User Manual for sellEton models including: SL-905 Digital Indicator With Numeric Keypad, SL-905, Digital Indicator With Numeric Keypad, Indicator With Numeric Keypad, Numeric Keypad, Keypad

SL-905 Indicator

Manuals & How To Guides — SellEton Scales


File Info : application/pdf, 31 Pages, 2.54MB

PDF preview unavailable. Download the PDF instead.

SL-905 Indicator-Manual fc5f8ba4-69dd-4c87-beda-23923b244ac0 ?v=1701195330
SL-905 INDICATOR USER'S MANUAL
www.selletonscales.com

TABLE OF CONTENTS
Safety Precautions Preparations and Set Up Features Specifications Power Supply Displays Display and Key Descriptions Operating Instructions Calibration Indicator Parameter Settings Helpful Definitions Connectors RS232 Serial Output Format 4-20 mA Analog Output Relay Output Troubleshooting Q&A Contact Us

1 1 2 3 4 5 6 7-9 10 12-16 17 18-19 20-23 24 25 26 27 28

SAFETY PRECAUTIONS
For safe operation of the weighing indicator, please follow these instructions:  Calibration inspection and maintenance of the indicator are prohibited by non-pro-
fessional staff  Please ensure that the indicator rests on a stable surface  The indicator is a piece of static sensitive equipment; Please cut off power during
electrical connections  Touching the internal components by hand is prohibited  DO NOT exceed the rated load limit of the unit  DO NOT step on the unit  DO NOT jump on the scale  DO NOT use this product if any of the components are cracked  DO NOT use for purposes other then weight taking  To avoid damaging the battery do not keep charger plugged in once battery is fully
charged  Make sure the weight is not over the Max capacity as it could damage the load cell
inside  Material that has a static electric charge could influence the weighing. Discharge
the static electricity of the samples, if possible. Another solution to the problem is to wipe both sides of the pan and the top of the case with an anti-static agent
Please take anti-static prevention measures Any accumulated charge on the body of the human operator should be discharged first before opening the protective container with ESDS devices inside. The discharge can be accomplished by:  Putting a hand on a grounded surface or, ideally, by wearing a grounded Anti-static
Wrist Strap and an Anti-static Mat
PREPARATION & SET UP
 Plug into a wall outlet to avoid interference with other wirings  Turn on the indicator while there is no load  Calibration may be required before weighing when the scale is initially installed or
moved from a location
1

FEATURES
Suitable for Bench scales, floor scales and truck scales, this indicator works with 32bit CPU and 24bit high precision ADC. It is made with a Stainless Steel housing. Excellent performance make it proper for varied interface function. Waterproof connector and cable connection fixed inside.
 Multiple weighing units: (lb/kg)  Gross/Tare/Pre-Set Tare/Zero  Multiple Hold functions  Count weighing  Accumulation weighing  Check weighing  Overload / Underload indication  Connects to printer  Splash proof keyboard and display  Connects to a Remote Display/Scoreboard  Power saving mode  Can connect to a PC or printer for data logging (optional)  Wireless capability (optional)  20 mm Letter LED display  Rechargeable battery  RS232 output
Technical Parameters
 Accuracy class: 5000 e  Resolution - Display: 30,000 ; ADC: 2,000,000  Zero stability error: TK0 < 0.1V//K  Span stability error: TKspn < ± 6 ppm//K  Sensitivity (internal): 0.3 V / d  Input voltage: -30 to +30mV DC  Excitation circuit: 5 VDC, 4 wire connection, 12 load cell of 350ohm max  AC power: AC 110V  Battery 6V4Ah  Operation temperature: -10 °C ~ +40 °C  Operation humidity: 85%RH  Storage temperature: -40 °C ~ +70 °C (32-104°F)  Weight: 7lbs (3.2kg)  Dimensions: 11.25" x 4" x 9.5" (260x160x80mm)
2

1.77" 6.25"

SPECIFICATIONS
FIGURE 1: INDICATOR MEASUREMENTS

10.25"

zero

Hi

OK

Lo

Gross Net Hold Total pcs

lb

kg Battery

Hold Total Tare

Unit Del Zero

Check Print

Count

Esc

Set

ON/OFF

SL-905
1 23 4 56 789
0#

3.25"

9.5"

11.25"

4"

3.14" 5.11"

3

POWER SUPPLY
AC Adapter
We recommend to plug into a wall outlet to avoid interference with other wirings.
Battery
Please charge the internal battery fully before first time use. To keep the battery in best condition, fully discharge the battery every month by leaving the indicator on until the indicator powers off, and then recharge fully. If the battery is not going to be used for a long period of time it is recommended to remove it to avoid leakage.
 When the Battery is low the battery indicator light flashes red  During charging the red light will stay lit  The light will turn green once fully charged

SL-905 (LED)
Keypad and Display

MODEL SL-905
OVER

ACCEPT

UNDER

0

Gross Net Hold Total Pcs LB KG Battery

HOLD UNIT CHECK PRINT

CLASS III
123 456 789 C0 #

TOTAL DEL COUNT ESC

TARE ZERO SET ON/OFF

Cap. _______x D_____

4

DISPLAY AND KEY DESCRIPTION

HOLD Holds the weight on the scale Depending on the parameter setting

UNIT Shifts between weighing units

CHECK Checking the record by number

PRINT TOTAL
DEL

Print the weighing data
In gross mode: get the tare weight In net mode: clear the tare, get the Gross Delete the previous weighing data and the saved record

COUNT Use the scale to count product based on a sample weight

ESC TARE ZERO SET ON/OFF

Back to weighing
1. Zero's the scale. Used when using a container to hold objects 2. Clears the tare to see the gross weight Zero's the scale
Confirm the action
Powers the Indicator On or Off if held for 2 seconds

C

Clear the input

#

Confirm the input

Gross Net Hold Total Pcs LB KG Battery Over Accept Under

The scale is at zero
The scale is stable
Shows you are in Gross weight mode (includes tare); default mode Shows you are in Net weight mode (weight without tared weight) Shows you are in Hold mode Shows you are in Accumulation mode Shows you are in Counting mode The weight is shown in pounds The weight is shown in kilograms Flashes red = low battery, Solid red = charging, Green = fully charged Flashes when weight is higher than set alarm parameter Flashes when weight is within the set alarm parameters Flashes when weight is lower than set alarm parameter

5

OPERATING INSTRUCTIONS

Power On
 Turn on the power by pressing the ON/OFF button for 2 seconds. Once on, the scale will flash the voltage and then begin to auto-check and count down from 0-9 sequentially before entering the weighing mode Note: Anything on the scale before powering on will automatically be tared out.

Zeroing
 The zero function is used only when the scale is empty and is not at zero due to
material build up  Pressing the ZERO key while the scale is stable will reset your scale to 0  Depending on what your manual zero range parameter is set to, you can zero out
any weight within your set selection, after that you will receive an error and will
need to tare out the weight

Unit Selection
 To switch between measuring units (kg/lb) press the UNITS key

Tare Function
To Tare: When in gross mode, press TARE to enter net mode and tare out the weight

TARE

 The Tare function is used when you only wish to see the current change in weight,

not the entire amount of weight that is on the scale  When the indicator is in gross mode (gross light is shown) pressing the TARE key

will Tare the current weight on the scale and enter the net mode (net light shown)

 For example if you are using a container add the container to the scale, press tare

and the display will show the tare symbol

and reset back to 0

 Add your product to the scale to weigh without the weight of the container

 To exit Tare mode press the TARE key again to enter gross mode and you will see

the total weight of the container and the product Note: If you remove the container the scale will show the minus weight of the container

To use a pre-set tare weight
 Press and hold the TARE key and SET key at the same time  Input the weight you want tared on the keypad  Press TARE key to confirm  Press TARE to reset back to 0

6

Accumulation
 The accumulated function is used to add multiple weights and total them together  You can add up to 999 different weights together  In weighing mode load the first weight, once stable press the TOTAL key to enter
the accumulated mode. The "total" light will display

TOTAL
 Remove the first weight and add the second weight to the scale  Once stable press TOTAL key to add the weight to the accumulated total

TOTAL

 Repeat previous steps until all desired weights have been added to the total (max 999
times)  When you are done and want to display the accumulated total, press the TOTAL and
SET key together. The accumulated number "n###" (the number of weights you
are adding together) will flash on the display followed by the total

TOTAL + SET
 If you want to print the accumulated total, hold the PRINT key for one second  To exit accumulated mode, press the ESC key  To delete the saved weights press the DEL key followed by the set key to confirm

DEL

SET

 Repete this process to cleaer all saved weights

Print
 If the indicator is connected to a printer and the weight on the scale is stable press the PRINT key to print the current weight
Note: In tare mode the printer can not print if negative weight is shown

7

Counting Function
 The counting function is used to count a high volume of identical parts. You can do
this by setting a sample and then either adding to the sample or taking away from
the sample to count the number of objects on the scale  In weighing mode: press COUNT to enter count mode COUNT  Then press COUNT and SET at the same time to set your sample with the keypad

COUNT + SET

 (ex. If you are counting pencils, you would add a sample of pencils to the scale, if

you add 50 pencils to the scale you would enter 50 on the keypad.)

#  Press

on the keypad to set your sample

 The scale is now ready to start counting, load your product on the scale and the

indicator will show the quantity

+ 450 pencils

 (ex. if you add 450 pencils to your sample of 50, the scale will read 500 pencils)  To see the weight on the scale press COUNT  To exit counting mode press the ESC key  If you want to count a different product hold the COUNT and SET keys together to
enter a new sample

Hold
There are 4 different hold functions you can choose from in the C11 parameter
1. Peak Hold: Grabs the highest weight (for materials testing, ie. tension and pulling force)  Press the HOLD key then add weight to the scale  The indicator will show the highest weight it recorded and hold it on the screen until
a higher weight is placed on the scale 2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate  While weighing, press HOLD and the indicator will hold the current weight on the
screen until HOLD is pressed again 3. Auto Hold: If the weight on the scale is above 20d (20 x division) and is stable, the indicator will hold that weight on the screen for 3 seconds then go back to general weighing  Pressing the hold key is unnecessary, holding is done automatically when the scale
is stable 4. Average Hold: Used for animal weighing, the indicator will display the average weight sampled from 3 seconds

8

Recipes
Recipes allows you to store formulas for different products you may continuously be weighing, and creates a quick easy way to streamline your weighing.
For example: If you are selling 2 lb bags of cherries, you would create a recipe by setting a tolerance using upper and lower limits. This tolerance allows you to quickly separate multiple 2 lb bags of cherries. For this example we are going to use a tolerance of 0.1 lbs. Meaning we would set our upper limit to 2.10 lbs and our lower limit as 1.90 lbs. We are telling the scale, that any weight taken out of the container between 1.9 lbs - 2.1 lbs is acceptable, anything higher or lower is not. Now that our recipe is set add your large container of cherries on the scale. Then tare out your container of cherries, setting the scale to 0. Now if you start scooping out cherries from the container into a bag, the scale will read in negative (ex. -1.5 lbs) until the amount set in our recipe is reached. When you have taken out ~2 lbs (what your tolerance is set to) the scale will reset back to 0. Then the process can be repeated again, take out ~2 lbs of cherries until the scale reads 0.
To set these formulas: 1. Press SET, the screen will show what recipe number you are on, the default is 1. If it
is not on 1, use the keypad to enter the recipe number you would like to set. (up to 99 recipes can be saved) 2. Press # to confirm that number 3. You will now be back in weighing mode the screen will show 0 4. Press SET and ESC at the same time to enter the settings mode 5. use the keypad to input 13 followed by # to confirm 6. Use the keypad to enter your Upper Limit, followed by # to confirm (ex. 2.10) 7. The screen will show C14, press # to enter this setting 8. Use the keypad to enter your Lower Limit, followed by the # key to confirm (ex. 1.90) 9. The screen will show C15, press ESC to save and return to weighing mode 10. Then place your container on the scale, The screen will show you the total weight 11. Press TARE to set the scale back to 0 12. After pressing tare you are able to remove product from the scale until you are within your set tolerance and the scale reads 0. (ex. 1.9 - 2.1 lbs)
To use already set formulas: 1. In weighing mode, press SET, followed by the number the recipe is saved under 2. Press # to confirm 3. Add your product to the scale, Then press TARE 4. You can now start removing product and the scale will read 0 when its within the set
tolerance 5. To change recipes, press SET followed by the number the recipe is saved under
9

CALIBRATION PROCEDURE
1. Turn on the scale by holding ON/OFF . 2. Press SET and ESC together to access the setup menu. 3. If done correctly, the display should now show
4. Press # to access the C1 channel. The display should show [C01 01].
5. On they keypad enter which unit you want to calibrate in (1 = kg, 2 = lb).
6. Press # to save your value. The display will now show
7. Press # to access the C2 channel. The display should show [C02 00].
8. Use the keypad enter the setting for the decimal places desired (The C2 channel is used to adjust the decimal point on the scale. A value of 1 means there is one digit behind the decimal point 1= 0.0, a value of 2 = 0.00)
9. On they keypad set your value. The display will now show
10. Press # to access the C3 channel. The display should show [C03 01].
11. On they keypad enter the value of your desired graduation. (The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set to 1, the scale will read in 0.1 lb. increments.)
12. Press # to set your value. The display will now show
13. Press # to access the C4 channel. The display will show [000000].
14. On they keypad enter the maximum capacity you want to use for your scale. (The C4 channel is used to enter in the max capacity of the scale; Make sure the
number you enter does not exceed the max capacity of your scale.)
15. Press # to set your value. The display will now show
16. Press # to access the C5 channel. The display should show [C05 00].
17. The C5 channel calibrates zero on the scale. Make sure the scale is empty. 18. Press 1 on the keypad to change the value to 1.
19. Press # to zero calibrate. The display will count down from 10-1 while the scale is
calibrating. When the display shows 0 the zero calibration is complete.
20. Press # to to continue. The display will now show
21. Press # to access the C06 channel. The display will show [C06 00].
22. The C6 channel is used to calibrate the scale with a known weight. Press 1 on the
keypad to set the value of C6 to [C06 01]. 23. Press # to to continue. The display will flash SPAN, and then show [000000].
24. On the keypad enter the weight of the calibration weight you will use (must be at least 10% of your max capacity you set in C04.
25. Place your calibration weight on the empty scale and # .
10

CALIBRATION CONT.
26. The scale will count down from 10 to 0. Once 0 has been reached, the display will
show CALEnd.
27. Press # to continue. The display will now show
28. Press ESC to save and exit the setup menu. 29. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale. 30. If the scale does not show the value of the calibration weight, check that the feet on
the platform are not screwed in too tightly, and verify that the platform is level.
31. Unload the scale; the display should read 000000.
32. If the scale does not display 000000, check that the feet on the platform are not screwed in too tightly, and verify that the platform is level.
11

INDICATOR PARAMETER SETTINGS

The parameter settings menu has a calibration section (C01 to C07 explained above) and a parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08 and up can be accessed by the user. If you break the official seal by opening the back of the indicator to access the seal switch, you may need to have the indicator recertified. Be sure to adjust the seal switch back to the original setting after calibration/configuration has been performed.

To enter calibration/parameter settings, follow the procedure below:

1. Press and hold the SET and ESC key at the same time to enter the parameter settings

2. Navigate through the settings (C01 to C45) by using the number keypad

# 3. Press the

key to enter/edit the parameter setting

Press the ESC key to save and exit settings at any time

Table 1. Indicator Parameter Settings

Function

Parameter Settings/Options

Weighing Unit Decimal Setting

C01 C02

1 = kg 2 = lb 0 = no decimal 1 = #.# 2 = #.## 3 = #.### 4 = #.####

Graduation Setting
(readability of the least significant digit)

C03

Maximum Capacity

C04

options: 1/2/4/10/20/50 Example with no decimal places (ie C02=0) 1 = 1 lb 2 = 2 lb 5 = 5 lb 10 = 10 lb 20 = 20 lb 50 = 50 lb set max capacity ex. 100kg = 0100.00

Zero Calibration

C05

0 = zero calibration not needed 1 = set the zero calibration (Please ensure scale is empty and the

Calibration
Restore Default Settings

C06 C07

stable light is on) 0 = calibration not needed 1 = ready to calibrate with calibration weight 0 = do not restore 1 = restore to default settings

12

Function Hold Function
Unit Switch Power Saving Mode
Automatic Power Off
Alarm Setting Upper Limit Alarm Lower Limit Alarm Inner Code Display Preserved menu Preserved menu Communication Setting for the RS232 Port
Preserved menu Manual Zero Range
Initial Zero Range

Parameter Settings/Options

C08
C09 C10 C11
C12

0 = turn off hold function 1 = Peak hold - Grabs the highest weight 2 = Manual hold - Grabs the current weight 3 = Auto hold - Automatically holds data when stable 4 = Average hold - for animal weighing, averages the weight from a
sample of 3 seconds 5 = Auto Average hold - Average hold without the need to press the
hold key
0 = turn off unit switch 1 = turn on unit switch 0 = turn off power saving setting 3 = turn off display if no change within 3 minutes 5 = turn off display if no change within 5 minutes 0 = turn off auto power off 1 = power off automatically if no change within 10 minutes 2 = power off automatically if no change within 20 minutes 3 = power off automatically if no change within 30 minutes 4 = power off automatically if no change within 40 minutes 5 = power off automatically if no change within 50 minutes 6 = power off automatically if no change within 60 minutes 0 = turn off alarm 1 = turn on alarm

C13 Set upper limit within the max. capacity

C14 Set lower limit within the max. capacity

C15 check the inner code (raw data)

C16

C17 C18

Set the serial interface data output method: 0 = Turn off serial interface data output 1 = Command request mode, connect computer. 2 = Print mode, connect printer 3 = Continuous sending mode, connect computer or display

C19 C20
C21

0 = turn off manually zero setting 1 = ±1% max capacity 2 = ±2% max capacity 4 = ±4% max capacity 0 = no initial zero setting 1 = ±1% max capacity 2 = ±2% max capacity 5 = ±5% max capacity 10 = ±10% max capacity 20 = ±20% max capacity

13

Function

Parameter Settings/Options

Zero Tracking
Zero Tracking Time Overload Range

C22
C23 C24

0= turn off zero tracking 0.5 = ±0.5d 1.0 = ±1.0d 2.0 = ±2.0d 3.0 = ±3.0d 4.0 = ±4.0d 5.0 = ±5.0d

d = division

Note: the zero tracking range can not be bigger than manual zero

range

0 = turn off zero tracking time 1 = 1 second 2 = 2 seconds 3 = 3 seconds

00 = turn off overload range 01-99d = overload range setting

d = division

Negative Display Standstill Time Standstill Range

C25 C26 C27

Digital Filter (for filtering moving weight such as animals)

C28

Noise Filter

C29

C41 Communication Format

Baud Rate

C42

0 = -9d 10 = -10% max. capacity 20 = -20% max. capacity

0 = quick

1 = medium

2 = slow

1 = 1d 2 = 2d 5 = 5d 10 = 10d

d = division

0 = turn off dynamic filter 1 = 1 digital filter strength 2 = 2 digital filter strength 3 = 3 digital filter strength 4 = 4 digital filter strength 5 = 5 digital filter strength 6 = 6 digital filter strength Note: The higher the number, the higher the filter strength

0 = turn off noise filter 1 = 1 digital filter strength 2 = 2 digital filter strength 3 = 3 digital filter strength

0 = turn off continuous transmit 1 = format 1 2 = format 2 3 = format 3 4 = format 4

0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19200 6 = 38400 7 = 57600 8 = 115200

14

Function Odd-Even Check Interval
Checksum Simple Command Support Commanding Format Baud Rate
Odd-Even Check Response Format Slave ID Print Mode Baud Rate

Parameter Settings/Options

C43

0 = 8n 1 = 70 2 = 7e

C44

0 = no limit 1 = 100ms 2 = 200ms 3 = 500ms 4 = 1s 5 = stable transmit

C45

0 = no 1 = yes

C46

0 = no 1 = yes

C51

0 = turn off commanding mode 1 = simple commanding 2 = standard commanding

C52

0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19200 6 = 38400 7 = 57600 8 = 115200

C53

0 = 8n 1 = 70 2 = 7e

C54

0 = format 1 1 = format 5 2 = format 4

C55 0-99

C61

0 = turn off printing mode 1 = turn on printing mode

C62

0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19200 6 = 38400 7 = 57600 8 = 115200

15

Table 3. Default Parameter Settings

Function Weighing Unit Decimal Setting Graduation Setting Maximum Capacity Zero Calibration Calibration Restore Default Hold Function Unit Switch Power Saving Mode Automatic Power Off Alarm Setting Upper Limit Alarm Lower Limit Alarm Inner Code Display Communication Setting Manual Zero Range Initial Zero Range Zero Tracking Zero Tracking Time Overload Range Negative Display Standstill Time Standstill Range Digital Filter Noise Filter Communication Format Baud Rate Odd-Even Check Interval Checksum Simple Command Commanding Format Baud Rate Odd-Even Check Response Format Slave ID Print Mode Baud Rate

Parameter
C01 C02 C03 C04 C05 C06 C07 C08 C09 C10 C11 C12 C13 C14 C15 C18 C20 C21 C22 C23 C24 C25 C26 C27 C28 C29 C41 C42 C43 C44 C45 C46 C51 C52 C53 C54 C55 C61 C62

Default 1 0 1 10000 0 0 0 0 9 0 0 01 000000 000000 999999 1 2 10 0.5 1 9 10 1 2 0 2 2 1 0 1 0 0 1 4 0 0 0 1 4

16

BATTERY INSTALLATION
The indicator works with AC power, please note the voltage is proper and check if the AC power side is earthing. Voltage transformer and stable circuit is build in the indiciator please dont open it without professional instruction.
Stable Circuit connection J1 Inditial level for AC power input and voltage transfomer J2 Second level for voltage transformer F1 220V 0.25A fuse
PCB indicator power connection Postive battery connect with BAT+ , negative batter connect with BAT-
17

SCALE INSTALLATION
SL-905 maxi. Supports 12 x 350ohm load cells. (4 wires or 6 wires load cell) The connection between load cells and indicator is by terminal block. the connection diagram as below.
4 Wire load cell connection
EXC+, EXC-, SIG+, SIG- (load cell) to EXC+, EXC-, SIG+, SIG- (indicator) EXC+ and SEN+ to be short circuit EXC- and SEN+ to be short circuit
18

6 Wire load cell connection
EXC+,SEN+, SEN-, EXC-,SIG+,SIG- (load cell) to EXC+,SEN+, SEN-, EXC-,SIG+,SIG(indicator)
Communication Interference SL-905 owns 3pcs RS232 , 1 pc RS485 and 1 pc USB-Com. 1 RS232 is electrical isolation with load cell and can be set in C18 parameter, DEFAULT off. RS485 and USB-COM are electrical isolation too. The communication function is set in C41 ~ C79 parameter. RS 232 connection as below.
19

USB-COM as below

RS485 as below.

PIN
RX1 TX1 GND RX2 TX2 GND 8X3 TX3 GND A B USBV+ DD+ GND

Definition
RS232-1 recieve RS232-1 transmit Earthing RS232-1 recieve RS232-1 transmit Earthing RS232-1 recieve RS232-1 transmit Earthing RS485 output A RS485 output B USB power DataData+ USB earthing

Interface
RS232-1

Function
Continues Tranmit

RS232-2

Printing

RS232-3
Electrical Isolation

C18: Commanding mode
Printing mode Continues transmit mode

RS485 Electrical isolation

Commanding mode

USB to COM Electrical isolation

Commanding mode

20

OUTPUT FORMAT

Computer Continuous Sending Mode (Format 1)

,

,

CR LF

{ { {

S 1

S 2

S 3

Data

S 4

S1:

weight status, ST=standstill, US=not standstill, OL=overload

S2:

weight mode, GS=gross mode, NT=net mode

S3:

weight of positive and negative, "+" or "-"

Data: weight value, including decimal point

S4:

"kg" or "lb"

CR:

carriage return

LF: line feed

21

RS232 SERIAL OUTPUT FORMAT

Follow the pin out of Table 5 below to connect the indicator the RS-232 Serial device

Table 5. DB9 Pin Description

DB9 Pin 2 3
5

Definition TXT RXD GND

Function Transmit Data Receive Data Ground Interface

The serial output format depends on the settings for parameter C18. The serial output consists of a string of ASCII characters. Here is a list of the serial parameters
 8 data bits  1 stop bits  No parity  No handshaking Note: With the RS232 Output option we have data logging software available as seen in Figure 5.

FIGURE 5. DATA LOGGING SOFTWARE
Below are the formats of the serial output  C18=0 - Turn off serial interface data output  C18=1 - Continuous sending mode, connect 2nd big display  C18=2 - Print mode, connect printer  C18=3 - Command request mode, connect computer.  C18=4 - PC continuous sending mode, connect computer  C18=5 - PC/big display, continuous sending mode  C18=6 - Print to adhesive label printer
22

Compatible for Yaohua A9 (Format 3)
ASCII code to show the data. The transmission data is the current weight. Each one is by 12 groups of data.

X bit 1 2 3
:
: 8 9 10 11 12

02(XON) start + or - symbol

Note

High bit of weighing data

Weighing data:

Weighing data:
Low bit of weighing data Decimal bits from right to left(0~4) Odd or even checking for high 4 bits Odd or even checking for low 4 bits 03(XOFF) end

PC or Remote Display Continuous Sending Mode (Format 4)

<STX> <POL> XXXXX.XX <L/K> <G/N> <STAT> <CR> <LF>

Start Transmission

Weight Data

Polarity: <SP> = Positive "-" = Negative

Gross/Net:

Units: lb = punt

GR = Gross NT = Net

kg = kilogram

pcs = pieces*

Carriage

Status:

Return

<SP> = Valid

Line Feed

M = Motion

O = Over/under range

PC or Remote Display Continuous Sending Mode (Format 5)

<STX> <POL> XXXXX.XX <sp> <Lb/KG> <sp> <GR/NT> <CR> <LF>

Start Transmission

Weight Data

Polarity: <SP> = Positive "-" = Negative

Space

Space

Units: lb = punt kg = kilogram pcs = pieces*

Gross/Net: GR = Gross NT = Net

Carriage Return
Line Feed

23

Commanding Format

Standard comand Use ASCII code to response

Transmit:

Format

STX

ADDRESS

Content

02

A-Z / 20

No address, use space (20hex), 6 bits in total.

COMMAND A-F,T,Z

VERIFY 3H 3L

Responsing:

Format

STX

Content

02

Content

02

ADDRESS A-Z / 20 A-Z / 20

COMMAND B-D A,T,Z,N,X

REPLY (W)

VERIFY 3H 3L 3H 3L

W: symbol+weight+decimal position Back 6/14 bit Command list: A: shaking hand B: gross weight C: tare weight D: net weight T: tare Z: zero N: command can't be done X: Non commanding

Simple commanding:

ETX 03
ETX 03 03

Code

Name

T

Tare

Z

Zero

P

Printing

R

Read gross/net weight

Note: If set salve ID is not 0, transmit with ID.

Function Reduce tare or clear tare
Tare to be zero Print the current weight
Response to format 1

24

PRINTING
The following are the printing formats on an Selleton ticket printer:

Standard Printing mode

S/N:
Date: Time: Net Tare Gross

XXX
05/01/2017 11:30:52 25.6lb 10.3lb 35.9lb

Accumulation Printing mode

Date: Time: N001: N002:

05/01/2017 11:30:52 25.6lb 10.3lb

Total Count

35.9lb 002

Counting Printing mode

Date: Time:
Pieces
APW Net: Tare: Gross:

05/01/2017 11:30:52
xxxxxx pcs
xxxx.xx kg 25.6lb 10.3lb 35.9lb

Checking Printing mode

Date: Time: Number Weight

05/01/2017 11:30:52 231 10.3kg

25

HELPFUL DEFINITIONS
Division: The amount of increments a scale offers. How accurate the scale can be
Capacity: the maximum amount the scale can contain
Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered on that will automatically zero. example: If initial zero range is set to 10% of the max. capacity and your max. capacity is 100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it will automatically zero out the weight.
Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let you manually zero (anything above this percent will be tared)
Zero Tracking Range: A subset to the manual zero range; if the weight on the scale is not stable, the zero tracking range still allows you to zero within a set division of the scale
Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero'd
Overload Range: Weight allowance that is out of the set calibrated range. Adds a tolerance to the calibrated max. capacity without having to recalibrate. example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set the overload range to 60, you can add 1060lbs of weight to the scale without it displaying an error code
Negative Display: How far you can go in the negative direction before displaying an error code
Standstill Time: How fast the scale will stabilize
Standstill Range: How much the scale can fluctuate before being determined stable
Digital Filter: For filtering moving weight, such as animals, It changes how sensitive the scale is to variations in movement.
Noise Filter: A filter for how susceptible the scale is to general variations
Baud Rate: The rate at which information is transferred in a communication channel. example: In the serial port context, "9600 baud" means that the serial port is capable of transferring a maximum of 9600 bits per second.
26

TROUBLESHOOTING

Error Codes

Error
UUUUUU

Reason 1. Overload 2. Wrong connection with load cell 3. Load cell has quality problem

1. Calibration is no good
nnnnnnn 2. Wrong connection with load cell 3. Load cell has quality problem

ERR1 ERR2

During calibration, weight is not used or the weight is above the max. capacity
During calibration, the weight is below the minimum required weight

ERR3

During calibration, the input signal is negative

ERR4 ERR5 ERR6 ERR9 ERR10

During calibration signal is unstable EEPROM Error Exceed Zero Range The number to chick is too big No information

Solution 1. Reduce the weight 2. Check load cell connection 3. Inspect load cell; Check the input/output 4. See Q&A section 1. Make sure scale is level 2. Check load cell connection 3. Check load cell input and output resistance 4. See Q&A section Use correct weight within the defined range
The calibration weight minimum is 10% of the max. capacity set in C04. Recommended to use 60%-80% of max. capacity if possible 1. Check all wire connections 2. Check load cell 3. Recalibrate 4. PCB replacement needed if steps 1-3 fail After the platform is stable, start calibration
Change PCB
Remove the load and See Q&A section
Input again with proper number
The accumulation is empty

27

Q&A
Q: The scale does not turn on
A: Make sure the power cord is plugged in, and that there is power. One easy way to test this is by connecting another appliance to the same outlet and see if it's
operational Q: The reading goes negative when a load is applied A: Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or
junction box (if one is used) Q: How do I resolve ERR6 error?
A: Please follow the procedure below: 1) Turn on the indicator and make sure nothing is on the scale, and that the scale
is level and not wobbling 2) Press and hold the "PRINT and HOLD" key simultaneously for a few of seconds 3) The screen will read "C01" 4) Using the arrow keys, change C01 to C20. You have to change the 1st digit from 0 to 2 first before you can change the 2nd digit 1 to a 0. 5) Press "PRINT" key to enter C20 parameter 6) Change the value of C20 on the right to 100 if possible using the up arrow key. If 100 is not available change to 20 7) Press "PRINT" key to enter your selection 8) The screen will read "C21" now 9) Press "PRINT" key to enter C21 parameter 10) Change the value on the right of C21 to 100 if available, 20 if not 11) Press "PRINT" key to enter your selection 12) he screen will read "C22" now 13) Press "TOTAL" key to save and exit 14) Power the indicator off and then on, and see if this resolves the ERR 6 issue. If not, then following the Q&A answers below for resolving "nnnnnn"and "uuuuuu"
errors Q: How do I resolve "nnnnnn" and "uuuuuu" error? A: 1) Check to see if the cable that runs from the indicator to the junction box is
damaged. If it is, replace the cable. 2) Open up the junction box (if available) and check to see if there is any water damage. If so, replace the junction box 3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block)
are not loose. Re-tighten the screws even if the wires seem to be connected 4) Recalibrate 5) If steps 1-4 do not work, there is a possibility one or more load cells are defective (consult with tech@selletonscales.com for further instructions)
28

CONTACT US
Please e-mail info@selletonscales.com for any sales related questions or call 844-735-5386
Don't forget to visit our website at: www.selletonscales.com



References

Adobe InDesign CC 2017 (Macintosh) Adobe PDF Library 15.0