NIDEK Tonoref III Quick Start Guide

Training - NIDEK Tonoref III

Download

NIDEK Tonoref III - Birmingham Optical

Nidek Tonoref III QUICK GUIDE

25 set 2024 — QUICK GUIDE. • 4 in 1 testing - Autorefraction ... The Nidek Tonoref III has a gentle nozzle design to create a quieter puff of air, again aiming.

PDF preview unavailable. Download the PDF instead.

NIDEK-Tonoref-III-Quick-Start-Guide
Nidek Tonoref III QUICK GUIDE

Features:
· 4 in 1 testing - Autorefraction, keratometry, pachymetry, non-contact tonometry · Fully automated capture and eye tracking · Tiltable, 7-inch, colour screen · Quick and comfortable measurements for patient and clinician · Connectivity to other Nidek devices (eg. RT phropter series, LM focimeter series) · Advanced assessment tools including accommodation measurement and opacity
imaging

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Autorefraction:
The Nidek Tonoref III calculates a quick and automatic measurement of patient's refractive error (prescription) without any patient input (objectively). The device uses a large pupil imaging method to simultaneously measure patient prescription at 2 pupil diameters (large and small).

This improves accuracy and allows an estimation of the effects on pupil size on night vision. If the patient has a large difference in their prescriptions at small and large pupil diameters, this may indicate a different spectacle prescription is needed at night (eg. Night driving).
The minimum pupil size required for autorefraction measurement using the Nidek Tonoref III is 2mm.
During measurement, the patient is asked to fixate on using a distance target (hot air balloon). A distance target is used to help relax patient accommodation during measurement.

The patient should be advised that the target will change focus throughout the measurement. This is to further relax their eyes.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Keratometry:
The Nidek Tonoref III calculates a quick and automatic measurement of patient's corneal curvature. This measurement is essential for the fitting of contact lenses. The device uses the same double ring method as used in autorefraction capture. By using 2 different measurement mires, the Nidek Tonoref II improves accuracy, as the keratometry measurement is less likely to be influenced by the patient eyelid.

As the keratometry mires are produced by infrared light, which is invisible to the human eye, the patient does not have a specific fixation target during measurement.
Instead they should be asked to fixate straight ahead.
Most commonly, keratometry measurements are taken at the same time as autorefraction on the Nidek Tonoref III.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Pachymetry:
Pachymetry is the measurement of corneal thickness. This measurement gives the optometrist important information about the health of the patient's cornea and the potential risk of developing eye disease such as glaucoma. Traditional pachymetry methods involve touching the patient's eye with an ultrasound probe. The Nidek Tonoref III uses infrared light to conduct quick and accurate pachymetry measurement without contact. As infra-red light is invisible to the human eye, the patient has no specific fixation target during measurement and instead should be asked to look straight ahead Most commonly, pachymetry measurements are taken at the same time as tonometry on the Nidek Tonoref III. The pachymetry and tonometry measurements can be used together to calculate a corrected IOP (intraocular pressure) reading.

Without pachymetry, tonometry devices assume that all patient's have the same corneal thickness. In reality, corneal thickness can vary greatly between patients. The thickness of the cornea effects how the eye will respond to the `puff of air' used in the tonometry measurement. The response of the cornea will affect the final tonometry result. This means that non-contact tonometry devices without pachymetry, could be under or overestimating a patient's IOP, and therefore could be miscalculating the patient's risk of having or developing glaucoma.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

NCT (Non-Contact Tonometry):
The Nidek Tonoref III performs non-contact measurements of patient IOP. Non-contact tonometry measurements are calculated by firing a `puff of air' at the patient's cornea. The IOP is calculated by the amount of air pressure needed to flatten the cornea. This technique can be uncomfortable and is often unpopular with patients. The Nidek Tonoref III uses Automated Puff Control (APC) to create a more `patient friendly' air puff. For a final IOP to be calculated, the measurement must be repeated several times. The Nidek APC technology adapts the amount of air pressure based on the first measurement taken. This ensures the minimum amount of air pressure is `puffed' at the patient.

The Nidek Tonoref III has a gentle nozzle design to create a quieter puff of air, again aiming to improve patient comfort.
During IOP measurement, the patient should be asked to look at the green fixation light within the air nozzle.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

TOUCH STCoRuEEcNh Screen:

The Nidek Tonoref III has a 7-inch, tilt-able touch screen display. This displays the measurement screen as below:

To change measurement
mode

Autorefraction/ Keratometry mode

Pachymetry/ NCT mode

To select individual measurement within measurement mode

To toggle automatic functions

Delete data
Print

Auto tracking Auto capture

Eye being tested will be highlighted
BLUE

Accommodation measure mode
Retroillumination
Opacity assessment
Manual measure
mode Vision compare mode
Summary
Settings

Base unit:

The Nidek Tonoref III base unit is simplified to assist with ease of use.

Brake / Stop
Chinrest adjust buttons

Lock
Joystick (and capture button)

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Measurement Capture:
STEP 1: Patient positioning
Before measurement can be captured, the patient needs to place their chin the chin rest and forehead against the bar.
The patient's eyes should be aligned with the eye level markers on the side of the forehead rest support.

The patient position can be adjusted by altering the height of the supporting table or the patient chinrest, using the directional arrows on the base unit.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

STEP 2: Capture mode selection
The capture mode and specifc measurement to be taken should be indentified, the test explained to the patient and the correct fixation target advised.

STEP 3: Patient Alignment To begin measurement, the joystick should be used to align the capture rings with the centre of the patient's pupil.
Once aligned with the patient's pupil, the base unit should be pushed forward towards the patient. Measurement will begin automatically when the capture rings are yellow.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Guide arrows will appear if device needs to be moved closer to, or further away from the patient.

STEP 4: RK Measurement Mode It is possible to measure autorefraction, keratometry, or both together in RK mode. It is possible to combine the measurements with NT mode by leaving all capture modes on the display screen before capture start.

The patient should be advised to fixate on the hot air balloon target during measurement.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

STEP 5: NT Measurement Mode
It is possible to measure pachymetry, tonometry, or both together in NT mode. It is possible to capture NT measurements directly after RK, if all capture modes are on display screen at start of measurement of capture.
After RK is complete, `PULL BACK' will display on the screen. The operator will need to pull the device away from the patient, where it will automatically lower ready NT measurement. Before NT measurement will begin, the operator will need to realign with the centre of the patient's pupil. The patient should be advised to fixate on the green light in the centre of the air nozzle. Measurement will begin when the alignment lines appear yellow.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

STEP 6: Summary and print out
Once measurements are complete, results can be viewed on the summary screen of the display unit or printed.
The summary page displays all the patient data on a single screen for the optometrist to analyse.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

The Printout explained:

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Advanced assessment features:
The Nidek Tonoref III can perform a number of advanced assessments.

Accommodation measure mode
Retroillumination
Opacity assessment
Manual measure
mode
Vision compare
mode

ACCOMODATION MODE
The Nidek Tonoref III can provide an assessment of a patient's accommodation using the `Accommodation Measure Mode'. The patient is advised to fixate on a distance target (hot air balloon).

The target will be moved between a distance and near focus in ± 0.25D increments. The device measures the patient's objective response to change in focus over a 30 second period. The Nidek Tonoref III will then plot the patient's accommodative response on a graph:

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

As with other capture modes on the Nidek Tonoref III, the operator will need to align with the centre of the patient pupil Measurement will be started by selecting the capture button in the centre of the joystick. TIP! - Accommodation measurements can only be taken after RK measurement is complete.
RETRO-ILLUMINATION OPACITY ASSESSMENT The Nidek Tonoref III contains a retro-illumination camera for the improved imaging of media opacities.

The Nidek cataract indices can give clinicians a rough indication of severity of opacities and allow easier monitoring.
TIP! - Cataract indices are a guide only and will be variable depending on exact alignment.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

MANUAL MEASURE MODE
The manual measure mode allows the operator to take measurements of the anterior eye structures such as cornea and pupil size under both photopic and mesopic conditions.
It also acts as a dispensing tool by allowing measurements such as PD and heights (NB. PD is measured automatically during auto-refraction stage).
Once the operator has aligned with patient pupil, the capture button on the centre of the joystick should be pressed to begin measurement.

The arrows can then be used to align markers. The central button of the joystick can then again be pressed to finalise measurement.
VISION COMPARE MODE
The Nidek Tonoref III `Vision Compare Mode', allows a quick comparison of patient's visual acuity with autorefractor reading and patient's current spectacles and / or their unaided visions. To compare visual acuity with autorefractor to patient's current spectacle prescription, the Nidek Tonoref III must be connected to the lens meter (LM) and the data must have been inputted.

www.birminghamoptical.co.uk

0808 0123 2020

www.birminghamopticaleducation.co.uk/

Once in `Vision Compare Mode', the internal scenery will change to distance view. The orange spectacles can then be pressed to change between patient's autorefraction prescription and inputted LM data. The same process can be done for reading by pressing the book icon on the RHS of the screen. TIP! - There is no facility to compare with patient's SUBJECTIVE (ie. Final) prescription that would be found during the sight test with the optometrist.
CONTACT LENS SELECTION MODE
The Nidek Tonoref III contains software to help with ideal contact lens selection based on patient's RK measurements. The contact lens selection mode can be accessed on the summary screen once all measurements have completed. The Nidek Tonoref III will then attempt to select the best lens size, BC (base curve / BOZR) and prescription for the patient compared to the internal contact lens database. TIP! - The current Nidek Tonoref III contact lens database does not contain many lenses commonly used in the UK market.