1. مقدمه
This manual provides essential instructions for the proper use, maintenance, and care of your Cynamed Self-Aspirating Dental Anesthetic Syringe. Designed for dental surgical procedures, this instrument facilitates the controlled delivery of anesthetic solutions. Each syringe is crafted from stainless steel, featuring a two-tip piston ring and a thumb handle for precise control. These instruments are German-grade, autoclave-able, sterilizable, and conform to CE & ISO 9001:2008 standards.
Please read this manual thoroughly before initial use to ensure safe and effective operation.

شکل 1: تمام شده استview of Cynamed Self-Aspirating Dental Anesthetic Syringes.
2. اجزای محصول
The Cynamed Self-Aspirating Dental Anesthetic Syringe consists of the following key components:
- بشکه: Holds the anesthetic cartridge (1.8ml capacity).
- Piston Rod with Harpoon: Engages the rubber stopper of the anesthetic cartridge for aspiration.
- Thumb Ring: Provides leverage for aspiration and injection.
- Finger Rests: Supports the fingers during operation.
- آداپتور سوزنی: Connects the dental needle to the syringe.

شکل 2: با جزئیات view of the syringe components.
3. راه اندازی و آماده سازی
- عقیم سازی: Before first use and after every subsequent use, thoroughly clean and sterilize the syringe. Refer to the "Maintenance and Sterilization" section for detailed instructions.
- بازرسی: Visually inspect the syringe for any signs of damage, corrosion, or wear. Ensure all moving parts operate smoothly. Do not use a damaged syringe.
- بارگیری کارتریج:
- Retract the piston rod fully by pulling back the thumb ring.
- Insert the anesthetic cartridge (1.8ml) into the barrel, ensuring the rubber stopper faces the piston rod.
- Gently advance the piston rod until the harpoon engages the rubber stopper of the cartridge.
- Needle Attachment:
- Carefully screw the dental needle onto the needle adapter of the syringe. Ensure a secure fit to prevent leakage.
- Remove the needle cap just prior to use.

Figure 3: Syringe prepared for cartridge insertion.
4. دستورالعمل های عملیاتی
The Cynamed Self-Aspirating Syringe is designed for single-handed aspiration, enhancing safety and efficiency during anesthetic delivery.
- موقعیت یابی: Hold the syringe firmly with your thumb in the thumb ring and fingers on the finger rests.
- تزریق: Slowly advance the piston rod to deliver the anesthetic solution.
- Aspiration (Self-Aspirating Mechanism): The self-aspirating feature relies on the elasticity of the rubber diaphragm in the anesthetic cartridge. To aspirate:
- After needle insertion, release the pressure on the thumb ring.
- The slight negative pressure created by the retraction of the rubber stopper will draw blood into the cartridge if the needle is in a blood vessel.
- Observe the cartridge for blood return. If blood is present, withdraw the needle slightly and re-aspirate.
- تکمیل: After the injection is complete, carefully withdraw the needle. Immediately recap the needle using a one-handed scoop technique or a needle recapping device to prevent needlestick injuries.
- دفع: Dispose of the used needle and anesthetic cartridge in appropriate sharps and biohazard containers according to local regulations.

Figure 4: Syringe with needle attached, demonstrating readiness for operation.
۷. نگهداری و استریلیزاسیون
Proper cleaning and sterilization are crucial for maintaining the functionality and longevity of your Cynamed dental syringe and for preventing cross-contamination.
- Immediate Post-Use Cleaning:
- Immediately after use, disassemble the syringe by removing the needle and cartridge.
- Rinse the syringe thoroughly under running water to remove any visible debris or anesthetic residue.
- تمیز کردن:
- Clean the syringe using an enzymatic cleaner or a neutral pH detergent, following the manufacturer's instructions for the cleaning agent.
- Use a soft brush to clean all surfaces, paying close attention to crevices and moving parts.
- Rinse thoroughly with distilled or deionized water to remove all cleaning solution residues.
- عقیم سازی:
- The Cynamed syringe is designed to be autoclave-able.
- Place the clean, dry syringe in a sterilization pouch or tray.
- Sterilize using a steam autoclave cycle (e.g., 134°C for 3-5 minutes or 121°C for 15-20 minutes), adhering to the autoclave manufacturer's guidelines and local infection control protocols.
- Ensure the syringe is completely dry before storage.
- روغن کاری (اختیاری): Periodically, a small amount of medical-grade lubricant can be applied to the piston rod to ensure smooth operation.
- ذخیره سازی: Store sterilized syringes in a clean, dry, and protected environment until ready for use.

Figure 5: Syringe disassembled for cleaning and maintenance.
6 عیب یابی
- Issue: Harpoon does not engage the rubber stopper of the anesthetic cartridge.
علت احتمالی: The harpoon may be bent, damaged, or incompatible with the specific cartridge's rubber stopper. Alternatively, the cartridge may not be fully seated.
راه حل:
- Ensure the anesthetic cartridge is correctly and fully inserted into the syringe barrel.
- Inspect the harpoon for any visible damage or deformation. If damaged, the syringe should be replaced.
- Verify that the anesthetic cartridge brand and type are compatible with the syringe.
- If the issue persists, discontinue use and contact Cynamed customer support.
- Issue: Leakage of anesthetic solution during injection.
علت احتمالی: Loose needle attachment, cracked cartridge, or damaged syringe barrel.
راه حل:
- Ensure the needle is securely screwed onto the needle adapter.
- Inspect the anesthetic cartridge for any cracks before loading.
- Examine the syringe barrel for any visible damage. If damaged, replace the syringe.
- Issue: Piston rod movement is stiff or jerky.
علت احتمالی: Accumulation of debris, lack of lubrication, or bent piston rod.
راه حل:
- Thoroughly clean and sterilize the syringe as per maintenance instructions.
- Apply a small amount of medical-grade lubricant to the piston rod.
- Inspect the piston rod for any bending. If bent, replace the syringe.
7. مشخصات
| نام محصول | Cynamed Self-Aspirating Dental Anesthetic Syringe |
| مدل | Self-Aspirating 1.8CC Dental Syringe |
| مواد | فولاد ضد زنگ |
| Item Volume Capacity | 1.8 Milliliters (for anesthetic cartridges) |
| Sterility Rating | Sterile (after proper sterilization procedures) |
| گواهینامه ها | CE & ISO 9001:2008 Certified Instruments |
| Autoclave-able | بله |
| سازنده | CYNAMED |
8. گارانتی و پشتیبانی
Cynamed instruments are manufactured to high-quality standards. For specific warranty information or technical support, please contact your authorized Cynamed distributor or visit the official Cynamed webسایت
سازنده: CYNAMED
Webسایت: Visit the Cynamed Store on Amazon (برای اطلاعات عمومی محصول)