KIMO KH 200 KISTOCK Datalogger for Humidity and Temperature
Technical Datasheet
KH 200 KISTOCK
Datalogger for humidity and temperature
- Measures up to 5 parameters simultaneously
- Large LCD display
- 2 external inputs
- Light sensor
- Fast data transfer (1,000 values per second)
- Up to 16,000 measurement points
- 2 configurable alarms per channel
- Compact and handy size
- Magnetic mount
- IP 67 or IP 40 enclosure and rubber protection on the edges
Technical Data
Displayed Units
°C, °F, %RH, mV, V, mA, A, Lux, °Ctd, °Ftd
Resolution
0.1°C, 0.1°F, 0.1%RH, 1mV, 0.001V, 0.001mA, 0.1A, 1 Lux
External Inputs
- 2 Jack connectors (2.5 stereo)
- 1 Mini-DIN connector (KH-200-D)
Alarms
2 points per channel
Measurement Interval
From 1s to 24h
Operating Temperature
-20 to +70°C (KH-200-A)
-20 to +80°C (KH-200-D)
Storage Temperature
-40 to +85°C
Battery Life
5 years* (*when storing 1 measurement value every 15 minutes at room temperature (20°C))
Thermo-hygrometer Sensor
- Sensor Type: CMOS
- Measurement Range: 5 to 95%RH
- Accuracy (GAL): ± 2.95%RH between 18°C and 28°C
- Response Time: t=50s (V =2m/s)
Temperature
- Measurement Range: -20 to +70°C (KH-200-A), -20 to +80°C (KH-200-D)
- Accuracy: ±1% of displayed value, ±0.4°C (+5°C≤T<+80°C); ±2% of displayed value, ±0.6°C (-20°C
- Response Time: t=25s (V =2m/s)
*Guaranteed accuracy (GAL) according to standard NFX 15-113 and table << 2000-2001 HYGROMETERS >>
EMC (GAL) = ±2.95%RH between 18 and 28°C (normal measurement range)
Measurement range: 5 to 95%RH, Drift: 1%RH /year
EMC = E + E + k (u²+u²+u²+u²)¹/²
E: linearity and hysteresis = ±1.33%RH
E: temperature coefficient = ± 0.42%RH with u: calibration uncertainty = ± 0.55%RH
u: uncertainty due to resolution = ± 0.003%RH
u: uncertainty due to production = ± 0.2%RH
u: uncertainty due to repeatability = 0.13%RH
k: factor = 2
External Temperature Sensors (accessories)
- Sensor Type: NTC
- Measurement Range: -40 to +120°C
- Accuracy: ±0.3°C (-25°C<T<+70°C); ±0.5°C
See technical datasheet << Measurement sensors and cables for Kistock Class 100/200 Datatoggers ».
Light Sensor
- Sensor Type: photodiode
- Measurement Range: 0 to 10,000 Lux
- Accuracy: ±10%
Current Signal Cable (accessories)
- Measurement Range: 0/4-20mA
- Accuracy: ±0.05mA
Voltage Signal Cable (accessories)
- Measurement Range: 0-2.5V
- Accuracy: ±0.002V
- Measurement Range: 0-10V
- Accuracy: ±0.02V
Current Clamp (accessories)
- Measurement Range: 0-600A
- Accuracy: ±1 to 2.5% of displayed value (depending on measurement range)
*All accuracy data in this datasheet are documented under laboratory conditions and can guarantee accurate measurements under similar circumstances, or with a calibration certificate with compensation for your environment.
Models
Model | Thermo-hygrometer | Display | External Input | Protection Class |
---|---|---|---|---|
KH-200-AN | Internal | Nei | 2 | IP 40 |
KH-200-AO | Internal | 2-line | 2 | IP 40 |
KH-200-DN | External sensor | Nei | 3 | IP 67 |
KH-200-DO | External sensor | 2-line | 3 | IP 67 |
KH 200 D is supplied with an external thermo-hygrometer sensor (Ref. KTHP 130).
Other Information
- Dimensions: 91 x 65 x 33 mm
- Weight: 85g
- Display: 2-line LCD
- Screen Size: 45 x 28.5 mm
- Controls: 2 buttons (SELECT and OK)
- Material: Polycarbonate casing, Elastomer protective sleeve
- Protection Class: IP 67 or IP 40
- Communication Port: 1 input for 3.5mm Jack connector (male)
- Electronics: Digital electronics, Lacquered printed circuit board, RoHS compliant
- Battery Life: 3.6V Lithium 1/2 AA
- Visible Alarm: 2 LEDs (green and red)
- Operating Conditions: Air and neutral gases
Dimensions
91 mm
65 mm
33 mm
Connections
KH-200-D External Inputs
- Mini-DIN connector
- Thermo-hygrometer connector
KH-200-A External Inputs
- Jack connector (2.5)
- Sensor inputs for: NTC temperature sensors, current signal cables, voltage signal cables, current clamps
PC Connection
- Jack connector (3.5)
- Input for PC data cable
Display
4
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
°C: Temperature in degrees Celsius
°F: Temperature in degrees Fahrenheit
%RH: Relative Humidity
°Ctd: Dew point temperature
V or mV: Volt in V or mV
A or mA: Current in A or mA
Logger Capabilities
5 logger options
KISTOCK can log measured values in 5 different ways:
- "Standard" mode => logs instantaneous value with the set interval
- "Minimum", "Maximum", and "Average" => logs the minimum, maximum, or average value over a given interval
- "Monitoring" => provides a historical and accurate report for troubleshooting intermittent faults, without stopping the logging. To do this, you need to: specify a logging interval for the period when the measured value is within your alarm limits; specify a logging interval for the period when the measured value has gone outside your alarm limits
You can also configure KISTOCK to log non-stop ("loop" when memory is full, the oldest measurement points are overwritten).
4 Methods to Start the Logger
Once you have set your logging interval, you can start the logger with:
- Delayed start (with predefined start and stop)
- Software
- By holding down the OK button for 5 seconds
- "Online". In this case, the measured values are saved directly and continuously to your PC in real-time.
6 Ways to Stop Logging
You can stop logging:
- Using a stop date
- Using a set period
- A set number of measurement points is reached
- When memory is full
- With the "Stop" option in the software
- By holding down the OK button for at least 5 seconds, if this function has been activated in the software.
Accessories and Cables
Wide selection of NTC temperature sensors (general, insertion, ambient, wire, Velcro sensors, and sensors with handles).
- Current and voltage cables
- Current clamps
See technical datasheet "Measurement sensors and cables for Kistock Class 100/200 Datatoggers".
KILOG Software
- Configuration and data processing software
- KILOG allows you to configure, save, and process data easily and quickly
- Software: Ref. KILOG
- USB interface (data cable): Ref. I-KIC2
- Complete set*: Ref. KIC2 KILOG (*includes KILOG software and USB data cable)
KISTOCK-PC Interface
Connect your datalogger to your PC via USB port with high transfer speed. Ref. I-KIC2
KILOG CFR Software
KILOG CFR software is a key tool for users requiring traceability according to 21CFR-Part 11 standards. Security and results are guaranteed: it is not possible to modify or alter measurement data.
- Interface (data cable): Ref. I-KIC2
- Complete set: KILOG CFR software + 1 interface: Ref. KIC2 CFR
Accessories
KNT Data Collector
The KNT data collector allows you to empty the memory of your dataloggers and start a new logging period without using a PC. The data transferred to the data collector is then transferred to your PC for further processing as usual. KNT can hold a total of 500,000 measurement values. Ref. KNT 300
Printer for KNT 300 Data Collector
Ref. ITP
Secure Wall Bracket
KIMO has designed a new anti-theft wall bracket without a padlock. The logger cannot be moved without the key. Ref. KAV
When your Kistock is mounted on the mounting bracket, insert the locking pin.
Unlock: Insert the key into the locking pin and turn 1¼ turns. Remove the key with the pin. Your Kistock is now free.
Mounting/Placement
KISTOCK can be mounted in various ways very easily:
- Magnetic mount (see image)
- Secure wall bracket (optional accessory, see accessories)
- Nail mount
Wall bracket and nail mount
Magnetic mount
Secure wall bracket
How to Change the Battery?
With a 5-year battery life (*), KISTOCK guarantees long-term measurements. To change the battery:
- Remove the instrument's rear screw with a screwdriver
- Remove the back cover
- Insert the new battery, observing polarity
- Replace the back cover and the screw
(*) At 1 logged measurement value every 15 minutes at 20°C
Calibration
KISTOCK dataloggers can be ordered with a factory-supplied calibration certificate as an option.
Warranty
KISTOCK dataloggers have a 1-year warranty against manufacturing defects.
Extension Cable for NTC Temperature Sensors
Made of PVC HT, 5m long, with Jack connectors (male and female) Ref. KRC 5
PS! You can connect several cables together (max 25 meters)
- Lanyard. Ref. KDC
- Lithium 1/2 AA battery. Ref. KBL
Importer
Instrumentcompaniet AS
Ryensvingen 1, 0680 Oslo
Tel: 23 30 21 00, Fax: 23 30 21 01
Epost: ic@instrumentcompaniet.no
www.instrumentcompaniet.no
instrument companiet as