Laserliner - logoKondensearje Spot XP
BrûkershânliedingLaserliner Condense Spot XPLaserliner Condense Spot XP 2

Laserliner Condense Spot XP - icon 2Read the operating instructions and the enclosed brochure „Guarantee and additional notices“ completely. Follow the instructions they contain. Safely keep these documents for future reference.

Funksje / Applikaasje

The CondenseSpot is an infrared temperature measuring instrument with an integrated hygrometer that permits non-contact temperature measurements on surfaces as well as the measurement of relative humidity and ambient temperature. The instrument measures how much electromagnetic energy is emitted in the infrared wavelength range and uses this information to calculate surface temperature. During this process, two additional integrated sensors detect relative humidity and ambient temperature. The dew point is also calculated.

Laserliner Condense Spot XP - figure 1Laserliner Condense Spot XP 3

1 low battery charge 6 min/max and average measurement values
2 stability progress bar for relative
humidity measurement
7 dew point temperature (D Temp) and ambient temperature  (A Temp) in °C or °F
3 laser beam switched on, temperature measurement (infrared) 8 condense water indicator
4 measurements in °C or °F 9 last measurement value will be displayed brie y (7s)
5 relative humidity in % 10 replace battery

Laserliner Condense Spot XP 4Laserliner Condense Spot XP 5

Laser: The laser is a targeting aid to sight the location for the infrared measurement. Only the surface‘s temperature is measured. Laserliner Condense Spot XP 6

7. Displaying Min/Max/Avg, Ambient temperature, Dew point temperatureLaserliner Condense Spot XP 7

8. Dew point temperature is the temperature at which current air conditions would produce condensation. The CondenseSpot calculates the dew point temperature from ambient temperature, relative humidity, and ambient atmospheric pressure. In combination with the infrared thermometer, thermal bridges are easily detected. If the temperature of the measured location drops below the dew point, condensation (water) will form on the surface.Laserliner Condense Spot XP 8

9. Relative fochtichheid
The relative humidity is indicated in relation to the maximum possible humidity (100 %) at which air forms water vapor. The ability of air to hold water is temperature-dependent. Thus humidity is the volume of water vapor in the air. The range for humidity is 0 … 100%. 100% = saturation point.
Under these conditions for temperature and atmospheric pressure, air cannot absorb any more water.Laserliner Condense Spot XP 9

When temperature fluctuations (± 1 °C) and/or changes to relative humidity (± 1%) are rapid, the sensor must adapt itself to the new conditions. An hourglass with a progress bar is displayed during this stabilization process.
During this process, displayed values are only approximations. Only after these symbols have been switched off is a stable value presented which represents maximum accuracy.

10. Setting the emission coef cientLaserliner Condense Spot XP 10

A built-in sensor head detects the material/surface-specific infrared rays emitted by every object. The level of these emissions is determined by the material‘s emission coefficient (0.01 to 0.99). This instrument is preset to an emission coefficient of 0.95, which is applicable to most organic materials as well as plastics,  ceramics, wood, rubber, and stone. For details of those materials with a different emission coefficient, please refer to the table.

Unbekende útstjitkoëffisjint:
Apply masking tape or black matt paint to the surface of the area whose temperature you wish to measure. Wait until the tape/paint has heated up.
The temperature of the surface can then be measured with an emission coefficient of 0.95.

11. Emission coef cient tables
Nonmetalen

asbest 0.93 Kalkstien 0.98
Asfalt 0.95 Marble
Black, dull finish Greyish, polished
0.94
0.93
Basalt 0.70
Stienread 0.93
Carborundum 0.90 muorrewurk 0.93
Ceramics 0.95 Fervje
Swart, mat
Heat-resistant White
0.96-0.98 0.92
0.85 -0.95
Sina
Brilliant white With glaze
0.7 - 0.75 0.92
Klaai 0.95 Papier
Alle kleuren
0.95-0.97
Coal Non-oxidised 0.8-0.9
Plastic
Translucent PE, P, PVC
0.95
0.94
Concrete, plaster, mortar 0.93
Katoen 0.77 Rubber
Hard
Sacht, griis
0.94-0.95 0.89
Ierdewurk, mat 0.93
Stof 0.95
Glês 0.85 -0.94 Kwartsglês 0.93
Graphite 0.7 -0.8 Snie 0.80
Grint 0.95 Ierde 0.9 -0.98
Gypsum 0.8-0.95 Tarre 0.79 -0.84
Heat sink Swart, anodisearre 0.98 Tar papier 0.91 -0.93
Transformer ferve 0.94
Minsklike hûd 0.98 Wallpaper, ljochtkleurich 0.88-0.90
Iis
Opklearje
Mei swiere froast
0.97
0.98
Wetter 0.93
Bosk
Unbehandele
Beuk, skaft
0.8 -0.95 0.94
Lime 0.3 -0.4
Lime maim brick 0.95

metals

Alloy A3003 Oxidised
Rûge
0.3
0.1 - 03
Izer, smeid Matt 0.9
Foarsprong
rough
Oxidearre
0.4
0.2 - 0.6
Aluminium Oxidised
Oppoetst
0.2 - 0.4
0.04 - 0.06
Molybdenum Oxidised 0.2 - 0.6
brass
Oppoetst
Oxidearre
0.3
0.5
Nickel Oxidised 0.2 - 0.6
Platina Swart 0.9
Koper
Oxidearre
Elect. terminal strips
0.4 - 0.8 0.6 Stiel
Cold rolled Ground plate
Gepolijst plaat
Alloy (8% nikkel,
18% chromium) Galvanised Oxidised
Heech oksidearre
Freshly rolled Rough, flat surface
Rusty, read
Sheet, nickel-plated Sheet, rolled
0.7 - 0.9 0.4 - 0.6 0.1
0.35
0.28
0.80
0.88
0.24
0.95- 0.98 0.69
0.11
0.56
Haynes Metal alloy 0.3 - 0.8
Inconel
Oxidearre
Sandblasted Electropolished
0.7 - 0.95 0.3 - 0.6 0.15
Izer
Oxidearre
Mei roest
With red rust
0.5 - 0.9 0.5 - 0.7 0.61 - 0.85
Izer, getten
Oxidearre
Non-oxidised
Gesmelte massa
0.6 - 0.95 0.2
02 - 03
Zinc Oxidised 0.1

technyske gegevens

Technical revisions reserved. 08.13

Measured quantities °C (°F) infrared temperature measurement, %rH relative humidity, °C (°F) ambient temperature measurement
Infrared measurement range -40 °C … 350 °C (-40 °F … 662 °F)
Ambient temperature measurement range -10 °C … 60 °C (14 °F … 140 °F)
Relative humidity measurement range 20 % … 90 % rH
Dew point indicator -20 ° C ... 60 ° C
Infraread krektens t 1 °C (-10 °C … 60 °C); t 1.5 °C (< 10 °C and > 60 °C) or t 1.5 % whichever value is greater
Ambient temperature accuracy t 2 °C
Relative humidity accuracy t 3%
Infrared resolution 0.1 ° C
Relative humidity resolution 1%
Dew point resolution 1 ° C
Emission koëffisjint Adjustable, 0.10 … 0.99
Operating temperatuer 0 ° C… 40 ° C
Storage temperatuer -20 ° C ... 70 ° C
Optyk 8:1 (distance: measured spot)
Laserwellenlange 650 nm
Laser type Class 2, < 1 mW
Streamtafier Type 9 V E block battery

General safety instructions.
Attention: Do not look directly into the laser beam! The laser may not get into the hands of children! Do not point this device unnecessarily toward persons.Laserliner Condense Spot XP - warning

EU rjochtlinen en beskikking
Laserliner Condense Spot XP - ce
Dit apparaat foldocht oan alle nedige noarmen foar it frije ferkear fan guod binnen de EU.
Dit produkt is in elektrysk apparaat en moat apart wurde sammele foar ôffieren neffens de Jeropeeske rjochtline oer ôffal fan elektryske en elektroanyske apparatuer.
Fierdere feiligens en oanfoljende meidielings op: www.laserliner.com/info
CondenseSpotLaserliner Condense Spot XP - CondenseSpot

Laserliner Condense Spot XP - iconBETSJINNING
Umarex GmbH & Co KG
- Laserliner -
Möhnestraße 149, 59755 Arnsberg, Dútslân
Tel.: +49 2932 638-300, Fax: +49 2932 638-333
[e-post beskerme]
Umarex GmbH & Co KG
Donnerfeld 2
59757 Arnsberg, Dútslân
Tel.: +49 2932 638-300, Fax: -333
www.laserliner.com
Laserliner - logoLaserliner Condense Spot XP - ce

Dokuminten / Resources

Laserliner Condense Spot XP [pdf] Hânlieding
Kondensearje Spot XP

Lit in reaksje efter

Jo e-mailadres wurdt net publisearre.