VENLAB VM-200M Multimeter User Manual

READ AND UNDERSTAND THIS MANUAL BEFORE USING THE INSTRUMENT. Failure to understand and comply with the WARNING and operating instructions can result in serious or fatal injuries and/or property damage.

Contact us: support@venlabtools.com

Made in China

Safety Information

The Meter conforms to IEC1010 600V (CAT II) over voltage safety standard and pollution level 2.

A Warning identifies conditions and procedures that are dangerous to the user.

To prevent possible electrical shock, fire, or personal injury:

Features

Display HOLD

In the Display HOLD mode, the Meter freezes the display.

  1. Press the 'HOLD' button, the meter display will hold the measured reading.
  2. Press the 'HOLD' button again to release the data, and the meter will return to the normal measurement state.

WARNING: To avoid electric shock, when Display 'HOLD' is activated, be aware that the display will not change when you apply a different voltage.

Backlight

  1. Press 'LIGHT' button to toggle the backlight on and off.
  2. The backlight automatically turns off after 30 seconds.

NOTE: When the battery power is low, the backlight brightness will not be enough.

Maintenance

Cleaning

Measurement

Note:

DC Voltage Measurement

  1. Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
  2. Switch to the ' " ' area, select suitable ranges and put the test leads to the loaded circuit.

AC Voltage Measurement

  1. Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
  2. Switch to the ' " ' area, select suitable ranges and put the test leads to the loaded circuit.

Voltage Measurement Note:

DC Current Measurement

  1. Put the black lead into the 'COM' port and red lead into the 'VΩmA' port if the current is under 200mA; put the red lead into the '10A' port if the current is between 200mA ~ 10A.
  2. Switch to the ' "" ' area, select suitable ranges and put the test leads to the loaded circuit.

Note:

Resistance Measurement

  1. Put the black lead into the 'COM'port and the red lead into the 'VΩmA' port.
  2. Switch to the ' Ω'area, select suitable ranges and put the test leads to the loaded circuit.

Note:

hFE Measurement

  1. Switch to the ' hFE' range.
  2. Please confirm whether the transistor is 'PNP' or 'NPN'. Then put the ‘E. B. C' feet into the corresponding sockets.

Note: The Meter tests the approximate value of hFE, and the test conditions are base current 10uA and Vce about 3V.

Diode Measurement

  1. Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
  2. Switch to the ' " ' ranges and touch the diode anode with the red probe, the black probe contacts the diode cathode.

Note:

Continuity Measurement

  1. Put the black lead into the 'COM'port and the red lead into the 'VΩmA' port.
  2. Switch to the ' " ' ranges and put the test leads to the circuit to be tested.
  3. The Meter will buzz if the resistance between the two points is less than 50 Ω.

Note:

Specifications

Maxim Display 2000 counts
Low Battery Indication " displayed
Operation Temperature 0℃---40℃, <85%RH
Storage Temperature -10℃---50℃, <85%RH
Battery Type 1.5V AA Batteries
Fuse 200mA/250V PPTC, 10A/250V fuse
Accessories 1 x Manual, 2 x AA Batteries, 2 x Test Leads
Input polarity indication Automatically displays “-”
Data Hold "H" displayed

DC Voltage

Range Resolution Accuracy Overload Protection
200mV 0.1mV ± (0.5%+5) 600V
2V 0.001V
20V 0.01V
200V 0.1V ± (0.8%+5)
600V 1V 600V

Input Impedance: 10MΩ

Overload Protection: 200mV 250V; Other: DC/AC 600V

AC Voltage

Range Resolution Accuracy
200V 0.1V ± (1.2%+10)
600V 1V

Frequency Response: 40Hz~400Hz; T-RMS

DC Current

Range Resolution Accuracy
200uA 0.1uA ± (1%+2)
2mA 1uA ± (1.5%+2)
20mA 0.01mA
200mA 0.1mA ± (3%+2)
10A 0.01A ± (3%+2)

Overload Protection: mA: 200mA / 250V PPTC 10A: 10A/250V fuse

Resistance

Range Resolution Accuracy
200Ω 0.1 Ω ± (0.8%+3)
2ΚΩ 0.001ΚΩ ± (0.8%+2)
20ΚΩ 0.01ΚΩ
200ΚΩ 0.1ΚΩ ± (1%+3)
2ΜΩ 0.001ΜΩ ± (1%+3)

Overload Protection: 250V AC

hFE

Range Instruction Condition
hFE NPN or PNP, 0-1000 Base current 10uA, Vce about 3V

Diode and Audible Continuity

Range Instruction Condition
" Show forward voltage OCV 3V DC
"" <50Ω, buzz OCV 3V

Three Year Warranty

3年間保証

For further detail of warranty coverage and warranty repair information, send email to support@venlabtools.com

YH Consulting Limited
C/O YH Consulting Limited Office 147, Centurion House,
London Road, Staines-upon-Thames, Staines, Surrey,
London, TW18 4AX
REP +44 07514-677868
H2YHUK@gmail.com

E-CrossStu GmbH
Mainzer Landstr.69, 60329, Frankfurt am Main, Germany
e-crossstu@outlook.com
+49 69332967674

Models: VM-200M Digital Multimeter, VM-200M, Digital Multimeter, Multimeter

PDF preview unavailable. Download the PDF instead.

B1TLxeTAFCL

Related Documents

Preview VENLAB VM-600A Multimeter User Manual
User manual for the VENLAB VM-600A Digital Multimeter, covering safety information, product features, measurement procedures, specifications, and maintenance.
Preview VENLAB VD-200 Non-Contact Voltage Tester Operation Manual
Comprehensive operation manual for the VENLAB VD-200 non-contact voltage tester. Includes safety precautions, product overview, detailed operating instructions for AC voltage detection, wire identification, flashlight/laser functions, battery replacement, and technical specifications.