ដើម្បីក្រិតតាមខ្នាត ឧបករណ៍វាស់សមត្ថភាពឌីជីថល Tektronix 6013អនុវត្តតាមជំហានទាំងនេះ៖
1. Make sure the meter is turned off.
2. Locate the Zero Adjust xxxx on the front panel of the meter.
3. Set the Function xxxx to the lowest capacitance range (200pF).
4. Connect the test alligator clips to each other to short them.
5. Turn on the meter by pressing the POWER switch.
6. While observing the LCD display, rotate the Zero Adjust xxxx until the display shows "0".
7. The meter is now calibrated for the lowest capacitance range.
8. Repeat the calibration process for the other capacitance ranges if necessary.
Note: The Zero Adjust function may not work when using the external alligator clips if the capacitance being measured is beyond 20nF.
For more detailed instructions and illustrations, refer to the "OPERATION" section of the សៀវភៅណែនាំអំពីការវាស់ស្ទង់សមត្ថភាពឌីជីថល Tektronix 6013.