Gentec-EO UD Series Thermal Sensor Disks

User Manual

Revision 1.1

Warranty

First Year Warranty

The Gentec-EO thermal disk carries a one-year warranty (from date of shipment) against material and/or workmanship defects when used under normal operating conditions. The warranty does not cover damages related to misuse.

Gentec-EO will repair or replace at its option any disk which proves to be defective during the warranty period; except in the case of product misuse.

Any unauthorized alteration or repair of the product is also not covered by the warranty.

The manufacturer is not liable for consequential damages of any kind.

In the case of a malfunction, contact your local Gentec-EO distributor or the nearest Gentec-EO office to obtain a return authorization number. Return the material to the appropriate address below.

Contacting Gentec Electro-Optics Inc.

To help us answer your calls more efficiently, please have the model number of the disk you are using ready before calling Customer Support.

Gentec Electro-Optics, Inc.
445, St-Jean-Baptiste, Suite 160
Québec, QC, G2E 5N7
Canada

Tel: (418) 651-8003

Fax: (418) 651-1174

E-mail: service@gentec-eo.com

Website: gentec-eo.com

1. Ultra Series UD Disk

1.1. Introduction

The Gentec-EO UD power detector series includes five opto-thermal sensors (UD10, UD12, UD19, UD25 and UD55). The high power surface absorber sensors are designed for use at high average power densities.

Unit Aperture (mm) Power range
UD10 10 0.1 mW to 2 W
UD12 12 1 mW to 70 W
UD19 19 0.2 mW to 200 W
UD25 25 10 mW to 250 W
UD55 55 5 mW to 700 W

For Gentec-EO's nearest office contact information, see p. ii, Contacting Gentec Electro-Optics Inc.

1.2. Warnings and Disclaimer

In no event shall Gentec-EO or any of its affiliates be liable for any indirect, special, incidental or consequential injury to persons or damage to property caused by the use of any of our products. By purchasing from Gentec-EO or any of its affiliates, you hereby indicate that you understand and agree to the following:

Disclaimer

I am fully responsible for the safe application and use of this product and agreed to such by completing the sales process.

I will not use a laser device without wearing approved laser safety goggles designed for such purpose.

I am aware and responsible of safely dealing with any back reflections.

I will not use the detector in violation of any local, state or federal law, and I understand that it is my responsibility to know and abide by those laws relating to the ownership and use of the product in my jurisdiction.

1.3. Specifications

The following specifications are based on an operating temperature of 15 to 28°C (59 to 82°F) and a relative humidity not exceeding 80%. Disks must be stored in an environment between 10°C to 65°C and a relative humidity not exceeding 90%.

UD10-2-H5-L

Measurement Capability
Effective Aperture Diameter10 mm
Spectral Range0.19 – 20 μm
Typical Power Noise Level¹0.1 mW
Typical Rise Time (0-95%)¹²3.0 s
Typical Sensitivity²2 mV/W
Energy Mode
Typical Sensitivity¹2.4 mV/J
Maximum Measurable Energy³3 J
Typical Noise Equivalent Energy¹5 mJ
Damage Thresholds & Laser Limits
Max. Average Power2 W
Max Average Power (Fan Cooled)2 W
Maximum Average Power Density36 kW/cm² (1064 nm, 2 W, CW)
11 kW/cm² (10.6 μm, 2 W, CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 360 μs, 5 Hz5 J/cm²14 kW/cm²
1.064 µm, 7 ns, 10 Hz1.0 J/cm²143 MW/cm²
532 nm, 7 ns, 10 Hz0.6 J /cm²86 MW/cm²
266 nm, 7 ns, 10 Hz0.3 J /cm²43 MW/cm²
Physical Characteristics
Dimensions440 x 3D mm
Weight (Head Only)7 g
Recommended Load Impedance100 kΩ
Product Number202832

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

UD12-70-H5

Measurement Capability
Effective Aperture Diameter12 mm
Spectral Range0.19 – 20 μm
Typical Power Noise Level¹1 mW
Typical Rise Time (0-95%)¹²1.6 s
Typical Sensitivity²0.53 mV/W
Energy Mode
Typical Sensitivity¹0.84 mV/J
Maximum Measurable Energy³5 J
Typical Noise Equivalent Energy¹20 mJ
Damage Thresholds & Laser Limits
Max. Average Power70 W
Max Average Power (Fan Cooled)30 W
Maximum Average Power Density36 kW/cm² (1064 nm, 10 W, CW)
11 kW/cm² (10.6 μm, 10 W, CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 360 μs, 5 Hz5 J/cm²14 kW/cm²
1.064 µm, 7 ns, 10 Hz1.0 J/cm²143 MW/cm²
532 nm, 7 ns, 10 Hz0.6 J /cm²86 MW/cm²
266 nm, 7 ns, 10 Hz0.3 J /cm²43 MW/cm²
Physical Characteristics
Dimensions360 x 2D mm
Weight (Head Only)4 g
Recommended Load Impedance100 kΩ
Product Number200382

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

UD19-50-W5

Measurement Capability
Effective Aperture Diameter19 mm
Spectral Range190 nm – 10 μm
Typical Power Noise Level¹1 mW
Typical Rise Time (0-95%)¹²5 s
Typical Sensitivity²0.65 mV/W
Energy Mode
Typical Sensitivity¹0.33 mV/J
Maximum Measurable Energy³200 J
Typical Noise Equivalent Energy¹23 mJ
Damage Thresholds & Laser Limits
Max. Average Power50 W
Max Average Power (Fan Cooled)50 W
Maximum Average Power Density100 kW/cm² (1064 nm, 10 W, CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 150 μs, 5 Hz100 J/cm²667 kW/cm²
1.064 µm, 7 ns, 10 Hz1.1 J/cm²157 MW/cm²
532 nm, 7 ns, 10 Hz1.1 J /cm²157 MW/cm²
266 nm, 26 ns, 10 Hz0.7 J /cm²27 MW/cm²
Physical Characteristics
Dimensions440 x 3D mm
Weight (Head Only)7 g
Recommended Load Impedance100 kΩ
Product Number200761

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

UD19-150-H5 / UD19-200-H9

Measurement Capability
Effective Aperture Diameter19 mm19 mm
Spectral Range190 nm – 10 μm190 nm – 10 μm
Typical Power Noise Level¹1 mW3 mW
Typical Rise Time (0-95%)¹²2.8 s4.5 s
Typical Sensitivity²0.65 mV/W0.23 mV/W
Energy Mode
Typical Sensitivity¹0.65 mV/J0.23 mV/J
Maximum Measurable Energy³15 J25 J
Typical Noise Equivalent Energy¹20 mJ60 mJ
Damage Thresholds & Laser Limits
Max. Average Power150 W200 W
Max Average Power (Fan Cooled)50 W110 W
Max. Average Power Density36 kW/cm² (1.064 μm, 10W CW)
11 kW/cm² (10.6 μm, 10W CW)
45 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 μm, 10W CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 150 μs, 5 HzH5: 5 J/cm², H9: 9 J/cm²H5: 14 kW/cm², H9: 25 kW/cm²
1.064 µm, 7 ns, 10 Hz1.0 J/cm²143 MW/cm²
532 nm, 7 ns, 10 Hz0.6 J /cm²86 MW/cm²
266 nm, 26 ns, 10 Hz0.3 J /cm²43 MW/cm²
Physical Characteristics
Dimensions440 x 3D mm
Weight (Head Only)7 g
Recommended Load Impedance100 kΩ
Product Number200262200576

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

UD25-200-H9 / UD25-350-H12

Measurement Capability
Effective Aperture Diameter25 mm25 mm
Spectral Range190 nm - 20 μm190 nm - 20 μm
Typical Power Noise Level¹3 mW10 mW
Typical Rise Time (0-95%)¹²5 s7.9 s
Typical Sensitivity²0.23 mV/W0.1 mV/W
Energy Mode
Typical Sensitivity¹0.14 mV/J0.05 mV/J
Maximum Measurable Energy³40 J40 J
Typical Noise Equivalent Energy¹200 mJ200 mJ
Damage Thresholds & Laser Limits
Max. Average Power200 W350 W
Max Average Power (Fan Cooled)150 W250 W
Max. Average Power Density45 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 µm, 10W CW)
25 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 µm, 10W CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 150 μs, 5 Hz9 J/cm²25 kW/cm²
1.064 µm, 7 ns, 10 Hz1.0 J/cm²143 MW/cm²
532 nm, 7 ns, 10 Hz0.6 J/cm²86 MW/cm²
266 nm, 26 ns, 10 Hz0.3 J/cm²43 MW/cm²
Physical Characteristics
Dimensions540 x 3D mm
Weight (Head Only)13 g
Recommended Load Impedance>100 kΩ
Product Number200263202378

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

UD55-200-H9 / UD55-500-H12 / UD55-700-HD

Measurement Capability
Effective Aperture Diameter55 mm55 mm55 mm
Spectral Range190 nm - 20 μm190 nm - 20 μm190 nm - 20 μm
Typical Power Noise Level¹5 mW15 mW45 mW
Typical Rise Time (0-95%)¹²11 s18 s14 s
Typical Sensitivity²0.12 mV/W0.06 mV/W0.03 mV/W
Energy Mode
Typical Sensitivity¹0.028 mV/J0.015 mV/J0.008 mV/J
Maximum Measurable Energy³200 J200 J200 J
Typical Noise Equivalent Energy¹250 mJ250 mJ250 mJ
Damage Thresholds & Laser Limits
Max. Average Power200 W500 W700 W
Max Average Power (Fan Cooled)150 W300 W400 W
Max. Average Power Density45 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 µm, 10W CW)
45 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 µm, 10W CW)
45 kW/cm² (1.064 μm, 10W CW)
14 kW/cm² (10.6 µm, 10W CW)
Pulsed Laser Damage Thresholds Max. Energy Density Peak Power Density
1.064 μm, 150 μs, 5 Hz9 J/cm²25 kW/cm²
1.064 µm, 7 ns, 10 Hz1.0 J/cm²143 MW/cm²
532 nm, 7 ns, 10 Hz0.6 J/cm²86 MW/cm²
266 nm, 26 ns, 10 Hz0.3 J/cm²43 MW/cm²
Physical Characteristics
Dimensions850 x 4D mm
Weight (Head Only)39 g39 g180 g
Recommended Load Impedance>100 kΩ
Product Number200264201220203757

¹ These characteristics depend on the thermal management and electronics provided by the user. Packaging, cooling and electronics similar to our UD Series will provide similar performances. See UD Series specifications sheets for more details. Actual performance depends on the tradeoffs in a user's design. It may be possible to enhance some performance parameters at the expense of others.
² Without anticipation algorithm or circuitry.
³ For 360 µs pulses. Higher pulse energy possible when customized for long pulses (ms), less for short pulses (ns).

2. Mounting and Cooling Recommendations

Important: Never touch directly the absorbing surface of the thermal disk

2.1. Cooling

The type of cooling (back cover, heat sink, fan, water) and the size of the cooling plate depend on the power applied on the thermal disk. Contact Gentec-EO to discuss your specific application.

Warning: The thermal disk temperature should never exceed 150 °C

2.2. Cooling Plate

The thermal disk should be supported by the cooling plate only on the edge of the thermal disk. The cooling plate rim should be made out of a material with a high thermal conductivity. It should also be flat and smooth to allow a good thermal contact with the disk. There should be a gap of at least 6 mm between the back of the disk and any other surface.

Thermal Disk Configuration Diagram

Diagram showing a cross-section of a thermal disk mounted on a cooling plate. Key elements include: THERMAL DISK, 6 mm MINIMUM AIR GAP BEHIND DISK, FRONT ABSORBER FACE OF DISK, BACK THERMOCOUPLE SIDE OF DISK, 2X MECHANICAL CONTACT AREAS FOR COOLING (BACK COVER, HEATSINK, FAN, WATER COOLING MODULE), THERMOCOUPLE STAY OUT ZONE, 2X THERMAL PASTE INTERFACE, COOLING MODULES (CUSTOMER SUPPLIED). A note indicates: DO NOT MAKE MECHANICAL CONTACT IN THE THERMOCOUPLE STAY OUT ZONE!

2.3. Thermal Disk Mounting

Before mounting the thermal disk in place, a thin film of thermal paste (For example: Wakefield Engineering Inc. thermal paste part no. 120-2.) should be applied on the support rim. Put the disk in place and turn it gently so that a good thermal contact is made between the disk and the support rim. (Be careful not to damage the contact pins.) The thermal disk must be maintained in place from the front (do not apply thermal paste on the front of thermal disk).

2.4. Laser Beam

The laser beam should always be centered with respect to the center of the thermal disk. The beam diameter should always be smaller than the optical aperture of the disk.

Laser Beam Alignment Diagram

An isometric view of a thermal disk showing a laser beam centered on its aperture. The text indicates the laser beam should be centered and its diameter smaller than the aperture.

3. Safety Instructions

3.1. General

To ensure a long lifetime of accurate measurements, the thermal disk should be maintained within the following ambient conditions:

WARNING: Be careful not to exceed the maximum levels and densities stated in the specifications.

It is possible to store and operate your thermal disk beyond this range. For any specific requirements, please contact your local Gentec-EO representative.

For the most accurate measurements, center the beam on the sensor.

3.2. Damage to the Optical Absorber Material

Damage to the optical absorber material is usually caused by exceeding the manufacturer's specifications, such as:

Refer to the specifications pages for the UD disk specifications. Damage may also be caused if the absorber surface is contaminated. A slight discoloration of the coating does not affect the calibration.

In any case, the beam's incident area should not be less than 10% of the disk's aperture. Please contact Gentec-EO to make measurements with such smaller beams.

In the event of significant damage to the coating, some UD Series disk can be recoated. Contact your local Gentec-EO representative for information on repair. See p. ii Contacting Gentec Electro-Optics Inc.

Declaration of Conformity

WEEE Compliance

All Gentec-EO products comply with the European Directive 2012/19/EU – WEEE.

RoHS Compliance

All Gentec-EO products comply with the European Directive 2011/65/EU – Restriction of Hazardous substance (RoHS 2), except TPM-300.

Appendix A: WEEE Directive

Recycling and Separation Procedure for WEEE Directive 2002/96/EC

This section is used by the recycling center when the UD disk reaches the end of its life.

Contents:

Separation:

Company Information

Gentec Electro-Optics, Inc.

445 St-Jean-Baptiste, Suite 160
Québec, QC, G2E 5N7
Canada

Tel: (418) 651-8003

Fax: (418) 651-1174

E-mail: info@gentec-eo.com

Website: www.gentec-eo.com

Leader in Laser Beam Measurement Since 1972

Product Categories: Power & Energy Meters, Beam Profiling, THz Measurement

Locations:

Models: UD Series, Thermal Sensor Disk, UD Series Thermal Sensor Disk

File Info : application/pdf, 16 Pages, 867.13KB

PDF preview unavailable. Download the PDF instead.

203756-UD-User-Manual-Rev1.1

References

Microsoft Word 2013 Microsoft Word 2013

Related Documents

Preview Gentec-EO UD Series Thermal Sensor Disks User Manual
User manual for Gentec-EO's UD Series thermal sensor disks, providing detailed specifications, mounting and cooling recommendations, safety instructions, warranty information, and compliance declarations.
Preview Gentec-EO BA Series Optical Attenuators User Manual | Specifications & Operation
Comprehensive user manual for Gentec-EO BA Series Optical Attenuators (BA16 and BA32). Covers specifications, operating instructions, warranty, safety, and compatibility for laser beam measurement and profiling applications.
Preview Gentec-EO BLU Bluetooth Wireless Detectors User Manual
This user manual provides detailed information on the Gentec-EO BLU series of Bluetooth wireless detectors. It covers product features, specifications, installation, operation, software integration, and troubleshooting for precise laser power and energy measurements.
Preview Gentec-EO UP19K-30H-W5-D0 Laser Power Detector Specifications
Technical specifications for the Gentec-EO UP19K-30H-W5-D0 thermopile laser power detector, capable of measuring laser power up to 30 W. Features include modular design, high damage threshold, compact size, and energy mode measurement.
Preview Gentec-EO HP280A-30KW-HD-IMP-D0: High Power Laser Detector Specifications
Detailed specifications for the Gentec-EO HP280A-30KW-HD-IMP-D0, a high-power laser detector capable of measuring up to 30 kW, featuring a large aperture and stable readings for demanding laser applications.
Preview Gentec-EO HP Series High Power Laser Detectors User Manual
Comprehensive user manual for Gentec-EO HP Series high power laser detectors, covering specifications, operating instructions, safety, and warranty for accurate laser power measurement.
Preview Gentec-EO BLU Series User Manual: Bluetooth Wireless Detectors
User manual for Gentec-EO BLU Series Bluetooth Wireless Detectors, detailing specifications, quick start, user interface, serial communication, and maintenance for laser power and energy measurements.
Preview Gentec-EO BLU Series: Bluetooth Wireless Detectors User Manual
Comprehensive user manual for the Gentec-EO BLU series of Bluetooth wireless power meters and detectors, covering installation, operation, specifications, and troubleshooting. Learn about connecting via Bluetooth or USB, using the mobile app and PC software, and understanding device features.