SCXI-1120

CALIBRATION PROCEDURE
SCXITM-1120/D
Introduction
This document contains information and step-by-step instructions for calibrating the National Instruments (NI) SCXI-1120 and SCXI-1120D modules.
What Is Calibration?
Calibration consists of verifying the measurement accuracy of a module and adjusting for any measurement error. Verification is measuring the performance of the module and comparing these measurements to the factory specifications. During calibration, you supply and read voltage levels using external standards, then you adjust the module calibration circuitry. This circuitry compensates for any inaccuracies in the module, and returns the accuracy of the module to the factory specifications.
Why Should You Calibrate?
The accuracy of electronic components drifts with time and temperature, which can affect measurement accuracy as a module ages. Calibration restores these components to their specified accuracy and ensures that the module still meets NI standards.
How Often Should You Calibrate?
The measurement requirements of your application determine how often the SCXI-1120/D module needs to be calibrated to maintain accuracy. NI recommends that you perform a complete calibration at least once every year. You can shorten this interval to 90 days or six months based on the demands of your application.

National InstrumentsTM, NITM, ni.comTM, and SCXI TM are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies.

370268B-01

© 2000, 2001 National Instruments Corp. All rights reserved.

September 2001

Equipment and Other Test Requirements

This section describes the test equipment, software, documentation, and test conditions required for calibrating the SCXI-1120/D modules.

Test Equipment

Calibration requires a high-precision voltage source with at least 50 ppm accuracy and a multiranging 5 1/2 digit digital multimeter (DMM) with 15 ppm accuracy.

Instruments
NI recommends the following instruments for calibrating SCXI-1120/D modules: · Calibrator--Fluke 5700A · DMM--NI 4060 or HP 34401A
If these instruments are not available, use the accuracy requirements listed previously to select substitute calibration instruments.

Connectors
If you do not have custom connection hardware, you need the following connectors: · Terminal block, such as the SCXI-1320 · Shielded 68-pin connector cable · 50-pin ribbon cable · 50-pin breakout box · SCXI-1349 adapter
These components give easy access to the individual pins on the SCXI-1120/D module front and rear connectors.

Software and Documentation
No special software or documentation is necessary to calibrate the SCXI-1120/D module. This calibration document contains all the information you need to complete the verification and adjustment procedures. If you would like more information on the module, refer to the SCXI-1120/D User Manual.

SCXI-1120/D Calibration Procedure

2

ni.com

Test Conditions
Calibration

Follow these guidelines to optimize the connections and environment during calibration: · Keep connections to the SCXI-1120/D module short. Long cables and
wires act as antennae, picking up extra noise and thermal offsets that can affect measurements. · Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. · Maintain the temperature between 18­28 °C. · Keep relative humidity below 80%. · Allow a warm-up time of at least 15 minutes for the SCXI-1120/D module to ensure the measurement circuitry is at a stable operating temperature.
The calibration procedure for the SCXI-1120/D module consists of the following steps: 1. Set up the module for testing. 2. Verify the existing operation of the module to determine whether it is
operating within its specifications. 3. Adjust the module with respect to a known voltage source. 4. Verify that the module is operating within its specifications after
adjustments.

© National Instruments Corporation

3

SCXI-1120/D Calibration Procedure

Setting Up the Module
Refer to Figures 1 and 2 while performing the following steps to set up the SCXI-1120/D module for verification: 1. Remove the grounding screw from the module. 2. Remove the cover on the module to access the potentiometers.

2

1

1 Grounding screw

2 Module cover
Figure 1. Grounding Screw and Module Cover Removal
3. Remove the side plate of the SCXI chassis. 4. Install the SCXI-1120/D into slot 4 of the SCXI chassis.

SCXI-1120/D Calibration Procedure

4

ni.com

®

SCXI MAINFRAME INNSATTRIOUNMAELNTS

2

3

1 Thumbscrews

1

2 Slot 4

3 Side plate

Figure 2. Side Plate Removal and Module Installation

The SCXI-1120/D module does not need to be connected to a data acquisition (DAQ) device. Leave the configuration of the digital jumpers W41­W43 and W46 unchanged because they do not affect this procedure.

Configuring the Gain Jumpers
Each input channel has two user-configurable gain stages. The first-stage gain provides gains of 1, 10, 50, and 100. The second-stage gain provides gains of 1, 2, 5, 10, and 20. Table 1 shows the jumper reference designators for gain selection associated with each channel. Table 2 shows how to position each jumper to select the desired gain for each channel.

Table 1. Gain Jumper Reference Designators

Input Channel Number
0

First-Stage Gain Jumper
W1

Second-Stage Gain Jumper
W9

1

W2

W10

2

W3

W11

© National Instruments Corporation

5

SCXI-1120/D Calibration Procedure

Table 1. Gain Jumper Reference Designators (Continued)

Input Channel Number

First-Stage Gain Jumper

Second-Stage Gain Jumper

3

W4

W12

4

W5

W13

5

W6

W14

6

W7

W15

7

W8

W16

Gain First Stage
Second Stage

Table 2. Gain Jumper Positions

Setting

Jumper Position

1

D

10

C

50

B

100

A (factory setting)

1

A

2

B

5

C

10

D (factory setting)

20

E

To change the gain setting of a specified channel on the module, move the appropriate jumper on the module to the position indicated in Table 2. Refer to Table 1 for jumper reference designators, and Figure 3 for the location of the jumpers.

SCXI-1120/D Calibration Procedure

6

ni.com

3 1

4

5

V25 V26 W27 W28 W29W30 W31V32V33 W34 W35 W36 V37 V38 V39 V40

6
2
W43
7
W45

W42

1 12

22 W14 U14

AD705 GAINDIP

AD

Q12 R18

W6 ABCDABCDEAB

8

9 10 11

12 Front

13 Rear

1 Thumbscrews 2 Front Connector 3 Product Name, Assembly Number,
and Serial Number 4 Output Null Adjust Potentiometers 5 Second-Stage Filter Jumpers 6 Rear Signal Connector

7 SCXIbus Connector 8 Input Null Adjust Potentiometers 9 First-Stage Gain Jumpers 10 Second-Stage Gain Jumpers 11 First-Stage Filter Jumpers 12 Terminal Block Mounting Hole 13 Grounding Screw

Figure 3. SCXI-1120/D Parts Locator Diagram

Note The SCXI-1120D module has an additional fixed pre-stage gain of 0.5.

© National Instruments Corporation

7

SCXI-1120/D Calibration Procedure

The order of the settings for the first- and second-stage gain does not matter as long as the first-stage gain multiplied by the second-stage gain--multiplied by 0.5 when using the SCXI-1120D--equals the desired final gain value. · SCXI-1120--To determine the overall gain of a given channel on the
SCXI-1120 module:
First-Stage Gain × Second-Stage Gain = Overall Gain · SCXI-1120D--To determine the overall gain of a given channel on the
SCXI-1120D module:
(First-Stage Gain × Second-Stage Gain) × 0.5 = Overall Gain

Configuring the Filter Jumpers
Each input channel also has two user-configurable filter stages. The SCXI-1120 module ships in the 4 Hz position and the SCXI-1120/D module ships in the 4.5 kHz position. Refer to Table 3 or 4 to find the correct jumper setting for the desired cutoff frequency. Figure 3 shows the locations of the jumper blocks on the SCXI-1120/D modules. Verify that both filter stages are set to the same filter setting to ensure that you achieve the required bandwidth.

Table 3. SCXI-1120 Filter Jumper Settings

Input Channel Number

First Filter Jumper

4 Hz (Factory Setting)

10 kHz

Second Filter Jumper

4 Hz (Factory Setting)

10 kHz

0

W17-A

W17-B

W25

W26

1

W18-A

W18-B

W27

W28

2

W19-A

W19-B

W29

W30

3

W20-A

W20-B

W31

W32

4

W21-A

W21-B

W33

W34

5

W22-A

W22-B

W35

W36

6

W23-A

W23-B

W37

W38

7

W24-A

W24-B

W39

W40

SCXI-1120/D Calibration Procedure

8

ni.com

Input Channel Number
0 1 2 3 4 5 6 7

Table 4. SCXI-1120D Filter Jumper Allocation

First Filter Jumper

Second Filter Jumper

4.5 kHz (Factory Setting)

22.5 kHz

4.5 kHz (Factory Setting)

22.5 kHz

W17-A

W17-B

W26

W25

W18-A

W18-B

W28

W27

W19-A

W19-B

W30

W29

W20-A

W20-B

W32

W31

W21-A

W21-B

W34

W33

W22-A

W22-B

W36

W35

W23-A

W23-B

W38

W37

W24-A

W24-B

W40

W39

Verifying the Operation of the Module
The verification procedure determines how well the SCXI-1120/D module is meeting its specifications. You can use this information to select the appropriate calibration interval for your application. Refer to the Setting Up the Module section for information on how to configure the channel filter and channel gain.
Complete the following steps to verify the operation of the SCXI-1120/D module:
1. Read the Test Conditions section in this document.
2. Refer to Table 7 for the SCXI-1120 module or Table 8 for the SCXI-1120D module for all acceptable settings for the module. Although NI recommends verifying all ranges and gains, you can save time by checking only those ranges used in your application.
3. Set the channel filter for all channels on the module to 4 Hz for the SCXI-1120 module or 4.5 kHz for the SCXI-1120D module.
4. Set the channel gain on all channels to the gain you want to test, starting with the smallest gain available for the module. Available gains are shown in Tables 7 and 8.
5. Connect the calibrator to the analog input channel you are testing, starting with channel 0.
If you do not have an SCXI terminal block such as the SCXI-1320, refer to Table 5 to determine the pins on the 96-pin front connector that correspond to the positive and negative inputs of the specified channel.

© National Instruments Corporation

9

SCXI-1120/D Calibration Procedure

For example, the positive input for channel 0 is pin A32, which is labeled CH0+. The negative input for channel 0 is pin C32, which is labeled CH0­.

Table 5. SCXI-1120/D Front Connector Pin Assignments

Pin Number 32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8

Column A CH0+ NP CH1+ NP NC NP CH2+ NP CH3+ NP NC NP CH4+ NP CH5+ NP NC NP CH6+ NP CH7+ NP NC NP NC

Column B NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP NP

Column C CH0­ NP CH1­ NP NC NP CH2­ NP CH3­ NP NC NP CH4­ NP CH5­ NP NC NP CH6­ NP CH7­ NP NC NP RSVD

SCXI-1120/D Calibration Procedure

10

ni.com

Table 5. SCXI-1120/D Front Connector Pin Assignments (Continued)

Pin Number Column A

7

NP

6

RSVD

5

NP

4

+5V

3

NP

2

CHSGND

1

NP

NP--No pin; NC--No connect

Column B NP NP NP NP NP NP NP

Column C NP
RSVD NP
MTEMP NP
DTEMP NP

6. Connect the DMM to the output of the same channel to which the calibrator was connected in step 5. Refer to Figure 4 to determine the pins on the 50-pin rear connector that correspond to the positive and negative outputs for the specified channel. For example, the positive output for channel 0 is pin 3, which is labeled MCH0+. The negative output for channel 0 is pin 4, which is labeled MCH0­.

© National Instruments Corporation

11

SCXI-1120/D Calibration Procedure

AOGND MCH0+ MCH1+ MCH2+ MCH3+ MCH4+ MCH5+ MCH6+ MCH7+ OUTREF
NC NC SERDATIN DAQD*/A SLOT0SEL* NC DIG GND NC SERCLK NC NC RSVD NC NC NC

12 34 56 78 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50

AOGND MCH0­ MCH1­ MCH2­ MCH3­ MCH4­ MCH5­ MCH6­ MCH7­ NC NC DIG GND SERDATOUT NC NC NC NC SCANCLK NC NC NC NC NC NC NC

Figure 4. SCXI-1120/D Rear Connector Pin Assignments
7. Set the calibrator voltage to the value specified by the Test Point entry listed in Table 7 for the SCXI-1120 module or Table 8 for the SCXI-1120D module.
8. Read the resulting output voltage on the DMM. If the output voltage result falls between the Upper Limit and the Lower Limit values, the module passed the test.
9. Repeat steps 5 through 8 for the remaining test points. 10. Repeat steps 5 through 9 for the remaining analog input channels. 11. Repeat steps 4 through 10 for the remaining gain settings specified in
the appropriate table. 12. Repeat steps 3 through 11, but set the channel filter to 10 kHz for the
SCXI-1120 module or 22.5 kHz for the SCXI-1120D module.
You have completed verifying the operation of the module.

SCXI-1120/D Calibration Procedure

12

ni.com

Adjusting Offset Null Values of the Module
Complete the following steps to adjust the offset null value:
1. Set the channel gain on all channels to a gain of 1. Set the filter value to 4 Hz for the SCXI-1120 module or 4.5 kHz for the SCXI-1120D module. Refer to the Setting Up the Module section in this document for information on how to set the channel gain.
2. Connect the calibrator to the analog input channel you want to adjust, starting with channel 0. Refer to Table 5 to determine the pins on the 96-pin front connector that correspond to the positive and negative inputs of the specified channel. For example, the positive input for channel 0 is pin A32, which is labeled CH0+. The negative input for channel 0 is pin C32, which is labeled CH0­.
3. Connect the DMM to the output of the same channel to which the calibrator was connected in step 2. Refer to Figure 4 to determine the pins on the 50-pin rear connector that correspond to the positive and negative outputs for the specified channel. For example, the positive output for channel 0 is pin 3, which is labeled MCH0+. The negative output for channel 0 is pin 4, which is labeled MCH0­.
4. Set the calibrator to output 0.0 V.
5. Adjust the output potentiometer of the channel until the DMM reading is 0 ±3.0 mV. Refer to Figure 3 for the potentiometer location and Table 6 for the potentiometer reference designator.

Table 6. Calibration Potentiometers Reference Designators

Input Channel Number

Input Null

Output Null

0

R08

R24

1

R10

R25

2

R12

R26

3

R14

R27

4

R16

R28

5

R18

R29

6

R20

R30

7

R21

R31

6. Set the channel gain on all channels to 1000.0.

© National Instruments Corporation

13

SCXI-1120/D Calibration Procedure

7. Adjust the input potentiometer of channel 0 until the DMM reading is 0 ±6.0 mV. Refer to Figure 3 for the potentiometer location and Table 6 for the potentiometer reference designator.
8. Repeat steps 1 through 7 for the remaining analog inputs.
You have completed adjusting the module.
Verifying Adjusted Values
After you complete the adjustment procedure, it is important to verify the accuracy of the adjusted values by repeating the procedure in the Verifying the Operation of the Module section. Verifying the adjusted values ensures the module is operating within its specifications after adjustments.
Note If the SCXI-1120/D module fails after calibration, return it to NI for repair or replacement.

SCXI-1120/D Calibration Procedure

14

ni.com

Specifications

Gain 0.011 0.011 0.011 0.021 0.021 0.021 0.051 0.051 0.051 0.11 0.11 0.11 0.21 0.21 0.21 0.51 0.51 0.51
1 1 1

Table 7 contains test specifications for SCXI-1120 modules. Table 8 contains test specifications for SCXI-1120D modules. If the module was calibrated within the last year, the output from the module should fall between the Upper Limit and Lower Limit values.

Table 7. SCXI-1120 Specifications

Test Point (V)
232.5
0

4Hz filter setting

Upper Limit Lower Limit

(V)

(V)

2.346996

2.303004

0.006888

­0.006888

10kHz filter setting

Upper Limit Lower Limit

(V)

(V)

2.349248

2.300752

0.009140

­0.009140

­232.5 186 0 ­186 93 0 ­93 46.5 0

­2.346996 3.751095 0.006922 ­3.751095 4.687000 0.006784 ­4.687000 4.686925 0.006709

­2.303004 3.688905 ­0.006922 ­3.688905 4.613000 ­0.006784 ­4.613000 4.613075 ­0.006709

­2.349248 3.753353 0.009180 ­3.753353 4.689236 0.009020 ­4.689236 4.689186 0.008970

­2.300752 3.686647 ­0.009180 ­3.686647 4.610764 ­0.009020 ­4.610764 4.610814 ­0.008970

­46.5 23.25
0 ­23.25

­4.686925 4.686775 0.006559 ­4.686775

­4.613075 4.613225 ­0.006559 ­4.613225

­4.689186 4.689056 0.008840 ­4.689056

­.610814 4.610944 ­0.008840 ­4.610944

9.3

4.686353

4.613647

4.688626

4.611374

0

0.006138

­0.006138

0.008410

­0.008410

­9.3

­4.686353

­4.613647

­4.688626

­4.611374

4.65

4.691704

4.608296

4.693926

4.606074

0

0.011488

­0.011488

0.013710

­0.013710

­4.65

­4.691704

­4.608296

­4.693926

­4.606074

© National Instruments Corporation

15

SCXI-1120/D Calibration Procedure

Gain 2 2 2 5 5 5 10 10 10 20 20 20 50 50 50
100 100 100 200 200 200 250 250 250 500 500

Table 7. SCXI-1120 Specifications (Continued)

4Hz filter setting

10kHz filter setting

Test Point (V)

Upper Limit Lower Limit Upper Limit Lower Limit

(V)

(V)

(V)

(V)

2.325

4.690653

4.609347

4.692876

4.607124

0

0.010437

­0.010437

0.012660

­0.012660

­2.325

­4.690653

­4.609347

­4.692876

­4.607124

0.93

4.690498

4.609502

4.692726

4.607274

0

0.010282

­0.010282

0.012510

­0.012510

­0.93

­4.690498

­4.609502

­4.692726

­4.607274

0.465

4.690401

4.609599

4.692626

4.607374

0

0.010185

­0.010185

0.012410

­0.012410

­0.465

­4.690401

­4.609599

­4.692626

­4.607374

0.2325

4.690139

4.609861

4.692416

4.607584

0

0.009924

­0.009924

0.012200

­0.012200

­0.2325

­4.690139

­4.609861

­4.692416

­4.607584

0.093

4.690046

4.609954

4.692331

4.607669

0

0.009831

­0.009831

0.012115

­0.012115

­0.093

­4.690046

­4.609954

­4.692331

­4.607669

0.0465

4.689758

4.610242

4.692066

4.607934

0

0.009542

-0.009542

0.011850

­0.011850

­0.0465

­4.689758

­4.610242

­4.692066

­4.607934

0.02325

4.689464

4.610536

4.691936

4.608064

0

0.009248

­0.009248

0.011720

­0.011720

­0.02325

­4.689464

­4.610536

­4.691936

­4.608064

0.0186

4.689313

4.610687

4.692016

4.607984

0

0.009097

­0.009097

0.011800

­0.011800

­0.0186

­4.689313

­4.610687

­4.692016

­4.607984

0.0093

4.689443

4.610557

4.692731

4.607269

0

0.009227

­0.009227

0.012515

­0.012515

SCXI-1120/D Calibration Procedure

16

ni.com

Table 7. SCXI-1120 Specifications (Continued)

4Hz filter setting

10kHz filter setting

Gain

Test Point (V)

Upper Limit Lower Limit Upper Limit Lower Limit

(V)

(V)

(V)

(V)

500

­0.0093

­4.689443

­4.610557

­4.692731

­4.607269

1000

0.00465

4.693476

4.606524

4.698796

4.601204

1000

0

0.013260

­0.013260

0.018580

­0.018580

1000

­0.00465

­4.693476

­4.606524

­4.698796

­4.601204

2000

0.002325

4.703044

4.596956

4.712556

4.587444

2000

0

0.022828

­0.022828

0.032340

­0.032340

2000

­0.002325

­4.703044

­4.596956

1Value available only when used with the SCXI-1327 high-voltage terminal block

­4.712556

­4.587444

Gain 0.011 0.011 0.011 0.0251 0.0251 0.0251 0.051 0.051 0.051 0.11 0.11 0.11 0.251

Test Point (V) 232.5 0
­232.5 186 0 ­186 93 0 ­93 46.5 0 ­46.5 18.6

Table 8. SCXI-1120D Specifications

4.5KHz filter setting

22.5KHz filter setting

Upper Limit Lower Limit Upper Limit Lower Limit

(V)

(V)

(V)

(V)

2.351764

2.298236

2.365234

2.284766

0.006230

­0.006230

0.019700

­0.019700

­2.351764 ­2.298236 ­2.365234 ­2.284766

4.698751

4.601249

4.733819

4.566181

0.007683

­0.007683

0.042750

­0.042750

­4.698751 ­4.601249 ­4.733819 ­4.566181

4.697789

4.602211

4.768769

4.531231

0.006720

­0.006720

0.077700

­0.077700

­4.697789 ­4.602211 ­4.768769 ­4.531231

4.698899

4.601101

4.841289

4.458711

0.007830

­0.007830

0.150220

­0.150220

­4.698899 ­4.601101 ­4.841289 ­4.458711

4.701669

4.598331

5.028819

4.271181

© National Instruments Corporation

17

SCXI-1120/D Calibration Procedure

Gain 0.251 0.251 0.5 0.5 0.5
1 1 1 2.5 2.5 2.5 5 5 5 10 10 10 25 25 25 50 50 50 100 100 100

Table 8. SCXI-1120D Specifications (Continued)

4.5KHz filter setting

22.5KHz filter setting

Test Point Upper Limit Lower Limit Upper Limit Lower Limit

(V)

(V)

(V)

(V)

(V)

0

0.010600

­0.010600

0.337750

­0.337750

­18.6

­4.701669 ­4.598331 ­5.028819 ­4.271181

9.3

4.697331

4.602669

4.703726

4.596274

0

0.006355

­0.006355

0.012750

­0.012750

­9.3

­4.697331 ­4.602669 ­4.703726 ­4.596274

4.65

4.697416

4.602584

4.710876

4.589124

0

0.006440

-0.006440

0.019900

­0.019900

­4.65

­4.697416 ­4.602584 ­4.710876 ­4.589124

1.86

4.697883

4.602117

4.732426

4.567574

0

0.006908

­0.006908

0.041450

­0.041450

­1.86

­4.697883 ­4.602117 ­4.732426 ­4.567574

0.93

4.698726

4.601274

4.768726

4.531274

0

0.007750

­0.007750

0.077750

­0.077750

­0.93

­4.698726 ­4.601274 ­4.768726 ­4.531274

0.465

4.700796

4.599204

4.841236

4.458764

0

0.009820

­0.009820

0.150260

­0.150260

­0.465

­4.700796 ­4.599204 ­4.841236 ­4.458764

0.18

5.070004

3.929996

4.870004

4.129996

0

0.530350

­0.530350

0.330350

­0.330350

­0.18

­5.070004 ­3.929996 ­4.870004 ­4.129996

0.086

4.360392

4.239608

4.825892

3.774108

0

0.022500

­0.022500

0.488000

­0.488000

­0.086

­4.360392 ­4.239608 ­4.825892 ­3.774108

0.038

3.879624

3.720376

4.810624

2.789376

0

0.039800

­0.039800

0.970800

­0.970800

­0.038

­3.879624 ­3.720376 ­4.810624 ­2.789376

SCXI-1120/D Calibration Procedure

18

ni.com

Table 8. SCXI-1120D Specifications (Continued)

4.5KHz filter setting

22.5KHz filter setting

Gain

Test Point Upper Limit Lower Limit Upper Limit Lower Limit

(V)

(V)

(V)

(V)

(V)

250

0.0125

3.277438

2.972563

4.830188

1.419813

250

0

0.091500

­0.091500

0.056751

­1.644250

250

­0.0125

­3.277438 ­2.972563 ­4.830188 ­1.419813

500

0.006

3.273770

2.726230

4.810770

1.189230

500

0

0.176000

­0.176000

1.713000

­1.713000

500

­0.006

­3.273770 ­2.726230 ­4.810770 ­1.189230

1000

0.0029

3.416058

2.383942

4.895058

0.904942

1000

0

0.342000

­0.342000

1.821000

­1.821000

1000

­0.0029

­3.416058 ­2.383942

1Value available only when used with the SCXI-1327 high-voltage terminal block

­4.895058

­0.904942

© National Instruments Corporation

19

SCXI-1120/D Calibration Procedure



References