To ensure accurate measurements with the EE150 Humidity and ឧបករណ៍បញ្ជូនសីតុណ្ហភាព, follow these guidelines as per the user manual:
1. **Temperature Equilibrium:** It is crucial for accurate measurement that the temperature of the probe and sensing head matches the air temperature being measured. Avoid temperature gradients along the probe by installing it entirely inside the duct. If there are temperature differences inside and outside the duct, thermally isolate the probe part outside the duct.
2. **Mechanical Stress:** Avoid exposing the device and sensing head to extreme mechanical stress. Do not touch the sensing head and do not attempt to remove the fixed filter cap.
3. **ESD Sensitivity:** The stainless steel probe is ESD sensitive. Handle it accordingly and do not connect it to ground potential.
4. **Installation:** Mount the EE150 sensor in a way that minimizes temperature gradients and ensures the sensing head is at the same temperature as the air being measured.
By following these guidelines, users can ensure accurate and reliable measurements with the EE150 Humidity and ឧបករណ៍បញ្ជូនសីតុណ្ហភាព for HVAC applications.