SKC GS-1 Single-inlet Cyclone

Cat. No. 225-105

Introduction

The GS-1 Cyclone is a 10-mm lightweight conductive plastic sampler with a single inlet. It is designed for use with either a Diesel Particulate Matter (DPM) Cassette or a standard three-piece 37-mm cassette and filter (25 mm with adapter accessory) for collecting respirable dust particles. The GS-1 Cyclone conforms to the ISO 7708/CEN criteria of a 4-µm 50% cut-point at 2 L/min with minimum bias. The cyclone can be operated at 1.7 or 2 L/min to meet MSHA DPM sampling requirements, or at 3 L/min for a 3.5-µm cut-point to meet the MSHA for silica standard. The GS-1 Cyclone is supplied with a 37-mm cassette adapter, bowl adapter, and grit pot.

Figure 1. GS-1 Cyclone with standard 37-mm cassette installed - exploded view

Components:

Required Equipment

Getting Started

Assembling Cyclone with Standard 37 or 25-mm 3-piece Cassette (Figure 1)

The GS-1 Cyclone is supplied fully assembled with bowl adapter and 37-mm cassette adapter. A 25-mm cassette adapter is sold separately as Cat. No. 225-101.

  1. Determine cassette diameter to be used with cyclone.
    • If using a 37-mm cassette, proceed to Step 2.
    • If using a 25-mm cassette, remove the 37-mm cassette adapter from the cyclone (pull it off the top) and replace it with the 25-mm cassette adapter accessory (align with cyclone top and press on until firmly seated).
  2. Disassemble a three-piece cassette and set aside the inlet section (usually marked "inlet"). Keep the inlet section for closing the cassette after sampling.
  3. Select a filter and support pad as specified in the sampling method. Place the support pad into the cassette outlet section and place the filter on top of the support pad. Insert the cassette ring (middle) section into the cassette outlet section, ensuring a firm seal.
  4. Hold the cyclone upside down (cassette adapter facing downward). Insert the cassette adapter into the cassette middle ring section. Press until a firm seal is established.
  5. Ensure the thumbscrew on the cyclone is secure and that the grit pot remains on the cyclone body during calibration and sampling.

Inserting Cyclone/Cassette Assembly in Cassette Holder (Figure 2)

Figure 2. GS-1 Cyclone/cassette assembly in holder (Cat. No. 225-1)

  1. Insert the cyclone body through the large opening of the cassette holder.
  2. Ensure the cyclone/cassette assembly is seated firmly in the holder by inserting the small round nodule on the rim of the bowl adapter into the notch in the cassette holder.
  3. Remove the plug from the cassette outlet.
  4. Secure the cyclone/cassette assembly in the holder by stretching the spring-loaded hold-down plate over the cassette outlet.
  5. Insert the holder's Luer adapter (located on the end of the spring-supported rubber tubing) into the cassette outlet.

See Form 40081 for using the GS-1 Cyclone with a DPM Cassette.

Operation

Calibration

Flow rate:

  1. Prepare a cyclone/cassette assembly (Figures 1 and 2). Ensure grit pot remains on the cyclone body during calibration.
  2. SKC recommends using the smallest calibration jar possible. To achieve this, do NOT use Cassette Holder Cat. No. 225-1 during calibration.
  3. Place cyclone/cassette assembly into an airtight calibration jar that contains an inlet and outlet (Cat. No. 225-111; Figure 3).
  4. Using flexible tubing, connect the outlet of the calibrator to the inlet of the calibration jar.
  5. Run tubing from the outlet of the cyclone/cassette assembly through the outlet of the calibration jar and to the inlet of a sample pump. If using the Calibration Jar (Cat. No. 225-111), connect the Luer adapter inside the jar lid to the cassette outlet. Connect the jar outlet tubing to the sample pump inlet.
  6. Turn on the pump and calibrate to the desired flow rate following directions in the pump and calibrator operating instructions.
  7. After calibration, disassemble the calibration jar, remove the cyclone/cassette assembly, and replace the cassette used for calibration with a fresh (unused) cassette to be used for sampling.

Figure 3. GS-1 Cyclone/cassette assembly in calibration jar

Calibration Setup:

Sampling

Check the sampling method being used for flow rate and sampling media requirements.

  1. Calibrate pump to desired flow rate using representative sampling media. See Calibration.
  2. Leave grit pot in place during calibration and sampling.
  3. Ensure a fresh cassette has been installed on the cyclone and the cyclone/cassette assembly has been inserted into the cassette holder. See Inserting Cyclone/Cassette Assembly in Cassette Holder.
  4. Using the Tygon tubing attached to the cassette holder, connect the holder containing the cyclone/cassette assembly to the inlet of the calibrated sample pump.
  5. Clip holder near the breathing zone and clip pump to the worker's belt.
  6. Start pump and record pertinent details.
  7. Sample for the time specified in the method.
  8. Turn off pump when sampling is complete and record pertinent details.
  9. Remove cyclone/cassette assembly from holder and immediately cap the cassette outlet.
  10. Separate cassette from cyclone and immediately replace the inlet section of the cassette. Cap the inlet with the provided plug.
  11. Package cassette and send it with all data to a laboratory for analysis.

Important: Discard particles that were collected in the cyclone's grit pot. ⚠️

Maintenance

Cleaning

After sampling, clean all parts of the cyclone, including the interior of the grit pot, with mild soapy water. The cyclone can be wiped with a clean dust-free tissue, air dried, blown dry, or wiped with isopropyl alcohol.

Caution: Do not use strong solvents to clean the cyclone. ❌

References

Accessories/Replacement Parts

Description Cat. No.
Filter Cassette/Cyclone Holder, for DPM cassette or standard 25 or 37-mm 3-piece filter cassettes with or without a cyclone, required when using DPM Cassette with GS-1 Cyclone 225-1
Replacement Cassette Adapters
25-mm cassettes 225-101
37-mm cassettes (supplied with cyclone) 225-102
Replacement Bowl Adapter 225-108
Replacement Grit Pots, pk/25 P225012
Standard-size Multi-purpose Calibration Jar 225-111

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC's sole liability and the buyer's exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to skcinc.com/warranty.

Models: 225-105, GS-1, GS-1 Single Inlet Conductive Plastic Cyclone, Single Inlet Conductive Plastic Cyclone, Conductive Plastic Cyclone, Plastic Cyclone

File Info : application/pdf, 8 Pages, 989.74KB

37098

References

Acrobat Distiller 21.0 (Windows)

Related Documents

Preview SKC Respirable Dust Cyclone Performance Guide - Air Sampling Equipment
Detailed performance guide for SKC Respirable Dust Cyclones, including optimal flow rates, collection efficiency data relative to ISO/CEN/OSHA/ACGIH standards, and key advantages. Features SKC Aluminum, GS-1, and GS-3 cyclone models.
Preview SKC PPI Samplers: Operating Instructions for Disposable Parallel Particle Impactors
Comprehensive operating instructions for SKC Disposable Parallel Particle Impactors (PPI) used in occupational hygiene for accurate respirable and thoracic dust sampling. Covers setup, calibration, and sampling procedures.
Preview SKC Multi-purpose Calibration Jar Operating Instructions (225-111, 225-112)
Operating instructions for the SKC Multi-purpose Calibration Jar (Cat. Nos. 225-111 and 225-112), detailing setup, flow rate verification, and required equipment for air sampling applications.
Preview SKC Preloaded Disposable Parallel Particle Impactors (PPI) Operating Instructions
This document provides operating instructions for SKC Preloaded Disposable Parallel Particle Impactors (PPI) samplers, designed for precise size-selection of respirable and thoracic dust in occupational hygiene applications. Learn about their features, performance, preparation, and sampling procedures.
Preview SKC Cyclone Samplers for Respirable Dust: Operating Instructions
Comprehensive operating instructions for SKC Cyclone Samplers (Models 225-69 and 225-69-37) designed for sampling respirable dust aerosols. Covers preparation, cassette loading, sampler assembly, flow calibration, sampling procedures, post-sampling steps, filter weighing, transportation, and includes a certificate of conformity.
Preview SKC Face Level Sampling Headset for Welding Fume and Hazardous Metal Monitoring
The SKC Face Level Sampling Headset enables personal air sampling inside masks and helmets, ideal for monitoring hazardous metals like manganese, nickel, and chrome in welding fumes. It features dual flexible arms for simultaneous sampling and meets ISO 10882-1 standards.
Preview SKC Deployable Particulate Sampler (DPS) System Operating Instructions
Comprehensive operating instructions for the SKC Deployable Particulate Sampler (DPS) System, detailing setup, calibration, sampling, maintenance, and ordering information for particulate matter monitoring.
Preview SKC Preloaded Coated Filters Cat. No. 225-9014 Operating Instructions
Operating instructions for SKC Preloaded Coated Filters, Cat. No. 225-9014. Details on usage, storage, sampling, and warranty for OSHA ID-214 Ozone filters.