Fridababy 3-in-1 Ear, Forehead + Touchless Thermometer
USER MANUAL
INFRARED TEMP CHECKS IN 1 SEC
This ultimate multi-use infrared thermometer is here to deliver accurate temp checks in one second. The 3-in-1 Ear, Forehead + Touchless Thermometer has a color-coded screen and optional sound alarm to alert you when fevers spike. Seamlessly move from in-ear temp checks when baby’s awake, to stealth forehead temp checks while baby’s sleeping. In addition to tiny humans, this thermometer works on adults, and even reads surface temperatures in touchless mode -- like the bottle you’re about to feed your feisty freeloader.
Product Illustration
The thermometer features a forehead reader, an ear probe, a power/measure/mute/unmute button, a mode/memory button, a battery cover, and an LCD screen displaying current and previous temperature values, Fahrenheit/Celsius degrees, mode icons, and low battery indication.
Modes:
- Adult Forehead Mode
- Child Forehead Mode (under 12yr old)
- Ear Mode
- Object Mode
Directions
- Press the big Power/Measure button for 1 second to turn the thermometer on. The LCD screen will show the last measurement taken.
- Select your mode by following the directions for ear, forehead, or touchless use.
- Press and release the Power/Measure button for 1 second again.
- Once the beep is heard, you can read the value. “T” indicates a temperature reading.
- The device automatically shuts off after being unused for 10 seconds.
For Ear Use:
- Remove the forehead reader, automatically selecting Ear mode.
- Fit the probe snugly into the ear canal.
- For children under 1 year: Pull the ear straight back.
- For children over 1 year to adults: Pull the ear up and back.
- Ensure the ear canal is straight for the probe to receive an infrared ray from the eardrum.
- Warning: Ensure the ear is clean, with no earwax or obstructions. The right ear reading may differ from the left ear; always take the temperature in the same ear. Do not force the thermometer into the ear canal, as it may get injured. Be careful when taking the temperature on a child, whose ear canal is small. Do not insert past the point where you feel resistance.
For Forehead Use:
- With the forehead reader on, press the Mode/ Memory button to choose “Adult forehead mode” or “Child forehead mode”.
- Touch the forehead reader to the center of the forehead.
- Warning: The forehead measurement is an indicative reading. The measured forehead temperature can fluctuate up to 1°F/0.5°C from your actual body temperature. Factors influencing accuracy include hair, sweat, or dirt on the forehead; ensure the area is clean. Always check if the lens is clean. Ensure the user and thermometer have been in the same room for at least 30 minutes prior to measurement.
For Touchless Use:
- Note: This mode is ONLY for measuring the temperature of the room or an object. Not for body temp checks.
- Press the Mode/ Memory button to choose the house icon ?.
- Hold the thermometer about 1-3 cm (0.39-1.18 inches) away from the object.
Normal Body Temperature / Fever Indicator
Body temperature can vary from 97°F to 100.0°F and still be considered normal. The medically accepted “normal” body temperature is 98.6°F. Body temperature is commonly lower upon waking. Readings vary based on the body location point.
- Oral: 97°F (36.1°C) - 100°F (37.7°C) are normal.
- Rectal: Generally 1°F (0.5°C) higher than oral.
- Underarm (Axillary): 1°F (0.5°C) lower than oral.
A fever is defined as a body temperature elevated above the normal for that person. Tracking an individual’s temperature consistently at the same body site and time will help determine their normal temperature.
Temperature Reading Indicators:
- Normal: 89.6°F - 99.5°F (32°C - 37.5°C) - Green light, one long beep.
- Low fever: 99.7°F - 100.4°F (37.6°C - 38.0°C) - Orange light, 3 short double-beeps.
- High fever: 100.6°F - 109.2°F (38.1°C - 42.9°C) - Red light, 3 short double-beeps.
- Touchless (Room/Object): 32.0°F - 199.0°F (0°C - 100°C) - White light, one long beep.
When to Call a Doctor:
- Infants under 3 months: Temperature of 100.4°F (38.0°C) or higher.
- At any age: Fever rises above 104°F (40°C) repeatedly.
When to Take Multiple Measurements:
It is recommended to take 3-5 measurements in the same location and use the highest one as the reading in the following situations:
- Newborn infants in the first 100 days.
- Children under three years of age with a compromised immune system.
- When the user is learning to use the thermometer for the first time.
Troubleshooting / Error Codes
Error | Problem/Cause | Solution |
---|---|---|
Er1 | The ambient temperature is not in range (outside 10°C-40°C / 50°F-104°F). | Take a measurement under an ambient temperature between 10°C (50.0°F) and 40°C (104°F). |
Hi | In ear or forehead mode, T>42.9°C (109.2°F) / In object mode, T>100°C (199°F) | One long beep, and red backlit for 3 seconds. |
Lo | In ear or forehead mode, T<32°C (89.6°F) / In object mode, T<0°C (32°F) | One long beep, and red backlight for 3 seconds. |
ErC | Memory error or calibration is not finished. Item will be turned off automatically after 3 seconds. | Contact dealer. |
Low Battery Indication (Battery icon flickers) | Power voltage is between 2.4V-2.7V. You can still use the thermometer. | Replace with a new battery when the icon is solid. |
Low Battery Indication (Battery icon flickers, item turns off) | Power voltage is lower than 2.4V. | Replace with a new battery. |
Failed to power on | Low battery level; Polarities of batteries reversed; Thermometer is damaged. | Replace with a new battery; Ensure batteries are in the right position; Contact dealer. |
Reading is too low | Dirty probe lens; Distance too far; User just came from a cold environment. | Clean the lens; Keep in contact with forehead or put probe into ear canal; Stay in a warm room for at least 30 min before taking a reading. |
Reading is too high | User just came from a hot environment. | Stay in an adequately cool room for at least 30 minutes before taking a reading. |
Specifications
- Type: Digital Infrared Thermometer
- Measure Range: Forehead + Ear: 32.0°C–42.9°C (89.6°F–109.2°F) / Object: 0°C–100°C (32°F–199°F)
- Accuracy: Ear + Forehead mode ±0.2° C/±0.4 °F; Object mode ±1.0°C / 2.0°F
- Display: Segment LCD, Four color LED backlight (white, green, orange, red)
- Memory: Stores 40 values
- Battery: AAA (1.5V DC) x 2
- Battery life: Up to 3,000 times
- Dimension: 143 x 35 x 41mm
- Weight: 80g (with battery)
- Ambient operating range: Temperature: 10°C-40°C (50°F-104°F); Humidity: 15-95% RH (non-condensing); Atmospheric pressure: 70-106kPa
- Storage and transportation condition: Temperature: -20°C-55°C (-4°F-131°F); Humidity: 0-95% RH (non-condensing); Atmospheric pressure: 50kPa to 106kPa
- Ingress Protection Rating: IP 22
- Classification: Type BF
- Model: AOJ-20A
Contents
1 Thermometer, 2 x AAA batteries, 1 Storage Pouch, 1 Owner's Manual, 1 Quick Start Guide
Care and Cleaning
- Use an alcohol swab or cotton swab moistened with 70% alcohol to clean the thermometer casing and probe. Allow alcohol to dry completely before use.
- Ensure no liquid enters the interior. Never use abrasive cleaning agents, thinners, or benzene. Never immerse the instrument in water or other cleaning liquids.
- Take care not to scratch the LCD screen.
- Clean the thermometer probe after each use.
Battery Insertion and Replacement
- First time use: Pull battery door off, open chamber, insert two AAA batteries, slide chamber back, reattach door.
- Replacement: Replace battery when low battery icon appears. Remove old batteries lawfully. Replace with two new AAA batteries.
- Remove batteries if the thermometer will not be used for more than two months.
Memory Mode
Press the Mode/ Memory button for 5 seconds to enter memory mode. Press the button again to check stored readings one by one. Displays “---” if no values are saved. All stored readings are reset when batteries are replaced.
Important Safety Information and Warnings
- This product is not intended to diagnose or treat any health problem or disease and should not be used as a substitute for professional medical advice.
- Thermometers should only be used under adult supervision.
- Keep out of reach of children under 12 years.
- Never immerse the thermometer into water or other liquids (not waterproof).
- Do not use the thermometer if there are signs of damage on the measuring sensor or instrument. Contact FridaBaby if damaged.
- Do not drop the instrument. Protect it from severe impact and shock. Do not twist the instrument or measuring sensor.
- Do not use the thermometer on newborns or for continuous temperature monitoring.
- Do not take a measurement while or immediately after nursing a baby.
- Patients should not drink, eat, or be physically active before/while taking the measurement.
- Warning: Consult your doctor for symptoms like unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc., even in the absence of fever.
- Seek professional advice immediately when there is a temperature elevation and you are taking the temperature of anyone who is:
- Over 60 years of age.
- Has diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer, chemotherapy, chronic steroid treatment, splenectomy).
- Bedridden (e.g., nursing home patient, stroke, chronic illness).
- A transplant patient (e.g., liver, heart, lung, kidney).
- This thermometer is not intended for preterm babies or small-for-gestational age babies.
- This thermometer is not intended to interpret hypothermic temperatures.
- Do not allow children to take their temperatures unattended.
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference; and (2) this device must accept any interference received, including interference that may cause undesirable operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. If this equipment does cause harmful interference to radio or television reception, the user is encouraged to try to correct the interference by reorienting or relocating the receiving antenna, increasing the separation between the equipment and receiver, connecting the equipment into an outlet on a circuit different from that to which the receiver is connected, or consulting a dealer or an experienced radio/TV technician.
CAUTION: Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Conformity
This appliance conforms to the following standards:
- ASTM E 1965-98 (Re-approved 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- ISO80601-2-56 First Edition 2009-10-01 Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement.
Disposal
Do not dispose of used batteries in the trash. Consult your local recycling center for the proper way to dispose of batteries in your area. Keep used batteries away from children, pets, and excessive heat.
Contact Information
Manufactured for Fridababy LLC, Miami, FL 33137
Love it? Leave a review. fridababy.com
Questions? frida.com/help
File Info : application/pdf, 1 Pages, 775.05KB