Thermo Scientific Small Benchtop Rotors

50143334-e - Thermo Scientific Small Benchtop Rotors - Instruction Manual

50143334-e - Thermo Scientific Small Benchtop Rotors - Instruction Manual - Thermo Fisher Scientific

PDF preview unavailable. Download the PDF instead.

50143334-e - Thermo Scientific Small Benchtop Rotors - Instruction Manual
Thermo Scientific Small Benchtop Rotors
for Thermo Scientific Megafuge 8 / 8R, Sorvall ST 8 / 8R and SL 8 / 8R Centrifuges Instruction Manual
50143334-e · 07 / 2020
Visit us online to register your warranty:
thermofisher.com/labwarranty

WEEE Compliance
This product is required to comply with the European Union's Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked with the following symbol:

Table of Contents
WEEE Compliance. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 2
Preface .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 7
Intended Use.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 7 Precautions.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 7 Symbols Used in the manual.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 8
Thermo Scientific Rotor Specifications .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 9
TX-150.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 10 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 10 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 10 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 14 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 15
TX-100S.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 16 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 16 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 16 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 17 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 18
TX-100.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 19 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 19 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 19 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 20
M10 .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 21 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 21 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 21 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 23 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 24
MT-12. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 25 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 25 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 25
HIGHConic III.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 27 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 27 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 27 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 28 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 29
CLINIConic.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 30 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 30 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 30 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 31
MicroClick 18 x 5 .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 32 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 32 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 32

Thermo Scientific

Small Benchtop Rotors | 3

Table of Contents

Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 33 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 33 MicroClick 24 x 2 .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 34 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 34 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 34 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 35 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 36 MicroClick 30 x 2 .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 37 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 37 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 37 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 38 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 39 Microliter 48 x 2.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 40 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 40 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 40 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 41 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 42 8 x 8 PCR Strip .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 43 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 43 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 43 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 44 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 45 8 x 50 mL Individually Sealed .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 46 Items Supplied .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 46 Technical Data .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 46 Accessories .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 47 Biocontainment Certificate.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 48
Thermo Scientific Auto-Lock Rotor Exchange .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 49
Rotor Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Removing the Rotor .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 50
Rotor Loading.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 51
Before a Run.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 51 Rotor Temperature Range. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 51 Rotor Loading.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 51
Correct Balances. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 52 Incorrect Balances. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 52 Maximum Loading .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 52 Rotor Life Time.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 53 Service Life Examples.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 53
Aerosol-tight Applications.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 55
Basic Principles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Placing O-Ring .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 55

4 | Small Benchtop Rotors

Thermo Scientific

Table of Contents
Fill Level. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 56 Checking the Aerosol-Tightness .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 56
Quick Test. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 56 Aerosol-tight Closure with ClickSeal.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 57 Removing of Aerosol-tight Rotors.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 58
Maintenance and Care.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 59
Cleaning Intervals.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 59 Basics .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 59
Rotor and Accessories Inspection. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 59 Cleaning. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 60 Disinfection.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 61 Decontamination.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 62 Autoclaving.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 63 Service of Thermo Fisher Scientific. .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 63 Storage .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 63 Shipping and Disposal.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 64
Chemical Compatibility Chart.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 65
Index.  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . 71

Thermo Scientific

Small Benchtop Rotors | 5

P
Preface
Before starting to use a rotor, read through this instruction manual carefully and follow the instructions. Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers warranty.
Intended Use
This centrifuge is a laboratory product used to separate substance mixtures of different densities. This centrifuge has to be operated by trained specialists only.
Precautions
In order to ensure safe operation of these Thermo Scientific small benchtop rotors, the following general safety regulations must be followed: · Do not remove any components of the rotor. · Do not use rotors which show any signs of corrosion and/or cracks. Do not touch the electronic components of the
centrifuge and do not make any changes to the electronic or mechanical components. · Use only with rotors which have been loaded properly. · Never overload these rotors. · Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this rule are
commercially available glass or plastic centrifuge tubes, provided they have been approved for the speed or the RCF value of the rotor. · Observe the safety instructions. Pay particular attention to the following aspects: · Rotor installation: Check that the rotor is locked properly into place before operating the centrifuge. · Always balance the samples. Maximum sample density at maximum speed: 1.2

Thermo Scientific

Small Benchtop Rotor | 7

Symbols Used in the manual
Symbols Used in the manual
This symbol refers to general hazards. CAUTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur.
This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe.

8 | Small Benchtop Rotors

Thermo Scientific

Thermo Scientific Rotor Specifications

1

There are several Thermo ScientificTM small benchtop rotors available for the Thermo Scientific MegafugeTM 8 / 8R centrifuge, the Thermo Scientific SorvallTM ST 8 / 8R centrifuge and the Thermo Scientific SL 8 / 8R centrifuge.

Thermo Scientific Rotors

Article No.

TX-150 swinging bucket rotor

75005701

 TX-150 round buckets

75005702

 TX-150 50mL conical buckets

75005703

TX-100S clinical swinging bucket rotor with sealed carriers

75005704

TX-100 clinical swinging bucket rotor with carriers

75005705

M10 microplate swinging bucket rotor

75005706

 M10 buckets

75005723

 M10 sealed buckets

75005721

MT-12 microtube swinging bucket rotor

75005600

HIGHConic III fixed angle rotor

75005709

CLINIConic fixed angle rotor

75003623

MicroClick 18 x 5 microtube rotor

75005765

MicroClick 24 x 2 microtube rotor

75005715

MicroClick 30 x 2 microtube rotor

75005719

Microliter 48 x 2 microtube rotor

75003602

8 x 8 PCR Strip rotor

75005720

8 x 50 mL Individually Sealed rotor

75003694

Hematocrit rotor*

75005733

* Look at the separate Thermo Scientific Hematocrit Rotor Instruction Manual for specific data and instructions.

Thermo Scientific

Small Benchtop Rotors | 9

Chapter 1 | TX-150

TX-150

Items Supplied

Article No. 75005701 75003786 50139640

Item TX-150 Rotor Bolt Grease Instruction Manual

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1

Technical Data
TX-150 with round buckets

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

4 x 190 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 12968 50000

Radius max. / min.

144 mm / 51 mm

Angle

90°

Acceleration / Braking Time

20 s / 30 s

Maximum Speed at 4 °C

50 Hz: 4500 rpm 60 Hz: 4500 rpm

Sample Temperature at Max. Speed
(Ambient temperature of 23 °C, run time 90 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Optional

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 4 x 190 g 4500 rpm 3260 x g 12968 50000 144 mm / 51 mm 90° 20 s / 30 s
60 Hz: 4500 rpm
60 Hz: < 4 °C
Optional 121 °C

100 V 2.9 kg 4 x 190 g 4500 rpm 3260 x g 12968 50000 144 mm / 51 mm 90° 20 s / 30 s 50 Hz: 4500 rpm 60 Hz: 4500 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Optional
121 °C

10 | Small Benchtop Rotors

Thermo Scientific

TX-150 | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

4 x 190 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 12968 50000

Radius max. / min.

144 mm / 51 mm

Angle

90°

Acceleration / Braking Time

12 s / 18 s

Sample Temperature at nmax

5 °C

(Ambient temperature of 23 °C, run time 60 minutes)

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 4 x 190 g 4500 rpm 3260 x g 12968 50000 144 mm / 51 mm 90° 13 s / 19 s 5 °C Yes
121 °C

100 V 2.9 kg 4 x 190 g 4500 rpm 3260 x g 12968 50000 144 mm / 51 mm 90° 13 s / 19 s 5 °C Yes
121 °C

Thermo Scientific

Small Benchtop Rotors | 11

Chapter 1 | TX-150
TX-150 with conical buckets

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

4 x 150 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14532 50000

Radius max. / min.

144 mm / 45 mm

Angle

90°

Acceleration / Braking Time

20 s / 30 s

Maximum Speed at 4 °C

50 Hz: 4500 rpm 60 Hz: 4500 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Optional

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 4 x 150 g 4500 rpm 3260 x g 14532 50000 144 mm / 45 mm 90° 20 s / 30 s
60 Hz: 4500 rpm
60 Hz: < 4 °C
Optional 121 °C

100 V 2.9 kg 4 x 150 g 4500 rpm 3260 x g 14532 50000 144 mm / 45 mm 90° 20 s / 30 s
50 Hz: 4500 rpm 60 Hz: 4500 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Optional
121 °C

12 | Small Benchtop Rotors

Thermo Scientific

TX-150 | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

4 x 150 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14532 50000

Radius max. / min.

144 mm / 45 mm

Angle

90°

Acceleration / Braking Time

12 s / 18 s

Sample Temperature at nmax

7 °C

(Ambient temperature of 23 °C, run time 60 minutes)

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 4 x 150 g 4500 rpm 3260 x g 14532 50000 144 mm / 45 mm 90° 13 s / 19 s 7 °C No
121 °C

100 V 2.9 kg 4 x 150 g 4500 rpm 3260 x g 14532 50000 144 mm / 45 mm 90° 13 s / 19 s 7 °C No
121 °C

Thermo Scientific

Small Benchtop Rotors | 13

Chapter 1 | TX-150
Accessories

1

2

3

4

5

6

7

8

Article No.

Description

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

75005703

50 mL Conical Buckets (unsealed, no adapter needed) (set of 4)

8 x 50

29.5 x 120

75005702

Round Buckets (set of 4)

4 x 145

50 x 100

75005707

Click Seal Biocontainment Lids for Round Buckets (set of 4)

75005724

Replacement O-rings for Lids (set of 4)

Adapters for 50 mL Conical Buckets (sets of 2)

75005808

15 mL Conical Tube

8 x 15

17 x 123

Adapters for Round Buckets (sets of 4)

Direct fit

145 mL Bottle (75005734)

4 x 145

50 x 100

1 75005735

100 mL Round Bottom Open-Top Tube

4 x 100

45 x 117

2 75005736

50 mL Conical or Skirted Tube

4 x 50

29.5 x 120

3 75005744

30 mL SterilinTM Universal Container

4 x 30

25 x 120

4 75005737

15 mL Conical Tube

8 x 15

17 x 122

4 75005737

11 mL IVF Tube

8 x 11

17 x 122

5 75003504

13 mL Urine Tube

16 x 13

17 x 110

5 75003504

12 mL Blood Collection Tube (GreinerTM)

16 x 12

17 x 110

5 75003504

10 mL Blood Collection or 15 mL CorexTM/KimbleTM Tube

16 x 15

17 x 110

6 75005739

5/7 mL Blood Collection Tube

24 x 5/7

13 x 110

7 75005740

3/5 mL Blood Collection Tube or Cryotube

28 x 3/5

13 x 110

8 75005743

1.5/2 mL Microtube (or MicrotainerTM tube)

40 x 2

11 x 65

Rotor Packages

75005760

Cell Culture Package,

4 x 50

29.5 x 120

INCLUDES: TX-150 Rotor (75005701), Round Buckets (75005702), Adapters for 50 mL Conical Tubes (75005736)

75005761

High Capacity Cell Culture Package,

8 x 50

18 x 124

INCLUDES: TX-150 Rotor (75005701), Conical Buckets (75005703), Adapters for 15 mL Conical Tubes (75005808)

75005762

Clinical Rotor Package,

24 x 5/7

18 x 124

INCLUDES: TX-150 Rotor (75005701), Round Buckets (75005702), ClickSeal Biocontainment Lids (75005707), Adapters for Blood Collection Tubes: 5/7 mL (75005739) and 10 mL (75005738)

14 | Small Benchtop Rotors

Thermo Scientific

Biocontainment Certificate

TX-150 | Chapter 1

Thermo Scientific

Small Benchtop Rotors | 15

Chapter 1 | TX-100S

TX-100S

Items Supplied

Article No. 75005704 75003786 50139640 20059135 50110911 50110924 20059364 76001803

Item TX-100S Rotor Bolt Grease Instruction Manual Carriers Metal Sleeves Biocontainment Lids O-Rings Rubber Pads

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 4 8 8 8 8

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

3.1 kg

Maximum Permissible Load

8 x 25 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14813 50000

Radius max. / min.

144 mm / 45 mm

Angle

90°

Acceleration / Braking Time

20 s / 30 s

Maximum Speed at 4 °C

50 Hz: 4500 rpm 60 Hz: 4500 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Optional

Permissible Temperature Range Autoclavable 121 °C

120 V 3.1 kg 8 x 25 g 4500 rpm 3260 x g 14813 50000 144 mm / 45 mm 90° 20 s / 30 s
60 Hz: 4500 rpm
60 Hz: < 4 °C
Optional 121 °C

100 V 3.1 kg 8 x 25 g 4500 rpm 3260 x g 14813 50000 144 mm / 45 mm 90° 20 s / 30 s
50 Hz: 4500 rpm 60 Hz: 4500 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Optional
121 °C

16 | Small Benchtop Rotors

Thermo Scientific

TX-100S | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

3.1 kg

Maximum Permissible Load

8 x 25 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14258 50000

Radius max. / min.

144 mm / 44 mm

Angle

90°

Acceleration / Braking Time

12 s / 18 s

Sample Temperature at nmax

7 °C

(Ambient temperature of 23 °C, run time 60 minutes)

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 3.1 kg 8 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 44 mm 90° 13 s / 19 s 7 °C Yes
121 °C

100 V 3.1 kg 8 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 44 mm 90° 13 s / 19 s 7 °C Yes
121 °C

Accessories

Article No.

Description

Adapters for TX-100S Clinical Rotor (each)

11172596

5/7 mL BD HemogardTM/BD VacutainerTM Tube

11172595

5 mL BD Hemogard Tube

11172287

3 mL Blood Collection Tube

11172288

1.5/2 mL Microtube (or MicrotainerTM tube)

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

16/8 x 5/7 16/8 x 5 16/8 x 3 16/8 x 1.5/2

13 x 110 13 x 75 11 x 70 10 x 41

Thermo Scientific

Small Benchtop Rotors | 17

Chapter 1 | TX-100S
Biocontainment Certificate

18 | Small Benchtop Rotors

Thermo Scientific

TX-100 | Chapter 1

TX-100

Items Supplied

Article No. 75005705 75003786 50139640 20059134 50110904 76001803

Item TX-100 Rotor Bolt Grease Instruction Manual Carriers Metal Sleeves Rubber Pads

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 4 16 16

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

3.3 kg

Maximum Permissible Load

16 x 25 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14258 50000

Radius max. / min.

144 mm / 46 mm

Angle

90°

Acceleration / Braking Time

20 s / 30 s

Maximum Speed at 4 °C

50 Hz: 4500 rpm 60 Hz: 4500 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 3.3 kg 16 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 46 mm 90° 20 s / 30 s
60 Hz: 4500 rpm
60 Hz: < 4 °C
No 121 °C

100 V 3.3 kg 16 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 46 mm 90° 20 s / 30 s
50 Hz: 4500 rpm 60 Hz: 4500 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C No
121 °C

Thermo Scientific

Small Benchtop Rotors | 19

Chapter 1 | M10

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

3.3 kg

Maximum Permissible Load

16 x 25 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4500 rpm 3260 x g 14258 50000

Radius max. / min.

144 mm / 46 mm

Angle

90°

Acceleration / Braking Time

12 s / 18 s

Sample Temperature at nmax

7 °C

(Ambient temperature of 23 °C, run time 60 minutes)

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 3.3 kg 16 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 46 mm 90° 13 s / 19 s 7 °C No
121 °C

100 V 3.3 kg 16 x 25 g 4500 rpm 3260 x g 14258 50000 144 mm / 46 mm 90° 13 s / 19 s 7 °C No
121 °C

Accessories

Article No.

Description

Adapters for TX-100 Rotor (each)

Direct fit

13 mL Urine Tube

11172596

5/7 mL BD HemogardTM/BD VacutainerTM Tube

11172595

5 mL BD Hemogard Tube

11172287

3 mL Blood Collection Tube

11172288

1.5/2 mL Microtube (or MicrotainerTM tube)

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

16/8 x 13 16/8 x 5/7 16/8 x 5 16/8 x 3 16/8 x 1.5/2

17 x 110 13 x 110 13 x 75 11 x 70 10 x 41

20 | Small Benchtop Rotors

Thermo Scientific

M10

Items Supplied

Article No. 75005706 75003786 76003500 50136940

Item M10 Rotor Bolt Grease O-Ring Grease Instruction Manual

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1

Technical Data
M10 with Standard Carriers

M10 | Chapter 1

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

2 x 125 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4400 rpm 2576 x g 5189 30000

Radius max. / min.

119 mm / 80 mm

Angle

90°

Acceleration / Braking Time

25 s / 25 s

Maximum Speed at 4 °C

50 Hz: 4400 rpm 60 Hz: 4400 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Optional

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 2 x 125 g 4400 rpm 2576 x g 5189 30000 119 mm / 80 mm 90° 20 s / 25 s
60 Hz: 4400 rpm
60 Hz: < 4 °C Optional 121 °C

100 V 2.9 kg 2 x 125 g 4400 rpm 2576 x g 5189 30000 119 mm / 80 mm 90° 30 s / 25 s
50 Hz: 4400 rpm 60 Hz: 4400 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Optional
121 °C

Thermo Scientific

Small Benchtop Rotors | 21

Chapter 1 | M10

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

2 x 125 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4400 rpm 2576 x g 5189 30000

Radius max. / min.

119 mm / 80 mm

Angle

90°

Acceleration / Braking Time

15 s / 23 s

Sample Temperature at nmax

5 °C

(Ambient temperature of 23 °C, run time 60 minutes)

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 2 x 125 g 4400 rpm 2576 x g 5189 30000 119 mm / 80 mm 90° 14 s / 23 s 5 °C No
121 °C

100 V 2.9 kg 2 x 125 g 4400 rpm 2576 x g 5189 30000 119 mm / 80 mm 90° 14 s / 23 s 5 °C No
121 °C

22 | Small Benchtop Rotors

Thermo Scientific

M10 with Biocontainment Carriers

M10 | Chapter 1

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.9 kg

Maximum Permissible Load

2 x 300 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4400 rpm 2576 x g 5189 30000

Radius max. / min.

119 mm / 63 mm

Angle

90°

Acceleration / Braking Time

25 s / 25 s

Maximum Speed at 4 °C

50 Hz: 4400 rpm 60 Hz: 4400 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Optional

Permissible Temperature Range Autoclavable 121 °C

120 V 2.9 kg 2 x 300 g 4400 rpm 2576 x g 5189 30000 119 mm / 63 mm 90° 20 s / 25 s
60 Hz: 4400 rpm
60 Hz: < 4 °C Optional 121 °C

100 V 2.9 kg 2 x 300 g 4400 rpm 2576 x g 5189 30000 119 mm / 63 mm 90° 30 s / 25 s
50 Hz: 4400 rpm 60 Hz: 4400 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Optional
121 °C

Accessories

Article No. 75005723 75005721

Description Unsealed Buckets (set of 2) Sealed Buckets (set of 2) Rmax 109 mm

Rotor Capacity
(places x volume, mL)
4 Standard or 2 Midi-Deepwell
4 Standard or 2 Midi-Deepwell

Max Tube Dimensions (Ø x L, mm) Height < 33 mm
Height < 33 mm

Thermo Scientific

Small Benchtop Rotors | 23

Chapter 1 | M10
Biocontainment Certificate

24 | Small Benchtop Rotors

Thermo Scientific

MT-12 | Chapter 1

MT-12

Items Supplied

Article No. 75005600 50136940 70056346 20056966

Item MT-12 Rotor Instruction Manual ClickSeal Biocontainment Lid Metal Sleeves

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 12

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.8 kg

Maximum Permissible Load

12 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

13000 rpm 16438 x g 954 50000

Radius max. / min.

87 mm / 46 mm

Angle

90°

Acceleration / Braking Time

40 s / 50 s

Maximum Speed at 4 °C

50 Hz: 13000 rpm 60 Hz: 13000 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.8 kg 12 x 4 g 13000 rpm 16438 x g 954 50000 87 mm / 46 mm 90° 40 s / 50 s
60 Hz: 13000 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 1.8 kg 12 x 4 g 13000 rpm 16438 x g 954 50000 87 mm / 46 mm 90° 45 s / 50 s
50 Hz: 13000 rpm 60 Hz: 13000 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

Thermo Scientific

Small Benchtop Rotors | 25

Chapter 1 | HIGHConic III

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.8 kg

Maximum Permissible Load

12 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

13000 rpm 16438 x g 954 50000

Radius max. / min.

87 mm / 46 mm

Angle

90°

Acceleration / Braking Time

27 s / 35 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

22 °C

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 1.8 kg 12 x 4 g 13000 rpm 16438 x g 954 50000 87 mm / 46 mm 90° 24 s / 33 s 22 °C No
121 °C

100 V 1.8 kg 12 x 4 g 13000 rpm 16438 x g 954 50000 87 mm / 46 mm 90° 24 s / 33 s 22 °C No
121 °C

26 | Small Benchtop Rotors

Thermo Scientific

HIGHConic III | Chapter 1

HIGHConic III

Items Supplied

Article No. 75005709 50139640 20059176 75003058 76003500

Item HIGHConic III Rotor Instruction Manual Biocontainment Lid O-Ring Set O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.7 kg

Maximum Permissible Load

6 x 75 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

9500 rpm 12108 x g 2087 50000

Radius max. / min.

120 mm / 57 mm

Angle

45°

Acceleration / Braking Time

40 s / 45 s

Maximum Speed at 4 °C

50 Hz: 8700 rpm 60 Hz: 8700 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.7 kg 6 x 75 g 9500 rpm 12108 x g 2087 50000 120 mm / 57 mm 45° 45 s / 45 s
60 Hz: 8700 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 2.7 kg 6 x 75 g 9500 rpm 12108 x g 2087 50000 120 mm / 57 mm 45° 55 s / 45 s
50 Hz: 8700 rpm 60 Hz: 8700 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

Thermo Scientific

Small Benchtop Rotors | 27

Chapter 1 | HIGHConic III

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.7 kg

Maximum Permissible Load

6 x 75 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

8700 rpm 10155 x g 2488 50000

Radius max. / min.

120 mm / 57 mm

Angle

45°

Acceleration / Braking Time

28 s / 40 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

14 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.7 kg 6 x 75g 8700 rpm 10155 x g 2488 50000 120 mm / 57 mm 45° 27 s / 40 s 14 °C Yes
121 °C

100 V 2.7 kg 6 x 75g 8700 rpm 10155 x g 2488 50000 120 mm / 57 mm 45° 27 s / 40 s 14 °C Yes
121 °C

Accessories

Article No.

Description

75005731

Replacement Lid (each)

75003058

Replacement Sealing O-rings (set of 2 with grease)

Adapters for HIGHConic III Rotor (set of 2)

Direct fit 75005802 75005803

50 mL Round Bottom Tube 38 mL Round Bottom Tube 16 mL Round Bottom Tube

75005808

15 mL Conical Tube

75005804 75005805 75005770

12 mL Round Bottom Tube 6.5 mL Round Bottom Tube 5 mL Conical Microtube

75005806 75005807

3.5 mL Round Bottom Tube 1.5/2 mL Microtube

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

6 x 50 6 x 38 6 x 16 6 x 15 6 x 12 6 x 6.5 6 x 5 12 x 3.5 12 x 2

30 x 115 25.5 x 110 18 x 123 17 x 123 16 x 95 13.5 x 114 17 x 100 11 x 100 11 x 40

28 | Small Benchtop Rotors

Thermo Scientific

Biocontainment Certificate

HIGHConic III | Chapter 1

Thermo Scientific

Small Benchtop Rotors | 29

Chapter 1 | CLINIConic

CLINIConic

Items Supplied

Article No. 75003623 50139640 50110904 76001803

Item CLINIConic Rotor Instruction Manual Metal Sleeves Rubber Pads

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 30 30

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

4.7 kg

Maximum Permissible Load

30 x 30 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4400 rpm 3030 x g 6521 50000

Radius max. / min.

140 mm / 85 mm

Angle

37°

Acceleration / Braking Time

25 s / 30 s

Maximum Speed at 4 °C

50 Hz: 4400 rpm 60 Hz: 4400 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 4.7 kg 30 x 30 g 4400 rpm 3030 x g 6521 50000 140 mm / 85 mm 37° 25 s / 30 s
60 Hz: 4400 rpm
60 Hz: < 4 °C
No 121 °C

100 V 4.7 kg 30 x 30 g 4400 rpm 3030 x g 6521 50000 140 mm / 85 mm 37° 30 s / 30 s
50 Hz: 4400 rpm 60 Hz: 4400 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C No
121 °C

30 | Small Benchtop Rotors

Thermo Scientific

MicroClick 18 x 5 | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

4.7 kg

Maximum Permissible Load

30 x 30 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

4400 rpm 3030 x g 6521 50000

Radius max. / min.

140 mm / 85 mm

Angle

37°

Acceleration / Braking Time

16 s / 27 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

14 °C

Aerosol-tight

No

Permissible Temperature Range Autoclavable 121 °C

120 V 4.7 kg 30 x 30g 4400 rpm 3030 x g 6521 50000 140 mm / 85 mm 37° 18 s / 26 s 14 °C No
121 °C

100 V 4.7 kg 30 x 30g 4400 rpm 3030 x g 6521 50000 140 mm / 85 mm 37° 18 s / 26 s 14 °C No
121 °C

Accessories

Article No.

Description

Adapters for CLINIConic Rotor (each)

Direct fit 75008817 11172596

15 mL Round/Conical Bottom Tube 10 mL Round Bottom Tube 5/7 mL BD Hemogard / BD Vacutainer Tube

11172595

5 mL BD Hemogard Tube

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

30 x 15 30 x 10 30 x 5/7 30 x 5

16.5 x 131 16.5 x 95 13 x 106 13 x 75

Thermo Scientific

Small Benchtop Rotors | 31

Chapter 1 | MicroClick 18 x 5

MicroClick 18 x 5

Items Supplied

Article No. 75005765 50139640 20059119 75005726 76003500

Item MicroClick 18 x 5 Rotor Instruction Manual ClickSeal Lid O-Ring Set O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.7 kg

Maximum Permissible Load

18 x 9 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

14000 rpm 22351 x g 486 50000

Radius max. / min.

102 mm / 70 mm

Angle

45°

Acceleration / Braking Time

55 s / 55 s

Maximum Speed at 4 °C

50 Hz: 12400 rpm 60 Hz: 13000 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.7 kg 18 x 9 g 14000 rpm 22351 x g 486 50000 102 mm / 70 mm 45° 50 s / 55 s
60 Hz: 13500 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 1.7 kg 18 x 9 g 14000 rpm 22351 x g 486 50000 102 mm / 70 mm 45° 60 s / 55 s 50 Hz: 13700 rpm 60 Hz: 14000 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

32 | Small Benchtop Rotors

Thermo Scientific

Accessories

Article No. Description

Adapters for MicroClick 18 x 5 Rotor (set of 2)

75005756

1.5/2 mL Microtube

Biocontainment Certificate

MicroClick 18 x 5 | Chapter 1

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

18 x 1.5/2

11 x 45

Thermo Scientific

Small Benchtop Rotors | 33

Chapter 1 | MicroClick 24 x 2

MicroClick 24 x 2

Items Supplied

Article No. 75005715 50139640 20058086 75003405 76003500

Item MicroClick 24 x 2 Rotor Instruction Manual ClickSeal Lid O-Ring Set O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.2 kg

Maximum Permissible Load

24 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

17850 rpm 30279 x g 406 50000

Radius max. / min.

85 mm / 51 mm

Angle

45°

Acceleration / Braking Time

35 s / 45 s

Maximum Speed at 4 °C

50 Hz: 15200 rpm 60 Hz: 16800 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.2 kg 24 x 4 g 17850 rpm 30279 x g 406 50000 85 mm / 51 mm 45° 30 s / 50 s
60 Hz: 17400 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 1.2 kg 24 x 4 g 17850 rpm 30279 x g 406 50000 85 mm / 51 mm 45° 40 s / 50 s
50 Hz: 17500 rpm 60 Hz: 17500 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

34 | Small Benchtop Rotors

Thermo Scientific

MicroClick 24 x 2 | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.2 kg

Maximum Permissible Load

24 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

16000 rpm 24328 x g 505 50000

Radius max. / min.

85 mm / 51 mm

Angle

45°

Acceleration / Braking Time

23 s / 35 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

18 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.2 kg 24 x 4 g 16000 rpm 24328 x g 505 50000 85 mm / 51 mm 45° 23 s / 35 s 18 °C Yes
121 °C

100 V 1.2 kg 24 x 4 g 16000 rpm 24328 x g 505 50000 85 mm / 51 mm 45° 23 s / 35 s 18 °C Yes
121 °C

Accessories

Article No.

Description

75005725

Replacement ClickSeal Biocontainment Lid (each)

75003405

Replacement O-ring for Lid (each)

Adapters for MicroClick 24 x 2 Rotor (sets of 30)

75005752 75005753 75005754

0.2 mL PCR Tube 0.5 mL Microtube 0.25 mL Microtube

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

24 x 0.2 24 x 0.5 24 x 0.25

6.5 x 20 8 x 44 6 x 46

Thermo Scientific

Small Benchtop Rotors | 35

Chapter 1 | MicroClick 24 x 2
Biocontainment Certificate

36 | Small Benchtop Rotors

Thermo Scientific

MicroClick 30 x 2 | Chapter 1

MicroClick 30 x 2

Items Supplied

Article No. 75005719 50139640 20059119 75005726 76003500

Item MicroClick 30 x 2 Rotor Instruction Manual ClickSeal Lid O-Ring Set O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.7 kg

Maximum Permissible Load

30 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

14000 rpm 21694 x g 563 50000

Radius max. / min.

99 mm / 64 mm

Angle

45°

Acceleration / Braking Time

40 s / 50 s

Maximum Speed at 4 °C

50 Hz: 13000 rpm 60 Hz: 13500 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.7 kg 30 x 4 g 14000 rpm 21694 x g 563 50000 99 mm / 64 mm 45° 40 s / 50 s
60 Hz: 14000 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 2.7 kg 30 x 4 g 14000 rpm 21694 x g 563 50000 99 mm / 64 mm 45° 50 s / 50 s 50 Hz: 14000 rpm 60 Hz: 14000 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

Thermo Scientific

Small Benchtop Rotors | 37

Chapter 1 | MicroClick 30 x 2

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.5 kg

Maximum Permissible Load

30 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

14000 rpm 21694 x g 563 50000

Radius max. / min.

99 mm / 64 mm

Angle

45°

Acceleration / Braking Time

28 s / 35 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

24 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.5 kg 30 x 4 g 14000 rpm 21694 x g 563 50000 99 mm / 64 mm 45° 26 s / 35 s 24 °C Yes
121 °C

100 V 1.5 kg 30 x 4 g 14000 rpm 21694 x g 563 50000 99 mm / 64 mm 45° 26 s / 35 s 24 °C Yes
121 °C

Accessories

Article No.

Description

75005725

Replacement ClickSeal Biocontainment Lid (each)

75003405

Replacement O-ring for Lid (each)

Adapters for MicroClick 30 x 2 Rotor (sets of 30)

75005752 75005753 75005754

0.2 mL PCR Tube 0.5 mL Microtube 0.25 mL Microtube

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

30 x 0.2 30 x 0.5 30 x 0.25

6.5 x 20 8 x 44 6 x 46

38 | Small Benchtop Rotors

Thermo Scientific

Biocontainment Certificate

MicroClick 30 x 2 | Chapter 1

Thermo Scientific

Small Benchtop Rotors | 39

Chapter 1 | Microliter 48 x 2

Microliter 48 x 2

Items Supplied

Article No. 75003602 50143707 76003500

Item Microliter 48 x 2 Rotor Instruction Manual O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.4 kg

Maximum Permissible Load

48 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

12900 rpm 18233 x g 771 50000

Radius max. / min.

98 mm / 59 mm

Angle

45°

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.4 kg 48 x 4 g 12900 rpm 18233 x g 771 50000 98 mm / 59 mm 45° Yes
121 °C

100 V 2.4 kg 48 x 4 g 12900 rpm 18233 x g 771 50000 98 mm / 59 mm 45° Yes
121 °C

40 | Small Benchtop Rotors

Thermo Scientific

Microliter 48 x 2 | Chapter 1

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

2.4 kg

Maximum Permissible Load

48 x 4 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

11800 rpm 15256 x g 922 50000

Radius max. / min.

98 mm / 59 mm

Angle

45°

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 2.4 kg 48 x 4 g 11800 rpm 15256 x g 922 50000 98 mm / 59 mm 45° Yes
121 °C

100 V 2.4 kg 48 x 4 g 11800 rpm 15256 x g 922 50000 98 mm / 59 mm 45° Yes
121 °C

Accessories

Article No.

Description

Adapters for Microliter 48 x 2 Rotor (sets of 24)

76003758 76003759 76003750

0.5 mL Microtube 0.25 mL Microtube 0.2 mL PCR Tube

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

48 x 0.5 48 x 0.25 48 x 0.2

8 x 44 6 x 46 6.5 x 20

Thermo Scientific

Small Benchtop Rotors | 41

Chapter 1 | Microliter 48 x 2
Biocontainment Certificate

42 | Small Benchtop Rotors

Thermo Scientific

8 x 8 PCR Strip | Chapter 1

8 x 8 PCR Strip

Items Supplied

Article No. 75005720 50139640 20059119 75005726 76003500

Item 8 x 8 PCR Strip Rotor Instruction Manual ClickSeal Lid O-Ring Set O-Ring Grease

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 1 1 1

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.4 kg

Maximum Permissible Load

64 x 0.5 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

15000 rpm 17860 x g 538 50000

Radius max. / min.

71 mm / 44 mm

Angle

45°

Acceleration / Braking Time

30 s / 45 s

Maximum Speed at 4 °C

50 Hz: 15000 rpm 60 Hz: 15000 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.4 kg 64 x 0.5 g 15000 rpm 17860 x g 538 50000 71 mm / 44 mm 45° 25 s / 45 s
60 Hz: 15000 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 1.4 kg 64 x 0.5 g 15000 rpm 17860 x g 538 50000 71 mm / 44 mm 45° 30 s / 45 s 50 Hz: 15000 rpm 60 Hz: 15000 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

Thermo Scientific

Small Benchtop Rotors | 43

Chapter 1 | 8 x 8 PCR Strip

Megafuge 8, Sorvall ST 8 and SL 8 Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

1.4 kg

Maximum Permissible Load

64 x 0.5 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

15000 rpm 17860 x g 538 50000

Radius max. / min.

71 mm / 44 mm

Angle

45°

Acceleration / Braking Time

20 s / 32 s

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

12 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 1.4 kg 64 x 0.5 g 15000 rpm 17860 x g 538 50000 71 mm / 44 mm 45° 20 s / 33 s 12 °C Yes
121 °C

100 V 1.4 kg 64 x 0.5 g 15000 rpm 17860 x g 538 50000 71 mm / 44 mm 45° 20 s / 33 s 12 °C Yes
121 °C

Accessories

Article No.
75005730 75005726

Description
Replacement ClickSeal Biocontainment Lid (each) Replacement O-ring for Lid (each)

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

44 | Small Benchtop Rotors

Thermo Scientific

Biocontainment Certificate

8 x 8 PCR Strip | Chapter 1

Thermo Scientific

Small Benchtop Rotors | 45

Chapter 1 | 8 x 50 mL Individually Sealed

8 x 50 mL Individually Sealed

Items Supplied

Article No. 75003694 50139640 50111786 50129119 20290564

Item 8 x 50 mL Individually Sealed Instruction Manual Metal Cannister Biocontainment Lid O-Ring

If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.

Quantity 1 1 8 8 8

Technical Data

Megafuge 8R, Sorvall ST 8R and SL 8R Centrifuges

Centrifuge Frequency

230 V

Weight (empty)

3.3 kg

Maximum Permissible Load

18 x 9 g

Maximum Speed nmax Maximum RCF-Value at nmax K-Factor at nmax Max. Cycle Number

5600 rpm 5014 x g 5879 50000

Radius max. / min.

143 mm / 63 mm

Angle

45°

Acceleration / Braking Time

35 s / 40 s

Maximum Speed at 4 °C

50 Hz: 5600 rpm 60 Hz: 5600 rpm

Sample Temperature at nmax
(Ambient temperature of 23 °C, run time 60 minutes)

50 Hz: < 4 °C 60 Hz: < 4 °C

Aerosol-tight

Yes

Permissible Temperature Range Autoclavable 121 °C

120 V 3.3 kg 18 x 9 g 5600 rpm 5014 x g 5879 50000 143 mm / 63 mm 45° 30 s / 40 s
60 Hz: 5600 rpm
60 Hz: < 4 °C
Yes 121 °C

100 V 3.3 kg 18 x 9 g 5600 rpm 5014 x g 5879 50000 143 mm / 63 mm 45° 35 s / 40 s
50 Hz: 5600 rpm 60 Hz: 5600 rpm
50 Hz: < 4 °C 60 Hz: < 4 °C Yes
121 °C

46 | Small Benchtop Rotors

Thermo Scientific

Accessories

Article No.

Description

75003011

Replacement Biocontainment Lids (set of 2)

75003789

Replacement O-ring Kit

Adapters for 8 x 50 mL Individually Sealed Rotor (set of 2)

75005808 75005804

15 mL Conical Tube 10 mL Blood Collection Tube (16 x 100 mm)

75005805 75005806

7 mL Blood Collection Tube (13 x 100 mm) 3.5 mL Blood Collection Tube

8 x 50 mL Individually Sealed | Chapter 1

Rotor Capacity
(places x volume, mL)

Max Tube Dimensions
(Ø x L, mm)

8 x 15 8 x 10 8 x 7 16 x 3.5

17 x 121 16 x 110 13 x 110 11 x 75

Thermo Scientific

Small Benchtop Rotors | 47

Chapter 1 | 8 x 50 mL Individually Sealed
Biocontainment Certificate

48 | Small Benchtop Rotors

Thermo Scientific

Thermo Scientific Auto-Lock Rotor Exchange
Rotor Installation
CAUTION
Unapproved or incorrectly combined accessories can cause serious damage to the centrifuge.

2

This rotor is equipped with a Thermo ScientificTM Auto-LockTM rotor exchange. This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have to be bolted onto the centrifuge spindle. Proceed as follows: 1. Open the door of the centrifuge and if necessary, remove any dust, foreign objects or residue from the chamber.
Auto-Lock and O-ring must be clean and undamaged.

1 2

1

Auto-Lock

2

O-Ring

2. Place the rotor over the centrifuge spindle and lower it vertically slowly down the centrifuge spindle. If necessary, a light push may be required to ensure connection. The rotor clicks automatically into place.

CAUTION
Do not force the rotor onto the centrifuge spindle. If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle with little pressure.

Thermo Scientific

Small Benchtop Rotors | 49

Chapter 2 | Removing the Rotor
3. Check if the rotor is properly installed by lifting it slightly on the handle or from beneath the rotor. If the rotor can be pulled up, then it must be reconnected to the centrifuge spindle.
WARNING
If the rotor cannot be properly locked in place after several attempts, then the Auto-Lock is defective, and you are not permitted to operate the rotor. Check for any damage to the rotor: Damaged rotors must not be used. Keep the centrifuge spindle area of the rotor clear of objects. If in doubt, please call Thermo Fisher Scientific Service.
CAUTION
Check that the rotor is properly locked on the centrifuge spindle before each use by pulling it at its handle or from beneath the rotor. The rotor has to be locked tight.
4. If available close the rotor with the rotor lid.
CAUTION
Be sure to check all seals before starting any aerosol-tight applications.
5. Close the centrifuge door.
Removing the Rotor
To remove the rotor, proceed as follows: 1. Open the centrifuge door. 2. Grab the rotor handle and fully depress the Auto-Lock button. At the same time, pull the rotor directly upwards and
remove it from the centrifuge spindle. Make sure not to tilt the rotor while doing this.

50 | Small Benchtop Rotors

Thermo Scientific

Rotor Loading

3

Before a Run
1. Read and observe the precautions and the safety instructions in this manual and in the instruction manual of the centrifuge.
2. Check the rotor and all accessory parts for damages such as cracks, scratches or traces of corrosion. 3. Check the rotor chamber, the centrifuge spindle and the Auto-Lock. 4. Check the rotor suitability. ("Chemical Compatibility Chart" on page 65)

Rotor Temperature Range

CAUTION
Operate the rotor in a temperature range between -9 °C and +40 °C only. A pre-tempering in a freezer below -9 °C is not allowed.

NOTICE
The rotors can warm up in ventilated centrifuges at high ambient temperatures. Temperature above 45 °C blood samples can be damaged. Let the rotor cool down between two runs.

Rotor Loading
CAUTION
Failure to load all buckets may lead to damage or accident. All buckets and sleeves must be in place before starting the rotor.
Always balance opposite loads. Balance opposite loads in number of tubes and by position to ensure safe and smooth operation.

Thermo Scientific

Small Benchtop Rotors | 51

Chapter 3 | Maximum Loading
Correct Balances
Fixed Angle Rotors
Swing Out Rotors
Incorrect Balances
Fixed Angle Rotors
Swing Out Rotors

Maximum Loading
The rotor can run at high speeds. Each rotor is specifically designed to run at its maximum speed with a defined load. For further details please refer to the rotor manual supplied with the rotor. The safety system of the centrifuge requires that you do not overload the rotor. The rotors are designed to work with solution with a density of up to 1.2 g/ml. Above this density or if total load is above the maximum weight the following steps should be taken: · Reduce the fill level. · Reduce the speed.

52 | Small Benchtop Rotors

Thermo Scientific

Use the table or the formula:

Maximum Loading | Chapter 3

nadm = admissible speed nmax = maximum speed

Explanation of RCF-Value

The relative centrifugal force (RCF) is given as a multiple of the force of gravity g. It is a unit less numerical value which is used to compare the separation or sedimentation capacity of various centrifuges, since it is independent of the type of device. Only the centrifuging radius and the speed are used for calculation:

RCF = 11.18 x

x r

r = centrifuging radius in cm

n = rotational speed in rpm

The maximum RCF value is related to the maximum radius of the tube opening.

Remember that this value is reduced depending on the tubes and adapters used.

This can be accounted for in the calculation above if required.

Once the rotor has been properly installed, the main switch turned on and the centrifuge door closed, you may start centrifuging.

Use of Tubes and Consumables

Care should be taken to ensure that the tubes and bottles used in the centrifuge are: · Rated to or above the selected rcf to be spun at · They are being used at or above their minimum fill volume · They are not being used above their design life (age or number of runs) · They are inspected for damage Please refer to manufacturers data sheets for further information.

Rotor Life Time

WARNING
Replace the rotor when the specified number of cycles is reached. Due to the mechanical load a rotor can break and damage the centrifuge. Replace the rotor and buckets when the specified number of cycles is reached.

The lifetime of rotors and buckets is dependent on the amount of mechanical load. Do not exceed them number of cycles recommended for rotors and buckets. The maximum number of cycles is given in the rotor table in the rotor specification section. The maximum number of cycles for buckets is marked on the buckets themselves.

Service Life Examples
Usage profile Frequent use 20 runs / day 220 days / year

Maximum service life at 50,000 cycles 7 years

Thermo Scientific

Small Benchtop Rotors | 53

4
Aerosol-tight Applications
Basic Principles
CAUTION
Biological seals are part of bio-containment systems as specified in international and national bio-safety guidelines and cannot relied on as the only means of safeguarding workers and the environment when handling pathogenic micro-organisms. Mind the "Laboratory Biosafety Manual" of the World Health Organization (WHO) and the regulations in your country.
CAUTION
When centrifuging hazardous samples, do not open aerosol-tight rotors or buckets unless placed in a safety cabinet. Always bear in mind the maximum permitted fill levels.
WARNING
Be sure to check all seals before starting any aerosol-tight applications.
Check that the sample containers are well suited for the desired centrifugation process.
Placing O-Ring
The O-ring fulfills its purpose best, when it is neither stressed nor bulked. Meaning the O-ring should be equally placed in the groove of the lid. Place the O-ring as follows: 1. Place the O-ring above the groove. 2. Push the O-ring on two opposite places into the groove. Make sure the rest of the O-ring is equally distributed. 3. Push the centers of the loose parts into the groove. 4. Push the remaining O-ring into place.
NOTE
If the O-ring seems to be too long or too short, take it off the lid and repeat the process.

Thermo Scientific

Small Benchtop Rotors | 55

Chapter 4 | Fill Level
Fill Level
The tubes are only to be filled to a level which ensures that the sample is unable to reach the top of the tube during centrifugation. Therefore, fill the tube only 2/3 of the rated level.
Checking the Aerosol-Tightness
The aerosol tightness testing of the rotors and buckets took place according to the microbiological test process in accordance with the EN 61010-2-020 Appendix AA. Whether or not a rotor is aerosol-tight depends primarily on proper handling. Check as needed to make sure your rotor is aerosol-tight. The careful inspection of the seals and seal surfaces for signs of wear and damage such as cracks, scratches and embrittlement are extremely important. Aerosol-tight applications are not possible if the bucket tops are open. Aerosol-tightness requires the correct operation when filling the sample vessels and closing the rotor lid.
Quick Test
As a quick test, it is possible to test the aerosol-tight beaker and fixed-angle rotors using the following process: 1. Lubricate all seals lightly.
Always use the special grease (part no. 76003500) when lubricating the seals. 2. Fill the bucket or the rotor with approx. 10 ml of carbonated mineral water. 3. Close the bucket or the rotor as explained in the handling instructions. 4. Shake the bucket or the rotor vigorously using your hands.
This releases the carbonic acid gas which is bound in the water, resulting in excess pressure. Do not apply pressure to the lid when doing so! Leaks can be detected by escaping water or the sound of escaping gas. Replace the seals if you detect any leaks. Then repeat the test. 5. Dry the bucket, rotor, rotor cover and the cover seal.
CAUTION
Prior to each use, the seals in the rotor are to be inspected in order to assure that they are correctly seated and are not worn or damaged. Damaged seals are to be replaced immediately. Replacement seals are supplied with the rotors and can also be re-ordered as a spare part set. When loading the rotor, ensure that the rotor lid closes securely. Damaged or clouded rotor covers are to be replaced immediately.

56 | Small Benchtop Roto

Thermo Scientific

Checking the Aerosol-Tightness | Chapter 4
Aerosol-tight Closure with ClickSeal
1. If necessary, grease the lid joint before closing the lid. Use grease (76003500) for this. 2. Raise the latch.
The cap can now be easily placed on the bucket.
3. Lower the latch to close the bucket aerosol-tight; be sure the latch clicks into place.
CAUTION
If the latch is not flipped down, the caps could be damaged during centrifuging. If the latch has not clicked into place, the bucket is not aerosol-tight. Never raise the bucket at its latch.

Thermo Scientific

Small Benchtop Rotors | 57

Chapter 4 | Removing of Aerosol-tight Rotors
Removing of Aerosol-tight Rotors
For your protection when using an aerosol tight lid, the rotor should only be removed with the lid closed.
NOTICE
Rotors supplied with a lid for aerosol-tight applications come with a mandrel, which belongs to the Auto-Lock. Be sure not to place the lid onto this mandrel to prevent it from being damaged.
CAUTION
Skin can be pierced by sharp mandrel tip. Do not touch the mandrel.

58 | Small Benchtop Roto

Thermo Scientific

Maintenance and Care

5

Cleaning Intervals

For the sake of personal, environmental, and material protection, it is your duty to clean and if necessary, disinfect the centrifuge on a regular basis.

Maintenance Rotor Chamber (Bowl) Rotor Accessories

Recommended Interval Daily or when polluted Daily or when polluted Daily or when polluted

CAUTION
Refrain from using any other cleaning or decontamination procedure than those recommended here, if you are not entirely sure that the intended procedure is safe for the equipment. Use only approved cleansers. If in doubt, contact Thermo Fisher Scientific.

Basics
CAUTION
Not rated procedures or agents could deteriorate the materials of the centrifuge and lead to malfunction. Refrain from using any other cleaning or decontamination procedure than those recommended here, if you are not entirely sure that the intended procedure is safe for the equipment. Use only approved cleansers. If in doubt, contact Thermo Fisher Scientific.
 Use warm water with a neutral detergent that is suitable for use with the materials of the centrifuge. If in doubt contact the manufacturer of the cleaning agents.
 Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or scrubbing powder.  Remove rotor and clean bowl with a small amount of cleaning agent, applied to a clean cloth.  Use a soft brush without metal bristles to remove stubborn residue.  Afterwards rinse with a small amount of distilled water and remove any excess with absorbent towels.  Use only disinfectants with a pH of 6-8.
Rotor and Accessories Inspection
After thoroughly cleaning rotors, they should be inspected for damage, wear and corrosion.

Thermo Scientific

Small Benchtop Rotors | 59

Chapter 5 | Cleaning
Metal Parts
Ensure that the black protective coating is complete. It can be removed through wear and chemical attack and can lead to unseen corrosions. Any signs of corrosions, such as rust or white / metallic pitting, the rotor or accessories should be immediately removed from service. Particular attention should be taken with the bottom of buckets on swing out rotors and tube cavities on fixed angle rotors.
Plastic Parts
Check for signs plastic crazing, fading, bruising or cracking.
CAUTION
Do not run any rotor or accessories with sign of damage. Ensure that the rotor, buckets and accessories are within the service life and number of cycles. It is recommend that you have rotors and accessories inspected yearly as part of your routine service to ensure safety.

Cleaning
CAUTION
Before using any cleaning methods except those recommended by the manufacturer. Users should check with the manufacturer of the cleaning agents that the proposed method will not damage the equipment.

Clean as follows: 1. Clean rotor, buckets and accessories outside of the centrifuge bowl. 2. Separate all rotors, buckets, lids, adapters and tubes to allow thorough cleaning. 3. Rinse rotor and all accessories with warm water and a neutral detergent that is suitable for use with the materials of
the centrifuge. If in doubt contact the manufacturer of the cleaning agents. Ensure grease on rotor trunnions (pivot point for swinging buckets) is cleaned away. 4. Use a soft brush without metal bristles to remove stubborn residue. 5. Rinse rotor and all accessories with distilled water. 6. Place the rotors on a plastic grate with their cavities pointing down, to allow to fully drain and dry. 7. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50 °C. If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures could damage the material and shorten the lifetime of the parts. Once clean and dry, inspect the rotor and accessories. After cleaning, treat the entire surface of aluminum parts including the cavities with corrosion protection oil (70009824). Treat the bolt of the swing out rotor with bolt grease (75003786).
CAUTION
Drive and lid lock can be damaged by entering liquids. Do not allow liquids, especially organic solvents, to get on the drive shaft, the drive bearings or the centrifuge door locks. Organic solvents break down the grease in the motor bearing. The drive shaft could lock up.

60 | Small Benchtop Rotors

Thermo Scientific

Disinfection | Chapter 5
Disinfection
WARNING
A hazard of infection is possible when touching the contaminated rotor and centrifuge parts. Infectious material can get into the centrifuge when a tube breaks or as a result of spills. In case of contamination, make sure that others are not put at risk. Disinfect the affected parts immediately.
CAUTION
Equipment can be damaged by inappropriate disinfection methods or agents. Before using any cleaning or disinfection methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Observe the safety precautions and handling instructions for the cleaning agents used.
The rotor chamber and the rotor should be treated preferably with a neutral disinfectant. Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other disinfectants. For details, see "Basics" on page 59. Disinfect as follows: 1. Disinfect rotor, buckets and accessories outside of the centrifuge bowl. 2. Separate all rotors, buckets, lids, adapters and tubes to allow thorough disinfection. 3. Treat the rotor and accessories according to the instructions for the disinfectant. Adhere strictly to the given
application times. Be sure the disinfectant can drain off the rotor. 4. Rinse the rotor and accessories thoroughly with water and then rub down. 5. Place the rotors on a plastic grate with their cavities pointing down, to allow to fully drain and dry. 6. Dispose the disinfectant according to the applicable guidelines. Clean the rotor after disinfecting as described in "Cleaning" on page 60.

Thermo Scientific

Small Benchtop Rotors | 61

Chapter 5 | Decontamination
Decontamination
WARNING
Radiation is possible when touching the contaminated rotor and centrifuge parts. Radioactive material can get into the centrifuge when a tube breaks or as a result of spills. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately.
CAUTION
Equipment can be damaged by inappropriate decontamination methods or agents. Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Observe the safety precautions and handling instructions for the cleaning agents used.
For general radioactive decontamination use a solution of equal parts of 70 % ethanol, 10 % SDS (Sodium Dodecyl Sulfate) and water. Decontaminate as follows: 1. Decontaminate rotor, buckets and accessories outside of the centrifuge bowl. 2. Separate all rotors, buckets, lids, adapters and tubes to allow thorough decontamination. 3. Treat the rotor and accessories according to the instructions for the decontamination solution. Adhere strictly to the
given application times. Be sure the decontamination solution can drain off the rotor. 4. Rinse the rotor first with ethanol and then with deionized water. Adhere strictly to the given application times. 5. Be sure the decontamination solution can drain off the rotor. 6. Rinse the rotor and accessories thoroughly with water. 7. Place the rotors on a plastic grate with their cavities pointing down, to allow to fully drain and dry. 8. Dispose of the decontamination solution according to the applicable guidelines. Clean the rotor after disinfecting as described in "Cleaning" on page 60.

62 | Small Benchtop Rotors

Thermo Scientific

Autoclaving | Chapter 5
Autoclaving
CAUTION
Never exceed the permitted temperature and duration when autoclaving.
NOTE
No chemical additives are permitted in the steam.
1. Before autoclaving clean rotor and accessories as described above. 2. Place the rotor on a flat surface.  Rotors and adapter can be autoclaved at 121 °C.  The maximum permissible autoclave cycle is 20 minutes at 121 °C. Clean the rotor before autoclaving and rinse it with distilled water. Remove all accessories (tubes, adapters) from the rotor. Place the rotor on a flat surface.
Service of Thermo Fisher Scientific
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an authorized service technician. The service technicians check the following: · electrical equipment safety; · suitability of the set-up site; · door lock, imbalance and safety system; · rotor; · fixation of rotor and drive shaft. Before service, centrifuge and rotors should be thoroughly cleaned and decontaminated to ensure full and safe inspection can be completed. Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge. This is only valid if the centrifuge has only been maintained by an authorized Thermo Scientific service technician.
Storage
Any moisture left on a metal rotor can initiate corrosion, so after cleaning ensure proper storage: · Remove all adapters from rotor cavities when not in use. · Dry and store upside-down on a plastic matting to allow for airflow or a ventilated shelf to avoid gathering
condensation in the cavity or bucket bottom.

Thermo Scientific

Small Benchtop Rotors | 63

Chapter 5 | Shipping and Disposal
Shipping and Disposal
WARNING
When removing the centrifuge and accessories from use for disposal you have to clean and if necessary disinfect or decontaminate the entire system. In doubt contact the Thermo Fisher Scientific customer service.
For the disposal of the centrifuge mind the regulations in your country. Contact the Thermo Fisher Scientific Customer Service for the disposal of the centrifuge. For contact information check the back page of this manual or visit www.thermofisher.com/centrifuge For the countries of the European Union the disposal is regulated by the European Union's Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. ("WEEE Compliance" on page 2)

64 | Small Benchtop Rotors

Thermo Scientific

Chemical Compatibility Chart

A

MATERIAL ALUMINIUM ANODIC COATING for ALUMINIUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE ROTOR PAINT COMPOSITE Carbon Fiber/Epoxy DELRINTM ETHYLENE PROPYLENE GLASS NEOPRENE NORYLTM NYLON PET¹, POLYCLEARTM,CLEARCRIMPTM POLYALLOMER POLYCARBONATE POLYESTER, GLASS THERMOSET POLYTHERMIDE POLYETHYLENE POLYPROPYLENE POLYSULFONE POLYVINYL CHLORIDE RULON ATM, TEFLONTM SILICONE RUBBER STAINLESS STEEL TITANIUM TYGONTM VITONTM

Thermo Scientific

CHEMICAL 2-mercaptoethanol Acetaldehyde Acetone
Acetonitrile AlconoxTM Allyl Alcohol Aluminum Chloride Formic Acid (100 %) Ammonium Acetate Ammonium Carbonate Ammonium Hydroxide (10 %) Ammonium Hydroxide (28 %) Ammonium Hydroxide (conc.) Ammonium Phosphate Ammonium Sulfate Amyl Alcohol Aniline Sodium Hydroxide (<1 %) Sodium Hydroxide (10 %) Barium Salts Benzene Benzyl Alcohol Boric Acid Cesium Acetate Cesium Bromide Cesium Chloride

S SU / SMS / SUS SUS S / S S S SUS S S S S S S / U U / / / M / U / / / MU U UMM / M S U / S / U M S U U S UM S S U U S U S U U U S S U U S MM S U U S S U / SMS / S S U S UMUU / SMUU S S S S UU UUS / S S S / S S S S S SMS S S S S S S S S S SU / / / U / / S / / / / S / S SMS S S / MS / / S / / UU S S S S U S S S SMS S S S / S S S S SMUU S S / SMU / / U / / / / U / SMUU S S / U S / U S / U
SSU / SSS / SSSSSSSU / SSSSSSSSSS
MSUS S S S S S S S S S SUU / S S S S S SMS S S
UU S U S SMS S S S S / S UMS S S S S S S S SMS
UU S U S UMS S S S S U S UMS S S S S S S S SMS
UUUU S UMS / S / S U S UU S S S / MS S S S / U
U / S / S S S S S S S S / S SM / S S S S S SMS S S UMS / S S U S S S S S S S S S / S S S S S S U S S U S / MU / / S S / M / S / MS S S SM / / / U / S / M S S UU S U SMS UUUUUUU / SMUU S S S S U S U / M S S S / / S M S S / S MM S S S S S S M S S / U
U / MU / / U / MM S S U S U U S S S S S S M S S / U MU S / S S S S S S S S S S SM / S S S S S SMS S S S S UU S UMU S UU S UUUMUMUUU S UU S U S S / U U / / MM / M / S U U U U U U U / M S M / S / S US SMS SUS S S S S S S S SUS S S S S S S S S S M / S / S S S / S S S S / S S / / S S S S S SMS S S MS S / S S S / S S S S S S S / / S S S S S SMS S S MS SUS S S / S S S S S S S / / S S S S S SMS S S

Small Benchtop Rotors | 65

Annex A | Chemical Compatibility Chart

MATERIAL ALUMINIUM ANODIC COATING for ALUMINIUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE ROTOR PAINT COMPOSITE Carbon Fiber/Epoxy DELRINTM ETHYLENE PROPYLENE GLASS NEOPRENE NORYLTM NYLON PET¹, POLYCLEARTM,CLEARCRIMPTM POLYALLOMER POLYCARBONATE POLYESTER, GLASS THERMOSET POLYTHERMIDE POLYETHYLENE POLYPROPYLENE POLYSULFONE POLYVINYL CHLORIDE RULON ATM, TEFLONTM SILICONE RUBBER STAINLESS STEEL TITANIUM TYGONTM VITONTM

CHEMICAL Cesium Formate Cesium Iodide Cesium Sulfate Chloroform Chromic Acid (10 %) Chromic Acid (50 %) Cresol Mixture Cyclohexane Deoxycholate Distilled Water Dextran Diethyl Ether Diethyl Ketone Diethylpyrocarbonate Dimethylsulfoxide Dioxane Ferric Chloride Acetic Acid (Glacial) Acetic Acid (5 %) Acetic Acid (60 %) Ethyl Acetate Ethyl Alcohol (50 %) Ethyl Alcohol (95 %) Ethylene Dichloride Ethylene Glycol Ethylene Oxide Vapor Ficoll-HypaqueTM Hydrofluoric Acid (10 %) Hydrofluoric Acid (50 %) Hydrochloric Acid (conc.) Formaldehyde (40 %) Glutaraldehyde Glycerol Guanidine Hydrochloride Haemo-SolTM Hexane

MS S / S S S / S S S S S S S / / S S S S S SMS S S MS S / S S S / S S S S S S S / / S S S S S SMS S S MS S / S S S / S S S S S S S / / S S S S S SMS S S U U U U S S MU S U UMUMU U UMMU U S U U UM S U / UU S UU / S S S U S SMUMS S UMSMU S S S
U / UU / UU / / / S UU SMUMS S UMS / UM / S SSU / / / S / SUUUUUU / / UU / USSSSUS S S S / S S S U S U S S U U UM S MUMM S UMMU S SSS / SSS / SSSSSSS / / SSSSSSSSSS SSSSSSSSSSSSSSSSSSSSSSSSSSS MS S S S S S / S S S S S S S S S S S S S S SMS S S S S UU S S S U S UU S UUUUUUUUU S S S SMU S / U U / / M / S U / S / MU U UMM / U S / / S U U S SU / S S S / S SUSUSU / / S S SMS S S S S S SSUUSSS / SUSSUSUU / SSUUSSSSUU M S U U S S MM S U U S UMU U / MMMU S S S S U U UU S / / / MS / M / S / S / / / S S / / / MU S / S S S UU S S UMS U S UUUUUMS UMU S UU S / U S SMS S SMS S S S SMS S S S S S SMS SMS SM S S UU S S U / SMS UUMU SMSMSMSMU SMU MMU U S S MM S S U S UMU U / S S U U S MM S U U S S S S S SMS S S S S U S U S S S S S S S SMSMU
S S S U S SMS S S S S U S U / S S SMS S S U SMU S / UU / / SM / UUSUUUUUUU / USU / S / S S S S S S S S S S S S S / SUS S S S S S S SMSMS S / U / / U / / SU / S / SM / / S S SUSUS S SU MS S / S S S / S S S S / S S / S S S S S S SMS S S UUUM / / U / / UU S / SMU S S S SMS UUU / /
U U U U / / U / / U U U U S U U U S S MM S U U U / M
UUUU / UUM / UMUUMUUU / S / U S UUU / /
MMM S S S S M S S S S M S S S U S S M S S S M S MU SSSS / / S / SSSSSSS / / SSS / / SSS / / MS S / S S S S S S S S S S S S / S S S S S S S S S S UUS / SSS / SSSSSSS / / SSSSSSUSSS SSS / / / S / SSSSSSS / / SSSSSSSSSS S S S / S S S / S S U S UMU S S U S SMS U S S U S

66 | Small Benchtop Rotors

Thermo Scientific

Chemical Compatibility Chart | Annex A

MATERIAL ALUMINIUM ANODIC COATING for ALUMINIUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE ROTOR PAINT COMPOSITE Carbon Fiber/Epoxy DELRINTM ETHYLENE PROPYLENE GLASS NEOPRENE NORYLTM NYLON PET¹, POLYCLEARTM,CLEARCRIMPTM POLYALLOMER POLYCARBONATE POLYESTER, GLASS THERMOSET POLYTHERMIDE POLYETHYLENE POLYPROPYLENE POLYSULFONE POLYVINYL CHLORIDE RULON ATM, TEFLONTM SILICONE RUBBER STAINLESS STEEL TITANIUM TYGONTM VITONTM

CHEMICAL Isobutyl Alcohol Isopropyl Alcohol Iodoacetic Acid Potassium Bromide Potassium Carbonate Potassium Chloride Potassium Hydroxide (5 %) Potassium Hydroxide (conc.) Potassium Permanganate Calcium Chloride Calcium Hypochlorite Kerosene Sodium Chloride (10 %) Sodium Chloride (sat'd) Carbon Tetrachloride Aqua Regia Solution 555 (20 %) Magnesium Chloride Mercaptoacetic Acid Methyl Alcohol Methylene Chloride Methyl Ethyl Ketone MetrizamideTM Lactic Acid (100 %) Lactic Acid (20 %) N/Butyl Alcohol N/Butyl Phthalate N, N-Dimethylformamide Sodium Borate Sodium Bromide Sodium Carbonate (2 %)

/ / MU / / S S / U / S U S SMS S S / S S S / S / S MMMU S S S S S U S S U S UM S S S S S S S MMM S S S M / S S S / S M S S M S S / M S S S S S M S S MM US S / S S S / S S S S S S S S S S S / S S SMS S S MU S S S S S / S S S S S S U S S S S S S S S S S S S USS / SSSSSSSSSSS / SSSSSSSUSSS UU S S S SM / S S S S / S U S S S S S S SMUMS U
UUMU / / M / MS S / UMUUU SM / MU / UU / U
S S S / S S S / S S SUS S SM / SMSUS SMSUS MU S S S S S S S S S S S SMS / S S S S S SMS S S M / U / S MM S / M / S / S M S / S S S M S MU S / S S S S / S S S U S MU S UMM S / MMM S S U S S U S S / S S S S S S / / / S S S S S / S S S S / S SM / S
U / SUS S S / / / / S S S S S / S S / S / S SM / S
U UM S S UMU S U U S UMU S S MM S MMMMU S S U / UU / / U / / / / / UUUUUUU / / / / / S /M SSS / / / S / SSSSSSS / / SSS / SSSSSS
MS S / S S S S S S S S S S S S S S S S S S SMS S S
USU / SMS / SMSUUUU / SUUSMSUS S S S S S S U S SMS S S S S U S UMS S S S S S SMSMU UUUUMS S U S UU S UUUUUMUUU S SMU S U S SUUS SMS SUUSUSUUUS SUUS S S SUU MS S / S S S / S S S S / S S / / S S S S S SMS S S / / S / / / / / / MSU / S S SMS S / MSMS S / S / / S S / / / / / MSM / S S S S S S SMSMS S / S S / S U / / S / / S M / U S M S S S S MM S M / S / S S S U / S S S / S U U S U U UM / U U S U S MM S U S
S S SUSMS / S SUSUSUU / S SUUSMS S SU
MS S S S S S S S S SUS S S S / S S S S S SMS S S US S / S S S / S S S S S S S S / S S S S S SMS S S MU S S S S S S S S S S S S U S S S S S S S S S S S S

Thermo Scientific

Small Benchtop Rotors | 67

Annex A | Chemical Compatibility Chart

MATERIAL ALUMINIUM ANODIC COATING for ALUMINIUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE ROTOR PAINT COMPOSITE Carbon Fiber/Epoxy DELRINTM ETHYLENE PROPYLENE GLASS NEOPRENE NORYLTM NYLON PET¹, POLYCLEARTM,CLEARCRIMPTM POLYALLOMER POLYCARBONATE POLYESTER, GLASS THERMOSET POLYTHERMIDE POLYETHYLENE POLYPROPYLENE POLYSULFONE POLYVINYL CHLORIDE RULON ATM, TEFLONTM SILICONE RUBBER STAINLESS STEEL TITANIUM TYGONTM VITONTM

CHEMICAL Sodium Dodecyl Sulfate Sodium Hypochlorite (5 %) Sodium Iodide Sodium Nitrate Sodium Sulfate Sodium Sulfide Sodium Sulfite Nickel Salts Oils (Petroleum) Oils (Other) Oleic Acid Oxalic Acid Perchloric Acid (10 %) Perchloric Acid (70 %) Phenol (5 %) Phenol (50 %) Phosphoric Acid (10 %) Phosphoric Acid (conc.) Physiologic Media (Serum, Urine) Picric Acid Pyridine (50 %) Rubidium Bromide Rubidium Chloride Sucrose Sucrose, Alkaline Sulfosalicylic Acid Nitric Acid (10 %) Nitric Acid (50 %) Nitric Acid (95 %) Hydrochloric Acid (10 %) Hydrochloric Acid (50 %) Sulfuric Acid (10 %) Sulfuric Acid (50 %) Sulfuric Acid (conc.) Stearic Acid Tetrahydrofuran
68 | Small Benchtop Rotors

SSS / SSS / SSSSSSS / SSSSSSSSSSS
UUMS SMU S SMS S SMS S S SMS S SMU SMS MS S / S S S / S S S S S S S / / S S S S S SMS S S SSS / SSSSSSSSSSSS / SSSSSUSSSS US S / S S S S S S S S S S S S S S S S S S SMS S S S / S S / / / S / / / S S SUU / / S / / / S SM / S S S S / S S S SMS S S S S SM / S S S S S S S S S S US S S S S / S S S / / S S S S / S S S S S SMS S S S S S / / / S U S S S S UUMSMUU S S S U S S S S S / S / / / SMS S S SUS S S SUS S S S / S SMS S / U S S S U U S U S S M S S S S S S S S S MU S MM UUMS S S U S S S S S U S U S S S S S S S S UMS S U / U / S U U / S MM / / MUM S MM / M S U / S / S
U U U / / U U / S UMU UMU U UMMUM S U U S U S U S U / S MM / S UMU U S UM S M S U U S UMMM S U S U / S UM / S UMUUUUU S UMUU S UUUMS UUMS S S U S S S S U / S S S S S S S S S UMU S S
U UMM / / U S / M S U UMM S S S M S M S UMU / S
MS S S / / S / S S S S S S S S S S S S S S S S S S S S S U / SMS S SMS U S S S U S S S S U S UMSMS U S UU S UU / U S S UUMUU / U SMU S S UUUU MS S / S S S / S S S S S S S / / S S S S S SMS S S MS S / S S S / S S S S S S S / / S S S S S SMS S S MS S / S S S S S S S S S S S S S S S S S S S S S S S MS S / S S S / S S S S S SUS S S S S S S SMS S S UUSSSSS / SSSUSSS / SSS / SSSUSSS USUS SUU / SUSU / S S S S S S S S SMS S S S U S UM S U U / S U S U UMMUMMM S S S U S S M S U / UU / UU / / UUUUMUUUUMUU S U S S / S UUMS S S U / S S S UU S U S S S S S S S S UMS S
U U U U S U U / S M S U UMU U S S S S M S MU UMM
MUU S S UU / S SMU S S S S S S S S S S UUU S S
MUUU S UU / S SMUU S UUMS S S S S UUUMS
MUUU / UUM / / MUU S UUUMS UMS UUU / S S / S / / / S M S S S S / S S S S S S S S S MM S S S S S UU S UUMS UU S UUU / MUUUU S U S S UU
Thermo Scientific

Chemical Compatibility Chart | Annex A

MATERIAL ALUMINIUM ANODIC COATING for ALUMINIUM BUNA N CELLULOSE ACETATE BUTYRATE POLYURETHANE ROTOR PAINT COMPOSITE Carbon Fiber/Epoxy DELRINTM ETHYLENE PROPYLENE GLASS NEOPRENE NORYLTM NYLON PET¹, POLYCLEARTM,CLEARCRIMPTM POLYALLOMER POLYCARBONATE POLYESTER, GLASS THERMOSET POLYTHERMIDE POLYETHYLENE POLYPROPYLENE POLYSULFONE POLYVINYL CHLORIDE RULON ATM, TEFLONTM SILICONE RUBBER STAINLESS STEEL TITANIUM TYGONTM VITONTM

CHEMICAL Toluene Trichloroacetic Acid Trichloroethane Trichloroethylene Trisodium Phosphate Tris Buffer (neutral pH) Triton X/100TM Urea Hydrogen Peroxide (10 %) Hydrogen Peroxide (3 %) Xylene Zinc Chloride Zinc Sulfate Citric Acid (10 %)

S S UU S SMU S UU S UUU S UMUUU S U S UUM UUU / S S UMS U S UU SM / MS S UU S UUUMU S / U / / / MU / U / S UUUUUUUUU S U / S / S / / UU / / / U / U / SUUUUUUUUUSU / U / S / / / S / /M/ / / / / / S / / SSS / / S / / S / S
USSSSSS / SSSSSSSSSSSSSSSSSSS SSS / SSS / SSSSSSSSSSSSSSSSSSS S / US S S S / / / / S S SMS S S S / S S SMS / S UUMS S UU / S S S U S S SMU S S S S S SMS U S
SMS S S / S / S S S S S S S SMS S S S S S S S S S S S U S S SMU S UUUUUUMUMUUU S UMS U S UUSSSSUSSSSSSSSSSSSSSSSUSSS USS / SSSSSSSSSSSSSSSSSSSSSSS MS SMS SMS S S S S S S S SMS S S S S S S S S S

¹Polyethlyeneterephtalate

Key
S ­ Satisfactory. M ­ Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved, etc.;
suggest testing under actual conditions of use. U ­ Unsatisfactory, not recommended. / ­ Performance unknown; suggest testing, using sample to avoid loss of valuable material.
NOTICE
Chemical resistance data is included only as a guide to product use. Because no organized chemical compatibility data exists for materials under the stress of centrifugation, when in doubt we recommend pretesting sample lots.

Thermo Scientific

Small Benchtop Rotors | 69

Index
Symbols
8x8 PCR Strip 40 8 x 50 mL Individually Sealed 43
A
Aerosol-tight Applications 53 Aerosol-tight Closure with ClickSeal 55 Autoclaving 60, 61
B
Before a Run 49
C
Checking the Aerosol-Tightness 54 Chemical Compatibility Chart 63 Cleaning 57, 58 Cleaning Intervals 57 CLINIConic 30
D
Decontamination 59, 60 Disinfection 59
E
Explanation of RCF-Value 51
H
HIGHConic III 27
I
Intended Use 7 Items Supplied 7, 10
M
M10 21 Maintenance and Care 57 Maximum Loading 50 MicroClick 18x5 32 MicroClick 24x2 34 MicroClick 30x2 37 MT-12 25
P
Precautions 7
Thermo Scientific

Preface 7
R
Removing the Rotor 48 Rotor Installation 47 Rotor Life Time 51 Rotor Loading 49 Rotor Specifications 9 Rotor Temperature Range 49
S
Service of Thermo Fisher Scientific 61 Shipping and Disposal 62 Storage 61 Symbols 8
T
Thermo Scientific Auto-Lock Rotor Exchange 47
Thermo Scientific Auto-Lock Rotor Exchange 47 TX-100 19 TX-100S 16 TX-150 10
W
WEEE 2
Small Benchtop Rotors | 71

Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg, 37520 Osterode am Harz Germany

thermofisher.com/rotors
© 2013-2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries unless otherwise indicated.
Delrin, TEFLON, and Viton are registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham Health As. RULON A and Tygon are registered trademarks of Saint-Gobain Performance Plastics. Alconox is a registered trademark of Alconox. Ficoll is a registered trademark of GE Healthcare. Haemo-Sol is a registered trademark of Haemo-Sol. Triton is a registered trademark of Union Carbide Corporation. Valox is a registered trademark of General Electric Co.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Shown pictures within the manual are examples and may differ considering the set parameters and language. Pictures of the user interface within the manual are showing the English version as example.

Australia +61 39757 4300 Austria +43 1 801 40 0 Belgium +32 9 272 54 82 China +800 810 5118, +400 650 5118 France +33 2 2803 2180 Germany national toll free 0800 1 536 376 Germany international +49 6184 90 6000 India toll free +1800 22 8374 India +91 22 6716 2200 Italy +39 02 95059 552

Japan +81 3 5826 1616 Korea +82 2 2023 0600 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Nordic / Baltic / CIS countries +358 10 329 2200 Russia +7 812 703 42 15, +7 495 739 76 41 Singapore +82 2 3420 8700 Spain / Portugal +34 93 223 09 18 Switzerland +41 44 454 12 12

UK / Ireland +44 870 609 9203 USA / Canada +1 866 984 3766
Other Asian Countries +852 3107 7600 Countries not listed +49 6184 90 6000

50143334 is the original instruction manual.

en


Adobe PDF Library 15.0 Adobe InDesign 15.0 (Windows)