To select the range and set the output voltagអ៊ី នៅលើ Time Electronics 1006 Portable DC Millivolt ប្រភពអនុវត្តតាមជំហានទាំងនេះ៖
1. ថាមពលនៅលើឯកតា។
2. Check the battery level indicator on the top of the unit. Green indicates that the batteries are ready for use, while red indicates that the batteries need to be changed or recharged.
3. Select the desired range using the push button. The 1006 offers three ranges: 0 to 999.9mV (0.1 mV steps), 0 to 99.99mV (10 µV steps), and 0 to 9.999mV (1 µV steps).
4. Dial the desired output voltage using the thumbwheel switches. The output voltage can be adjusted within the selected range.
១.៧ ។ វ៉ុលលទ្ធផលtage polarity can be selected using the front panel switch. The "off" position on the 1006 disconnects the output and provides a short circuit on the output terminals.
សូមយោងទៅ Time Electronics 1006 Portable DC Millivolt ប្រភព សៀវភៅណែនាំអ្នកប្រើប្រាស់ សម្រាប់ការណែនាំលម្អិតបន្ថែមទៀត និងការប្រុងប្រយ័ត្នសុវត្ថិភាព។