ad4411 simplified
AD-4411 Weighing Indicator Simplified Instruction Manual

1WMPD4004632B

 All rights reserved. No part of this manual may be reprinted or copied without the prior written permission of A&D.
 The specifications and other information contained in this manual are subject to change without notice in order to make improvements.
 Every precaution has been taken in the preparation of this manual. Nevertheless, A&D assumes no responsibility for errors or omissions. If you discover any problems with this manual, please notify A&D.
 A&D Company, Limited shall not be liable for special, indirect, incidental, or consequential damages, loss of profits or production or commercial loss in any way connected with the products, whether such claim is based in contract, warranty, negligence, or strict liability.

3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-1566
Detailed instruction manual
This manual provides simplified precautions and operating instructions for AD-4411. For further information about the AD-4411, please refer to the "AD-4411 Instruction Manual" which is available for download from the A&D website (http://www.aandd.jp).
Introduction
The AD-4411 is a weighing indicator that can convert signals from strain gauge load cells and connect them to an Ethernet-based field network. It contributes to an efficient system by connecting weighing instruments to industrial control systems in plants and factories. - Daisy-chain connection is possible without a switching hub, thanks to two
communication ports. - 7-segment green LED display with a character height of 10mm and display
resolution of ±999999. - High-speed AD conversion of 1200 times/second and digital filter enable high
speed and accuracy weighing. - DIN96x48 panel mount type with IP65 protection on the front panel. - PC can update the settings via USB port.
Safety precaution
Read the following precautions carefully before using the indicator for safe and correct usage.
WARNING  Provide an external safety circuit to the indicator so that the safety of the whole
system can be secured even if errors occur in the external power supply or the indicator.  Do not use the indicator in the following environment: -where the temperature and the humidity exceed the specifications -where corrosive gases or flammable gases exist -where the indicator gets wet with oil, chemicals, or water -where the indicator is exposed to direct sunlight  Turn off all the external power supplies used in the system before installing or removing the indicator.  Turn off all the external power supplies used in the system before wiring.  Be sure to ground the indicator.

Part names
Front panel

(4)

(1)

(5)

(2)

(3)

No. Name (1) Main display
ZERO status NET status (2) STABLE status S1 / S2 / S3 status [ZERO/] key
[TARE/] key
(3) [F1/] key
[F2/ ] key
[MODE/ESC] key (4) Capacity label (5) Unit label

Description Displays measured value or various settings. The LED is ON when the measured value is within 1/4 the minimum division. The LED is ON when the net value is displayed. The LED is ON when the measured value is stable. The LED is ON when the S1 / S2 / S3 status ON condition (FncF07 / 08 / 09) is met. Zeros the gross value. Moves the flashing digit to the left when not in measurement mode. Performs tare. Increases the flashing digit by one when not in measurement mode. Performs the function set for the F1 key function (FncF05). Decreases the flashing digit by one when not in measurement mode. Performs the function set for the F2 key function (FncF06). Updates the setting value entered when not in measurement mode. Changes the operation mode. Cancels the setting value entered when not in measurement mode. Attach the included capacity label, if necessary. Attach the included unit label, if necessary.

Rear panel

(2)

(3) (1)

(4)

(5)

No. Name

Description

(1) DC power input terminals Terminals for connection of a DC24V power

supply.

(2) Load cell input terminals Terminals for connection of load cells.

(3) USB connector

Connector for connection with setting PC.

(Type-C)

(4) Field network status LEDs Notifies field network status.

(5) Field network connector Connector for connection of PLC via field

network. Dual ports can be used for daisy chain

wiring (RJ-45).

Accessories Waterproof packing, Panel mount bracket x2, Capacity label, Unit label, Power connector, Load cell connector.
Mounting to control panel
Insert the waterproof packing around the Unit, and insert the Unit through the front of the panel. Insert the left and right mounting brackets into the case grooves and push until they reach the panel.
Waterproof packing Control panel
Panel mounting brackets

Connection to power supply and connection to load cell
Attach the accessory power connector and wire as shown in the figure below.

Applicable wire

Item

Specifications

Wire size

0.14 to 1.5 mm² (AWG 26 to 16)

Wire strip length

7 mm

24V

Tightening torque 0.22 to 0.25 Nm

0V
In the case of the 4-wire connection type, attach the accessory load cell connector and wire as shown below. Change the load cell connection type (CALF17) in the calibration function to 0: 4-wire type (default value = 1: 6-wire type).

Shield Power supply (+)
Signal (+) Power supply (-) Signal (-)

6-wire connection Set load cell connection type (CALF17) to 1: 6 wire type (Default). When you connect the load cells in parallel, use a summing box. Attach the accessory load cell connector and wire as shown below.
Shield Power supply (+) Signal (+) Power supply (-) Signal (-)
Summing box

Calibration
Calibrate the AD-4411 to properly convert the signal from the load cell to a load value. Please prepare a calibration weight.
After Power-On, press the [MODE/ESC] key more than 3s.
Press the [F1/] key twice.

Press the [F2/ ] key.

×2

Press the[F1/] key.

Actual load calibration Press the [F2/ ] key.
Press the [F2/ ] key.

The current load cell input signal (mV/V) will be displayed. Press the [F2/ ] key to execute Zero calibration.
If Zero calibration is successful, "PASS" will be displayed, and zero calibration will be completed. Press the [F2/ ] key.
Press the [F2/ ] key.
Set a calibration weight value by the following key operations. [ZERO/] key: Moves the flashing digit to the left. [TARE/] key: Increases the flashing digit by one. [F1/] key: Decreases the flashing digit by one. [F2/ ] key: Confirm the setting value.
The current load cell input signal (mV/V) will be displayed. Place the calibration weight or apply a load on the load cell. Press the [F2/ ] to execute Span calibration.
If span calibration is successful, "PASS" will be displayed, and span calibration will be completed. × 4 Press the [MODE/ESC] key four times to return to the measurement mode.

Operation mode

Power ON

Press more than 3s

Measurement mode

Basic function

Item selection

Setting value input

For details, see the "AD-4411 Instruction Manual" on the A&D website (http://www.aandd.jp).

Press to return to the previous screen.

Checking software version Checking display Checking key switches Checking load cell input signal (mV/V)

Calibration function

Item selection

Actual load calibration

Zero calibration

Span calibration
Digital calibration function Item selection Initialization mode

Setting value input

Executing zero calibration

Checking signal (mV/V) Executing span calibration

Calibration weight value input

Checking signal (mV/V)

Setting value input

Function list

Calibration function list

CALF Setting item

01

Unit

02

Decimal point position

03

Minimum division d

04

Maximum capacity

05

Zero setting range

06

Zero tracking time

07

Zero tracking width

08

Stability detection time

09

Stability detection width

10

Zero-setting when unstable

11

Taring when unstable

12

Taring when the gross is

negative

13

Zero clear

14

Power-on zero

15

Condition of negative

overload

16

NTEP

17

Load cell connection type

Setting value 0: None / 1: g / 2: kg / 3: t 0: 0 (No decimal point) / 1: 0.0 / 2: 0.00 / 3: 0.000 / 4: 0.0000 / 5: 0.00000 1: 1 d / 2: 2 d / 3: 5 d / 4: 10 d / 5: 20 d / 6: 50 d 1 to 999999 0 to 100 % 0.0 to 5.0 s 0: Disable / 1: 0.5 d / 2: 1.0 d / 3: 1.5 d / 4: 2.0 d / 5: 2.5 d / 6: 3.0 d / 7: 3.5 d / 8: 4.0 d / 9: 4.5 d 0.0 to 9.9 s 0 to 100 d 0: Disable / 1: Enable 0: Disable / 1: Enable 0: Disable / 1: Enable

Default 2 0
1 999999 100 0.0 0
1.0 2 1 1 1

0: Disable / 1: Enable

1

0: Disable / 1: Enable

0

0: Gross < -(Maximum capacity + 8d) /

0

1: Gross < -19d

0: Disable / 1: Enable

0

0: 4-wire type / 1: 6-wire type

1

Digital calibration function list

DCAL Setting item

01

Load cell input signal at Zero Calibration

02

Load cell input signal (at Span Calibration

­ at Zero Calibration)

03

Weight value at Span Calibration

Setting value -7.00000 to 7.00000 mV/V 0.00001 to 7.00000 mV/V
1 to 999999

Default 0.00000 2.00000
20000

Basic function list

FncF Setting item

Setting value

01

Locking [ZERO/] key

0: Disable / 1: Enable

02

Locking [TARE/] key

0: Disable / 1: Enable

03

Locking [F1/] key

0: Disable / 1: Enable

04

Locking [F2/ ] key

0: Disable / 1: Enable

05

Function of [F1/] key

0: None / 1: Tare clear / 2: Zero clear /

06

Function of [F2/ ] key

3: Gross / net display selection 4: High-resolution display selection

07

Condition to turn S1 status ON 0: None / 1: Hi / 2: OK / 3: Lo /

08

Condition to turn S2 status ON 4: Zero setting error / 5: Taring error /

09

Condition to turn S3 status ON 6: High resolution display

10

Digital filter cut-off frequency 0: 273.0 12: 20.0 24: 2.8 36: 0.40

[Hz]

1: 120.0 13: 17.0 25: 2.4 37: 0.34

2: 100.0 14: 14.0 26: 2.0 38: 0.28

3: 84.0 15: 12.0 27: 1.7 39: 0.24

4: 70.0 16:10.0 28: 1.4 40: 0.20

5: 68.0 17: 8.4 29: 1.2 41: 0.17

6: 56.0 18: 7.0 30: 1.0 42: 0.14

7: 48.0 19: 6.8 31: 0.84 43: 0.12

8: 40.0 20: 5.6 32: 0.70 44: 0.10

9: 34.0 21: 4.8 33: 0.68 45: 0.08

10: 28.0 22: 4.0 34: 0.56 46: 0.07

11: 24.0 23: 3.4 35: 0.48

11

Upper limit value

-999999 to 999999

12

Lower limit value

-999999 to 999999

13

Comparison target for Upper 1: Gross / 2: Net

limit value / Lower limit value

Default 0 0 0 0 0 0 0 0 0 30
10 -10 1

For functions other than those listed above, see the "AD-4411 Instruction Manual" on the A&D website (http://www.aandd.jp).

Specifications
Dimension Installation method Operating temperature and humidity range IP rating
Power supply Load cell input
Excitation voltage
Signal input range minimum input sensitivity Nonlinearity Temperature coefficient
Sampling rate Display
Main display Status display Unit Key switches External interface AD-4411-EIP AD-4411-PRT AD-4411-ECT USB

96(W) x 48(H) x 98.5(D) mm Panel mount -10°C to +40°C Less than 85%RH, non-condensing (When the indicator is installed to the control panel) Front panel: IP65. Inside the panel: IP2X DC24V -15% to +10%, 4.5W max.
DC5V ±5% 90 mA Up to six 350  load cells can be connected in parallel. 6-wire type with remote sensing -7.0 mV/V to +7.0 mV/V 0.15 V/d or more (d=minimum division) 0.005% of F.S. max. Zero drift: ±0.02 V/°C typ. ±0.1 V/°C max. Span drift: ±3 ppm/°C typ. ±15 ppm/°C max. 1200 times / second
7-digit LED (green) with a character height of 10 mm LED (red) x 6 Attach a label of g / kg / t x 5
EtherNet/IP PROFINET EtherCAT Type-C connector, USB 2.0 (Full-speed)

External dimension

96

9

89.5

4 8

44. 6

91

92

+0.8 -0

+0.6 - 0

Panel thickness 1.6 to 3.2

4 5

Unit: mm

