User Guide Terrameter LS

torleif

ABEM Terraloc Pro User manual

Resource Center - Guideline Geo

PDF preview unavailable. Download the PDF instead.

ABEM-Terraloc-Pro-User-Manual
User Manual
Terraloc Pro
ABEM Product Number 33 700 89 ABEM 20161209, based on release 2.2.8 of SeisTW

ABEM Terraloc Pro
Thank you for choosing ABEM Terraloc Pro

Trademarks
Terraloc® is a registered trademark of ABEM Instrument AB. Microsoft® and Windows® are registered trademarks of Microsoft Corporation. All other trademarks belong to their respective holder.
General information
Information in this manual is subject to change without notice and constitutes no commitment by ABEM Instrument AB.
ABEM Instrument AB takes no responsibility for errors in this manual or problems that may arise from the use of this material.
In general, e-mail correspondence gives the fastest response. In view of our policy of progressive development, we reserve the right to alter specifications without prior notice.

ABEM will be pleased to receive occasional reports from you concerning the use and experience of the equipment. We also welcome your comments on the contents and usefulness of this manual. In all communication with ABEM be sure to include the instrument types and serial numbers.

Contact details:

Address:

ABEM Instrument AB Löfströms Allé 6A

SE-17266 Sundbyberg

Sweden

Phone:

+46 8 564 88 300

Fax:

+46 8 28 11 09

Web site: www.guidelinegeo.com

Email:

sales@guidelinegeo.com support@guidelinegeo.com

© Copyright 2011 ABEM Instrument AB. All rights reserved.

ABEM Terraloc Pro

Table of Contents

Section

Page

About This Manual .....................................................................................................iii

1 Get ready - Unpacking your new Terraloc® Pro ............................................... 5 1.1 Welcome To Refraction, Reflection And Tomography.................................5

1.2 Features of the ABEM Terraloc Pro .............................................................. 5

1.3 The Delivered Instrument .............................................................................. 6

1.4 Inspection ....................................................................................................... 7

1.5 Shipping Damage Claims .............................................................................. 7

1.6 Shipping/Repacking instructions ................................................................... 8

1.7 Registration .................................................................................................... 8

1.8 Take Time to Read The Technical Documentation ....................................... 8

1.9 Software ......................................................................................................... 8

2 Overview of the Instrument ................................................................................ 9 2.1 The Connector Panel ...................................................................................... 9

2.2 The Power Panel .......................................................................................... 11

2.3 The Built-in GPS Receiver .......................................................................... 12

2.4 The User Interface Panel..............................................................................12

2.5 The Power Supply........................................................................................13

2.6 Interconnecting Two or More Instruments .................................................. 13

3 Quick Start ......................................................................................................... 14

4 The User Interface ............................................................................................. 16 4.1 The Display..................................................................................................16

4.2 Keyboard and Mouse ................................................................................... 17

4.3 Using SeisTW .............................................................................................. 19

4.4 SeisTW Layout Parts ................................................................................... 23

4.5 Menus...........................................................................................................33

4.6 Dialogs ......................................................................................................... 37

5 Data Processing .................................................................................................. 69 5.1 Unfilter Data ................................................................................................ 69

5.2 First Breaks .................................................................................................. 69

5.3 FIR Filter......................................................................................................70

5.4 Cross Correlate.............................................................................................71

5.5 Moving Average .......................................................................................... 73

i

ABEM Terraloc Pro
6 Triggering Methods ........................................................................................... 75 6.1 Make/Break Switch Input ............................................................................ 75 6.2 Using the Trigger Coil ................................................................................. 75 6.3 Radio Triggering .......................................................................................... 75
7 Measurement ...................................................................................................... 76 7.1 Basic Operations .......................................................................................... 76 7.2 Data Transfer ............................................................................................... 77 7.3 Optimizing ................................................................................................... 78
8 Troubleshotting and Diagnostics ...................................................................... 79 8.1 General SeisTW Program Problems ............................................................ 79 8.2 Data Acquisition Problems .......................................................................... 79 8.3 Trigger Problems ......................................................................................... 80 8.4 Remote Diagnostics (VPN)..........................................................................80 8.5 In Case of Malfunction ................................................................................ 83
9 Appendix A. Technical Specification ............................................................... 84 10 Appendix B. Connectors................................................................................86
10.1 Seismic Input Connectors ............................................................................ 86 10.2 Power Connector..........................................................................................88 10.3 TTL Arm/Trig Connector ............................................................................ 88 10.4 Alarm Connector..........................................................................................88 10.5 Cascade Connector.......................................................................................89 11 Appendix C. SeisTW Installation ................................................................. 90 11.1 Install Procedure for SeisTW.......................................................................90 12 Appendix D. Printout Examples ................................................................... 93 13 Appendix E. The First Arrivals File Format (FIR) .................................... 95 13.1 General ......................................................................................................... 95 13.2 Description ................................................................................................... 95 14 Appendix F. Seismic Methods.......................................................................96 14.1 Refraction.....................................................................................................96 14.2 Reflection ..................................................................................................... 97 14.3 Optimum Offset ........................................................................................... 97 14.4 Tomography.................................................................................................97 14.5 VSP .............................................................................................................. 97 14.6 Vibroseis ...................................................................................................... 98 15 Appendix G. Bibliography ............................................................................ 99
ii

ABEM Terraloc Pro

About This Manual

The conventions and formats of this manual are described in the following paragraphs:
 Typographical conventions used in this manual:

Italic Bold Blue Italic

Names of objects, figure descriptions In-line minor headers, emphasis URL links

 Formats used in this manual for highlighting special messages:
 Use of the internal keyboard is given in this format  A sequence of steps will have two or more of these parts
Further information about this particular usage is given like this

Note!

This format is used to highlight information of importance or special interest

Warning! Ignoring this type of notes might lead to loss of data or a malfunction

These notes warn for things that can lead to people or animals getting hurt or to equipment getting damaged

iii

ABEM Terraloc Pro iv

ABEM Terraloc Pro
1 Get ready - Unpacking your new Terraloc® Pro
1.1 Welcome To Refraction, Reflection And Tomography Welcome to the ABEM Terraloc®Pro, the multi-channel digital seismograph for cost-effective refraction and high-resolution reflection surveys, tomography, vibration measurements, and more, anywhere in the world in all weather conditions.
The basic Terraloc Pro is a self-contained multi-channel seismograph with internal PC-compatible computer, a hard disk and a daylight visible 8.4 " TFT color display with SVGA resolution. Operating power comes from an internal battery, or any external battery pack or power source that delivers from 10 - 30 volts DC. Typically this means a re-chargeable battery pack, a car (or truck) battery, or AC/DC power supply (office power supply unit). The inbuilt battery charger charges the internal battery pack when an external power source is connected.
The Terraloc Pro has a hard disk with a size of at least 100 GB. It also has 3 USB 2.0 ports, an Ethernet port and a VGA monitor port.
The physical dimensions are the same for all models, 12 ­ 48 Channels.
After a survey you may process data stored on the internal hard disk using Terraloc Pro internal PC or an external computer. Large amounts of data can be transferred between the Terraloc Pro and an external PC using the built in Ethernet port in the Terraloc Pro. For filtering and basic processing you can use the Terraloc Pro internal software called SeisTW, which is the software that controls the functions of the Terraloc Pro. Third party software packages for seismic data processing can be run directly on the Terraloc Pro. Please ask your authorized ABEM Distributor for details about the seismic interpretation and processing packages that are available.
Your Terraloc Pro was carefully checked at all stages of production. It was thoroughly tested before being approved for delivery. If you handle and maintain it according to the instructions in the technical documentation, you will get many years of satisfactory service from it.
1.2 Features of the ABEM Terraloc Pro Examples of features of the ABEM Terraloc Pro are:
 SeisTW for Windows XP, ABEM developed measurement software (Included and factory installed)
 3 USB ports for connecting external accessories such as USB CD/DVD, USB memory sticks, keyboard, mouse, card reader etc.
 Ethernet port for fast transfers of data and networking capabilities  Daylight visible color 8.4" TFT SVGA display  Excellent resolution thanks to a 24 bit ADC (analog/digital converter)  In-field quality control of measurements thanks to geophone tests, noise monitoring, and a wide
choice of single- or multi-trace view modes  Excellent results for tomography and high resolution seismic thanks to selectable sampling rates
from 25 µs to 2 ms in seven steps  Full on-screen display of recorded traces with software roll-along, automatic pick of first arrivals,
list of first arrival times, velocity calculation, frequency analysis of single traces.
5

ABEM Terraloc Pro
1.3 The Delivered Instrument Your Terraloc Pro arrives in a wooden transport box. Open it and unpack all items carefully. Check the contents of the box or crate against the packing list. If you ordered optional equipment, check the invoice/packing list for details and compare with your original order. A standard ABEM Terraloc Pro system includes the following (Figure 1): 1 Terraloc Pro field unit with a number of channels as shown on the packing list 1 External power cable with connector and crocodile clips,
ABEM part no. 33 3000 42 1 Internal battery pack, ABEM part no. 33 3000 77 1 Office power supply unit, ABEM part no. 39 0450 08 1 Cable for office power supply, ABEM part no. 33 7000 58 1 External USB-Keyboard-Mouse Kit, ABEM part no. 33 5000 35 1 Trigger cable 250m on reel, ABEM part no. 33 0011 25, (packed in own box) 1 Terraloc Pro Accessories & Tools kit, ABEM part no.33001193 (small carton box)
comprising:
2 2 m connection cables (for trigger coil) ABEM part no. 39 7101 04 1 Insulating tape roll 1 Engineer pliers 1 Pair of cutting pliers 1 Torx key T-20 1 Torx key T-25 1 Philips No.1 Screwdriver 1 Trigger coil, ABEM part no. 33 0011 26 1 LAN cable RJ45 connectors 5m (for Ethernet), ABEM no. 39 7101 69
1 Terraloc Pro Documentation kit ABEM part no. 33 5000 93, comprising:
1 Terraloc Pro Instruction manual 1 USB memory stick for software recovery 1 Warranty registration card
6

ABEM Terraloc Pro

Transport Crate

Terraloc Pro field unit
External Power Cable LAN Network Cable, 5 m
Connection Cables, 2 m Documentation kit: -User Manual -CD -Warranty Registration Card
Philips Screwdriver
Figure 1

Trigger Cable 250 m on reel

Office Power Supply Unit
Insulating tape 12 V NiMH Battery Pack
Keyboard and mouse kit

Cable for Office Power Supply

Software on USB Memory Stick

Torx keys T20 and T25

Trigger Coil

Pliers

Standard Terraloc Pro system

1.4 Inspection Inspect the instrument and accessories for loose connections and inspect the instrument case for any damage that may have occurred due to rough handling during shipment.
The instrument is delivered in a reusable plywood box. The box is designed to offer a convenient and safe transport option. All packing materials should be carefully preserved for future re-shipment, should this become necessary. Always make sure to use the transport box provided, or an alternative of at least equivalent mechanical protection and shock absorption whenever the instrument is shipped.
1.5 Shipping Damage Claims File any claim for shipping damage with the carrier immediately after discovery of the damage and before the equipment is put into use. Forward a full report to ABEM, making certain to include the ABEM delivery number, instrument type(s) and serial
7

ABEM Terraloc Pro
number(s). If it is a question of short shipment you must make a claim in writing to ABEM within 14 days of your receipt of shipment.
1.6 Shipping/Repacking instructions The ABEM packing kit is specially designed for the Terraloc Pro. The packing kit should be used whenever shipping is necessary. If original packing materials are unavailable, pack the instrument in a wooden box that is large enough to allow some 80 mm of shock absorbing material to be placed all around the instrument. This includes top, bottom and all sides. Never use shredded fibers, paper or wood wool, as these materials tend to pack down and permit the instrument to move inside its packing box. Please read our shipping instructions before returning instruments to ABEM. The instructions can be found on our website. For further assistance please contact ABEM or its authorized distributor. Contact information can be found in the beginning of this document.
1.7 Registration When you have checked the packing list, the next important thing to do is to register your Terraloc Pro. To register send an email with your contact information to support@guidelinegeo.com. Once registered, you will able to receive software updates and product information.
1.8 Take Time to Read The Technical Documentation To ensure you get optimum results with the ABEM Terraloc Pro, please take time to read this instruction manual thoroughly. If you should, for any reason, have difficulties in operating ABEM Terraloc Pro or in getting satisfactory seismic survey results, please contact your authorized ABEM distributor. ABEM always listens to end-user comments about their experience with ABEM products. So please send occasional reports on field usage as well as your ideas on how the Terraloc Pro and its technical documentation can be improved to help you do an even better job of seismic surveying.
1.9 Software Terraloc Pro is delivered with all necessary software installed at the factory. If the software needs to be updated, or re-installed, the procedure is described in 11 Appendix C. SeisTW Installation. What is SeisTW? SeisTW (Seismograph Terraloc Windows) is a Windows XP application that is used to control the Terraloc Pro. It can also be installed on any PC running Windows XP and used to view and manage seismic records. However, when installed on a PC all functions accessing the Terraloc Pro hardware will be disabled. SeisTW is included and factory installed in all Terraloc Pro instruments.
8

ABEM Terraloc Pro

2 Overview of the Instrument

2.1 The Connector Panel
All connectors except for the external power are situated on the right side panel of the Terraloc Pro (Figure 2). Some of the connectors are described in more detail in chapter 10 Appendix B. Connectors.

Note!

Always have the connector protection dust caps in place whenever a connector is not used

Figure 2 The Connector panel 9

ABEM Terraloc Pro

The connectors:

Label

Function

A

Ethernet

B

USB 1

C

USB 2

D

USB 3

E

VGA

F

Cascade

G

TTL Trig/Arm: To connect two or more Terraloc Pro as Master and

Slave(s), for radio shot, and vibrator hand-shaking.

Mating connector: see 10.3 TTL Arm/Trig Connector

H

Alarm: This connector can activate alarm units

I

Trigger input: for a trigger geophone shot instant contacts, a wire loop

around the explosive charge, or trigger output from a mechanical energy

source.

Mating connectors: 4 mm banana plug or bare wire

J

Reference channel 2: (up hole channel). Connector for a single geophone

or vibrator reference (signature).

Mating connectors: 4 mm banana plug or bare wire

K

Reference channel 1: (up hole channel). Connector for a single geophone

or vibrator reference (signature).

Mating connectors: 4 mm banana plug or bare wire

L

Signal: for connecting geophone spread cables to channel 13-24 (24-

channel) or 25-48 (48-channel).

The connector is wired to industry standard.

For wiring and mating connector: see

10.1.1 12 and 24 Channel Terraloc Pro and

10.1.2 48 Channel Terraloc Pro

M

Signal: for connecting geophone spread cables to channel 1-12 (24-

channel) or 1-24 (48-channel).

The connector is wired to industry standard.

For wiring and mating connector: see

10.1.1 12 and 24 Channel Terraloc Pro and

10.1.2 48 Channel Terraloc Pro

10

ABEM Terraloc Pro
2.2 The Power Panel The power panel of the Terraloc Pro is shown in Figure 3. The Power Input connector is described in more detail in chapter 10 Appendix B. Connectors.

Figure 3 The Power panel

The connectors:

Label

Function

N

Power Input: for connecting an external power source.

For wiring and mating connector see chapter 10.2 and for specifications

see chapter 1

Use External Power cable with clips for a car battery, or

Office power supply unit with Cable for office power supply

O

Internal battery lid

11

ABEM Terraloc Pro
2.3 The Built-in GPS Receiver Terraloc Pro has a built-in GPS receiver (Figure 4). In order to function well the builtin antenna in the handle of the instrument must be able to receive signals from a sufficient number of satellites. This will normally not function indoors and in outdoor areas with limited viewing angle towards the sky the function can be limited, for example in a forest. Positioning data is automatically saved in the header of the current record. The GPS receiver status is shown on the display (see chapter 4.4.8 Application Status Bar).

Figure 4 The GPS antenna is integrated in the left side of the handle

2.4 The User Interface Panel
All interaction with the Terraloc Pro is done through the user interface panel. Figure 5 points out the parts of the user interface panel.

Colour Display

Built-in Keyboard

LED Window

Figure 5 The user interface panel There are two LED's shown through the LED Window:
- The green LED indicates disk activity - The yellow LED indicates if sampling is on or off
12

ABEM Terraloc Pro
2.5 The Power Supply The Terraloc Pro can use an external power source as well as an internal battery as power supply. The external source can be a battery or a PSU (Power Supply Unit). If possible use the supplied cable set for the external power source. Both external and internal power sources can be attached at the same time. In this case the internal battery will be charged if the external battery is charged enough. The power supply status is shown on the display (see chapter 4.4.8 Application Status Bar).
For field operations a good, adequate in capacity and recently charged battery is vital for the best performance. It is possible to fully run the Terraloc Pro without the internal battery but for your convenience you should always have one installed.
The internal battery is primarily designed as a backup power source for operating the instrument during set up, data transfer etc, hence it cannot be used alone to power the instrument for a days work. It has quite a snug fit in the battery compartment. If the protective liner that keeps the cells together is found defective during inspection, please contact ABEM support for further information.
Once the instrument has been turned on and the external battery for any reason is disconnected the instrument will automatically switch to the internal battery. This useful feature makes it possible to disconnect the external battery temporarily without shutting off the instrument when for instance moving from one place to another.
2.6 Interconnecting Two or More Instruments Should more channels be needed than can be supplied by the use of a single instrument, it is possible to connect (virtually) any number of Terraloc Pro instruments. The Arm, Disarm, and Trigger events can be synchronized with interconnected instruments. The TTL Arm/Trig connector is used to connect the instruments, see chapter 10.3.
Figure 6 shows an example from a survey where four Terraloc Mk6 were used to comprise a 96-channel system. The same can be done with Terraloc Pro instruments.

Figure 6

96-channel record, made using four interconnected Terraloc Mk6 13

ABEM Terraloc Pro

3 Quick Start

In this section we will make a measurement of noise. It will give you an insight to how easy it is to set your Terraloc Pro up for operation. You will need no more equipment than the instrument itself and the power supply. However, before starting any fieldwork it is wise to invest time to go through and familiarize yourself with the various menus, dialogs and options that exist. These are described in detail in the following chapters. Should you feel uncertain during any of the steps below you can press <HLP> to get access to the help screen for explanations about which key command does what.
Now follow these steps:
 Connect the power supply (see Figure 3 connector N) and switch on the instrument by pressing <POWER>
 Some diagnostic messages show up on the screen during the start up tests and then Windows XP is started
 SeisTW starts automatically
 Press <ARM> to create a new acquisition record using the last active acquisition mode. To verify/change the acquisition settings press:
<1> for Acquisition setup
<2> for Trig setup
<3> for Noise monitor
<4> for Acquisition (analog) filters
<5> for Receiver spread
<6> for Layout geometry
<7> for Header information (job ID, line ID, notes, etc.)

Note!

The built-in keyboard cannot be used to input text. For this an external USB-keyboard is needed

<9> for View options (trace style, time compression, scale factor)
 Now press <ARM> again. This arms the instrument and makes it ready to trigger and record a trace. The status bar (at the bottom of the screen) displays the message "<<<ARMED>>>"
 Press <CTR> + <ARM> to force the instrument to trig. The message "<<< TRIGGERED >>>" is displayed in the status bar, shortly followed by "Transferring data...", "Data in memory" and then "<<<ARMED>>>".
 The recorded data is displayed in the three frames at the center of the screen. To change view options, press <9>
 Trigging once more by pressing <CTR> + <ARM> will replace the traces on the screen with a new set that looks a little bit different. What you see now is the average of the two measurements made so far

14

ABEM Terraloc Pro  Press <SAVE> to save the data (the message "No data" will be displayed) or
press <ESC> to disarm the instrument (the message "Data in memory" will be displayed)  When you are finished getting acquainted with the instrument, you may shut it down. Press <CTR> + <SPACE> for the quick menu and select "Power Off" among the menu items. Press <ENT> when the confirming dialog appears  Now you should have learned a little about how to operate the instrument. Do not be afraid to test different settings and modes. There is no risk of causing any damage. Should you somehow get problems with the Terraloc Pro software SeisTW, it can be reinstalled (see 11 Appendix C. SeisTW Installation)
15

ABEM Terraloc Pro
4 The User Interface
The user interacts with the instrument through the User Interface Panel and possibly connected USB input devices. This chapter explains the basics of this interaction. 4.1 The Display SeisTW will normally be shown on the display. Figure 7 shows a normal start-up view of the SeisTW window.
Figure 7 The SeisTW main window For more information about the layout parts of SeisTW please see chapter 4.4.
16

ABEM Terraloc Pro

4.2 Keyboard and Mouse Commands from the user are entered through a keyboard and/or a mouse. There is a built-in keyboard (see Figure 5) but an external USB keyboard can also be used and as well an external USB mouse.
4.2.1 The Built-in Keyboard Table 1 lists the names of the buttons as referenced in this document.

<1>

<2>

<3>

<POWER>

<4>

<5>

<6>

<7>

<8>

<9>

<.>

<BACKSPACE>
<LEFT>

<0> <UP> <TAB>

<-> <RIGHT>

<ESC> <ARM> <SAVE>

<SHIFT>

<DOWN>

<+>

<PRN>

<CTR>

<SPACE>

<HLP>

Table 1 Names used for the built-in keyboard buttons

<ENTER>

Note!

Where <ARROWS> is used in the text it means all four arrow keys (up, down, left and right) Where <NUMBERS> is used in the text it means all numerical keys (0-9)

Note!

The <UP> and <SHIFT> keys are similar in appearance but the arrow of the <SHIFT> key is wider

17

ABEM Terraloc Pro

4.2.2 An External Keyboard
A standard USB computer keyboard can be connected to one of the USB ports of the Terraloc Pro and used as a complement to the built-in keyboard. The mapping between the built-in buttons and the computer keyboard is listed in Table 2.

Note!

The only way to enter and edit text is to use an external keyboard

1

2

3

(none)

4

5

6

7

8

9

.

0

Backspace

Up

Left

Tab

Esc

-

F2

Right

F3

Shift

Down

+

F4

Ctrl

Space

F1

Enter

Table 2 Mapping between built-in keyboard and external keyboard
4.2.3 An External Mouse A standard USB mouse can be connected to one of the USB ports of the Terraloc Pro and used as a normal mouse in the Windows XP environment.

18

ABEM Terraloc Pro 4.3 Using SeisTW SeisTW is a normal Windows program and using the program with external keyboard and mouse is like using any other Windows program. However using the built-in keyboard naturally brings with it some limitations. Some measures have been taken within SeisTW to remedy this and the rest of this chapter explains some of the more general of these measures. More information about the use of the built-in keyboard can be found in the chapters that describe the various functions of SeisTW. Please see Figure 13 on page 23 for a descriptive overview of the layout of SeisTW.  Highlighting different views (Record View ­ Trace View ­ Frequency View). This
is useful for working with the different views  Press <TAB> to highlight the next view  Press <SHIFT> + <TAB> to highlight the previous view
Figure 8 SeisTW with the Trace View highlighted  Changing the sizes of the views. That is, moving the separators between the views
(Figure 9 and Figure 10)  Press <CTR> + <UP> to move the horizontal separator upwards  Press <CTR> + <DOWN> to move the horizontal separator downwards  Press <CTR> + <LEFT> to move the vertical separator to the left  Press <CTR> + <UP> to move the vertical separator to the right
19

ABEM Terraloc Pro
Figure 9 SeisTW with the horizontal separator moved upwards
Figure 10 SeisTW with the vertical separator moved to the right  Hiding the Trace and Frequency Views. The Record View will enlarge to cover the
hidden area  Press <SHIFT> + <0> to alternately hide and show the two views 20

ABEM Terraloc Pro  Showing or hiding the Logging Window.
 Press <SHIFT> + <SPACE> to alternately hide and show the Logging Window
Figure 11 The Logging Window on the left side  Opening and stepping through Menu Bar items
 Press <CTR> + <BACKSPACE> to set focus on the Menu Bar  Press <DOWN> to open the File menu list  Press <DOWN> or <UP> to highlight a menu item  Press <LEFT> or <RIGHT> to open another top level menu list  Press <ENT> to execute the highlighted menu item Or, if a record is opened or created  Press <SPACE> two times (the first opens the Context Menu and the second
opens the System Menu)  Press <RIGHT> and the File menu list is opened  Press <ARROWS> as described above to select the wanted menu item  Navigating between input fields on dialogs  Press <TAB> to highlight the next input field  Press <SHIFT> + <TAB> to highlight the previous input field
21

ABEM Terraloc Pro
 Changing settings on dialogs. The way to change a setting depends on the type of input field. See Figure 12 for examples of input field types  Drop-down list (see Trig input mode): Press <PRN> to open the list Press <PRN> again to close the list Press <UP> or <DOWN> to change the value  Track-bar (see Trig input level; Trig input mode must be Analog or Channel): Press <LEFT> or <RIGHT> to change the value  Check-box (see Ext. arm verify): Press <SHIFT> to change the value  Up-down field (see Verify timeout [ms]): Press <UP> to increment the value with 1 Press <DOWN> to decrement the value with 1 Press the <NUMBERS> keys to directly enter digits Press <BACKSPACE> to delete the digit before the input marker
Drop-down list
Track bar
Check box Up-down field
Figure 12 Part of Trig setup dialog as input field example
 Closing an opened dialog  Press <ENT> to close the dialog and save possible changes Or  Press <ESC> to close the dialog without saving possible changes
22

ABEM Terraloc Pro 4.4 SeisTW Layout Parts
Figure 13 The SeisTW layout The purpose and specific functions of each layout part will be described below. 4.4.1 Title Bar The Title Bar displays the application name and version. It will also display the file name of an open record. 4.4.2 Menu Bar The Menu Bar presents the main menu items to the user. Some of the displayed short cuts on the menu items are only applicable to an external keyboard. 4.4.3 Tool Bar The Tool Bar presents the user with some buttons for actions that can be performed.  Hide or show the Tool Bar.
 Press <SHIFT> + <1> to alternately hide or show the Tool Bar Hiding the tool bar will free more of the screen area for displaying data. 4.4.4 Record View The Record View shows all traces vertically. A time scale is displayed on the left side. This timescale adjusts according to sample interval and view options. Tic lines across the screen (Figure 14) can be enabled in the View options dialog (see chapter 4.6.13).
23

ABEM Terraloc Pro
Figure 14 The Record View; Left: without Tic lines Right: with Tic lines At the top of the view there is a trace marker. This marker points out the current trace, which is the trace that is shown in the Trace and Frequency Views (Figure 15).
Figure 15 Trace Marker; Left: For an opened record file Right: For a new record  Moving the Trace Marker between traces
 Press <LEFT> to move the marker to the previous trace or from the first to the last trace (wrap around)
 Press <RIGHT> to move the marker to the next trace or from the last to the first trace (wrap around)
 Press <SHIFT> + <LEFT> to move the marker to the first trace  Press <SHIFT> + <RIGHT> to move the marker to the last trace When a record has been created the top of the view also displays the current Stack On status, and polarity. The Stack On is displayed by squares above each trace (Figure 16). If the square is filled the stack for that trace is on, and if the square is open, the same stack is off (see chapter 4.6.5.1 for information about the stack function). If negative polarity has been selected for a trace, a minus sign is displayed under the square (Figure 17).
Figure 16 Stack On Status; Traces 1 and 3 are off
Figure 17 Negative Polarity; Traces 1 and 3 have negative polarity 24

ABEM Terraloc Pro
 Scrolling the view  Press <UP> to scroll the view upwards  Press <DOWN> to scroll the view downwards  Press <SHIFT> + <UP> to scroll the view upwards a whole page  Press <SHIFT> + <DOWN> to scroll the view downwards a whole page
A timeline can be moved across the view. The time and A/D-value for the current trace and timeline position will be displayed in the status field just below the views. The timeline can be used to position a first break marker at the location of the timeline on the current trace.
 Moving a timeline across the view (Figure 18)  Press <+> to move the timeline downwards  Press <-> to move the timeline upwards  Press <SHIFT> + <+> to move the timeline downwards with a large step  Press <SHIFT> + <-> to move the timeline upwards with a large step

Note!

Figure 18 The red timeline
Keeping the key pressed will accelerate the movement of the timeline

 Positioning a first break marker (Figure 19)  Press <.> to position a first break marker. The marker will be positioned on the current trace. A similar marker is also positioned in the Trace View
25

ABEM Terraloc Pro
Figure 19 First break marker  Positioning a first break marker on trace 2 (Figure 20)
 Press <RIGHT> to select trace 2  Press <.> to position a first break marker
Figure 20 First break marker on trace 2  Removing an existing first break marker
 Select the wanted trace by pressing <LEFT> and/or <RIGHT>  Press and hold <-> until the timeline is invisible  Press <.> to remove the first break marker 4.4.5 Trace View The trace view displays an enlarged view of the current trace and its frequency content.  Change the trace to view  Press <UP> to change to the next trace
26

ABEM Terraloc Pro
 Press <DOWN> to change to the previous trace
 Scrolling the view  Press <LEFT> to scroll the view to the left  Press <RIGHT> to scroll the view to the right  Press <SHIFT> + < LEFT > to scroll the view to the left a whole page  Press <SHIFT> + < RIGHT > to scroll the view to the right a whole page
A timeline can be moved across the view. The time and A/D-value for the current trace and timeline position will be displayed in the status field just below the views. The timeline can be used to position a first break marker at the location of the timeline on the current trace.
 Moving a timeline across the view (Figure 21)  Press <+> to move the timeline to the right  Press <-> to move the timeline to the left  Press <SHIFT> + <+> to move the timeline to the right with a large step  Press <SHIFT> + <-> to move the timeline to the left with a large step

Note!

Figure 21 The red timeline
Keeping the key pressed will accelerate the movement of the timeline

A reference time marker can be positioned at the location of the time line. If the time line is moved when the reference time marker is active, the status bar will display, in
27

ABEM Terraloc Pro addition to the normal information, the relative time and the corresponding frequency (i.e. reciprocal time).  Position a reference time marker (Figure 22)
 Press <0> to position a reference time marker
Figure 22 The red dotted reference time marker  Move the timeline and show relative time (Figure 23)
 Press <+> to move the timeline to the right
Figure 23 A reference time marker with timeline  Removing an existing reference time marker
 Press <-> until the timeline is invisible  Press <0> to remove the reference time marker  Positioning a first break marker (Figure 24)  Press <.> to position a first break marker
Figure 24 First break marker (timeline moved on the second figure) 28

ABEM Terraloc Pro
 Removing an existing first break marker  Press <-> until the timeline is invisible  Press <.> to remove the first break marker
4.4.6 Frequency View The Frequency View displays the frequency components of the trace. Here it is possible to check the amplitudes of the frequency components with the frequency line. The frequency and the corresponding amplitude value are displayed on the Record Status Bar just below the Frequency View.  Change the trace to view
 Press <UP> to change to the next trace  Press <DOWN> to change to the previous trace
 Moving a frequency line across the view (Figure 25)  Press <+> to move the frequency line to the right  Press <-> to move the frequency line to the left  Press <SHIFT> + <+> to move the frequency line to the right with a large step  Press <SHIFT> + <-> to move the frequency line to the left with a large step

Figure 25 The Frequency View with the frequency line

Note!

Please be aware that the values displayed, mostly are interpolated, as the frequency line represents a frequency calculated from the pixel coordinate, which can be in-between samples.

29

ABEM Terraloc Pro
4.4.7 Record Status Bar The Record Status Bar consists of two fields that displays trace centric information (Figure 26).

Figure 26 The Record Status Bar The leftmost field contains information as described in Table 3.
#nn Trace number T Sample interval in microseconds D Pre-trig/delay in milliseconds Len Length of trace in number of samples S Number of stacks
Table 3 Leftmost field information

The rightmost field displays different data depending on which view is highlighted. The following tables describe the three cases.

Note!

There will only be data displayed in the rightmost field if the timeline or frequency line respectively is visible

-

The Record View

t

Position of the timeline (ms)

A/D Measured value at timeline. Unit is available as raw A/D-value, V, mV, mm/s or cm/s. This is selectable in the view options dialog

-

The Trace View

t

Position of the timeline (ms)

A/D Measured value at timeline. Unit is available as raw A/D-value, V, mV, mm/s or cm/s. This is selectable in the view options dialog

dT The relative time (ms), the corresponding frequency within parenthesis. Only displayed when the reference time marker is used

f Level

The Frequency View Frequency (Hz) Amplitude (db)
Table 4 Rightmost field information

30

ABEM Terraloc Pro

Figure 27 The Record Status Bar with Trace View delta time

4.4.8 Application Status Bar The Application Status Bar displays general status information. There are seven separate fields on the bar:

Field

Description

The current record number

Is used the next time an acquired record is saved

The active acquisition Standard, Roll-along, or Optimum offset mode

The current instrument state

For possible states see Table 5 below

Power source status Internal with voltage

External

Activated reference The field is blank if no reference channel is activated channel

Error or warning alert for channels

Each board in the instrument has a one-character place in this field. See the three dashes in Figure 13. The About dialog (chapter 4.6.1) shows more information on each board.
Possible alerts: - = No error or warning B = Broken channel E = Warning for early trig (see the Warn for early trig setting in chapter 4.6.5.2) N= Warning for noisy trig (see the Warn for noisy trig setting in chapter 4.6.5.2)

GPS signal indication

Green background with dB value if fully functional
Red background with text "No GPS signal" if no signal is detected (usual behavior indoors)
Red background with text "No GPS device" if SeisTW cannot get contact with the GPS

31

ABEM Terraloc Pro

No data

There is no data in memory and the instrument is ready to be armed. In this state all acquisition parameters can be changed

<<< ARMED >>> The instrument is armed and ready for a trigger. In this state no acquisition parameters can be changed

When multiple instruments are connected and synchronized, this <<< Pending arm >>> state is activated when the user arms one instrument, and it
awaits arm confirmation from the other instrument(s)

<<< Triggered >>> The instrument has triggered and data acquisition is proceeding

Transferring data ... The data has been acquired and is being transferred to the memory

Data in memory There is data in the memory; the instrument is ready to be armed. Some, but not all, acquisition parameters can be changed

<<<SAVING>>>

Data is being saved. When the save operation has finished the memory will be cleared, the record number incremented, and the instrument ready to be armed

<<< Testing >>> The geophone test is active

Geophone test data The memory contains geophone test data. Press <SAVE> to save the data, or <ESC> to reject

Accept or reject? Waiting for the user to accept or reject the acquired data for stack in preview mode. Press <ENT> to accept, <ESC> to reject

WARNING

A minor error occurred, or an informational message has to be displayed. Details will be displayed in a separate message

ERROR

A fatal or major error occurred. Detailed information is displayed in a separate error message

Table 5 Instrument states

32

ABEM Terraloc Pro
4.5 Menus SeisTW has a normal Windows main menu. Since this is easier to use with external mouse and keyboard than with the built-in keyboard there are also two complementing menu choices added, the Quick Menu and the Context Menu. These duplicates selected items from the Main Menu. There is also a separate pop-up menu, Clear Traces, that is used for clearing recorded data when needed.
4.5.1 The Main Menu The Main Menu is a normal Windows main menu.

Figure 28 The Main Menu ­ File menu item opened

Submenu File

Submenu items
- New: Create a new record. Opens the Select Acquisition Mode dialog (see chapter 4.6.3)
- Open: Open a previously saved record. A standard open file dialog is shown
- Close/Close All: Close one or all open record(s) - Save: Saves the current record. The current working
directory will be used. The filename has the form "DAT_xxxx.sg2" where xxxx is substituted with the next record number - Save As: Same as Save but the user can choose filename and which directory to save in. A standard Save As-file dialog is shown - Change Working Directory: A Browse For Folderdialog is shown from which the user can choose a new working directory - Page Setup: Opens the standard Page Setup-dialog
33

Edit View Setup
Action Process Window

ABEM Terraloc Pro
where page orientation, margins etc can be set. - Print: Opens the standard Print-dialog where printer
can be chosen. See 12 Appendix D. Printout Example for result examples - Exit: A confirmation dialog is shown and then SeisTW is closed - Reboot system: The instrument is rebooted (restarted) - Power off system: The instrument is turned off - Header info: Displays the Header info dialog (chapter 4.6.11) - Source/receiver locations: Displays the Source/receiver locations dialog (chapter 4.6.11) - Preferences: Displays the Preferences dialog (chapter 4.6.3) - Toolbar: Hides/Shows the Toolbar - Logging: Hides/Shows the Logging Window - Details: Hides/Shows Trace/Frequency - Refresh: Refreshes the SeisTW window - Velocity analysis: Displays the Velocity Analysis dialog (4.6.14) - Options: Displays the View Options dialog (4.6.13) - Sampling: Displays the Acquisition Setup dialog (4.6.5.1) - Trig: Displays the Trig Setup dialog (4.6.5.2) - Noise Monitor: Displays the Noise Monitor dialog (4.6.5.3) - Filters: Displays the Acquisition Filter Setup dialog (4.6.5.4) - Receiver spread: Displays the Receiver Spread dialog (4.6.6) - Layout geometry: Displays the Layout Geometry dialog (4.6.9) - Header info: Displays the Header Info dialog (4.6.11) - Arm: Arms the instrument - Geophone test: Starts a geophone test (4.6.7) - Force trig: Forces a trigger - Disarm: Disarms the instrument
- Auto pick: Performs an automatic first break pick (5.1)
- Clear picks: Clears all first break picks (5.1) - FIR filter: Displays the FIR Filter dialog (5.3) - Moving average filter: Displays the Moving average
dialog (5.5) - Unfilter data: Reloads the original unfiltered data - Cross Correlate: Displays the Cross Correlate dialog
(5.4) Standard Windows Window submenu
34

ABEM Terraloc Pro

Help

- Help: Displays the help file - Keyboard help: Displays a specific part of the help
file - System info: Displays the System Information dialog
(4.6.2) - About: Displays the About dialog (4.6.1)
Table 6 Main Menu items

4.5.2 The Quick Menu Duplicates most of the menu items from the File submenu of the Main Menu (Figure 29). See chapter 4.5.1 for specifics on each menu sub item.
 Opening the Quick Menu
 Press <CTR> + <SPACE> to open the Quick Menu

Figure 29 The Quick Menu

4.5.3 The Context Menu
The Context Menu exists in two similar versions, a compact and a data version. The compact version is shown when no data exists in the current record (Figure 30). Consequently the data version is shown when data exists (Figure 31).
The compact Context Menu duplicates some menu items from three submenus of the Main Menu (Process, View and Actions) and also from the Clear Traces pop-up menu. See chapter 4.5.1 and 4.5.4 for specifics on each menu sub item.
The data Context Menu on the other hand duplicates the entire Process submenu as well as some menu items from the View submenu and also the Clear Traces pop-up menu.

Note!

The Context Menu will not be shown if no record is created or opened

35

ABEM Terraloc Pro  Opening the Context Menu
 Press <SPACE> to open the Context Menu Or  Right-click with a mouse
Figure 30 The Compact Context Menu

Figure 31 The Data Context Menu

See chapter 5.1 for more on the First breaks submenu functions.

4.5.4 The Clear Traces Menu
Used to clear one or more traces of recorded data. In contrast to the Delete last shot command these clear traces commands will clear all stackings, if any.

Note!

The menu items of the Clear Traces Menu are not available from the Main Menu.

Note!

The Clear Traces Menu will only be shown when data has been recorded

 Opening the Clear Traces Menu
 Press <ESC> to open the Clear Traces Menu Or  Via the Clear Traces submenu of the compact Context Menu

Figure 32 The Clear Traces Menu 36

ABEM Terraloc Pro 4.6 Dialogs 4.6.1 The About Dialog The About dialog displays information about the serial number, software versions, number of boards, number of measurement channels, the health of the boards etc (Figure 33).  Opening the About dialog
 Press <SHIFT> + <HLP> to open the About dialog
Figure 33 The About Dialog 4.6.2 The System Information Dialog Displays information about the GPS system status (Figure 34). This dialog can only be accessed from the Help submenu of the Main Menu.
Figure 34 The System Information Dialog 4.6.3 The Preferences Dialog Various general settings can be accessed from this dialog. The settings are divided into four areas, each with its own tab on the dialog.  Opening the Preferences dialog  Press <CTR> + <9> to open the Preferences dialog
37

ABEM Terraloc Pro - The next record number is normally incremented
automatically but the next number to use can be set here - The format of the saved recorded data can be set - By default SeisTW prompts for an exit confirmation but this can be turned off
- Various colors can be set here. The four colored areas are buttons that, when pressed, will show a standard Windows color select dialog.
- The stretching along the timeline can be set here. Values between 1 and 8 are allowed. A higher value result in increased stretch. See 12 Appendix D. Printout Example for result examples
- The impedance and resistivity of the receivers are set here
38

ABEM Terraloc Pro
4.6.4 The Select Acquisition Mode Dialog The Select Acquisition Mode dialog is used to change the acquisition mode and to change the number of traces to be used (Figure 35).  Opening the Select Acquisition Mode dialog
 Press <CTR> + <SPACE> to show the Quick Menu  Press <1> to execute the New menu item, which will open the Select
Acquisition Mode dialog

Figure 35 The Select Acquisition Mode Dialog

The different acquisition modes:

Standard

- All acquisition is performed according to the current settings. The only automatic actions are clearing the memory and updating the record after a save & update operation. The number of traces to use can be changed in this mode only from this dialog

Roll-along - When first pressing <ARM>, a new record is created containing the number of traces defined by the Roll-along start/end parameters in the Layout Geometry Dialog. Pressing <SAVE> will cause the record to be saved and the rollalong parameters to be updated according to the Roll-along step size as defined in the Layout Geometry Dialog. How the Roll-along parameters are updated is determined by the Roll-along reverse direction check box

Optimum - When a record is created it will initially only have the first trace's

offset

stack enabled. Besides, only the currently active trace and traces

containing data will be visible. When the data for the currently active

trace has been acquired, the user can press <SAVE>, which will

advance the active trace one trace. Pressing <SAVE> when the last

trace is active will save and update the record.

It is still possible for the user to modify acquisition parameters,

including receiver spread parameters, but be careful. Modifying

receiver spread parameters, may lead to acquiring data on a trace that

already contains data, but should not be updated

Table 7 Acquisition Modes

39

ABEM Terraloc Pro
4.6.5 The Acquisition Setup Dialog The Acquisition Setup dialog is a container for four different categories of settings for data acquisition: sampling, trig, noise and filters (Figure 36). Each category has its own tab on the dialog and they will be described in separate sub-chapters below. It is also possible to access them all without closing the dialog in-between.  Switching between setting categories when the dialog is displayed
 Press <CTR> + <TAB> to switch to the next category (tab) Or  Press <SHIFT> + <CTR> + <TAB> to switch to the previous category (tab)
SeisTW will remember the latest used combined acquisition settings between sessions. It is also possible to save the settings to disk and later reload them. The settings are stored in acquisition settings files (*.acq), which are text files with an ini-file format.  Saving acquisition settings to disk
 Press <TAB> until the Save button is selected  Press <ENTER> (or <SPACE>) to open a save as dialog  Name the file by pressing <NUMBERS>  Press <ENTER> to save the file
 Reloading acquisition settings from disk  Press <TAB> until the Load button is selected  Press <ENTER> (or <SPACE>) to open a select file dialog  Press <SHIFT> + <TAB> to move the focus to the file list  Press <ARROWS> to select the wanted file  Press <ENTER> to reload the file
 Restore default acquisition settings  Press <TAB> until the Default button is selected  Press <SPACE>
4.6.5.1 The Sampling Settings Category
These settings control how SeisTW will sample data.  Opening the Sampling Settings (Acquisition Setup dialog with the Setup tab
selected)  Press <1>
40

ABEM Terraloc Pro

Figure 36 The Acquisition Setup Dialog; Setup tab selected

The resultant record length will vary from short (5.1 ms) to long (80 minutes) depending on your choice of sampling interval and number of samples to be recorded.
Record length = `Sampling interval' x `Number of samples'
If a long sampling interval is combined with a low number of samples, the resulting record file will be small (takes up less disk space), but will contain less information and your interpretation possibilities will be reduced. Conversely, a short sampling interval with a high number of samples will give you good information for interpretation, but file size will be larger. Your choice will always be a compromise
Usually the sampling interval is determined by other factors than the record length. Thus, changing the number of samples to record usually varies the record length. However, if the number of samples available cannot give a suitable record length you may have to change the sampling interval
Stacking is a function to enhance the quality of the recorded data. Samples from more than one shot are added to each other giving a suppression of noise in comparison to the relevant data.

Setting

Description

Sampling interval No of samples Pretrig/delay (ms)

- Available sample intervals are: 20, 40, 100, 200, 400, 1000, 2000, 4000 and 10000 microseconds
- Number of samples to acquire. Available choices are: 256, 512, 1024, 2048, 4096, 8192, 16384, 32768, 65536, 131072, 262144, 480000
- Selects the pre-trig or delay for the trig event. Pre-trig is set by entering a negative time, and will save the corresponding amount of data before the trig event. Delay is set by entering a

41

No of stacks Stack mode

ABEM Terraloc Pro

positive time, and will delay data acquisition after the trig corresponding to the delay. The pre-trig/delay is measured in milliseconds. The pre-trig can be set from 1 ms to the record length. The delay can be set to the following ranges:

Sample interval (ms)
20 40 100 200 400 1000 2000 4000 10000

Delay range (s)
0 - 9.6 0 - 19.2 0 - 48.0 0 - 96.0 0 - 192.0 0 - 480.0 0 - 960.0 0 - 1920.0 0 - 4800.0

- If this number is greater than zero, the record will be saved automatically when this number of stacks has been acquired. If you type 0 (zero), stacking will continue until you press <SAVE>. Even if you type a number higher than 0 (zero) you may always interrupt stacking by pressing the <SAVE> key. When the record has been saved, the next record will be initialized

- The stack mode determines how the acquired data is added to the stack and how it is displayed. The following stacking modes are available:

Name Description
Fast Adds the acquired data to the stack as soon as the data is available. Does not display the data. The instrument is automatically armed for the next shot. This mode gives the highest rate for data collection as no screen update takes place
Auto The same as the Fast stack, but the stacked data is displayed. The instrument is automatically armed for the next shot
Preview Displays the acquired data and prompts the user to accept or reject the data. When the data is accepted, it is added to the stack, and the stacked data is displayed. Press <ENT> to accept or <ESC> to reject the acquired data. The instrument is automatically armed for the next shot. If a new shot is received before the <ENT> key is pressed the previous shot is lost. The last shot added to the stack cannot be removed by the "delete last shot" feature
Single Same as Auto stack, but the instrument has to be manually armed again for the next shot

42

ABEM Terraloc Pro

Re-arm mode

- If it is set to Auto, the instrument is automatically armed after a record has been saved. This is useful in, for example, marine surveys. If set to Manual the user has to arm the instrument by pressing <ARM>, or some external arm source has to set the arm input to its armed state.

Input voltage range - Available choices are: 500 mV, 5.0 V and 12.5 V

Input gain (dB)

- This setting complements the Input voltage range setting Available choices are: 0, 12, 24, 36, 48

Note!

Setting the input gain to 0 dB makes it possible to measure frequencies down to 0 Hz whereas higher settings gives measurement down to 1 Hz

Input impedance

- Set up for different types of sensors. Examples are 3000 Ohm for ABEM sensors and High for hydrophones Available choices are: 3000 Ohm, 20 kOhm, High

Table 8 Sampling settings

4.6.5.2 The Trig Settings Category
These settings control when SeisTW will sample data i.e. how sampling will be triggered (Figure 37).
 Opening the Trig Setup dialog (Acquisition Setup dialog with the Trig tab selected)
 Press <2>

43

ABEM Terraloc Pro Figure 37 The Acquisition Setup Dialog; Trig tab selected

Setting
Trig input mode

Description
- Selects trig input source, and its mode The following modes are available:

Trig input level

Name Analog
Make/Break
TTL Rising Edge TTL Falling Edge Channel

Description
When using the trigger input connector, the instrument is triggered when the signal exceeds the trig input level on the analog trig input. Select Analog when you use a standard trigger geophone or a trigger coil. If you use Analog triggering, you should check and/or set the trig input level.
The instrument is triggered when a trigger circuit connected to the trigger input connector is closed (make) or opened (break). The trigger circuit can for example be a twisted pair of insulated wires inserted in a dynamite charge. The wires are then shorted when the charge explodes (make switch). A break switch can be a single wire, which has been wound a few turns around the charge and the explosion cuts the wire (break switch). The instrument detects the state change from opened to closed (make), or from closed to opened (break), depending on the state at the time of arm. Set Trig Input Level to a low value to avoid inadvertent triggering by spurious signals.
The instrument is triggered when the TTL signal on the digital trig input goes from low to high
The instrument is triggered when the TTL signal on the digital trig input goes from high to low
The instrument is triggered when the signal on any channel input, including the reference channels, exceeds the trig input level. If you use Channel triggering, you should check and/or set the trig input level.

Manual only The instrument will only trigger manually from the keyboard (internal or external)

- The trig input level can be set from 0 to 100%. Increasing the trigger input level increases the sensitivity, which means that a lower signal level is needed to trig the Terraloc Pro. Decreasing the trigger input level on the other hand decreases the sensitivity, which means that a higher signal level is needed to trig the Terraloc Pro. Sensitivity level needs to be high enough to ensure triggering by the trigger signal, but not so high that spurious signals will trigger in advance of the actual shot impulse. For example, when a geophone is used as the source of the trigger signal, a time

44

ABEM Terraloc Pro
delay will always be present between the shot instant and the triggering time. There are two main causes for this:
1. The propagation delay from the shot point to the geophone 2. The rise time of the geophone output signal to the triggering level
Figure 38 illustrates the relationship between trigger sensitivity and the rise time of the receiver output signal to the triggering level.

Figure 38 Trig signal from a geophone and the trig event
To reduce the propagation delay the only way is to move the geophone closer to the shot point. This cannot always be done due to physical limitations in which case you will have to accept the delay. The rise effect is another matter, because it is influenced by a number of conflicting requirements. If the trigger sensitivity is increased, the result is of course an earlier trig event, but increasing the sensitivity also means that the risk of triggering the system by a noise signal increases. If the sensitivity is too low, noise triggering will not occur, but instead a considerable and poorly defined delay is introduced. This can seriously degrade the performance of the stacking of signals, since any signal with a period time comparable to, or less than this trig event uncertainty, will be attenuated. So in conclusion, you will have to find a suitable compromise between high sensitivity to false triggering and large timing errors.

Warn for noisy trig

- The meaning is to warn when there is a risk that sampling was triggered on noise instead of signal level. A possible warning is shown in the status bar (see chapter 4.4.8). This setting, together with its three sub settings (below), decides how the evaluation is done. If the signal level is higher than the given level in connection with the trigger point then the warning is raised. Figure 39 illustrates the meaning of the involved settings.

45

ABEM Terraloc Pro

Evaluation time [ms]
Guard time [ms]

Figure 39 Trig signal from a geophone and the trig event
- The time window during which the signal level is checked against the Noise warning level. See Figure 39 above
- A time window where the signal level is not checked. This is to avoid false warnings from the time just before the trig point. See Figure 39 above

Noise

- The threshold level for the noise warning. See Figure 39 above

warning level

[%]

Warn for early trig

- The meaning is to warn when there is a risk that sampling was triggered before a stable measurement was possible. A possible warning is shown in the status bar (see chapter 4.4.8).

Table 9 Trig settings

External Arm Input
External arm is used when interconnecting two or more Terraloc Pros using the TTL Arm/Trig connector (Figure 69 chapter 10.3). There is no limit for how many Terraloc Pros may be connected in this way. When external arm input is on the Terraloc Pro monitors the input continuously and if a correct signal is received the Terraloc Pro will arm.

Note!

If you have several instruments or devices connected in a "daisy chain", you must ensure that both Arm Input mode and Arm Out mode are properly defined on each instrument (i.e. they must all be set to either TTL rising edge or TTL falling edge)

46

ABEM Terraloc Pro
External Arm/Trig Output Use this to inform other electronic devices (seismographs, vibrators, computers, etc) that the Terraloc Pro has triggered. The signal is in TTL standard and uses the TTL Arm/Trig connector (Figure 69 chapter 10.3)

Setting

Description

External trig - The following modes are available: out mode

Name

Description

Off

The Trig-out is off

TTL Rising The instrument will make the trig-out signal go from

Edge

low to high when the instrument gets armed

TTL Falling The instrument will make the trig-out signal go from

Edge

high to low when the instrument gets armed

External arm - The following modes are available: out mode

Name

Description

Off

The Arm-out is off

TTL Rising The instrument will make the arm-out signal go from

Edge

low to high when the instrument gets armed

TTL Falling The instrument will make the arm-out signal go from

Edge

high to low when the instrument gets armed

External arm - The following modes are available: input mode

Name

Description

Off

The TTL Arm/Trig input is not monitored

TTL Rising The instrument is armed when the TTL signal on the

Edge

TTL Arm/Trig input goes from low to high

TTL Falling The instrument is armed when the TTL signal on the

Edge

TTL Arm/Trig input goes from high to low

Ext. arm verify

- When several instruments are interconnected, the external arm inputs and outputs can be connected in such a way that when one instrument is armed it in turn will arm the next instrument. If this choice is checked when the user presses <ARM> on one instrument, it will wait until it receives an external arm from the last instrument in the chain before actually accepting the arm event. If no external arm is received within the timeout set, the instrument will disarm and the disarm event will propagate to all the other instruments

Verify

- The time to wait for an external arm before disarming and showing

timeout [ms] an error message

Table 10 External Arm/Trig settings

47

ABEM Terraloc Pro
4.6.5.3 The Noise Monitor Settings Category
The Noise Monitor Setup dialog (Figure 40) has settings that control the Noise Monitor dialog (chapter 4.6.6).  Opening the Noise Monitor Setup dialog (Acquisition Setup dialog with the Noise
tab selected)  Press <3>

Figure 40 The Acquisition Setup Dialog; Noise tab selected

Setting

Description

Noise monitor - When the noise monitor is On, it will be displayed when the

status

instrument is armed. Available choices are: On, Off

Attenuation [dB]

- Attenuation in decibels of the displayed signal. The maximum displayed signal level is always 120 dB, but the lowest signal level displayed can be set by changing the Attenuation

Threshold level [dB]

- Sets a threshold level in decibels. Size depends on amplitude scale of the noise monitor. When the monitored signal exceeds this threshold level, a warning is displayed in the noise monitor window

Show noise

- Press this button to directly display the noise monitor. Press <ESC> to close it

Table 11 Noise Monitor settings

48

ABEM Terraloc Pro
4.6.5.4 The Filters Settings Category
These settings control how SeisTW filter data to be sampled (Figure 41).  Opening the Filters Setup (Acquisition Setup dialog with the Filters tab selected)
 Press <4>

Figure 41 The Acquisition Setup Dialog; Filters tab selected

Signals usually contain noise from sources such as wind and traffic. This noise often has low frequency. Analog filter removes these frequencies from the signals. However, the filter may also deteriorate original signals. The higher the cut-off frequency and filter damping, the worse possible distortions become. Using an analog filter is always a compromise.
If the noise level is high, record it. Use the frequency view to analyze it, and to see the actual noise frequency. Thereafter select and use an appropriate analog filter.
If the noise level is not high, do not use the analog filters.
Analog filters affect all channels.

Note!

Note that you will not be able to recover any incoming signals that are filtered out. Use analog filters only to remove low frequency ground roll. Generally be cautious about using these filters, as there is always a risk that they may eliminate valuable signal information

Setting Notch filter

Description - Turns the notch filter on or off. The notch filter is calibrated at
factory for either 50 or 60 Hz. Use this when working in vicinity of
49

ABEM Terraloc Pro

Lowcut filter Status
Lowcut filter Slope
Lowcut filter Cutoff freq.

power lines otherwise leave it off. A spectrum analysis of a noise recording may often show if power line noise is present
- Turns the analog low-cut filter on or off

- Select the slope of the filter. Available choices are 12 dB/octave and 24 dB/octave

- Selects the low cutoff (3 dB rejection) frequency in Hz. The possible frequency choice depends on the selected slope. You have 16 different cut-off frequencies for each filter slope to choose from, see the table below. When choosing filter slope, remember that generally 24 dB/octave filters distort more than 12 dB/octave, but will also damp noise more effectively. Use as low a cut off frequency as possible generally twice the maximum noise frequency. A good rule is to start with the 12-dB/octave filter. If the recorded signal is acceptable then keep the filter, otherwise try again with the 24dB/octave filter.

12 dB/octave
12 24 36 48 60 72 84 96 108 120 132 144 156 168 180 192

24 dB/octave
15 30 45 60 75 90 105 120 135 150 165 180 195 210 225 240

Table 12 Filter settings

4.6.6 The Noise Monitor Dialog There is a real-time noise monitor integrated in the system. It is displayed in the Noise Monitor dialog (Figure 42). The Noise Monitor can be used to just inspect the noise level, or monitor the noise so that the operator is able to fire the shot at the right moment.
 Opening the Noise Monitor dialog
 The Noise Monitor is opened either directly from Noise Monitor Setup dialog with the Show Noise button or when the Noise Monitor Status is turned on and
50

ABEM Terraloc Pro the instrument is armed
Figure 42 The Noise Monitor Dialog  Adjusting the attenuation (Figure 43)
 Press <SHIFT> + <+> to increase the attenuation in 6 dB step  Press <SHIFT> + <-> to decrease the attenuation in 6 dB step
Figure 43 Increased attenuation value  Adjusting the threshold (Figure 44)
 Press <+> to increase the threshold in 1 dB step  Press <-> to decrease the threshold in 1 dB step
51

ABEM Terraloc Pro
Figure 44 Threshold adjustments  Activate geophone test relays (Figure 44)
 Press <SPACE> to activate test relays. It is not possible to do this when the instrument is armed
Figure 45 After activation of test relays 4.6.7 The Geophone Test Result Dialog This geophone test is more extensive than the geophone check available in the Noise Monitor. This test records the response from the geophones to an impulse signal. A DC-current is sent to the geophones dislocating the seismic mass of the geophone. When the DC-current is switched off, the mass performs a damped oscillation with its resonance frequency while it comes to rest. Thus, you will get a report on the maximum amplitude of the response, resonance frequency and damping. The recording of the response starts just before the DC-current is switched off. The response is recorded and SeisTW then analysis the recorded test data and determines the status of each channel.
52

ABEM Terraloc Pro After the analysis of the data has been performed the result is displayed as a normal record and as a report-log in a Textfile Viewer dialog. Furthermore these results are also saved in the current working directory as a record in SG2-format and as a report in text format. The files have file extensions of ".sg2" and ".log" respectively. The filename pattern is: TEST_xxxx-n where xxxx is the current record number and n is a serial number.  Starting the geophone test, which eventually displays the Geophone Test Result
dialog  Press <SHIFT> + <ARM>
Figure 46 The Geophone Test Result Dialog 4.6.8 The Receiver Spread Dialog The Receiver Spread dialog is used to set up the traces, including input channel mapping and polarity (Figure 47).  Opening the Receiver Spread dialog
 Press <5>
53

ABEM Terraloc Pro

Figure 47 The Receiver Spread Dialog

Note!

Default settings can be reloaded by using the Default button

Channel
Specifies the channel mapped to each trace. It is possible to map any channel to any trace, and one channel can be mapped to any number of traces. If the reference channel is enabled, it will be mapped to the trace as specified by the corresponding channel.
 Changing channel mapping
 Press <ARROWS> to select the wanted trace in the channel column
 Press <NUMBERS> to change the input channel of the trace

 Map all channels in forward direction (channel 1 to trace 1, 2 to 2 etc)  Press <ARROWS> to select any trace in the channel column  Press <SHIFT> + <+>

 Map all channels in reverse direction (channel 24 to trace 1, 23 to 2 etc for a 24 channel instrument)
 Press <ARROWS> to select any trace in the channel column  Press <SHIFT> + <->

Polarity Specifies the polarity of the recorded signal. If the polarity is positive, the signal will be stored as is. If the polarity is negative, the signal will be inverted before it is stored.  Changing polarity on one trace
54

ABEM Terraloc Pro
 Press <ARROWS> to select the wanted trace in the polarity column  Press <SPACE> to toggle the polarity Or  Press <+> to set a positive polarity Or  Press <-> to set a negative polarity
 Changing polarity on all traces  Press <ARROWS> to select any trace in the polarity column  Press <SHIFT> + <SPACE> to toggle the polarity on all traces Or  Press <SHIFT> + <+> to set all traces to a positive polarity Or  Press <SHIFT> + <-> to set all traces to a negative polarity
Stack Enables or disables stacking for the specified trace. If the stack for a trace is disabled (non-checked), data cannot be added (or subtracted) from that stack.  Changing stack state for one trace
 Press <ARROWS> to select the wanted trace in the stack column  Press <SPACE> to toggle the value Or  Press <1> to set a checked value Or  Press <0> to set a non-checked value
 Changing stack state for all traces  Press <ARROWS> to select any trace in the stack column  Press <SHIFT> + <SPACE> to toggle the value on all traces Or  Press <SHIFT> + <1> to set all traces to a checked value Or  Press <SHIFT> + <0> to set all traces to a non-checked value
55

ABEM Terraloc Pro Trace Enables or disables viewing of the specified trace.
Note! All traces will be recorded regardless of the Trace value  Changing trace state for one trace
 Press <ARROWS> to select the wanted trace in the trace column  Press <SPACE> to toggle the value Or  Press <1> to set a checked value Or  Press <0> to set a non-checked value  Changing trace state for all traces  Press <ARROWS> to select any trace in the trace column  Press <SHIFT> + <SPACE> to toggle the value on all traces Or  Press <SHIFT> + <1> to set all traces to a checked value Or  Press <SHIFT> + <0> to set all traces to a non-checked value
56

ABEM Terraloc Pro 4.6.9 The Layout Geometry Dialog The Layout Geometry dialog is divided into five different sections (Figure 48). Each section is described separately below.  Opening the Layout Geometry dialog
 Press <6>
Figure 48 The Layout Geometry Dialog Source location X, Y, and Z are coordinates of the source location.  Changing a value
 Press <TAB> to select the X, Y or Z value to be changed  Press <NUMBERS> and possibly <-> and <.> to construct a valid value  Press <TAB> to set the value and move to the next value Receiver locations X, Y, and Z are coordinates for the receivers.
57

ABEM Terraloc Pro
 Changing a value
 Press <ARROWS> to select the X, Y or Z value to be changed  Press <NUMBERS> and possibly <-> and <.> to construct a valid value  Press <ENT> to set the value and move down to the next value
When the first and second value has been given then the following values can be entered quicker given that the distances are the same.  Quick completion
 Press <CTR> + <DOWN> This will add the difference between the value in the first row and the second row to the value in the second row and enter this value in the third row. This can then be repeated for the following rows. Keep holding <CTR> + <DOWN> and all values for the current column will be filled in. If the value on the second row is larger than the one in the first row the difference will be added to the value in the second row and entered in the third row etc, for example starting with 0 on the first row and 5 on the second row will produce 10, 15, 20 etc in the following rows. If the value on the second row is smaller than the one in the first row the difference will be subtracted from the value in the second row and entered in the third row etc, for example starting with 100 on the first row and 95 on the second row will produce 90, 85, 80 etc in the following rows.
Note! Both positive and negative values are allowed

 Opening the Layout Helper dialog (see chapter 4.6.10) when the marker is located in the receiver locations part of the dialog  Press <SPACE>
Move-ups Describes how the source, receivers and the receiver connected to the reference channel (if any) are updated when a record has been finished.  Changing a value
 Press <TAB> to select the dX, dY or dZ value to be changed  Press <NUMBERS> and possibly <-> and <.> to construct a valid value  Press <TAB> to set the value and move to the next value

Note!

The ref. channel values are only available when the Ref. channel drop-down field is set to either Ch. A or Ch. B
58

ABEM Terraloc Pro
Roll-along The settings used to control roll-along measurements (see chapter 4.6.4 for more on roll-along). Note that it is assumed that the lowest numbered trace is on the left side and the highest numbered is on the right side.

Setting Roll-along reverse direction First trace
Length Step size

Description - If checked, the roll-along segments will be shifted to the left
(normally they are shifted to the right)
- The first (left-most) trace of the current cable layout for the current record
- Number of traces comprising one roll-along segment - Number of steps to shift the roll-along after finishing a record

General These are general settings for all sections of the dialog.

Setting

Description

Units

- Defines the linear units used for all location data. Possible values are: None, Meters, Centimeters, Feet and Inches. If None is specified it will be up to the user to interpret location data

Source type (*)

- An appropriate text string describing the source used to acquire this record. Pre-defined values are: Untitled, Hammer, Weight Drop, Seismic Gun, Explosives, and Vibrator.
The asterisk means that the user may enter any text string in this field

Receiver type - An appropriate text string describing the receivers used to acquire

(*)

this record. Pre-defined values are:

Untitled, Vertical_Geophone, SH_Horizontal_Geophone,

SV_Horizontal_Geophone, and Accelerometer

The asterisk means that the user may enter any text string in this field

Ref. channel

- Terraloc Pro with 12, 24 and 48 Channels have two extra reference channels. These channels connect thru banana terminals on the connector panel and have their own traces shown on screen Possible values are: Off, Ch. A and Ch. B

4.6.10 The Layout Helper Dialog The Layout Helper dialog (Figure 49) can be used to quickly fill in the receiver locations in the Layout geometry dialog.
 Opening the Layout Helper dialog
59

ABEM Terraloc Pro
 Press <SPACE> when the marker is in the Receiver locations section of the Layout Geometry dialog
Figure 49 The Layout Helper Dialog It is possible to enter values in any two of the entries layout start, layout end, and receiver separation. The third entry is calculated automatically.  Setting an entry to be calculated automatically
 Press <TAB> to select the entry to be automatically calculated  Press <SPACE>
 Accepting the values and exit  Press <TAB> to select the OK button  Press <SPACE> or <ENT> to exit from the dialog and automatically fill in the receiver locations grid
4.6.11 The Source/Receiver Locations Dialog This dialog is best used to get a view of source and receiver locations in existing record data. It also displays the locations per trace.  Opening the Source/Receiver Locations dialog
 Press <SHIFT> + <6> to open the Source/Receiver Locations dialog

Figure 50

The Source/Receiver Locations Dialog 60

ABEM Terraloc Pro
4.6.12 The Header Info Dialog The Header Info dialog enables input of general header information (Figure 51).  Opening the Header Info dialog
 Press <7> to open the Header Info dialog

Figure 51 The Header Info Dialog

Note!

Remember that an external USB-keyboard is needed to be able to write letters. Therefore it can be practical to fill in the header info before going out in field

Setting

Description

Job ID Line ID Client

- A text string identifying the job - A text string identifying the seismic line - A text string naming the client of the job

Company - A text string naming the company of the client

Observer - A text string naming the observer(s)

Note

- A free form text string

4.6.13 The View Options Dialog The View Options dialog handles settings for how data is viewed in SeisTW (Figure 52). The dialog is divided into six sections. The View Mode setting decides which one of the Normalize, AGC, Enhanced and Hyperbolic sections that is available for setting up. The Frequency Analysis section affects how the Frequency View (4.4.6) will present the frequency data.
 Opening the View Options dialog
61

 Press <9>

ABEM Terraloc Pro

Figure 52 The View Options Dialog

Setting

Description

View mode - The view mode determines how data is scaled for the display. The following modes are available:

Trace style

Name

Description

Enhanced Normalize AGC
Hyperbolic

The data is not scaled, but the amplitudes are used directly. It is, however, possible to attenuate the displayed signals using the enhanced attenuation parameter The max value in each trace is used to scale all samples in the trace. See also global scaling Uses the average amplitude calculated from a running window (which length is specified by the AGC window parameter). This means that each sample is scaled according to the average signal level in the samples vicinity Applies a hyperbolic scaling to the data. If the logarithmic choice is selected, the function ArcSinH will be used, otherwise the function TanH is used

- The following styles are available:

Name
VAR+ VARWiggle Dotted

Description
This is a wiggle trace with the positive side filled-in This is a wiggle trace with the negative side filled-in The trace is plotted as a wiggle Each sample value is plotted as a dot

62

ABEM Terraloc Pro

Time

- Selects compression in time. This makes more of the record visible.

compression Available values are:

1x, 2x, 4x and 8x

Scale factor - A general factor by which every sample is multiplied. Range: 1 ­ 1000

Trace clip

- How many traces the plotted curve may overlap before it is clipped. With trace clip = 1, no overlap will occur. If trace clip is 2, a trace may overlap the positive part of the trace on the left, and the negative part of the trace on the right Range: 1 ­ number of channels

Remove DC - If enabled, the DC offset is removed before the trace is scaled. It is

offset

recommended to keep this enabled

Show tic lines

- If enabled, major and minor tic lines will be plotted. The time interval between the tic lines is determined by sample interval and time compression

Figure 53 Tic lines On (left) and Off (right)

A/D conv. units

- Decides the unit type for the A/D-value that is displayed on the Record Status Bar (4.4.7). Available values are: None, V, mV, mm/s, cm/s (None = Raw A/D-value)

Global scaling

- If enabled, the maximum value in the whole record is used to normalize every sample of all traces. Only available when View Mode is set to Normalize

AGC window (ms)

- The length of the window, in milliseconds, used to calculate the average value to use for scaling of a sample value. The window moves along the trace with each sample that is scaled. Only available when View Mode is set to AGC Range 1 ­ 32000

Average

- If enabled, the average values on the stack is used for each trace, otherwise the summed stack is used Only available when View Mode is set to Enhanced

Attenuation - Used to attenuate the signals. This will bring out weaker signals, while

[dB]

hiding stronger signals.

Only available when View Mode is set to Enhanced

Logarithmic - If enabled, ArcSinH is used as the scaling function otherwise TanH is

63

ABEM Terraloc Pro

used. Only available when View Mode is set to Hyperbolic

Linearity range [dB]

- This value sets the amplitude level that is within the linear part of the scaling function. Both scaling functions are linear in the beginning (for small amplitudes), while compressing larger amplitudes. Only available when View Mode is set to Hyperbolic

Windowing function

- Selects the function to be used for data windowing. Available values are: No window, Hanning, Hamming, Blackman, Bartlett, Kaiser, 4th order Blackman-Harris, Flat top

Max frequency [Hz]

- The maximum frequency to display. The displayed spectrum will go from 0 Hz up to the selected maximum frequency. Available values are: 50, 100, 200, 500, 1000, 2000, 5000, 10000, 25000

Dynamic range [dB]

- The maximum frequency component is used as reference when calculating the spectrum. The displayed spectrum will go from 0 dB up to the selected maximum dB value. Available values are: from 6 dB up to 198 dB in steps of 6 dB

4.6.14 The Velocity Analyzer
The Velocity Analyzer consists of a dialog, which is displayed at the top of the screen, and the Velocity Marker, a red line that is displayed in the Record View (Figure 54). The Velocity Analyzer can be used to estimate the apparent seismic velocity in refraction records.

Figure 54 The Velocity Analyzer; Dialog and Velocity Marker 64

ABEM Terraloc Pro
When the Velocity Analyzer starts, it checks the receiver locations and calculates the receiver separation. If the receiver separation seems to be erroneous, or the receiver locations are not specified, a valid separation value can be entered in the dialog (Figure 55). When the Velocity Marker is tilted a velocity value is displayed in the dialog. The value is calculated from the slope of the line as a function of receiver separation. This way, you can move and tilt the Velocity Marker in such a manner that it correlates with for example first arrivals in a refraction record. Thus it is easy to find out velocities for different layers.

Figure 55 The Velocity Analyzer Dialog

Setting
Receiver separation Units

Description - The calculated value can be changed.
- The unit for the velocity value is set here. Available values are: None, m/s, cm/s, ft/s and in/s

 Opening the Velocity Analyzer  Press <SHIFT> + <8>

 Closing the Velocity Analyzer  Press <ESC>

 Moving the Velocity Marker (see the figures below)  Press <+> to move the line down  Press <-> to move the line up  Press <SHIFT> + <+> to move the line to the right  Press <SHIFT> + <-> to move the line to the left
The end of the Velocity Marker that has the little circle is the anchor point of the line. The other end is called the free end. This end moves when the Velocity Marker is being tilted and stretched.  Tilting the Velocity Marker (see the figures below)
 Press <CTR> + <RIGHT> to move the free end to the right  Press <CTR> + <LEFT> to move the free end to the left
65

ABEM Terraloc Pro  Press <CTR> + <DOWN> to move the free end down  Press <CTR> + <UP> to move the free end up Start position
Moved down
Moved to the right
Free end to the left
Free end further to the left
66

ABEM Terraloc Pro Free end to the right Free end up Free end down Free end left Free end up
67

Moved to the right

ABEM Terraloc Pro

68

ABEM Terraloc Pro

5 Data Processing

The data processing discussed below works on data in memory, not to a previously saved file.

Note!

If a Save command is done then previously stored data will be overwritten and lost. Use Save As and choose a different file name to keep both the original data and the processed data

5.1 Unfilter Data
The Unfilter data menu item on the data Context Menu (Figure 31) discards any processing results and reads back the original data from disk.

5.2 First Breaks Chapter 14.1 Refraction discusses First Breaks / First Arrivals. These entries are accessed via a sub menu in the Context Menu (Figure 56).

Figure 56 The First Breaks submenu
There are two formats available for pick files, ABEM's FIR-format, and REFLEXW's pck-format. Choose format before loading or saving the first break picks.
More information about the first arrival file format (FIR-format, FIrst aRrival) can be found in chapter 13 Appendix E. The First Arrivals File Format (FIR).

Menu item

Description

Auto pick Load picks Save picks

- Performs an automatic first break pick. Automatic computing of first arrivals works best on data with small pre-signal noise. You should always check the picked arrivals and edit any bad picks. If there are one or more picks for this record, you will be warned before times are picked automatically.
- Loads the first break picks from a pick file to the currently active record. If there are more picks in the pick file than traces in the record, the superfluous picks are discarded. If there are fewer picks, only the first traces loads the picks
- Saves the first break picks to a text file in the current directory.

Clear picks - Clears the first break picks

69

ABEM Terraloc Pro

5.3 FIR Filter
The FIR filters (Finite Impulse Response) are used to reduce noise from the recorded data.

Note!

The abbreviation FIR in this filter context is not the same as FIR used in the first breaks context

The FIR filter dialog (Figure 57) is accessed from the data Context Menu (Figure 31).

Figure 57 The FIR filter dialog Clicking OK will apply the filter to the current data.

Setting

Description

Filter type - Selects the type of filter to apply to data. The following types are available:

Name Low pass High pass Band pass
Band reject

Description
Rejects frequencies higher than the high cut-off Rejects frequencies lower than the low cutoff Rejects frequencies lower than the low cut-off and higher than the high cut-off Rejects frequencies between the low cut-off and high cut-off

Windowing - Data windowing function to apply to the data when filtering. Available

function

values are: Hanning and Blackman

Cut-off - The cut-off frequencies are specified as the frequency where the pass frequencies band signal has been reduced by 3 dB and the transition band starts.
The low- and high pass filters only specifies a single cut-off frequency, while the band pass and band reject filters specifies two frequencies, low- and high cut-off

Filter length

- The number of filter coefficients used to realize the filter. The longer the filter, the steeper its slope, i.e. it will cut the signal more abruptly. A longer filter also takes longer to apply, especially to long records

70

ABEM Terraloc Pro
5.4 Cross Correlate Chapter 14.6 Vibroseis discusses the use of cross correlation. Selecting the Cross correlate menu item from the data Context Menu (Figure 31) opens the Reference trace selection dialog (Figure 58).

Figure 58 The Reference trace selection dialog

Setting Ref. trace

Description - Value range: 1 ­ number of channels

Enter the trace number that was used for the reference signal and then press OK. The cross correlation may take several minutes, so be patient. The Cross correlation progressing status dialog will be shown (Figure 59). The progress is shown partly as a progress bar and partly time values. Updates to the dialog are a bit uneven but occur every 10th to 15th second.

Figure 59 The Cross correlation dialog
The next two figures (Figure 60 and Figure 61) first display raw data from a record acquired using vibration seismic and then after cross correlation has been applied to the data.

71

ABEM Terraloc Pro
Figure 60 Opened record ­ before processing
Figure 61 The same record ­ now cross-correlated 72

ABEM Terraloc Pro
5.5 Moving Average Moving average is used to analyze a set of data points by creating a series of averages of different subsets of the full data set. It can be used to smooth out short-term fluctuations and highlight longer-term trends. Selecting the Moving Average menu item from the data Context Menu (Figure 31) opens the Enter filter length dialog (Figure 62).

Figure 62 The Enter filter length dialog

Setting Filter length - Value range: 1 - 1023

Description

Enter the wanted filter length (the number of samples to use) and then press OK.

The next two figures (Figure 63 and Figure 64) first display raw data from a record acquired using vibration seismic and then after being processed with a moving average filter with a filter length of 15.

73

ABEM Terraloc Pro
Figure 63 Opened record ­ before processing
Figure 64 The same record ­ after moving average filter 74

ABEM Terraloc Pro

6 Triggering Methods

To make a recording with the Terraloc Pro seismic system, an initiating trigger signal is required. The trigger pulse defines the start of the data recording and is the reference for all timing.

6.1 Make/Break Switch Input
The system will trigger if the wires of the trigger cable are shorted together (make switch), or if the shorted wires are opened (break switch).
When you use explosives, one trigger method is to put a few turns of wire around the charge. The wire is cut by the explosion and triggers the seismograph (break switch). You can also twist a pair of insulated wires together and insert the twisted part into the dynamite. The explosion will compress the wires and crush/melt the insulation causing the leads of the wires to short together. This will trigger the seismograph (make switch).
Besides, this make switch triggering method can be applied when you use falling weight or hammer as energy source. However, you must use a metallic shock plate and the falling weight or the hammerhead must be made out of metal. Connect one lead of the trigger wire to the shock plate and the other lead to the falling weight or hammerhead. When the hammer hits the shock plate, the trigger circuit is shorted and the instrument triggers.

6.2 Using the Trigger Coil
If you want to trigger Terraloc Pro with the ignition current going out to the charge, you can use the Trigger Coil (current detector unit) included in the Terraloc Pro accessories. To use this you merely feed one of the two shot wires through the hole in the trigger coil. The trigger coil is either connected directly to the trigger input or to the extension connectors on the trigger cable reel. Then set the Terraloc Pro in the "Analog" trigger input mode with the sensitivity control set at about 50 %. When the charge is fired, the ignition current will trigger Terraloc Pro instantly. The current pickup trigger method is very convenient since you only need to bring the shot cable to the shot hole.

Note!

You have to use (seismic) blasting caps with no built-in delay to be able to use this method. If you use ordinary blasting caps the ignition delay will be included in your record. There are seismic blasting caps of the safety type available. Their delay is only some 50 µs if fired with a high power-blasting machine.

6.3 Radio Triggering
In case you need to trigger the Terraloc Pro in places where you cannot use a trigger cable, you can use a simple radio equipment to transmit the trigger pulse.

75

ABEM Terraloc Pro
7 Measurement
7.1 Basic Operations See chapter 3 Quick Start for an introduction to the most basic operations on the Terraloc Pro.  Initiating
 Press <ARM> Or  Press <CTR> + <SPACE> to open the Quick Menu  Press <1> to select New  Press <ENT>
 Arming  Press <ARM>
 Trigging  Press <CTR> + <ARM> to force a trig Or  Set up an automatic trigging (chapter 4.6.5.2)
 Saving  Press <SAVE> to save the current file (prompting for overwrite if the file already exists) Or  Press <SHIFT> + <SAVE> to open a "Save As"-dialog Or  Press <CTR> + <SAVE> to force a save of the current file (overwriting any existing file)
 Disarming  Press <ESC> to disarm an armed instrument
 Delete recorded data  Press <BACKSPACE> to delete the last acquired shot  Or  Press <ESC> to show the Clear Traces Menu  Press <1> or <2> or <3> to delete the wanted data
76

ABEM Terraloc Pro

 Open a stored record  Press <CTR> + <SPACE> to open the Quick Menu  Press <2>  Press <SHIFT> + <TAB> to move focus to the file list  Press <ARROWS> to select the wanted file  Press <ENTER>
 Close the current record  Press <CTR> + <SPACE> to open the Quick Menu  Press <3>
 Switch between opened records  Press <CTR> + <TAB> to switch forward  Press <CTR> + <SHIFT> + <TAB> to switch backward

7.2 Data Transfer
It is highly advisable to make backup copies of recorded data. As with every computerized system there is always a slight risk that data could be lost due to hardware failure or corrupted data. ABEM cannot take responsibility for recorded data that is lost.

7.2.1 Data Transfer Using the Ethernet Port
This is a function of Microsoft Windows XP Professional and not a specific function of the ABEM Terraloc Pro. Hence, ABEM cannot be responsible for any problems that may occur that isn't associated with the Terraloc Pro hardware or measurement programs developed by ABEM.
File transfers from your Terraloc Pro to a PC can be done using a network cable. You will also need an external USB-keyboard and USB-mouse for the Terraloc Pro. These parts are supplied with the Terraloc Pro at delivery.

Note!

If the Terraloc Pro is connected directly to a PC, rather than connected to an existing LAN (Local Area Network), it might be necessary to use a crossed network cable. This is not supplied with the instrument but is available in most computer stores.

One way to do file transfer over the network cable connection in Windows is to set up a small network and allow sharing of a folder, which measurement data can be copied to. By default the Terraloc Pro is set up with the computer name Terraloc Pro and to be a member of a workgroup called WORKGROUP. The IP number of the Terraloc Pro can be found by writing the "ipconfig" command in the command prompt accessible under "Start menu /All programs / Accessories".

77

ABEM Terraloc Pro

To access the shared folder in the Terraloc Pro from the PC, go to "Start / Search / select Computers or people" from the menu on the left and then "A computer on the network". At this point enter the IP number of the Terraloc Pro in the Computer name search field and click the search button. You may have to login, but then you should be able to access the files.

Note!

There are security issues with sharing folders. Never put any sensitive data in a shared folder since it is easy accessible when the Terraloc Pro is connected to a network or to the Internet. It is also advisable to only allow reading of files and not writing.

Another option is to configure the Terraloc Pro for use in a normal office network (LAN) wired or wireless (WiFi); this will not be described in this manual.

7.2.2 Data Transfer Using an USB Memory Stick/Drive
Terraloc Pro has built in USB 2.0 ports for fast and easy file copying to a USB memory stick/hard disk.

7.3 Optimizing
Many of the settings you select affect the performance of the system. You can set up the system to do the data acquisition as fast as possible, or to give you as much information during the acquisition as possible, which often means a more secure operation.

7.3.1 For Speed Sometimes, for example in marine seismic surveys, it is important to obtain fast data acquisition. There are some operations that can be modified, or even skipped, to enhance the acquisition speed. Still, there are some operations that are fixed, and to this category belong the actual data acquisition (sampling procedure), transfer of the data from acquisition memory to trace memory, and writing of the data to disk.
However, the following should be considered:
 Do not display data after shooting, i.e. set Stack Mode to "Fast stack" since scaling of traces takes a considerable time (chapter 4.6.5.1).
 Use as short records as possible (No of samples) (chapter 4.6.5.1).

7.3.2 For Security When you optimize for security, you set the instrument up to give you as much information about the data acquisition as possible. This means that, e.g. data and progress are displayed.
 Set Stack Mode to "Preview" or "Single" (chapter 4.6.5.1).
 Use the noise monitor if you are in "´Single" Stack Mode (chapter 4.6.5.3).

78

ABEM Terraloc Pro
8 Troubleshotting and Diagnostics
Although great care has been taken to make Terraloc Pro as reliable as possible, there is always a small risk that something does not work properly. Should you have trouble getting things to work please refer to this chapter. This is a guide to common problems and how to work them out.
8.1 General SeisTW Program Problems These errors are generally related to the software.
8.1.1 The Program Does Not Start There should normally be no problem starting SeisTW in the Terraloc Pro once it has been installed. However, if SeisTW does not start when starting the instrument the program might need to be reinstalled. Follow the installation instructions in 11 Appendix C. SeisTW Installation.
8.2 Data Acquisition Problems The data acquisition problems can range from errors in the setting up of the system for measurement, over hardware problems, to errors in the settings in the software.
8.2.1 Terraloc Pro Only Waits For Confirmation When Arming If the instrument shows the status message "<<<Pending ARM ...>>>" when you try to arm the instrument, it means that External Arm Input mode is set to TTL Rising edge or TTL Falling edge. If it is, it will wait for an external arm signal to arrive before it arms itself. This is used when you interconnect two or more Terraloc Pro instruments (chapter 2.6). If you use one Terraloc Pro only, no external arm will arrive, hence the Terraloc Pro never arms (unless you have some other external device that confirms the arm command). Set External Arm Input to Off and the arming will be normal (chapter 4.6.5.2).
8.2.2 Dead Channels/Traces Check the Receiver Spread dialog for the settings of the "Stack On" and "Trace On" parameters (chapter 4.6.8). You should also check the reference channel setting in the Layout Geometry dialog (chapter 4.6.9).
8.2.3 Data Is Not Displayed Check if you have selected Fast Stack as stacking mode, as this causes data not to be displayed on the screen (chapter 4.6.5.1). In case of Auto Stack, Preview and Single, check that the "Trace" parameters in the Receiver Spread dialog are activated (chapter 4.6.8).
8.2.4 Large Offset Check offset level and do not worry if it is less than 2000 units.
79

ABEM Terraloc Pro
8.2.5 Incorrect Channel Order Either one cable, at least, has been reversed in the layout or the channel assignments are erroneous. Check the cable and/or the channel assignments in the Receiver Spread dialog (chapter 4.6.8). Please be aware that a reversed cable can be corrected for in this dialog.
8.3 Trigger Problems Correct triggering is essential for the quality of the data from the acquisition, especially when it comes to timing. This means that you should be very careful when selecting triggering method and setting up the triggering system. It may not always be obvious that there is something wrong with the trigger.
8.3.1 Triggering Too Late or Too Early Erroneous setting of the trigger sensitivity usually causes this when analog triggering is used. Adjust the sensitivity level so the trigger pulse is detected correctly (chapter 4.6.5.2).
8.3.2 Spurious Triggering This is usually caused by too high trigger sensitivity, resulting in triggering on pretrig event noise. Adjust the sensitivity level so the trigger pulse is detected correctly (chapter 4.6.5.2).
If you are using radio triggering, also check the signal levels of the transmitter and receiver respectively.
8.3.3 Unable To Trigger The trigger sensitivity might have been set too low, or the type of trigger input does not agree with the trigger method used. Check the trigger settings (chapter 4.6.5.2) and the trigger cable; there may be a break in the cable or a bad connection somewhere.
Select "Make/Break" trig input mode (the trigger input level should be about 50%) and try to short the trigger input by a bare wire. The instrument should trigger when you make or when you break connection. If the instrument does not trigger, then you might have a fault in the internal triggering electronics.
8.3.4 Triggering Immediately When Arming If you are using analog triggering, the trig sensitivity might have been set too high. At the highest sensitivity level, even internal electronic circuitry noise may cause triggering (chapter 4.6.5.2).
8.4 Remote Diagnostics (VPN) The Terraloc Pro can be connected to ABEM for remote diagnostics over a VPN (Virtual Private Network). To connect the instrument to a VPN you need a standard Ethernet based TCP/IP LAN (Local Area Network) that is connected to the Internet. The instrument is connected to the LAN either wired with a RJ-45 cable or wireless with WiFi.
If the LAN has a DHCP service, the instrument will acquire an IP number and most likely the other required network settings from the DHCP server when the network
80

ABEM Terraloc Pro

service starts. Note that the DHCP server must allow unregistered MAC addresses. If it does not, the instrument's MAC address must be registered in it. Please contact your local network administrator if this is necessary.
Restrictions: The LAN router or firewall must not block outgoing traffic on port 1194, and must allow incoming traffic that is initiated from inside the LAN to be returned to the instrument. It must also allow VPN communication with the ABEM office (www.abemoffice.com). Further, if the LAN is using NAT, it must not use the private IP network 10.17.23.x since the VPN will be using it. Most office LANs will meet these specifications.
Please note: Some countries/companies have firewall rules that blocks access to this type of service.

Customer Network

ABEM

Terraloc Pro

Router

Internet

Allow outgoing traffic to www.abemoffice.com port 1194

Router

Server

Figure 65 Remote diagnostics over VPN
If you are not familiar with the terminology in this section, and experience problems with the connection, please contact your local network administrator.
 Establishing a connection  Double-click on the
desktop icon named Connect to ABEM VPN

81

ABEM Terraloc Pro The instrument will try to establish a connection
Tray icons will change like this during the connection phase
Connection established!

 Disconnecting
 Right-click on the OpenVPN tray icon to open its context menu

OpenVPN tray icon

82

 Select the Disconnect menu item

ABEM Terraloc Pro

Connection disconnected!
8.5 In Case of Malfunction In case of malfunction please carry out applicable tests as described in this manual. If it is not possible to find the cause of the problem, follow the instructions in Section 8.4 Remote Diagnostics (VPN) to connect the instrument to ABEM's technical support, and send a description of the problem via e-mail to support@guidelinegeo.com.
Should a fault occur that is not correctable on site, please send full details to ABEM. It is essential that the instrument type and serial number is included and, if possible, the original ABEM delivery number. On receipt of this information, disposition instructions will be sent by return. Freight to ABEM must be prepaid. For damage or repairs outside the terms of the Warranty, ABEM will submit an estimate before putting the work in hand.
Be sure to fill in the warranty registration card (included with the equipment) correctly and return it to ABEM promptly. This will help us process any claims that may be made under the warranty. It will also help us keeping you informed about for instance free software upgrades. ABEM welcomes your response at any time. Please let us know your name and address, and the serial number of the instrument.

83

ABEM Terraloc Pro

9 Appendix A. Technical Specification
General

Number of channels

12, 24 or 48

Additional channels

Easily obtained by linking two or more units together

Up-hole channel Sampling rate (selectable)

Yes, 2 additional independent 100 sps ­ 50 ksps (20 s ­ 10 ms)

Record length (selectable) Up to 480 k samples / ch. equivalent to: 9,6 s ­ 80 min

Pre-trig record (selectable) 0 ­ 100 % of record length

Delay time

Up to 2 minutes

Stacking

32 bits, up to 999 impacts

Unstack

Remove last shot from stack

Trigger inputs

Trigger coil, make/break, geophone, TTL

A/D converter resolution

24 bits

Dynamic range

(theoretical/measured) 144 dB / >120 dB

Max input signal/ impedance 0,5 Vpp/3 k, 5 Vpp/20 k, 12,5 Vpp/3 k, hi impedance

Frequency range

DC to 20 kHz hi imp

Total harmonic distortion 0,0005%

Crosstalk

-120 dB

Noise monitor

Amplitude

Anti-alias filters

Set automatically based on sampling rate

Connectors

NK-27 / KPT 55

Power

10 ­ 34 V DC external power, 12 V 8Ah NiMh internal battery

Power consumption

30/60 W (man/acq)

Ambient temp (operating) -20 to + 55 °C

Ambient temp (storage)

-30 to + 70 °C

Casing

Rugged Al alloy; Meets IEC IP 66

Weight, 24 channels

10 kg

Weight, 48 channels

11 kg

Dimensions (W x L x H)

39 x 21 x 32 cm

84

Post recording features
Digital filters Spectrum analysis Velocity Analysis First-arrivals picking Pre-stack correlation

ABEM Terraloc Pro
Band-, low-, high- pass band-reject, remove DC offset Any single trace, FFT analysis On-screen analysis of refractor velocity Automatic or manual. Times can be saved with record. Yes, cross correlation with reference or any other ch.

Processor, RAM and hard disk

Processor Operating System Internal RAM Hard disk capacity Display External display port I / O port Network interface

Low power Intel Atom, 1,6 GHz Windows XP Pro 2GB (DDR SO-DIMM module) at least 100 GB 8,4" Active TFT LCD, full color, daylight visible, 800x600 VGA output 3 x USB 2.0 ports 1 x IEEE 802.3 TP-10/100/1000 (RJ-45 IP67) 2xTP-10/100 KPT 08 Inbuilt WLAN Antenna in handle

85

ABEM Terraloc Pro
10 Appendix B. Connectors
10.1 Seismic Input Connectors 10.1.1 12 and 24 Channel Terraloc Pro Connector type: Cannon NK-27-32P Panel connector (mating side) (fits to NK-27-21C-1/2 " cable connector)

Figure 66 Input Connector 12- and 24-channel Terraloc Pro

Connector 1-12

Pin

Channel

1

1+

2

1-

3

2+

4

2-

"

"

"

"

23

12+

24

12-

Connector 13-24

Pin

Channel

1

13+

2

13-

3

14+

4

14-

"

"

"

"

23

24+

24

24-

10.1.2 48 Channel Terraloc Pro Connector type:
86

ABEM Terraloc Pro
Cannon KPT-02-A22-55P Panel connector (mating side) (fits to KPT-06 A22-55S cable connector)

Figure 67 Input Connector 48-channel Terraloc Pro

Connector 1-24

Pin

Channel

A

24+

B

24-

C

23+

D

23-

"

"

"

"

Z

13+

a

13-

b

12+

c

12-

"

"

"

"

z

1+

AA

1-

Connector 25-48

Pin

Channel

A

25+

B

25-

C

26+

D

26-

"

"

"

"

Z

36+

a

36-

b

37+

c

37-

"

"

"

"

z

48+

AA

48-

87

ABEM Terraloc Pro
10.2 Power Connector Connector type: Tyco Electronics 788189-02 (mating side), (fits to 788188-1 cable connector).
(Mating side view)
Figure 68 Power Connector 1 Positive power supply (+) 2 Negative power supply (-) 10.3 TTL Arm/Trig Connector Connector type: KPT 02-E10-6P (fits to KPT 06-E10-6S cable connector.)
(Mating side view)
Figure 69 TTL Arm/Trig Connector
A Trigger Output B Arm Input C Trigger Input D GND (Ground) E No Connection F Arm Output
TTL stands for Transistor-Transistor-Logic. It is used in connection to digital signals. A digital signal is considered to be either a logical 0 or a logical 1 (hereafter only called 0 and 1). Physically a 0 corresponds to a voltage of 0-0.7 V, while a 1 corresponds to a voltage of 2.8-5.0 V. Alternatively, a 0 might be called "low", and a 1 called "high". 10.4 Alarm Connector Connector type: KPT 02-E08-4P (fits to KPT 06-E08-4S cable connector.)
88

ABEM Terraloc Pro (Mating side view)
Figure 70 Alarm Connector A Alarm output +12V at alarm (output max. 0.5A) B Alarm relay output pole 1 C Alarm relay output pole 2 D Alarm output ground
10.5 Cascade Connector Connector type: KPT 02-E12-8P (fits to KPT 06-E12-8S cable connector.)
(Mating side view)
Figure 71 Cascade Connector A Tx+ 1 B Rx+ 1 C Tx+ 2 D Rx+ 2 E Tx- 1 F Rx- 1 G Tx- 2 H Rx- 2 The Cascade connector is used when measurements are done with several Terraloc Pros. Connection between instruments is made with a special Cascade connection cable.
89

ABEM Terraloc Pro
11 Appendix C. SeisTW Installation
Terraloc Pro is delivered with all necessary software installed at the factory. However, if the software needs to be updated, or re-installed, the procedure is described below. When the installation has finished, verify that the SeisTW version is correct, and that the device driver and API version are identical. This information can be found in the About dialog (press <CTR>+<HLP>). 11.1 Install Procedure for SeisTW Copy the setup file "SetupSeisTW-x.x.x.xxxx.exe" (the x:s represents the version number) to the hard drive of the Terraloc Pro and double-click the file to start the setup procedure. Follow the instructions in the setup wizard and verify that all three boxes are checked to install Terraloc Pro drivers, SeisTW and examples (Figure 72).
Figure 72 SeisTW Setup ­ Component choice
90

ABEM Terraloc Pro Now click on the next button, input the serial number of the instrument and set the number of channels to the number of channels installed in the instrument (Figure 73).
Figure 73 SeisTW Setup ­ Configuration Then click on the Install button and the setup program will now install all necessary files for SeisTW (Figure 74).
Figure 74 SeisTW Setup ­ Installation ongoing
91

ABEM Terraloc Pro Click the Next button when the Installation Complete screen is displayed (Figure 75).
Figure 75 SeisTW Setup ­ Installation completed And then click the Finish button (Figure 76).
Figure 76 SeisTW Setup ­ Completing Click the Yes button to complete the installation procedure.
Figure 77 SeisTW Setup ­ Reboot The Terraloc Pro will now restart and automatically launch SeisTW.
92

ABEM Terraloc Pro
12 Appendix D. Printout Examples
Figure 78 Printout example ­ Stretch factor 3 (default) 93

ABEM Terraloc Pro
Figure 79 Printout example ­ Stretch factor 8 (max) 94

ABEM Terraloc Pro

13 Appendix E. The First Arrivals File Format (FIR)

13.1 General
This is an ASCII text file format, containing first arrivals for a record. The file is formatted, so it can be printed on any printer that prints ASCII text. Trace number 1 always starts on line 22. The last trace is succeeded by a line that contains dash characters (-) only. If you are going to edit this file, be sure to keep the proper format.
Following the format for this type of files it is of course possible create new first arrival files manually using a text editor. Be sure however, to save the text in ASCII format and not in any word-processing format (e.g. WordPerfect or Microsoft Word).

13.2 Description

<BEGINNING-OF-FILE>

1: Creator: <Instrument - Program version> [Serial number: <xxxxxxx>]

2:

3: Record:

<record ID>

Date: <DD/MMM/YYYY>

4: Sampling interval:

<x.xxx ms>

Time: <HH/MM/SS>

5: Number of samples:

<xxxx (xxxx.x ms)>

6: Delay:

<xxx ms>

7: Highpass:

<OFF|xx dB/octave, xx cutoff Hz>

8: Notch:

<OFF|ON>

9: Digital filters:

<None|Bandpass|Highpass...>

10: [Low|high cutoff:

<xx Hz, xx dB/octave>]

11: [high cutoff:

<xx Hz, xx dB/octave>]

12:

13:==================================================================

14:

15: SHOT LOCATION: X = <x.xxx m> Y = <x.xxx m> Z = <x.xxx m>

16:

17:

18: RECEIVER LOCATIONS AND FIRST ARRIVALS:

19: -----------------------------------------------------------------

20: Trace X (m) Y(m) Z(m) First arrivals (ms)

21: -----------------------------------------------------------------

22: 1

x.xxx x.xxx x.xxx

x.xxx

23: 2

x.xxx x.xxx x.xxx

x.xxx

24: 3

x.xxx x.xxx x.xxx

x.xxx

. .

.

.

.

.

. .

.

.

.

.

. .

.

.

.

.

. .

.

.

.

.

. .

.

.

.

.

21+N: N

x.xxx x.xxx x.xxx

x.xxx

22+N: ---------------------------------------------------------------

<END-OF-FILE>

<> Represents a value

[] Denotes optional text

|

Delimits possible values

95

ABEM Terraloc Pro

14 Appendix F. Seismic Methods

There are a variety of seismic methods used. The objective of the survey controls which specific method to use. This section will give you an overview of some commonly used methods. Please refer to the bibliography at the end of the manual. If a more detailed and thorough description of seismic methods is needed then Butler (2005) can be recommended as it is relatively new and has an extensive and updated listing of references.
In the refraction and reflection methods there is usually a division between shallow and deep surveys.

14.1 Refraction
The objective is to find out the arrival times of the head waves to map the depth to the refractors in which the waves travel. The refraction method is based on the assumption that the earth is made of layers of materials that increase in seismic velocity with each successively deeper layer. The key element is that an incident ray is critically refracted along the boundaries between layers, before returning to the surface. From the first arrival times it is possible to calculate the seismic velocity for each layer and the depth to the boundaries. The seismic velocity gives information about material properties, and what kind of material comprises each layer. Additionally, frequency analysis of the recorded signals can give more information about the material properties.
The principles for seismic refraction techniques can be found in most geophysical textbooks. For a more detailed description of both theory and practice, see Sjögren (1984).
Investigations performed with the refraction method can yield a variety of reliable data such as depth of various overburden layers, depth to bedrock, rock quality, soil compositions and solidity, rip ability, excavatability, water tables and rock structure.
The refraction seismic method can be used for a wide range of applications, for example:

Underground
Foundations
Excavations Resource searches Water prospecting
Ore prospecting

Tunnels and their entrances, machinery halls, gas and oil storage facilities, air raid shelters
Heavy industrial buildings, bridges, harbor quays and breakwaters, dams, piling, airfields
Harbor basins and entrances, pipelines, canals, roads, railways
Gravel, sand and quarry sites
Groundwater table in the overburden, water bearing sections of rock
Mineralized weathered zones, buried channels with high mineral content

96

ABEM Terraloc Pro
14.2 Reflection In this method, the arrival time events are attributed to seismic waves that have been reflected from interfaces where changes in acoustic impedance occur, and of wave shape changes.
The seismic reflection method has mainly been used for deep investigations (depth > 30 m) in oil prospecting. During recent years however, shallow reflection investigations have become common for engineering and environmental purposes. It is now an important complement to refraction investigations, and has even sometimes replaced refraction. The main reasons for the increase in use of the reflection method is the development of lightweight, high-performance seismographs and the possibility of advanced data processing on inexpensive personal computers. Thus, the cost for reflection investigations has decreased considerably.
Both acquisition and processing of reflection data are more complex and time consuming than they are for refraction data.
14.3 Optimum Offset This is a special case of the seismic reflection method, in which data are recorded with a fixed source-receiver offset. It is a method for shallow investigations. The offset is chosen to be an optimum value (hence the name), and typically, it is a window where the reflection from the target is located between the refracted first arrivals and the ground roll in the seismogram.
14.4 Tomography The general idea for tomography is that information about the properties of the interior of a region can be obtained through measurements at the boundary. Thus, this is a method for finding the (2-dimensional) distribution of some physical property (e.g. velocity, reflectivity, bulk modulus, etc.). It can involve borehole-to-borehole, surface-to-borehole, or surface-to-surface measurements. The main restriction is that the source and receiver positions, and hence any boreholes, must be confined to the same plane. This plane can have any orientation.
Usually the travel times for a large number of ray paths through the rock volume is measured and, sometimes, even amplitudes (direct or reflected) are analyzed. Then the dataset goes through an inversion process where the spatial distribution of the physical property is estimated. The technique is very computational intensive and is costly because of the need for boreholes.
The final results are usually presented as maps or plots where the values of the physical property are coded in color or grayscale.
An introduction to this method can be found in Worthington (1984).
14.5 VSP VSP is short for Vertical Seismic Profiling, i.e. measurements with the receivers located in a borehole and the source located on the ground. If the source is moved away from the head of the borehole, it is called "offset VSP". In "Reversed VSP", the receivers are located on the ground and the source is located in the borehole.
The VSP technique is seldom used alone, but is rather used to provide better interpretation of seismic reflection data. VSP allows accurate determination of one-
97

ABEM Terraloc Pro way travel time to various geologic units and analysis of attenuation and acoustic impedances, which are needed for construction of synthetic seismograms. A brief introduction to this method is given in Cassel (1984). 14.6 Vibroseis Vibroseis is a seismic method in which a vibrator is used as an energy source to generate a controlled wave train, instead of the usual impulsive sources (e.g. hammer, explosives, shot guns, etc.). This method requires recording of the source signal for reference. A sinusoidal vibration of continuously varying frequency is applied during a sweep period typically lasting for several seconds (>10 s.). The sweep may start at either low, or high frequencies, and it can be linear or nonlinear. The recorded data, comprising many super positioned wave trains, has to be correlated with the source signal. The correlated record resembles a conventional seismic record such as results from an impulsive source.
98

ABEM Terraloc Pro
15 Appendix G. Bibliography
Butler, Dwain K., Editor. 2005. Near Surface Geophysics. Volume 13 in Series: Investigations in Geophysics. Society of Exploration Geophysicists. ISBN 1-56080-130-1.
Cassel, Bruce. 1984. Vertical Seismic Profiling - an Introduction. First Break Vol. 2 No. 11. European Association of Exploration Geophysicists.
Palmer, Derecke. 1980. The Generalized Reciprocal Method of Seismic Refraction Interpretation. Society of Exploration Geophysicists. ISBN 0-931830-14-1.
Parasnis, D.S. 1986. Principles of Applied Geophysics. 4th Ed. Chapman and Hall. ISBN 0-412-283301.
Penoyer, Robert 1993. The Alpha-Beta filter. The C Users Journal Vol. 11 No. 7 (73-86). R&D Publications.
SEG 1990. Seismic (/Radar) Files in Personal Computers Standard. Engineering and Groundwater Geophysics Subcommittee, Society of Exploration Geophysicists. ISBN 1-56080-020-8
Sheriff, Robert E. 1991. Encyclopedic Dictionary of Exploration Geophysics. 3rd Edition. Society of Exploration Geophysicists. ISBN 0-931830-47-8 (Series), ISBN 1-56080-018-6 (Volume).
Sjögren, Bengt. 1984. Shallow Refraction Seismics. Chapman and Hall. ISBN 0-412-24210-9. Ward, Stanley H., Editor. 1990. Geotechnical And Environmental Geophysics Vol I-III. Volume 5 in
Series: Investigations in Geophysics (Edwin B. Neitzel, Series Editor). Society of Exploration Geophysicists. ISBN 0-931830-99-0. Worthington, M.H. 1984. An Introduction To Geophysical Tomography. First Break Vol. 2 No. 11. European Association of Exploration Geophysicists. Yilmaz, Özdogan. 1987. Seismic Data Processing. Volume 2 in Series: Investigations in Geophysics (Edwin B. Neitzel, Series Editor). Society of Exploration Geophysicists. ISBN 0-931830-40-0
99


Microsoft Word 2013 Microsoft Word 2013