User Guide for Honeywell models including: FP5000, Pressure Transducer
MODEL FP5000 Configurable Pressure Transducer 008927 Issue 3 DESCRIPTION The Honeywell Model FP5000 Series is a media-isolated piezoresistive silicon
MODEL FP5000 High Performance Configurable Pressure Transducer 008927 Issue 3 DESCRIPTION The Honeywell Model FP5000 Series is a media-isolated piezoresistive
File Info : application/pdf, 7 Pages, 475.62KB
DocumentDocumentMODEL FP5000 High Performance Configurable Pressure Transducer 008927 Issue 3 DESCRIPTION The Honeywell Model FP5000 Series is a media-isolated piezoresistive silicon pressure sensor offering multiple output options for reading pressure over the specified full-scale pressure span and temperature range. It is compensated for sensor offset, sensitivity, temperature effects and non-linearity to offer improved thermal stability and accuracy. Hastelloy® C276 and 316L stainless steel wetted parts provide durability with abrasive or corrosive media. DIFFERENTIATION · Offers improved accuracy and reliability · Configurable platform enables a sensor to be built to customer requirements. Simplified nomenclature and order codes make ordering easier · Many pressure and operating temperature range options · Built from stocked components; most configurations are shipped within ten business days · Extensive history of pressure measurement know-how · Bi-directional differential output available FEATURES · Pressure ranges from 10 in-H2O [0.36 psi] up to 5000 psi · Gage, absolute, vacuum, barometric, compound, and differential wet/wet pressure types · Higher accuracy to 0.1 %FSS BFSL · Multiple output types: 0 Vdc to 5 Vdc, 0 Vdc to 10 Vdc, 4 mA to 20 mA, 5 Vdc ±5 Vdc, 12 mA ±8 mA · Multiple electrical and pressure connection options · Zero adjustment through potentiometer · Operating temperature ranges from -40°C to 125°C [-40°F to 250°F] · Multiple compensation temperature ranges · Faster response and higher resolution · Fully analog reduced-noise signal path provides continuous output resolution · Stainless steel construction · Ha C276 and 316L stainless steel wetted parts offer more enhanced durability with abrasive or corrosive media · CE, RoHS, REACH compliant · Intrinsic Safety Certifications: cFMus, ATEX, IEC Ex certified 2AR option (4 mA to 20 mA) VALUE TO CUSTOMERS · Built on the Honeywell history of higher-quality pressure sensing technologies · Next-gen design of the popular FP2000 pressure sensor · Offers more repeatable, reliable, and accurate pressure measurements over time · Rugged, stainless steel pressure sensors are built and tested to perform and survive in many demanding environments · Configurable platform creates a wide range of standard configurations · Stocked components enable shipping within ten business days on most configurations APPLICATIONS · Test stands (Automotive, Aerospace, Industrial, and Medical) · R&D test labs · Hydraulic and pneumatic system monitoring · Leak detection · Manufacturing mold pressure control · Pump and compressor control · Liquid level measurement · Oil & gas process control · Fluid flow measurement · Valve testing PORTFOLIO Model FP5000 pressure transducers are part of a comprehensive line of Honeywell pressure sensors. HIGH PERFORMANCE CONFIGURABLE PRESSURE TRANSDUCER MODEL FP5000 TABLE 1. PERFORMANCE SPECIFICATIONS Characteristic Measure Operating pressure ranges Accuracy1,7 Gage: 0.36 psi to 5000 psi Absolute: 5 psi to 5000 psi Vacuum: 0.36 psi to 15 psi Barometric: 0 to 30 in-Hg, 16 to 32 in-Hg, 26 to 32 in-Hg Compound ranges available consult factory Differential: 0.36 psi to 1000 psi Equivalent ranges are available in other pressure units also: kPa, bar, mm-Hg, in-Hg, mbar, torr, in-H2O 0.2 %FSS BFSL (Standard accuracy) 0.1 %FSS BFSL (High accuracy) Output (selectable) 0 Vdc to 5 Vdc, 0 Vdc to 10 Vdc, or 4 mA to 20 mA (two wire) Resolution Continuous (Fully analog signal path) TABLE 2. ENVIRONMENTAL SPECIFICATIONS Characteristic Measure Operating temperature range See Table 3 (Electrical connectors) Compensated temperature range See Table 4 (Thermal effects error band) Thermal effects error band (TEEB)2,3 See Table 4 (Thermal effects error band) Sealing See Table 3 (Electrical connectors) Notes: 1. Accuracies stated are with respect to best fit straight line (BFSL) for all errors including linearity, hysteresis, and non-repeatability through zero. 2. Thermal Effects Error Band - The maximum deviation in output due to changes in temperature over the entire compensated temperature range, relative to output measured at reference temperature. Includes all errors due to: Thermal Effect on Offset and Thermal Effect on Span. 3. Thermal effects error band (TEEB) increases pro-rata for pressure ranges below 5 psi [0.35 bar]. 4. True Zero Output: The voltage output versions have onboard circuitry that allows the output signal to swing all the way to ground (True Zero) and even a little below (~-0.2 V). This mitigates increased error at lower voltage measurements. 5. Over pressure: The absolute maximum rating for pressure which may be safely applied to the product for it to remain in specification once pressure is returned to the operating pressure range. Exposure to higher pressure may cause permanent damage to the product. 6. Burst pressure: The maximum pressure that may be applied to the product without causing escape of the pressure media. The product should not be expected to function after exposure to any pressure beyond the rated burst pressure. 7. All specifications apply at 25°C [77°F] and under operating conditions unless otherwise noted. 8. Full Scale Span (FSS): The algebraic difference between output signal measured at the upper and lower limits of the operating pressure range. Also known as "span". 9. Offset: The output signal obtained when the reference pressure is applied to all available pressure ports. Also known as "null" or "zero". 10. Reference pressure: The pressure used as a reference (zero) in measuring product performance. Unless otherwise specified, this is vacuum (0 psia) for absolute pressure sensors and local ambient atmospheric pressure (0 psig) for gage/vacuum pressure sensors. 11. Minimum operating pressure: The lower limit of the operating pressure range. 12. Maximum line pressure: The maximum pressure that can be applied simultaneously to both ports of a Differential Pressure Sensor without causing changes in specified performance 13. Line pressure effect on zero: The change in the output at zero pressure caused by change in line pressure on both ports of differential pressure sensor. TABLE 3. ELECTRICAL CONNECTORS Connector Operating Temperature Range PT-02A-10-6P -40°C to 125°C [-40°F to 250°F] DIN FORM A -40°C to 125°C [-40°F to 250°F] DIN FORM C -40°C to 90°C [-40°F to 194°F] Integral cable -40°C to 105°C [-40°F to 221°F] Conduit fitting -40°C to 105°C [-40°F to 221°F] M12 x 1, 4-pin -40°C to 85°C [-40°F to 185°F] Sealing IP67 IP65 IP65 IP67 IP67 IP67 TABLE 4. THERMAL EFFECTS ERROR BAND (TEEB) Compensated For Standard Temperature Range Accuracy For High Accuracy 0° C to 60° C [40° F to 140° F] < ±0.75 %FSS < ±0.5 %FSS -20° C to 80° C [0° F to 180° F] < ±1.5 %FSS < ±1 %FSS -40° C to 85° C [-40° F to 185° F] < ±2.25 %FSS < ±1.5 %FSS -40° C to 125° C [-40° F to 250° F] < ±2.25 %FSS < ±1.5 %FSS TABLE 5. MECHANICAL SPECIFICATIONS Characteristic Measure Media Gas, liquid Weight (approx.) gage, absolute, vacuum, barometric, compound 150 g [5.3 oz] Weight (approx.) differential 300 g [10.6 oz] Wetted parts material Ha C276 and 316L stainless steel Labels Laser engraved Maximum allowed line pressure12 Operating range < 15 psi (1 bar): 50 psi 15 psi (1 bar) < Operating range 50 psi (3.5 bar): 250 psi 50 psi (3.5 bar) < Operating range 250 psi (17 bar): Pressure range: +500 psi 250 psi (17 bar) < Operating range 1000 psi (70 bar): Pressure range: +1000 psi Line pressure effect on zero13 Operating range < 15 psi (1 bar): ±1 % FSS 15 psi (1 bar) < Operating range 50 psi (3.5 bar): ±1 % FSS 50 psi (3.5 bar) < Operating range 250 psi (17 bar): ±1.5 % FSS 250 psi (17 bar) < Operating range 1000 psi (70 bar): ±1.5 % FSS Model FP5000 High Performance Configurable Pressure Transducer Datasheet | sps.honeywell.com/ast | 2 HIGH PERFORMANCE CONFIGURABLE PRESSURE TRANSDUCER MODEL FP5000 TABLE 6. ELECTRICAL SPECIFICATIONS7 Specifications 2AM, 2AR 2AY 4 mA to 20 mA 12 mA ±8 mA 2AN 0 V to 5 V 2AP 0 V to 10 V 2AW 5 V ±5 V Supply voltage 9 Vdc to 28 Vdc 9 Vdc to 28 Vdc 14 Vdc to 28 Vdc Current consumption 4 mA to 24 mA < 6 mA Output at reference pressure 10 (absolute, gage, vacuum, differential) Output at minimum operating pressure11 - (compound, barometric) 4 mA ±0.5 %FSS 12 mA ±0.5 %FSS 0 V ±0.5 %FSS 0 V ±0.5 %FSS 5 V ±0.5 %FSS Full scale span (FSS)8 16 mA ±1 %FSS 16 mA ±1 %FSS 5 V ±1 %FSS 10 V ±1 %FSS 10 V ±1 %FSS Frequency response 3500 Hz Reverse voltage protection Yes, 28 V Load impedance < 950 Ohm at 28 V decreasing linearly to 0 Ohm at 9 V > 10K Ohms Insulation resistance >500 MOhm to case GND at 33 V Overvoltage protection >32 V Power up time < 1 sec Zero adjustment potentiometer Yes, > ±5 %FS adjustment, accessible from top after demounting connector 2AR not available for differential pressure sensing 2AY and 2AW are available only for bi-directional differential pressure sensing TABLE 7. OVERPRESSURE AND BURST PRESSURE RATINGS Pressure type/Port Operating Range Over Pressure Burst Pressure Gage, absolute, vacuum, barometric, compound, differential version positive (high side) 15 psi (1 bar) 15 psi to 1000 psi (1 bar to 70 bar) > 1000 psi (70 bar) 6X FS 4X FS 3X FS or 10,000 psi (700 bar) whichever is less 10X FS 6X FS 4X FS or 10,000 psi (700 bar) whichever is less Differential version negative (low side) with respect to positive (high side) 15 psi (1 bar) 15 psi to 1000 psi (1 bar to 70 bar) 6 X 10X 4X FS or 250 psi whichever is less 6X FS or 500 psi whichever is less Model FP5000 High Performance Configurable Pressure Transducer Datasheet | sps.honeywell.com/ast | 3 HIGH PERFORMANCE CONFIGURABLE PRESSURE TRANSDUCER MODEL FP5000 TABLE 8. DIN FORM A (6M), DIN FORM C (6BO) WIRING STANDARD PIN 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation 0 V to 5 V/0 V to 10 V/ 5 V ± 5 V (2AN, 2AP, 2AW) Designation 1 (+) Supply (+) Supply 2 (+) Output (+) Output 3 No connection Supply return/ (-) Output E No connection No connection ALTERNATIVE 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation (+) Supply (+) Output No connection Case GND 0 V to 5 V/0 V to 10 V/5 V ± 5 V (2AN, 2AP, 2AW) Designation (+) Supply Supply return/ (-) Output (+) Output Case GND TABLE 9. PT02A-10-6P, 6-PIN (6A) WIRING STANDARD PIN 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation 0 V to 5 V/0 V to 10 V/ 5 V ± 5 V (2AN, 2AP, 2AW) Designation A (+) Supply (+) Supply B No connection Supply return C No connection (-) Output D (+) Output (+) Output E No connection No connection F No connection No connection ALTERNATIVE 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation (+) Supply (+) Output No connection No connection No connection No connection 0 V to 5 V/0 V to 10 V/5 V ± 5 V (2AN, 2AP, 2AW) Designation (+) Supply (+) Output No connection Supply return/ (-) Output No connection No connection TABLE 10. INTEGRAL CABLE (6Q), CONDUIT FITTING (6R) WIRING STANDARD ALTERNATIVE PIN 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation 0 V to 5 V/0 V to 10 V/ 5 V ± 5 V (2AN, 2AP, 2AW) Designation 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation Red (+) Supply (+) Supply (+) Supply Black (+) Output Supply return Not available Green Not available (-) Output Not available White Not available (+) Output (+) Output 0 V to 5 V/0 V to 10 V/5 V ± 5 V (2AN, 2AP, 2AW) Designation (+) Supply Supply return/ (-) Output Not available (+) Output TABLE 11. M12 X 1, 4-PIN (6BJ) WIRING STANDARD PIN 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation 0 V to 5 V/0 V to 10 V/ 5 V ± 5 V (2AN, 2AP, 2AW) Designation 1 (+) Supply (+) Supply 2 No connection (+) Output 3 (+) Output Supply return/ (-) Output 4 Case GND Case GND ALTERNATIVE 4 mA to 20 mA / 12 mA ±8 mA (2AM, 2AR, 2AY) Designation (+) Supply (+) Output No connection Case GND 0 V to 5 V/0 V to 10 V/5 V ± 5 V (2AN, 2AP, 2AW) Designation (+) Supply No connection Supply return/ (-) Output (+) Output TABLE 12. INTRINSICALLY SAFE APPROVALS FOR OPTION 2AR (NOT AVAILABLE ON 2AM, 2AN, 2AP, 2AY, 2AW) Agency Approvals cFMus Class I, Div 1, Groups A, B, C, D Class I, Zone 0, AEx/Ex ia IIC T4/T5 Ga Ta= -40°C to 40°C (T5), -40°C to 85°C (T4) ATEX II 1 G Ex ia IIC T4/T5 Ga Ta= -40°C to 40°C (T5), -40°C to 85°C (T4) IEC Ex Ex ia IIC T4/T5 Ga Ta= -40°C to 40°C (T5), -40°C to 85°C (T4) UKCA II 1 G Ex ia IIC T4/T5 Ga Ta= -40°C to 40°C (T5), -40°C to 85°C (T4) (Pending Approval) See Honeywell's website (http://sps.honeywell.com/ast) for up-to-date information regarding intrinsically safe approvals. Refer to Installation manual #008-0751-00 for installation/wiring instructions, cautions and warnings. Model FP5000 High Performance Configurable Pressure Transducer Datasheet | sps.honeywell.com/ast | 4 HIGH PERFORMANCE CONFIGURABLE PRESSURE TRANSDUCER MODEL FP5000 Figure 1. Product Nomenclature NFG 1 BR, 1AK, 2AM, 5F, 6M, 7BA Type NFG Gage NFA Absolute NFV Vacuum NFB Barometric (UG, UQ, UR only) NFC Compound (consult factory) NDW Differential, wet/wet Accuracy 1 0.10 %* 2 0.20 % *Not available in ranges < 2 psi *Not available for 2AY and 2AW Range kPa bar KD 35 kPa KF 100 kPa KG 200 kPa KH 300 kPa KJ 700 kPa KL 1000 kPa KM 1500 kPa KQ 3000 kPa KR 5000 kPa KS 7000 kPa KT**10000 kPa in-H2O WA* 10 in-H2O WC* 20 in-H2O WE* 30 in-H2O WG* 50 in-H2O ME 1 bar MF 2 bar NA 3.5 bar MG 5 bar NB 7 bar MH 10 bar MI 20 bar MJ 30 bar NC 35 bar MK 50 bar ND 70 bar ML** 100 bar NE** 135 bar MY** 200 bar NG** 350 bar in-Hg UG 30 in-Hg UQ 16-32 in-Hg UR 26-32 in-Hg More ranges are available. Please consult Honeywell for options. psi AN* 0.5 psi AP* 1 psi AT 5 psi AV 10 psi BJ 15 psi BL 25 psi BM 30 psi BN 50 psi BP 75 psi BR 100 psi CJ 150 psi CL 200 psi CN 250 psi CP 300 psi CQ 400 psi CR 500 psi CS 600 psi CT 750 psi CV 1000 psi DJ** 1500 psi DL** 2000 psi DM** 2500 psi DN** 3000 psi DR ** 5000 psi * Not available in absolute pressure type ** Not available in differential pressure type Ranges with blue highlight (greater than 100 kPa, 15 psi, 1 bar) are NOT available in Vacuum. Calibration Data* 5-point calibration at 77°F (default option) 9A 9-point calibration at 77°F * Only positive side calibration data provided for differential sensors Wiring 7BA Standard 7BB Alternative Electrical Connection 6A PT02A-10-6P, 6-pin 6M DIN A 43650, 4-pin 6BO DIN C, 4-pin 6Q Integral cable, 5 feet# 6R Conduit fitting with 5 feet# cable 6BJ M12 x 1 (IEC 61076-2-101 style AM), 4-pin #Consult factory for other cable lengths Pressure Port 5A 1/4-18 NPT female 5B 1/4-18 NPT male 5D 7/16-20 UNF male 5F G1/4 B female 5G 5P 5AE 5I G1/4 B male M12 x 1.5 male 1/2-14 NPT male 1/8-27 NPT male Output 2AM 4 mA to 20 mA 2AN 0 Vdc to 5 Vdc 2AP 0 Vdc to 10 Vdc 2AR 4 mA to 20 mA Intrinsically Safe (see Table 11 for certifications) 2AY 12 mA ±8 mA 2AW 5 V ±5 V 2AR not available for differential pressure sensing 2AY and 2AW are available only for bi-directional differential pressure sensing Temperature Compensation 1AK 0°C to 60°C [40°F to 140°F] 1Y1 -20°C to 80°C [0°F to 180°F] 1AP1 -40°C to 85°C [-40°F to 185°F] 1BA2 -40°C to 125°C [-40°F to 250°F] 1 1Y and 1AP not available for pressure ranges below 5 psi [0.35 bar] 2 1BA not available for pressure ranges below 50 psi [3.5 bar] SAMPLE CATALOG LISTINGS Order Code Description NFA1BM,1AK,2AP,5A,6A,7BA Model FP5000; 0.10% accuracy; 30 psi absolute; compensated across 0°C to 60°C [40°F to 140°F]; 0 Vdc to 10 Vdc output; 1/4-18 NPT female port; PT02A-10-6P 6-pin electrical connector; standard wiring; 5-point calibration data at 77°F NFG2DR,1Y,2AR, 5G,6Q,7BB Model FP5000; 0.20% accuracy; 5000 psi gage; compensated across -20°C to 80°C [0°F to 180°F]; Intrinsic Safety Certifications; 4 mA to 20 mA output; G 1/4 B male port; 5 ft long integral cable; alternative wiring; 5-point calibration data at 77°F NDW1BR,1AK,2AM,5F,6A,7BA Model FP5000; 0.10 % accuracy; 100 psi differential; compensated across 0°C to 60°C [40°F to 140°F]; 4 mA to 20 mA output; G 1/4B female port, PT02A-10-6P 6-pin electrical connector; standard wiring; 5-point calibration data at 77°F Model FP5000 High Performance Configurable Pressure Transducer Datasheet | sps.honeywell.com/ast | 5 HIGH PERFORMANCE CONFIGURABLE PRESSURE TRANSDUCER MODEL FP5000 Figure 2. Mounting Dimensions Model FP5000 High Performance Configurable Pressure Transducer Datasheet | sps.honeywell.com/ast | 6 ELECTRICAL TERMINATION CODE 6BO: 4-PIN, MICRO DIN FORM C CODE 6M: 4-PIN, STANDARD DIN 43650 FORM A CODE 6A: 6-PIN, BENDIX CONNECTOR CODE 6Q: 4-CONDUCTOR, CODE 6R: 4-CONDUCTOR, CODE BJ: 4-PIN, M12 x 1 VENTED, INTEGRAL CABLE VENTED, 1/2-14 NPT CONDUIT (IEC 61076-2-101) STYLE AM EXIT WITH INTEGRAL CABLE 26,0 [1.1] Ø 31,8 [1.3] 10,4 [0.4] 7,9 [0.3] 18,2 [0.7] 8,8 [0.3] Ø 31,8 [1.3] 21,3 [0.8] n31,8 [1.3] 10,4 [0.4] (35,8 ) [1.4] Ø 25,0 [1.0] 41,1 [1.6] 10,4 [0.4] Ø 31,8 [1.3] 27,2 [1.1] Ø 31,8 [1.3] Ø 25,0 [1.0] 8,8 [0.3] Ø 25,0 [1.0] 37,1 [1.5] 37,1 [1.5] 22,2 [0.9] CHOOSE FROM PRESSURE PORTS CODES: 5F OR 5A 22,2 [0.9] 16,8 [0.7] PRESSURE PORTS CODE 5F: G 1/4 B FEMALE 22,2 [0.9] CODE 5A: 1/4-18 NPT FEMALE 22,2 [0.9] CODE 5G: G 1/4 B MALE 15,1 [0.6] CHOOSE FROM ELECTRICAL TERMINATION CODES: 6BO, 6M, 6A, 6Q, 6BJ, OR 6R Ø 31,8 [1.3] Ø 25,0 [1.0] CHOOSE FROM PRESSURE PORTS CODES: 5AE, 5G, 5P, 5B, 5D, OR 5I CODE 5B: 1/4-18 NPT MALE 18,4 [0.7] CODE 5D: 7/16-20 UNF MALE 16,8 [0.7] CODE 5AE: 1/2-14 NPT MALE 18,0 [0.7] Ø 25,0 [1.0] 27,0 [1.1] CODE 5P: M12 X 1.5 MALE 15,2 [0.6] 63,8 [2.5] 69,0 [2.7] CHOOSE FROM PRESSURE PORTS CODES: 5F OR 5A CODE 5I: 1/8-27 NPT MALE 13,2 [0.5] , CAUTION PRODUCT DAMAGE DUE TO MECHANICAL ISSUES · Ensure torque specifications are determined for the specific application. Mating materials and thread sealants can result in significantly different torque values from one application to the next. · When using mating parts made of stainless steel, use a thread sealant with anti-seize properties to prevent thread galling. Ensure the sealant is rated for the application. · Use appropriate tools (such as an open-ended wrench or deep-well socket) to install transducers. · Always hand-start transducers into the hole to prevent cross threading and damage. · Ensure that torque is not applied to the electrical connector. · Ensure that the proper mating electrical connector with a seal is used to connect the transducer. Improper or damaged seals can compromise ingress protection, leading to short circuits. · To ensure proper environmental sealing and electrical connections when using a connector, follow the connector manufacturer's installation guidelines. · All terminal cavities must be sealed using the correct wire gauge and seal combination. · If only two leads are used, any additional terminal cavities should be sealed per the connector manufacturer's installation guide. · Honeywell recommends using a crimping tool for crimping wires to the connector terminals. · Contact the individual connector manufacturer for connector wiring. Failure to comply with these instructions could result in product damage. , NOTICE Refer installation manual #008-0751-00 for installation/ wiring instructions, cautions and warnings related to Intrinsically Safe FP5000 (option 2AR). FOR MORE INFORMATION Honeywell Advanced Sensing Technologies services its customers through a worldwide network of sales offices and distributors. For application assistance, current specifications, pricing, or the nearest Authorized Distributor, visit sps.honeywell.com/ast or call: USA/Canada Latin America Europe Japan Singapore Greater China +302 327 8920 +1 305 805 8188 +1 302 327 8920 +81 (0) 3-6730-7152 +65 6355 2828 +86 4006396841 m WARNING PERSONAL INJURY DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury. Failure to comply with these instructions could result in death or serious injury. m WARNING MISUSE OF DOCUMENTATION · The information presented in this product sheet is for reference only. Do not use this document as a product installation guide. · Complete installation, operation, and maintenance information is provided in the instructions supplied with each product. Failure to comply with these instructions could result in death or serious injury. Warranty/Remedy Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship during the applicable warranty period. Honeywell's standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgment or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items that Honeywell, in its sole discretion, finds defective. The foregoing is buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages. While Honeywell may provide application assistance personally, through our literature and the Honeywell website, it is buyer's sole responsibility to determine the suitability of the product in the application. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this writing. However, Honeywell assumes no responsibility for its use. Honeywell Advanced Sensing Technologies 830 East Arapaho Road Richardson, TX 75081 sps.honeywell.com/ast Hastelloy® is a registered trademark of Haynes International, Inc. 008927-3-EN | 3 | 08/21 © 2021 Honeywell International Inc.