Operations Manual For Research Use Only
PH: 908-769-5555 EM: info@BenchmarkScientific.com WEB: www.BenchmarkScientific.com

Thank you for choosing our steam sterilizers. Prior to operating this instrument, please read the operations manual carefully and follow all installation instructions. IMPORTANT NOTICE: If you can't open the door, please unlock the door according to the instruction "How to open the door in the case of power outage" in the manual. Need Maintenance
If this picture appears on the screen when power on or E88 appears on the report, please call your dealer or local service maintenance. Your steam sterilizer needs a regular maintenance.
Instructions Manual
B4000-28 Document: Version 07T10013 v2.8
Subjects to technical changes

Table of Contents
1. General .................................................................................................................................................... 1
1.1 Scope of manual ............................................................................................................................................. 1 1.2 Intended use .................................................................................................................................................... 1 1.3 General Safety Instructions .............................................................................................................................. 1 1.4 Standards and directives .................................................................................................................................. 1 1.5 Symbols ........................................................................................................................................................ 1
2. Description of the autoclave ...................................................................................................................... 2
2.1 Autoclave views ................................................................................................................................................ 2 2.2 Control panel ................................................................................................................................................... 2 2.3 Technical Specifications ..................................................................................................................................... 3 2.4 Packing content .............................................................................................................................................. 3
3. Installation .............................................................................................................................................. 4
3.1 General conditions ...................................................................................................................................... 4 3.2 Power supply connection ............................................................................................................................ 4
4. Setup ................................................................................................................................................. 4 4.1 Basic set .......................................................................................................................................................... 5 4.2 Fill the distilled Water Tank .......................................................................................................................... 5 4.3 Preparation of the sterilization materials ........................................................................................................ 5
5. Operation .............................................................................................................................................. 6
5.1 Select the program ........................................................................................................................................... 6 5.2 Running the sterilization program .................................................................................................................. 6 5.3 Start the sterilization program ..................................................................................................................... 6 5.4 End of cycle .................................................................................................................................................. 6 5.5 Manual abort the program ................................................................................................................................. 7 5.6 Record of the cycle ....................................................................................................................................... 7 5.7 Printer ......................................................................................................................................................... 7 5.8 Report ............................................................................................................................................................ 7
6. Advance settings ......................................................................................................................................... 9 6.1 Parameter ...................................................................................................................................................... 9 6.2 Ambient pressure .......................................................................................................................................... 9 6.3 Unit ................................................................................................................................................... 9 6.4 Preheat ............................................................................................................................................................ 9
7. Maintenance ............................................................................................................................................ 10
7.1 Clean the distilled water tank ........................................................................................................................ 10 7.2 Clean Chamber, door seal, trays and tray Rack ........................................................................................... 10 7.3 Door adjustment ...................................................................................................................................... 10 7.4 Replacement of the door seal ...................................................................................................................... 11 7.5 The drain valve .................................................................................................................................... 11 7.6 How to open the door in the case of power outage .................................................................................. 11
8. Error codes ............................................................................................................................................. 12
9. Transportation and storage ......................................................................................................................... 13
10. Safety devices ....................................................................................................................................... 13
Appendix 1 - Water properties / Characteristics .............................................................................................. 14 Appendix 2 - Diagrams of the sterilization programs ................................................................................... 15 Appendix 3 - Wiring diagram ................................................................................................................ 17 Appendix 4 - Hydraulic diagram .................................................................................................................. 18

Instructions Manual
1. General
1.1 Scope of Manual
This manual contains information concerning the installation, operation and maintenance of the steam sterilizers. To ensure proper performance of the autoclave, the instructions given in this manual should be thoroughly understood and followed. Keep the manual near to the sterilization in a readily accessible location for future reference.
1.2 Intended Use
The sterilizer described in this manual is intended for the sterilization of laboratory research tools and media. It operates automatically with 134°C and 121°C sterilization temperatures.
1.3 General Safety Instructions
-Read and understand this manual before attempting to install or operate the sterilizer. -Make sure that all the installation conditions are fully complied with. -Ensure that the supply voltage agrees with the supply voltage specified on the type plate of the sterilizer. -This appliance must be grounded. Connect only to a properly grounded outlet. -Do not cover or block any openings on this appliance. -Use this appliance only for its intended use a described in this manual. -Do not exceed the maximum weight limit of the loads specified in this manual. -Do not operate this appliance if it has a damaged cord or plug, if it is not working properly, or if it has been damaged or dropped. -Never must put into the sterilizer inflammables or explosives products. -The sterilizer may not be operated in areas in which gas or any other explosive volatile substance is present. -Installation and repair work should only be performed by authorized service technician. Work by unqualified persons could be dangerous and void warranty.
1.4 Symbols
For safe operation, please pay close attention to the alert symbols below which can be found in the sterilizer or throughout this manual.
This symbol represents an electrical caution - ground protection. Hot Surface This symbol represents a warning of a potential hot surface. Important safety information. This symbol represents a warning for extra caution.
1

2. Description of the sterilizer
2.1 Sterilizer views

Instructions Manual

1. Used water tank 2. Distilled water tank 3. LCD 4. Control Panel 5. Printer (Optional) 6. Main power switch 7. USB port
2.2 Control panel
.

8. Drain connector (Used water tank) 9. Drain connector (Distilled tank) 10. Door 11. Safety valve 12. Condenser ventilation 13. Rating plate 14. Circuit breaker 15. Power socket

Date of manufacture

SN

Manufacture number

Manufacturer Name

Manufacturer Address

1

Temperature of the cycle

2

Program

3

Holding time

4

Pressure

5

Fill distilled water tank

6

Drain used water tank

7

Printer is connected

8

Door is closed

9

Time

10

Current temperature

11

Up button

12

Down button

13

Start button

14

Set button

2

2.3 Technical specifications

Instructions Manual

Item
Chamber Rated Voltage Circuit breaker Nominal power Sterilization temperature
Capacity of the distilled water tank
Operation temperature Operation relative Humidity
Overall dimensions Net weight
Max. Noise level Atmospheric pressure

Parameter
319mm x 417 mm (12.5 x 16.4 in.) 220V (200V to 245V) 50/60Hz F16A/400V 2100VA 121°C /134°C
Approx 8.0 L (Water at level Max) Approx 1.5 L (Water at level Min)
5°C-40°C Max. 80%, non condensing 640mm(W)*560mm(H)*640mm(D)
92 kg <70 dB 76 kPa ­ 106 kPa

2.4 Packing content

No.

Accessory

1

Instrument tray rack

2

Draining hose

3

Instructions manual

4

Door seal

Quantity
1 2 1 1

3

Instructions Manual
3. Installation
3.1 General conditions
Position the device on a flat and level surface with minimum capacity 60 kgs. The sterilizer should be placed on a level worktable. Improper water level in the chamber could cause a sterilizer malfunction. Leave at least 10cm between the device rear part and the wall. The clearance required to open the door is 40cm. Position the autoclave at such a height as to make it possible for the operator to check the whole sterilization chamber and carry out the normal cleaning operations. The room where the device is installed must be enough ventilated. Do not install the device near washing basins, taps, etc. where it is likely to be splashed. Do not lean on the door when it is opened. Do not place trays , papers, fluid containers, etc. on the sterilizer.

3. 2 Power supply connection
Check the label on back panel o sterilizer to verify voltage rating for the unit. Failure to connect the autoclave to an appropriate power supply could result in damage to the unit, and electrical shock to personnel. Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only used for the sterilizer is recommended. Never connect the device pin to reductions of any type.

4 Setup
Open the door and remove all of the inner contents for unpacking. Connect the power cord to an outlet of the appropriate voltage. Turn on the main switch on the right side. After switching on, the machine turns on the LCD and shows the door position, water level, working program, date, time and etc.
Holding the up button for about 3 seconds, it will unlock the door.

Note: The control panel will be locked for the initial 10 seconds after powering up for system initialization.

Notice: Before using the sterilizer or at any time the low water level icon distilled water.

blinks, fill the distilled water tank with

4

Instructions Manual 4.1 Basic Setting
The "Basic Setting" Menu permits to set the following options: *Date *Time *Language Select the "Basic Setting" from the main menu and press the"Set" button. Select the item by pressing the "Set" button. The unit you selected will be lighted. Adjust the value by pressing the up or down buttons. Press the "Set" button to proceed to the next item. After the language is set, then the data is permitted to be saved by pressing the "Set" button until the main menu appears.

Program Basic Set Report
DATE:23-04-17 TIME:09:10:08 LANGUAGE:ENG Counter:12

Note: If you want to cancel the setting and return to the main menu by pressing the "Start" button. Note: The Counter (cycle No) can not be set by the operator.

Abbreviation of language options

CHN

Chinese

ENG

English

DEU

PL

Polish

FR

French

HUN

NL

Dutch

LTU

Lithuanian

LAT

ITA

Italian

RUS

Russian

4.2 Fill the distilled water tank
Ensure that the drain valve is closed and open the water tank cover.

German Hungary Latvian

ESP ROM CZE

Spanish Romanian
Czech

Use only distilled water. (see Appendix 1)
4.3 Preparation of sterilization materials
For the most effective sterilization and to preserve the sample, please follow below: *Clean instruments immediately after use. *Residual chemicals left over after cleaning and disinfecting process may damage and corrode parts of the autoclave, always rinse off the instruments using distilled water. *Follow instrument manufacturer's guidelines and recommendations for handing and cleaning instruments prior to sterilization. *Check the manufacturer's instructions as to proper procedure for sterilizing of each item. *Arrange the samples of different materials on different trays or with at least 3cm of space between them. *Clean and dry instruments thoroughly before placing them into tray. *We recommend using sterilization paper (or cloth) between the tray and sample to avoid direct contact. *Arrange the containers (glasses, cups, test-tubes, etc) on one side or inverted position, avoiding possible water stagnation. *Don't stack the trays one above the other or put them in direct contact with the walls of the sterilization chamber. *Always use the instrument tray handle. *Wrap the samples one by one or, if more tools have to be set in the same bag, verify that these are made of the same material.
5

Instructions Manual
*Don't use metallic clips, pins or other, as this jeopardizes the maintenance of the autoclave. *Don't overload the trays over the stated limit (see appendix 2).

5. Operation 5.1 Select the program
Press the "Up" button to open the main menu, select "Program". You will see the available sterilization programs. See Appendix 2.

Program Basic Set
Report

SOLID(134)
WRAPPED(121) WRAPPED(134)
PRION(134)

Select the program by up or down buttons and confirm it by pressing the "Set" button. If you don't want to select a program you may press the "Return" button to exit.

5.2 Running the sterilization program.
After selecting program, the instruments to be sterilized can now be placed on the tray placed inside the chamber. 5.2.1 After the instruments are loaded, you may close the door.
Push the door to be closed, holding for about three seconds, it will lock automatically. Caution: The door can not be locked if the pressure is higher than 0 kPa. Please wait for the chamber cooling down.

5.3 Start the sterilization program.
After the "Start" button is pressed, the stage and the status of the current cycle will appear on the display. The sterilizers will perform the program automatically. (see appendix 2).

Total time or count down until completion
5.4 End of cycle
After cycle is completed, the printer will be activated and print out a report of the cycle (if the optional printer has been connected) or save the report in the USB drive (optional).

Caution: Always use the tray handle to load or unload the tray into the autoclave. Failure to do so can result in burning.
6

Instructions Manual 5.5 Manual abort of the program
It is possible to interrupt a started cycle prematurely. If you need to interrupt a cycle and remove the items urgently, you may hold the "Start" button for 3 seconds. (If held for 3 seconds during the drying time, this will skip the dry cycle.

If you interrupt any cycle before it reaches the "Drying" step, then the items inside the autoclave must be considered not sterile. If you need to interrupt a cycle after the holding time of the sterilization cycle and during the drying step then the items inside the autoclave can be considered sterilized.

Caution: Depending on the status of the Cycle, when the door is opened, steam may escape the chamber.

5.6 Record of the cycle
USB Flash memory (Optional) A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive into the slot located on the service door of the sterilizer. The information will automatically output directly to the USB drive after the cycle has completed. The name of the file is determined by the serial number of the machine and the cycle number. For example: The serial number is E00001. The cycle number is 0012. The file name in the USB stick is 01001200.txt. The first two numbers represent machine number. The middle four numbers represent cycle number. The last two numbers represent error code. E.g. 00:no error;01: error E01

5.7 Printer (Optional)
If installed you can see the Icon in the screen stop flashing. The printer (Optional) will produce a report of the cycle that just ended. T the end of each cycle the printer will print out a report of the cycle.

5.8 Report
Internal Memory In this menu you can read the latest information of the last 20 cycles stored in the internal memory of the sterilizer. 5.9.1 Select "Report" from the main menu and press the Set" button, you will see the list of records. 5.9.2 Select the record by pressing the up or down buttons. 5.9.3 Press the "Set" button. Then the record is printed (if the optional printer has been connected) the report or/and is saved in the USB drive (optional). Note: It can save in the internal memory only the reports of the last 20 cycles. Note: The storage system is based on the principle of "first In-first-\out". Tap start button to exit.

Program Basic Set Report
00012 00013 00014 00015

When vising printed data records, refer to the diagram below:

7

Instructions Manual

=========================

Program:

WRAPPED

Temperature: 134C

Pressure: 206.0 kPa

Dry Time: 02Min

Ster Time: 4.0Min

-------------------------------------

time temp. pressure

Start 15:24:20 042.0C

T1: 15:32:11 070.0C 053.2kPa

T2 : 15 :36 :08 075.3C 009.7kPa

T3 : 15 :39 :21 090.3C 050.4kPa

T4 : 15 :44 :32 094.3C 009.6kPa

T5 : 15 :47 :12 119.0C 102.4kPa

T6 : 16 :00 :11 110.2C 009.3kPa

TS :

134.8C 221.6kPa

MAX.Temperature:135.1C

MIN.Temperature:134.5C

MAX.Pressure :230.4kPa

MIN.Pressure :212.9kPa

T7 : 16 :04 :02 135.0C 223.5kPa

T8: 16:06:32 134.8C 214.1kPa

End 16:14:12 78.2C

--------------------------------------

Cycle No: 0005

Ster Value: Success

Date: 2011-01-18

SN:E00001

Operator:

=======================

8

Instructions Manual
6. Advance setting
The advance setting interface permits to set the following Options: *Parameter: Permits to change the time of Holding time and Drying time. *Unit: Permits to change the unit of measure temperature and pressure. *Preheat: This option allows you to maintain the required temperature in the sterilization chamber and the steam generator to start a new cycle immediately for the next 60 minutes so to run a new cycle should expect preheating time from 3 to 5 minutes upon start. If the option is disabled (OFF) once the sterilization cycle end immediately the Autoclave is no longer heated, so when you start a new cycle should expect a preheating time between 3 to 5minutes upon start. Note: To maintain the temperature for longer time it is recommended that after each cycle hold the door closed.

Enter the setting Power on the machine while holding the set button and hold for 5 seconds. This will enter into the advanced setting mode.
6.1 Parameter
Press up or down buttons to select Parameter. Press "Set" button to enter the menu. Select the program that you need to change by pressing the up or down buttons. Press "Set" button to enter the setting. Adjust the parameter by pressing the up or down buttons. Drying time:0-20 minutes 121 holding time: 1-60 minutes 134 holding Time: 1-20 minutes After you finish adjusting the parameter Press the "Set" button to save and return to the above menu. Press the "Start" button to cancel and exit.

Parameter Unit
Preheat
SOLID(121) SOLID(134) WRAPPED(121) WRAPPED(134)
Holding Time:20.0 Dry Time:02

6.2 Ambient pressure
Select the "Ambient Pressure" to calibrate the local ambient pressure to 0. Tap set button to enter the menu. Make sure the pressure inside the chamber had been balanced. Press the "Set" button to calibrate the pressure. Press the "Return" button to cancel and exit.

Ambient Pressure Offset: 0.1

6.3 Unit
Select the "Unit" to adjust the unit of temperature and pressure. Press" the "Set" button to enter the menu. Select the item by pressing the "Set" button. Change the unit by pressing the up or down buttons.
Pressure: kPa/bar/psi Temperature: /F Press the "Set" button to save and exit. Cancel and Exit by pressing the "Return" button.

Pressure: kPa Temperature: C

6.4 Preheat
Select the "Preheat" to adjust the Preheat. Press the "Set" button to enter the menu. If you prefer not to activate the automatic preheating, set the value to "OFF" by pressing the up or down buttons. Press the "Set" button to save. Tap start button to cancel and exit.

Preheat: ON

9

Instructions Manual

7.Maintenance

To assure proper operation and maximum autoclave life, carefully follow all recommendations for periodic maintenance.

One of the MOST important steps you can take to prevent problems with your sterilizer is to use ONLY distilled water.

Frequency

Number of cycles

Maintenance operation

Clean the door seal

Clean the filter inside the chamber and in the clean

Monthly

50

water tank

Clean the chamber the trays and the rack

Clean the external surface

Every 3 months

200

Clean the distilled water tank

Every year

800

Replace the door seal

7.1 Clean the distilled water tank
Disconnect the main cable. Drain the tank completely using the drain tube and leave it connected into the connector in a open position. Clean the internal surface with a soft sponge and a small soft brush for the areas that are difficult to reach using and a mild soap. Remove the filter and clean it with a small soft brush and mild soap, rinse it with distilled water and put it back in to the position.

7.2 Clean Chamber, door seal, trays and tray Rack.

Remove the trays and tray rack from the chamber.

Clean trays, rack and inside of chamber with mild soap.

Rinse the trays, rack and inside of chamber with a smooth cloth

and

distilled water.

Examine door seal for possible damage.

Clean door seal and mating surfaces with a damp cloth.

Note: Do not use bleaching agents or any abrasive materials / substances in chamber. Failure to comply may result in damage to the chamber and/or other components.

Caution: To prevent burns, let unit to cool before cleaning gaskets and touch the surface.

7.3 Door adjustment
Under normal circumstances the chamber door does not require adjustments. However, if the seal fails (resulting in steam leaking from the front of the chamber), you may use the spanner tool to tighten the door seal. 7.4.1 Open the door. 7.4.2 Insert the spanner in the hole. Loose the screw anticlockwise.
Tighten the door by rotate the lid counter-clockwise. Loosen the door by rotate the lid clockwise as the figure below. 7.4.3 Tighten the screw clockwise.
Caution: Never adjust the chamber door while the door is closed.

10

Instructions Manual 7.4 Replacement of the door seal
Open the chamber door. Remove the door seal ring carefully by hand. Clean the door seal ring carefully with a smooth cloth with distilled water. Moisten the new seal with medical disinfectant or isopropyl alcohol. Insert the new seal and press in sequence as follows:

2) Press in the top and
bottom of the door seal.

2) Press in the left and right sides of the door seal.

3) Press the remaining sections of the seal.

Caution: Please ensure the chamber and the door are cooled prior to replacing the seal ring.

7.5 The drain valve

1.Press the included hose on 2.Pull the drain valve to the drain valve firmly. outward to drain the tank.

3. Push the drain valve back after draining the tank.

11

8.Error codes

Instructions Manual

Code E1 E2 E3 E5 E6 E9 E11 E12 E20 E21 E23 E24
E25
E26 E28 E32 E33

Description

Proposed solution

Steam generator temperature sensor Power off & run a new cycle

error

Contact your supplier if error persist

Inner temperature sensor error

Power off & run a new cycle Contact your supplier if error persist

Temperature sensor of the chamber Carefully ensure that the chamber wall is heated and

wall error

contact your supplier

Fail to release the pressure

Power off & run a new cycle Contact your supplier if error persist

Door lock problem during the cycle Check the door close switch.

Failure to hold temperature

Ensure the distilled tank isn't empty. Check the inner temperature sensor. Check somewhere for leaking.

Failure to preheat the steam

Power off & run a new cycle

generator

Contact your supplier if error persist

Failure to preheat the chamber

Power off & run a new cycle Contact your supplier if error persist

Program manually interrupted

Holding the set button for 3 seconds

Failure to reach the holding time.

Check somewhere leaking inside the autoclave.

(sterilization time)

Contact your supplier if error persist

Result of vacuum test is void

The temperature of the chamber is high. Try again after the chamber has cooled down.

It takes too long time to enter the next Check somewhere leaking.

status.

Or contact your supplier if error persists.

The door is too tighten.

Or Check the door locking switch. There is a problem of locking the door.
Check the door motor.

Contact your supplier if error persists.

There is a problem of unlocking the door.

Check the door unlocking switch. Check the door motor. Contact your supplier if error persists.

The pressure is over 240kPa

Power off and contact your supplier if error persists

The pressure is not lower than -1 kPa Power off and contact your supplier if error persists
in 30 seconds during locking the door.

The pressure is not lower than 1 kPa in Power off and contact your supplier if error persists
30 seconds during unlocking the door.

Caution: You may cancel the voice of alarm by taping any button. And cancel the alarm by holding the set button for 3 seconds after you repair it. Then swith off and switch on again.

12

Instructions Manual
9. Transportation and storage
9.1 Switch off the sterilizer before transportation or storage. 9.2 Pull out the plug. Let the machine cool down. 9.3 Drain the distilled water tank and the used water tank. Condition for transport and storage Temperature: -20°C ~ +50°C Relative humidity:  85% Atmospheric pressure: 50kPa~ 106kPa.
10. Safety devices
1.Main fuses: Protection the instrument against possible failures of the heating resistor. Action: Interruption of the electric power supply. 2.Thermal cutouts on the main transformer windings: protection against possible short circuit and main transformer primary winding overheating Action: Temporary interruption of winding. 3.Safety valve: Protection against possible sterilization chamber over-pressure. Action: Release of the steam and restoration of the safety pressure. 4.Safety micro-switch for the door status: Comparison for the correct closing position of the door. Action: Signal of the wrong position of the door 5.Thermostat on chamber heating resistors: Protection for possible over heating of the chamber heating resistors. Action: Interruption of the power supply of the chamber resistors. 6.Thermostat on steam generator heating resistors: Protection for possible overheating of the steam generator heating resistors. Action: Interruption of the power supply of the steam generator resistors. 7.Door safety lock: Protection against accidental opening of the door. Action: Impediment of the accidental opening if the door during the program. 8.Self-leveling hydraulic system: Hydraulic system for the natural pressure leveling in case of manual cycle interruption, alarm or black-out. Action: Automatic restoration of the atmospheric pressure inside chamber.
13

Appendix 1

Instructions Manual

Water properties / Characteristics

Description Evaporate residue Silicium oxide sio2 Iron Cadmiun Lead Rest of heavy metals Chloride Phosphates Conductivity PH Value Appearance Hardness

Feed water  10mg/ I  1mg/ I  0.2mg/ I  0.005mg/ I  0.05mg/ I  0.1mg/ I  2mg/ I  0.5 mg/ I  15s /cm 5 ­ 7.5
Colorless, clean
0.02 mmol/ I

Condensate  1.0mg/kg  1.0mg/kg  0.1mg/kg  0.05mg/kg  0.1mg/kg  0.1mg/kg  0.1mg/kg  0.1mg/kg  3 s /cm 5-7
Colorless, clean
0.02 mmol/ I

14

Appendix 2

Instructions Manual
Diagrams of the sterilization programs

Temperature (°C) Pressure (Kpa) Holding time ( i) Total time (min) Max load (kg)
Max load per tray (kg)

Programs

Type

134 210
UNWRAPPED
121 110

4 25-40 20 25-40

Unwrapped solid material

6.0 2.0

134 210 6 33-53

Unwrapped porous material

3.5 2.0

Single-wrapped porous material 3.0 2.0

WRAPPED 121 110

20 30-50

Dual-wrapped porous material 2.5 2.0
Single-wrapped hollow material 4.0 2.0
Dual-wrapped solid and hollow
3.5 2.0
material

Unwrapped porous material

3.5 2.0

134 210 10 37-57

Single-wrapped porous material 3.0 2.0

TEXTILE

Dual-wrapped porous material 2.5 2.0 Single-wrapped hollow material 4.0 2.0

121 110 30 40-60

Dual-wrapped solid and hollow
3.5 2.0
material

Unwrapped porous material

6.0 2.0

Single-wrapped porous material 3.5 2.0

(Extensive) 134 210
PRION

18 45-65

Dual-wrapped porous material 3.0 2.0 Single-wrapped hollow material 2.5 2.0

Dual-wrapped solid and hollow
4.0 2.0
material

LIQUID 134 210 10 60-80

Liquid

4.0 4.0

LIQUID DRYING

121 110
 

30 60-80
 1-20

Liquid, agar


4.0 4.0
 

The time required for sterilizer to be ready for routine use after the power is switched is less than 5

minutes.

The max. Temperature of the 134°C sterilization cycle is 137°C

The max. Temperature of the 121°C sterilization cycle is 124°C

15

Instructions Manual Unwrapped

Wrapped Textile Prion

LIQUID

Drying

16

Appendix 3
Thermal protector Chamber heater Steam generator heater Chamber heater Steam generator heater

Instructions Manual
Wiring diagram

Water pump Fan

Printer power Motor power

TP1: Steam generator temperature sensor TP2: Inner temperature sensor TP3: Temperature sensor of chamber wall V1: Air release valve V4: Water release valve

Printer Port
USB drive Tank Min. Level Public Tank Max. Level Used water tank Door lock switch Door unlock switch Door close switch
TP3
TP2 TP1
LCD

17

Appendix 4

Instructions Manual
Hydraulic diagram

Safety valve

Pressure sensor

Condenser

Test Port

Chamber

Distilled water tank

Used water tank

V1: Air release valve V4: Water release valve

18

