User Guide for ResMed models including: Astral Series, Invasive and Non-Invasive Vantilator

Quick Setup Guide

ResMed COVID-19: Our response

Quick start guide


File Info : application/pdf, 13 Pages, 2.57MB

PDF preview unavailable. Download the PDF instead.

astral-quick-setup-guide-row-eng
Quick start guide
English

Read your entire AstralTM User Guide or Clinical Guide before use.
The Astral device

1 2

3

4 56

78

1. Adapter port Can be fitted with single limb adapter, single limb leak adapter or double limb adapter (Astral 150 only)
2. Handle
3. Inspiratory port (to patient)
4. Ethernet connector (service use only)
5. USB connector (for download to ResScan and connection of approved accessories)
6. Mini USB connector for ResMed Connectivity Module (RCM) or ResMed Connectivity Module Hospital (RCMH)
7. DC power inlet
8. Device on/off push button
9. SpO2 Sensor connector 10. Remote alarm five pin
connector
11. Low flow oxygen input (up to 30L/min)
12. Air inlet (complete with hypoallergenic filter)
Note: Some accessories may not be available in all regions.

12

11

10 9

1

The Astral device interface

1

5

4

Touch screen

1

2

3

11 10

9

8

2

1. Touch screen

2. Power source indicators

AC (mains power supply) DC (external battery, car accessory adapter or ResMed Power Station II) Internal battery
3. Therapy on/off indicator

Device ready

Constant green display when the device is turned on but not ventilating.

Device ventilating

Flashes blue when the

device is ventilating and the

Ventilation LED setting is

`ON'. Otherwise is `OFF'

3

4. Alarm mute/reset button

Illuminates when an alarm is

triggered and flashes when

the sound is muted.

5. Alarm bar

45

1. Clinical mode access button

Locked

Unlocked

2.

Manual breath

button only shown

if enabled

6 3. Information bar 4. Internal battery indicator

5.

Lock touch screen button

6. Menu bar

7. Bottom bar
7
8. Start ventilation button

Stop ventilation button 9. Main screen 10. Sub menus 11. Pressure bar
2

Patient circuits
The Astral device supports single and double limb circuits with the use of interchangeable circuit adapters. Breathing circuits may be of 10, 15 or 22 mm diameter. For recommended patient circuit components and compatible accessories, refer to www.resmed.com/astral/circuits.
There are three circuit adapters:

Adapter
1. Single limb leak

For use with
Single limb circuit with intentional leak or Mouthpiece circuit

2. Single limb
3. Double limb (Astral 150 only)

Single limb circuit with expiratory valve (expiratory valve integrated into the circuit)
Double limb circuit (expiratory valve integrated into the adapter) OR single limb circuit with intentional leak
or Mouthpiece circuit

Fitting the circuit adapter
Before connecting the patient circuit, the adapter specific to the required circuit type must be fitted. To fit the adapter:

1

2

3

4

5

1. Turn over the device and place on a soft surface (to protect the LCD screen).
2. Press and hold the eject button. Pull the cover out towards you.
3. Lift the adapter out of the socket.
4. Replace with the new adapter, ensuring it sits firmly in the socket.
5. Place the cover over the enclosure, ensuring the runners on the device and the cover are aligned. Slide the cover back into place until the latch clicks.

3

Setting up the circuit
For complete details on setting up the circuit see Assembling patient circuits in the Astral Clinical Guide.

Single limb circuit with intentional leak or Mouthpiece circuit

Single limb circuit with expiratory valve using ResMed proprietary circuit

Single limb circuit with expiratory valve using standard circuit

Double limb circuit (Astral 150 only) or single limb circuit with intentional leak and Mouthpiece circuit

To patient
Performing a Learn Circuit
4

Do not connect patient interfaces prior to performing the Learn Circuit.
1. From the Setup menu, select the Circuit sub menu.
2. Press Start and follow the on-screen prompts.
A Results screen is displayed once Learn Circuit is complete. Error codes are provided if Learn Circuit fails and can be found in the Clinical Guide for help with troubleshooting. The previous test results can be accessed by selecting the Results button in the Circuit sub menu.

Powering on the device

To power on the device, simply press the green power on/off switch at the back of the device. The device will perform a system check.
On completion of the system check, the Patient Home screen and active program is displayed.
If more than one program displays, the active program will be highlighted orange.
Note: Settings configured in the active program will be used when ventilation is started.

Accessing Clinical mode

To access Clinical mode:
Clinical mode can be accessed from any screen, whether or not the Astral device is ventilating.
1. From the patient home screen, press and hold for 3 seconds.
2. Select 20 minutes or Unlimited. The padlock is unlocked and the Main Settings screen is displayed.

To exit Clinical mode:

1. Press

. The Exit Clinical Mode screen

is displayed.

2. Press Confirm. The padlock is locked and the Patient Home screen is displayed.

Note: If you do not make a selection within 7 seconds, the device returns to the previous screen.

5

Using the Setup Assistant

To quickly set up the Astral device and start ventilation for a new patient, use the Setup Assistant.

1. Access Clinical mode.

2. From the main menu, press .

3. Select Setup Assistant and press Start.

4. A warning message will display. Select Continue.

5. Select the patient type to automatically configure alarm and setting ranges.

6. Select the circuit type to be connected.

7. Select Start to run the Learn Circuit.

8. Follow the on-screen prompts to attach and test the circuit.

9. Select your required ventilation mode. The default settings for the mode are displayed.

10. Review and adjust the settings as required.

11. Press

to start ventilation.

For further information on adjusting setting parameters and alarm settings, see Adjusting patient settings and Adjusting alarm settings in the Astral Clinical Guide.

6

Navigating the menus
Monitors

Waveforms Monitoring Trends

Displays the last 15 seconds of patient airway pressure and flow in a graph. The graph updates in real-time.
Displays all measured parameters in numerical form.
Displays trending of various parameters for the last 30 days.

Setup Alarms Information

Programs*

Setup Assistant*

Circuit

Settings

Main Settings Manual Breath Sigh Breath

USB

Patient Data

Settings

Upgrade**

Device Config.

Device Date / Time Units Patient access

Alarm 1 & 2

Tidal Volume (Vt) Minute Ventilation (Mv) Respiratory Rate Pressure Low PEEP Ventilation Stop Leak NV mask / Rebreathing

Alarm 3

FiO2 SpO2 Pulse

Apnoea Response

Apnoea Response Apnoea Detection

Alarm Volume Alarm Level Alarm Minimum Volume Alarm Test

Disconnection Disconnection Alarm Alarm

Events

Alarms Settings System Delete*

Device

Basic Advanced

Battery

Charge Maintenance

*Available in Clinical mode only. **Software upgrades can only be performed by an authorised service representative. Astral has the ability to be upgraded with a USB stick containing the appropriate software. Please contact your authorised agent for further information.
7

Enabling additional programs

The Astral device comes factory preset with one active program. Additional programs can be enabled.

1. Access Clinical mode.
2. From the Setup main menu, select Programs.
3. Enable additional programs by pressing the slider.

Note: The current active program is highlighted in orange and cannot be disabled.
4. Rename the program by pressing the Rename button.

Within the Rename window, a name for the current active program can be chosen from the list provided (eg, DAY).

Note: A selected name can be removed

by pressing the delete button

.

5. When selected, the program name is displayed on the program label and on the information bar when that program is active.

8

Enabling the `Big buttons' mode

The Astral device offers an enhanced access feature (`Big buttons' mode) to provide easier usability and accessibility for starting and stopping ventilation as well as Alarm mute.
The `Big buttons' mode can be switched on and off as required in either Clinical or Patient mode.
1. From the main menu, press .
2. Select the Patient Access tab from the Device Config. menu.
3. Move the Big buttons slider to On.
Your enhanced access feature is now enabled.

9

Alarms
When an alarm is activated, the Astral device provides both audible and visual alerts, and displays an alarm message in the Alarm display on the Information bar.

Alarm priority High Medium
Low

Alarm bar Red flashing light
Yellow flashing light

Audible alert
10 beeps every 5 seconds
3 beeps every 15 seconds

Yellow steady

2 beeps every 25 seconds

Indicator
1. Alarm display

Description
Displays either the alarm message for the highest priority active alarm, or the last alarm not yet reset.
Press the Alarm display for further alarm information.
Certain conditions may result in
multiple alarms. indicates that there are multiple active alarms.

Press when displayed to view all active alarms and respond appropriately. Alarms are displayed in order of priority.

2. Active Alarms screen

Displays the full set of active alarms. Will automatically display upon activation of an alarm in patient mode.

3. Information menu

Some alarms clear automatically. To view a history of alarms, view the alarm log through the Information menu.

4. Alarm mute/reset button

State: · no light ­ no active alarms · steady light ­ active alarm/s · flashing light ­ alarm mute
on.
This button also allows you to: · mute the audible alert · reset the currently displayed
alarm (if permitted).

5. Alarm bar Indicates the priority of the alarm in the Alarm display.

10

Attaching a humidifier

1. Connect a length of air tubing to the inspiratory port of the device.
2. Connect the other end of the air tubing to the inlet port of the humidifier.
3. Connect the patient circuit to the outlet port on the humidifier.

Attaching a Heat Moisture Exchange (HME)
Too ppaattieiennt t
Attaching an antibacterial filter

An HME can be used with the Astral device with a double limb circuit or single limb circuit with integrated valve. Place the HME between the patient end of the circuit and the patient interface. Do not connect patient interfaces prior to performing the Learn Circuit.
1. Fit the antibacterial filter to the inspiratory port of the device.
2. Connect the air tubing to the other side of the filter.
3. Perform the Learn Circuit function.
4. Attach the patient interface to the free end of the air tubing.
For complete details on attaching patient circuit accessories, see Accessories in the Astral Clinical Guide.

For a full list of accessories, see the Ventilation accessories guide on www.resmed.com. If you do not have internet access, please contact your ResMed representative.

11

ResMed Ltd 1 Elizabeth Macarthur Drive, Bella Vista NSW 2153 Australia See ResMed.com for other ResMed locations worldwide. Astral is a trademark and/or registered trademark of the ResMed family of companies. For patent and other intellectual property information, see ResMed.com/ip. © 2018 ResMed Ltd. 278119/3 2018-07
ResMed.com

