Test Report-SAR Part1

ONYX INTERNATIONAL INC. GO6 E Ink Tablet, ePaper Tablet, Digital Paper, E reader, Paper tablet, eBook reader XR3-GO6 XR3GO6 go6

PDF preview unavailable. Download the PDF instead.

GetApplicationAttachment.html?id=7555566
Shenzhen CTA Testing Technology Co., Ltd.
Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai
Street, Bao'an District, Shenzhen, China

TEST REPORT

Report Reference No. .....................: CTA24050900405 FCC ID ........................................... : XR3-GO6
Compiled by ( position+printed name+signature)..: File administrators Jinghua Xiao

Supervised by ( position+printed name+signature)..: Project Engineer Lushan Kong

Approved by ( position+printed name+signature)..: RF Manager Eric Wang

Date of issue.....................................: May. 23, 2024

Testing Laboratory Name ..............: Shenzhen CTA Testing Technology Co., Ltd.

Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Address ............................................. :
Fuhai Street, Bao`an District, Shenzhen, China

Applicant's name ............................: ONYX INTERNATIONAL INC.

Address ............................................. : Test specification........................... :

Room 101, Building 4, No. 202 Shiyu Road, Nansha District, Guangzhou City, Guangdong Province, China

IEC 62209-2:2010; IEEE 1528:2013; FCC 47 CFR Part 2.1093; Standard ...........................................: ANSI/IEEE C95.1:2005; Reference FCC KDB 447498;
KDB 865664; KDB 248227; KDB 616217 Shenzhen CTA Testing Technology Co., Ltd. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the Shenzhen CTA Testing Technology Co., Ltd. is acknowledged as copyright owner and source of the material. Shenzhen CTA Testing Technology Co., Ltd. takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context.
E Ink Tablet, ePaper Tablet, Digital Paper, E reader, Paper tablet, Test item description ..................... :
eBook reader
Trade Mark........................................: BOOX

Manufacturer.....................................: Onyx International Inc.

Model/Type reference .......................: Go 6

Listed Models ..................................: Refer to page 2

Rating ...............................................: DC 3.8V From battery and DC 5.0V From external circuit

Result................................................: PASS

Report No.: CTA24040900105
TEST REPORT

Page 2 of 69

Equipment under Test

: E Ink Tablet, ePaper Tablet, Digital Paper, E reader, Paper tablet, eBook reader

Model /Type

: Go 6

Listed Models

: BOOX Go 6, BOOX Go 6 Plus, BOOX Go 6 Pro, BOOX Go 6 Lite, BOOX Go Color 6, BOOX Go Color 6 Plus, BOOX Go Color 6 Pro

Applicant

: Onyx International Inc.

Address

: Room 101, Building 4, No. 202 Shiyu Road, Nansha District, Guangzhou City, Guangdong Province, China

Manufacturer

: Onyx International Inc.

Address

: Room 101, Building 4, No. 202 Shiyu Road, Nansha District, Guangzhou City, Guangdong Province, China

Test Result:

PASS

The test report merely corresponds to the test sample. It is not permitted to copy extracts of these test result without the written permission of the test laboratory.

Report No.: CTA24040900105
  Revision History  

Page 3 of 69

REV. Rev.1.0

ISSUED DATE May. 13, 2024

DESCRIPTION Initial Test Report Release

Report No.: CTA24040900105

Page 4 of 69

Contents

1 Statement of Compliance ............................................................................................... 6 2 General Information........................................................................................................ 7
General Remarks ............................................................................................................................................. 7 Description of Equipment Under Test (EUT) ..................................................................................................... 7 Device Category and SAR Limits ..................................................................................................................... 8 Applied Standard .............................................................................................................................................. 8 Test Facility....................................................................................................................................................... 8 Environment of Test Site................................................................................................................................... 9 Test Configuration............................................................................................................................................. 9
3 Specific Absorption Rate (SAR) ................................................................................... 10
Introduction..................................................................................................................................................... 10 SAR Definition ................................................................................................................................................ 10
4 SAR Measurement System .......................................................................................... 11
E-Field Probe...................................................................................................................................................11 Data Acquisition Electronics (DAE)................................................................................................................. 12 Robot .............................................................................................................................................................. 13 Measurement Server ...................................................................................................................................... 13 Phantom ......................................................................................................................................................... 14 Device Holder ................................................................................................................................................. 14 Data Storage and Evaluation.......................................................................................................................... 15
5 Test Equipment List...................................................................................................... 17 6 Tissue Simulating Liquids............................................................................................. 18 7 System Verification Procedures ................................................................................... 20 8 EUT Testing Position.................................................................................................... 22
Body-Supported Device Configurations.......................................................................................................... 22
9 Measurement Procedures ............................................................................................ 23
Spatial Peak SAR Evaluation ......................................................................................................................... 23 Power Reference Measurement ..................................................................................................................... 23 Area Scan Procedures ................................................................................................................................... 24 Zoom Scan Procedures .................................................................................................................................. 24 Volume Scan Procedures ............................................................................................................................... 25 Power Drift Monitoring .................................................................................................................................... 25
10 TEST CONDITIONS AND RESULTS........................................................................... 26
Conducted Power Results .............................................................................................................................. 26 Transmit Antennas.......................................................................................................................................... 28 SAR Test Exclusion and Estimated SAR ........................................................................................................ 29 SAR Test Results............................................................................................................................................ 31 SAR Measurement Variability ......................................................................................................................... 33 Simultaneous Transmission Analysis.............................................................................................................. 34
11 Measurement Uncertainty ............................................................................................ 35 Appendix A. EUT Photos and Test Setup Photos ......................................................... 37 Appendix B. Plots of SAR System Check ..................................................................... 38

Report No.: CTA24040900105

Page 5 of 69

Appendix C. Plots of SAR Test Data............................................................................. 41

Appendix D. DASY System Calibration Certificate........................................................ 44

Report No.: CTA24040900105
1 Statement of Compliance
<Highest SAR Summary>

Page 6 of 69

This device is in compliance with Specific Absorption Rate (SAR) for general population/uncontrolled exposure limits (1.6 W/kg) specified in FCC 47 CFR part 2 (2.1093) and ANSI/IEEE C95.1-2005, and had been tested in accordance with the measurement methods and procedures specified in IEEE 1528-2013 The maximum results of Specific Absorption Rate (SAR) found during testing are as follows.

Frequency Band
WLAN2.4G WLAN5.2G WLAN5.8G SAR Test Limit (W/Kg) Test Result

<Highest SAR Summary>
Highest Reported 1g-SAR(W/Kg)

Body (0mm)

0.461 0.475 0.526

1.60 PASS

Simultaneous Reported SAR
(W/Kg)
N/A

This device is in compliance with Specific Absorption Rate (SAR) for general population/uncontrolled exposure limits (1.6 W/kg) specified in FCC 47 CFR part 2 (2.1093) and ANSI/IEEE C95.1-2005, and had been tested in accordance with the measurement methods and procedures specified in IEEE 1528-2013

Report No.: CTA24040900105
2 General Information

General Remarks
Date of receipt of test sample

: May. 06, 2024

Page 7 of 69

Testing commenced on

: May. 06, 2024

Testing concluded on

: May. 13, 2024

Description of Equipment Under Test (EUT)

Product Name:

E Ink Tablet, ePaper Tablet, Digital Paper, E reader, Paper tablet, eBook reader

Model/Type reference:

Go 6

Power supply:

DC 3.8V From battery and DC 5.0V From external circuit

Testing sample ID: Hardware version:

CTA240509004-1# (Engineer sample) CTA240509004-2# (Normal sample) V1.0

Software version:

V1.0

Tx Frequency:
Type of Modulation: Category of device:

SRD: BT:2402~2480MHz 2.4G WIFI: 2412~2462MHz 5G WIFI: 5180~5240MHz, 5745~5825MHz
BT: GFSK, /4DQPSK, 8DPSK 2.4G WIFI: BPSK, QPSK,16QAM,64QAM 5G WIFI: BPSK, QPSK,16QAM,64QAM, 256QAM Portable device

Remark: The above DUT's information was declared by manufacturer. Please refer to the specifications or user's manual for more detailed description.

Report No.: CTA24040900105
Device Category and SAR Limits

Page 8 of 69

This device belongs to portable device category because its radiating structure is allowed to be used within 20 centimeters of the body of the user. Limit for General Population/Uncontrolled exposure should be applied for this device, it is 1.6 W/kg as averaged over any 1 gram of tissue.

Applied Standard
The Specific Absorption Rate (SAR) testing specification, method, and procedure for this device is in accordance with the following standards:
 FCC 47 CFR Part 2 (2.1093:2013)  ANSI/IEEE C95.1:2005  IEEE Std 1528:2013  KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz v01r04  KDB 865664 D02 RF Exposure Reporting v01r02  KDB 447498 D01 General RF Exposure Guidance v06  KDB 248227 D01 802 11 Wi-Fi SAR v02r02  KDB 616217 D04 SAR for laptop and tablets v01r02

Test Facility
FCC-Registration No.: 517856 Designation Number: CN1318 Shenzhen CTA Testing Technology Co., Ltd. has been listed on the US Federal Communications Commission list of test facilities recognized to perform electromagnetic emissions measurements.

A2LA-Lab Cert. No.: 6534.01 Shenzhen CTA Testing Technology Co., Ltd. has been listed by American Association for Laboratory Accreditation to perform electromagnetic emission measurement.

ISED#: 27890 CAB identifier: CN0127 Shenzhen CTA Testing Technology Co., Ltd. has been listed by Innovation, Science and Economic Development Canada to perform electromagnetic emission measurement.

The 3m-Semi anechoic test site fulfils CISPR 16-1-4 according to ANSI C63.10 and CISPR 16-1-4:2010.

Report No.: CTA24040900105
Environment of Test Site
Items Temperature () Humidity (%RH)

Required 18-25 30-70

Page 9 of 69
Actual 22~23 55~65

Test Configuration
The device was controlled by using a base station emulator. Communication between the device and the emulator was established by air link. The distance between the EUT and the antenna of the emulator is larger than 50 cm and the output power radiated from the emulator antenna is at least 30 dB smaller than the output power of EUT. The EUT was set from the emulator to radiate maximum output power during all tests. For WLAN SAR testing, WLAN engineering testing software installed on the EUT can provide continuous transmitting RF signal.

Report No.: CTA24040900105
3 Specific Absorption Rate (SAR)

Page 10 of 69

Introduction
SAR is related to the rate at which energy is absorbed per unit mass in an object exposed to a radio field. The SAR distribution in a biological body is complicated and is usually carried out by experimental techniques or numerical modeling. The standard recommends limits for two tiers of groups, occupational/controlled and general population/uncontrolled, based on a person's awareness and ability to exercise control over his or her exposure. In general, occupational/controlled exposure limits are higher than the limits for general population/uncontrolled.
SAR Definition
The SAR definition is the time derivative (rate) of the incremental energy (dW) absorbed by (dissipated in) an incremental mass (dm) contained in a volume element (dv) of a given density ().The equation description is as below:
     =  () =  ()
SAR is expressed in units of Watts per kilogram (W/kg) SAR measurement can be either related to the temperature elevation in tissue by
  =  (  )
Where: C is the specific head capacity, T is the temperature rise and tisthe exposure duration, or related to the electrical field in the tissue by
||  = 
Where: is the conductivity of the tissue,  is the mass density of the tissue and E is the RMS electrical field strength. However for evaluating SAR of low power transmitter, electrical field measurement is typically applied.

Report No.: CTA24040900105
4 SAR Measurement System

Page 11 of 69

DASY System Configurations
The DASY system for performance compliance tests is illustrated above graphically. This system consists of the following items:
 A standard high precision 6-axis robot with controller, a teach pendant and software  A data acquisition electronic (DAE) attached to the robot arm extension  A dosimetric probe equipped with an optical surface detector system  The electro-optical converter (EOC) performs the conversion between optical and electrical
signals  A measurement server performs the time critical tasks such as signal filtering, control of the
robot operation and fast movement interrupts.  A probe alignment unit which improves the accuracy of the probe positioning  A computer operating Windows XP  DASY software  Remove control with teach pendant and additional circuitry for robot safety such as warming
lamps, etc.  The SAM twin phantom  A device holder  Tissue simulating liquid  Dipole for evaluating the proper functioning of the system
components are described in details in the following sub-sections.
E-Field Probe
The SAR measurement is conducted with the dosimetric probe (manufactured by SPEAG).The probe is specially designed and calibrated for use in liquid with high permittivity. The dosimetric probe has special calibration in liquid at different frequency. This probe has a built in optical surface detection system to prevent from collision with phantom.

Report No.: CTA24040900105

 E-Field Probe Specification

<EX3DV4 Probe>

Construction

Symmetrical design with triangular core

Built-in shielding against static charges

PEEK enclosure material (resistant to organic

solvents, e.g., DGBE)

Frequency

10 MHz to 6 GHz; Linearity: ±0.2 dB

Directivity

±0.3 dB in HSL (rotation around probe axis)

± 0.5 dB in tissue material (rotation normal to

probe axis)

Dynamic Range 10 µW/g to 100 W/kg; Linearity: ±0.2 dB (noise:

typically< 1 µW/g)

Dimensions

Overall length: 330 mm (Tip: 20 mm)

Tip diameter: 2.5 mm (Body: 12 mm)

Typical distance from probe tip to dipole

centers: 1 mm

Page 12 of 69 Photo of EX3DV4

 E-Field Probe Calibration
Each probe needs to be calibrated according to a dosimetric assessment procedure with accuracy better than ±10%. The spherical isotropy shall be evaluated and within ±0.25dB. The sensitivity parameters (NormX, NormY, and NormZ), the diode compression parameter (DCP) and the conversion factor (ConvF) of the probe are tested. The calibration data can be referred to appendix C of this report.

Data Acquisition Electronics (DAE)
The data acquisition electronics (DAE) consists of a highly sensitive electrometer-grade preamplifier with auto-zeroing, a channel and gain-switching multiplexer, a fast 16 bit AD-converter and a command decoder and control logic unit. Transmission to the measurement server is accomplished through an optical downlink for data and status information as well as an optical uplink for commands and the clock. The input impedance of the DAE is 200MOhm; the inputs are symmetrical and floating. Common mode rejection is above 80dB.

Photo of DAE

Report No.: CTA24040900105
Robot

Page 13 of 69

The SPEAG DASY system uses the high precision robots (DASY5: TX60XL) type from Stäubli SA (France). For the 6-axis controllersystem, the robot controller version (DASY5: CS8c) from Stäubli is used. The Stäublirobot series have many features that are important for our application:
 High precision (repeatability ±0.035 mm)  High reliability (industrial design)  Jerk-free straight movements  Low ELF interference (the closed metallic construction shields against motor control fields)

Photo of DASY5 Measurement Server
The measurement server is based on a PC/104 CPU board with CPU (DASY5: 400 MHz, Intel Celeron), chipdisk (DASY5: 128 MB), RAM (DASY5: 128 MB). The necessary circuits for communication with the DAE electronic box, as well as the 16 bit AD converter system for optical detection and digital I/O interface are contained on the DASY I/O board, which is directly connected to the PC/104 bus of the CPU board. The measurement server performs all the real-time data evaluation for field measurements and surface detection, controls robot movements and handles safety operations.
Photo of Server for DASY5

Report No.: CTA24040900105

Phantom

<SAM Twin Phantom>

Shell Thickness

2 ±0.2 mm;

Center ear point: 6 ±0.2 mm

Filling Volume

Approx. 25 liters

Dimensions

Length: 1000 mm; Width: 500 mm;

Height: adjustable feet

Measurement Areas Left Hand, Right Hand, Flat Phantom

Page 14 of 69

Photo of SAM Phantom

The bottom plate contains three pair of bolts for locking the device holder. The device holder positions

are adjusted to the standard measurement positions in the three sections. A white cover is provided to

tap the phantom during off-periods to prevent water evaporation and changes in the liquid parameters.

On the phantom top, three reference markers are provided to identify the phantom position with respect

to the robot.

<ELI4 Phantom>

Shell Thickness

2 ±0.2 mm (sagging: <1%)

Filling Volume

Approx. 30 liters

Dimensions

Major ellipse axis: 600 mm

Minor axis:400 mm

Photo of ELI4 Phantom
The ELI4 phantom is intended for compliance testing of handheld and body-mounted wireless devices in the frequency range of 30 MHz to 6 GHz. ELI4 is fully compatible with standard and all known tissue simulating liquids.
Device Holder
The SAR in the phantom is approximately inversely proportional to the square of the distance between the source and the liquid surface. For a source at 5 mm distance, a positioning uncertainty of ±0.5mm would produce a SAR uncertainty of ± 20%. Accurate device positioning is therefore crucial for accurate and repeatable measurements. The positions in which the devices must be measured are defined by the standards. The DASY device holder is designed to cope with different positions given in the standard. It has two scales for the device rotation (with respect to the body axis) and the device inclination (with respect to the line between the ear reference points). The rotation center for both scales is the ear reference point (ERP).Thus the device needs no repositioning when changing the angles.

Report No.: CTA24040900105

Page 15 of 69

The DASY device holder is constructed of low-loss POM material having the following dielectric

parameters: relative permittivity  = 3 and loss tangent  = 0.02. The amount of dielectric material has

been reduced in the closest vicinity of the device, since measurements have suggested that the

influence of the clamp on the test results could thus be lowered.

Device Holder

Data Storage and Evaluation

 Data Storage

The DASY software stores the assessed data from the data acquisition electronics as raw data (in

microvolt readings from the probe sensors), together with all the necessary software parameters for the

data evaluation (probe calibration data, liquid parameters and device frequency and modulation data) in

measurement files. The post-processing software evaluates the desired unit and format for output each

time the data is visualized or exported. This allows verification of the complete software setup even after

the measurement and allows correction of erroneous parameter settings. For example, if a measurement

has been performed with an incorrect crest factor parameter in the device setup, the parameter can be

corrected afterwards and the data can be reevaluated.

The measured data can be visualized or exported in different units or formats, depending on the selected

probe type (e.g., [V/m], [A/m], [W/kg]). Some of these units are not available in certain situations or give

meaningless results, e.g., a SAR-output in a non-lose media, will always be zero. Raw data can also be

exported to perform the evaluation with other software packages.

 Data Evaluation

The DASY post-processing software (SEMCAD) automatically executes the following procedures to

calculate the field units from the microvolt readings at the probe connector. The parameters used in the

evaluation are stored in the configuration modules of the software

Probe parameters - Sensitivity

Normi, ai0, ai1, ai2

- Conversion factor

ConvFi

- Diode compression point dcpi

Device parameters - Frequency

f

- Crest factor

cf

Media parameters - Conductivity



- Density



Report No.: CTA24040900105

Page 16 of 69

These parameters must be set correctly in the software. They can be found in the component documents

or they can be imported into the software from the configuration files issued for the DASY components.

In the direct measuring mode of the multi-meter option, the parameters of the actual system setup are

used. In the scan visualization and export modes, the parameters stored in the corresponding document

files are used.

The first step of the evaluation is a linearization of the filtered input signal to account for the compression

characteristics of the detector diode. The compensation depends on the input signal, the diode type and

the DC-transmission factor from the diode to the evaluation electronics. If the exciting field is pulsed, the

crest factor of the signal must be known to correctly compensate for peak power.

The formula for each channel can be given as



=



+





 

with Vi= compensated signal of channel i, (i = x, y, z) Ui = input signal of channel i, (i = x, y, z) cf = crest factor of exciting field (DASY parameter) dcpi = diode compression point (DASY parameter)
From the compensated input signals, the primary field data for each channel can be evaluated

E-field Probes = 

H-field

Probes

=





++ 

with Vi = compensated signal of channel i,(i= x, y, z) Normi= sensor sensitivity of channel i, (i= x, y, z), V/(V/m)2 for E-field Probes ConvF= sensitivity enhancement in solution aij= sensor sensitivity factors for H-field probes f = carrier frequency [GHz] Ei= electric field strength of channel iin V/m Hi= magnetic field strength of channel iin A/m

The RSS value of the field components gives the total field strength (Hermitian magnitude)

 =  +  + 

The primary field data are used to calculate the derived field units.



=









 

with SAR = local specific absorption rate in W/kg Etot= total field strength in V/m  = conductivity in [mho/m] or [Siemens/m]  = equivalent tissue density in g/cm3
Note that the density is set to 1, to account for actual head tissue density rather than the density of the tissue simulating liquid.

Report No.: CTA24040900105
5 Test Equipment List

Page 17 of 69

Manufacturer

Name of Equipment

Type/Model

Serial Number

Calibration Last Cal. Due Date

SPEAG 2450MHz System Validation Kit D2450V2

745

Aug. 28,2023 Aug. 27,2026

SPEAG

5GHz System Validation Kit

D5GHzV2

1031

Feb.16, 2023 Feb.15, 2026

Rohde & Schwarz

UNIVERSAL RADIO COMMUNICATION TESTER

CMW500

1201.0002K50Nov.05, 2023 Nov.04, 2024
104209-JC

SPEAG

Data Acquisition Electronics

DAE3

428

Aug.30,2023 Aug.29,2024

SPEAG

Dosimetric E-Field Probe

EX3DV4

7380

June 21,2023 June 20,2024

Agilent

ENA Series Network Analyzer

E5071C

MY46317418 Oct.25, 2023 Oct.24, 2024

SPEAG

DAK

DAK-3.5

1226

NCR

NCR

SPEAG

SAM Twin Phantom

QD000P40CD

1802

NCR

NCR

SPEAG

ELI Phantom

QDOVA004AA

2058

NCR

NCR

AR

Amplifier

ZHL-42W

QA1118004

NCR

NCR

Agilent

Power Meter

N1914A

MY50001102 Oct.25, 2023 Oct.24, 2024

Agilent

Power Sensor

N8481H

MY51240001 Oct.25, 2023 Oct.24, 2024

R&S

Spectrum Analyzer

N9020A

MY51170037 Oct.25, 2023 Oct.24, 2024

Agilent

Signal Generation

N5182A

MY48180656 Oct.25, 2023 Oct.24, 2024

Worken

Directional Coupler

0110A05601O-10 COM5BNW1A2 Oct.25, 2023 Oct.24, 2024

Note:

1. The calibration certificate of DASY can be referred to appendix C of this report.

2. The dipole calibration interval can be extended to 3 years with justification. The dipoles are also not physically

damaged, or repaired during the interval.

3. The Insertion Loss calibration of Dual Directional Coupler and Attenuator were characterized via the network analyzer

and compensated during system check.

4. The dielectric probe kit was calibrated via the network analyzer, with the specified procedure (calibrated in pure water)

and calibration kit (standard) short circuit, before the dielectric measurement. The specific procedure and calibration

kit are provided by Agilent.

5. In system check we need to monitor the level on the power meter, and adjust the power amplifier level to have precise

power level to the dipole; the measured SAR will be normalized to 1W input power according to the ratio of 1W to

the input power to the dipole. For system check, the calibration of the power amplifier is deemed not critically required

for correct measurement; the power meter is critical and we do have calibration for it

Report No.: CTA24040900105
6 Tissue Simulating Liquids

Page 18 of 69

For the measurement of the field distribution inside the SAM phantom with DASY, the phantom must be filled with around 25 liters of homogeneous body tissue simulating liquid. For head SAR testing, the liquid height from the ear reference point (ERP) of the phantom to the liquid top surface is larger than 15 cm, which is shown in Fig. 6.1. For body SAR testing, the liquid height from the center of the flat phantom to the liquid top surface is larger than 15 cm, which is shown as followed:

Photo of Liquid Height

The following table gives the recipes for tissue simulating liquid.

Frequency Water Sugar Cellulose Salt Preventol DGBE Conductivity Permittivity

(MHz)

(%) (%)

(%)

(%)

(%)

(%)

()

(r)

For Head

835

40.3 57.9

0.2

1.4

0.2

0

0.90

41.5

1800,1900,2000 55.2

0

0

0.3

0

44.5

1.40

40.0

2450

55.0

0

0

0

0

45.0

1.80

39.2

2600

54.8

0

0

0.1

0

45.1

1.96

39.0

For Body

835

50.8 48.2

0

0.9

0.1

0

0.97

55.2

1800,1900,2000 70.2

0

0

0.4

0

29.4

1.52

53.3

2450

68.6

0

0

0

0

31.4

1.95

52.7

2600

65.5

0

0

0

0

31.5

2.16

52.5

Report No.: CTA24040900105

The following table shows the measuring results for simulating liquid.

Measured

Target Tissue

Measured Tissue

Frequency r
(MHz)

Dev.

Dev.



r



(%)

(%)

2450

39.2

1.80

38.809 -1.00% 1.775 -1.39%

5250

35.9

4.71

35.355 -1.52% 4.670 -0.85%

5750

35.4

5.22

36.190 2.23% 5.080 -2.68%

Page 19 of 69

Liquid Temp.
22.6 22.2 22.6

Test Data
05/21/2024 05/22/2024 05/22/2024

Report No.: CTA24040900105
7 System Verification Procedures

Page 20 of 69

Each DASY system is equipped with one or more system validation kits. These units, together with the predefined measurement procedures within the DASY software, enable the user to conduct the system performance check and system validation. System validation kit includes a dipole, tripod holder to fix it underneath the flat phantom and a corresponding distance holder.

 Purpose of System Performance check
The system performance check verifies that the system operates within its specifications. System and operator errors can be detected and corrected. It is recommended that the system performance check be performed prior to any usage of the system in order to guarantee reproducible results. The system performance check uses normal SAR measurements in a simplified setup with a well characterized source. This setup was selected to give a high sensitivity to all parameters that might fail or vary over time. The system check does not intend to replace the calibration of the components, but indicates situations where the system uncertainty is exceeded due to drift or failure.

 System Setup
In the simplified setup for system evaluation, the EUT is replaced by a calibrated dipole and the power source is replaced by a continuous wave that comes from a signal generator. The calibrated dipole must be placed beneath the flat phantom section of the SAM twin phantom with the correct distance holder. The distance holder should touch the phantom surface with a light pressure at the reference marking and be oriented parallel to the long side of the phantom. The equipment setup is shown below:

Spacer

3D Probe positioner

s

Field probe

Flat Phantom

Dipole

Signal

Amp

Generator

3dB
Att3

Dir.Coupler Cable

Att2 PM3

x Att1
PM1

PM2

System Setup for System Evaluation

Report No.: CTA24040900105

Page 21 of 69

Photo of Dipole Setup
 Validation Results
Comparing to the original SAR value provided by SPEAG, the verification data should be within its specification of 10%. The table below shows the target SAR and measured SAR after normalized to 1W input power. It indicates that the system performance check can meet the variation criterion and the plots can be referred to Appendix A of this report.

Date

Frequency (MHz)

Power fed onto reference dipole
(mW)

Targeted SAR 1g (W/kg)

05/21/2024

2450

250

52.7

05/22/2024

5250

100

77.7

05/22/2024

5750

100

78.0

Measured SAR1g (W/kg)
12.90 7.89 7.99

Normalized SAR (W/kg)

Deviation (%)

51.60 78.90 79.90

-2.09% 1.54% 2.44%

Report No.: CTA24040900105
8 EUT Testing Position

Page 22 of 69

Body-Supported Device Configurations
According to KDB 616217 section 4.3, SAR should be separately assessed with each surface and separation distance positioned against the flat phantom that correspond to the intended use as specified by the manufacturer. The antennas in tablets are typically located near the back (bottom) surface and/or along the edges of the devices; therefore, SAR evaluation is required for these configurations. Exposures from antennas through the front (top) surface of the display section of a full-size tablet, away from the edges, are generally limited to the user's hands. Exposures to hands for typical consumer transmitters used in tablets are not expected to exceed the extremity SAR limit; therefore, SAR evaluation for the front surface of tablet display screens are generally not necessary, except for tablets that are designed to require continuous operations with the hand(s) next to the antenna(s).
 To position the device parallel to the phantom surface with either keypad up or down.
 To adjust the device parallel to the flat phantom.
 To adjust the distance between the device surface and the flat phantom to 0 mm.
 When each surface is measurement, the SAR Test Exclusion Threshold in KDB 447498 should be applied.

Fig.81 Illustration for Body Position

Report No.: CTA24040900105
9 Measurement Procedures

Page 23 of 69

The measurement procedures are as follows: (a) Use base station simulator (if applicable) or engineering software to transmit RF power continuously (continuous Tx) in the middle channel. (b) Keep EUT to radiate maximum output power or 100% duty factor (if applicable) (c) Measure output power through RF cable and power meter. (d) Place the EUT in the positions as setup photos demonstrates. (e) Set scan area, grid size and other setting on the DASY software. (f) Measure SAR transmitting at the middle channel for all applicable exposure positions. (g) Identify the exposure position and device configuration resulting the highest SAR (h) Measure SAR at the lowest and highest channels attheworst exposure position and device configuration if applicable.
According to the test standard, the recommended procedure for assessing the peak spatial-average SAR value consists of the following steps: (a) Power reference measurement (b) Area scan (c) Zoom scan (d) Power drift measurement
Spatial Peak SAR Evaluation
The procedure for spatial peak SAR evaluation has been implemented according to the test standard. It can be conducted for 1g and 10g, as well as for user-specific masses. The DASY software includes all numerical procedures necessary to evaluate the spatial peak SAR value. The base for the evaluation is a "cube" measurement. The measured volume must include the 1g and 10g cubes with the highest averaged SAR values. For that purpose, the center of the measured volume is aligned to the interpolated peak SAR value of a previously performed area scan. The entire evaluation of the spatial peak values is performed within the post-processing engine (SEMCAD). The system always gives the maximum values for the 1g and 10g cubes. The algorithm to find the cube with highest averaged SAR is divided into the following stages:
(a) Extraction of the measured data (grid and values) from the Zoom Scan (b) Calculation of the SAR value at every measurement point based on all stored data (A/D values
and measurement parameters) (c) Generation of a high-resolution mesh within the measured volume (d) Interpolation of all measured values form the measurement grid to the high-resolution grid (e) Extrapolation of the entire 3-D field distribution to the phantom surface over the distance from
sensor to surface (f) Calculation of the averaged SAR within masses of 1g and 10g
Power Reference Measurement
The Power Reference Measurement and Power Drift Measurements are for monitoring the power drift of the device under test in the batch process. The minimum distance of probe sensors to surface determines the closest measurement point to phantom surface. This distance cannot be smaller than the distance of sensor calibration points to probe tip as defined in the probe properties.

Report No.: CTA24040900105
Area Scan Procedures

Page 24 of 69

The area scan is used as a fast scan in two dimensions to find the area of high field values, before doing a fine measurement around the hot spot. The sophisticated interpolation routines implemented in DASY software can find the maximum found in the scanned area, within a range of the global maximum. The range (in dB0 is specified in the standards for compliance testing. For example, a 2 dB range is required in IEEE standard 1528 and IEC 62209 standards, whereby 3 dB is a requirement when compliance is assessed in accordance with the ARIB standard (Japan), if only one zoom scan follows the area scan, then only the absolute maximum will be taken as reference. For cases where multiple maximums are detected, the number of zoom scans has to be increased accordingly. Area scan parameters extracted from FCC KDB 865664 D01 SAR measurement 100 MHz to 6 GHz.

Zoom Scan Procedures
Zoom scans are used assess the peak spatial SAR values within a cubic averaging volume containing 1 gram and 10gram of simulated tissue. The zoom scan measures points (refer to table below) within a cube shoes base faces are centered on the maxima found in a preceding area scan job within the same procedure. When the measurement is done, the zoom scan evaluates the averaged SAR for 1 gram and 10 gram and displays these values next to the job's label. Zoom scan parameters extracted from FCC KDB 865664 D01 SAR measurement 100 MHz to 6 GHz.

Report No.: CTA24040900105
Volume Scan Procedures

Page 25 of 69

The volume scan is used for assess overlapping SAR distributions for antennas transmitting in different frequency bands. It is equivalent to an oversized zoom scan used in standalone measurements. The measurement volume will be used to enclose all the simultaneous transmitting antennas. For antennas transmitting simultaneously in different frequency bands, the volume scan is measured separately in each frequency band. In order to sum correctly to compute the 1g aggregateSAR, the EUT remain in the same test position for all measurements and all volume scan use the same spatial resolution and grid spacing. When all volume scan were completed, the software, SEMCAD postprocessor can combine and subsequently superpose these measurement data to calculating the multiband SAR.

Power Drift Monitoring
All SAR testing is under the EUT install full charged battery and transmit maximum output power. In DASY measurement software, the power reference measurement and power drift measurement procedures are used for monitoring the power drift of EUT during SAR test. Both these procedures measure the field at a specified reference position before and after the SAR testing. The software will calculate the field difference in dB. If the power drift more than 5%, the SAR will be retested.

Report No.: CTA24040900105
10 TEST CONDITIONS AND RESULTS

Page 26 of 69

Conducted Power Results
<WLAN 2.4GHz Conducted Power>

Mode

Channel

Frequency (MHz)

1

802.11b

6

11

1

802.11g

6

11

1

802.11n(HT20)

6

11

3

802.11n(HT40)

6

9

2412 2437 2462 2412 2437 2462 2412 2437 2462 2422 2437 2452

Conducted Peak Output
Power(dBm)
13.68 13.75 13.28 12.70 13.59 12.70 13.07 13.13 12.62 13.07 13.56 13.14

Conducted
Average Output
Power(dBm)
12.56 12.69 12.22 11.41 12.11 11.59 11.68 11.89 11.37 10.91 11.33 11.11

Tune-up limit
(dBm)
13.00 13.00 13.00 12.50 12.50 12.50 12.50 12.50 12.50 12.00 12.00 12.00

<WLAN 5.2GHz Conducted Power>

Type
802.11a
802.11n(HT20) 802.11n(HT40) 802.11ac(HT20) 802.11ac(HT40) 802.11ac(HT80)

Channel
36 40 48 36 40 48 38 46 36 40 48 38 46 42

Frequency
(MHz)
5180 5200 5240 5180 5200 5240 5190 5230 5180 5200 5240 5190 5230 5210

Conducted Average
Output Power(dBm)
12.38 10.13 9.04 12.27 10.04 8.97 12.32 10.52 12.30 10.05 8.98 12.36 10.58 11.25

Tune-up limit
(dBm)
13.00 11.00 10.00 13.00 11.00 10.00 13.00 11.00 13.00 11.00 10.00 13.00 11.00 12.0

<WLAN 5.8GHz Conducted Power>

Type 802.11a
802.11n(HT20) 802.11n(HT40) 802.11ac(HT20) 802.11ac(HT40) 802.11ac(HT80)

Channel
149 157 165 149 157 165 151 159 149 157 165 151 159 155

Frequency (MHz)
5745 5785 5825 5745 5785 5825 5755 5795 5745 5785 5825 5755 5795 5775

Conducted Average Output Power(dBm)
12.60 12.10 11.60 12.53 12.04 11.50 12.90 12.04 12.56 12.07 11.56 12.93 12.10 12.83

Tune-up limit
(dBm) 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00 13.00

Report No.: CTA24040900105 <Bluetooth Conducted Power>

Mode BLE 1M GFSK /4DQPSK 8DPSK

Channel
0 19 39 0 39 78 0 39 78 0 39 78

Frequency (MHz)
2402 2440 2480 2402 2441 2480 2402 2441 2480 2402 2441 2480

Conducted Peak
Output
Power(dBm)
-2.12 -1.70 -1.02 -2.20 -1.86 -1.10 -3.50 -3.14 -2.38 -3.50 -3.16 -2.35

BLE 1M

Conducted
Average Output
Power(dBm)
-2.45 -2.22 -1.69 -2.76 -2.33 -1.89 -3.91 -3.67 -2.81 -3.85 -3.45 -2.98

Page 27 of 69
Tune-up limit (dBm)
-1.0 -1.0 -1.0 -1.0 -1.0 -1.0 -2.5 -2.5 -2.5 -2.5 -2.5 -2.5

Report No.: CTA24040900105
Transmit Antennas

Page 28 of 69

Bottom

Back View

Antennas WLAN

Distance of The Antenna to the EUT surface and edge

Front

Back

Top Side Bottom Side Left Side

<5mm

<5mm

127mm

0mm

0mm

Right Side 80mm

Report No.: CTA24040900105
SAR Test Exclusion and Estimated SAR

Page 29 of 69

SAR Test Exclusion Considerations Per KDB 447498 D01v06, the 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances  50 mm are determined by: [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)] ·[f(GHz)]  3.0 for 1-g SAR and  7.5 for 10-g extremity SAR  f(GHz) is the RF channel transmit frequency in GHz.  Power and distance are rounded to the nearest mW and mm before calculation.  The result is rounded to one decimal place for comparison.

Per KDB 447498 D01v06, at 100 MHz to 6 GHz and for test separation distances > 50 mm, the SAR test exclusion threshold is determined according to the following: a) [Threshold at 50mm)+(test separation distance-50mm)*(f(MHz)/150)]mW, at 100MHz to 1500MHz b) [Threshold at 50mm)+(test separation distance-50mm)*10]mW at > 1500MHz and  6GHz

Estimated SAR Per KDB447498 requires when the standalone SAR test exclusion of section 4.3.1 is applied to an antenna that transmits simultaneously with other antennas, the standalone SAR must be estimated according to the following to determine simultaneous transmission SAR test exclusion;  (max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)]·[ f(GHz)/x] W/kg for test separation distances  50 mm; where x = 7.5 for 1-g SAR, and x = 18.75 for 10-g SAR.  0.4 W/kg for 1-g SAR and 1.0 W/kg for 10-g SAR, when the test separation distances is > 50 mm

The below table, exemption limits for routine evaluation based on frequency and separation distance was according to SAR-based Exemption ­ §1.1307(b)(3)(i)(B).

Standalone SAR Test Exclusion and Estimated SAR

Wireless Interface

Frequency (MHz)

Configuration

Max. Power

With tune-up

dBm

mW

Distance (mm)

Calculation Result

SAR Exclusion Thresholds

Standalone SAR Exclusion

Estimated SAR
(W/Kg)

Rear Side

13.0

19.95

5

6.25

3

No

N/A

Left edge

13.0

19.95

5

6.25

3

2.4GHz

2450

Right edge

13.0

19.95

127

19.95

866

WLAN

Top edge

13.0

19.95

80

19.95

396

No

N/A

Yes

0.400

Yes

0.400

Bottom edge

13.0

19.95

5

6.25

3

No

N/A

Rear Side

13.0

19.95

5

9.14

3

No

N/A

Left edge

13.0

19.95

5

9.14

3

5.2 GHz

5250

Right edge

13.0

19.95

127

19.95

836

WLAN

Top edge

13.0

19.95

80

19.95

366

Yes

N/A

Yes

0.400

Yes

0.400

Bottom edge

13.0

19.95

5

9.14

3

No

N/A

Rear Side

13.0

19.95

5

9.60

3

No

N/A

5.8 GHz

Left edge

13.0

19.95

5

9.60

3

5785

WLAN

Right edge

13.0

19.95

127

19.95

662

Yes

N/A

Yes

0.400

Top edge

13.0

19.95

80

19.95

712

Yes

0.400

Report No.: CTA24040900105

Page 30 of 69

Bottom edge

13.0

19.95

5

9.60

3

No

N/A

Rear Side

-1.0

0.79

5

0.25

3

Yes

0.03

Left edge

-1.0

0.79

5

0.25

3

Yes

0.03

Bluetooth

2450

Right edge

-1.0

0.79

127

0.79

832

Yes

0.400

Top edge

-1.0

0.79

80

0.79

362

Yes

0.400

Bottom edge

-1.0

0.79

5

0.25

3

Yes

0.03

Remark: 1. Maximum average power including tune-up tolerance; 2. When the minimum test separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR
test exclusion 3. when the distance is < 50 mm exclusion threshold is "Ratio", when the distance is > 50 mm exclusion
threshold is "mW".

Report No.: CTA24040900105

Page 31 of 69

SAR Test Results

General Note:
1 Per KDB 447498 D01v06, the reported SAR is the measured SAR value adjusted for maximum tune-up

tolerance.

a) Tune-up scaling Factor = tune-up limit power (mW) / EUT RF power (mW), where tune-up limit is the

maximum rated power among all production units.

b) For SAR testing of WLAN signal with non-100% duty cycle, the measured SAR is scaled-up by the

duty cycle scaling factor which is equal to "1/ (duty cycle)"

c) For WLAN/Bluetooth: Reported SAR(W/kg)= Measured SAR(W/kg)* Duty Cycle scaling factor * Tune-

up scaling factor

2 Per KDB 447498 D01v06, for each exposure position, testing of other required channels within the

operating mode of a frequency band is not required when the reported 1-g or 10-g SAR for the mid-band

or highest output power channel is:
  0.8 W/kg or 2.0 W/kg, for 1-g or 10-g respectively, when the transmission band is  100 MHz   0.6 W/kg or 1.5 W/kg, for 1-g or 10-g respectively, when the transmission band is between 100 MHz
and 200 MHz   0.4 W/kg or 1.0 W/kg, for 1-g or 10-g respectively, when the transmission band is  200 MHz
3 Per KDB 865664 D01v01r04, for each frequency band, repeated SAR measurement is required only when

the measured SAR is 0.8W/kg.

Report No.: CTA24040900105 <Body SAR>

Page 32 of 69

SAR Values [WIFI 2.4G]

Plot Mode
No.

Test Position

Average Tune-Up

Power Measured Reported

Freq.

Scaling

Ch.

Power Limit

Drift

SAR1g

SAR1g

(MHz)

Factor

(dBm) (dBm)

(dB) (W/kg)

(W/kg)

Measured / Reported SAR numbers-Body distance 0mm

DSSS

Front Side 06 2437 12.69

13.00 1.074 0.02

0.212

0.228

#1

DSSS

Rear Side 06 2437 12.69

13.00 1.074 0.03

0.429

0.461

DSSS

Bottom Side 06 2437 12.69

13.00

1.074 -0.04

0.387

0.416

DSSS

Left Side

06 2437 12.69

13.00 1.074 -0.02

0.401

0.431

Remark: The highest reported SAR for DSSS is adjusted by the ratio of OFDM to DSSS specified maximum output power was

0.625 W/Kg(0.788*(11.5/14.5)=0.625) So ODFM SAR test is not required.

SAR Values [WIFI 5.2G]

Plot Mode
No.

Test Position

Average Tune-Up

Power

Freq.

Scaling

Ch.

Power Limit

Drift

(MHz)

Factor

(dBm) (dBm)

(dB)

Measured / Reported SAR numbers-Body distance 0mm

802.11a(HT20) Front Side 36 5180 12.38

13.00

1.153 -0.03

#2 802.11a(HT20) Rear Side 36 5180 12.38

13.00

1.153 0.02

802.11a(HT20) Bottom Side 36 5180 12.38

13.00

1.153 -0.07

802.11a(HT20) Left Side 36 5180 12.38

13.00

1.153 -0.04

Measured SAR1g (W/kg)
0.201 0.412 0.365 0.385

Reported SAR1g (W/kg)
0.232 0.475 0.421 0.444

SAR Values [WIFI 5.8G]

Plot Mode
No.

Test Position

Average Tune-Up

Power

Freq.

Scaling

Ch.

Power Limit

Drift

(MHz)

Factor

(dBm) (dBm)

(dB)

Measured / Reported SAR numbers-Body distance 0mm

802.11a(HT20) Front Side 149 5745 12.60

13.00

1.096 -0.01

#3 802.11a(HT20) Rear Side 149 5745 12.60

13.00

1.096 0.02

802.11a(HT20) Bottom Side 149 5745 12.60

13.00

1.096 0.03

802.11a(HT20) Left Side 149 5745 12.60

13.00

1.096 -0.05

Measured SAR1g (W/kg)
0.242 0.480 0.432 0.462

Reported SAR1g (W/kg)
0.265 0.526 0.473 0.506

Report No.: CTA24040900105
SAR Measurement Variability

Page 33 of 69

SAR measurement variability must be assessed for each frequency band, which is determined by the SAR

probe calibration point and tissue-equivalent medium used for the device measurements. When both head and body tissue-equivalent media are required for SAR measurements in a frequency band, the variability measurement procedures should be applied to the tissue medium with the highest measured SAR, using

the highest measured SAR configuration for that tissue-equivalent medium. The following procedures are

applied to determine if repeated measurements are required.

1 Repeated measurement is not required when the original highest measured SAR is < 0.80 W/kg; steps

2) through 4) do not apply. 2 When the original highest measured SAR is  0.80 W/kg, repeat that measurement once.

3 Perform a second repeated measurement only if the ratio of largest to smallest SAR for the original and first repeated measurements is > 1.20 or when the original or repeated measurement is 1.45

W/kg (~ 10% from the 1-g SAR limit).

4 Perform a third repeated measurement only if the original, first or second repeated measurement is 

1.5 W/kg and the ratio of largest to smallest SAR for the original, first and second repeated

measurements is > 1.20.

SAR Measurement Variability

Band

Mode

Test Position

First

Second

Original

The

Ch.

Repeated

Repeated SAR

SAR (W/kg)

Ratio

SAR (W/kg)

(W/kg)

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

Report No.: CTA24040900105
Simultaneous Transmission Analysis

Page 34 of 69

Per FCC KD B447498 D01, simultaneous transmission SAR test exclusion may be applied when the sum of the 1-g SAR for all the transmitting antenna in a specific a physical test configuration is 1.6 W/Kg. When the sum is greater than the SAR limit, SAR test exclusion is determined by the SAR to peak location separation ratio.

Ratio=

(SAR1+SAR2 )1.5

 0.04

(peak location separation,mm)

The following procedures adopted from "FCC SAR Considerations for Cell Phones with Multiple Transmitters" are applicable to handsets with built-in unlicensed transmitters such as 802.11 a/b/g/n and Bluetooth devices which may simultaneously transmit with the licensed transmitter.

Application Simultaneous Transmission information: N/A

Evaluation of Simultaneous SAR

The device only support WLAN and BT function and they share the same antenna and cannot transmitting at the same time.

Report No.: CTA24040900105
11 Measurement Uncertainty

Page 35 of 69

NO

Source

Uncert. ai %

Prob. Dist.

Div. k

ci (1g)

ci (10g)

Stand.U ncert. ui
(1g)

Stand.U ncert. ui
(10g)

Veff

1

Repeat

0. 4

N

1

1

1

0. 4

0. 4

9

Instrument

2

Probe calibration

7

N

2

1

1

3.5

3.5



3

Axial isotropy

4.7

R

3 0.7 0.7

1.9

1.9



4 Hemispherical isotropy

9.4

R

3 0.7 0.7

3.9

3.9



5

Boundary effect

1.0

R

3

1

1

0.6

0.6



6

Linearity

4.7

R

3

1

1

2.7

2.7



7

Detection limits

1.0

R

3

1

1

0.6

0.6



8

Readout electronics

0.3

N

1

1

1

0.3

0.3



9

Response time

0.8

R

3

1

1

0.5

0.5



10

Integration time

2.6

R

3

1

1

1.5

1.5



11

Ambient noise

3.0

R

1

1

1.7

1.7



3

12

Ambient reflections

3.0

R

3

1

1

1.7

1.7



13

Probe positioner mech. restrictions

0.4

R

3

1

1

0.2

0.2



Probe positioning with

14

respect to phantom

2.9

shell

R

3

1

1

1.7

1.7



15

Max.SAR evaluation

1.0

R

3

1

1

0.6

0.6



Report No.: CTA24040900105

16

Device positioning

17

Device holder

18

Drift of output power

19

Phantom uncertainty

20 Liquid conductivity (target) 21 Liquid conductivity (meas) 22 Liquid Permittivity (target) 23 Liquid Permittivity (meas)

Combined standard

Expanded uncertainty(P=95%)

Test sample related

3.8

N

1

1

1

5.1

N

1

1

1

5.0

R

3

1

1

Phantom and set-up

4.0

R

3

1

1

5.0

R

3 0.64 0.43

2.5

N

1 0.64 0.43

5.0

R

3 0.6 0.49

2.5

N

1 0.6 0.49

RSS

UC =

n
 Ci 2U i 2
i =1

U = kU

,k=2

C

3.8 5.1 2.9
2.3 1.8 1.6 1.7 1.5 11.4%
22.8%

Page 36 of 69

3.8

99

5.1

5

2.9



2.3



1.2



1.2



1.5



1.2



11.3%

236

22.6%

Report No.: CTA24040900105
Appendix A. EUT Photos and Test Setup Photos
Please refer to separated files for Test Setup Photos of the SAR.

Page 37 of 69

Report No.: CTA24040900105
Appendix B. Plots of SAR System Check
2450MHz System Check

Page 38 of 69
Date: 05/21/2024

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: 745 Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2450 MHz;  = 1.775 S/m; r = 38.809;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(7.50, 7.50, 7.50) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)

Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 20.9 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 83.22 V/m; Power Drift = 0.09 dB Peak SAR (extrapolated) = 33.47 W/kg

SAR(1 g) = 12.90 W/kg; SAR(10 g) = 6.52 W/kg Maximum value of SAR (measured) = 20.5 W/kg

0 dB = 20.5 W/kg System Performance Check 2450MHz 250mW

Report No.: CTA24040900105
5250MHz System Check

Page 39 of 69
Date: 05/22/2024

DUT: Dipole 5GHz; Type: D5GHzV2; Serial: 1031 Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1 Medium parameters used (interpolated): f = 5250 MHz;  = 4.670 S/m; r = 35.355;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(5.45, 5.45, 5.45) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)

Area Scan (101x101x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm Maximum value of SAR (interpolated) = 23.2 W/kg

Zoom Scan (7x7x13): Measurement grid: dx=4mm, dy=4mm, dz=2mm Reference Value = 26.39V/m; Power Drift = -0.06 dB Peak SAR (extrapolated) = 48.40 W/kg

SAR(1 g) = 7.89 W/kg; SAR(10 g) = 2.28 W/kg Maximum value of SAR (measured) = 19.30 W/kg

0 dB = 19.20 W/kg System Performance Check 5250MHz 100mW

Report No.: CTA24040900105
5750MHz System Check

Page 40 of 69
Date: 05/22/2024

DUT: Dipole 5GHz; Type: D5GHzV2; Serial: 1102 Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1 Medium parameters used (interpolated): f = 5750 MHz;  =5.080 S/m; r = 36.190;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(4.96, 4.96, 4.96) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)

Area Scan (101x101x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm Maximum value of SAR (interpolated) = 13.78 W/kg

Zoom Scan (7x7x13): Measurement grid: dx=4mm, dy=4mm, dz=2mm Reference Value = 19.88 V/m; Power Drift = -0.07 dB Peak SAR (extrapolated) = 33.12 W/kg

SAR(1 g) = 7.99 W/kg; SAR(10 g) = 2.30 W/kg Maximum value of SAR (measured) = 13.70 W/kg

0 dB = 13.70 W/kg System Performance Check 5750MHz 100mW

Report No.: CTA24040900105
Appendix C. Plots of SAR Test Data
#1 Date: 05/21/2024

Page 41 of 69

WIFI2.4G_DSSS_Rear side_0mm_Ch01

Communication System: UID 0, Generic WIFI (0); Frequency: 2437 MHz;Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2437 MHz;  = 1.780 S/m; r =38.810;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(7.50, 7.50, 7.50,) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)
Front /Area Scan (91x91x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.500 W/kg

Front /Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 11.22 V/m; Power Drift = -0.03 dB Peak SAR (extrapolated) = 0.612 W/kg

SAR(1 g) = 0.429 W/kg; SAR(10 g) = 0.257 W/kg Maximum value of SAR (measured) = 0.452 W/kg

Report No.: CTA24040900105 #2 Date: 05/22/2024

Page 42 of 69

WLAN 5.2GHz_802.11ac(HT40)_Rear side_0mm_CH38

Communication System: UID 0, Generic WLAN (0); Frequency: 5190 MHz;Duty Cycle: 1:1 Medium parameters used (interpolated): f = 5190 MHz;  =4.651 S/m; r = 36.433;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(5.45, 5.45, 5.45) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)

Area Scan (61x61x1): Measurement grid: dx=1.000mm, dy=1.000mm Maximum value of SAR (interpolated) = 0.480 W/Kg

Zoom Scan (8x8x21): Measurement grid: dx=4mm, dy=4mm, dz=1.4mm Reference Value = 0.32V/m; Power Drift = 0.02 dB Peak SAR (extrapolated) = 0.660 W/kg

SAR(1 g) = 0.412 W/kg; SAR(10 g) = 0.227 W/kg Maximum value of SAR (measured) = 0.430 W/kg

Report No.: CTA24040900105 #3 Date: 05/22/2024

Page 43 of 69

WLAN 5.8GHz_802.11ac(HT20)_Rear side_0mm_CH149
Communication System: UID 0, Generic WLAN (0); Frequency: 5745 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 5745 MHz;  = 5.354 S/m; r = 34.621;  = 1000 kg/m3 Phantom section: Flat Section

DASY5 Configuration: · Probe: EX3DV4 - SN7380; ConvF(4.96, 4.96, 4.96) ; Calibrated: 6/21/2023 · Sensor-Surface: 1.4mm (Mechanical Surface Detection) · Electronics: DAE3 Sn428; Calibrated: 08/30/2023 · Phantom: Twin-SAM V8.0 ; Type: QD 000 P41 AA; Serial: 1974 · DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)

Area Scan (61x61x1): Measurement grid: dx=1.000mm, dy=1.000mm Maximum value of SAR (interpolated) = 0.510 W/Kg

Zoom Scan (8x8x21): Measurement grid: dx=4mm, dy=4mm, dz=1.4mm Reference Value = 0.53 V/m; Power Drift = 0.02 dB Peak SAR (extrapolated) = 0.689 W/kg

SAR(1 g) = 0.450 W/kg; SAR(10 g) = 0.286 W/kg Maximum value of SAR (measured) = 0.470W/kg

Report No.: CTA24040900105
Appendix D. DASY System Calibration Certificate

Page 44 of 69

Report No.: CTA24040900105

Page 45 of 69

Report No.: CTA24040900105

Page 46 of 69

Report No.: CTA24040900105

Page 47 of 69

Report No.: CTA24040900105

Page 48 of 69

Report No.: CTA24040900105

Page 49 of 69

Report No.: CTA24040900105

Page 50 of 69

Report No.: CTA24040900105

Page 51 of 69

Report No.: CTA24040900105

Page 52 of 69

Report No.: CTA24040900105

Page 53 of 69

Report No.: CTA24040900105

Page 54 of 69

Report No.: CTA24040900105

Page 55 of 69

Report No.: CTA24040900105

Page 56 of 69

Report No.: CTA24040900105

Page 57 of 69

Report No.: CTA24040900105

Page 58 of 69

Report No.: CTA24040900105

Page 59 of 69

Report No.: CTA24040900105

Page 60 of 69

Report No.: CTA24040900105

Page 61 of 69


Microsoft Word 2016 Microsoft Word 2016