Catalyst Dx Chemistry Analyzer Quick Reference Guide

Catalyst Dx Chemistry Analyzer Quick Reference Guide

Note: For detailed instructions on using your Catalyst Dx* Chemistry Analyzer, see the IDEXX Catalyst Dx* Chemistry Analyzer Operator's Guide.

Catalyst Dx Analyzer Components

The diagram illustrates the Catalyst Dx Analyzer, labeling key components such as the touch-screen interface, sample drawer, waste drawer, maintenance access doors, and tip/diluent drawer(s).

Touch-screen interface: Displays the graphical user interface of the instrument.

Sample drawer: Holds new slides, sample cups, and whole blood separators.

Waste drawer: Holds used tips and slides.

Maintenance access doors: Provide access to the maintenance areas of the analyzer.

Tip/diluent drawer(s): Holds up to 12 tips and all diluent and wash materials.

Sample Volume Required

Whole Blood

Fill to the lowest line on the separator (0.7 cc [700 μL]).

Serum/Plasma/Urine

300 μL is the recommended sample volume when running a Chem 17 CLIP and electrolytes.

Note: The minimum sample volume required varies depending on the number of slides being run.

Running a Sample

  1. Enter the sample information on the IDEXX VetLab* Station.
  2. Once the sample information appears in the Pending list on the Catalyst Dx Home screen, tap it and then tap Select.
  3. Select the Sample Type.
  4. If you are running a NH3 slide, UPC slides, or PHBR slide, select the applicable check box.
  5. Specify the desired type of dilution, and then use the [up/down arrows] to specify the desired dilution factor (total parts).
  6. Tap Next.
  7. Load the slides and sample in the sample drawer. Note: If you are running a Lyte 4 CLIP, be sure to load it in the sample drawer before any other CLIPs or slides.
  8. Tap Run. When the analysis is complete, the results are sent to the IDEXX VetLab Station automatically.

Dilutions

The Catalyst Dx analyzer supports automated dilutions (the analyzer mixes the sample and diluent for you) and manual dilutions (you prepare the dilution outside of the analyzer).

Note: Do not dilute samples that are undergoing NH3, PHBR, FRU, TT4, SDMA, PL, PROG, BA, or electrolyte testing. Do not perform an automated dilution on CRP samples.

Special Slides (NH3, UPC, or PHBR)

  • If you are running a UPC panel or a PHBR slide, do not load any other CLIPs or slides in the sample drawer.
  • If you are running a NH3 slide, be sure to load it in the sample drawer before any other CLIPs or slides.
  • If you are running a UPC panel, be sure to load both the UCRE and UPRO slides into the sample drawer.
  • If you are running a special slide without selecting the applicable special slides check box and/or you do not follow the on-screen instructions, your results will be flagged and you may receive inaccurate results.

Opening the Maintenance Access Doors

  1. Press up on the door panel below the touch screen.
  2. If the area above the door panel has vertical plastic slats (see photo 2a below), push down firmly on the door panel. The maintenance access doors are released. OR If the area above the door panel has a metal handle (see photo 2b below), pull down on both sides of the metal handle above the door panel until you hear a click. The maintenance access doors are released.
  3. Place a finger beneath the center of the maintenance access doors and push up until the doors lock in place.

Visual aids (photos 2a, 2b, 3) demonstrate the different mechanisms for opening the doors.

Closing the Maintenance Access Doors

  1. If the area above the door panel has vertical plastic slats (see photo 1a below), push and hold down on the door panel. The maintenance access doors close automatically. OR If the area above the door panel has a metal handle (see photo 1b below), pull down on both sides of the metal handle above the door panel until you hear a click. The maintenance access doors close automatically.
  2. Press up on the door panel below the touch screen until it clicks.

Visual aids (photos 1a, 1b, 2) demonstrate the different mechanisms for closing the doors.

Cleaning and Maintenance

Monthly

  • Clean optics windows, incubator ring, and white reference tile.
  • If there is sample on the centrifuge shield, remove the shield from the analyzer and gently clean it with mild soap and water to remove the residue. Rinse, dry thoroughly, and replace.
  • Run quality control on the analyzer after cleaning.

Quarterly

Remove the fan filter and vacuum it thoroughly to remove dirt or dust.

As Needed

  • Upgrade your analyzer software promptly.
  • Clean the internal components when prompted by the analyzer.
  • Clean the outside of the analyzer with a damp (not wet) lint-free cloth. A mild liquid soap will remove grease.
  • Always disconnect the power cable before cleaning the analyzer.
  • If the screen gets dirty, apply an antistatic screen cleaning agent (NOT ammonia-based) to a clean cloth or paper towel and wipe the screen. Do not spray the cleaner directly onto the screen.
  • Clean the interior of the waste drawer with a lint-free cloth dampened with 70% isopropyl alcohol.

To Clean the Internal Components

  1. On the Catalyst Dx touch screen, tap Tools and then tap Clean Analyzer.
  2. Open the maintenance access doors.
  3. Remove the black whole blood separator carriers, any whole blood separators or sample cups from the sample drawer, and the white centrifuge shield. Then, clean the black carriers and white shield with an IDEXX-supported alcohol prep pad and return them to their positions.
  4. Lift the sample drawer assembly by pressing the center latch and lifting up.
  5. Remove the carousel by lifting up.
  6. Using an IDEXX-supported alcohol prep pad, wipe the incubator ring track in a counterclockwise direction (do not wipe the optics or ion windows at this time). Repeat this step at least three times using a new alcohol prep pad for each wipe.
  7. Clean the optics, ion windows, and reference tile on the carousel using the instructions in step 6.
  8. Using an IDEXX-supported dry optical tissue, dry the optics, ion windows, and reference tile, ensuring all signs of dampness have evaporated from the cleaned components. If streaks or smudges remain, repeat the cleaning process.
  9. Replace the carousel.
  10. Lower the sample drawer assembly and ensure that it is locked into place.
  11. Close the maintenance access doors.
  12. On the Catalyst Dx touch screen, tap Done. The analyzer initializes (approximately 5 minutes).

Visual aids show the black whole blood separator carrier, white centrifuge shield, sample drawer assembly, and carousel components.

Running Quality Control

  1. Tap the Catalyst Dx icon on the IDEXX VetLab Station Home screen.
  2. Tap Quality Control.
  3. Tap the quality control lot number you are using and tap Run QC.
  4. Tap the QC information in the Pending list on the Catalyst Dx Home screen and then tap Load.
  5. Follow the on-screen instructions for loading the quality control materials and then tap Run. Once the results are complete, you can view the results using these methods:
    • Tap View in the View Results message (this message displays by default when a run is complete) to access the Records: Test Results screen.
    • Tap the QC run in the Results list.
    • Tap Records Search on the IDEXX VetLab Station Home screen, search for and select the QC run you want to view, and then tap View.
Models: Catalyst Dx Chemistry Analyzer, Dx Chemistry Analyzer, Chemistry Analyzer, Analyzer

PDF preview unavailable. Download the PDF instead.

catalyst-dx-quick-ref-guide-en Adobe PDF Library 17.0 PDFsam Enhanced 7

Related Documents

PreviewCatalyst Dx Chemistry Analyzer Quick Reference Guide
A concise guide for operating and maintaining the IDEXX Catalyst Dx Chemistry Analyzer, covering sample preparation, analysis, dilutions, and cleaning procedures.
PreviewCatalyst SmartQC Control Quick Reference Guide
A quick reference guide for Catalyst SmartQC Control, detailing storage and handling requirements, and frequently asked questions about its use with Catalyst One and Catalyst Dx analyzers.
PreviewIDEXX ProCyte Dx Hematology Analyzer Quick Reference Guide
A concise guide for operating the IDEXX ProCyte Dx Hematology Analyzer, covering analyzer components, sample preparation, running patient samples and quality control, exiting standby mode, monthly rinse procedures, powering on/off, and changing reagent kits and stain packs.
PreviewIDEXX Catalyst One Chemistry Analyzer Operator's Guide
This operator's guide provides comprehensive instructions for using the IDEXX Catalyst One Chemistry Analyzer. It covers setup, operation, sample preparation, quality control, maintenance, and technical specifications for veterinary diagnostic use.
PreviewIDEXX ProCyte Dx Hematology Analyzer Operator's Guide
Comprehensive guide for operating, maintaining, and troubleshooting the IDEXX ProCyte Dx Hematology Analyzer, detailing its features, sample analysis, quality control, and technical specifications for veterinary use.
PreviewCatalyst SDMA Test Quick Reference Guide
A quick reference guide for the Catalyst SDMA Test, covering storage, handling, and frequently asked questions about its use with Catalyst One and Catalyst Dx analyzers.
PreviewCatalyst One Chemistry Analyzer: Quick Reference Guide for Veterinary Labs
A quick reference guide for the IDEXX Catalyst One Chemistry Analyzer, covering component identification, sample preparation, running samples, quality control, monthly cleaning, and available chemistry tests with CLIP compatibility.
PreviewIDEXX VetLab UA Analyzer Operator's Guide - Veterinary Urinalysis Testing
Comprehensive operator's guide for the IDEXX VetLab UA Analyzer, detailing installation, operation, maintenance, and troubleshooting for veterinary urinalysis testing.