3 www.watchgas.eu Detection User Manual Single Sustainable Gas Detector PDM D e t e c t i on Description PDM is a single sustainable gas detector designed to detect oxygen deficiency and

Single Sustainable

Contents. Contents Description Warning Caution 1 Product Overview 2 Activation And Deactivation 3 Mode 3.1 Detection Mode 3.2 Display Mode 3.3 Alarms / Battery / Test.. View


File Info : application/pdf, 14 Pages, 1.68MB

PDF WatchGas PDM User Manual EN V1 4
Detection
USER MANUAL
PDM+
Single Sustainable Gas Detector
www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

Contents

Contents

Description

Warning

Caution

1

Product Overview

2

Activation And Deactivation

3

Mode

3.1 Detection Mode

3.2 Display Mode

3.3 Alarms / Battery / Test Failure display

3.4 Default Alarm Set Points

3.5 Default Calibration Gas Concentrations

4

Event Log

5

Calibration

5.1. Fresh Air Calibration

5.2. Standard Gas Calibration

5.3. Remaining Calibration Day

5.4. Return to the Detection Mode

6

Self Test & Bump Test

6.1. Self Test

6.2. Bump Test

7

Sensor & Battery Replacement

7.1. Sensor Replacement

7.2. Battery Replacement

8

Specifications

8.1. General Specifications

8.2. Sensor Specifications

9

Certificates

10 Limited Warranty

User Manual
2 3 3 4 5 6 7 7 7 7 7 7 8 8 8 9 9 10 10 10 11 11 11 12 12 12 12 13 14

2

User Manual Single Sustainable Gas Detector PDM+

DD ee tt ee c t i o n

Description
PDM+ is a single sustainable gas detector designed to detect oxygen deficiency and presence of toxic gas in the ambient environment. The PDM+ is sustainable, hence its battery and sensor can be replaced. When turned on, PDM+ continuously monitors ambient air for the presence of a specific gas and alerts the user to potentially unsafe exposure with LED, vibrating, and audible alarms in the event that gas concentration exceeds alarm set points. The alarm set point, calibration range, and display configuration can be changed via WatchGas IR Link (Optional). After use, the instrument can be turned off.

Warning
· Any unauthorized attempt to repair or modify the product, or any other cause of damage beyond the range of the intended use, including damage by fire, lightening, or other hazard, voids liability of the manufacturer
· Activate this product only if sensor, visual, detection, and audible cover are clear from contaminants such as dirt and debris that could block the area where gas is to be detected
· Do not clean and rub the LCD screen of the products with a dry cloth or hands in hazardous environment to prevent the static electricity
· Perform cleaning and maintenance of the products in fresh air that is free of hazardous gases · Test the response of a sensor regularly by the gas concentration exceeding alarm set point · Test LED, audio and vibration manually · Gas concentration measurements by the sensor can vary based on the environment (pressure
and humidity) Therefore, calibration of PDM+ should be performed in the same (or similar) environment of the device's actual use · If the temperature changes sharply during use of the device (e g indoors vs outdoors), the value of the measured gas concentration can suddenly change Please use the PDM+ after the gas concentration value has stabilized · Severe vibration or shock to the device may cause a sudden reading change Please use PDM+ after the value of gas concentration has stabilized Excessive shock to PDM+ can cause the device and/or sensor to malfunction · All alarm value is set based on the alarm standard that is required by international standard. Therefore, alarm values should be changed only under the responsibility and approval of the administration of the work site where the instrument is used · Use IR communications in the safety zone which is free of hazardous gases · Replace the battery and sensor in clean environment, which is free of hazardous gas

3

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

User Manual

Caution

· Before operating this device, please read the manual carefully. · This device is not a measurement device, but a gas detector. · If calibration and self-test fails continuously, please do not use the device. · For the O2 detector, perform calibration every 30 days in the fresh air environment. · Before use, please check the activation date, and if the activation date is past, please do not use
the device. · Clean detectors with a soft cloth and do not use chemvical substances for cleaning. · To maintain 24 months life time, avoid the below activities except the necessary cases to check
events(Max/Min), lifetime/concentration, and alarm set points. Otherwise, the frequent use of the button will deplete the battery lifetime less than 24 months. 1. Push the button frequently without valid reasons. 2. Frequent alarm operation or alarms are remained for a long time. *Normal Alarm Use: 1 time
and 2 minutes per day.

4

User Manual Single Sustainable Gas Detector PDM+

7 3
6 5

1. Product Overview
DETECTOR COMPONENTS 1. Gas sensor 2. Gas type sticker 3. LCD display 4. Button 5. Buzzer
6 6. Optical alarm LEDs 7. IR communication port
4

DD ee tt ee c t i o n

1 2

TWA STEL

DISPLAY SYMBOLS Alarm
Low Alarm
High Alarm Stabilization Success Stabilization Fail Fresh Air Calibration Standard Gas Calibration TWA Time Weighted Average STEL Short Term Exposion Limit Remaining Months Remaining Days Remaining Time (Hours) Max Peak Value Min Peak Value
Unit of Measurement
Remaining battery life

5

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

2. Activation & Deactivation

User Manual

Note
Information about production dates? Please contact WatchGas by calling +31 (0)85 01 87 709 or by sending us
an e-mail to: info@watchgas.nl

ACTIVATION AND DEACTIVATION
In a safe environment, when pressing and holding the botton ( ) for 3 seconds, gas type and firmware version (ex. v2.2) will be displayed. For 10 seconds countdown, the device will be stabilized. After stabilization is completed, the device will move to Detection mode.

In the event that stabilization of the device fails, will appear on the display and gas measurement mode will not be entered. In this case, perform calibration or contact authorized reseller for repair/ return information. To deactivate the device, pleas press and hold the key( ) for countdown.
Caution
Although the PDM+ is calibrated before it leaves the factory, a proper calibration is advisable before using it at the work site. The user should check whether the device is properly detecting hazardous concentrations and make sure that the detecting section of
the device is not blocked with materials impairing the detection.
6

User Manual Single Sustainable Gas Detector PDM+

DD ee tt ee c t i o n

3. Mode
3.1 MEASUREMENT MODE

When activated, in measuring mode, gas concentration will appear on the screen. Oxygen concentration is displayed in percent by volume (%vol) and toxic concentration is displayed in parts per million (ppm).

3.2 DISPLAY MODE In the Measuring Mode, by pressing key ( ) for one second, the following icons will appear in order:
(only for oxygen) -> -> STEL value -> TWA value-> Clr Max-> Clr TWA/STEL ->1st alarm set point --> 2nd alarm set point -> STEL setpoint -> TWA Set point-> Firmware version -> Remaining Calibration Day -> Calibration
If you press the key or do not push any button for a second, the PDM+ will return to Measuring Mode.
3.3. ALARM ACTIVATION & CONFIGURING ALARM SET POINTS When a gas concentration exceeds alarm set points, or will be displayed and the device will vibrate, flash its LEDs, and beep. To stop alarms, evacuate immediately to a clean air location.
When bump test or calibration failed, will be displayed and the PDM+ will beep intermittently.

For high and low alarm

For TWA and STEL Alarm

To configurate the alarm setpoints, please follow the steps below.

· Press the button until the above alarm setpoint is displayed

· Press and hold the button for three seconds and the first digit of alarm setpoint starts to blink

· To increase the value, press the button for one second

· To save the alarm setpoints, press the key for 3 seconds

* Ensure that the second alarm setpoint must be greater than first alarm setpoint.

* Ensure Standard Factory alarm setpoints vary depending on countries, states, and companies.

Caution
Before changing alarm setpoints, please ensure the alarm set points are in compliance with your local guidelines.

3.4. DEFAULT ALARM SET POINTS

Gas

O2

CO

H2S

Low alarm 19% 25 ppm 5 ppm

H2

SO2

100 ppm 1 ppm

Cl2

NH3

NO2

1.5 ppm 20 ppm 5 ppm

High Alarm 23% 25 ppm 5 ppm 100 ppm 1 ppm 1.5 ppm 20 ppm 5 ppm

STEL

-

100 ppm 3.2 ppm -

0.3 ppm unknown 50 ppm 1

TWA

-

20 ppm 1.6 ppm -

0.3 ppm unknown 20 ppm 0.5 ppm

Disclaimer: The default Set Points of these gases: H2, SO2, Cl2, NH3 and NO2, are subject to change without notice.
NOTE: Option to manually change alarm setpoints can be unlocked using the IR-link.

7

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

User Manual

3.5. DEFAULT CALIBRATION GAS CONCENTRATIONS

Gas

O2

CO

H2S

H2

SO2

Concentration

18% Vol. (99,9% N2)

50 ppm

10 ppm 25 ppm

700 ppm

5 ppm 10 ppm

Cl2 5ppm 10 ppm

NH3 50 ppm

NO2 5 ppm

H2O2
See our documentation on "Bump Test H2O2"

4. Event Log

Last 30 events are stored on the PDM+. Once 30 events are stored, the oldest log events get overwritten. Stored log events can be transferred via WatchGas-IR Link. Each alarm event is recorded as follows: · Types of alarms · Alarm concentration in ppm or % · Peak concentration · Alarm duration
5. Calibration

Caution
Initial calibration is performed on all devices prior to shipment. Once received, calibration should be regularly depending on frequency of use.

1. Enter calibration menu.

5.1. FRESH AIR CALIBRATION

Fresh AIr Calibration

Standard Gas Calibration

2. Press and hold the key for 5 seconds to enter the calibration mode ( ), icon and `CAL' mark will appear on the LCD. 3. Press the key for three seconds to initiate calibration. When calibration begins, a countdown (starting at 10) will appear on the screen.

Once completed, will appear on the LCD. The device will return to the calibration menu. If calibration fails, will appear on the LCD. Check that the air is clean and that no contaminants are blocking the sensor opening and try again. If fresh air calibration fails repeatedly, contact WatchGas.
8

User Manual Single Sustainable Gas Detector PDM+

DD ee tt ee c t i o n

5.2. STANDARD GAS CALIBRATION Attach the calibration adapter to the PDM+ and to a gas cylinder with a concentration matching the calibration concentration. Check 3.1. Display Mode to check the calibration concentration.
Press and hold the key for 5 seconds to enter calibration mode ( ), icon and `CAL' mark will appear on the LCD. Press the key again for a second when CAL is displayed, to switch to standard gas
calibration, appears. Start the flow of the gas cylinder by opening the valve. Press the key for three seconds to initiate calibration. When calibration begins, a countdown will appear on the screen. The duration of the countdown depends on the sensor type and can be change by the IR link (configurable with WatchGas IR-Link). Confirm the calibration value by pressing the button once.

To set the calibration setpoints, please follow the steps below.
· Press the button until the above calibration setpoint is displayed. · Press and hold the button for three seconds and the first digit of calibration setpoint starts to blink. · To increase the value, press the button for one second . · To save the calibration setpoints, press the button for 3 seconds. · Calibration will start Once completed, icon will appear several seconds on the display. Then, the device will return to Detection mode. 5.3. REMAINING CALIBRATION DAY In Measuring mode, scroll through the menu by pressing the button until the following is displayed:
The default setting is "N/A". To activate the remaining calibration day, set an interval on the "Cal Interval(day)" via IR LINK.
If you set the calibration interval via IR LINK, the remaining day will be displayed. To check the remaining day, press the button until the above image is displayed.
Caution
Calibration should be performed in a fresh-air environment that is free for any contaminants and other gases. Preferably, do not perform calibration in a confined space.

9

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

User Manual

Once completed, will appear on the LCD. After a few seconds, the PDM+ will return to Detection mode.
If calibration fails, will appear on the LCD. Check that the gas cylinder is not empty and that is has not expired. Also make sure that no contaminants are blocking the sensor opening and try again. If standard gas calibration fails repeatedly, contact WatchGas. 5.4. RETURN TO DETECTING MODE.
In the standard calibration mode , press the key for a second to toggle fresh air calibration, standard calibration, and ESC. In the ESC mode, press the key for 3 seconds, the PDM+ will get out of the calibration mode and return to Detection mode. NOTE: Span calibration value can be change in the device and IR link software

CALIBRATION CONCENTRATION.

Gas

O2

CO

Concentration

0.0%Vol. (99,9% N2)

100 ppm

H2S 50 ppm

6. Self Test & Bump test

H2

SO2

500 ppm 10 ppm

Cl2 10 ppm

NH3

NO2

100 ppm 10 ppm

6.1 SELF TEST The default interval of the Self-Test is 20hr, meaning the PDM+ will ask for a self test after each 20 hours of use. By default the Self-Test setting is turned off. The interval is configurable via IR-Link between 8~20 hours. The self test can also be switched off via IR-Link. When the interval is activated, STS message will flash. The message will flash until users perform the Self test. Once you press the button, it will test buzzer, LED, Vibration, LCD, and show alarm thresholds. After the test is completed, END message with icon will be displayed. (Users are required to check the test processes.)
10

User Manual Single Sustainable Gas Detector PDM+

DD ee tt ee c t i o n

6.2. BUMP TEST The interval of Bump test is 1~365days, and the default is switched off. To initiate the bump test, set the bump test interval. Once the bump test interval expires, Bts message will flash.

Put the PDM+ in the docking station with a valid and full gas cylinder. Alternatively, attach the calibration adapter to PDM+ and a valid and full gas cylinder. Press and hold the key for 3 seconds, the TST message will be displayed for 45 seconds (To cancel, press the button for one second). Within the 45 seconds, start the flow from the gas cylinder. If no gas is applied, the bts message will flash again.

If the test is successful, SUC message with will be displayed for 30 seconds. Stop the flow, remove the calibration adapter. If the test fails, FA message with , will be displayed and BTS message will be flashing until the test is successful. If the bump test repeatedly fails, contact WatchGas.

7. Sensor & Battery Replacement
Caution
It is absolutely prohibited to replace battery at potential explosion or dangerous regions. Replace the battery in a clean environment, which has no hazardous gases. Replacement of components can invalidate the intrinsic safety function.
Replacing the sensor and battery should be performed by authorized sellers, agents, distributors, or managers. Contact WatchGas if replacement is needed.
Disassembly should be necessary only for sensors & battery replacement. After the sensor replacement, Fresh air and SPAN calibration should be done. Before disassembling, please
turn off the power and remove screws. Only use SB-AA02 3.6V battery.
7.1. SENSOR REPLACEMENT 1. Deactivate the detector 2. Remove the 6 screws on the back case and carefully open te case. 3. Remove the 2 screws on the PCB board. 4. After removing the battery, replace with the new sensor matching with the gas type. For instance,
If you have the PDM+ CO, the CO sensor should be used for the replacement. 5. Assemble the detector. 6. Have the sensor stabilized for 5 minutes before use. 7. After assembling, perform the fresh air calibration and standard calibration with the concentration
in this manual. 8. Check the settings.

11

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

User Manual

7.2. BATTERY REPLACEMENT 1. Deactivate the detector 2. Remove the 6 screws on the back case. 3. Replace battery with original SB-AA02 3.6V battery only 4. Assemble the detector. 5. After assembling, perform the fresh and standard calibration. 6. Before use, have the sensor stabilized for 5 minutes. Some sensor may require al longer
stabilization time.

8. Specifications

8.1. GENERAL SPECIFICATIONS

Size

48mm(W) x 85mm(H) x 22mm(D)

Weight Sensor technology

93g (Toxic), 104g (O2) (Battery, clip included) Electrochemical Cell

Temperature Humidity

-40ºC ~ +50ºC (for Toxic) / -35ºC ~ +50ºC (for O2) 5% ~ 95% RH (Non-condensing)

Alarm type

High Alarm, Low Alarm, TWA Alarm, STEL Alarm, Over range Alarm, Battery low alarm, Bumptest and calibration due notification

Alarm signal

Acoustic: 95dB @ 30cm Visual: Red flashing LED's Vibration: Vibration

Display

LCD Display

Calibration

2-point calibration, zero and span

Event log

30 Most recent events

Battery

Lithium Primary Battery SB-AA02(P) 3.6V, 1.2Ah

Measurement

Diffusion

Housing

Polycarbonate and rubber

Accuracy deviation

2-3%

IP-Rating

IP67

Safety certifications

ATEX: II 1G Ex ia IIC T4 Ga INMETRO: Ex ia IIC T4 Ga IECEx: Ex ia IIC T4 Ga CE: Conformité Européenne

Warranty

24 Months

8.2. SENSOR SPECIFICATIONS

Model

Detectable Gas Ranges

PDM+ O

0 - 30 %vol

PDM+ CO

0 - 500 ppm

PDM+ CO High Range

0-2000 ppm

PDM+ HS

0 - 100 ppm

PDM+ H2 PDM+ SO

0 - 1000 ppm 0 - 50 ppm

PDM+ NH3

0 - 100 ppm

PDM+ HO NON-ATEX

0 - 99 ppm

PDM+ NO2

0 - 20 ppm

Resolution 0.1 %vol 1 ppm 1 ppm 0.1 ppm 1 ppm 0.1ppm 1 ppm 0.1 ppm 0.1 ppm

Article Numbers 7192002 7192001 7192009 7192000 7192005 7192004 7192003 7192007 7192011

12

User Manual Single Sustainable Gas Detector PDM+

DD ee tt ee c t i o n

9. Certificates

Intrinsic Safety: The detector is in conformity of the following standards

IECEx:

Ex ia IIC T4 Ga 12 3 4 5

1. Explosion Protected 2. Protection Concept

IECEx

IECEx KTL 19.0019X

3. Gas Group

4. Temperature Classification

ATEX:

2198 II 1 G Ex ia IIC T4 Ga IP67

5. Equipment Protection level

KRH 19 ATEX 1022X

Directive 2014/34/EU

KCS:

Ex ia IIC T4

KTL 19-KA2BO-0491X

INMETRO

Ex ia IIC T4 Ga BRA-19-GE-0022X

Compliance: Electromagnetic Compatibility Directive 2014/30/EU Standards: The electrical apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents, was found to comply with the following standards: · IEC 60079-0: 2011 Ed. 6 · IEC 60079-11: 2011 Ed 6 · IEC 60079-26 : 2014-10 Ed 3 · EN 60079-0: 2012+A11:2013 · EN 60079-11: 2012

Manufacturing Approval: The detector manufacturer is certified compliant with ISO 9001:2000 provisions

13

www.watchgas.eu

DD ee tt ee cc t i o n

PDM+ Single Sustainable Gas Detector

10. Limited Warranty

User Manual

WATCHGAS warrants this product to be free of defects in workmanship and materials-under normal use and service-for two years from the date of purchase from the manufacturer or from the product's authorized reseller.
The manufacturer is not liable (under this warranty) if its testing and examination disclose that the alleged defect in the product does not exist or was caused by the purchaser's (or any third party's) misuse, neglect, or improper installation, testing, or calibrations. Any unauthorized attempt to repair or modify the product, or any other cause of damage beyond the range of the intended use, including damage by fire, lightening, water damage or other hazard, voids liability of the manufacturer.
In the event that a product should fail to perform up to manufacturer specifications during the applicable warranty period, please contact the product's authorized reseller or WATCHGAS service center at +31 (0)85 01 87 709 for repair/return information.

Detection

WatchGas B.V. Klaverbaan 121 2908 KD Capelle aan den IJssel +31 (0)85 01 87 709 The Netherlands info@watchgas.eu - www.watchgas.eu
15-09-21 V1.4