DreamStation 2
CPAP Auto CPAP User manual

© Koninklijke Philips N.V., 2021. All rights reserved.

Contents 1. Safety Information..................................................................................................................................................................... 1
Intended use........................................................................................................................................................................... 1 Important................................................................................................................................................................................. 1 Contraindications.................................................................................................................................................................. 1 Warnings..................................................................................................................................................................................2 Cautions...................................................................................................................................................................................6 Symbols glossary................................................................................................................................................................. 9 System contents.................................................................................................................................................................. 10 How to contact Philips Respironics................................................................................................................................ 11 2. System overview....................................................................................................................................................................... 11 Device with integrated humidifier...................................................................................................................................12 Humidifier water tank........................................................................................................................................................ 14 3. Device Setup............................................................................................................................................................................. 16 Where to place the device................................................................................................................................................ 16 Supplying AC power to the device................................................................................................................................. 16 Setting up the integrated humidifier............................................................................................................................. 18 Removing the humidifier water tank............................................................................................................................. 19 Filling and connecting the water tank..........................................................................................................................20 Installing/replacing the air filters................................................................................................................................... 22 Connecting the breathing circuit....................................................................................................................................24 Starting the device.............................................................................................................................................................. 25 Performance check.............................................................................................................................................................26 Bluetooth wireless technology.......................................................................................................................................26 Pairing your therapy device to your Bluetooth-enabled mobile device............................................................ 27 Therapy Off display.............................................................................................................................................................31 Device settings (language, time, airplane mode, Bluetooth reset)...................................................................... 32 4. Pop-up messages and troubleshooting...........................................................................................................................33 Device pop-up messages.................................................................................................................................................33 Alerts and notifications.....................................................................................................................................................33 Status pop-up messages..................................................................................................................................................36 Troubleshooting.................................................................................................................................................................. 37 5. Connectivity..............................................................................................................................................................................39 6. Accessories...............................................................................................................................................................................40 SD card...................................................................................................................................................................................40 Updating software using an SD card.............................................................................................................................40 Oximeter................................................................................................................................................................................ 41 Roll stand..............................................................................................................................................................................42 Supplemental oxygen......................................................................................................................................................................42 DC power...............................................................................................................................................................................42 Carrying case and airline travel......................................................................................................................................43
- i -

7. Cleaning, disinfection, and care..........................................................................................................................................44 Home cleaning: device exterior, heater plate, and humidifier air inlet/outlet seal.........................................44 Home cleaning: humidifier water tank (lid, water tank base, and water tank seal) ........................................45 Home disinfection: humidifier water tank (lid, water tank base, and water tank seal)..................................46 Requirements:......................................................................................................................................................................46 Home cleaning: tubing......................................................................................................................................................46 Caring for the filters............................................................................................................................................................ 47 Service.................................................................................................................................................................................... 47
8. Technical specifications........................................................................................................................................................48 Environmental.....................................................................................................................................................................48 Physical.................................................................................................................................................................................48 Service life.............................................................................................................................................................................48 Standards compliance ......................................................................................................................................................48 IEC 60601-1 classification................................................................................................................................................49 Electrical................................................................................................................................................................................49 Water capacity.....................................................................................................................................................................49 Heater plate..........................................................................................................................................................................49 Humidity................................................................................................................................................................................49 Declared dual-number noise emissions values (in accordance with ISO 4871)...............................................50 Pressure accuracy...............................................................................................................................................................50 Maximum flow rate (typical).............................................................................................................................................51 ISO 80601-2-70 Pneumatic Diagram............................................................................................................................51 Wireless notices................................................................................................................................................................... 52 Radio Specifications..........................................................................................................................................................53 Cellular................................................................................................................................................................................... 53 Bluetooth............................................................................................................................................................................... 53 Intake Port Filters................................................................................................................................................................53 Heated Tubing Specifications..........................................................................................................................................54 Disposal................................................................................................................................................................................. 54
9. EMC information.....................................................................................................................................................................55 FAA Compliance...........................................................................................................................................................................59 Limited Warranty......................................................................................................................................................................... 61
- ii -

1. Safety Information
Intended use The DreamStation 2 CPAP/DreamStation 2 Auto CPAP system delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30 kg (66 lbs). It is for use in the home or hospital/institutional environment.
Important The device is to be used only on the instruction of a licensed physician. Your home care provider will set the correct pressure settings and will provide accessories according to your health care professional's prescription. Several accessories are available to make your OSA treatment with the DreamStation 2 system as convenient and comfortable as possible. To ensure that you receive the safe, effective therapy prescribed for you, use only Philips Respironics approved accessories.
Contraindications When assessing the relative risks and benefits of using this equipment, the clinician should understand that this device can deliver pressures up to 20 cmH2O. In the event of certain fault conditions, a maximum pressure of 40 cmH2O is possible. Studies have shown that the following pre-existing conditions may contraindicate the use of CPAP therapy for some patients: · Bullous Lung Disease · Pathologically Low Blood Pressure · Bypassed Upper Airway · Pneumothorax · Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway
Pressure. Caution should be used when prescribing CPAP for susceptible patients such as those with: cerebral spinal fluid (CSF) leaks, abnormalities of the cribriform plate, prior history of head trauma, and/or pneumocephalus. (Chest 1989; 96:1425-1426) The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a sinus or middle ear infection. Not for use with patients whose upper airways are bypassed. Contact your health care professional if you have any questions concerning your therapy.
- 1 -

Warning: Use only the cleaning methods outlined in your user manual. Philips is unable to verify the safety or performance of any device if ozone or other unapproved cleaning and disinfection methods are used.

Warnings A warning indicates the possibility of injury to the user or operator.

Device usage This device is not intended for life support.

Contact your health care professional if symptoms of sleep apnea recur.

Personnel

This manual serves as a reference. The instructions in this manual are not intended to

qualifications supersede your health care professional's instructions regarding the use of the device.

The prescription and other device settings should only be changed on the order of the supervising physician.

The operator should read and understand this entire manual before using the device.

Operating

Do not operate the device in direct sunlight or near a heating appliance because these

temperature conditions can increase the temperature of the air coming out of the device.

Bacteria filter If the device is used by multiple persons in a hospital environment (such as rental devices), a low-resistance, main flow bacteria filter should be installed in-line between the device and the circuit tubing to prevent contamination.

Humidification can increase the resistance of the bacteria filter and the operator must monitor the bacteria filter frequently for increased resistance and blockage to ensure the delivery of the therapeutic pressure.

Improperly functioning device

If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if it has been dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is cracked or broken, discontinue use and contact your home care service provider.

- 2 -

Power cord
Patient circuits and tubing

Route the power cord to the outlet in a way that will prevent the cord from being tripped over or interfered with by chairs or other furniture.
To avoid strangulation hazards, ensure that all cords connected to the device are properly routed.
This device is activated when the power cord is connected.
Use only power cords supplied by Philips Respironics for this device. Use of power cords not supplied by Philips Respironics may cause overheating or damage to the device.
The device should only be used with compatible patient interfaces (e.g., masks, circuits and exhalation ports). Proper operation of the device with other circuits has not been verified by Philips Respironics and is the responsibility of the health care professional. A mask should not be used unless the device is turned on and operating properly. The exhalation port(s) associated with the mask should never be blocked. Explanation of warning: The device is intended to be used with special masks or connectors that have exhalation ports to allow continuous flow of air out of the mask. When the device is turned on and functioning properly, new air from the device flushes the exhaled air out through the mask exhalation port. However, when the device is not operating, enough fresh air will not be provided through the mask, and exhaled air may be rebreathed. Rebreathing of exhaled air can, in some circumstances, lead to suffocation.
If you are using a full face mask (a mask covering both your mouth and your nose), the mask must be equipped with a safety (entrainment) valve.
An exhalation port is required. Do not block the exhalation port. This can reduce airflow and result in rebreathing of exhaled air.
At low expiratory pressures, the flow through the exhalation port may be inadequate to clear all exhaled gas from the tubing ­ some rebreathing may occur.
Do not pull or stretch the tubing. This could result in circuit leaks.
Do not cover the tubing with a blanket or heat it in an incubator or with an overhead heater. This can affect the quality of the therapy or injure the patient.
The device should only be used with a compatible Philips Respironics mask, as prescribed by your Healthcare Provider.

- 3 -

Accessories
Pulse oximeter
Integrated humidifier
Cleaning

To ensure that you receive the safe, effective therapy prescribed for you, use only Philips Respironics accessories. The use of accessories, transducers, and cables other than those specified by Philips Respironics may result in increased emissions or decreased immunity of the device.
Use only Philips Respironics recommended pulse oximeter and sensors. Use of incompatible sensors can result in inaccurate pulse oximeter performance.
Do not use a damaged pulse oximeter or sensor.
Before use, carefully read these instructions and the instructions for use provided with the pulse oximeter and sensor.
For safe operation when using the integrated humidifier, the device must always be positioned below the breathing circuit connection at the mask. The device must be level for proper operation.
Allow the heater plate and water to cool down for approximately 15 minutes before removing the humidifier water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the humidifier water tank pan.
To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the device.
Do not immerse the device in any fluids or spray the device with water or cleaners. Clean the device with a cloth dampened with an approved cleaner.
Empty and clean the humidifier water tank daily to prevent mold and bacteria growth.

- 4 -

Oxygen EMC

When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen.
Do not connect the device to an unregulated or high pressure oxygen source.
When using oxygen with this system, a Philips Respironics pressure valve must be placed in-line with the patient circuit between the device and the oxygen source. The pressure valve helps to prevent the back flow of oxygen from the patient circuit into the device when the unit is off. Failure to use the pressure valve could result in a fire hazard.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame.
Do not use the device in the presence of a flammable anaesthetic mixture in combination with oxygen or air, or in the presence of nitrous oxide..
Do not use the device near a source of toxic or harmful vapors.
When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off before turning the device off. This will prevent oxygen accumulation in the device. Explanation of the Warning: When the device is not in operation and the oxygen flow is left on, oxygen delivered into the tubing may accumulate within the device's enclosure. Oxygen accumulated in the device enclosure will create a risk of fire.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. See the EMC section of this manual for distances to observe between RF Generators and the ventilator to avoid interference.
Do not use this device near active high frequency surgical equipment and the Radio Frequency shielded room of a Medical Electrical system for magnetic resonance imaging, where the intensity of electromagnetic disturbances is high.
The Health Industry Manufacturers Association recommends that a minimum separation of six inches be maintained between a wireless phone and a pacemaker to avoid potential interference with the pacemaker. The DreamStation 2 on-board Bluetooth communication should be considered a wireless phone in this regard.
- 5 -

Maintenance Never operate the device if any parts are damaged or if it is not working properly. Replace damaged parts before continuing use.
Periodically inspect electrical cords, cables, tubing, and accessories for damage or signs of wear. Discontinue use and replace if damaged.
Repairs and adjustments must be performed by Philips Respironics-authorized service personnel only. Unauthorized service could cause injury, invalidate the warranty, or result in costly device damage. Contact your home care service provider for maintenance.
Do not use the device if the humidifier water tank is leaking or damaged. Replace any damaged parts before continuing use.

Cautions A caution indicates the possibility of damage to the device.

Electrostatic Discharge (ESD)

Do not use antistatic or conductive hoses or conductive patient tubing with the device.
Pins of connectors marked with the ESD warning symbol shall not be touched and connections shall not be made without special precautions. Precautionary procedures include methods to prevent build-up of electrostatic charge (e.g., air conditioning, humidification, conductive floor coverings, non-synthetic clothing), discharging one's body to the frame of the equipment or system or to earth. It is recommended that all individuals that will handle this device understand these precautionary procedures at a minimum as part of their training.

Condensation

Condensation may damage the device. If the device has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature (operating temperature) for 24 hours before starting therapy. Do not operate the device outside of the environmental operating ranges listed in the "Technical specifications" section later in this manual.

- 6 -

Filters
DC power
Tobacco use Device placement

A properly installed, undamaged Philips Respironics reusable pollen filter is required for proper operation.
Dirty inlet filters may cause high operating temperatures that may affect device performance. Regularly examine the inlet filters as needed for integrity and cleanliness.
Never install a wet filter into the device. You must ensure sufficient drying time for the cleaned filter.
Make sure the air inlet (slotted) area on the side of the device is not blocked by bedding, curtains, or other items. Air must flow freely around the device for the system to work properly.
Always ensure that the DC power cord securely fits into your therapy device prior to use. Contact your home care service provider to determine if you have the appropriate DC cord for your specific therapy device.
When DC power is obtained from a vehicle battery, the device should not be used while the vehicle's engine is running. Damage to the device may occur.
Only use a Philips Respironics DC power cord and battery adapter cable. Use of any other system may cause damage to the device.
Tobacco smoke may cause tar build up within the device.
Do not place the device in or on any container that can collect or hold water. Take precautions to protect furniture from water damage.
Do not place the device directly onto carpet, fabric, or other flammable materials.
Do not plug the device into an outlet controlled by a wall switch.
Do not move the device while the humidifier water tank has water in it.

- 7 -

Humidifier water tank

Remove the humidifier water tank, empty all water, and replace the empty humidifier water tank before transporting the device.

Do not fill the humidifier water tank above the maximum fill line. If the humidifier water tank is overfilled, water may leak into the therapy device, humidifier, or onto your furniture. Damage to the humidifier or therapy device may occur.

Use only room temperature distilled water in the humidifier water tank. Do not put any chemicals or additives into the water. Possible airway irritation or damage to the humidifier water tank may result.

If using the integrated humidifier, do not start therapy without the humidifier water tank installed.

Do not attempt to fill the humidifier water tank while it is still connected to the device.

Cleaning

Do not immerse the device or allow any liquid to enter the enclosure or the inlet filter.

Do not steam autoclave the device. Doing so will destroy the device.

Do not use harsh detergents, abrasive cleaners, or brushes to clean the system.

Only the cleaning procedures listed in this manual are recommended by Philips Respironics. Use of other cleaning processes, not specified by Philips Respironics, may affect the performance of the product.

Notice: Any serious incident that has occurred in relation to this device should be reported to Philips and the competent authority of the Member State in which the user and/or patient is established.

Note: An electronic copy of these instructions can be found at: www.philips.com/IFU.

- 8 -

Symbols glossary The following symbols may appear on the device, power supply, accessories, and packaging.

Symbol Title and Meaning Consult instructions for use.

Symbol

www.philips.com/IFU

Electronic instructions for use Indicates that relevant information for use of the product is available in electronic form

For indoor use only Equipment is designed primarily for indoor use.

MR unsafe Do not use device in a Magnetic Resonance (MR) environment.

Bluetooth® symbol Indicates the device has Bluetooth capabilities.

Unique Device Identifier Indicates the Unique Device Identifier information.

AC power (Alternating current)

IP22

Title and Meaning Type BF applied part To identify a type BF applied part complying with IEC 60601-1. DC power (Direct current)
Drip proof equipment
Approved for airline use.
Maximum fill line
Warning: Hot surface
Packaging unit To indicate the number of pieces in the package.

Caution, consult accompanying documents.

Class II equipment (Double Insulated) To identify equipment meeting the safety requirements specified for Class II equipment.

- 9 -

Symbol Title and Meaning

Symbol

Single patient use Indicates that the tubing is for single patient use only.

Importer Indicates the entity importing the medical device into the EU.

Title and Meaning Do not disassemble.
Keep away from sunlight Indicates the medical device needs protection from light sources.

CC

Crossed-out wheeled bin Marking of EEE (electrical and electronic equipment). Follow local requirements for proper disposal.

Medical Device Indicates that the item is a medical device.

Date of Manufacture: to indicate the date on which a product was manufactured Country of Manufacturer: to indicate the country of manufacture of the product Note: When applied to the label, "CC" is replaced by the two letter country code

System contents
Your system may include the following items:

· Device

· Carrying case

· Humidifier water tank · SD Card

· User manual

· Reusable filter

· Quick start guide

· Disposable ultra-fine filter (optional)

· Flexible Tubing · 5 ft. (1.52 m) power cord · Power supply

- 10 -

How to contact Philips Respironics Should you experience trouble with this equipment or require assistance setting up, using, or maintaining the device or accessories, please contact your home care provider. If you need to contact Philips Respironics directly, contact customer service at +1-724-387-4000, or go to www.respironics.com to find your local customer service contact information. 2. System overview The DreamStation 2 CPAP/DreamStation 2 Auto CPAP is a Continuous Positive Airway Pressure therapy device designed for the treatment of Obstructive Sleep Apnea (OSA). The integrated humidifier and optional heated tubing is designed to deliver humidification to provide added comfort during therapy. This humidification level is controlled through the output of the heated humidifier as well as the temperature of the optional heated tubing. Use of the integrated humidifier with the heated tubing allows for a comfortable level of humidity to be maintained at the mask. Several accessories are also available for use with your device. Contact your home care provider to purchase any accessories not included with your system.
- 11 -

Device with integrated humidifier

3

4

5

1

4 6

8 11
10 7 12 9

2

- 12 -

# Feature

Description

1 Air outlet port Connect the flexible tubing here

2 Heated tubing Line up and connect the pin connector heated tubing connector here

3 Therapy on/off Starts and stops the

button

airflow for therapy.

4 Display screen This is the User Interface for the therapy device.

5 Humidifier water tank
6 Air inlet

Removable water tank that holds the water for humidification (shown installed)
Delivers air to the device

# Feature 7 Filter access

Description Access the filter here.

8 SD card access

Access the SD card here.

9 Air inlet for

Connects to the

the humidifier humidifier water tank

10 Humidifier air inlet/outlet seal

Provides a seal between the device and the humidifier water tank.

11 Heater plate

Warms the water in the water tank

12 Power inlet

Connect the power cord here

- 13 -

Humidifier water tank

1

3

2

4

5

4

8

6

6

7

- 14 -

# Feature

1

Lid

2

Lid front tab

3

Water tank release

indent

4

Lid hooks

5

Water tank latch

Description Removable to fill the water tank. Remove lid from here Unlatches water tank from the device when pressed
Connect over water tank base tabs Latches the water tank to the device

6

Water tank base tabs

Connect water tank lid hooks here to attach the water tank lid

7

Maximum fill lines

Indicate the maximum water level for safe operation. The fill lines also appear on the sides of the water tank base.

8

Water tank seal

Provides a seal between the water tank lid and base. Removable for ease of cleaning.

9

Grip indent (not shown) Grip indent, located on the bottom of the water tank base, for tank

removal

- 15 -

3. Device Setup
Where to place the device Place the device on a firm, flat surface somewhere within easy reach of where you will use it at a level lower than your sleeping position. Make sure the device is away from any heating or cooling equipment (e.g., forced air vents, radiators, air conditioners). Note: When positioning the device, make sure that the power cable is accessible
because removing power is the only way to turn off the device.
Cautions · Make sure the air inlet (slotted) area on the side of the device is not blocked by bedding,
curtains, or other items. Air must flow freely around the device for the system to work properly. · Do not place the device directly onto carpet, fabric, or other flammable materials. · Do not place the device in or on any container that can collect or hold water.
Supplying AC power to the device Warning: Periodically inspect electrical cords and cables for damage or signs of wear.
Discontinue use and replace if damaged. Caution: Do not use extension cords with this device. Complete the following steps to operate the device using AC power and refer to the following images for guidance:
1. Ensure that the humidifier tank is empty. 2. Plug the power supply cord's connector into the power inlet on the bottom of the
device 1 or 2 . For convenience, the power connection allows for the cord to connect and route in the direction that works best for your setup. An inset area on the bottom of the device allows the cord to be routed under the device 3 .
- 16 -

1
OR
2 3
- 17 -

3. Plug the socket end of the AC power cord into the power supply 4 .

4. Plug the pronged end of the AC power cord

into an electrical outlet that is not controlled

by a wall switch 5 .

4

5. Verify that the connections on the bottom of the device, at the power supply, and at the electrical outlet are fully inserted. This will help to ensure that a secure, reliable electrical connection has been made.

Note: Example only shown here. Local

electrical outlet and power cord

may vary.

5

Important: To remove AC power, disconnect the power supply cord from the electrical outlet.

Setting up the integrated humidifier
Warning: Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan.
Cautions
· Do not attempt to fill the tank while it is still attached to the device. · Use only room temperature distilled water in the tank. Do not put any chemicals or
additives into the water. Possible airway irritation or damage to the water tank may result. · Do not fill the water tank above the maximum fill line. If the water tank is overfilled,
water may leak into the therapy device, humidifier, or onto your furniture. Damage to the humidifier or therapy device may occur. Note: Clean the humidifier water tank before first use. Refer to the "Home cleaning:
humidifier water tank" section in this manual.

- 18 -

Removing the humidifier water tank 1. Gently press down on the indented area
on the top of the humidifier water tank to unlatch the tank from the device.
2. Grip the top and bottom of the humidifier water tank and slide it out away from the device.
- 19 -

3. Remove the lid by pulling up on the lid tab while holding the water tank base.

4. Pour out any remaining water and rinse the water tank base.

Filling and connecting the water tank

1. Place the water tank base on a firm, flat

surface. Fill the water tank with distilled

water no higher than the maximum fill

lines

, which are located on the back

and sides of the water tank base.

- 20 -

2. Reattach the lid. Place the hooks on the lid 1 over the small tabs on the water tank base 2 .

1

1

2

2

3. Press down 3 until the lid snaps securely over the lid front tab 4 on the front of the water tank base.
3

4

- 21 -

4. Slide the humidifier water tank back into the device, using the tracks on the bottom of the water tank base for proper alignment and connection. Ensure that it is latched into place before proceeding.
Installing/replacing the air filters Caution: A properly installed, undamaged Philips Respironics reusable pollen filter is
required for proper operation. The device uses a grey reusable pollen filter that can be rinsed and a blue disposable ultrafine filter. The reusable filter screens out normal household dust and pollens, while the ultrafine filter provides more complete filtration of very fine particles. The reusable filter must be in place at all times when the device is operating. The ultra-fine filter is recommended for people who are sensitive to tobacco smoke or other small particles. The grey reusable filter is supplied with the device. A disposable blue ultra-fine filter may also be included. If a filter is not installed when you receive your device, you must at least install the reusable filter before using the device. 1. Remove the humidifier water tank from the device. See the "Removing the humidifier
water tank" section later in this manual.
- 22 -

2. To remove an existing filter, pull the white tab on the end of the filter and pull the filter out of the device
3. If applicable, place a dry, grey reusable pollen filter on top of a new, optional disposable blue ultra-fine filter, align the arrows on both filters, and firmly snap them together.
4. Place the filter into the therapy device.

5. Reinstall the humidifier water tank.

- 23 -

Connecting the breathing circuit To use the system, you will need the following accessories in order to assemble the recommended breathing circuit: · Philips Respironics interface (nasal mask or full face mask) with integrated exhalation
port, or Philips Respironics interface with a separate exhalation device (such as the Whisper Swivel II) · Philips Respironics flexible tubing (12, 15, or 22 mm standard tubing or 12 or 15 mm heated tubing), 1.83 m (6 ft.) · Philips Respironics headgear (for the mask) Warnings · Do not pull or stretch the tubing. This could result in circuit leaks. · Inspect the tubing for damage or wear. Discard and replace the tubing as necessary. · If you are using a full face mask (a mask covering both your mouth and your nose), the mask must be equipped with a safety (entrainment) valve. · If the device is used by multiple persons (such as rental devices), a low-resistance, main flow bacteria filter should be installed in-line between the device and the circuit tubing to prevent contamination.
To connect your breathing circuit to the device, complete the following steps. Note: If you are using the optional 12 mm (non-heated) performance tubing, an
adapter is required to connect to the therapy device.
Note: Tubing is identified on the cuff with the tubing identifier symbol: "12", "15", "HT12", or "HT15". 22 mm tubing contains no symbol.
1. Connect the flexible tubing to the air outlet on the therapy device.
To connect heated tubing (shown), line up the pin connectors on the heated tube with the bottom of the air outlet port on the device. The clips at the end of the tubing should be aligned to the sides of the port.
- 24 -

2. Press the heated tubing into place over the air outlet port until the tabs on the side of the tube click into place in the slots on the sides of the outlet port. If you are using standard tubing (not shown), simply slide the tubing over the air outlet port on the device.
3. If you are using the optional 12 mm heated tubing or 12 mm performance tubing, connect the provided mask adapter to the mask connection end of the tubing. The 12mm performance tubing also requires a device adapter. When connecting the adapter to the CPAP, the clips should be aligned to the sides with the slight bump facing straight down.
4. Connect the tubing to the mask. For proper placement and positioning, refer to the instructions that came with your mask.
5. Attach the headgear to the mask if necessary. Refer to the instructions that came with your headgear.
Starting the device 1. Ensure power is supplied to the device as indicated by a white ring around the therapy
button. The first screen to display is the Philips logo, followed by the device model screen. Once fully powered on, the screen displays "To begin therapy click below". Note: The device may prompt you to set the language and time. 2. Put on your mask assembly. Refer to the instructions supplied with the mask. A small amount of mask leak is normal and acceptable. Correct large amount of mask leaks or eye irritation by adjusting your mask headgear. Note: If you are using the device in a bed with a headboard, try placing the tubing over the headboard. This may reduce tension on the mask.
- 25 -

3. Press the Therapy button on top of the device to turn on airflow and begin therapy. The Therapy button ring will illuminate blue. Note: If your home care provider has enabled the Automatic On feature, the device will automatically turn the airflow on when you put on your mask and breathe.
4. Press the Therapy button again to turn off the therapy. The Therapy button ring will illuminate white. Note: If your home care provider has enabled the Automatic Off feature, the device will automatically turn the airflow off when you take off your mask and the device recognizes inactivity.
Performance check To initiate a performance check, unplug the device and plug it back in. An error will appear on the screen if a performance issue is detected. Bluetooth wireless technology* Bluetooth wireless technology is one method by which you can transfer your therapy device's data to DreamMapper. DreamMapper is a mobile system designed to help Obstructive Sleep Apnea (OSA) patients enhance their sleep therapy experience.
*Bluetooth wireless technology and DreamMapper are not available in all markets. For more information, please consult your local Philips Respironics representative.
- 26 -

Pairing your therapy device to your Bluetooth-enabled mobile device Note: The blower must be off to allow Bluetooth pairing. Note: You can only pair your therapy device to one mobile device at any given time. Note: Pairing works best when your therapy device and mobile device are in the same
room. Note: The current version of DreamMapper will guide you through these instructions. Note: After initiating pairing in DreamMapper, you will have 30 seconds to complete
the setup. After this time, it will be cancelled automatically. Follow the steps below to manually pair to your mobile phone or tablet. 1. With your therapy device powered up, initiate Bluetooth Setup from the DreamMapper
mobile app. Note: From DreamMapper you may need to select from a list of available Bluetooth devices. The therapy device will appear as "PR BT XXXX" (XXXX will be the last four digits of the serial number listed on your therapy device). 2. The pairing code will display on the device. 3. In DreamMapper, confirm the 6-digit pairing code displayed on the therapy device. Note: When pairing from certain mobile devices you may need to enter the 6-digit pairing code in DreamMapper. Once entered, pairing will automatically begin. 4. Click the Therapy button to complete pairing. The therapy device displays "Pairing to Device" and then "Success Device Paired".
- 27 -

Therapy On display
While the device is delivering therapy, the prescription pressure or RAMP PLUS pressure is displayed. You can also view and adjust your humidification and RAMP PLUS settings.

1

5

2

3

4

#

Symbol Feature

Description

1

cmH2O Therapy

Displays the prescribed pressure setting. If RAMP PLUS is running, the

pressure

current delivered pressure is displayed.

2

Adjustable Tap the humidification symbol to view the current humidifier setting.

humidifier To change the setting, tap the humidification icon again until the

setting (if desired setting is displayed. Each tap will display a different setting.

enabled) After approximately 3 seconds the display returns to the therapy

screen and the setting is saved.

The available settings are: Off, Low, Medium, and High.

Note: If you adjust the humidifier setting during therapy, the new setting will automatically be active for current therapy, and it will be used the next time you turn therapy on.

- 28 -

#

Symbol Feature

Description

3

RAMP

The device is equipped with a RAMP PLUS feature that allows

PLUS

you to adjust your starting pressure for a set period of time

(Ramp+) (default of 30 minutes) for added comfort when you are trying

to fall asleep. During the set time, the air pressure will remain at

your set starting pressure unless the device detects an event and

identifies the need to increase your pressure. When RAMP PLUS

has concluded, your prescribed therapy pressure will resume.

To activate RAMP PLUS, tap the RAMP PLUS symbol. The first time RAMP PLUS is activated, the setting will default to Low. To change the setting, continue to tap the RAMP PLUS symbol until the desired setting is displayed. After approximately 3 seconds the display returns to the therapy screen and the setting is saved. Every therapy session thereafter will automatically start RAMP PLUS with the start of therapy.

The available settings are: Off, Low (4 cmH2O), Medium (6

cmH O), High (8 cmH O), and Max (10 cmH O).

2

2

2

Note: There is no need to tap the RAMP PLUS symbol again

unless you want to make a change to the RAMP PLUS pressure

or restart RAMP PLUS.

Note: The device screen goes dark after 60 seconds of inactivity. To wake up the screen, tap the RAMP PLUS icon or the humidification icon. This will simultaneously reactivate your RAMP PLUS pressure (if set), or it will take you to your prescribed minimum pressure level and continue to automatically adjust your delivered pressure as needed (if in Auto CPAP mode). Once the display appears, you can continue to adjust settings as desired.

4

Therapy

The therapy button ring illuminates blue to indicate therapy is

button ring turned on. The ring does not illuminate when therapy is on and

the screen has gone dark.

The ring illuminates white when the device is plugged in and therapy is off.

- 29 -

# 5
Not shown Not shown
Not shown

Symbol

Feature

Description

Oximetry

The therapy device can pair to a pulse oximeter through a Bluetooth connection. When the therapy device successfully pairs to the pulse oximeter, the Bluetooth symbol and "SpO2" will display on the therapy screen. See the "Oximeter" section later in this manual for additional information.

Cell modem Displays signal strength of cellular modem on devices that have

signal

the cellular modem available except when in Airplane Mode.

strength

Bluetooth

Displays when the device has Bluetooth available. When the symbol is grey, it indicates that Bluetooth is available. If the device is connected to a Bluetooth client (phone, tablet, etc.) it will appear white. This symbol does not appear when Airplane Mode is enabled.

Airplane mode

Displays when Airplane Mode is enabled. It will replace the Cell and/or Bluetooth symbols when active.

- 30 -

Therapy Off display After Therapy is turned off a series of screens that display a summary of your therapy may appear. A description of each screen is below. Each screen displays three times and then the screen displays:
To begin therapy
click below

Note: You can also restart therapy at any time by pressing the therapy button.

Display text Description

Last Session This screen displays the amount of time you received therapy on the device for

Duration

the most recent one day time frame.

Last Session This screen displays the nightly Apnea/Hypopnea index (AHI) value for the

AHI

most recent one day time frame.

Last Session Mask fit

This screen displays how your mask fit during the most recent one day time frame. A display of "Good fit" indicates that the leak found allows for optimal performance of the device. A display of "Adjust fit" indicates that the leak may affect device performance, however, the device will remain functional and deliver therapy. If "Adjust fit" displays, adjust your mask to reduce the leak before starting therapy again.

- 31 -

Device settings (language, time, airplane mode, Bluetooth reset) The device settings screens allow you to reset the device language and time and Bluetooth connections. You also can turn Airplane Mode on and off. Follow the steps below to navigate through the device settings screens. For each setting, press and hold the Therapy button to complete the action, or click the Therapy button to cancel the action. Note: To reset Bluetooth connections and change Airplane Mode you must follow the
steps below to first pass through the language and time screens.
1. With therapy off, press and hold to the left and right of the display screen simultaneously until the display shows the language and time reset screen.
2. After approximately 3 seconds, the screen displays: Settings reset Hold to start Click to cancel
3. Press and hold the Therapy button to start the language and time reset, or click the Therapy button to cancel reset. The device displays the reset or cancelling screens and then displays the Bluetooth setting. Note: If you reset language and time, the device will prompt you to set the language and time before the Bluetooth selection screen displays.
4. Press and hold the Therapy button to reset the Bluetooth connections, or click the Therapy button to cancel reset. The device displays the reset or cancelling screens and then displays the Airplane Mode setting.
5. Press and hold the Therapy button to turn Airplane Mode on or off. The device displays the reset or cancelling screens and then exits back to the Therapy off screen. Note: Airplane mode will automatically turn off after three days.
- 32 -

4. Pop-up messages and troubleshooting

Device pop-up messages
Device pop-up messages may appear on your device to alert you of a device status update or required action. When the message appears, follow the action or reference the following table for an explanation of the pop-up and any action that should be taken. When "Click to Dismiss" displays with the message, you can click the Therapy button to dismiss and close the pop-up screen.

Alerts and notifications

Condition/Screen text
Blocked inlet

Description Blocked airway

Possible cause
Blockage at device inlet

Action
Check device air inlet is not obstructed. Check air filter(s) are installed properly; replace if needed.

Low leak Humidifier Error Tube Error

Blocked airway
Humidification control is disabled.
Heated tube error (only when heated tube is present)

Blockage at tube or mask
There may be a problem with your humidifier. Therapy will run without humidification. Heated tube may be overheated or damaged.

Check tube is not crushed or folded such that air flow is restricted. Check mask is attached properly and without any obstruction.
Turn off device and disconnect from power. Visually check that electrical contacts are clear, then reconnect power cord. If alert continues, contact your provider.
Turn off device. Detach heated tube from device, make sure that tube is not covered or obstructed, and then reattach to device. If alert continues, contact your provider.

- 33 -

Condition/Screen text Humidifier Off
Check Power (light blinks continuously) Low Voltage (light blinks continuously)
Device Update Hold to Confirm Click to Cancel
SD Error Remove/Reinsert

Description

Possible cause

The attached battery does not support humidification.
Indicates an incompatible power supply is attached.
Low voltage

Use of battery pack for power.
Incorrect power supply.
Incompatible power supply is attached.

A new software version is available.

New software update.

There was a problem reading your SD card.

Device cannot read the SD card. A problem may exist with the SD card, it was removed during a writing activity, or it was inserted incorrectly.

Action
Disconnect the battery pack and use a compatible power supply or use the device without humidification.
Switch to a Philips-provided power supply that is capable of supporting therapy. An 80W power supply is required to support humidification. Confirm a compatible Philips Respironics power supply is attached. Switch to compatible power supply if needed. If battery is being used, ensure battery is adequately charged. Press and hold the therapy button to update the device software.
Or
Click the therapy button to cancel the update. Click the Therapy button to clear the notification. Remove and reinsert the SD card. If alert continues, contact your provider.

- 34 -

Condition/Screen Description text

SD Full Contact provider

Your SD card is full.

Possible cause Your SD card is full.

SD Removed Reinsert

SD card is not inserted into the device.

Indicates SD card has been removed from therapy device and not reinserted before the start of the current therapy session.

Action
Remove SD card and replace with a new card, or contact your provider for a new SD card. Reinsert the SD card to record therapy data.

Service Required Contact support (light blinks continuously)

Indicates an

Device error.

error

which enters

device into "Safe

State." This

allows power

to remain on

but airflow is

disabled.

Disconnect device from power. Reattach power cord to restore power. If the alert continues to occur, contact your home care provider.

- 35 -

Status pop-up messages
Status messages do not require action. The pop-up will automatically time out or you can click the Therapy button to clear the message.

Screen Text
Compliance Achieved
SD Card Activity Do not remove.
Prescription Updated
Prescription Invalid

Description You have reached your therapy compliance goal.
SD card read/write underway.
Your prescription has been updated.
A prescription or setting update was unsuccessful. Contact your home care provider.

Therapy Auto Off
SpO2 Successful
SpO2 Unsuccessful

Your device has automatically turned off due to inactivity.
You have achieved at least 4 hours of therapy and oximetry use.
Minimum therapy hours not met. Your pulse oximetry recording was unsuccessful.

- 36 -

Troubleshooting

Problem
Nothing happens when you apply power to the device.

Why it happened
There's no power at the outlet or the device is unplugged.

The airflow does not turn on.

There may be a problem with the blower.

The device's display is erratic.

The device has been dropped or mishandled, or the device is in an area with high Electromagnetic Interference (EMI) emissions.

What to do
If you are using AC power, check the outlet and verify that the device is properly plugged in. Make sure there is power available at the outlet. Make sure the AC power cord is connected correctly to the power supply and the power supply cord is securely connected to the device's power inlet. If the problem continues to occur, contact your home care provider. Return both the device and power supply to your provider, so they can determine if the problem is with the device or power supply. If you are using DC power, make sure your DC power cord and battery adapter cable connections are secure. Check your battery. It may need recharged or replaced. If the problem persists, check the DC cord's fuse following the instructions supplied with your DC cord. The fuse may need to be replaced. If the problem still occurs, contact your home care provider.
Make sure the device is powered correctly. Press the Therapy button on top of the device to start airflow. If the airflow does not turn on, there may be a problem with your device. Contact your home care provider for assistance.
Unplug the device. Reapply power to the device. If the problem continues, relocate the device to an area with lower EMI emissions (away from electronic equipment such as cellular phones, cordless phones, computers, TVs, electronic games, hair dryers, etc.). If the problem still occurs, contact your home care provider for assistance.

- 37 -

Problem The airflow is much warmer than usual.
The water in the humidifier water tank runs out before morning.
I hear a leak or whistling sound coming from my therapy device (not related to mask leak).

Why it happened The air filters may be dirty. The device may be operating in direct sunlight or near a heater.
Humidifier water tank was not full at start of session. Mask leak is excessively high. The ambient conditions are very dry/cool.
The therapy device air inlet may be obstructed. The tube is not fully connected. The humidifier seals are not fully seated or are missing.

What to do
Rinse or replace the reusable air filter or replace the disposable ultra-fine filter. The temperature of the air may vary somewhat based on your room temperature. Make sure that the device is properly ventilated. Keep the device away from bedding or curtains that could block the flow of air around the device. Make sure the device is away from direct sunlight and heating equipment. If the problem continues, contact your home care provider.
Under most conditions, a full humidifier water tank should last for a typical sleep session if the humidifier tank is filled to the maximum fill line at the beginning of the sleep session. However, many factors impact water consumption, including: your humidifier or heated tube settings, the level of mask leak, and the duration of your sleep session. Make sure that the humidifier water tank is filled to the maximum fill line at the start of your sleep session. Check that your mask is fitted properly, and adjust as needed to reduce mask leak to normal levels.
Check therapy device air inlet is not obstructed, and filters are free of debris and properly inserted. Confirm that the device, humidifier water tank, and tube are connected properly and not leaking. Confirm that the water tank lid seal is present and properly seated.

- 38 -

5. Connectivity

QoS: Wireless Quality of Service (QoS) refers to the necessary level of service and performance needed for the wireless functions of the device. It involves parameters such as reliability of data transmission, effective transfer rate, error rate, and mechanisms to define priority levels for time-critical signals.

Bluetooth® QoS: Bluetooth uses frequency hopping, channel coding, and error correction to address interference and is designed to operate with other devices that occupy the same spectrum. In addition to the measures defined in the Bluetooth standard, the DreamStation 2 radio incorporates other methods to minimize likelihood of QoS problems. These include:

-- Data sent between the CPAP and any external devices use an additional checksum verification to ensure that data is correctly received without errors.

-- For all Bluetooth applications except pulse oximeter: The CPAP is a portable device and will not always be near the mobile device when the CPAP is ready to transfer data. The mobile device attempts to reconnect until it successfully connects and the data transfer is complete.

--

For use second

with pulse and stores

oximeter: that data

aTlhoengrawdiitohrtehceeitvimesepsutalsmepra. TtehaenddatOa2

level from

once per the pulse

oximeter is checked for validity and thrown away if not valid. This data is not

displayed locally but is logged, and after a minimum of 4 hours of data is collected

it is sent to the remote server. If the CPAP is not able to get 4 hours of good data it

alerts the patient, and the patient will need to use it again the next night. The data

is not used to diagnose the patient's condition, and will continue to reconnect to

the pulse oximeter until a connection is achieved.

Cellular QoS

The Cellular modem is designed for use with select Philips Respironics therapy devices. It automatically transfers data between the DreamStation 2 device and Philips Respironics proprietary compliance software. The DreamStation 2 modem incorporates various methods to minimize likelihood of QoS problems. These include:

-- No real time data monitoring is used in this application. If the data transfer is unsuccessful, the modem attempts to reconnect until it successfully connects and the data transfer is complete.

-- Data sent between the CPAP and any external devices uses an additional checksum verification to ensure that data is correctly received without errors.

For information on Bluetooth use and pairing, see the "Bluetooth wireless technology" and "Pairing your therapy device to your Bluetooth-enabled mobile device" sections of this manual.

- 39 -

6. Accessories
There are several optional accessories available for your DreamStation 2 device. For a full list of accessories that can be used with this device, see the DreamStation 2 accessory list at www.philips.com/IFU. Contact your home care service provider for additional information on the available accessories. When using optional accessories, always follow the instructions enclosed with the accessories. SD card The device may come with an SD card inserted in the SD card slot behind the water tank. The SD card stores information for the home care provider. Your home care provider may ask you to periodically remove the SD card and send it to them for evaluation.
Updating software using an SD card You can update the device software using an SD card. The software update must be done when the therapy is off. 1. Connect the device to power. 2. Insert an SD card with the new software version into the device. After approximately 15
to 20 seconds, a pop-up screen appears displaying: Device Update Hold to Confirm Click to Cancel
- 40 -

3. Press and hold the Therapy button for at least 3 seconds to continue with the software upgrade. To cancel the software upgrade, click the Therapy button.
4. When the update begins, the screen displays "Software Update Active" and the therapy button ring flashes. The screen will then appear black or may display a progress bar until the update is complete.
5. The device User Interface (UI) will restart when the upgrade is complete. Remove the SD card.
6. If an error occurs, contact Philips Respironics for a new SD card.
Oximeter
The DreamStation 2 device can pair to a Nonin BT Pulse Oximeter, using Bluetooth, for measuring %SpO2 and pulse rate. To use a pulse oximeter with your therapy device, follow these steps. 1. Follow the setup instructions provided with your pulse oximeter and finger sensor. 2. Attach the finger sensor to your forefinger. 3. Press the Therapy button on your device to begin therapy. 4. Once the device detects a good connection with the pulse oximeter, both the Bluetooth
symbol and "SpO2" will display on the top of the display screen. Note: It may take up to 30 seconds for the device to recognize the pulse oximeter.
When you have achieved at least 4 hours of therapy and oximetry use, a "SpO2 Successful" pop-up message will display. If you do not reach the minimum therapy hours, a "SpO2 Unsuccessful" message will display.
- 41 -

Roll stand
There is a roll stand available for use with your DreamStation 2 device. Please see the instructions included with your roll stand for more information.
Supplemental oxygen
Oxygen can be added to the patient circuit. Please note the warnings listed below when using oxygen with the device. Warnings · When using oxygen with this system, the oxygen supply must comply with local
regulations for medical oxygen. · Oxygen supports combustion. Oxygen should not be used while smoking or in the
presence of an open flame. · Do not use the device in the presence of a flammable anaesthetic mixture in combination with
oxygen or air, or in the presence of nitrous oxide.. · If supplemental oxygen is added at the exit of the flow generator or humidifier, a Philips
Respironics Pressure Valve must be placed in-line with the patient circuit between the device and the oxygen source. The pressure valve helps prevent the backflow of oxygen from the patient circuit into the device when the unit is off. Failure to use the pressure valve could result in a fire hazard. · When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off before turning the device off. This will prevent oxygen accumulation in the device. · Do not connect the device to an unregulated or high pressure oxygen source.
DC power
A Philips Respironics DC power cord can be used to operate this device in a stationary recreational vehicle, boat, or motor home. In addition, a Philips Respironics DC battery adapter cable, when used with a DC power cord, allows the device to be operated from a 12 VDC free-standing battery. Warning: Periodically inspect electrical cords and cables for damage or signs of
wear. Discontinue use and replace if damaged.
- 42 -

Cautions · Always ensure that the DC power cord securely fits into your therapy device prior to use.
Contact your home care provider or Philips Respironics to determine if you have the appropriate DC cord for your specific therapy device. · When DC power is obtained from a vehicle battery, the device should not be used while the vehicle's engine is running. Damage to the device may occur. · Only use a Philips Respironics DC Power Cord and Battery Adapter Cable. Use of any other system may cause damage to the device. Refer to the instructions supplied with the DC power cord and adapter cable for information on how to operate the device using DC power.
Carrying case and airline travel
When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect the system if it is put through checked baggage. Do not travel with water in the water tank. The device is suitable for use on airlines when the device is operating from an AC or DC power source. For your convenience at security stations, there is a symbol on the bottom of the device indicating that it is medical equipment and is suitable for airline use. It may be helpful to bring this manual along with you to help security personnel understand the device. If you are traveling to a country with a line voltage different than the one you are currently using, a different power cord or an international plug adaptor may be required to make your power cord compatible with the power outlets of the country to which you are traveling. Contact your home care provider for additional information.
- 43 -

7. Cleaning, disinfection, and care
Warning: Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank base.
Home cleaning: device exterior, heater plate, and humidifier air inlet/outlet seal Clean the device exterior surface weekly. Clean the air inlet/outlet seal daily. 1. To avoid electrical shock, make sure that the device is disconnected from all outlets and
power sources. Remove any cables attached to the device or battery pack. 2. Remove the humidifier tank from the device. 3. Use a lint-free cloth dampened (not dripping) with a liquid soap solution (5 ml of liquid
dish soap per 3.8 liters of warm potable water) to clean the exterior of the device, heater plate, and humidifier air inlet seal. 4. Work the cloth into the areas around the therapy button, humidifier air inlet/outlet seal, and any other areas where soil may be difficult to remove. Ensure that you remove all visible soil. 5. Use a lint-free cloth dampened (not dripping) with potable water to remove all detergent residue. 6. Inspect the device for cleanliness. If necessary, repeat the cleaning steps until all surfaces are visibly clean. 7. Allow the device to dry completely before reconnecting it to a power source. 8. Inspect the device and all circuit parts (filter, tube, and mask) for damage, such as cracks, tears, or broken pieces. Ensure that air inlet/outlet seal is properly installed and not dislodged. Replace any damaged parts.
- 44 -

Home cleaning: humidifier water tank (lid, water tank base, and water tank seal) Clean the humidifier water tank before first use. Hand wash daily. The humidifier water tank can also be washed in the top rack of a dishwasher weekly. 1. Press the therapy button to stop the airflow, and allow the heater plate and water to
cool. 2. Remove the water tank from the device and remove the lid and water tank seal from the
tank base. 3. Wash the humidifier water tank (lid, water tank base, and water tank seal) in the
dishwasher (top shelf only) or in a solution of warm potable water and a mild liquid dishwashing detergent (5 ml of liquid dish soap per 3.8 liters of warm water) using a soft bristle brush to remove adhering substances. Note: Pay close attention to all corners and crevices. 4. Fully immerse and rinse each item separately with potable water for one minute and agitate vigorously. 5. Allow all parts to air dry. 6. Inspect the humidifier water tank for damage. If any parts show signs of wear or damage, contact your home care provider for a replacement. 7. Re-install the water tank seal. To install, insert the seal into the back of the tank base and then press the seal down until it is fully seated into the water tank base.
8. Before reinstalling the humidifier water tank, fill it with distilled water no higher than the maximum fill line.
- 45 -

Home disinfection: humidifier water tank (lid, water tank base, and water tank seal) Disinfect the humidifier water tank weekly.
Requirements:
· 70% isopropyl alcohol (70% solution of isopropyl alcohol in water)
1. Before disinfecting the device, be sure that it has been cleaned as instructed in the previous "Home cleaning: humidifier water tank" section of device user manual.
2. Immerse the humidifier tank (lid, water tank base, and water tank seal) in 70% isopropyl alcohol for 5 minutes.
3. Rinse all parts of the humidifier tank with potable water for at least 1 minute. 4. Allow all parts to air dry. 5. Inspect the humidifier water tank for damage. If any parts show signs of wear or damage,
contact your home care provider for a replacement.
Home cleaning: tubing Hand wash the tubing, mask adapter (if included), and connectors (if included) before first use and weekly. Discard and replace the tubing every 6 months. 1. Disconnect the flexible tubing from the device. 2. Gently wash the tubing, including any adapters or connectors, in a solution of warm
potable water and a mild liquid dishwashing detergent (1 teaspoon of liquid dish soap per gallon of warm water)to adequately remove adhering substances from the tube, adapters, and connectors. Gently agitate the tubing by hand so that the inner surface of the tube is cleaned. 3. Rinse thoroughly to remove all soap residue from the tube, adapters, and connectors with water and allow to air dry. Make sure all parts are dry before next use. 4. Visually inspect the tubing for cleanliness. Repeat the cleaning if not visually clean. 5. Inspect the tubing for damage or wear (cracking, tears, punctures, etc.). Discard and replace if necessary.
- 46 -

Caring for the filters Notes · Only Philips supplied filters should be used as replacement filters. · Replace the disposable, blue ultra-fine filter if it is damaged or has accumulated debris. In the home environment, the disposable, ultra-fine filter should be replaced after 30 nights of use, or sooner, if it appears clogged. DO NOT rinse the ultra-fine filter. Under normal usage in the home environment, you should rinse the grey reusable filter at least once every two weeks and replace it with a new one every six months. 1. If the device is operating, press the Therapy button to stop the airflow. Unplug the power
cord from the wall outlet. 2. Remove the water tank to access the filter area, then remove the filter from the device.
Refer to the "Install/replace the air filters" section earlier in this manual. 3. Take the reusable filter to a sink and run warm tap water through the white middle
portion of the filter to rinse away any debris. 4. Shake the filter to remove as much water as possible. 5. Allow the filter to air dry completely before reinstalling it. If the filter is damaged, replace
it. 6. Reinstall the filter into the filter access area on the device.
Service The device does not require routine servicing. Warning: If you notice any unexplained changes in the performance of this device,
if it is making unusual or harsh sounds, if it has been dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, disconnect the power cord and discontinue use. Contact your home care provider.
- 47 -

8. Technical specifications Environmental
Operating Temperature Storage Temperature Relative Humidity (operating & storage) Atmospheric Pressure:
Physical
Dimensions

Device: 5° to 35°C (41° to 95°F) -20° to 60°C (-4° to 140°F) 15 to 95% (non-condensing) Device: 101 to 77 kPa (0 - 2286 m / 0 - 7500 ft)
273.81 mm L x 158.5 mm W x 84.83 mm H (10.78 in L x 6.24 in W x 3.34 in H)

Weight (empty water tank without power supply) 1040 g (2.29 lbs)
Service life The expected service life of the DreamStation 2 device is five (5) years. The expected service life of the DreamStation 2 humidifier water tank is one (1) year. The expected service life of the DreamStation 2 humidifier air inlet/outlet seal is one (1) year.
Standards compliance This device is designed to conform to the following standards: · IEC 60601-1 General Requirements for Basic Safety and Essential Performance of
Medical Electrical Equipment · IEC 60601-1-11 General Requirements for Basic Safety and Essential Performance in the
Home Healthcare Environment · IEC 60601-1-6 General Requirements for Safety - Usability · IEC 62366 Application of Usability Engineering in Medical Devices · IEC 62304 Medical Device Software ­ Software Life-cycle Processes · ISO 80601-2-70 Sleep Apnea Breathing Therapy Equipment · ISO 80601-2-74 Respiratory Humidifying Equipment · EN 60601-1-2 Electromagnetic Compatibility

- 48 -

IEC 60601-1 classification

Type of Protection Against Electric Shock

Class II Equipment

Degree of Protection Against Electric Shock

Type BF Applied Part

Degree of Protection Against Ingress of Water

Device: Drip Proof, IP22 First characteristic numeral - 2 - Protection against ingress of solid foreign objects  12.5 mm diameter. Explanation: Protected against access to hazardous parts with a finger and protected against solid foreign objects of 12.5 mm diameter and greater. Second characteristic numeral - 2 - Protection against ingress of water with harmful effects dripping (15° tilted). Explanation: Protected against vertically falling water drops when enclosure tilted up to 15°.

Mode of Operation

Continuous

Electrical Water capacity Heater plate Humidity
Maximum delivered gas temperature

AC Power Consumption: 100­240 VAC, 50/60 Hz, 2.0-1.0 A
325 ml (11 oz.) at recommended water level
Max Temperature: 68°C (154°F)
Humiditymin Output:  12 mg H2O/L Measured at expected leak across therapy pressures , 17 °C­ 35°C, 15% RH for all compatible tubes.
<43°C (109.4°F)

- 49 -

Declared dual-number noise emissions values (in accordance with ISO 4871)

Tube Size

Sound Pressure Level

Uncertainty

Sound Power Uncertainty Level

15 and 22 (mm) tubing type
12 (mm) tubing type

27 dB(A) 27 dB(A)

2 db(A) 2 db(A)

35 dB(A) 35 dB(A)

2 dB(A) 2 dB(A)

Pressure accuracy Pressure Increments: 4-20 cmH2O (in 0.5 cmH2O increments)

Maximum static and dynamic pressure accuracy, according to ISO 80601-2-70:2015:

Tube Type

Static

Dynamic

4 to 20 cmH2O

15 and 22 (mm) tubing type

± 0.5 cmH2O

± 1.0 cmH2O

12 (mm) tubing type

± 1.0 cmH2O

± 2.0 cmH2O

Static pressure accuracy has a measurement uncertainty of 1.31%.

Dynamic pressure accuracy has a measurement uncertainty of 2.70%.

- 50 -

Maximum flow rate (typical)

Tube Type

Flow 4

12 (mm) tubing type

Average flow at the patient

120

(heated or non-heated) connection port (l/min)

15 (mm) tubing type

Average flow at the patient

121

(heated or non-heated) connection port (l/min)

22 (mm) tubing type

Average flow at the patient

122

connection port (l/min)

ISO 80601-2-70 Pneumatic Diagram

12

3

4

Test pressures (cmH2O)

8

12

16

20

117

108

101

92

135

125

116

107

141

131

122

112

5

6

# 1 Air 2 Filter & air inlet 3 Blower 4 Air outlet 5 6 ft. tubing 6 Mask

Feature

- 51 -

Wireless notices · The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG,
Inc. and any use of such marks by Philips Respironics is under license. Other trademarks and trade names are those of their respective owners. · The DreamStation 2 device is capable of transmitting data between the therapy device and a mobile device. This connection between the therapy device and a mobile device is encrypted. · A small portion of the firmware that performs data encryption on the DreamStation 2 device is being utilized under the Apache 2.0 and Mozilla 2.0 licenses. These licenses are available at:
www.apache.org/licenses/LICENSE-2.0 and https://www.mozilla.org/en-US/MPL/2.0/ · This product meets RF exposure requirements when it is positioned with a separation
distance of at least 20 cm away from the body. · This device is a certified Bluetooth radio device with: FCC ID (USA): THO1141623 · Use of non-original manufacturer-approved accessories may violate your local RF
exposure guidelines and should be avoided. · This device complies with part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio, TV reception, or other devices which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interference by one or more of the following measures: -- Reorient or relocate the receiving antenna (on the radio, TV, or other device). -- Increase the separation between the equipment and receiver. -- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected. -- Consult the dealer of the device for help.
- 52 -

· Hereby, Respironics, Inc. declares that the DreamStation 2 family of products are in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: http://incenter.medical.philips.com/PMSPublic
Radio Specifications
Cellular

Operating Frequency Range

3G UMTS: 800, 850, 900, 1900, 2100 MHz

Maximum Output Power
Modulation
Bluetooth Operating Frequency Range Maximum Output Power Modulation Bandwidth: Intake Port Filters

2G GSM: 850, 900, 1800, 1900 MHz 3G UMTS: 24 dBm 2G GSM: 33 dBm 3G UMTS: QPSK 2G GSM: GMSK, 8-PSK
2.4-2.4835 GHz 3 dBm GFSK 2 MHz

Pollen Filter Ultra-fine Filter

100% Polyester 88% Efficient @ 7-10 micron size
Blended Synthetic Fiber 95% Efficient @ 0.5-0.7 micron size

- 53 -

Heated Tubing Specifications

Maximum Recommended Pressure

20 cmH2O

Inner Diameter
Length Heated Tubing Temperature Range Heated Tubing Temperature Cut-out Material

15 mm (HT15) 12 mm (HT12)
1.83 m (6 ft.)
16° to 30° C (60° to 86° F)
 41° C ( 106° F)
Flexible plastic and electrical components

Disposal
Dispose of this device in accordance with local collections and recycling regulations. For more information, visit www.philips.com/recycling.

- 54 -

9. EMC information
Your unit has been designed to meet EMC standards throughout its Service Life without additional maintenance. There is always an opportunity to relocate your DreamStation 2 Therapy Device within an environment that contains other devices with their own unknown EMC behavior. If you believe your unit is affected by locating it closer to another device, simply separate the devices to remove the condition. Pressure and Flow Accuracy If you suspect that the pressure and/or flow rate accuracy is affected by EMC interference remove power and relocate the device to another area. If performance continues to be affected discontinue use and contact your home care provider. SpO2 and Pulse Rate Accuracy The DreamStation 2 therapy device is designed to capture the SpO and pulse rate oximetry
2
data within the accuracy specification described in the sensor manufacturer's instructions for use. When 4 hours of successful oximetry data have been achieved the device indicates this to the user by displaying "SpO2 Successful". If you suspect that your unit is affected by EMC interference, remove power and relocate the device to another area. If performance continues to be affected discontinue use and contact your home care provider.
- 55 -

Guidance and Manufacturer's Declaration - Electromagnetic Emissions ­ This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions CISPR 11

Group 1

The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/Flicker emissions IEC 61000-3-3

Class B Class A Complies

The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network.

Emission of Radio Frequency Energy RTCA/DO-160G Section 21

Category M

This device is suitable for use onboard commercial airplanes inside passenger cabin.

Guidance and Manufacturer's Declaration - Electromagnetic Immunity ­ This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.

Immunity Test

IEC 60601 Test Level Compliance Level

Electrostatic Discharge (ESD) IEC 61000-4-2

±8 kV contact
±2 kV, ±4 kV, ±8 kV, & ±15 kV air discharges

±8 kV contact
±2 kV, ±4 kV, ±8 kV, & ±15 kV air discharges

Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 35%.

- 56 -

Immunity Test

IEC 60601 Test Level Compliance Level

Electromagnetic Environment Guidance

Electrical fast Transient/ burst IEC 61000-4-4

±2 kV for power supply lines, 100 kHz repetition rate
±1 kV for input-output lines; 100 kHz repetition rate

±2 kV for power supply lines, 100 kHz repetition rate
N/A- Device does not have user I/O lines that are longer than 3m in length.

Mains power quality should be that of a typical home or hospital environment.

Surge IEC 61000-4-5

±1 kV differential mode ±2 kV common mode

±1 kV differential mode
N/A: Device is Class 2 and does not have earth connection.

Mains power quality should be that of a typical home or hospital environment.

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

<5% U (>95% dip in U )

T

T

for 0.5 cycle at 45 degree

increments

<5% UT (>95% dip in UT) for 1 cycle

70% UT (30% dip in UT) for 0.5 seconds

<5% UT (>95% dip in UT) for 5 seconds

<5% U (>95% dip in U )

T

T

for 0.5 cycle at 45 degree

increments

<5% UT (>95% dip in UT) for 1 cycle

70% UT (30% dip in UT) for 0.5 seconds

<5% UT (>95% dip in UT) for 5 seconds

Mains power quality should be that of a typical home or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.

Power frequency (50/60 30 A/m Hz) magnetic field IEC 61000-4-8

30 A/m

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical hospital or home environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

- 57 -

Guidance and Manufacturer's Declaration - Electromagnetic Immunity ­ This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.

Immunity Test IEC 60601 Test Level Compliance Level

Electromagnetic Environment - Guidance

Conducted RF 3 Vrms IEC 61000-4-6 150 kHz to 80 MHz

6 Vrms Amateur Radio & ISM Bands between 150 kHz and 80 MHz

Radiated RF

10 V/m

IEC 61000-4-3 80 MHz to 2.7 GHz

3 Vrms 150 kHz to 80 MHz
6 Vrms Amateur Radio & ISM Bands between 150 kHz and 80 MHz
10 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended 30 cm (12 inches) separation distance.
Interference may occur in the vicinity of equipment marked with the following symbol:

Telecommunication frequencies as specified in clause 8.10 of IEC 60601-1-2:2014:

450, 810, 870, 930, 1720, 1845, 1970, and 2450 MHz at 28 V/m

28 V/m

385 MHz at 27 V/m
710, 745, 780, 5240, 5500, and 5785 MHz at 9 V/m

27 V/m 9 V/m

- 58 -

If you plan on using your DreamStation 2 on an airplane, please detach this letter and bring it with you.

FAA Compliance
To Whom It May Concern:
The following Philips DreamStation 2 devices are in compliance with commercial airline EMI/RFI requirements:
· DreamStation 2 CPAP · DreamStation 2 Auto CPAP
Philips has designed and tested the identified devices for compliance with section 21, Category M, RTCA DO-160 EMI/RFI requirements as specified in the Code of Federal Regulations 14 CFR 382 "Nondiscrimination on the Basis of Disability in Air Travel; Final Rule". In accordance with these requirements, the aforementioned components may be used onboard an aircraft without further testing by the carrier. If you have any other questions regarding our products, please feel free to call the Philips Respironics Customer Services department 1-724-387-4000. You can also use the following address:

Respironics Inc. 1001 Murry Ridge Lane Murrysville, PA 15668
www.philips.com/respironics

- 59 -

- 60 -

Limited Warranty
Respironics, Inc., a Philips company ("Philips Respironics") provides this non-transferable, limited warranty for DreamStation 2 ("Product") to the customer who originally purchased the Product directly from Philips Respironics. What this Warranty Covers: Philips Respironics warrants each new Product will be free from defects in materials and workmanship and will perform in accordance with the Product specifications under normal and proper use and maintenance in accordance with applicable instructions, subject to the exclusions below. How Long does this Warranty Last: Two (2) years from the longer of the date of shipment to the purchaser or date of setup by purchaser for the end user, except: The warranty period for accessories, replacement parts, and disposables including, but not limited to, tubing, filters, carrying case, and power cord is 90 days from the date of shipment to the original purchaser. What this Warranty does not cover: This warranty does not apply to any software included with the Product as the software warranty is included in the software license. This warranty does not cover damage or injury whether to the Products, personal property, or persons caused by accident, misuse, abuse, Acts of God, water ingress, repair or alteration by anyone other than Philips Respironics or its authorized service center, failure to operate in accordance with the terms of the operating manual and instructions, lack of reasonable care, the discontinuance of a network (e.g. 2G, 3G, etc.) by a carrier (e.g. ATT, Verizon, etc.), or other defects not related to material or workmanship. This warranty is not transferable. If Philips Respironics finds that a Product returned for service or the issue raised is not covered under this limited warranty, Philips Respironics may charge an evaluation fee and return shipping. What Philips Respironics will do: If a Product does not meet the warranty above in the first 90 days after the original shipment date, Philips Respironics will replace the device with a new Product. Thereafter, if a Product fails to conform to the warranties set forth above during the applicable warranty period, Philips Respironics will repair or replace the Product or refund the original purchase price, in Philips Respironics sole discretion. Philips Respironics may use new or remanufactured assemblies, components, and parts in repair and new or recertified refurbished devices for replacement. The balance of the original warranty period will apply to any Product or component of a Product repaired or replaced under this warranty.
- 61 -

Warranty Disclaimer; Limitation of Liability: EXCEPT AS SET FORTH IN THIS LIMITED WARRANTY, PHILIPS RESPIRONICS MAKES NO WARRANTIES, EXPRESSED OR IMPLIED, STATUTORY OR OTHERWISE, REGARDING THE PRODUCT OR ITS QUALITY OR PERFORMANCE. PHILIPS RESPIRONICS SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTY OF MERCHANTABILITY AND THE IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT WILL PHILIPS RESPIRONICS MAXIMUM LIABILITY UNDER THESE WARRANTIES EXCEED THE ORIGINAL PURCHASE PRICE OR WILL PHILIPS RESPIRONICS BE LIABLE FOR ANY ECONOMIC LOSS, LOSS OF PROFITS, OVERHEAD, OR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES. Repair, replacement, or return of purchase price by Philips Respironics is the original purchaser's sole and exclusive remedy under this warranty. This warranty gives you specific legal rights, and you may also have other rights that vary from state to state. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above exclusion and limitations may not apply to you. How to get warranty support: Patients contact your local authorized Philips Respironics dealer and dealers contact Respironics, Inc. at:
1001 Murry Ridge Lane Murrysville, Pennsylvania 15668-8550 +1-724-387-4000
Note: For Australian and New Zealand customers this warranty replaces the warranty contained above.
1. The following statement is provided to a customer who is a consumer under the Australian Consumer Law: Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the good repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.
2. The following statement is provided to a customer who is a consumer under the Consumer Guarantees Act 1993, New Zealand: Our goods come with guarantees that cannot be excluded under the Consumer Guarantees Act 1993. This guarantee applies in addition to the conditions and guarantees implied by that legislation.
- 62 -

3. Philips warrants that the products shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of five (5) years from the date of purchase from an authorised Philips Homecare Provider. This Warranty covers the replacement or repair at the option of Philips, of any product that has a manufacturing or material defect that is not the result of normal wear and tear, or a natural characteristic of the material used. This Warranty is not transferable and does not cover products used for commercial purposes, and it does not apply to any consumable items (including but not limited to filters, masks, tubes and humidifier chambers).
4. The customer is responsible for returning the product to an authorised Philips Homecare Provider, and collecting the product from the Homecare Provider after repair or replacement, at its own cost. Philips is responsible only for the freight cost of transporting the product between the Homecare Provider and Philips. Philips reserves the right to charge an evaluation and postage fee for any returned product where no problem is found following evaluation.
5. This Warranty does not cover: · products purchased outside of Australia and New Zealand · any damage caused as a result of misuse or abuse, modification, tampering with or alteration of the product, pest infestation, or liquid egress into the product · contamination due to cigarette, pipe, cigar or other smoke · failure to follow manufacturer's instruction for use as per user's manual · defects that are a consequence of repairs to a product made or attempted by a service provider other than one approved by Philips · products that have been subjected to incorrect electrical supply or inconsistent electrical supply or used with inappropriate accessories.
6. This Warranty is not transferrable in the event of any resale or transfer of products. 7. To make a claim under this Warranty, contact your Homecare Provider. Alternatively, contact: Philips
Electronics Australia Limited, 65 Epping Road, North Ryde NSW 2113 Australia. Tel: 1300 766 488, Email: repairs-src@philips-easyconnect.com
AUSTRALIAN SPONSOR DETAILS: Philips Electronics Australia Ltd. 65 Epping Road, North Ryde, NSW 2113 Australia