VENLAB VM-200M Multimeter User Manual
READ AND UNDERSTAND THIS MANUAL BEFORE USING THE INSTRUMENT. Failure to understand and comply with the WARNING and operating instructions can result in serious or fatal injuries and/or property damage.
Contact us: support@venlabtools.com
Made in China
Safety Information
The Meter conforms to IEC1010 600V (CAT II) over voltage safety standard and pollution level 2.
A Warning identifies conditions and procedures that are dangerous to the user.
To prevent possible electrical shock, fire, or personal injury:
- Do not alter the Product and use only as specified, or the protection supplied by the Product can be compromised.
- Do not use the Product around explosive gas, vapor, or in damp or wet environments.
- Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible.
- Do not apply more than the rated voltage, between the terminals or between each terminal and earth ground.
- Do not use test leads if they are damaged. Examine the test leads for damaged insulation, exposed metal, or if the wear indicator shows. Check test lead continuity.
- Use the correct terminals, function, and range for measurements.
- The range switch should be placed in the correct range position.
- The input signal is not allowed to exceed the specified limit value to prevent electric shock and damage to the instrument.
- When measuring televisions or switching power supplies, be aware that there may be pulses in the circuit that can damage the circuit.
- Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50°C.
Features
Display HOLD
In the Display HOLD mode, the Meter freezes the display.
- Press the 'HOLD' button, the meter display will hold the measured reading.
- Press the 'HOLD' button again to release the data, and the meter will return to the normal measurement state.
WARNING: To avoid electric shock, when Display 'HOLD' is activated, be aware that the display will not change when you apply a different voltage.
Backlight
- Press 'LIGHT' button to toggle the backlight on and off.
- The backlight automatically turns off after 30 seconds.
NOTE: When the battery power is low, the backlight brightness will not be enough.
Maintenance
Cleaning
- Turn off the power to the Meter and remove the test leads.
- Wipe the case with a damp cloth and mild detergent.
- Dirt or moisture in the terminals can affect readings.
- If there is any abnormality, stop using the meter immediately.
Measurement
Note:
- Check the battery first, if ' " ' is displayed, the battery needs to be changed.
- "" signal besides sockets warns the maximum input voltage and current.
- Switch to the right range before measurement.
- The red lead mentioned in this manual is positive in polarity, the black lead is negative.
DC Voltage Measurement
- Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
- Switch to the ' " ' area, select suitable ranges and put the test leads to the loaded circuit.
AC Voltage Measurement
- Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
- Switch to the ' " ' area, select suitable ranges and put the test leads to the loaded circuit.
Voltage Measurement Note:
- If the correct range is unknown, choose the highest range, then switch from the high range to the low range.
- 'OL' means overload, please switch to a higher range.
- The Maximum input voltage is 600V.
- Avoid electric shock when measuring high voltage.
DC Current Measurement
- Put the black lead into the 'COM' port and red lead into the 'VΩmA' port if the current is under 200mA; put the red lead into the '10A' port if the current is between 200mA ~ 10A.
- Switch to the ' "" ' area, select suitable ranges and put the test leads to the loaded circuit.
Note:
- If the correct range is unknown, choose the highest range, then switch from the high range to the low range.
- When the red lead is in the "10A" port, the measuring time should be less than 10 seconds to avoid the circuit heating affecting the accuracy.
- 'OL' means overload, please switch to a higher range.
- When the fuse is blown, replace it with the same specification.
- Do not measure AC current.
Resistance Measurement
- Put the black lead into the 'COM'port and the red lead into the 'VΩmA' port.
- Switch to the ' Ω'area, select suitable ranges and put the test leads to the loaded circuit.
Note:
- 'OL' means overload, please switch to a higher range.
- 'OL' will also be displayed if open-circuited.
- If the measured resistance is > 1MΩ, it takes several seconds to get a stable reading, which is a normal phenomenon for the measurement of high resistance.
- Before testing a resistor in the circuit, please make sure the circuit is powered off and the capacitor is completely discharged.
hFE Measurement
- Switch to the ' hFE' range.
- Please confirm whether the transistor is 'PNP' or 'NPN'. Then put the ‘E. B. C' feet into the corresponding sockets.
Note: The Meter tests the approximate value of hFE, and the test conditions are base current 10uA and Vce about 3V.
Diode Measurement
- Put the black lead into the 'COM' port and the red lead into the 'VΩmA' port.
- Switch to the ' " ' ranges and touch the diode anode with the red probe, the black probe contacts the diode cathode.
Note:
- 'OL' will be displayed if open-circuited.
- If the test leads are connected reversely with the diode polarity, 'OL' will also be displayed.
- "" ' is in the same range.
Continuity Measurement
- Put the black lead into the 'COM'port and the red lead into the 'VΩmA' port.
- Switch to the ' " ' ranges and put the test leads to the circuit to be tested.
- The Meter will buzz if the resistance between the two points is less than 50 Ω.
Note:
- 'OL' will be displayed if open-circuited.
- The circuit should be tested when the power is off because any loaded signal would make a buzzer.
- "" ' is in the same range.
Specifications
Maxim Display | 2000 counts |
Low Battery Indication | " displayed |
Operation Temperature | 0℃---40℃, <85%RH |
Storage Temperature | -10℃---50℃, <85%RH |
Battery Type | 1.5V AA Batteries |
Fuse | 200mA/250V PPTC, 10A/250V fuse |
Accessories | 1 x Manual, 2 x AA Batteries, 2 x Test Leads |
Input polarity indication | Automatically displays “-” |
Data Hold | "H" displayed |
DC Voltage
Range | Resolution | Accuracy | Overload Protection |
200mV | 0.1mV | ± (0.5%+5) | 600V |
2V | 0.001V | ||
20V | 0.01V | ||
200V | 0.1V | ± (0.8%+5) | |
600V | 1V | 600V |
Input Impedance: 10MΩ
Overload Protection: 200mV 250V; Other: DC/AC 600V
AC Voltage
Range | Resolution | Accuracy |
200V | 0.1V | ± (1.2%+10) |
600V | 1V |
Frequency Response: 40Hz~400Hz; T-RMS
DC Current
Range | Resolution | Accuracy |
200uA | 0.1uA | ± (1%+2) |
2mA | 1uA | ± (1.5%+2) |
20mA | 0.01mA | |
200mA | 0.1mA | ± (3%+2) |
10A | 0.01A | ± (3%+2) |
Overload Protection: mA: 200mA / 250V PPTC 10A: 10A/250V fuse
Resistance
Range | Resolution | Accuracy |
200Ω | 0.1 Ω | ± (0.8%+3) |
2ΚΩ | 0.001ΚΩ | ± (0.8%+2) |
20ΚΩ | 0.01ΚΩ | |
200ΚΩ | 0.1ΚΩ | ± (1%+3) |
2ΜΩ | 0.001ΜΩ | ± (1%+3) |
Overload Protection: 250V AC
hFE
Range | Instruction | Condition |
hFE | NPN or PNP, 0-1000 | Base current 10uA, Vce about 3V |
Diode and Audible Continuity
Range | Instruction | Condition |
" | Show forward voltage | OCV 3V DC |
"" | <50Ω, buzz | OCV 3V |
Three Year Warranty
3年間保証
For further detail of warranty coverage and warranty repair information, send email to support@venlabtools.com
YH Consulting Limited
C/O YH Consulting Limited Office 147, Centurion House,
London Road, Staines-upon-Thames, Staines, Surrey,
London, TW18 4AX
REP +44 07514-677868
H2YHUK@gmail.com
E-CrossStu GmbH
Mainzer Landstr.69, 60329, Frankfurt am Main, Germany
e-crossstu@outlook.com
+49 69332967674