BRTSys Modbus Salinity Sensor Adapter Datasheet
Version 1.0
1 Introduction
The Modbus Salinity Sensor Adapter is designed to work with a Salinity probe to form a complete Salinity sensor. It uses the electrical conductivity method for its measurement. This adapter has a BNC connector for attaching the Salinity probe. A 2-point calibration method is used to calibrate the adapter and probe and allows measurements of salinity between 1 ppt and 120ppt with a precision of 1ppt. These adapters and probes are suitable for use in applications such as aquaculture fishponds, shrimp ponds, sea water applications, nutrient tanks, water treatment plants, and swimming pools.
1.1 Features
- Salinity Sensor Adapter to integrate with cell constant K=10 Salinity probe with BNC connector
- Measures Salinity range of 1ppt to 120ppt with linearized output and 1ppt resolution
- 2 Point calibration
- Implements Modbus RTU protocol
- High report rate of every 5 seconds
- Low power consumption 5V-91mW
- Operating temperature range: 0°C to +70°C
- Flush mount and DIN Rail Mount options
Visit https://brtsys.com/resources/ for more information.
Certifications: CE, FCC, REACH, RoHS Compliant
2 Part Numbers / Ordering Information
Part Number | Description |
---|---|
MS-1301-01A | Modbus Salinity Sensor Adapter |
MA-0102-01A | Modbus RS485-RJ11 Cable (30cm) |
LA-1201-01A | LDSBus DIN Rail Mount Set |
3 Specifications
Features | Labels | Description |
---|---|---|
Interface | BNC (Connect to Salinity probe) RS485 Modbus RTU | |
LED Indicator (RGB) | System Status Indicator (Please refer to LED section) | |
Mounting | Flush Mount, DIN Rail Mount | |
Power | Modbus Voltage | 9-24V DC Bus Power |
Device Input Voltage | 5V DC Bus Power | |
Typical Power | 5V 91mW | |
Max. Power | 266mW | |
Salinity Sensor Input module | Detection Range | 1 - 120ppt |
Resolution | 0.01ppt | |
Response Time | <1 Minute | |
Calibration | 2 Point Calibration | |
Physical Characteristics | Color | White |
Housing | Polycarbonate | |
Dimensions | L117.6mm x W42.9mm x H29.7mm | |
Environmental Limits | Operating Temperature | 0 to 70°C |
Storage Temperature | -20 to 85°C | |
Ambient Relative Humidity | 5 to 95% (non-condensing) | |
Package Contents | Device | 1x Modbus Salinity Sensor Adapter |
Wire Assembly | 1X Modbus RS485-RJ11 Cable(30cm) | |
Optional | Mounting Accessories | 1x LDSBus DIN Rail Mount set |
4 FCC Compliance Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- These devices may not cause harmful interference, and
- These devices must accept any interference received, including interference that may cause undesired operation.
NOTE: The equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If the equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Re-orient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
To maintain compliance with FCC's RF exposure guidelines, at least 20cm of separation distance between the device and the user's body must be always maintained.
FCC Radiation Exposure Statement
This device complies with FCC radiation exposure limits set forth for an uncontrolled environment and it also complies with Part 15 of the FCC RF Rules. This equipment must be installed and operated in accordance with the instructions provided, and the antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-users and installers must be provided with antenna installation instructions and consider removing the no-collocation statement.
Caution
Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
The FCC logo is displayed.
5 Hardware Features
The Modbus Salinity Sensor Adapter features:
- BNC Connector: For connecting the Salinity Electrode Probe (not included).
- System Status LED Indicator: Provides visual feedback on the device's status.
- RJ12 Port: Modbus data and power interface port.
- Flush Mount Slot: Allows for direct mounting on a wall or flat surface.
- DIN Rail Plate: For mounting on a standard DIN rail using an optional mount set.
Function | Labels | Description |
---|---|---|
BNC Connector | Probe | Probe Interface |
Salinity Electrode Probe | Sensing probe | |
System Status LED Indicator | LED | Modbus status LED |
RJ12 Port | Modbus | Modbus data and power interface port. The physical port is RJ12. The connection interface can be RJ11/RJ12. |
6 Sensor Adapter Configuration and Installation
Please visit https://brtsys.com/resources/software/utility-tools to access the Modbus Configuration Utility guide on how to configure the device name, device address and termination settings before using it for your specific application.
6.1 Connection Diagram for Standard Modbus Power Supply
The diagram shows a Modbus network setup:
- Device 1 (Modbus Host Controller): Connected via Twisted Pair Cable.
- Device 2 (Modbus Device): Connected via RS485-RJ11 Cable or RS485-JST Cable.
- Device 3 (Modbus Device): Connected via RS485-RJ11 Cable or RS485-JST Cable.
- Connections include RS485 A, RS485 B, Ground, and 9-24VDC VIN.
Setup Instructions:
- Use a Cat5e/Cat6e RJ45 Twisted Pair Cable to connect the Modbus controller (Host) to the network for RS485 communication and power.
- Connect each Modbus device to the network using either an RS485-JST cable or an RS485-RJ11 cable, as provided with the device.
- BRTSys Modbus devices have built-in bus termination resistors. These resistors can be enabled or disabled by using the BRTSys Modbus Configuration Utility. When installing the device as the last device on the bus, these terminations may be used to terminate the bus.
6.2 RS485-RJ11 Cable(30cm)
The RS485-RJ11 cable has an RJ11 connector on one end and an RS485 Modbus interface on the other. The pin configuration is as follows:
PIN Legend | Function |
---|---|
VIN | Modbus Input Voltage 9-24VDC |
GND | Ground |
GND | Ground |
B | RS485-B |
A | RS485-A |
7 Mounting Instructions
7.1 Flush Mount
The device can be flush mounted directly on a wall or any flat surface using 2 M3.5*16mm (thread) screws.
A diagram shows the adapter with screw holes labeled '2xM3.5*16mm'.
7.2 DIN Rail Mount
The device can be mounted on a DIN Rail using the LDSBus DIN Rail Mount set. This set is optional and includes the bracket and mounting screws.
A diagram shows the adapter mounted on a DIN Rail.
8 Modbus Registers
The following table details the Modbus registers for the adapter:
Address | Parameter | Starting Address | Quantity of Registers | Supported Function Code | Parameter Range and Description | Default |
---|---|---|---|---|---|---|
RS485 Termination | 0000H | 1 | 0x03/0x10 | 0 - Termination OFF 1 - Termination ON | 126 OFF | |
Baud Rate | 0001H | 1 | 0x03/0x10 | 0 - 1200 bps 1 - 2400 bps 2 - 4800 bps 3 - 9600 bps 4 - 19200 bps 5 - 38400 bps 6 - 115200 bps | 9600 bps | |
Parity | 0002H | 1 | 0x03/0x10 | 0 - None 1 - Odd 2 - Even | Even | |
Status LED Enable | 0003H | 1 | 0x03/0x10 | 0 - LED OFF 1 - LED ON | LED ON | |
Measuring Temperature | 0004H | 1 | 0x03/0x10 | Measuring Temperature | 25°C | |
Sensor Calibration Version | 0005H | 1 | 0x03/0x10 | Calibration version | 0x0090 | |
Sensor Calibrated Date | 0006H | 2 | 0x03/0x10 | Calibration date YYYYMMDD | N/A | |
Sensor Calibrated Constant | 0007H | 1 | 0x03/0x10 | Constant value = Calibrated 1st buffer solution ADC value | 0x0020 | |
Sensor Calibrated EC (1st Point) | 0009H | 2 | 0x03/0x10 | Calibrated 1st buffer solution EC value | N/A | |
Sensor Calibrated ADC (2nd Point) | 000AH | 2 | 0x03/0x10 | Calibrated 2nd buffer solution ADC value | N/A | |
Sensor Calibrated EC (2nd Point) | 000CH | 2 | 0x03/0x10 | Calibrated 2nd buffer solution EC value | N/A | |
Sensor Calibrated Slope | 000EH | 2 | 0x03/0x10 | Calibrated slope | N/A | |
Sensor Calibrated Origin | 0010H | 2 | 0x03/0x10 | Calibrated origin | N/A | |
Sensor Calibrated Probe | 0012H | 2 | 0x03/0x10 | Calibrated probe | N/A | |
Sensor Calibrated Temperature Checksum | 0014H | 2 | 0x03/0x10 | Calibrated temperature XOR Checksum | N/A | |
Device UUID | 0016H | 1 | 0x03/0x10 | Calibrated XOR Checksum to 19H | N/A | |
Device Firmware Version | 0018H | 8 | 0x03 | MSxxxxxxxxxxxxyy where x is ASCII character and yy is 16-bit running number | N/A | |
Device Part Number | 0026H | 1 | 0x03 | XX - Not concerned M - Major N - Minor | N/A | |
Reserved | 002EH | 1 | 0x03 | Device ID | 0x800F | |
Salinity | 002FH | N/A | N/A | N/A | 1ppt to 120ppt (100 to 12000) | N/A |
Salinity ADC data | 0030H | 1 | 0x03 | 0 to 1023 | N/A | |
Reset | 0031H | 1 | 0x03 | Write 1 to reset | N/A | |
Reserved | 0032H | 1 | 0x06 | Write 1 to reset | N/A | |
Reserved | 0150H | 1 | 0x06 | Write 1 to reset | N/A | |
Identify | 0151H | 1 | 0x06 | Write 1 to start blinking the device @1Hz for 10 seconds | N/A |
9 Mechanical Dimensions
All dimensions are in millimetres.
Diagrams show the top, side, and bottom views of the adapter with dimensions:
- Top View: Length 117.6mm, Width 42.9mm.
- Side View: Height 29.7mm.
- Bottom View: Mounting hole spacing 60.0mm.
10 System Status LED Indicators
The system status LED indicates the device's operational state:
Device Status | LED Color | Flashing Frequency | Description |
---|---|---|---|
Termination ON | BLUE | Steady - Non- flashing | |
Termination OFF | GREEN | Steady - Non- flashing | |
Device Configuration Error | RED | Steady - Non- flashing | Device configuration error |
Communication | RED/GREEN/ BLUE/YELLOW | Blink twice (Short blink) | Device in communication |
Firmware update | YELLOW | Steady – Non- flashing | Device firmware update |
Note:
- For reliable communication, ensure that the power supply and the RS485 termination settings are correct.
- Ensure that the Modbus address and baud rate are configured correctly before deployment.
11 Probe Selection
The following specifications are recommended for selecting a Probe:
- Detection Range: 1ppt to 120ppt
- Cell Constant: K=10
- Connector: BNC
For more information on calibration, please refer to the Modbus Configuration Utility User Guide.
For information related to probes recommendation and selection criteria, please refer to LDSBus Probe Specifications.
12 Contact Information
Refer to https://brtsys.com/contact-us/ for contact information.
System and equipment manufacturers and designers are responsible to ensure that their systems, and any BRT Systems Pte Ltd (BRTSys) devices incorporated in their systems, meet all applicable safety, regulatory and system-level performance requirements. All application-related information in this document (including application descriptions, suggested BRTSys devices and other materials) is provided for reference only. While BRTSys has taken care to assure it is accurate, this information is subject to customer confirmation, and BRTSys disclaims all liability for system designs and for any applications assistance provided by BRTSys. Use of BRTSys devices in life support and/or safety applications is entirely at the user's risk, and the user agrees to defend, indemnify, and hold harmless BRTSys from any and all damages, claims, suits, or expense resulting from such use. This document is subject to change without notice. No freedom to use patents or other intellectual property rights is implied by the publication of this document. Neither the whole nor any part of the information contained in, or the product described in this document, may be adapted, or reproduced in any material or electronic form without the prior written consent of the copyright holder. BRT Systems Pte Ltd, 1 Tai Seng Avenue, Tower A, #03-01, Singapore 536464. Singapore Registered Company Number: 202220043R.
Appendix A – References
Document References
- Modbus Configuration Utility User Guide
- LDSBus Probe Specifications
Acronyms and Abbreviations
Terms | Description |
---|---|
DC | Direct Current |
LED | Light Emitting Diode |
PPT | Parts Per Thousand |
Appendix B – List of Figures and Tables
List of Figures
- Figure 1 - Modbus Salinity Sensor Adapter Hardware Features
- Figure 2 - Modbus Salinity Sensor Adapter - Connection Diagram
- Figure 3 - RS485-RJ11 Cable(30cm)
- Figure 4 - Modbus Salinity Sensor Adapter Flush Mount
- Figure 5 - Modbus Salinity Sensor Adapter DIN Rail Mount
- Figure 6 - Modbus Salinity Sensor Adapter Dimension – Top View
- Figure 7 - Modbus Salinity Sensor Adapter Dimension – Side View
- Figure 8 - Modbus Salinity Sensor Adapter Dimension – Bottom View
List of Tables
- Table 1 - Part Numbers / Ordering Information
- Table 2 - Modbus Salinity Sensor Adapter Specifications
- Table 3 - Modbus Salinity Sensor Adapter Hardware Features
- Table 4 - RS485-RJ11 Cable(30cm) Pin Configuration
- Table 5 - Modbus Registers
- Table 6 - System Status LED Indicators
Appendix C – Revision History
Document Title: Modbus Salinity Sensor Adapter Datasheet
Document Reference No.: BRTSYS_000187
Clearance No.: BRTSYS#108
Product Page: https://brtsys.com/product-category/adapters/
Document Feedback: Send Feedback
Revision | Changes | Date |
---|---|---|
Version 1.0 | Initial Release | 18-06-2025 |
Related Documents
![]() | BRTSys Modbus CO2 Multi-Sensor Datasheet Comprehensive datasheet for the BRTSys Modbus CO2 Multi-Sensor. Details features, specifications, part numbers, installation, Modbus registers, and dimensions for this environmental monitoring device. |
![]() | BRTSys Modbus Trailing Edge Light Dimmer Datasheet Datasheet for the BRTSys Modbus Trailing Edge Light Dimmer, featuring trailing edge dimming technology, Modbus communication, and adjustable brightness control for LED lighting. Includes technical specifications, installation guides, and hardware features. |
![]() | BRTSys LDSBus Device Quick Start Guide: Setup and Configuration Get started with your BRTSys LDSBus device. This guide provides step-by-step instructions for connecting the USB adapter, configuring the device, and accessing resources for setup and installation. |
![]() | BRTSys IoTPortal User Guide: Sensor-to-Cloud Connectivity Comprehensive user guide for the BRTSys IoTPortal, a versatile and scalable sensor-to-cloud connectivity platform. Learn about features, setup, and prerequisites for integrating IoT devices. |
![]() | BRTSys IoT Portal User Guide: Managing Sensor-to-Cloud Connectivity Comprehensive user guide for the BRTSys IoT Portal Web Application, detailing features for gateway management, data visualization, event creation, and organization administration. Designed for system integrators and technical users. |
![]() | Onboarding Guide for BRTSys IoTPortal Gateway Step-by-step instructions for setting up and onboarding your BRTSys IoTPortal Gateway using the mobile application and network connection. |
![]() | BRTSys IoT Portal Gateway Onboarding Guide A step-by-step guide to easily onboard your BRTSys IoT Portal Gateway using the dedicated mobile application. Learn how to power up, connect, and register your gateway. |
![]() | BRTSys PANL Desk Manager 3.1.0: Installation and Configuration Guide Install and configure BRTSys PANL Desk Manager 3.1.0 with this comprehensive user guide. Learn about server setup, network requirements, Exchange and Microsoft 365 integration, and authentication for efficient workspace management. |