1587 cmeng0100
®
1587/1577 Insulation Multimeters
Calibration Manual
PN 2456783 September 2005 Rev. 1, 6/09
© 2005, 2009 Fluke Corporation, All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.

Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product's specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).

THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.

Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands

11/99

Table of Contents

Title

Page

Introduction........................................................................................................ 1 Safety Information ............................................................................................. 1 Contacting Fluke................................................................................................ 2 General Specifications ....................................................................................... 2 Electrical Specifications .................................................................................... 3
AC Voltage Measurement ............................................................................. 3 1587 Accuracy .......................................................................................... 3 1587 Low-pass Filter Voltage ................................................................... 3 1577 Accuracy .......................................................................................... 3
DC Voltage Measurement ............................................................................. 4 DC Millivolts Measurement .......................................................................... 4 DC and AC Current Measurement ................................................................ 4 Ohms Measurement....................................................................................... 4 Diode Test (1587 Only)................................................................................. 4 Continuity Test .............................................................................................. 5 Frequency Measurement (1587 Only)........................................................... 5 Frequency Counter Sensitivity ...................................................................... 5 Capacitance (1587 Only)............................................................................... 5 Temperature Measurement (1587 Only) ....................................................... 5 Insulation Specifications ............................................................................... 5
Model 1587 ............................................................................................... 6 Model 1577 ............................................................................................... 6 Basic Maintenance ............................................................................................. 6 Cleaning......................................................................................................... 6 Testing the Batteries Under Load.................................................................. 6 Testing the Fuse............................................................................................. 7 Replacing the Batteries and/or Fuse .............................................................. 8 Testing the Display........................................................................................ 9 Testing the Backlight..................................................................................... 9 Testing the Keypad........................................................................................ 9 Disassembling and Reassembling the Meter ..................................................... 10 Removing the Boot........................................................................................ 11 Removing the Battery Door........................................................................... 11 Removing the Bottom Case........................................................................... 12 Removing the PCA........................................................................................ 12

i

1587/1577
Calibration Manual

Removing the LCD........................................................................................ 13 Replacing the LCD ........................................................................................ 13 Reassembling the Meter ................................................................................ 15 Required Tools and Equipment ......................................................................... 15 Performance Test ............................................................................................... 16 Testing the 1577 and 1587 DMM Functions................................................. 16 Testing Temperature Function (1587 Only).................................................. 18 Testing the Discharge Circuit ........................................................................ 18 Testing the Insulation Function ..................................................................... 18
Insulation Resistance Accuracy Test......................................................... 18 Insulation Function, External Sense.......................................................... 19 Insulation Function, Source Voltage Accuracy Test................................. 20 I Nominal Test........................................................................................... 21 Low-Battery Test - DMM Function .............................................................. 21 Low-Battery Test - INSULATION Function ................................................ 22 Calibration Adjustment Procedure..................................................................... 23 Calibration Adjustment Counter.................................................................... 23 Calibration Adjustment Password ................................................................. 23 Changing the Password ................................................................................. 23 Restoring the Default Password .................................................................... 23 Meter Keys Used in the Calibration Steps..................................................... 25 Calibration Adjustment ................................................................................. 25 Service and Parts................................................................................................ 27

ii

List of Tables

Table

Title

Page

1. Description of Symbols Used in the Manual.......................................................... 1 2. Required Tools and Equipment.............................................................................. 15 3. DMM Accuracy Tests ............................................................................................ 16 4. Insulation Resistance Accuracy Test...................................................................... 19 5. Insulation Mode, Source Voltage Accuracy Tests ................................................. 20 6. Calibration Adjustment Steps................................................................................. 26 7. 1587/1577 Replacement Parts................................................................................ 28

iii

1587/1577
Calibration Manual
iv

List of Figures

Figure

Title

Page

1. Testing the Fuse ..................................................................................................... 7 2. Replacing the Batteries and Fuse ........................................................................... 8 3. LCD Display Test .................................................................................................. 9 4. Disassembling the Meter........................................................................................ 10 5. Insulation Terminal Clips....................................................................................... 13 6. Accessing the LCD ................................................................................................ 14 7. Source Voltage Accuracy Test............................................................................... 20 8. I Nominal Test........................................................................................................ 21 9. Low-Battery Test Connections............................................................................... 22 10. Restoring the Default Password ............................................................................. 24 11. 1587/1577 Replacement Parts................................................................................ 27

v

1587/1577
Calibration Manual
vi

Insulation Multimeters Introduction

Introduction
The Fluke Model 1587, and Model 1577 are battery-powered, true-rms insulation multimeters (hereafter, Meter or UUT) with a 6000-count and a 3 ¾ digit display. Although this manual contains calibration information for Models 1587 and 1577, all illustrations and examples assume use of the Model 1587. Table 1 provides a description of all the symbols used in this manual.
The Meter meets CAT III and CAT IV IEC 61010 standards. The IEC 61010 standard defines four measurement categories (CAT I to IV) based on the magnitude of danger from transient impulses. CAT III meters are designed to protect against transients in fixed equipment installations at the distribution level; CAT IV meters are designed to protect against transients from the primary supply level (overhead or underground utility service).
Safety Information
XWWarning
To avoid possible electric shock or personal injury, follow these guidelines: · Use the Meter only as specified in this manual or the protection provided by the
Meter might be impaired. · Do not use the Meter or test leads if they appear damaged, or if the Meter is not
operating properly. If in doubt, have the Meter serviced. · Always use the proper terminal, switch position, and range for measurements
before connecting the Meter to circuit under test. · Verify that the Meter is operational by measuring a known voltage. · Do not apply more than the rated voltage as marked on the Meter, between the
terminals or between any terminal and earth ground. · Use caution with voltages above 30 V ac rms, 42 V ac peak, or 60 V dc. These
voltages pose a shock hazard. · Replace the battery as soon as the low battery indicator (b) appears. · Disconnect circuit power and discharge all high-voltage capacitors before testing
resistance, continuity, diodes, or capacitance. · Do not use the Meter around explosive gas or vapor. · When using the test leads, keep your fingers behind the finger guards. · Remove test leads from the Meter before opening the Meter case or battery door.
Never operate the Meter with the back case removed or the battery door open. · Comply with local and national safety requirements when working in hazardous
locations. · Use proper protective equipment, as required by local or national authorities when
working in hazardous areas. · Avoid working alone. · Use only the replacement fuse specified or the protection may be impaired. · Check the test leads for continuity before use. Do not use if the readings are high or
noisy.

B F X b
~

Table 1. Description of Symbols Used in the Manual

AC (Alternating Current) DC (Direct Current) WARNING: risk of electric shock.

J

Earth Ground

I Fuse

T Double Insulated

Battery (Low battery when shown on display.)

W Important information; see manual

Do not dispose of this product as unsorted municipal waste. Go to Fluke's website for recycling information.

1

1587/1577 Calibration
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
· Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) · Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) · Canada: 1-800-36-FLUKE (1-800-363-5853) · Europe: +31 402-675-200 · Japan: +81-3-3434-0181 · Singapore: +65-738-5655 · Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
General Specifications
Maximum Voltage Applied to any Terminal ........ 1000 V ac rms or dc Storage Temperature.......................................... -40 °C to 60 °C (-40 °F to 140 °F) Operating Temperature ...................................... -20 °C to 55 °C (-4 °F to 131 °F) Temperature Coefficient ..................................... 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28
°C (< 64 °F or > 82 °F) Relative Humidity................................................ Noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F) 0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F) 0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F) Vibration.............................................................. Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument Shock .................................................................. 1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor) Electromagnetic Compatibility ............................ In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: accuracy = specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997). Safety.................................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSAC22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category III 1000 V (CAT III) and CAT IV 600 V. Certifications ....................................................... CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard EN 61010 Part 1-1002 Batteries.............................................................. Four AA batteries (NEDA 15A or IEC LR6) Battery Life.......................................................... Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 M with a duty cycle of 5 seconds on and 25 seconds off. Size ..................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L) Weight................................................................. 550 g (1.2 lb.) IP Rating ............................................................. IP40 Altitude Operating ........................................................ 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V Storage ........................................................... 12,000 m Over-Range Capability ....................................... 110% of range except for capacitance which is 1 % Compliance to EN 61557 .................................... IEC61557-1, IEC61557-2
2

Insulation Multimeters Electrical Specifications

Electrical Specifications

AC Voltage Measurement

1587 Accuracy
Range
600.0 mV 6.000 V 60.00 V 600.0 V 1000 V [1] 1 kHz bandwidth.

Resolution
0.1 mV 0.001 V 0.01 V 0.1 V 1 V

50 Hz to 60 Hz ±(% of Rdg + Digits)
±(1 % + 3) ±(1 % + 3) ±(1 % + 3) ±(1 % + 3) ±(2 % + 3)

60 Hz to 5000 Hz ±(% of Rdg + Digits)
±(2 % + 3) ±(2 % + 3) ±(2 % + 3) ±(2 % + 3) [1] ±(2 % + 3) [1]

1587 Low-pass Filter Voltage

Range

Resolution

600.0 mV

0.1 mV

6.000 V

0.001 V

60.00 V

0.01 V

600.0 V

0.1 V

1000 V

1 V

50 Hz to 60 Hz ±(% of Rdg + Digits)
±(1 % + 3) ±(1 % + 3) ±(1 % + 3) ±(1 % + 3) ±(2 % + 3)

60 Hz to 400 Hz ±(% of Rdg + Digits)
+(2 % + 3) -(6 % - 3) +(2 % + 3) -(6 % - 3) +(2 % + 3) -(6 % - 3) +(2 % + 3) -(6 % - 3) +(2 % + 3) -(6 % - 3)

1577 Accuracy

Range
600.0 mV 6.000 V 60.00 V 600.0 V 1000 V

Resolution
0.1 mV 0.001 V 0.01 V 0.1 V 1 V

50 Hz to 60 Hz ±(% of Rdg + Digits)
±(2 % + 3) ±(2 % + 3) ±(2 % + 3) ±(2 % + 3) ±(2 % + 3)

AC Conversion.................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at up to 500 V, decreasing linearly to crest factor <= 1.5 at 1000 V. For non-sinusoidal waveforms add ±(2 % reading + 2 % FS) typical, for a crest factor up to 3.
Input Impedance ................................................. 10 M (nominal), <100 pF, ac-coupled
Common Mode Rejection Ratio
(1 k unbalanced) .......................................... >60 dB at dc, 50 or 60 Hz
Overload Protection ............................................ 1000 V rms or dc, 107 V Hz Max

3

1587/1577 Calibration

DC Voltage Measurement

Range

Resolution

6.000 V dc

0.001 V

60.00 V dc

0.01 V

600.0 V dc

0.1 V

1000 V dc

1 V

[1] Accuracies apply to ± 100% of range.

Accuracy 1587 [1] ±(% of Rdg + Digits)
±(0.09 % + 2) ±(0.09 % + 2) ±(0.09 % + 2) ±(0.09 % + 2)

Accuracy 1577 [1] ±(% of Rdg + Digits)
±(0.2 % + 2) ±(0.2 % + 2) ±(0.2 % + 2) ±(0.2 % + 2)

Input Impedance ................................................. 10 M (nominal), <100 pF Normal Mode Rejection Ratio............................. >60 dB @ 50 Hz or 60 Hz Common Mode Rejection Ratio.......................... >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) Overload Protection ............................................ 1000 V rms or dc

DC Millivolts Measurement

Range 600.0 mV dc

Resolution 0.1 mV

Accuracy 1587 ±(% of Rdg + Digits)
±(0.1 % + 1)

Accuracy 1577 ±(% of Rdg + Digits)
±(0.2 % + 1)

DC and AC Current Measurement

Range

AC 45 Hz to 1000 Hz

400 mA 60 mA

400 mA DC
60 mA

[1] 1 kHz bandwidth.

Resolution
0.1 mA 0.01 mA
0.1 mA 0.01 mA

Accuracy 1587 ±(% of Rdg+Digits)
±(1.5 % + 2) [1] ±(1.5 % + 2) [1]
±(0.2 % + 2) ±(0.2 % + 2)

Accuracy 1577 ±(% of Rdg+Digits)
±(2 % + 2) [1] ±(2 % + 2) [1]
±(1.0 % + 2) ±(1.0 % + 2)

Burden Voltage (Typical) 2 mV/mA
2 mV/mA

Overload ............................................................ 600 mA for 2 minutes maximum
Overload Protection ............................................ 440 mA, 1000 V, FAST fuse
AC Conversion.................................................... Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor <= 1.5 at 600 mA. For non-sinusoidal waveforms add +(2 % reading + 2 % FS) typical, for a crest factor up to 3.

Ohms Measurement

Range

Resolution

Accuracy 1587 [1] +(% of Rdg+Digits)

600.0 

0.1 

6.000 k

0.001 k

60.00 k

0.01 k

600.0 k

0.1 K

6.000 M

0.001 M

50.0 M

0.01 M

[1] Accuracies apply from 0 to 100% of range.

±(0.9 % + 2) ±(1.5 % + 3)

Overload Protection ............................................ 1000 V rms or dc Open Circuit Test Voltage................................... <8.0 V dc Short Circuit Current ........................................... <1.1 mA

Accuracy 1577 [1] +(% of Rdg+Digits)
±(1.2 % + 2)
±(2.0 % + 3)

Diode Test (1587 Only)
Diode Test Indication .......................................... Display voltage drop: 0.6 V at 1.0 mA nominal test current: Accuracy ............................................................. ±(2 % + 3)

4

Insulation Multimeters Electrical Specifications

Continuity Test
Continuity Indication ........................................... Continuous audible tone for test resistance below 25  and off above 100 . Maximum Reading; 1000 
Open Circuit Voltage........................................... <8.0 V Short Circuit Current ........................................... 1.0 mA typical Overload Protection ............................................ 1000 V rms Response Time................................................... >1 m sec

Frequency Measurement (1587 Only)

Range
99.99 Hz 999.9 Hz 9.999 kHz 99.99kHz

Resolution
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz

Accuracy ±(% of Rdg+Digits)
±(0.1 % + 1) ±(0.1 % + 1) ±(0.1 % + 1) ±(0.1 % + 1)

Frequency Counter Sensitivity

Input Range

V ac Sensitivity (RMS Sine Wave) [1]

5 Hz to 20 kHz

20 kHz to 100 kHz

DC Trigger Levels [1] to 20 kHz [2]

600.0 mV ac

100.0 mV

150.0 mV

na

6.0 V

1.0 V

1.5 V

-400.0 mV and 2.5 V

60.0 V

10.0 V

36.0 V

1.2 V and 4.0 V

600.0 V

100.0 V

-

12.0 V and 40.0 V

1000.0 V

300.0 V

-

12.0 V and 40.0 V

[1] Maximum input for specified accuracy = 10x range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy.

[2] Usable to 100 kHz with full scale input.

Capacitance (1587 Only)

Range
1000 nF 10.00 F 100.0 F 9999 F

Resolution
1 nF 0.01 F 0.1 F 1 F

Accuracy ±(% of Rdg+Digits)
±(1.2 % + 2)
±(1.2 % +/- 90 counts)

Temperature Measurement (1587 Only)

Range

Resolution

Accuracy [1] ±(% of Rdg+Digits)

-40 ° C to 537 ° C

0.1 °C

±(1 % + 10 counts)

-40 ° F to 998 ° F

0.1 °F

±(1 % + 18 counts)

[1] Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument.

Insulation Specifications
Measurement Range .......................................... 0.1 M to 2 G model 1587, 0.1 M to 600 M model 1577 Test Voltages ...................................................... 50, 100, 250, 500, 1000 V model 1587, 500 and 1000 V model
1577 Test Voltage Accuracy........................................ +20 %, - 0 % Short-Circuit Test Current................................... 1 mA nominal Auto Discharge ................................................... Discharge time <0.5 second for C = 1 F or less Live Circuit Detection:......................................... Inhibit test if terminal voltage >30 V prior to initialization of test. Maximum Capacitive Load ................................. Operable with up to 1 F load.

5

1587/1577 Calibration

Model 1587
Output Voltage
50 V (0 % to + 20 %)
100 V (0 % to + 20 %)
250 V (0 % to + 20 %)
500 V (0 % to + 20 %)
1000 V (0 % to + 20 %)

Display Range
0.01 to 6.00 M 6.0 to 50.0 M 0.01 to 6.00 M 6.0 to 60.0 M 60 to 100 M 0.1 to 60.0 M 60 to 250 M 0.1 to 60.0 M 60 to 500 M 0.1 to 60.0 M 60 to 600 M 0.6 to 2.0 G

Resolution
0.01 M 0.1 M 0.01 M 0.1 M 1 M 0.1 M 1 M 0.1 M 1 M 0.1 M 1 M 100 M

Test Current 1 mA @ 50 k

Resistance Accuracy ±(% of Rdg + Digits)
±(3 % + 5 counts)

1 mA @ 100 k

±(3 % + 5 counts)

1 mA @ 250 k 1 mA @ 500 k 1 mA @ 1 M

±(1.5 % + 5 counts) ±(1.5 % + 5 counts) ±(1.5 % + 5 counts) ±(10 % + 3 counts)

Model 1577
Output Voltage
500 V (0 % to + 20 %)
1000 V (0 % to + 20 %)

Display Range
0.1 to 60.0 M 60 to 500 M 0.1 to 60.0 M

Resolution
0.1 M 1 M 0.1 M

Test Current 1 mA @ 500 k 1 mA @ 1 M

Resistance Accuracy ±(% of Rdg + Digits) ±(2.0 % + 5 counts)
±(2.0 % + 5 counts)

Note
Throughout this manual "rotary switch" refers to the rotary function switch on the Meter.

Basic Maintenance
This basic maintenance section of the manual contains tests and procedures that require no equipment other than the Meter and some consumables such as fuses and batteries. Also, internal access is limited to the battery and fuse compartment.

Cleaning
When cleaning is necessary, wipe the Meter with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture on the terminals can affect readings.

Testing the Batteries Under Load
XWWarning To avoid electrical shock or personal injury, replace the batteries as soon as the battery indicator (b) appears. A weak battery can cause false readings.
Before testing the batteries, turn the Meter to B. If b appears on the display, the batteries are weak; replace them. If the display is blank, batteries may not be present in the Meter, or they may be completely discharged; install new batteries.
The following procedure tests the batteries under load:
1. Depress the h while turning the rotary switch to INSULATION.
2. Read the display. The displayed voltage should be 5.2 V or greater. If voltage is lower than 5.2 V, replace the batteries and repeat this test.

6

static awareness
A Message From
Fluke Corporation
Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by:
1. Knowing that there is a problem. 2. Learning the guidelines for handling them. 3. Using the procedures, packaging, and
bench techniques that are recommended.
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.

1. MINIMIZE HANDLING

3. DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES. USE A HIGH RESISTANCE GROUNDING WRIST STRAP.

2. KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE.

4. HANDLE S.S. DEVICES BY THE BODY.

5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT.

8. WHEN REMOVING PLUG-IN ASSEMBLIES HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES.

6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE.

9. HANDLE S.S. DEVICES ONLY AT A STATIC-FREE WORK STATION.
10. ONLY ANTI-STATIC TYPE SOLDERSUCKERS SHOULD BE USED.
11. ONLY GROUNDED-TIP SOLDERING IRONS SHOULD BE USED.

7. AVOID PLASTIC,VINYL AND STYROFOAM® IN WORK AREA.
PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC. AND GERNER DYNAMICS, POMONA DIV.
® Dow Chemical

Insulation Multimeters Basic Maintenance
Testing the Fuse
XWWarning To avoid electrical shock or injury, remove the test leads and any input signals from the Meter before replacing the fuse. Test the fuse as described below and illustrated in Figure 1. If the fuse is defective, see Replacing the Batteries and/or Fuse later in this manual.
1. Insert a test probe in the J input terminal. 2. Turn the rotary switch to N and verify the Meter is in Auto Range. 3. Insert the probe in the mA input terminal. If the display reading is 0L, the fuse is
defective; replace the fuse.

LO

HOLD

MIN MAX

TEST

440 mA
OK

Figure 1. Testing the Fuse

ebp01f.eps

7

1587/1577 Calibration
Replacing the Batteries and/or Fuse Use the following procedure and Figure 2 to replace the batteries and/or the fuse:
XWWarning To avoid electrical shock, personal injury, or damage to the Meter observe the following warnings: · Replace the batteries as soon as the battery indicator (b)
appears. A weak battery can cause false reading. · Use ONLY fuses with the amperage, interrupt, voltage, and
speed ratings specified. · Before removing the battery cover, turn the rotary switch to
OFF, and then, remove the test leads from the front-panel terminals. 1. Remove the yellow rubber boot from the Meter. Use the thumb-hole to press on the rear of the Meter and peel the boot from the Meter.
2. Using a standard screwdriver, release the battery door from the Meter; turn the battery-door lock until the unlock symbol aligns with the arrow.
3. Lift the bottom of the battery door away from the Meter to access the fuse and battery compartment.
4. Remove and replace the batteries and/or fuse as shown in Figure 2.
5. Re-install the battery door and secure it by turning the battery door lock until the lock symbol (L) aligns with the arrow.
6. Position the bottom of the Meter in the bottom of the boot, and press the top of the instrument firmly into of the boot.

F440 mA 1000V Min interrupt rating
10 000 A

Fuse, Fast, 440 mA, 1000 V, Min Interrupt Rating 10000 A

Fluke PN 943121

Battery, 1.5 V AA Alkaline, NEDA 15A, IEC LR6

Fluke PN 376756

Figure 2. Replacing the Batteries and Fuse

ebp02f.eps

8

Insulation Multimeters Basic Maintenance
Testing the Display The display is an LCD comprised of a series of characters and segments for providing the user with a variety of information. To test the LCD and all of its segments proceed as follows: 1. Set the rotary switch to OFF. 2. Depress h and turn the function switch to B. 3. While continuing to depress h, verify that the lit display segments match those shown in Figure 3.

Figure 3. LCD Display Test

bav01f.eps

Testing the Backlight The display backlight is a toggle function controlled by the H key. Each press of H causes the backlight to change states, i.e., on to off or off to on. To test the backlight, press H twice to verify that the toggle function is working. Turn the backlight off.
Testing the Keypad The keypad consists of the seven keys located above the rotary function switch. To test the keypad, turn the rotary switch to B and momentarily press each of the seven keys. Each press of an operational key will cause the Meter to beep. No beep in response to a key press indicates a defective keypad.
Reset the Meter by turning the rotary switch to OFF and, then, to any other position.

9

1587/1577 Calibration
Disassembling and Reassembling the Meter
This section of the manual provides instructions for disassembling and reassembling the Meter. The instructions are limited to major replaceable assemblies and do not include component-level detail. See Figure 4 for an exploded view of the major assemblies. Also, the emphasis is placed on disassembly. However, when appropriate, an italicized entry at the end of each disassembly procedure provides critical hints for reassembly.

Yellow Boot
Battery Door Four Screws
Bottom Case

Fuse

PCA Screw

Serial Number

Label

Plastic Shields Black Red

PCA
Light-dispersing Back Panel
Elastomeric Strip Gray Plastic Bezel Keypad

Figure 4. Disassembling the Meter

Top Case

ebp04f.eps

10

Insulation Multimeters Disassembling and Reassembling the Meter
Removing the Boot The standard Meter comes equipped with a snug-fitting yellow rubber boot. The boot helps protect the Meter from rough handling and is normally left on the Meter. The first step in disassembling the Meter is to remove the boot. Use the following procedure to remove the boot: 1. Looking at the front of the Meter, place your thumbs on the top corners of the boot and firmly grasp the Meter. 2. Using both thumbs, push the boot up and over both corners of the Meter. 3. Continue pushing on the boot until both of its inside corners are resting on top of the Meter. 4. Now, rest the heel of one hand behind the display, and place all four fingers of the same hand along the upper front edge of the boot. 5. Firmly grasp the Meter with the other hand, and using your fingers, peel the boot over the top of the Meter. 6. Slide the Meter up and out of the boot. When installing the boot, position the bottom of the Meter in the bottom of the boot, and press the top of the instrument firmly into of the boot.
Removing the Battery Door XWWarning
To avoid the risk of electrical shock, turn the rotary switch to OFF, and remove the test leads from the front-panel terminals before removing the battery cover.
With the boot removed, the next step in disassembling the Meter is to remove the battery door. Use the following procedure to remove the door: 1. Locate the black slotted lock on the lower rear of the Meter. 2. Using a standard screwdriver, turn the battery-door lock until the unlock symbol
(l) aligns with the arrow. The door is now unlocked. 3. Lift the bottom of the battery door away from the Meter. Removing the battery
door provides access to the fuse and battery compartment. 4. If necessary, remove and replace the batteries and/or fuse as shown in Figure 2. When installingl the battery door, first, slide the top of the door into position, and then, secure it by turning the battery door lock until the lock symbol (L) aligns with the arrow.
11

1587/1577 Calibration
Removing the Bottom Case With the battery door removed, the next step in disassembling the Meter is to remove the bottom case. Use the following procedure to remove the bottom case:
Note When removing the back case, it is not necessary to remove the fuse or the batteries. 1. Locate the four Phillips head screws on the bottom case, two next to the batteries
and two next to the fuse. 2. Using a Phillips screwdriver, remove the four screws. 3. Separate the front and bottom cases at the fuse-end of the Meter. Tilt the freed
end of the case up, and lift it away from the top case. To install the bottom case, first, position and press together the display-end of the top and bottom cases. Then, prior to installing all four screws, press the fuse-end of the cases together.
Removing the PCA With the bottom case removed, the next step in disassembling the Meter is to remove the printed circuit assembly (PCA). Refer to Figure 4 and use the following procedure to remove the PCA: 1. One Phillips screw attaches the PCA to the top case. Locate the screw near the center of the PCA, and remove it. 2. Two chrome-plated springs on the PCA form an electrical connection (clip) to the two recessed INSULATION(+) terminals on the top case. Break these connections by carefully pulling each of the springs back and away from the terminals. See Figure 5. 3. With one hand over the PCA, roll the top case over (face up) and lift it away from the PCA.
Note Three red and one black plastic shields are used to isolate the user from the input terminals. With the PCA removed, these shields are loose and can fall away from the PCA. 4. Remove and set aside all four shields for use during reassembly.
XWWarning To avoid risk of electrical shock, make sure the plastic input terminals are properly positioned on the PCA before attaching it to the top case. When installing the PCA, proceed as follows: 1. With the PCA face-up, place all four plastic shields into position on the PCA. 2. Lower the top case onto the PCA, and roll both parts over (PCA up). 3. Connect (snap) the chrome-plated springs to the INSULATION terminals. 4. Install the screw that attaches the PCA to the top case.
12

Insulation Multimeters Disassembling and Reassembling the Meter

Figure 5. Insulation Terminal Clips

ebp11f.eps

Removing the LCD With the PCA removed, the final step in disassembling the Meter is to remove the LCD assembly from the PCA. Refer to Figure 6 and use the following procedure to remove the LCD assembly:
1. Remove the two screws from the display end of the PCA.
2. Hold the PCA face down with the fuse-end of the PCA toward you. Locate the screw-hole near the center of the assembly. This hole is the one used to attach (with a screw) the PCA to the top case.
3. Now, locate the gray plastic tab just above and to the left of the hole. Using your thumbnail press the tab down and toward the display end of the PCA. This will release the LCD assembly from the PCA.
W Caution To avoid damaging the plastic guide pins on the LCD assembly, keep the LCD assembly parallel to the PCA when separating the two parts.
4. Without tilting the PCA, lift it straight up and away from the LCD assembly.
5. To keep loose parts from falling away from the LCD assembly, keep it facedown, and set it in a safe place.

Replacing the LCD The LCD assembly consists of four pieces as shown in Figure 6: · Translucent light-dispersing back panel · Flexible elastomeric conducting strip · Plastic Bezel for housing the assembly's components · Glass LCD display
With the LCD assembly removed from the PCA, use the following procedure to replace the LCD:
1. Lift the translucent light-dispersing back panel off of the plastic bezel.
2. Remove the elastomeric strip from its slot in the bezel.
13

1587/1577 Calibration

3. Remove the old glass LCD display from the bezel.
Note Make sure the new LCD display is clean (free of lint and finger prints) before placing it in position in the bezel.
4. Position the new glass LCD display in the bezel; the silver face should face the rear, and the stepped portion of the glass should be directly under the elastomeric slot on the bezel.
5. Drop the elastomeric strip into its slot on the bezel.
6. Position the translucent light-dispersing back panel over the silver side of the glass LCD display. Make sure the guide pins on the light dispersing back panel are facing up and that they are on the side opposite the elastomeric strip.
7. While holding the LCD assembly (face down) in one hand, position the PCA (fuse side up) over the bezel; match the guide holes in the PCA with the plastic guide pins on the bezel and the translucent light dispersing back panel. After the LCD assembly is in position, lock it in place by pressing (below the display) the bezel against the PCA; listen for the tab on the bezel to snap (lock) into position on the PCA.
8. Secure the PCA to the LCD assembly by installing the two screws that attach the PCA to the bezel. These screws also ensure contact between the glass LCD display, the elastomeric strip, and the PCA.

Screws

PCA

Light-dispersing Back Panel
Glass LCD Display
Elastomeric Strip

Gray Plastic Tab
Figure 6. Accessing the LCD

Plastic Bezel

ebp05f.eps

14

Insulation Multimeters Required Tools and Equipment

Reassembling the Meter
To reassemble the Meter, logically reverse the previous disassembly procedures. In the process, make sure to re-establish all electrical connections; specifically, the elastomeric strip for the LCD, the red and black plastic shields for the input terminals, and the two spring contacts for the Insulation Test Probe. Also make sure all parts are correctly aligned and positioned; do not force-fit any of the parts into position.

Required Tools and Equipment
Table 1 lists the require equipment for performing the Performance Test and Calibration adjustments. If a recommended model is not available, use a substitute with equivalent or better specifications.

Equipment Calibrator
DMM (Test DMM) Power Supply HV Probe (Divider) Thermocouple Thermocouple MiniConnector

Table 2. Required Tools and Equipment

Required Characteristics
AC Voltage Range: 0-1000V Accuracy: +/- 0.25% Frequency Range: 50Hz - 5 kHz Accuracy: +/-3% DC Voltage: 0-1000V Accuracy: +/-0.075% Current Range: 0-600 mA Accuracy: AC mA = +/-0.38% Frequency: 45Hz -1 kHz DC mA =+/-0.058% Frequency Source: 100Hz - 100 kHz Accuracy: +/-0.028% Amplitude: 5 Volt Accuracy: +/-5% Ohms Range: 0-50 M Accuracy: +/-0.23% Capacitance Range: 700 nF - 8000 uF Accuracy: +/-0.35% Insulation Resistance: 500 k - 1.7 G Accuracy: +/-0.9% DC Volts: 0 to 1000V Accuracy: +/-5% DC Current: 0 - 2mA Accuracy: +/-1% DC Volts: 0 to 15V DC Voltage Range: 1kV -5 kV Accuracy: 1% Division ratio: 1000:1 Input resistance: 1000 M K Type with Dual Banana connector K Type

Recommended Model Fluke 9100 w/135 option
Fluke 189 Generic Lab Supply Fluke 80k-6 Fluke 80BK Fluke 80CK-M

15

1587/1577 Calibration

Performance Test
The following series of tests comprise a performance test for verifying the accuracy of the Meter (UUT) and its performance level. The performance test is recommended as an acceptance test for incoming inspection and as a calibration procedure for periodically ensuring the accuracy of the Meter. Fluke recommends running the performance test at least once a year.
No adjustments are required during the performance test, and it is not necessary to open the case. If the Meter does not pass all parts of the performance tests, repair and/or calibration adjustment are required. A calibration adjustment procedure is given later in this manual. If significant repairs are required, contact Fluke as described toward the front of this manual. If user repairs are appropriate, refer to the list of user-replaceable parts toward the end of this manual
Testing the 1577 and 1587 DMM Functions Use the following procedure to verify the accuracy of the measurement functions. 1. Connect the Calibrator to the J and COM input terminals on the UUT.
2. Turn the rotary switch to the function listed in each step of Table 2.
3. Apply the input level for each step listed in Table 2.
4. Compare the reading on the UUT display with the Display Reading in Table 2.
If the display reading falls outside of the range shown in Table 2, the UUT does not meet specification.

Table 3. DMM Accuracy Tests

Step

UUT Function

UUT Range

UUT Input Level

UUT Display Reading

Fluke 1577

Fluke 1587

1 mV ac

600.0 mV 30 mV, 60Hz

29.4 to 30.6

29.4 to 30.6

2 V ac

6.000V

3V, 60Hz

2.937 to 3.063

2.967 to 3.033

3 V ac

6.000V

3V, 400Hz

n/a

2.937 to 3.063

Push G (blue) for 1 second for low-pass filter band-pass test

4 V ac

6.000V

3V, 400Hz

n/a

2.817 to 3.063

5 V ac

6.000V

3V, 800Hz

n/a

1.680 to 2.670

Push G (blue) for 1 second to remove low-pass filter

6 V ac

60.00V

60V, 50Hz

58.77 to 61.23

59.37 to 60.63

7 V ac

600.0V

600V, 50Hz

587.7 to 612.3

593.7 to 606.3

8 V ac

1000V

1000V, 50 Hz

977 to 1023

n/a

9 V ac

1000V

1000V, 1 kHz,

n/a

977 to 1023

10 Hz, V ac

99.99 Hz *10 Hz, 0.15V

n/a

9.98 to 10.02

11 Hz, V ac

99.99 kHz *95kHz, 0.15V

n/a

94.89 to 95.11

12 Hz, V ac

99.99 kHz *90 kHz, 2V

n/a

89.90 to 90.10

13 V dc

6.000V

0.3V

0.298 to 0.302

0.298 to 0.302

14 V dc

6.000V

6V

5.986 to 6.014

5.993 to 6.007

15 V dc

60.00V

60V

59.86 to 60.14

59.93 to 60.07

16

Insulation Multimeters Performance Test

Table 3. DMM Accuracy Tests (cont)

Step

UUT Function

UUT Range

UUT Input Level

UUT Display Reading

Fluke 1577

Fluke 1587

16 V dc

600.0V

-600.0V

-601.4 to -598.6 -600.3 to -599.7

17 V dc

1000V

1000V

996 to 1004

997 to 1003

18 mV dc

600.0 mV 3 mV

2.9 to 3.1

2.9 to 3.1

19 mV dc

600.0 mV 30 mV

29.9 to 30.1

29.9 to 30.1

20 mV dc

600.0 mV -600 mV

-601.3 to 598.7 -600.7 to -599.3

21 Ohms

600.0 

0

22 Ohms

600.0 

600 

23 Ohms

6.000 k 6000 

24 Ohms

60.00 k 60.00 k

25 Ohms

600.0 k 600.0 k

26 Ohms

6.000 M 6.000 M

27 Ohms

50.00 M 50.00 M

28 Capacitance 1000 nF

700 nF

-0.2 to 0.2 592.6 to 607.4 5.926 to 6.074 59.26 to 60.74 592.6 to 607.4 5.926 to 6.074 48.97 to 51.03
n/a

-0.2 to 0.2 594.4 to 605.6 5.944 to 6.056 59.44 to 60.56 594.4 to 605.6 5.944 to 6.056 49.22 to 50.78 690 to 710

29 Capacitance 9999 uF

8000 uF

n/a

7814 to 8186

30 Continuity

n/a

103

Beeper off

Beeper off

31 Continuity

n/a

23

Beeper on

Beeper on

32 Diode

n/a

2.5 V

n/a

2.447 to 2.553

33 mA ac

60.00

3 mA, 45Hz

2.92 to 3.08

2.93 to 3.07

34 mA ac

600.0

400 mA, 60Hz

391.8 to 408.2

393.8 to 406.2

35 mA dc

60.00

60.00 mA

59.38 to 60.62

59.86 to 60.14

36 mA dc

600.0

400.0 mA DC

395.8 to 404.8

399 to 401

* To ensure accurate frequency use the Calibrator Frequency mode.

17

1587/1577 Calibration
Testing Temperature Function (1587 Only) Use the following procedure to verify that the UUT measures temperature within the published specification. 1. Connect the K-Type thermocouple to the temperature input of the UUT (J) and temperature calibrator.
Note To ensure an accurate measurement the UUT and thermocouple connector must be at the same temperature. 2. After connecting the thermocouple to the UUT and the Calibrator, allow the
junctions to stabilize before recording the display reading. Depending on temperature gradients this could take several minutes. 3. Turn the rotary switch to D, and press G to select temperature mode. 4. Put the calibrator in the thermocouple function, and set the output temperature to 23 °C. The UUT display reading should be within the limits of 21.8 °C to 24.2 °C.
Testing the Discharge Circuit The following Discharge Circuit Test is a safety-related test that verifies input jack wiring to the PCA, the RSOB contacts, RSOB pads on the PCA, and other active discharge components on the PCA.
W Caution To prevent damage to the Test DMM DO NOT press t during the following procedure. 1. Connect the Test DMM to the INSULATION terminals (+ and -) on the UUT. 2. Set the Test DMM to measure resistance. 3. Turn the rotary switch to INSULATION. 4. Verify that the Test DMM reading is between 2500  and 3500  .
Testing the Insulation Function
XWWarning To avoid electric shock hazard at the INSULATION output terminals, do not press t when the rotary switch is set to INSULATION.
W Caution To prevent damage to the calibrator do not use the 5500A or other standard calibrators for the insulation tests.
W Caution To prevent damage to the 9100A make sure it is in the Insulation Test mode prior to pushing t.
Insulation Resistance Accuracy Test To test the Insulation Resistance accuracy, sequentially complete the test steps in Table 4, using the following procedure. 1. Connect the INSULATION output terminals on the UUT to the calibrator. 2. Set the calibrator to the Insulation Resistance Test Mode.
18

Insulation Multimeters Performance Test

3. Turn the rotary switch to INSULATION. 4. Set the voltage range on the UUT; see Table 4. 5. Apply the Calibrator Output to the UUT; see Table 4. 6. Press t on the UUT. 7. Verify that the display reading is within the limits given in Table 4.

Step
1 2 3 4 5
6 7 8 9 10 11 12

Table 4. Insulation Resistance Accuracy Test

UUT Voltage Range

Calibrator Resistance
Range

Calibrator Output

UUT Display Reading

Limit Low

Limit High

500 V 500 V 500 V 1000 V 1000 V
1000 V 50 V 50 V 100 V 100 V 250 V 250 V

60.0 M 500 M 500 M 2.0 G 600 M

0.500 M 450 M 600 M 1.7 G 600 M

1587 Only

60.0 M 6.00 M 50.0 M 6.00 M 100 M 60.0 M 250 M

1 M 100 k 45 M 100 k 95 M 250 k 225 M

0.0 438 >550 1.2 586
0.5 0.05 43.1 0.05 87 0.0 217

1.0 462 >550 2.2 614
1.5 0.15 46.9 0.15 103 0.8 233

Insulation Function, External Sense The following test verifies that the UUT will sense a voltage > 30 V when present on the circuit under test.
1. Connect the UUT INSULATION output terminals to the calibrator voltage output terminals.
2. Turn the rotary switch to the INSULATION function.
3. Apply 35 V, 50 Hz to the UUT.
4. Verify that the UUT displays > 30 V in the primary display.

19

1587/1577 Calibration
Insulation Function, Source Voltage Accuracy Test The following test verifies the accuracy of the source voltage for the insulation function.
WCaution To prevent damage to the Test DMM use a HV Probe (Divider) with the Test DMM when testing the 1000V range of the UUT.
1. Connect the output of the HV Probe to the V ­ Com inputs on the Test DMM and turn the rotary switch to E.
2. Connect the input of HV Probe to the INSULATION output terminals on the UUT as shown in Figure 7.
3. Set the rotary switch to INSULATION.
4. Refer to Table 5, and complete steps 1 through 5. For each step, select the specified range on the UUT, press t, and verify that the voltage readings on both the UUT (lower right hand corner) and the DMM are within the limits shown in the table.

UUT

LO

HOLD

MIN MAX

TEST

INSULATION

Test DMM
18

SETUP

HOAuLtoDHOLD

HnzS%CAMNmCIsNELMAXRANSGAREVEELLOGGING NO

YES

ac+dc mV V ac+dc dB mV

CF

dB V

mAA

OFF
A mA A

A A macA+Adc CLEAVRIEMWEM ac+dc

COM

TEMPERATURE

V

Figure 7. Source Voltage Accuracy Test

ebp06f.eps

Table 5. Insulation Mode, Source Voltage Accuracy Tests

Step Function

Range

Test DMM Display Reading

UUT Display Reading

1 Insulation 2 Insulation

500 V 1000V

3 Insulation 4 Insulation 5 Insulation

50V 100V 250V

500.0 to 600.0 1000.0 to 1200.0
1587 Only 50.0 to 60.0 100.0 to 120.0 250.0 to 300.0

500 to 600 1000 to 1200
50 to 60 100 to 120 250 to 300

20

Insulation Multimeters Performance Test
I Nominal Test The following test verifies the ability of the UUT to maintain the nominal insulation test current while loaded. 1. Connect the Test DMM and Calibrator to the UUT INSULATION terminals as shown in Figure 8. 2. Set the Calibrator to the Insulation Resistance Test Mode. 3. Set the output resistance on the Calibrator 1 M . 4. Set the Test DMM to measure A DC 5. Set the UUT to INSULATION, 1000V range. 6. Press t on the UUT. The Test DMM reading should be greater than 1000.0 A.

UUT

LO

HOLD

MIN MAX

TEST 1000V

DANGER sHi

HIGH VOLTAGE sLo

15Vpk max

!

!

Hi

Lo

1500Vpk max

I+

I-

15Vpk max

Calibrator

Test DMM

A B C D E F CLR 7 89
GH I J KL _ / 4 56E

MN OP QR 1 23

ST 0

UV

WX Ñ

YZ

V OUTPUT
A OFF


Hz

Aux

ON

Mode

A

189 LOGGING MULTIMETER

SETUP

HOAuLtoDHOLD

HnzS%CAMNmCIsNELMAXRANSGAREVEELLOGGING NO

YES

ac+dc mV V ac+dc dB mV

CF

dB V

mAA

OFF
A mA A

A A macA+Adc CLEAVRIEMWEM ac+dc

COM

TEMPERATURE

V

>1000 A

Figure 8. I Nominal Test

ebp07f.eps

Low-Battery Test - DMM Function The following test verifies the accuracy of the low-battery indicator (b) on the UUT during DMM operations.
1. Remove the batteries from the battery compartment on the UUT.
2. Connect a 6-volt adjustable Power Supply between the positive and negative battery terminals of the UUT as shown in Figure 9.
3. Turn the rotary switch to C.
4. Slowly reduce the Power Supply output while observing the display on the UUT. Verify that the low-batter indicator (b) appears when the voltage is at 4.0 V
5. Now slowly increase the Power Supply output while observing the display on the UUT. Verify that the low-battery indicator (b) on theUUT goes out at 4.4 V.
6. Slowly reduce the Power Supply output while observing the display on the UUT. Verify that the UUT displays bAtt at 3.5 V.
7. Turn the rotary switch to OFF to reset the display.

21

1587/1577 Calibration

Power Supply

6.0 Vdc

UUT (Rear)

Power Supply

UUT (Front)
LO

A

B

Figure 9. Low-Battery Test Connections

ebp08f.eps

Low-Battery Test - INSULATION Function The following procedure verifies the accuracy of the low-battery indicator (b) during insulation operation.
XWWarning To avoid shock hazard DO NOT press t during the following procedure.
1. Remove the batteries from the battery compartment.
2. Connect a 6-volt adjustable Power Supply between the positive and negative battery terminals of the UUT as shown in Figure 9-A.
3. Set the Power Supply output to 6 V, and turn the rotary switch to INSULATION.
4. Slowly turn the Power Supply output down while observing the display on the UUT. Verify that the low-battery indicator (b) appears when the voltage is at 5.2 V
5. Continue to reduce the Power Supply output while observing the display on the UUT. Verify that bAt appears on the display when the supply voltage reaches 4.0 V.
Note bAt will appear in the lower display. bAtt will still appear in the prinmary display when the supply reaches 3.5 V just as it does in DMM functions.
6. Turn the rotary switch to OFF to reset the display.
7. Remove the Power Supply, and install the batteries and the battery door.

22

Insulation Multimeters Calibration Adjustment Procedure

Calibration Adjustment Procedure
The following sections comprise the Calibration Adjustment Procedure. The procedure is meant to bring the UUT back into specification following repair of the UUT or when the UUT fails the Performance Test. The required equipment is listed earlier in Table 2.
The Meter features closed-case calibration adjustment using known reference sources. During the procedures, known reference source values are applied to the meter; the Meter calculates correction factors and stores them in nonvolatile memory.

Calibration Adjustment Counter The Meter contains a calibration adjustment counter. The counter is incremented each time a Calibration Adjustment Procedure is completed. The value in the counter can be recorded and used to show that no adjustments have been made during a calibration cycle.
Use the following steps to view the calibration counter on the UUT.
1. While holding down H on the UUT, turn the rotary switch from OFF to B. The UUT should display Z CAL.
2. Press h once to see the calibration counter. For example n001.
3. Turn the rotary switch to OFF.

Calibration Adjustment Password
To start the Calibration Adjustment Procedure, the correct four-digit password must be entered. The password can be changed or reset to the default as described in following paragraphs. The default password is 1234.

Changing the Password Use the following steps to change the password on the UUT:
1. While holding down H on the UUT, turn the rotary switch from OFF to B. The UUT displays Z CAL.
2. Press h once to see the calibration counter. 3. Press h again to start the password entry. The UUT displays ?.?.?.?.
The following keys represent the digit indicated below when entering or changing the password:

h = 1 H = 3

m Unused f Unused t = 4

r = 2 G = 5

4. Press the four keys to enter the old password. If changing the password for the first time, enter h (1), r (2), H (3), and t (4).
5. Press r to change the password. The UUT displays //// if the old password is correct. If the password is not correct, the UUT emits a double beep, displays ?.?.?.? and the password must be entered again. Repeat step 4.
6. Press the four keys representing the new password.
7. Press h to store the new password.

Restoring the Default Password If the calibration password is forgotten, the default password (1234) can be restored using the following steps. 1. Turn the rotary switch from OFF to B.
23

1587/1577 Calibration

2. Remove the back case from the UUT. Leave the PCA in the top case. 3. Apply 6.0 V across the battery pads (XBT1) + and ­ on the back of the PCA. See
Figure 10.
XWWarning To avoid electrical shock or personal injury, remove the test leads and any input signal before removing the back case from the UUT. 4. Short across the CAL keypad on the back of the PCA. See Figure 10. The UUT should beep. The default password is now restored. 5. Remove the 6.0 V supply and install the back case on the UUT.
Cal Keypad

Battery Pads

Figure 10. Restoring the Default Password

ebp10f.eps

24

Insulation Multimeters Calibration Adjustment Procedure

Meter Keys Used in the Calibration Steps

The Meter keys behave as follows when performing the Calibration Adjustment Procedure. This may be of help determining why a calibration step is not accepted and for determining the input value without referring to Table 6.

H

Press and hold to show the measured value. The measurement value is not

calibrated so it may not match the input value. This is normal.

t Press and hold to display the required input amplitude.

h

Press and hold to display the frequency of the required input.

G

Store the calibration value and advance to the next step. This key is also

used to exit the calibration mode after the calibration-adjustment sequence

is complete.

Calibration Adjustment
Use the following steps to make calibration adjustments to the UUT. Complete the adjustment procedure before turning the UUT off; otherwise, the new calibration constants will not be saved.
1. While holding down H, turn the rotary switch from OFF to B. The UUT displays Z CAL.
2. Press h once to see the calibration counter, e.g., n0001.
3. Press h again to start the password entry. The UUT displays ?.?.?.?.
4. Press four keys to enter the password.
5. Press h to go to the first calibration step. The UUT displays C-01 if the password is correct. If the password is not correct, the UUT emits a double beep, displays ?.?.?.? and the password must be entered again. Repeat step 4.
6. Using Table 6, apply the input value listed for each calibration adjustment step. For each step, position the rotary switch and apply the input to the terminals as indicated in the table.
7. After each input value is applied, press G to accept the value and proceed to the next step (C-02 and so forth).
Note After pressing G, wait until the step number advances before changing the calibrator source or turning the rotary switch. If the rotary switch is not in the correct position, or if the measured value is not within the anticipated range of the input value, the UUT emits a double beep and will not continue to the next step. Some adjustment steps take longer to execute than others (10 to 15 seconds). For these steps, the UUT will beep when the step is complete. Not all steps have this feature.
8. After the final step, the display shows End to indicate that the calibration adjustment is complete. Press G to go to meter mode.
Note Set the calibrator to Standby prior to changing the function switch position and/or after completing adjustment of each function. If the calibration adjustment procedure is not completed correctly, the UUT will not operate correctly.

25

1587/1577 Calibration

Switch Position (Function)
AC Volts
DC Volts DC Millivolts Ohms
Continuity Diode Test milliamps
Insulation

Table 6. Calibration Adjustment Steps

Input Terminal

Calibration Adjustment Step

V/COM

C-01 C-02 (Model 1587 only) [1]

C-03

C-04 (Model 1587 only) [1]

C-05

C-06 (Model 1587 only) [1]

C-07

C-08 (Model 1587 only) [1]

C-09

C-10

C-11

C-12

C-13

C-14

C-15

C-16

C-17

C-18

C-19

C-20

C-21

C-22(Model 1587 only) [1]

mA/COM

C-23

C-24

C-25

C-26

mA/COM

C-27

Input Value
600.0 mV, 60 Hz 600.0 mV, 5 kHz 6.000 V, 60 Hz 6.000 V, 5 kHz 60.00 V, 60 Hz 60.0 V, 5 kHz 600.0 V, 60 Hz 600.0 V, 2 kHz 6.000 V, 0 Hz 60.00 V, 0 Hz 600.0 V, 0 Hz 600.0 mV, 0 Hz 60.00 mV, 0 Hz 600.0  6.000 k 60.00 k 600.0 k 6.000 M 0.000  50.0 M 600.0  6.000 V, 0 Hz 60.00 mA, 60 Hz 400.0 mA, 60 Hz 60.00 mA, 0 Hz 400.0 mA, 0 Hz 1.0 mA, 0 Hz

XWWarning
To avoid electrical shock hazard, remove all test leads from the UUT before performing the following step.

Insulation

None [2]

C-28

None (press BLUE to start internal cal)

Notes: 1. Calibration step will be skipped on Model 1577. 2. Internal HV output calibration. Ensure no lead is inserted in Insulation (+) jack.

26

Insulation Multimeters Service and Parts
Service and Parts
User service is limited to replacing parts. Table 7 identifies the parts available for replacement and Figure 11 shows the location of each part. To order replacement parts refer to Contacting Fluke earlier in this manual.

Figure 11. 1587/1577 Replacement Parts

ebp09f.eps

27

1587/1577 Calibration

Table 7. 1587/1577 Replacement Parts

Item

P/N

Qty Ref Des

Description

1 2277944 1

MP1

1577, Case Top

1 2277926 1

MP1

1587, Case Top

2 2277971 1

MP2

1577, Bracket, Mask

2 2277967 1

MP2

1587, Bracket, Mask

3 2281339 1

MP3

1577, Keypad

3 2281342 1

MP3

1587, Keypad

4 2277892 1

MP4

15x7, Shield, Bottom

5 2386438 1

MP5

1577, Door, Battery

5 2277998 1

MP5

1587, Door, Battery

6 2278007 1

MP6

15x7, Knob

7 2168609 1

MP7

15x7, Backlight

9 2277903 1

MP9

15x7, Bottom Case

10 2278018 1

MP10

15x7, Housing Assy, RSOB

11 2278029 1

MP11

15x7, Detent Spring

12 2278128 3 MP12-14 15x7, Insulator, Receptacle, Red

13 2278137 1

MP15

15x7, Insulator, Receptacle, Black

14 2278143 1

MP16

15x7, Tilt Stand

16 2278162 1

MP18

15x7, Holster

17 2281317 1

MP19

15x7, Battery Contact, Negative

18 2281321 1

MP20

15x7, Battery Contact, Pos

19 2141295 1

MP21

15x7, LCD, Display

20 666435

3 MP22-24 15x7, Battery Contact, Dual

21 376756

4 MP40-43 15x7, Battery, 1.5 V, 2.24 AH, 15 A, Alkaline, AA,14X50 mm

22 832246

4

H26-29 15x7, Screw, 5-14, 0.750, Pan, Phillips, Blk, Thread Forming

23 642931

3

H30-32 15x7 Screw, 4-14, 0.312, Pan, Phillips, Zinc, Thread Forming

24 822676

1

25 943121

1

26* 2070140 1

27* 1273113 1

MP32 W F1 MP26 MP25

15x7, Contact, PTF 15x7, Fuse, 0.406X1.375 Inches, 0.440A, 1000V, Fast 15x7, TL224 Test Leads RA2S 15x7, Thermocouple Assembly, Type K, 3 feet

28* 2000757 1 30* 2099044 1

MP28 MP30

15x7, Probe, Multifunction 15x7, Probe, Test, Banana Jack, 4mm Tip, Red

31* 2427138 1

MP31

15x7, Probe, Test, Banana Jack, 4mm Tip, Black

32* 1958646 1 33* 1958654 1 34 2396462 1

MP47 MP33 MP34

15x7, Alligator Clip, Ex-Large Blk, IEC1010 15x7, Alligator Clip, Ex-Large Red, IEC1010 15x7, Connector, Elastomeric

37 2401066 1 41* 2416504 1 43* 1942029 1 44 2416177 1 45* 2401027 1 46* 2456783 1 47* 2401011 1 * Not Shown

MP37 MP51 MP53 MP54 MP55 MP56 MP57

15x7, Absorber, Shock 15x7, Hard Case, Molded 15x7, Probe Cap, Gs-38 Red 15x7, Shield, Top 15x7 Manual, Users 15x7 Manaul. Calibration 15x7, CD, Manual Set

28

