INSTRUCTION MANUAL

BLE-YC01 Bluetooth 6-IN-1 Pool Tester (PH/EC/TDS/CL/ORP/TEMP Meter)

Product Overview and Parameters

The BLE-YC01 is a Bluetooth 6-in-1 pool tester designed to measure pH, EC, TDS, CL, ORP, and Temperature. It features an HD LCD display for clear readings.

Parameters

ParameterItemRangeResolutionAccuracy
pH0.1-14.0 pH0.1 pH±0.1 pH
CL0.1-4.0 mg/L0.1 mg/L
ECRange0-1999 µS/cm; 2000-10000 µS/cm; 10.01-20.0 mS/cm1 µS/cm; 10 µS/cm; 0.01 mS/cm±2% F.S
Resolution
Accuracy
TDSRange0-999 ppm; 1000-9990 ppm1 ppm; 10 ppm±2% F.S.
Resolution
Accuracy
ORPRange-999~+999 mV1 mV±15 mV
Resolution
Accuracy
Temp.Range0.1°C-80.0°C; 32.1°F-176.0°F0.1°C; 0.1°F±0.5°C
Resolution
Accuracy
CalibrationpH6.86/4.00/9.2 or 7.0/4.0/10.0
EC/TDS1413 µS/cm; 12.88 mS/cm; 111.8 mS/cm
ORP256 mVCalibration records can be viewed
ATC Temp.0.1°C-60.0°C
Environment0.1°C-80.0°C RH: max 90%
TDS Factor0.1-3.0 (Factory Default 0.5)
EC Temp. Compensation0.1-3.0 %/°C (Factory Default 2.0 %/°C)
Bluetoothmobile bluetooth 5.2 Version
Backlightswith backlights, support mobile control
DataSave, Printingsave as Excel sheet
WaterproofIP67
Battery3*AAA low battery indicator
Standby timeone year
Data StorageCan store the latest 1000 records of data
Dimensions/WeightHeight: 155mm; Diameter: 85mm (6.1*3.35in)/210g

Buttons and Display

The device features physical buttons for operation: "HOLD/TEMP" (to hold readings or switch temperature units), "MODE/CAL" (to switch modes or initiate calibration), and "ON/OFF" (power on/off). The HD LCD Display shows current measurements, such as pH (e.g., 5.2) and temperature (e.g., 22.5), along with the model name BLE-YC01.

App Interface and Data Management

The BLE-YC01 connects via Bluetooth 5.2 to a mobile application, allowing for advanced data management and control.

Saving Data Manually

To save data manually, click the "Save Reading" button in the app interface. A screen will appear showing the saved data, including Date, EC, TDS, ORP, CL, pH, and Temperature. Notes can also be added.

Viewing Measurement Data

Press the "Saved" icon in the app to view measurement data. Manually saved data records are displayed in black, while automatically saved data records are displayed in blue.

Backlight Control

The app allows control of the backlight. Toggle between "Back light on" and "Back light off" options.

Power Display

The app displays the current power level, e.g., "80%".

Hold Reading

The "Hold Reading" function can be activated from the app interface to freeze the current measurement on the screen.

Automatic Saving

The device supports automatic data saving. This feature can be toggled "on" or "off". The automatic saving interval can be set between 0 and 9999 seconds.

Settings and Configuration

Setting Max/Min Limits

Users can define maximum or minimum limit values for PH, EC, ORP, and CL based on their requirements. If test results fall outside these ranges, the numbers will be displayed in red.

  • pH Limits: Max 8.00, Min 1.00
  • EC Limits: Max 10000 µS/cm, Min 500 µS/cm
  • ORP Limits: Max 999 mV, Min -300 mV
  • CL Limits: Max 0.2 mg/L

TDS Factor Definition

The default TDS factor is 0.5. It can be set within a range of 0.0 to 3.0.

EC Temperature Compensation Factor Definition

The factory default EC temperature compensation factor is 2.0%/°C. It can be adjusted within a range of 0.0 to 3.0%/°C.

Save, Share, and Print

Press the "Saved" icon to view records. Press the "Share" icon to share the file. If connected to a printer, test records can be printed. Export options include Excel spreadsheets and sharing via various social media platforms.

Calibration Procedures

The device supports two sets of calibration points: 6.86/4.00/9.18 and 7.00/4.00/10.01. These can be selected in the settings.

PH Calibration

Preparation: Dissolve each buffer solution in 3 cups filled with 250 ml of distilled water. Convert the device to PH mode.

  1. Calibrating 6.86: Place the electrode in the 6.86 solution. Wait for the test to stabilize. Press "CAL" for 5 seconds, then release the button. Wait for "6.86" to flash 3 times. Calibration 6.86 is complete. Clean the electrode with distilled water.
  2. Calibrating 4.00: Place the electrode in the 4.00 solution. Wait for the test to stabilize. Press "CAL" for 5 seconds, then release the button. Wait for "4.00" to flash 3 times. Calibration 4.00 is complete. Clean the electrode with distilled water.
  3. Calibrating 9.18: Place the electrode in the 9.18 solution. Wait for the test to stabilize. Press "CAL" for 5 seconds, then release the button. Wait for "9.18" to flash 3 times. Calibration 9.18 is complete.

Note: The calibration steps for 7.00/4.00/10.01 are the same as above.

TDS/EC Calibration

  1. Calibrating 1413 µS/cm: Press "MODE/CAL" to convert to EC mode. Place the electrode in a 1413 µS/cm standard solution. Wait for the test to stabilize. Press "MODE/CAL" for 5 seconds; the screen will display test values. Use the up arrow (▲) to increase digits or the down arrow (▼) to decrease digits until the displayed value is 1413. Release the button. Wait for "1413" to flash 3 times. Calibration 1413 is complete. Clean the electrode with distilled water.
  2. Calibrating 12.88 mS/cm: Press "MODE/CAL" to convert to EC mode. Place the electrode in a 12.88 mS/cm standard solution. Wait for the test to stabilize. Press "MODE/CAL" for 5 seconds; the screen will display test values. Use the up arrow (▲) to increase digits or the down arrow (▼) to decrease digits until the displayed value is 12.88. Release the button. Wait for "12.88" to flash 3 times. Calibration 12.88 is complete. Clean the electrode with distilled water.

ORP Calibration

  1. Press the "MODE" button to convert to ORP Mode.
  2. Place the electrode in the 256mV solution. Wait for the value to stabilize.
  3. Press "CAL" for 5 seconds, then release the button.
  4. Use the up arrow (▲) to add digits or the down arrow (▼) to reduce digits until the displayed value is 256. Release the button.
  5. Wait for "256" to flash 3 times.
  6. ORP calibration is complete. Clean the electrode.