote wireless notification, Combustible Gases detector, continuous datalogging, industrial hygiene and safety, oil and gas, chemical plants, pharmaceutical plant, government health, safety, security, environmental agencies, hazardous materials teams, first responders, environmental applications, environmental consulting, ToxiRAEPro PID, detect 300 VOCs, RAE Systems by Honeywell

RAE Systems by Honeywell

ToxiRAE Pro LEL QuickStart Guide RevD - Honeywell

6 days ago — 5HDG %HIRUH 2SHUDWLQJ. The ToxiRAE Pro LEL User's Guide must be carefully read by all individuals who have or will have the responsibility of using, ...

ToxiRAE Pro LEL QuickStart Guide RevD

ToxiRAE Pro LEL | Honeywell

Not Your Device? Search For Manuals / Datasheets:

File Info : application/pdf, 2 Pages, 313.87KB

Document
Quick-Start-Guide ToxiRAE-Pro-LEL G02-4022-000 RevC ?download=false
7R[L5$(3UR/(/ 4XLFN6WDUW*XLGH 


RAE Systems Inc. 1349 Moffett Park Drive Sunnyvale, CA 94089 USA 3KRQH 408-952-8200 www.raesystems.com

:$51,1*6 
5HDG%HIRUH2SHUDWLQJ 
The ToxiRAE Pro LEL User's Guide must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer's instructions.
&$87,21 
Never operate the monitor when the cover is removed. Remove monitor top cap, rear cover, and battery only in an area known to be non-hazardous.

8VHU,QWHUIDFH 

The ToxiRAE Pro LEL's user interface consists of

the display, four LED lights, an alarm buzzer, and two

keys, [MODE] and [Y/+]. The LCD displays real-time

reading in specified measurement units, sensor type,

alarm type (when in alarm, including cal. or bump

overdue), Man Down status, battery status, datalog (if on),

and radio and connection quality (if available).


*DVLQOHW 

%HOWFOLS RQEDFN 

/('V 

/(' 

'LVSOD\ 

$ODUP  %X]]HU 

>02'(@NH\  

><@NH\ 


&KDUJLQJ7KH7R[L5$( 3UR/(/ 
Always fully charge the battery before use. Contacts on the bottom of the ToxiRAE Pro LEL meet the cradle's contact pins, transferring power. While charging, the LED on the cradle glows red. When the battery is fully charged, the LED glows green.

$

3ODFH 7R[L5$(
3UR/(/ LQ



FKDUJL QJFUDGOH 

5 DLVHGOLS  ORFNVLQWR  FUDGOHV ODWFK 

%  &RQQHFW$&  DGDSWHUWR FKDUJLQJ  FUDGOHDQG  SOXJLWLQWR$& 
SRZHUVRXUFH 

7XUQLQJ7KH7R[L5$( 3UR/(/ 2Q
Press and hold [MODE] for 3 seconds. During startup, the battery, buzzer, vibration alarm, and LEDs are tested, and then the ToxiRAE Pro LEL performs self-testing of its functions. When the main measurement screen appears, the ToxiRAE Pro LEL is ready for calibration or use.
Note: If the battery is almost fully discharged, the message "Battery too low! Needs charging. Powering off!" shows in the display and the ToxiRAE Pro LEL shuts off. You must charge the battery before you can use the instrument.

7XUQLQJ7KH7R[L5$( 3UR/(/ 2II
Press and hold [MODE]. A 5-second countdown to shutoff begins. You must continue pressing on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the ToxiRAE Pro LEL continues normal operation.
When the countdown beeps stop and you see "Unit Off," release your finger from the [MODE] key. The ToxiRAE Pro LEL is now off.

&KDUJLQJ 

/DWFK 

VWDWXV/(' 

UHOHDVH 







Calibration

The ToxiRAE Pro LEL can be automatically bump tested and calibrated using the AutoRAE 2 Test and Calibration Station (refer to its User's Guide for instructions). Manually calibrate using a fixed-flow regulator (flow rate between 0.5 and 1.0 liters per minute) and the supplied special calibration adapter that covers the gas inlet:

1. To begin calibration, connect the zero air or

calibration gas cylinder, flow regulator, and calibration

adapter to the ToxiRAE Pro LEL.

2. Enter Programming Mode by pressing and holding

[MODE] and [Y/+] simultaneously until the password

screen appears.

3. Input the 4-digit password. (The default password is

"0000." If you do not know the password, highlight

and select "OK.") Then follow the instructions for

zero calibration and span calibration.

Important! After a bump test or calibration, remove the

calibration adapter to ensure correct readings.

7.

Zero (Fresh Air) Calibration
The ToxiRAE Pro LEL should be zero-calibrated in clean air with 20.9% oxygen. If the ambient air has impurities, a cylinder of clean zero air should be used for zero calibration. With the ToxiRAE Pro LEL in Programming Mode and "Zero Calib" highlighted:
1. Press [Y/+]. "Apply zero gas..." is displayed. 2. Start the flow of the zero gas, if used, and press
[Y/+] to start the fresh air or zero calibration. 3. The screen says, "Zeroing..." and shows the
countdown. 4. When done, it says, "Zero is done! Reading = 0.0%
LEL" (this reading should be 0.0% LEL or very close to it). 5. "Span Calib" is highlighted. 6. Shut off the flow of zero air (if used). 7. Remove the calibration adapter.
8.

Manual Span Calibration

In Programming Mode, and with "Span Calib" highlighted:

1. Press [Y/+]. The screen displays the current (or

default) calibration gas and its concentration.

 Press [MODE] if you do not want to change it.

 Press [Y/+] if you do want to change it.

If you did not change it, you will see a message like this:

Cal. Gas:

Methane

Span=50% LEL

Apply gas...

2. Attach the calibration adapter, and connect the

calibration gas cylinder and flow regulator to the

ToxiRAE Pro LEL, and start the gas flow.

3. Press [Y/+] to start calibration.

4. Upon completion, "Span is done!" and the reading

are shown. Make sure it is within 10% of the span

gas value.

5. Remove the calibration adapter.

9.

Bump (Functional) Testing
RAE Systems recommends periodic bump testing to confirm that the sensor is working and that the alarms are functional.

Pressing [MODE] repeatedly allows you to step through the screens as shown here.

Menu Navigation In Normal Mode

With the ToxiRAE Pro LEL in Normal Mode:
1. Connect the calibration gas cylinder, flow regulator, and calibration adapter to the ToxiRAE Pro LEL and start the gas flow.
2. Make sure the unit goes into alarm and that the buzzer produces at least two beeps per second, the LED lights flash on and off, and the vibration alarm functions. The display backlight should illuminate and an alarm message should be shown in the display.
3. Turn off the gas flow. 4. Remove the calibration adapter.