FCC SAR

Arashi Vision Inc. CINSAAHA Camera 2AWWH-CINSAAHA 2AWWHCINSAAHA cinsaaha

PDF preview unavailable. Download the PDF instead.

GetApplicationAttachment.html?id=8352113
Correction1:May 30,2025

No.: GJWSZ2025-0188-SAR

TEST REPORT

NAME OF SAMPLE
CLIENT CLASSIFICATION OF TEST FCC ID
Max. SAR (1g):

: Camera : Arashi Vision Inc. : N/A : 2AWWH-CINSAAHA
: Body : 0.67 W/kg

CVC Testing Technology (Shenzhen) Co., Ltd.

CVC Testing Technology (Shenzhen) Co., Ltd.

Correction1:May 30,2025 Test Report No.: GJWSZ2025-0188-SAR Name : Arashi Vision Inc.

Page 2 of 33

Applicant

Address : 11th Floor, Building 2, Jinlitong Financial Center, Bao'an District, 518000 Shenzhen, Guangdong, China

Manufacturer

Name : Arashi Vision Inc.
Address : 11th Floor, Building 2, Jinlitong Financial Center, Bao'an District, 518000 Shenzhen, Guangdong, China

Name : Camera

Model/Type: CINSAAHA

Equipment Under Test Trade mark : Insta360

SerialNO.: GTS2024110147-6

Date of Receipt.

Sampe NO.:1-1 Mar.28,2025

Date of Testing

Apr.01,2025

Test Specification

Test Result

ANSI/IEEE Std. C95.1

FCC 47 CFR Part 2 (2.1093) IEEE 1528: 2013

Pass

Published RF exposure KDB procedures The equipment under test was found to comply with the

requirements of the standards applied.

Evaluation of Test Result

Seal of CVC

Date of issue:Apr.08,2025,Correction 1:May.30,2025

Compiled by:

Reviewed by:

Approved by:

Liang Jiatong

Name

Signature

Other Aspects: NONE.

Mo Xianbiao

Name

Signature

Dong Sanbi

Name

Signature

Abbreviations: Pass= passed

Fail = failed

N/A= not applicable

EUT= equipment, sample(s) under tested

This test report relates only to the EUT, and shall not be reproduced except in full, without written approval of CVC.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 3 of 33

TABLE OF CONTENTS
RELEASE CONTROL RECORD .........................................................................................................................................4
1 GENERAL INFORMATION...........................................................................................................................................5
1.1 GENERAL PRODUCT INFORMATION .......................................................................................................................... 5 1.2 TEST ENVIRONMENT...................................................................................................................................................... 6 1.3 TEST LOCATION............................................................................................................................................................. 6 1.4 TEST STANDARDS AND LIMITS .......................................................................................................................................... 7 1.5 STATEMENT OF COMPLIANCE ........................................................................................................................................ 8
2 SAR MEASUREMENT SYSTEM....................................................................................................................................9
2.1 DEFINITION OF SPECIFIC ABSORPTION RATE (SAR) ............................................................................................................... 9 2.2 SAR SYSTEM............................................................................................................................................................... 10 2.3 PROBE ....................................................................................................................................................................... 11 2.4 DATE ACQUISITION ELECTRONICS 4 (DAE4) ...................................................................................................................... 11 2.5 WRIST PHANTOM ........................................................................................................................................................ 14 2.6 DEVICE HOLDER........................................................................................................................................................... 14 2.7 SYSTEM VALIDATION DIPOLES ......................................................................................................................................... 15
3 TISSUE SIMULATING LIQUIDS ..................................................................................................................................16
3.1 SIMULATING LIQUIDS PARAMETER CHECK ......................................................................................................................... 16 3.2 LIQUIDS MEASUREMENT RESULTS ................................................................................................................................. 18
4 SAR SYSTEM VALIDATION........................................................................................................................................19
4.1 VALIDATION SYSTEM ..................................................................................................................................................... 19 4.2 SYSTEM VALIDATION RESULT ........................................................................................................................................ 20
5 SAR EVALUATION PROCEDURES ..............................................................................................................................21
6 SAR MEASUREMENT EVALUATION ..........................................................................................................................21
6.1 EUT CONFIGURATION AND SETTING ................................................................................................................................ 23 6.2 EUT TESTING POSITION ................................................................................................................................................ 25
7 MAXIMUM OUTPUT POWER...................................................................................................................................26
7.1 MAXIMUM CONDUCTED POWER..................................................................................................................................... 26 7.2 MEASURED CONDUCTED POWER RESULT....................................................................................................................... 26
8 SAR TESTING RESULTS .............................................................................................................................................27
8.1 SAR RESULTS FOR BODY EXPOSURE CONDITION (SEPARATION DISTANCE IS 0 CM GAP).............................................................. 27 8.2 SAR MEASUREMENT VARIABILITY ................................................................................................................................... 28
9 EQUITPMENT LIST ...................................................................................................................................................29
10 MEASUREMENT UNCERTAINTY .............................................................................................................................30
11 APPENDIXES..........................................................................................................................................................32
11.1 APPENDIX A: SAR PLOTS OF SYSTEM VERIFICATION .......................................................................................................... 32 11.2 APPENDIX B: SAR PLOTS OF SAR MEASUREMENT............................................................................................................ 32 11.3 APPENDIX C: CALIBRATION CERTIFICATE FOR PROBE AND DIPOLE ......................................................................................... 32 11.4 APPENDIX D: PHOTOGRAPHS OF EUT AND SETUP............................................................................................................. 32

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Correction1:May 30,2025 Test Report No.: GJWSZ2025-0188-SAR

Page 4 of 33

RELEASE CONTROL RECORD

ISSUE NO. GJWSZ2025-0188-SAR GJWSZ2025-0188-SAR

REASON FOR CHANGE Original release Modify the applicant's address and product name

DATE ISSUED Apr.08.2025 May 30.2025

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Correction1:May 30,2025 Test Report No.: GJWSZ2025-0188-SAR

Page 5 of 33

1 GENERAL INFORMATION 1.1 GENERAL PRODUCT INFORMATION

PRODUCT

Camera

BRAND

Insta360

MODEL

CINSAAHA

CINSAAHY (where Y would be any English letters or blank, different

ADDITIONAL MODEL

packing method, model designations on the marking plate for different

commercial purpose)

MODEL DIFFERENCES

Only model name is different

POWER SUPPLY MODULATION MODE
OPERATING FREQUENCY

Rechargeable Li-ion Battery 802.11b: DSSS WLAN 2.4G 802.11g/n (HT20): OFDM WLAN 5G 802.11a/n (HT20)/HT40 /ac(VHT80): OFDM Bluetooth GFSK, /4-DQPSK, 8-DPSK WLAN 2.4GHz Band: 2412 MHz ~ 2462 MHz WLAN 5.2GHz Band: 5180 MHz ~ 5240 MHz WLAN 5.8GHz Band: 5745 MHz ~ 5825 MHz

BATTERY

Bluetooth: 2402 MHz to 2480 MHz Rated Voltage:3.87V Charge Limit Voltage:4.50V Capacity: 2800mAh

ANTENNA TYPE

FPC Antenna

ANTENNA GAIN

2.4G:1.2dBi; 5.2G:2.6dBi; 5.8G:2.0dBi

Remark:

1. For more detailed features description, please refer to the manufacturer's specifications or the User's

Manual.

2. The EUT battery must be fully charged and checked periodically during the test to ascertain uniform

power

3. This is provided by the manufacturer. The laboratory is not responsible for technical data provided by the

customer

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 6 of 33

1.2 TEST Environment
Ambient conditions in the SAR laboratory: Items Temperature () Humidity (%RH)

Required 20.8 47

1.3 TEST Location
The tests and measurements refer to this report were performed by EMC testing Lab. of CVC Testing Technology (Shenzhen) Co., Ltd.
Lab Address: No. 1301-14&16, Guanguang Road, Xinlan Community, Guanlan Subdistrict, Longhua District, Shenzhen, Guangdong,China Post Code518110 Tel0755-23763060-8805 Fax0755-23763060 E-mailsz-kf@cvc.org.cn FCC(Test firm designation numberCN1363) IC(Test firm CAB identifier numberCN0137) CNAS(Test firm designation numberL16091)

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 7 of 33

1.4 TEST Standards and Limits

No.

Identity

1 FCC 47 CFR Part 2

2 ANSI/IEEE Std. C95.1-1992

3 IEEE Std. 1528-2013

Document Title
Frequency Allocations and Radio Treaty Matters; General Rules and Regulations
IEEE Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz
Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement Techniques

4 FCC KDB 447498 D01 v06

Mobile and Portable Device RF Exposure Procedures and Equipment Authorization Policies

5 FCC KDB 865664 D01 v01r04 SAR Measurement 100 MHz to 6 GHz

6 FCC KDB 865664 D02 v01r02 RF Exposure Reporting

7

FCC KDB 941225 D06 v02r01

SAR EVALUATION PROCEDURES FOR PORTABLE DEVICES WITH WIRELESS ROUTER CAPABILITIES

8 KDB 248227 D01 v02r02

802 11 Wi-Fi SAR

9 FCC KDB 447498 D04 v01

Interim General RF Exposure Guidance

(A). Limits for Occupational/Controlled Exposure (W/kg)

Whole-Body Partial-Body

Hands, Wrists, Feet and Ankles

0.4

8.0

20.0

(B). Limits for General Population/Uncontrolled Exposure (W/kg)

Whole-Body Partial-Body

Hands, Wrists, Feet and Ankles

0.08

1.6

4.0

NOTE: Whole-Body SAR is averaged over the entire body, partial-body SAR is averaged over any 1 gram of tissue defined as a tissue volume in the shape of a cube. SAR for hands, wrists, feet and ankles is averaged over any 10 grams of tissue defined as a tissue volume in the shape of a cube. Population/Uncontrolled Environments: Are defined as locations where there is the exposure of individuals who have no knowledge or control of their exposure. Occupational/Controlled Environments: Are defined as locations where there is exposure that may be incurred by people who are aware of the potential for exposure, (i.e. as a result of employment or occupation).

NOTE GENERAL POPULATION/UNCONTROLLED EXPOSURE
PARTIAL BODY LIMIT 1.6 W/kg

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 8 of 33

1.5 STATEMENT of Compliance

The maximum results of Specific Absorption Rate (SAR) found during testing for the EUT are as follows:

Mode

Highest Reported Body SAR1g (0 cm Gap) (W/kg)

2.4G WLAN

0.62

5.2G WLAN

0.56

5.8G WLAN

0.67

Bluetooth

0.10

Max. SAR

0.67

Note: 1. This device is in compliance with Specific Absorption Rate (SAR) for general population/uncontrolled
exposure limits (1.6W/kg as averaged over any 1 gram of tissue; 10-gram SAR for Product Specific 10g SAR, limit: 4.0W/kg) specified in FCC 47 CFR part 2 (2.1093) and ANSI/IEEE C95.1-1992 and had been tested in accordance with the measurement methods and procedures specified in IEEE 1528-2013 and FCC KDB publications.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 9 of 33

2 SAR Measurement System
2.1 Definition of Specific Absorption Rate (SAR)
SAR is related to the rate at which energy is absorbed per unit mass in an object exposed to a radio field. The SAR distribution in a biological body is complicated and is usually carried out by experimental techniques or numerical modeling. The standard recommends limits for two tiers of groups, occupational/controlled and general population/uncontrolled, based on a person's awareness and ability to exercise control over his or her exposure. In general, occupational/controlled exposure limits are higher than the limits for general population/uncontrolled.
The SAR definition is the time derivative (rate) of the incremental energy (dW) absorbed by (dissipated in) an incremental mass (dm) contained in a volume element (dv) of a given density (). The equation description is as below:

SAR is expressed in units of Watts per kilogram (W/kg) SAR measurement can be related to the electrical field in the tissue by
Where:  is the conductivity of the tissue;  is the mass density of the tissue and E is the RMS electrical field strength.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR
2.2 SAR System
DASY System Diagram:

Page 10 of 33

DASY is a system that is able to determine the SAR distribution inside a phantom of human being according to different standards. The DASY system consists of the following items: - Main computer to control all the system - 6 axis robot - Data acquisition Electronics - Miniature E-field probe - Phone holder - Head simulating tissue The following figure shows the system.
The EUT under test operating at the maximum power level is placed in the phone holder, under the phantom, which is filled with head simulating liquid. The E-Field probe measures the electric field inside the phantom. The Open SAR software computes the results to give a SAR value in a 1g or 10g mass.
LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 11 of 33

2.3 Probe
EX3DV4 ­ Smallest isotropic dosimetric probe for high precision SAR measurements in any exposure scenario (e.g., very strong gradient fields); the only probe that enables compliance testing for frequencies up to 10 GHz with a precision of better than 30% - Frequency range: 4 MHz ­ 10 GHz -Dynamic range: 0.01 W/kg ­ >100 W/kg -Tip diameter: 2.5 mm -Scanning distance: 1.4 mm

Figure 1-Speag COMOSAR Dosimetric E field Dipole
2.4 Date Acquisition Electronics 4 (DAE4)
High precision 3-channel differential voltmeter for use with SPEAG's field, SAR, and temperature probes. Serial optical link for communication with the DASY8 measurement server. Two-step probe touch detector for mechanical surface detection and emergency robot stop. - Measurement range: -100 ­ +300 mV (16-bit resolution and two range settings: 4 mV, 400 mV) - Input offset voltage: <5 µV (with auto zero) - Input resistance: 200 MOhm - Input bias current: <50 fA - Battery power: >10 hours of operation (with two 9.6 V NiMH batteries) - Dimensions (L × W × H): 60 × 60 × 68 mm - Calibration: ISO/IEC 17025 calibration service available.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 12 of 33

2.4.1 SAM-Twin Phantom
The shell corresponds to the specifications of the Specific Anthropomorphic Mannequin (SAM) phantom defined in IEC/IEEE 62209-1528. It enables the dosimetric evaluation of left and right hand phone usage as well as body-mounted usage at the flat phantom region. A cover prevents evaporation of the liquid. Reference markings on the phantom allow the complete setup of all predefined phantom positions and measurement grids by teaching three points with the robot.SAM-Twin V5.0 and higher has the same shell geometry and is manufactured from the same material as SAM-Twin V4.0 but with reinforced top structure. - Material: Vinyl ester, fiberglass reinforced (VE-GF) - Shell Thickness: 2 ± 0.2 mm (6 ± 0.2 mm at ear point) - Dimensions:Length: 1000 mm
Width: 500 mm Height: adjustable feet

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 13 of 33

2.4.2 ELI Phantom
The ELI phantom is used for compliance testing of handheld and body-mounted wireless devices in the frequency range of 4 MHz to 10 GHz. ELI is fully compatible with the IEC/IEEE 62209-1528 standard and all known tissue simulating liquids. ELI has been optimized regarding its performance and can be integrated into our standard phantom tables. A cover prevents evaporation of the liquid. Reference markings on the phantom allow installation of the complete setup, including all predefined phantom positions and measurement grids, by teaching three points. The phantom is compatible with all of SPEAG's dosimetric probes and dipoles.The latest ELI V8.0 phantom shell has optimized pretension in the bottom surface during production, such that the phantom is more robust and with reduced sagging. - Material:Vinyl ester, fiberglass reinforced (VE-GF) - Shell Thickness:2.0 ± 0.2 mm (bottom plate) - Dimensions:Major axis: 600 mm,
Minor axis: 400 mm - Filling Volume:approx. 30 liters.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 14 of 33

2.5 Wrist Phantom
The Wrist Phantom V10 is shape-compatible with the CTIA approved OTA GFPC-V1 and optimized for specific absorption rate evaluation of watches and other wireless hand accessories. - Material:Photosensitive epoxy acrylates - Shell Thickness:2 ± 0.2 mm - Wrist Shape:Design compatible with CTIA forearm.

2.6 Device Holder
The SAR in the phantom is approximately inversely proportional to the square of the distance between the source and the liquid surface. For a source at 5 mm distance, a positioning uncertainty of ±0.5 mm would produce a SAR uncertainty of ±20 %. Accurate device positioning is therefore crucial for accurate and repeatable measurements. The positions in which the devices must be measured are defined by the standards.
LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 15 of 33

2.7 System Validation Dipoles
Symmetrical dipole with l/4 balun. Enables measurement of feed point impedance with NWA. Matched for use near flat phantoms filled with tissue simulating solutions. - Frequency: 300 MHz to 10 GHz - Return loss: >20 dB - Power capability: >40 W

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 16 of 33

3 TISSUE Simulating Liquids
3.1 Simulating Liquids Parameter Check
For SAR measurement of the field distribution inside the phantom, the phantom must be filled with homogeneous tissue simulating liquid to a depth of at least 15 cm. For head SAR testing, the liquid height from the ear reference point (ERP) of the phantom to the liquid top surface is larger than 15 cm. For body SAR testing, the liquid height from the center of the flat phantom to the liquid top surface is larger than 15 cm. The nominal dielectric values of the tissue simulating liquids in the phantom and the tolerance of 5% are listed.

Liquid Height for Head Position

Liquid Height for Body Position

The dielectric properties of the tissue simulating liquids are defined in IEC 62209-1528. The dielectric properties of the tissue simulating liquids were verified prior to the SAR evaluation using a dielectric assessment kit and a network analyzer.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 17 of 33

Dielectric properties of Tissue Simulating Liquid

Frequency (MHz)
750 835 900 1450 1640 1750 1800 1900 2000 2100 2300 2450 2600 3000 3500 4000 4500 5000 5200 5300 5500 5600 5800 6000

Target Permittivity
41.9 41.5 41.5 40.5 40.3 40.1 40.0 40.0 40.0 39.8 39.5 39.2 39.0 38.5 37.9 37.4 36.8 36.2 36.0 35.9 35.6 35.5 35.3 35.1

Target Conductivity
0.89 0.90 0.97 1.20 1.29 1.37 1.40 1.40 1.40 1.49 1.67 1.80 1.96 2.40 2.91 3.43 3.94 4.45 4.66 4.76 4.96 5.07 5.27 5.48

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR
3.2 LIQUIDS Measurement Results
The measuring results for tissue simulating liquid are shown as below.

Page 18 of 33

Tissue Type
HSL600-10000

Frequency (MHz)
2450 5200 5800

Measured Conductivity
()
1.840 4.570 5.240

Measured Permittivity
(r)
39.400 37.200 36.100

Target Conductivity
()
1.80 4.66 5.27

Target Permittivity
(r)
39.20 36.00 35.30

Conductivity Deviation (%)
2.22 -1.93 -0.57

Permittivity Deviation
(%)
0.51 3.33 2.27

Test Date
Dec. 06, 2024 Dec. 06, 2024 Dec. 06, 2024

Note: 1. The dielectric properties of the tissue simulating liquid must be measured within 24 hours before the SAR
testing and within ±5% of the target values. Liquid temperature during the SAR testing must be within ±2 . 2. Since the maximum deviation of dielectric properties of the tissue simulating liquid is within 5%.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 19 of 33

4 SAR System Validation 4.1 Validation System
Each DASY system is equipped with one or more system validation kits. These units, together with the predefined measurement procedures within the DASY software, enable the user to conduct the system performance check and system validation. System kit includes a dipole, and dipole device holder.
The system check verifies that the system operates within its specifications. It's performed daily or before every SAR measurement. The system check uses normal SAR measurement in the flat section of the phantom with a matched dipole at a specified distance. The system validation setup is shown as below.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR
4.2 SYSTEM Validation Result
The measuring result for system verification is tabulated as below.

Page 20 of 33

Test Date

Frequency (MHz)

1W Target SAR-1g (W/kg)

Measured SAR-1g (W/kg)

Normalized to 1W SAR-1g (W/kg)

Deviation (%)

Dipole S/N

Probe S/N

DAE S/N

Dec. 06, 2024 Dec. 06, 2024 Dec. 06, 2024

2450 5200 5800

51.40 77.80 80.40

12.90 7.87 7.91

51.60 78.70 79.10

0.39 1.16 -1.62

1081 1353 1353

7628 7628 7628

1557 1557 1557

Note:

Comparing to the reference SAR value provided by SPEAG, the validation data should be within its

specification of 10 %. The result indicates the system check can meet the variation criterion and the plots can

be referred to Appendix A of this report.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 21 of 33

5 SAR Evaluation Procedures
To evaluate the peak spatial-average SAR values with respect to 1 g or 10 g cubes, fine resolution volume scans, called zoom scans, are performed at the peak SAR locations identified during the area scan. If the cube volume within the zoom scan chosen to calculate the peak spatial-average SAR touches any boundary of the zoom-scan volume, the zoom scan shall be repeated with the center of the zoom-scan volume shifted to the new maximum SAR location. For any secondary peaks found in the area scan that are within 2 dB of the maximum peak and are not within this zoom scan, the zoom scan shall be performed for such peaks, unless the peak spatial-average SAR at the location of the maximum peak is more than 2 dB below the applicable SAR limit (i.e., 1 W/kg for a 1.6 W/kg 1 g limit, or 1.26 W/kg for a 2 W/kg 10 g limit). The zoom scan resolutions specified in the table below must be applied to the SAR measurements.

Table 3 - Area scan parameters
Parameter
Maximum distance between the measured points (geometric centre of the sensors) and the inner phantom surface (zM1 in Figure 20 in mm)
Maximum spacing between adjacent measured points in mm (see O.8.3.1) b

DUT transmit frequency being tested

f 3 GHz

3 GHz <f 10 GHz

5 ±1

 ln(2)/2 ± 0,5a

20, or half of the corresponding zoom scan length, whichever is
smaller

60/f, or half of the corresponding zoom scan length,whichever is smaller

Maximum angle between the probe axis and the phantom surface normal (a in Figure 20)c

5°(flat phantom only) 30°(other phantoms)

5°(flat phantom only) 20°(other phantoms)

Tolerance in the probe angle

1°

1°

a  is the penetration depth for a plane-wave incident normally on a planar half-space.

b See Clause O.8 on how x and  y may be selected for individual area scan requirements.

c The probe angle relative to the phantom surface normal is restricted due to the degradation in the measurement accuracy in fields with steep spatial gradients. The measurement accuracy decreases with increasing probe angle and increasing frequency. This is the reason for the tighter probe angle restriction at frequencies above 3 GHz.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 22 of 33

Table 4 - Zoom scan parameters
Parameter

DUT transmit frequency being tested

f 3 GHz

3 GHz <f 10 GHz

Maximum distance between the closest measured points and the phantom surface (zM1 in Figure 20 and Table 3, in mm)

5

 ln(2)/2 ± 0,5a

Maximum angle between the probe axis and the phantom surface normal ( in Figure 20)
Maximum spacing between measured points in the x- and
y-directions ( x and  y, in mm)

5°(flat phantom only) 30°(other phantoms)
8

5°(flat phantom only) 20°(other phantoms)
24/f b

For uniform grids:

Maximum spacing between measured points in the

direction normal to the phantom shell

5

( z1 in Figure 20, in mm)

For graded grids:

Maximum spacing between the two closest 4
measured points in the direction normal to the phantom
shell ( z1 in Figure 20, in mm)

For graded grids:

Maximum incremental increase in the spacing 1,5
between measured points in the direction normal to the
phantom shell (Rz =  z2/ z1 in Figure 20)

Minimum edge length of the zoom scan volume in the x-

and y-directions (Lz, in O.8.3.2, in mm)

30

10/(f - 1) 12/f 1,5 22

Minimum edge length of the zoom scan volume in the

direction normal to the phantom shell

30

22

(Lh in O.8.3.2 in mm)

Tolerance in the probe angle

1°

1°

a  is the penetration depth for a plane-wave incident normally on a planar half-space.

b This is the maximum spacing allowed, which might not work for all circumstances.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 23 of 33

6 SAR Measurement Evaluation 6.1 EUT Configuration and Setting
<Considerations Related to WLAN for Setup and Testing> In general, various vendor specific external test software and chipset based internal test modes are typically used for SAR measurement. These chipset based test mode utilities are generally hardware and manufacturer dependent, and often include substantial flexibility to reconfigure or reprogram a device. A Wi-Fi device must be configured to transmit continuously at the required data rate, channel bandwidth and signal modulation, using the highest transmission duty factor supported by the test mode tools for SAR measurement. The test frequencies established using test mode must correspond to the actual channel frequencies. When 802.11 frame gaps are accounted for in the transmission, a maximum transmission duty factor of 92 - 96% is typically achievable in most test mode configurations. A minimum transmission duty factor of 85% is required to avoid certain hardware and device implementation issues related to wide range SAR scaling. In addition, a periodic transmission duty factor is required for current generation SAR systems to measure SAR correctly. The reported SAR must be scaled to 100% transmission duty factor to determine compliance at the maximum tune-up tolerance limit.
According to KDB 248227 D01, this device has installed WLAN engineering testing software which can provide continuous transmitting RF signal. During WLAN SAR testing, this device was operated to transmit continuously at the maximum transmission duty with specified transmission mode, operating frequency, lowest data rate, and maximum output power.
Initial Test Configuration An initial test configuration is determined for OFDM transmission modes in 2.4 GHz and 5 GHz bands according to the channel bandwidth, modulation and data rate combination(s) with the highest maximum output power specified for production units in each standalone and aggregated frequency band. When the same maximum power is specified for multiple transmission modes in a frequency band, the largest channel bandwidth, lowest order modulation, lowest data rate and lowest order 802.11a/g/n/ac mode is used for SAR measurement, on the highest measured output power channel in the initial test configuration, for each frequency band.
Subsequent Test Configuration SAR measurement requirements for the remaining 802.11 transmission mode configurations that have not been tested in the initial test configuration are determined separately for each standalone and aggregated frequency band, in each exposure condition, according to the maximum output power specified for production units. Additional power measurements may be required to determine if SAR measurements are required for subsequent highest output power channels in a subsequent test configuration. When the highest reported SAR for the initial test configuration according to the initial test position or fixed exposure position requirements, is adjusted by the ratio of the subsequent test configuration to initial test configuration specified maximum output power and the adjusted SAR is  1.2 W/kg, SAR is not required for that subsequent test configuration.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 24 of 33

SAR Test Configuration and Channel Selection When multiple channel bandwidth configurations in a frequency band have the same specified maximum output power, the initial test configuration is using largest channel bandwidth, lowest order modulation, lowest data rate, and lowest order 802.11 mode (i.e., 802.11a is chosen over 802.11n then 802.11ac or 802.11g is chosen over 802.11n). After an initial test configuration is determined, if multiple test channels have the same measured maximum output power, the channel chosen for SAR measurement is determined according to the following. 1) The channel closest to mid-band frequency is selected for SAR measurement. 2) For channels with equal separation from mid-band frequency; for example, high and low channels or two mid-band channels, the higher frequency (number) channel is selected for SAR measurement.

<Considerations Related to Bluetooth for Setup and Testing> This device has installed Bluetooth engineering testing software which can provide continuous transmitting RF signal. During Bluetooth SAR testing, this device was operated to transmit continuously at the maximum transmission duty with specified transmission mode, operating frequency, lowest data rate, and maximum output power.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 25 of 33

6.2 EUT Testing Position
6.2.1 Body Exposure Conditions
This variant report is made for verification. All the worst SAR configurations specified in the original SAR report was repeated and verified to ensure the device remains compliant.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 26 of 33

7 MAXIMUM Output Power 7.1 Maximum Conducted Power
Refer to the original report

7.2 MEASURED Conducted Power Result
Refer to the original report

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 27 of 33

8 SAR Testing Results

8.1 SAR Results for Body Exposure Condition (Separation Distance is 0 cm Gap)

Plot No.

Band

P1 802.11b P2 802.11ac P3 802.11n P4 BLE

Mode
VHT80 HT40
1M

Test Position
Top Side Top Side Top Side Top Side

Separation Distance
(cm)
0 0 0 0

Ch.
1 42 151 19

Max. Tune-up Power (dBm)
15.0 13.0 14.0 4.5

Measured Conducted
Power (dBm)
14.52 12.76 13.67 4.11

Scaling Factor
1.12 1.06 1.08 1.09

Power Drift (dB)

Measured SAR-1g (W/kg)

-0.06 0.03 0.02 0.00

0.555 0.527 0.621 0.087

Scaled SAR-1g (W/kg)
0.62 0.56 0.67 0.10

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 28 of 33

8.2 SAR Measurement Variability
According to KDB 865664 D01, SAR measurement variability was assessed for each frequency band, which is determined by the SAR probe calibration point and tissue-equivalent medium used for the device measurements. When both head and body tissue-equivalent media are required for SAR measurements in a frequency band, the variability measurement procedures should be applied to the tissue medium with the highest measured SAR, using the highest measured SAR configuration for that tissue-equivalent medium. Alternatively, if the highest measured SAR for both head and body tissue-equivalent media are  1.45 W/kg and the ratio of these highest SAR values, i.e., largest divided by smallest value, is  1.10, the highest SAR configuration for either head or body tissue-equivalent medium may be used to perform the repeated measurement. These additional measurements are repeated after the completion of all measurements requiring the same head or body tissue-equivalent medium in a frequency band. The test device should be returned to ambient conditions (normal room temperature) with the battery fully charged before it is re-mounted on the device holder for the repeated measurement(s) to minimize any unexpected variations in the repeated results.

Since all the measured SAR are less than 0.8 W/kg, the repeated measurement is not required.

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 29 of 33

9 EQUITPMENT List

Equipment

Manufacturer

Model

SN

System Validation Dipole System Validation Dipole Dosimetric E-Field Probe Data Acquisition Electronics Wideband Radio Communication Tester Signal Analyzer Vector Network Analyzer Dielectric assessment Kit Signal Generator Power Sensor Power Sensor DC Power Supply Cavity Coupler Directional Couper
Coaxial attenuator
Coaxial attenuator
Digital Thermometer temperature and humidity indicator Power Amplifier Mini circuit
Head Tissue Simulating Liquid
PHANTOM PHANTOM

SPEAG SPEAG SPEAG SPEAG
R&S R&S R&S SPEAG R&S R&S R&S Topward
/ /
R&S
R&S
LKM LINI-T mini-circuits
SPEAG
SPEAG SPEAG

D2450V2

1081

D5GHzV2

1353

EX3DV4

7628

DAE4

1557

CMW500

168558

FSV

104408

ZNB 40

101544

DAK-3.5

1327

SMB 100B

101440

NRP18S-10

101843

NRP18S-10

101845

3303D /

810984 LS0300103

SHX-DC04/12-20N 2206171042

8491A

1424.6721k02101845-HX

8491A

1424.6721K02101843-aM

DTM3000

3946

A10T

C193561473

ZVA-183W-S+

726202215

SL AAH U16 HBBL600-10000V6
BC

ELI V8.0

2171

SAM-Twin V8.0

2097

Cal. Data
May. 25, 2022 May. 27, 2022 July. 03, 2024 Oct. 08, 2024 May. 25, 2024 May. 22, 2024 May. 25, 2024 Oct. 22, 2022 Sep. 23, 2024 Sep. 20, 2024 Sep. 20, 2024 Sep. 20, 2024 Jan. 08, 2025 Jan. 08, 2025
Sep. 20, 2024
Sep. 20, 2024
Jan. 15, 2025 Apr. 28, 2024 Jan. 08, 2025
Jun.01,2022
N/A N/A

Cal. interval 3 years 3 years 1 year 1 year 1 year 1 year 1 year
N/A 1 year 1 year 1 year 1 year 1 year 1 year
1 year
1 year
1 year 1 year 1 year
N/A
N/A N/A

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 30 of 33

10 MEASUREMENT Uncertainty

This uncertainty represents an expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor of k=2.

Source of Uncertainty

Tolerance (± %)

Measurement System Errors

Probability Distribution

Standard

Standard

Ci

Ci

Vi

Divisor

Uncertainty Uncertainty

(1g) (10g)

Veff

(1g)

(10g)

Probe Calibration

±11.0%

Normal (k=2)

2

1

1

± 5.5 %

± 5.5 %



Probe Calibration Drift

±1.7%

Rectangular

3

1

1

±1.0%

Probe Linearity

±4.7%

Rectangular

3

1

1

±2.7%

Broadband Signal

±3.0%

Rectangular

3

1

1

±1.7%

Probe Isotropy

±7.6%

Rectangular

3

1

1

±4.4%

Other Probe + Electronic

±0.7%

Normal

1

1

1

±0.7%

RF Ambient

±1.8%

Normal

1

1

1

±1.8%

Probe Positioning

±0.006mm

Normal

1

0.14 0.14

±0.10%

Data Processing

±1.2%

Normal

1

1

1

±1.2%

Phantom and Device Errors Conductivity (meas.)DAK Conductivity (temp.)BB

±2.5% ±3.3%

Normal Rectangular

1

0.78 0.71

3 0.78 0.71

±2.0% ±1.5%

Phantom Permittivity

±14.0%

Rectangular

3

0

0

±0%

Distance DUT ­ TSL

±2.0%

Normal

1

2

2

±4.0%

Device Positioning

±2.4%/±2.8%

Normal

1

1

1

±2.8%

Device Holder DUT Modulationm

±3.4%/±3.5%

Normal

1

±2.4%

Rectangular

3

1

1

1

1

±3.5% ±1.4%

Time-average SAR

±1.7%

Rectangular

3

1

1

±1.0%

DUT drift Val Antenna Unc.val Unc. Input Powerval

±2.5% ±0.0% ±0.0%

Normal Normal Normal

1

1

1

±2.5%

1

1

1

±0%

1

1

1

±0%

Correction to the SAR results

C(,) SAR scalingp

±1.9% ±0.0%

Normal

1

Rectangular

3

1 0.84

1

1

±1.9% ±0%

Combined Standard Uncertainty (K = 1) Expanded Uncertainty (K = 2)

±12.54% ±25.1%

Uncertainty budget for frequency range 300 MHz to 3 GHz

±1.0%



±2.7%



±1.7%



±4.4%



±0.7%



±1.8%



±0.10%



±1.2%



±1.8%

100

±1.4%



±0%



±4.0%



±2.8%

30

±3.5%

30

±1.4%



±1.0%



±2.5%

30

±0%

±0%

±1.6% ±0% ±12.44% ±24.9%

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 31 of 33

Source of Uncertainty Measurement System Errors

Tolerance (± %)

Probability Diviso Ci

Distribution

r

(1g)

Standard

Standard

Ci

Vi

Uncertainty Uncertainty

(10g)

Veff

(1g)

(10g)

Probe Calibration

±13.1%

Normal

2

1

1

± 6.55 %

± 6.55 %



(k=2)

Probe Calibration Drift

±1.7%

Rectangular

3

1

1

±1.0%

Probe Linearity

±4.7%

Rectangular

3

1

1

±2.7%

Broadband Signal

±2.6%

Rectangular

3

1

1

±1.5%

Probe Isotropy

±7.6%

Rectangular

3

1

1

±4.4%

Other Probe + Electronic

±1.2%

Normal

1

1

1

±1.2%

RF Ambient

±1.8%

Normal

1

1

1

±1.8%

Probe Positioning

±0.005mm

Normal

1

0.29

0.29

±0.15%

Data Processing

±2.3%

Normal

1

1

1

±2.3%

Phantom and Device Errors

Conductivity (meas.)DAK Conductivity (temp.)BB

±2.5% ±3.3%

Normal

1

Rectangular

3

0.78

0.71

0.78

0.71

±2.0% ±1.5%

Phantom Permittivity

±14.0%

Rectangular

3

0.25

0.25

±2%

Distance DUT ­ TSL

±2.0%

Normal

1

2

2

±4.0%

Device Positioning

±2.4%/±2.8%

Normal

1

1

1

±2.8%

Device Holder DUT Modulationm

±3.4%/±3.5%

Normal

1

1

1

±2.4%

Rectangular

3

1

1

±3.5% ±1.4%

Time-average SAR

±1.7%

Rectangular

3

1

1

±1.0%

DUT drift Val Antenna Unc.val

±2.5% ±0.0%

Normal

1

1

1

Normal

1

1

1

±2.5% ±0%

Unc. Input Powerval

±0.0%

Normal

1

1

1

±0%

Correction to the SAR results

Deviation to Target SAR scalingp

±1.9% ±0.0%

Normal

1

Rectangular

3

1

0.84

1

1

±1.9% ±0%

Combined Standard Uncertainty (K = 1)

±12.8%

Expanded Uncertainty (K = 2)

±26.1% Uncertainty budget for frequency range 3 GHz to 6 GHz

±1.0%



±2.7%



±1.5%



±4.4%



±1.2%



±1.8%



±0.15%



±2.3%



±1.8%

60

±1.4%



±2%



±4.0%



±2.8%

30

±3.5%

30

±1.4%



±1.0%



±2.5%

30

±0%

±0%

±1.6% ±0% ±12.7% ±25.9%

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 32 of 33

11 APPENDIXES
All attachments are integral parts of this test report. This applies especially to the following appendix:
11.1 Appendix A: SAR Plots of System Verification
The plots for system verification with largest deviation for each SAR system combination are shown as follows.
11.2 Appendix B: SAR Plots of SAR Measurement

11.3 Appendix C: Calibration Certificate for probe and Dipole
Refer the appendix Calibration Report.
11.4 Appendix D: Photographs of EUT and setup

LT13-R-1101-SAR_FCC-A1

CVC Testing Technology (Shenzhen) Co., Ltd.

Test Report No.: GJWSZ2025-0188-SAR

Page 33 of 33

Important
(1) The test report is invalid without the official stamp of CVC; (2) Any part photocopies of the test report are forbidden without the written permission from CVC; (3) The test report is invalid without the signatures of Approval and Reviewer; (4) The test report is invalid if altered; (5) Objections to the test report must be submitted to CVC within 15 days. (6) Generally, commission test is responsible for the tested samples only. (7) As for the test result "-" or "N" means "not applicable", "/" means "not test", "P" means "pass" and "F" means "fail"

Address: No. 1301-14&16, Guanguang Road, Xinlan Community, Guanlan Subdistrict, Longhua District, Shenzhen, Guangdong,China Post Code518110 Tel0755-23763060-8805 Fax0755-23763060 E-mailsz-kf@cvc.org.cn http://www.cvc.org.cn


Microsoft Word 2010; modified using iTextSharp™ 5.5.1 ©2000-2014 iText Group NV (AGPL-version) Microsoft Word 2010