ℹ️ Document Conversion Notice: This page was converted from the original file for easier reading. Diagrams/images may appear only in the original PDF below.
Document preview
File info: application/pdf · 2 pages · 1.01MB

FSP Protek Power 80-100 Watt Medical & ITE Power Supplies

Model Series: PM101

Description

The PM101 series comprises AC-DC switching power supplies designed in a compact 2 x 4 x 1.29 inch package. These units can deliver 100 watts of continuous power with 7.5 CFM forced air cooling or 80 watts with convection cooling. Constructed on a printed circuit board, they are suitable for medical applications, information technology, and industrial applications. Approvals to IEC60601-1 and IEC62368-1 safety standards streamline design-in processes and reduce end-equipment compliance costs.

Certifications: [CE Mark] [RoHS Compliant]

Features

  • BF class insulation
  • Operation altitude up to 5000 meters
  • 2 x 4 inch footprint with 1.29 inch low profile
  • Less than 175 μA leakage current
  • Wide input range: 80-264 VAC
  • Meets EN55011 / EN55032 and FCC Class B standards
  • Short-circuit protection
  • Compliant with RoHS requirements
  • No load power consumption less than 0.15W

Safety Standard Approvals

[UL Certified] UL ES 60601-1, CSA C22.2 No. 60601-1 (File No. E178020)

[TÜV Certified] TÜV EN 60601-1

[UL Certified] UL 62368-1, CSA C22.2 No. 62368-1

[TÜV Certified] TÜV EN 62368-1

Input Specifications

  • Input voltage: 80-264 VAC
  • Power derating: Derates linearly from 100% at 90VAC to 90% at 85VAC and 80% at 80VAC.
  • Input frequency: 47-63 Hz
  • Input current: 2.0 A (rms) for 115 VAC, 1.2 A (rms) for 230 VAC
  • Earth leakage current: 175 μA max. @ 264 VAC, 63 Hz
  • Touch current: 100 μA max. @ 264 VAC, 63 Hz

Output Specifications

  • Output voltage/current: See rating chart.
  • Maximum output power: See rating chart.
  • Ripple and noise: 1% peak to peak maximum.
  • Over voltage protection: Set at 112-140% of its nominal output voltage; latches by recycle input to reset.
  • Short circuit protection: Automatic recovery.
  • Over temperature protection: Latching by recycle input to reset.
  • Temperature coefficient: All outputs ±0.04% /°C maximum.
  • Transient response: Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 μs after a 25% step load change.

Environmental Specifications

  • Operating temperature: -20°C to +70°C
  • Storage temperature: -40°C to +85°C
  • Relative humidity: 5% to 95% non-condensing
  • Temperature derating: Derates from 100% at +50°C linearly to 50% at +70°C, applicable to convection and forced-air cooling conditions.

General Specifications

  • Switching frequency: 65 KHz (typical)
  • Efficiency: See rating chart.
  • Hold-up time: 10 ms minimum at 100 W load and 115 VAC; 10 ms minimum at 80 W load and 100 VAC.
  • Line regulation: ±0.5% maximum at full load.
  • Inrush current: 80 A @ 115 VAC or 160 A @ 230 VAC, at 25°C cold start.
  • Withstand voltage: 4000 VAC from input to output, 1500 VAC from input to ground, 1500 VAC from output to ground.
  • MTBF: 150,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F.
  • EMC Performance:
    • EN55011/ EN55032: Class B conducted, class B radiated
    • EN61000-3-2: Harmonic distortion, class A
    • EN61000-3-3: Line flicker
    • EN60601-1-2, EN55024:
    • EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
    • EN61000-4-3: Radiated immunity, 9-28 V/m
    • EN61000-4-4: Fast transient/burst, ±2 KV
    • EN61000-4-5: Surge, ±1 KV diff., ±2 KV com
    • EN61000-4-6: Conducted immunity, 10 Vrms
    • EN61000-4-8: Magnetic field immunity, 30 A/m
    • EN61000-4-11: Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms

Output Voltage/Current Rating Chart

ModelV1Min. loadMax. Current at convectionMax. Current at 7.5 CFMTol.Ripple & Noise(2)Max. Power(1)Efficiency (typical) 115/230 Vac
PM101-12A12 V0 A6.67 A8.34 A±2%120 mV80 W / 100 W87/90%
PM101-13A15 V0 A5.34 A6.67 A±2%150 mV80 W / 100 W87/90%
PM101-13-1A18 V0 A4.45 A5.56 A±2%180 mV80 W / 100 W87/90%
PM101-14A24 V0 A3.34 A4.17 A±2%240 mV80 W / 100 W88/90%
PM101-15A28 V0 A2.86 A3.58 A±2%280 mV80 W / 100 W88/90%
PM101-16-1A32 V0 A2.50 A3.13 A±2%320 mV80 W / 100 W88/90%
PM101-17A36 V0 A2.23A2.78 A±2%360 mV80 W / 100 W88/90%
PM101-18A48 V0 A1.67A2.09A±2%480 mV80 W / 100 W88/90%

Notes:

  1. The first value of max. power is at convection cooling. The second value is with 7.5 CFM forced air provided by user.
  2. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum (or electrolytic) capacitor in parallel with a 0.1 μF ceramic capacitor across the output, except model PM101-12A which uses a 22 μF tantalum (or electrolytic) capacitor in parallel with a 0.1 μF ceramic capacitor across the output.

Mechanical Specifications

Dimensions: 2 x 4 x 1.29 inches [50.8 x 101.60 x 32.8 mm]. Tolerance: 0.02 [0.5] maximum.

Input Connector (CN1): Molex header 09-65-2038, mating with Molex housing 09-50-1031 or equivalent.

Output Connector (CN2): Molex header 09-65-2048, mating with Molex housing 09-50-1041 or equivalent.

Ground Tab: 0.25 [6.35] x 0.032 [0.8] fast-on connector.

Mounting: Four mounting holes, 0.157 [4.00] diameter.

Emissions Compliance: To ensure compliance with Level B emissions, connect the three "*" marked mounting holes with metallic standoffs to the chassis.

Weight: 155 grams (0.34 lbs.) approx.

Diagram Description: The diagram shows a rectangular power supply unit with connectors labeled CN1 and CN2. CN1 is a 3-pin connector on one side, and CN2 is a 4-pin connector on the opposite side. A ground tab is indicated. Mounting holes are shown around the perimeter. Dimensions are provided in inches and millimeters.

Output Power Derating Curves

Curve 1: Output Power vs. Input Voltage

This graph shows how output power is derated based on input voltage. The Y-axis represents Output Power (%) from 80% to 100%. The X-axis represents Input Voltage (V) from 80 to 264. The curve indicates full power (100%) is available from approximately 90V to 264V, with derating occurring below 90V, reaching 80% at 80V.

Curve 2: Output Power vs. Ambient Temperature

This graph illustrates output power derating based on ambient temperature. The Y-axis represents Output Power (%) from 50% to 100%. The X-axis represents Ambient Temperature (°C) from -20 to 70. The curve shows full power (100%) is available up to 50°C, and derates linearly to 50% power at 70°C.

Pin Chart

CN1CN2
PIN NO.1231234
PolarityLiveVoidNeutralV1V1Common ReturnCommon Return

Company Information

FSP Power Solution GmbH

Address: Jakobshöhe 16 / 41066 Mönchengladbach / Germany

Email: info@fsp-ps.de

Website: www.fsp-ps.de, www.fsp-group.com.ru

Phone: +49 (0) 2161 4952490

Fax: +49 (0) 2161 495249 21

Original Document

If the viewer doesn’t load, open the PDF directly.